introduction: - rochester institute of...
TRANSCRIPT
TRANSCAT Pipette Calibration Project 05417
1 INTRODUCTION:..........................................................................................2
2 RECOGNIZE AND QUANTIFY NEED...........................................................2
2.1 Project Mission Statement:..............................................................................2
2.2 Company Background.......................................................................................2
2.3 Calibration: An element of Metrology...........................................................2
2.4 Product Description...........................................................................................2
2.5 Scope Limitations...............................................................................................2
2.6 Stakeholders:.......................................................................................................2
2.7 Key Business Goals:..........................................................................................2
2.8 Top Level Critical Financial Parameters.......................................................2
2.9 Financial Analysis..............................................................................................2
2.10 Primary Market....................................................................................................2
2.11 Secondary Market...............................................................................................2
2.12 Order Qualifiers...................................................................................................2
2.13 Order Winners......................................................................................................2
2.14 Formal Statement of Work................................................................................2
3 CONCEPT DEVELOPMENT.........................................................................2
3.1 Concept Development Process......................................................................2
3.2 Brainstorming Session......................................................................................2
3.3 Conceptual Level Drawing...............................................................................2
3.4 Mechanical Concepts........................................................................................23.4.1 Plunger Depression..........................................................................................2
3.4.1.1 Stepper Motor..............................................................................................23.4.1.2 Pneumatic Systems......................................................................................2
3.4.2 Stand................................................................................................................23.4.2.1 Stand with vertical movement.....................................................................2
1
TRANSCAT Pipette Calibration Project 05417
3.4.2.2 Stand with radial movement........................................................................23.4.2.3 The Basic Stand...........................................................................................2
3.4.3 Vial Holder......................................................................................................23.4.3.1 The Simple Block........................................................................................23.4.3.2 The CMM Set up.........................................................................................2
3.5 Electrical Concept..............................................................................................23.5.1 Micro-controller...............................................................................................2
3.5.1.1 8051 Micro-controller..................................................................................23.5.1.2 DIOS Micro-controller................................................................................23.5.1.3 PIC Micro-controller...................................................................................23.5.1.4 BASIC Stamp Micro-controller...................................................................2
3.5.2 Vertical Movement Sensors.............................................................................23.5.2.1 Hall Effect Sensor........................................................................................23.5.2.2 Reed Switch.................................................................................................23.5.2.3 Magneto Resistive Devices..........................................................................2
3.5.3 Radial Movement Sensors...............................................................................23.5.3.1 Infra Red Proximity Detector (IRPD)..........................................................23.5.3.2 Potentiometer...............................................................................................23.5.3.3 Optical Interrupter.......................................................................................2
4 FEASIBILITY ASSESSMENT........................................................................2
4.1 The Stand..............................................................................................................24.1.1 Stand Concepts:...............................................................................................24.1.2 Attributes:........................................................................................................24.1.3 Level of Attainment Analysis..........................................................................24.1.4 Critical/Important Attribute List......................................................................24.1.5 Second Level of Attainment Analysis.............................................................24.1.6 Performance Feasibility...................................................................................24.1.7 Economic Feasibility.......................................................................................24.1.8 Technical Feasibility........................................................................................24.1.9 Schedule Feasibility.........................................................................................24.1.10 Final Decision..................................................................................................2
4.2 Plunger Depression............................................................................................24.2.1 Plunger Depression Attributes:........................................................................24.2.2 Concepts:.........................................................................................................24.2.3 Level of Attainment Analysis..........................................................................24.2.4 Performance Feasibility...................................................................................24.2.5 Economic Feasibility.......................................................................................24.2.6 Technical Feasibility........................................................................................24.2.7 Schedule Feasibility.........................................................................................24.2.8 Final Decision..................................................................................................2
4.3 The Micro-controller...........................................................................................2
2
TRANSCAT Pipette Calibration Project 05417
4.3.1 Micro-controller Concepts:..............................................................................24.3.2 Attributes:........................................................................................................24.3.3 Level of Attainment Analysis..........................................................................24.3.4 Performance Feasibility...................................................................................24.3.5 Economic Feasibility.......................................................................................24.3.6 Technical Feasibility........................................................................................24.3.7 Schedule Feasibility.........................................................................................24.3.8 Final Decision..................................................................................................2
4.4 Sensors for Vertical Tracking..........................................................................24.4.1 Concepts for Vertical Tracking:......................................................................24.4.2 Vertical Tracking Attributes:...........................................................................24.4.3 Level of Attainment Analysis..........................................................................24.4.4 Performance Feasibility...................................................................................24.4.5 Economic Feasibility.......................................................................................24.4.6 Technical and Schedule Feasibility.................................................................24.4.7 Final Decision..................................................................................................2
4.5 Radial Movement Sensors................................................................................24.5.1 Concepts for radial tracking:...........................................................................24.5.2 Attributes:........................................................................................................24.5.3 Level of Attainment Analysis..........................................................................24.5.4 Performance Feasibility...................................................................................2
5 SPECIFICATIONS, ANALYSIS AND SYNTHESIS.....................................26
5.1 Mechanical Analysis & Synthesis..................................................................25.1.1 Problem Statement...........................................................................................25.1.2 Known Information.........................................................................................25.1.3 Desired Information.........................................................................................25.1.4 Assumptions....................................................................................................25.1.5 Data and Analysis............................................................................................25.1.6 Quality Review................................................................................................2
5.2 Electrical Analysis..............................................................................................25.2.1 Design Specifications:.....................................................................................25.2.2 Micro-controller Specifications:......................................................................25.2.3 Sensor Specifications:......................................................................................25.2.4 Power Supply Specifications:..........................................................................25.2.5 Micro-controller...............................................................................................2
5.2.5.1 Selection of Micro-controller......................................................................25.2.5.2 Integration with Electrical System...............................................................25.2.5.3 Inputs:..........................................................................................................25.2.5.4 Outputs:........................................................................................................25.2.5.5 Serial:...........................................................................................................2
3
TRANSCAT Pipette Calibration Project 05417
6 BILL OF MATERIALS...................................................................................2
APPENDIX A: SYSTEM SCHEMATIC.................................................................2
APPENDIX B: ELECTRICAL SYSTEM................................................................2
APPENDIX C: FLOW CHART OF SYSTEM........................................................2
APPENDIX D: BEER-LAMBERT’S LAW...........................................................61
Appendix E: Hall Effect........................................................................................61
4
TRANSCAT Pipette Calibration Project 05417
1 Introduction:
Pipettes are widely used for delivering liquids in chemistry and life
sciences laboratories, where their accuracy and precision are critical to achieving
good results. In recent years, much attention has been focused on how best to
calibrate these devices in centralized metrology laboratories, but little attention
has been given to assuring good quality of liquid delivery in the place of use.
Transcat has noted this and is currently looking into ways to calibrate these
devices so as to expand the range of services.
Current calibration procedures require a technician with many hours of
training to manually operate a pipette and measure how much fluid is transported
compared to how much is metered on the pipette. Tools that are used for this
form of calibration are the Artel PCS ® colorimetric calibrator and the Artel
Pipette Tracker® software. These tools take the guesswork out of measuring the
fluid delivered and also perform the statistical equations needed to find the
accuracy and precision of the micropipette. To be specific the technician inserts
the fluid from the pipette into a vial that is inside of the PCS®, which then uses
Beer-Lambert’s Law to find the volume that is inside the vial. We will explain this
law later in the paper. The PCS® then uses serial communication to send the
results to the Pipette Tracker® software which then perform statistical equations.
Unfortunately the human interface tends to create quite a bit of variability
in the calibration process. That is why Transcat desired to create a system which
will automate the system as much as possible so as to reduce the variability and
increase the reliability of calibration process. Our current design will accomplish
this primarily by use of pneumatics.
The semi automatic pipette calibrator will require the technician to start by
inputting the unique code for the pipette in computer interface. The technician will
then be prompted to move the device over the vial containing the fluid, which is
the starting position. The device will then submerge the pipette tip into the fluid
5
TRANSCAT Pipette Calibration Project 05417
and, once told to by the technician; will aspirate the liquid into the pipette tip. The
pipette will then be raised and the technician will be prompted to move the device
over the calibrator. Again, the device will insert the pipette tip into a vial. Then the
technician will prompt the device to first dispense the fluid and then clear out the
tip. The pipette will be raised again and the process will be repeated. This cycle
will go on for 15 times, with three runs each at five equidistant volumes within the
range of the pipette.
2 Recognize and Quantify Need2.1 Project Mission Statement:
The goal of student engineering design team was to investigate viable
processes and approaches to semi automate the pipette calibration process for
as many pipette as possible. The final design consist of several pneumatic
cylinders, a manually turning stand, several sensors to find the precise
positioning of pipette, a computer interface for the technician, as well as the Artel
PCS® and the Pipette Tracker® software.
2.2 Company BackgroundTranscat is one of North America's leading providers of calibration services and
instrumentations. Since its' incorporation in 1964, Transmation (now Transcat)
concentrated on developing, servicing and distributing electronic instrumentation
used in the monitoring, calibration, and supervision. The primary markets for
Transcat's services are process, life sciences, manufacturing, communications,
automotive, and aerospace industries.
Transcat, Inc.'s former manufacturing organization located in Rochester, New
York, was comprised of the Transmation Instrument Division, established in
1964, and Altek Industries, which was purchased by Transmation, Inc. in 1996.
The two groups combined in 1999 to form the Products Group in 1980, the
Transcat (short for "Transmation Catalog") Division was established as a catalog
sales operation to offer customers a single source for calibration and test
6
TRANSCAT Pipette Calibration Project 05417
instrumentation. This operation has grown into a full-fledged industrial distribution
network for not only Transmation and Altek products, but also those of more than
200 other manufacturers.
As catalog sales increased at a dramatic pace, it became evident that this type of
equipment would require periodic recalibration and general maintenance in order
to perform at peak level. Thus, in 1988, Transcat opened the first of many
calibration laboratories to service customer equipment. The purchase of E.I.L. in
1997 and MeterMaster in 1999 (both distributors & calibration service
organizations with laboratories located throughout the United States and
Canada) established Transcat as the leading calibration service provider in the
U.S. Finally, the newest business unit of Transcat established in 1999,
MetersandInstruments.com, provides customers with an "Internet" channel for
the purchase of calibration equipment and tools. In late 2001, as part of their
strategy to divest non-core businesses, Transcat sold both the Products Group
Division and the MAC (Measurement and Control) Division. This allowed them to
focus on providing innovative, quality products and calibration services to their
customers. Upon completion of the sale, the company name was changed from
Transmation, Inc. to Transcat, Inc. Today, Transcat, Inc. employs 230 talented
individuals throughout the U.S., Canada, and China.
2.3 Calibration: An element of Metrology
Metrology is a branch of science that was created in France and consists
of the quantification of weights and measurement. Metrology is used in everyday
life to make sure that the instruments and the systems that we use can
accurately and reliably perform its designated task. Due to many cases of fraud
in market place there were laws that were created to regulate these
measurements. Today scientists instead of lawyers handle the regulation of
measurement, and they are formed at an international basis so as to create a
common base for all researchers to work on. With this cooperation, the level of
precision of measurements has risen considerably, improving not only the
7
TRANSCAT Pipette Calibration Project 05417
methods of research but also improving the quality of the products created by
this research.
Of particular interest to quality control (QC) is the element of metrology
called calibration. Calibration is the process of comparing measurements, made
by an instrument with a standard. The instrument, which is of unconfirmed
accuracy, is referred as the unit under test (UUT) and the instrument of known
accuracy is known as a measurement standard. Calibration is performed to
establish the accuracy of the UUT for measurements. Instruments that do not
meet the standards are adjusted and then tested again until they meet the
standards.
2.4 Product Description
The primary goal of students on the Transcat design team was to design, build,
test, and debug a working prototype device that incorporates a majority of the
design and will work under the specifications listed below:
Test Stand: A mechanical setup that will hold the pipette in place as the
calibration is performed. The test stand will move in two directions,
vertically and either rotationally or horizontally.
Plunger Depression Device: A mechanical device that will insure the
proper amount of force will be applied for the depression of the plunger for
both the dispensing and tip blow out phases of pipette evacuation.
Automated Digital Controls for some of the above features and for the
data interface to the PCS® and Pipette Tracker® software.
User Interface for the technician. This interface will allow the technician to
properly use the semi-automated system. This could be either a physical
or graphical interface.
8
TRANSCAT Pipette Calibration Project 05417
2.5 Scope Limitations
The prototype was fully designed by the end of the fall quarter and a
working device will be completed by the end of the spring quarter. At the end of
the fall quarter the senior design team presented the preliminary design, detail
sketches, and cost for the components needed to build the prototype. During the
spring quarter, the design team will participate in on-site testing, data collection,
and evaluation of prototype device.
As with any design, though, there are limitations to the scope. This simply
means the design will only be able to perform a specific set of tasks. The tasks
that we are designing for will meet the customer requirements as discussed
between the team and the customer.
The design team will be responsible for a device that:
Automates the human element of the calibration process
o Depressing the plunger
o Raising and lowering the pipette
Maintains a .25% accuracy in liquid dispensed during the calibration
process
Performs three measurements at each of five equidistant points within the
volumetric range of the pipette.
Keeps the pipette vertical within +/- 1 degree
Prevent light from contaminating the photosensitive fluids
Reduce human vibrations by use of semi-automation.
Allows human control over the processes being performed by use of the
Graphical User Interface.
Has a complete user’s manual for future reference.
9
TRANSCAT Pipette Calibration Project 05417
The design team will not be responsible for:
Taking apart the micropipettes to perform manual replacement of parts
Calibrating the Artel PCS® colorimetric pipette calibrator.
Initiating changes to the Artel Pipette Tracker® software.
Training individual users on the setup after the product has been
integrated into the everyday process of calibration.
2.6 Stakeholders:
The primary stakeholder for the research and development of the semi
automated system is Transcat, since the device will be integrated with the PCS®
and Pipette Tracker® software. The students are the secondary stakeholders in
this project because it will satisfy the engineering curriculum requirements for the
students in the design team and it will serve as a learning experience for them.
Eventually, the entire calibration industry could be a stakeholder, especially
seeing as not much work has been done in the field of automating colorimetric
calibration.
2.7 Key Business Goals:
In the field of calibration slight differences in precision can make the
difference between success and failure. After all, everyday more and more
companies are providing calibration services making the competition very tight.
Because of this is necessary for the variation to be as small as possible. Seeing
as the laboratories are at controlled environments, one can rule out variation due
to surrounding and concentrate instead on the variations caused by the
technicians. The main way to get rid of this is, of course, by removing as much
interaction as possible by use of automation. This prototype aspires to do as
much of this as possible, giving Transcat the competitive edge needed to
succeed in this field of calibration.
10
TRANSCAT Pipette Calibration Project 05417
2.8 Top Level Critical Financial Parameters
The top level critical financial parameters related to the project are
associated with the following components that are needed to build the prototype.
The pneumatic cylinders
The controls for the pneumatic cylinders
The micro-controller for the system
2.9 Financial Analysis
The team has been given a definite budget of $500 by Transcat.
Therefore, it is important that the team create a low cost system to fit this budget.
The most costly component will likely be the pneumatic cylinders that will actuate
the system. The total cost of the system will depend on the following
components:
The cylinder
Microprocessor
The sensors
The pneumatic controls
Bearings for rotation
The pipette holding fixture
Raw materials needed to create the base
Magnets to be put on the cylinder piston head
Miscellaneous electrical and mechanical components
11
TRANSCAT Pipette Calibration Project 05417
2.10 Primary Market
The primary market will be Transcat because the system is customized to
be used in their lab.
2.11 Secondary Market
The secondary market will be any other company that is interested in providing
colorimetric calibration services for micropipettes. In order for this system to be
part of the secondary market, modification will have to be made to the design to
enable it to be mass-produced.
2.12 Order Qualifiers
The primary market requires is that this prototype performs calibration runs with a
variation at or under 0.25% of volume dispensed. The setup will have to be able
to fit in the laboratory.
2.13 Order Winners Universal acceptance of pipettes
Reduce test time
Simpler calibration procedure
Simpler setup
Meeting all the project requirements
2.14 Formal Statement of Work
The RIT engineering designing team shall work closely with Transcat
technical staff to understand customer’s needs and incorporate these
requirements into the final product. The team will research through the
appropriate means to identify the various methods to which they will be able to
12
TRANSCAT Pipette Calibration Project 05417
automate the colorimetric calibration process. The design team will take part
constructive and on going discussion to understand the need for the balance
between the desires of the engineers, the needs of the customer, and product
feasibility. The team will also create appropriate design and specification options
that will be reviewed by Transcat. The design team is primarily concerned with
building a working prototype that incorporates all the project requirements that
were agreed upon with Transcat. In addition, the team will perform testing, data
collection, and evaluation of the prototype period. The design team will prepare a
professional written report including complete technical specification,
construction information, and experimental results, along with other information
that will be presented to the corporate managers.
Transcat will provide the engineering design team with a technical point of
contact that will provide guidance, along with feedback and field engineering
information pertaining to the operation requirements and use of the system. In
addition, they will provide access to relevant support information, including
documentation and case studies, which will be needed to complete the project,
along with access to their local calibration laboratory for insight operation and
testing evaluation. Financial support for all components needed in the build will
also be provided. All measurement instruments and components associated with
project are subject to Transcat review and approval.
Agreed by:
Jeff Youngs ____________________ Date:_______(Design Team Leader)
Jon Schneider___________________Date:_______(ME Design Team Member)
Ashish Rathour__________________Date:_______( EE Design Team Member)
Glenn Carroll____________________Date:_______(EE Design Team Member)
Tai To_________________________Date:________(EE Design Team Member)
13
TRANSCAT Pipette Calibration Project 05417
Mayank Rathour_________________Date:________(EE Design Team Member)
Howard Zion____________________Date:________(Customer Representative)
Rainer Stellrecht_________________Date:________(Customer Representative)
George Slack___________________Date:________(Faculty Coordinator)
Mark Hopkins___________________Date:________(Faculty Mentor)
Wayne Walter___________________Date:________(Faculty Mentor)
14
TRANSCAT Pipette Calibration Project 05417
3 Concept Development3.1 Concept Development Process
The concept development process has the main objective of developing
several design objectives that will each meet the design specifications and to
improve these ideas through interaction among team members. The design team
use brainstorming technique to produce variety of concepts that would be used in
further discussion. Team members also sketch conceptual drawings to provide a
clear picture of some of the proposed concepts. Finally, the team use the design
specifications to develop the ideas into a useful form.
3.2 Brainstorming Session
In the brainstorming session each of the team members pitched in to
create a large quantity of concept that could be used in creation of prototype. The
design team listed concepts on different design components of the project. A list
of these concepts was made and each of the team members was given two
votes for each component so that they could identify which concept that they
thought best met the design of specifications. The voting also helped in
consolidation of concepts. Table 1 shows the ideas generated during the
mechanical portion of the brainstorming session and how popular each concept
was.
15
TRANSCAT Pipette Calibration Project 05417
Design Component Concept Votes Rank
Plunger depression 1 Stepper motor 7 1
2 Pneumatic system 5 2
3 Hydraulic system 0 3
4 Spring Loading 0 3
5 Ratcheting mechanism 0 3
Stand 1 No stand 1 3
2 Basic stand 3 2
3
Stand with vertical
movement 4 1
4
Stand with radial
movement 4 1
5 Clamps 0 4
Vial Holder 1 Block 12 1
2 CMM setup (Grid) 0 2
Table 1 Mechanical Brainstorming Session
Afterwards the team voted on the electrical concepts of the prototype. The concept and
their popularity are listed in Table 2.
Design Component Concept Votes Rank
Microprocessor 1 8051 2 2
2 Pic Micro 7 1
3 Basic Stamp 1 3
4 DIOS 2 2
Vertical Movement
Sensors 1 Hall Effect Sensor 11 1
2 Reed Switch 0 3
3 Magneto Resistive Device 1 2
16
TRANSCAT Pipette Calibration Project 05417
Radial Movement
Sensors 1 Infra Red ( IRPD) 1 2
2 Sonar 0 3
3 Potentiometer 1 2
4 Optical Interrupter 10 1
5 Hair Trigger Switch 0 3
Table 2 Electrical Brainstorming Session
3.3 Conceptual Level Drawing
The team members that have the most popular designs then created
conceptual level drawings of these designs. These drawings were used to help
the team obtain an idea of how the concept would work. These conceptual
drawings can be found in appendix A.
3.4 Mechanical Concepts
Each of the design components has the list of concepts related to them.
The design team discussed top two or three concepts for each design
component.
3.4.1 Plunger Depression
In order to transport the fluids, the operator needs to depress the plunger
and release it in the fluid to draw it into the pipette tip, and then dispense the fluid
and blow out the tip. As one can imagine this was one of the most important
parts of the project. Therefore, much emphasis was put into in depth
explanations of the top two concepts.
3.4.1.1 Stepper Motor
This concept had two possible implementations. Each of them used the
precise movement of the stepper motor to its advantage. The first idea was to
17
TRANSCAT Pipette Calibration Project 05417
attach a gear to the shaft of the stepper motor and use the gear to drive a pinion
gear. This pinion would actuate vertically in order to either depress or release the
plunger of the pipette. The simplicity of this design was rather attractive to the
team. The other idea was to attach a lead screw to the shaft of the stepper
motor. A nut would be attached to the lead screw. In order to prevent the nut
from merely turning along with the screw, the nut had to be properly restrained.
There were several ideas as how to do this, but the winning idea had the nut
constrained by two arms that would run on two tracks that are part of the setup.
Several team members noted that this setup would be rather difficult to repair
should any problems arise. This was noted and brought up later in the feasibility
assessment.
3.4.1.2 Pneumatic Systems
This concept was originally unpopular due to the misconception that the system
would be very expensive. This was due to the fact that most of the pneumatic
systems that the team had viewed included an expensive system of machine
made servos. After this misunderstanding was cleared, the team was more eager
to work with this concept. At the time, the team felt that a simple pneumatic
cylinder with a spring return system would suffice. As per the control of the
pneumatic cylinder, the team felt that there were only two points of interest; that
of the first phase of depression and the blowout phase. Each of these phases
had their own unique range of pressures allowed and these pressures could
easily be converted to forces by taking into account the surface area of the piston
head. Therefore, one could control the system by using two distinct pressures,
each near the separation point between the two phases. All one would need to
go from one phase to the other would be a simple switching mechanism for the
two distinct pressures.
18
TRANSCAT Pipette Calibration Project 05417
3.4.2 Stand
Seeing as this part of the design would be holding the actual pipette in
place, this is easily one of the most important parts of the design. At the time the
team was divided on whether or not to actually have this stand move. The top
three concepts are listed below:
3.4.2.1 Stand with vertical movement
This concept was one of the hardest to conceptualize. The reason for this
was the fact that it would have to act against the forces of gravity and moving in
accurate manner. One of the first ideas that came up was to use a stepper motor
with a gear attach to its shaft and to have this gear interact with a pinion gear.
This time, however, the stepper would be the moving part instead of the pinion.
After consideration, we found that this would not be usable due to two reasons:
the resolution of movement would be too inaccurate and even if the gear teeth
were small enough to accommodate for accuracy, the teeth would then not be
able to support the relatively heavy load. It was at this time that we considered
the use of a pneumatic cylinder for the linear actuation. The piston would be
double acting, in which there would be two air supplies, one in the inlet orifice
and the other in the exhaust orifice. These two supplies would act against each
other so as to let the piston inside the cylinder move in a gradual manner instead
of slamming it into its fully opened and fully closed positions.
3.4.2.2 Stand with radial movement
Though the title of this concept refers only to radial movement, there was
talk of horizontal movement along with radial movement as we discussed this
concept.
When it comes to horizontal movement, we had quite a few ideas. One of
the first ideas was to use a manually operated trolley system. In this idea, a
simple “trolley”, which was actually a piece of metal with four wheels attached to
19
TRANSCAT Pipette Calibration Project 05417
it, would have its wheels inside of an extruded piece of plastic. One could then
move the trolley back and forth with a fairly smooth movement. One of the main
problems, however, was the fact that there would be no speed control, thereby
allowing for the possibility of the trolley coming to an extremely abrupt stop. We
found, after a few rounds of experimentation, that abrupt stops such as these
tend to cause extreme errors in calibration, mainly due to the fluid becoming
stuck in the tip. After we done with this idea, we advanced to the possibility of
moving the calibration fluid and calibrator instead of moving the pipette. To be
specific, we thought it would be a good idea to attach a table to the top of a
pneumatic cylinder. This cylinder would also be double acting so as to reduce the
possibility of jerky movements in the process. However, as will be seen in the
feasibility assessment, this option was rather expensive, especially for a small
budget such as ours.
When it came to radial movement we had a fair share of ideas as well.
One of the first ideas was to automate this by use of a rotating turntable. In
theory, this idea would work rather well. We would merely need to use a DC
Motor to turn the whole set up and use a system of bearings to make the
movements as smooth as possible. However, as with the other table idea, we
found that this was a rather expensive method and that it would be fairly hard to
control. The team’s other idea was that of manually moving the stand by use of a
simple handle. Again, we would have to use a system of bearings in order to
make the movements smooth, but with this setup, it would be rather easy to
create a system of stops to properly restrain the movement. The simplicity of the
set up made particularly attractive.
3.4.2.3 The Basic Stand
This setup merely called for the use of a rod that was fixed to the ground.
We thought of this concept when we were putting together an experimental setup
for preliminary data collection. In the experimental setup, we had a ring stand
that we attached specially made clamps to. Seeing as the results that were
20
TRANSCAT Pipette Calibration Project 05417
obtained from this experimental setup weren’t very inaccurate, this made the
possibility of using such a simple set up in our final design seem very plausible.
However, we also had to think about how inconvenient such a set up would be
for the technician that would have to use it.
3.4.3 Vial Holder
Though it may seem a very simple component to the system, the restraint
of the vial plays a very critical role in the calibration process. After all, one needs
to have a common place for the vial to be located at if one wants to efficiently use
their time when calibrating. Also, a restraint would help to prevent the vial from
spilling in case an accident should happen.
3.4.3.1 The Simple Block
As the title suggests, this setup is quite simple. One merely has to get a
block of certain material, possibly wood, and drill a hole into it, so that the vial will
fit into the block. This will be enough to satisfy the two criteria that the team set
for this design component.
3.4.3.2 The CMM Set up
This set up is based on the system used by Coordinate Measurement
Machinery (CMM). In this set up, there is a grid on the table and each of the
coordinates has a screw hole in it. The purpose of these holes is to allow various
attachments to be placed on the table in a secure fashion. This is very useful for
securing parts that are unusual shapes, but we are merely working with a
commonly shaped vial. Thus, the design would be too advanced for our needs.
Still, the grid is rather useful for a coordinate system for the movement of the
pipette. However, the two positions that the pipette would be moving to will be
fixed, negating the need for a coordinate system.
21
TRANSCAT Pipette Calibration Project 05417
3.5 Electrical Concept3.5.1 Micro-controller
In order to maintain control of the system, we found it necessary to include a
micro-controller. This component would keep track of all the actions and,
depending on state of the system, would tell the system to perform certain
actions. This is the heart of the electrical system.
3.5.1.1 8051 Micro-controller
The 8051 is an 8-bit microprocessor originally designed in the 1980's by
Intel that has gained great popularity since its introduction. Its standard form
includes several standard on-chip peripherals, including timers, counters, and
UART's, plus 4k bytes of on-chip program memory and 128 bytes (note: bytes,
not Kbytes) of data memory, making single-chip implementations possible. Its
hundreds of derivatives, manufactured by several different companies (like
Philips) include even more on-chip peripherals, such as analog-digital converters,
pulse-width modulators, I2C bus interfaces, etc. Costing only a few dollars per
IC, the 8051 is estimated to be used in a large percentage of all embedded
system products.
3.5.1.2 DIOS Micro-controller
The DIOS Micro-controller made by Kronos Robotics is an 8 bit micro
controller that comes in both 28 and 40 pin packages. The 40-pin package has
33 I/O pins, 8 A/D ports, and a 40 MHz internal clock. It includes 32 Kbytes of
program space (16 Kbytes for user programs), 256 bytes of internal memory, and
a programming port that, along with the RS232 interface included on the carrier
board can be used to communicate with the PC. The DIOS uses a form of the
BASIC language and can interface with a simple Visual Basic GUI on a PC. The
compiler software is available for free from Kronos Robotics.
3.5.1.3 PIC Micro-controller
22
TRANSCAT Pipette Calibration Project 05417
The PIC Micro-controller is yet another 8-bit micro-controller. PICs are
available in a wide range of packages and amount of pins. They are also
available in a wide range of internal clock speeds and a considerably large range
of memory size. There is a wide range of compilers available for them, however
the better ones are not free. An evaluation board is not necessary for its
operation, however a programming board or socket is required to program the
PIC. These range in price from the affordable $50, designed for on PIC in
particular, to the extreme $1000 programmers that can program any PIC you
throw at it
3.5.1.4 BASIC Stamp Micro-controllerBASIC Stamp is a micro-controller made by Parallax. They are available in
24 and 40 pin packages. Like the DIOS, they run on a form of the BASIC
language, making them easy to program. They have a serial port for
programming and interfacing with a PC.
3.5.2 Vertical Movement Sensors
This form of sensor would keep track of the vertical movement of the
stand. After much research, we found that some of the easiest and most widely
accepted sensors had to do with magnetism. This formed our concept
development, as one can see from table 2.
3.5.2.1 Hall Effect Sensor
This proximity sensor is very cheap and very compact. These sensors are
basically mounted to the side of the stand and keep track of any magnetic fields
around them. The theory behind the Hall effect is talked about in more detail in
Appendix E. These sensors are especially useful in pneumatic setups, because
most suppliers readily supply air cylinder with magnets on piston heads. This
way, one can see where the piston head is by keeping track of the magnetic field
around it.
23
TRANSCAT Pipette Calibration Project 05417
3.5.2.2 Reed Switch
This sensor is not a sensor at all. It is, as the name implies, merely a
switch. Its resolution is rather poor, with it closing the circuit it is on when a
magnetic field is up to 0.2 inches away and opening the circuit when the
magnetic field is at least 0.3 inches away. While these measurements may seem
small, they could easily ruin our system if they presented themselves.
3.5.2.3 Magneto Resistive Devices
These devices are quite similar to Hall Effect sensors. The main difference
between these two sensors is that the MRD sensors measure the change in a
magnetic field instead of the presence of one. This is accomplished by a variable
resistor in the device that changes value as the strength of the magnetic field
changes. This would be very useful if we were to keep track of how far away a
piston head is, if that piston head has a magnet on it.
3.5.3 Radial Movement Sensors
At this point of our project, we had decided to use radial movement
instead of horizontal movement. The fact that the path of motion is curved
instead of straight may for a fairly difficult concept development. However, this
difficulty inspired us to think of some fairly inventive sensor setups, as one can
see below.
3.5.3.1 Infra Red Proximity Detector (IRPD)
The IRPD is a proximity sensor, that is to say that it merely detects if
object is there or not. The IRPD acts both as a transmitter and receiver of an
infra red light beam. The beam reflects back to the sensor and is only noticed by
the receiver if the distance of reflection is small enough so that the infra red light
does not diffract too much. This is a fairly cheap and easy to mount sensor, but it
is very limited to its uses in a real world scenario.
24
TRANSCAT Pipette Calibration Project 05417
3.5.3.2 Potentiometer
The potentiometer is a variable resistor that gains resistance as the knob
is turned further away from its starting point. This is achieved by setting one end
of a circuit at the starting point and the other end of the circuit on a piece of metal
that is attached to the knob of the potentiometer. The piece of metal travels on a
resistive film, and as it goes further away from the starting point, there is more
and more resistive film in the circuit, increasing the resistance. One can keep
track of just how far the knob has rotated by either keeping track of voltage in the
circuit or finding the RC Time constant for a circuit containing the potentiometer
and a common capacitor. Either way, one has to experimentally find some
constant values for certain known angles of rotation. One can then linearly
interpolate values based on these known constants. We can use this to our
advantage by attaching a potentiometer to the stand and seeing how much the
knob turns as the shaft it is attached to moves.
3.5.3.3 Optical Interrupter
This sensor is very similar to the IRPD in concept. The difference, though,
is that the transmitter and receiver of the infrared light are two different parts. The
Optical Interrupter keeps track of any objects in its limited vicinity by seeing if any
infra red light reaches the receiver. If not, something is in the path of the infrared
light. Usually people use incremental encoder discs in conjunction with the
optical interrupter. To be specific, the incremental encoder discs have a series of
translucent lines that are etched into an opaque surface. Whenever the opaque
portion of the disc is inside of the proximity of the optical interrupter, there is no
infra red light that gets to the receiver. When the translucent portion of the disc is
between the two parts of the optical encoder, some infra red light gets to the
receiver. One can easily use this to his or her advantage by making the
translucent portions into thin lines that stand for certain discrete positions that the
25
TRANSCAT Pipette Calibration Project 05417
stand can be at. For our setup, we would permanently attach the incremental
encoder disc to the stand. The optical interrupter would be mounted so as to
allow the encoder disc to be equidistantly between the transmitter and receiver.
As the shaft rotates, the disc rotates and lines in the disc rotate as well.
26
TRANSCAT Pipette Calibration Project 05417
4 Feasibility Assessment
As a result of brainstorming and initial research, multiple concepts were
created for each of the design components. All of these concepts were
submitted for a feasibility assessment. The objective of the feasibility assessment
was to understand what was involved with each concept and decide which
concept had the highest probability of success. Each concept was evaluated on
the feasibility of its’ technical, economic, scheduling and performance aspects.
Each concept was assigned a rating between -, 0 and + for each of the attributes.
A score of 0 was a representation of a concept grading the same as the baseline.
A score of - represents a design that is slightly worse that the baseline design
and + represents a design that is an improvement over the baseline design. A
concept that received a - meant that it had very low feasibility. These parameters
were used to judge the concepts. Afterwards, we described the feasibility of the
designs in more detail. The discussion took on four aspects:
1. Performance Feasibility
2. Economic Feasibility
3. Technical Feasibility
4. Schedule Feasibility
The following sections present the feasibility for each of the design component
concepts.
4.1 The Stand
The first design component that we performed feasibility assessment on
was the most important component in the design. Because of this, we put the
most effort into this feasibility assessment, as can be seen by the number of
attributes used in the Pugh Analysis.
27
TRANSCAT Pipette Calibration Project 05417
4.1.1 Stand Concepts:
Baseline Concept: Stationary Stand
A: Pneumatic Cylinder/Handle
B: Hanging Box /Trolley
C: Rotating Table
D: Pneumatic table movement/Pneumatic Stand
4.1.2 Attributes:1. Horizontal Movement
2. Vertical Movement
3. Radial Movement
4. Smoothness of operation
5. Stability
6. Repeatability
7. User-friendly
8. Ability to support weight
9. Cost-efficient
10.Ease of setup
11.Ease of control
12.Ease of build
28
TRANSCAT Pipette Calibration Project 05417
4.1.3 Level of Attainment Analysis
Attribute A B C D
1 0 + 0 +
2 + 0 0 +
3 + 0 + 0
4 + 0 + +
5 - - - -
6 + + + +
7 + + + +
8 0 - 0 0
9 - - - -
10 0 - 0 0
11 + 0 + +
12 - - - -
Table 3 Pugh Analysis of The Test Stands: Baseline = Stationary Stand
As one can see from the Pugh analysis in Table 3, there was a tie in
feasibility between concepts A & D: the two pneumatic stands. To take care of
this, we decided to go through another round of Pugh Analysis. This time, we
decided to reduce the amount of concepts to only A & D and reduce the
attributes to the most important ones. This more specific Pugh analysis can be
seen in Table 4. We agreed that the following attributes were most important:
4.1.4 Critical/Important Attribute List1. Horizontal Movement
2. Vertical Movement
3. Radial Movement
29
TRANSCAT Pipette Calibration Project 05417
4. Smoothness of operation
5. Stability
6. Cost-efficient
7. Ability to support weight
8. Cost-efficient
4.1.5 Second Level of Attainment Analysis
A D
1 0 +
2 + +
3 + 0
4 + +
5 - -
6 - -
7 0 0
8 + +
Table 4 Second Pugh Analysis of The Test Stands: Baseline = Stationary Stand
Again, there was no clear winner, so we decided to go further into the
analysis. We considered all the aspects of assessment as were discussed
beforehand. Seeing as we had already ruled out two of the concepts, we decided
to only analyze the two pneumatic stands.
4.1.6 Performance FeasibilityBoth of these designs require the use of pneumatic cylinders. The first
concept requires only two cylinders, while the second requires three. Seeing as
we decided to use a valve bank, this was not a true problem. The main problem
was that of the pressure required.
30
TRANSCAT Pipette Calibration Project 05417
After talking with an industrial representative, we found that all of the
pneumatic components would be able to run properly on 60-80 psi, a level that is
usually found in laboratories. While we were there, though, the representative
brought up a problem that we had not thought of yet.
The problem was that putting a table on a pneumatic cylinder introduces
quite a bit of vibration when the cylinder gets to its two end points. This is due to
the fact that the cylinder operates, on a base level, with two mechanical stop
points. When these stop points are hit, the system comes to an abrupt stop and,
as a result, has quite a bit of vibration. We would be able to combat this by
making the cylinder double acting, in which both the exhaust and inlet ports
would have air pressure sent through them. In this setup, as one slowly reduces
the pressure on one side, one can either slowly speed up or slow down the
system. Still, there would be some vibration. Still, the other setup is mainly
hands-on and can also have quite a bit of vibration, depending on how hard the
technician swings it. However, one can handle this kind of vibration, seeing as
this only affects the performance of a single pipette. The vibration from the table,
however, would affect the calibrator, which would effect all future calibrations.
Thus, in this assessment, we decided that the concept of the manually moved
stand would be the best.
4.1.7 Economic FeasibilityAs was said before, both setups center on the pneumatic cylinders
Therefore, each of the designs have the same base cost for the vertically and
radial moving cylinder that holds the pipette and the other cylinder. The main
difference will be in the cost of the pneumatically controlled table.
When we talked to the pneumatic representative, we found that a common
table setup, including pressure control, tracks and sensors, would cost in a range
from $200-$300. This would consume most of our budget, leaving little room for
expenses that may come up later in the design. The manually controlled setup,
however, would have a main cost of bearings so that the setup can rotate
properly without wearing out the shaft of the pneumatic cylinder. The cost for this
31
TRANSCAT Pipette Calibration Project 05417
is much less than that of the table setup. Thus, the manually controlled setup has
the definitive advantage in this feasibility, as well.
4.1.8 Technical FeasibilityIn this form of feasibility, we assessed how well we could apply our
knowledge to creating a working component. For the automated setup, we would
have to know how to machine a serviceable track from extruded metal so as to
reduce costs. This is a very rudimentary process, so we could easily perform this
task. Also, we would have to create a form of locomotion for the table, most likely
through wheels that would run on the track. We were fairly unclear as to how to
do this, so our knowledge in this aspect was lacking. Finally, we would have to
find a way to smoothly actuate both cylinders. We had done some research on
this and, with the help of our Mechanical Engineering Advisor, Professor Wellin,
and the pneumatic representative; we felt that we had a solid base of knowledge
from which we could easily control the system. When it came to the manual
setup, we had to know how to move the main cylinder manually. Seeing as this
will be a simple setup in which the cylinder will move with the help of bearings in
the table on which it stands, this would not be too hard for our level of
knowledge. As per the actuation of the cylinder, it would be the same as the
automated setup, but we would only have to control one piston, making it easier
as a whole. Again, the manual setup has the advantage in the feasibility
assessment.
4.1.9 Schedule FeasibilitySeeing as both designs are quite alike, the time schedules for the two
setups will also be similar. However, as was talked about in the Technical
Feasibility, we would have to do more machining for the automated setup,
creating a longer time schedule for setup. Also, the automated setup would have
to have extra programming and sensors to control the pneumatic cylinder on
which the table stands. This would also increase the time schedule for setup.
Seeing as we need to concentrate more on improving the setup than creating the
32
TRANSCAT Pipette Calibration Project 05417
initial setup, it would be best if we had as little initial setup time as is possible.
Thus, the manual setup has the time advantage.
4.1.10 Final DecisionAfter going through all four forms of feasibility assessment, we can
conclude that the manual setup is the best for what our project.
4.2 Plunger Depression
This design component was a very important component in the design.
After all, if the pipette plunger cannot be depressed, one cannot transport fluid
from the vial to the calibrator. However, this setup is fairly simple, so we didn’t
have to account for too many attributes, as can be seen in the Pugh Analysis in
Table 5.
4.2.1 Plunger Depression Attributes:
1. Repeatability
2. Two stokes
3. Accuracy I force/distance
4. Ease of replacement
5. Ease of setup
6. Ease of control
4.2.2 Concepts: Baseline = Pneumatic Cylinder
A. Stepper motor with a lead screw
B. Spring loaded actuation
33
TRANSCAT Pipette Calibration Project 05417
4.2.3 Level of Attainment Analysis
Attribute A B
1 0 -
2 0 -
3 + -
4 - +
5 - +
6 - -
Table 5 Pugh Analysis of Plunger Depression: Baseline = Pneumatic Cylinder
As one can see from the Pugh Analysis, the baseline concept was the
winning concept. In order to reinforce this finding, we went through the four steps
of feasibility assessment.
4.2.4 Performance FeasibilityWhen we looked at the concepts, we decided right away that the spring-
loaded actuation was not feasible. From a mechanical engineering viewpoint,
whenever a spring is included in the system, there are many variances that come
into play. Also, any performance from the system will be fairly rough due to the
fact that the springs may not move unless a certain load is achieved, reducing
the gradual nature of the actuation. Seeing as our experimentation has
concluded that gradual actuation is preferred over rough actuation, we decided
that this concept isn’t feasible from a performance standpoint.
Afterward, we compared the pneumatic cylinder to the stepper motor.
When it comes to the precision of movement, the stepper motor has the
advantage. After all, the stepper has a system of discrete steps that, depending
on the controller, can be controlled to an accuracy of 1/64 of a step, or about
34
TRANSCAT Pipette Calibration Project 05417
1/32 of a degree. The true resolution, however, depends on the number of
threads per inch on the lead screw. Seeing as this parameter can be rather
small, this is not a problem.
When it comes to the cylinder, our control of the actuation is through force
instead of length measurement. For our means, we can control this parameter by
finding the two pressures needed to get to each level of actuation. After
experimentation, we found two values that would work for all models. This
method of actuation is preferable, seeing as the pipettes may not be mounted the
same way for each case. By using force stops, we can ignore differences in
length. We can do this with the stepper as well, but a force sensor will need to be
added to the design. Therefore, the slight advantage for performance goes to the
stepper motor.
4.2.5 Economic FeasibilityFor this case, we decided to create a miniature Bill of Materials so as to
make a more concrete analysis of our design. The approximate costs are listed in
Table 6 for the pneumatic system, and Table 7 for the stepper motor system.
Part Qty Cost
Total
Cost
Cylinder 1 ~$35 ~35
Regulators 2 ~$10 ~20
Switch Between
Regulators 1 ~$5 ~5
Total ~$60
Table 6 Estimated Cost of Pneumatic System
35
TRANSCAT Pipette Calibration Project 05417
Part Qty Cost
Total
Cost
Stepper Motor 1 ~$20 ~$20
Force Sensor 1 ~$25 ~$25
Stepper Controller 1 ~$5 ~$5
Lead Screw 1 ~$10 ~$10
Total ~$60
Table 7 Estimated Cost of Stepper Motor System
As one can see, there is no clear advantage to either side. The only
advantage that one could possibly think of would be the fact that the pneumatic
system would be able to be bought all from the same supplier, making it possible
to get bulk discounts.
4.2.6 Technical FeasibilityIn this form of feasibility assessment, we thought not only about how much
we knew about the subjects, but also about the simplicity of the design. When it
comes to knowledge of the subjects, most of the team has had no real
experience with either pneumatic cylinders or stepper motors. The team leader,
though, has had extensive experience with use of stepper motors in robotic
design. Also, some of our advisors are knowledgeable about the use of stepper
motors. On the other hand, the team has advisors that are very knowledgeable
about pneumatics and all of the principles of pneumatic use have been covered
in the Mechanical Engineering syllabus.
When it comes to the simplicity of the design, however, the pneumatic
system has the clear advantage. One merely needs to attach tubing to the
regulators; the switch and the cylinder to do most of the setup. The stepper
setup, however, tends to have a more complex setup. This is due to the fact that
one has to constrain a nut that is on the lead screw, so that the nut moves
vertically instead of having radial movement with the rotating screw. Thus, when
we finished analyzing this feasibility, we found that the pneumatic setup had the
advantage.
36
TRANSCAT Pipette Calibration Project 05417
4.2.7 Schedule FeasibilityAs was said in the technical assessment, the setup for the pneumatic
system will be much simpler. This is important, seeing as we need to limit the
initial setup time as much as possible to increase the time that can be spent on
improving the system. Seeing as a simpler setup decreases setup time, we can
say that the Pneumatic system has the clear advantage.
4.2.8 Final DecisionAfter going through all four forms of feasibility assessment, we can
conclude that the pneumatic setup is the best for what our project.
4.3 The Micro-controllerThis design component was a very important component in the design.
This component would keep track of all the actions and, depending on state of
the system, would tell the system to perform certain actions. We have taken an
account of four commonly used micro-controllers for our Pugh Analysis.
4.3.1 Micro-controller Concepts: Baseline Concept: PIC Micro
A. 8051
B. Basic stamp-2
C. DIOS
4.3.2 Attributes:1. Input/outputs
2. Applicability of code
3. Compatibility with PC
4. Ease of programming
5. Clock speed
6. Memory
7. Voltage requirement (small?)
37
TRANSCAT Pipette Calibration Project 05417
8. Documentation on controller
9. Knowledge of controller
10.Cost
11.Cost of accessories
4.3.3 Level of Attainment Analysis
Attribute A B C
1 - - 0
2 0 - 0
3 0 0 +
4 0 + +
5 + 0 +
6 0 0 -
7 0 0 0
8 0 0 +
9 - + 0
10 0 - -
11 0 - +
Table 8 Pugh Analysis of Micro controller: Baseline = PIC Micro
As one can see from the Pugh Analysis in Table 8, the DIOS micro-
controller was the winning concept overall. Also, one can see that the 8051 and
PicMicro micro-controllers were about equal. To further understand why each
concept was feasible or not, we performed the four types of Feasibility
Assessment.
4.3.4 Performance FeasibilityWhen it comes to interfacing with a PC, all of the micro-controllers start off
with an equal footing. However, certain micro-controllers, such as the 8051 and
38
TRANSCAT Pipette Calibration Project 05417
the PicMicro, need to have a serial communications port added on to a custom-
made breadboard in order to properly communicate with the computer. The Basic
Stamp and DIOS, however, both have readily available and in-stock breadboards
that have the capability for serial communication already built in. However, the
DIOS controller board has the unique advantage of direct solenoid/servo control
from the board.
When it comes to the amount of I/O pins, the more the better, seeing as
more pins means more commands that can be sent out. The PicMicro and DIOS
both have up to 40 pins, around 32 of which are programmable. The other
concepts have less total pins and, therefore, less I/O pins. Thus, the team found
that the DIOS had the largest advantage in this form of feasibility.
4.3.5 Economic FeasibilityWhen it comes to the price of the controller alone, the 8051 Micro
controllers are much cheaper than Basic Stamp or DIOS micro-controllers. In
fact, their prices are comparable with PIC micro controller. Still, this is not the
only form of cost that the system can have.
There is also the cost of accessories for the system. In this case, the 8051
and the PicMicro again have equivalent prices. The Basic Stamp, being a
Parralax product, has very expensive accessories when compared to the other
micro-controllers. Finally, the cost of accessories for the DIOS micro-controller is
relatively cheap when compared to the others. Seeing as we will have to add a
fair number of accessories to the micro-controller, we can say that the DIOS,
8051 and PicMicro are on almost equivalent footing.
4.3.6 Technical FeasibilityOne of the first attributes that come to mind when thinking of Technical
Feasibility would definitely be the ease of programming. Seeing as the team
leader has had extensive experience in programming systems with the Basic
Stamp, the team decided that this would be one of the easiest micro-controllers
to program. Still, the DIOS micro-controller also works with the Basic language.
39
TRANSCAT Pipette Calibration Project 05417
Therefore, the DIOS micro-controller could also easily be programmed. The other
micro-controllers have a base language of Assembly, a language of which the
Electrical Engineers had a good amount of experience, but not much on system-
level programming.
Another attribute to consider would be the amount of documentation
available for the micro-controller. All of the micro-controllers have some
documentation, including books in the library. Still, the DIOS has an extensive
web community behind it that is willing to give support to DIOS users free of
charge. Therefore, one could say that, all in all, the DIOS micro-controller has the
advantage.
4.3.7 Schedule FeasibilityAs was said before, both the DIOS and the Basic Stamp had readily made
carrier boards that could do a majority of the things we wanted to do. Seeing as,
for the 8051 and the PicMicro, the team would have to modify a breadboard with
all sorts of components, each which would have to be purchased with a lead
time, one can easily see that the time for setting up the controllers is much longer
for the 8051 and the PicMicro.
When it comes to programming time, the DIOS and Basic Stamp have
very simple commands and the team leader knows most of the commands that
would need to be used in any functionality. Seeing as not all the commands were
known for the 8051 and the PicMicro, the team would have to lose valuable time
learning how to program the controller. Thus, in this form of assessment, the
DIOS and the Basic Stamp are tied.
4.3.8 Final DecisionAs one can see in the four forms of feasibility assessment, the DIOS is the
clear winner for the micro-controller.
40
TRANSCAT Pipette Calibration Project 05417
4.4 Sensors for Vertical TrackingIn order to move the pipette from one place to another, one needs to clear
any objects that lie in its path. In order to keep track of whether the pipette is high
enough or not, one needs to install sensors. In the case of pneumatics, there are
a few sensors that are accepted as the industry standards. We chose these
sensors for the feasibility assessment.
4.4.1 Concepts for Vertical Tracking:Baseline Concept: Hall Effect
A. Reed Switch
B. Magneto Resistive Device
4.4.2 Vertical Tracking Attributes:1. Send signal to controller (feedback)
2. Easily mounted
3. Cost
4. Resolution/accuracy
5. Voltage/current needs
6. Simplicity of use
7. Knowledge of sensor
4.4.3 Level of Attainment Analysis
Concept A B
1 0 -
2 0 -
3 + -
4 - -
5 0 0
6 + -
7 0 -
41
TRANSCAT Pipette Calibration Project 05417
Table 9 Pugh Analysis of Vertical Tracking Sensors: Baseline = Hall Effect
By looking at Table 9, one would think that the reed switch would be the
most feasible concept. This will change, however, once we go further in-depth in
the feasibility assessment.
4.4.4 Performance FeasibilityWhen it comes to performance, the reed switch has a clear disadvantage.
The switch activates when a strong enough magnetic field is about ¼ of an inch
away and deactivates when the field goes about ½ of an inch away. This is not
very good resolution at all, especially seeing as how the pipette requires a higher
level of accuracy due to the need for the tip to go exactly into the fluid and not go
too high or too low. Thus, the reed switch is not very viable.
The Hall effect sensor and the Magneto-resistive Device both have a
better accuracy than the reed switch, but the two devices operate in different
ways. The hall effect sensor senses if a magnetic field of a certain predetermined
strength is present or not. If the field is any weaker than the predetermined
strength, the sensor will not pick it up. This makes it so the sensor doesn’t pick
up the magnet until it is right next to the sensor.
The Magneto-resistive Device, on the other hand, is not a proximity sensor
at all. This device measures the change in a magnetic field. Thus, it can keep
track of how far away a magnet is and how fast it is moving. The main problem is
that of the speed of the signal. When it sends out a signal, it takes a while to go
through its entire range and for the device to interpret this signal. Thus, the
resolution of position is not that good for the Magneto-resistive Device. Thus, one
can say that the Hall Effect sensor would be most feasible in this setup.
4.4.5 Economic FeasibilityAll of the sensors are not very expensive. In addition, they can be
mounted and attached to the micro-controller in almost the same way. Therefore,
one could say that, unless the budget for the project was to be near its limit, one
42
TRANSCAT Pipette Calibration Project 05417
could choose either option at an almost equivalent economic feasibility. If one
had to choose an option that was most feasible, however, the reed switch would
be the obvious choice, as one can see from the Pugh Analysis.
4.4.6 Technical and Schedule FeasibilitySeeing as most of the team has worked with basic sensors and controls,
none of these sensors should be outside of the expertise of the team members.
However, the least easily set up option would be the Magneto-resistive Device.
This is due to the fact that, instead of sending out a simple binary signal for the
user to read, the user has to make some sort of setup to utilize the change in
resistance in the MRD. One of the most common ways to do this is to create an
RC circuit by using a common capacitor in conjunction with the MRD. Then, one
can calculate the RC time constant, find several known position for certain time
constants, and then linearly interpolate the data to find the positions for the RC
time constants that one receives from the circuit. As one can see, this setup
takes much more effort than the others, so this is the least feasible option, both in
expertise and time taken to set it up.
4.4.7 Final Decision
After going through all four forms of feasibility assessment, we see that
there is a tie between the Reed Switch and the Hall Effect Sensor as per how
many forms of feasibility they were best in. However, for our project, the
performance is the most important aspect of the design, at least as sensors go.
Thus, one can say that the Hall Effect Sensor is the most feasible option for our
setup.
43
TRANSCAT Pipette Calibration Project 05417
4.5 Radial Movement SensorsThis component keeps track of the turning of the stand. Depending on the
readout from the sensor, one can see if the pipette is over the vial or over the
calibrator. From this, the micro-controller can tell the system to perform a certain
set of actions that is suitable for this position. Therefore, we had to be careful
when we performed the Pugh Analysis shown in Table 10.
4.5.1 Concepts for radial tracking:
Baseline Concept: Encoder disk and optical interrupter
A. IRPD (Infrared Proximity Device/Detector)
B. Sonar
C.Potentiometer
D.Hair trigger switch
4.5.2 Attributes:1. Cost
2. Accuracy/resolution
3. Feedback
4. Ease of mounting
5. Simplicity of use
6. Knowledge of sensor
4.5.3 Level of Attainment Analysis
Attribute A B C D
1 + - 0 +
2 - - 0 -
3 0 - + 0
4 + - + +
5 0 - - +
6 0 0 - 0
Table 10 Pugh Analysis of Sensors: Baseline = Optical Interrupter Setup
44
TRANSCAT Pipette Calibration Project 05417
According to this analysis, the hair trigger switch is the winner. Still, as
was shown in the last feasibility assessment, this does not always show what is
the true winner. Therefore, we performed the four types of feasibility assessment
again.
4.5.4 Performance FeasibilityThe IRPD and hair trigger switches both share a common problem: that of
resolution. For the IRPD, the range that it covers is about two inches, with much
variability that can occur with the sensing. The hair trigger sensor also has a two-
inch range, but the range is confined to the trigger on the switch. Seeing as the
vials are less than an inch in diameter apiece, these ranges in sensing are not
acceptable. Therefore, if the sensor is triggered at a distance of 1 ½ inches
instead of the possible 2 inch range, the tip of the pipette would likely miss its
mark.
When it comes to sonar, there is one key problem. This problem is the fact
that once an object gets within the range of about one inch, the sonar sensor no
longer notices it. This is not suitable for our system, seeing as our design would
need to have the sensor close to the object to save space in the design.
The potentiometer and the optical interrupter both have very accurate
sensing apparatus. The resistive film in the potentiometer lets one track the
position very well with RC Time Constants and linear interpolation. The optical
interrupter’s encoder disk lets one instantly know when one is at a reference
point. Either way, they both do the job well.
4.5.5 Economic FeasibilityAs one could see from the Pugh Analysis, the cheapest of the sensors are
the IRPD and the hair trigger sensors. To be specific, the hair-trigger sensor is
the least expensive of them all. When it comes to the IRPD, however, it is not
much less expensive than the optical interrupter setup.
45
TRANSCAT Pipette Calibration Project 05417
As per the sonar, not only is it the most expensive, but experience has
shown that they are very temperamental and tend to need to be replaced
regularly. Thus, this is far from a good choice for our system.
4.5.6 Technical FeasibilityWhen it comes to simplicity of use, one can’t get more simplistic than the
switch. After all, it is purely a physical touch sensor with binary output. The IRPD
is in the same vein; only it is a light wave sensor.
The sonar can be controlled with the proper conversion factor, either to
inches or mm. Still, as was said before, the sensor is quite temperamental, so
use of this sensor tends to give the user improper data that must be adjusted
each time a new type of error is introduced.
When it comes to the potentiometer, one has to figure out the RC Time
constants for certain positions and these Time constants change as the
capacitors either wear out or are replaced. Thus, it is fairly hard to keep track of.
The optical interrupter has a physical encoder disk that has lines etched
into it that are fixed in place. Thus, once one has mounted the encoder disk, one
no longer has to worry about adjusting for position error.
4.5.7 Schedule FeasibilityAll of the parts have to be mounted. The fastest mountings will be for the
IRPD and the switch. The second fastest would likely be sonar, seeing as one
also merely needs to mount it to be in place, but one also has to take time to find
the RC Time Constants. Third would be the potentiometer, which one has to
mount the shaft of directly to the cylinder. Lastly, the slowest to mount would be
the optical encoder, seeing as one has to mount the encoder disk to the cylinder
and then mount the mouth of the optical encoder around the disk.
When it comes to time spent correcting the errors in the system, the least
time would be spent on the optical encoder, seeing as it has very little ability to
have variance in it. The other sensors, however, would have plenty of time spent
46
TRANSCAT Pipette Calibration Project 05417
correcting the errors, perhaps even more than the time saved on mounting. Thus,
the team decided that the optical encoder would be the most feasible.
4.5.8 Final Decision
After looking through all forms of feasibility assessment, we found that the
optical encoder setup was the most feasible.
47
TRANSCAT Pipette Calibration Project 05417
5 Specifications, Analysis and Synthesis5.1 Mechanical Analysis & Synthesis5.1.1 Problem Statement
During the calibration of a pipette there are many factors that affect the
precise calibration of the pipette. Our goal is to minimize the factors that are
causing the error in the calibration.
5.1.2 Known InformationThe colorimetric or photometric method involves the analysis of volumes of diluted dye
in a cell of known path length. According to the Beer- Lambert Relationship,if a beam of
monochromatic light passes through homogeneous solutions of equal pathlength,the
absorbance measured is proportional to the dye concentration.So,with this in mind, an
unknown volume of dye can be pipetted into a known volume of diluent,the resulting dye
concentration can be measured photometrically, and the volume can be calculated.
There are many techniques and tips available that will optimize pipetting performance
and increase the reproducibility of results.
1. The equipment
1. Tips – It is advocated that only high quality tips, which optimize the pipette’s
performance, be used. A high quality tip is one that has a smooth uniform interior
with straight even sides that prevents the retention of liquids and minimizes
surface wetting. Also, the tip should have a clean, hydrophobic surface and a
perfectly centered opening in order to ensure the complete dispensing of the
sample.
48
TRANSCAT Pipette Calibration Project 05417
2. Liquid viscosity – since the pipette was originally factory calibrated using water,
any liquid that has viscosity higher or lower than water will impact the volume
dispensed.
2. The operator
1. Technique
a) Position – Pipette should be held vertical during the aspiration of liquids.
Holding a pipette 30deg. off vertical can cause as much as 0.7% more
liquid to be aspirated due to the impact of hydrostatic pressure.
b) Pre – Wetting /Pre – Rinsing Tips- Failing to pre-wet tips can cause
inconsistency between samples since liquid in the initial samples adhere to
the inside surfaces of the pipette tip, but liquid from later samples does
not. Also, if a new volume is dialed in on pipette’s micrometer, better
results will receive at the new volume by taking the old tip off and placing
a new one on the shaft before commence pipetting.
c) Release of Plunger – Releasing the plunger abruptly can cause liquid to
be bumped inside the pipette during a liquid transfer application. This can
cause liquid to accumulate inside the instrument which in turn can be
transferred to other samples causing variability in sample volume and the
potential for cross contamination.
d) Immersion Depth- The pipette tip should only be inserted into the vessel
containing the liquid to be transferred about 1-3mm. If the tip is immersed
beyond this, the results could be erroneously high.
49
TRANSCAT Pipette Calibration Project 05417
e) Thermal conductance- Thermal energy can be transferred from the
operator’s hand to the air within the pipette (dead air) or even to the
internal components themselves. This can have a dramatic impact on the
amount of liquid dispensed due to effects of expansion and/ or contraction.
To lessen this effect, it is recommended that some type of thermally
insulated gloves like latex of cloth be worn.
2. Pipette micrometer setting – It is important to avoid significantly over dialing or
under dialing the recommended range of pipette. Volume delivery performance
may change radically and may become completely undefined.
3. The Environment
a) Temperature – The volume delivery performance specifications of
pipettes have been referenced by most manufactures at room temperature
which is defined as (20- 25) deg.Celsius.Any deviation from this
specification can affect the amount of liquid dispensed due to the
expansion or contraction of the internal components.
b) Barometric Pressure – Pressure is reduced by 1.06” Hg for every 1000’
of elevation, however, barometric pressure has only a small effect on the
density formula, so the error encountered is not correcting for elevation is
often ignored.
c) Relative humidity – This is the percentage of moisture in the air at a
measured dry bulb temperature compared to the amount of moisture that
the air can hold at that temperature if the air is 100%saturated. Under dry
conditions, which are defined as less than 30% RH, it is extremely
50
TRANSCAT Pipette Calibration Project 05417
difficult to ensure an accurate measurement due to the rapid evaporation
rate. Conversely, excessive humidity, which is defined as greater than
75%, can cause a measurement to be erroneously high due to
condensation. Therefore, generally accepted guidelines for pipette volume
delivery specify that relative humidity be maintained within the range of
45%-75%.
Concerning the operator, there are many factors in proper pipette techniques
that play a major role in the calibration process. Factors that the operator have
control of include the position of the pipette, pre-wetting the tips, release of the
plunger, immersion depth, and the thermal conductance.
The environment is also a factor that affects the calibration of the pipette
and such factors include vibration, evaporation, and temperature. All materials
involved in the photometric must be at the same temperature as not to get any
temperature gradients. Let the dye solutions, spectrophotometer, pipette, tips,
etc. reach thermal equilibrium at room temperature overnight if possible to make
results more accurate. Vibration creates air within the solution and can affect the
calibration; therefore vibration is a concern of the system.
5.1.3 Desired InformationDuring the pipette calibration the goal is to figure out the largest error factor
and then once that is taken care of work the way down as to minimize all errors
and make the calibration more precise. The largest error factor right now is the
“speed” at which the plunger is actuated. Some information needed is the
amount of force to actuated the plunger at first stroke and clean out stroke.
5.1.4 AssumptionsThere are a few assumptions that need to be listed within the system. The
thermal equilibrium concern should be taken care of and by this one is to say
there is no need for a controlled environment for the temperature of the system.
The materials of the calibration should all be in the lab for at least a work day to
make sure the equipment is all at thermal equilibrium. The system can also
51
TRANSCAT Pipette Calibration Project 05417
ignore any hydrostatic pressures caused when the tip of the pipette is placed into
the test fluid. The pressure created during this process is so minimal that it can
be neglected. Another important assumption is that there is no reason to concern
the system with any evaporation concerns. The test vial will be very close to the
calibration device and therefore with the laboratory environment there should be
no issue of evaporation. The last assumption that is that the air supply in the lab
is going to be no less then fifty pounds per square inch, as to design the air
cylinders bore size around the supplied air pressure and the force required.
5.1.5 Data and AnalysisThe problem with the pipette is the human interference and therefore if
that can be diminished then the error may be able to also be diminished. In the
tables below the experiment was held using different types of strategies to
operate the pipette.
52
TRANSCAT Pipette Calibration Project 05417
Pipette Calibration Result
Oxford
Pipette
Range 10 - 100 uL
SN# 014756
8885-
500945 Average for 80 uLAccuracy +/-0.5 - +/-1uL 81.78571429
http://www.novamedpipcal.com/OxfordVar.htm Average for 10uL11.594
Reading (uL) Suck in/ Dispense/ Angle Actual Reading (uL) % Error Percent Error with
respect to averages
80 slow/slow/vertical 81.5 1.875 0.349344978
80 slow/fast/vertical 82 2.5 0.262008734
80 fast/fast/vertical 81.7 2.125 0.104803493
80 slow/slow/angle 81.3 1.625 0.593886463
80 slow/fast/angle 82.1 2.625 0.384279476
80 fast/slow/angle 81.9 2.375 0.139737991
80 fast/fast/angle 82 2.5 0.262008734
10 slow/slow/vertical 11.74 17.4 1.259272037
10 fast/slow/vertical 11.52 15.2 0.63826117
10 slow/fast/vertical 11.55 15.5 0.379506641
10 fast/fast/vertical 11.72 17.2 1.086769018
10 slow/slow/angle 11.44 14.4 1.328273245
Comment: All reading were off by 1-2 uL
The pipette was out of tolerance.
Varying the technique did not seem to matter at all or where dispense the fluid at an
angle.
Table 11 Tests for the Oxford Brand Pipettes
53
TRANSCAT Pipette Calibration Project 05417
Fisherbrand pipette Average for 100 uLSN# N30775 99.725
Range 20 - 200 uL Average for 45 uLAccuracy 44.5075
Average for 20 uL19.7075
Reading (uL) Suck in/ Dispense/ Angle Actual Reading (uL) % Error Percent Error based on averages
100 slow/slow/vertical 99.4 0.6 0.325896215
100 slow/slow/vertical 99.8 0.2 0.075206819
100 fast/fast/vertical 100.1 0.1 0.376034094
100 slow/fast/vertical 99.6 0.4 0.125344698
45 slow/slow/vertical 44.26 1.64 0.556086053
45 slow/slow/vertical 44.49 1.13 0.039319216
45 fast/fast/vertical 44.65 0.78 0.320170758
45 slow/fast/vertical 44.63 0.82 0.275234511
20 slow/slow/vertical 19.66 1.7 0.24102499
20 slow/slow/vertical 19.72 1.4 0.063427629
20 fast/fast/vertical 19.84 0.8 0.672332868
20 slow/fast/vertical 19.61 1.95 0.494735507
Comment: No need to put the tip of the pipette at an angle, it
doesn’t make a difference. Readings were very close to expected
value.
Table 12 Tests for the Fisherbrand Pipette
As seen in Tables 11 and 12, there is much difference when the pipette is
actuated and released in different manners. In Table 13, the different pipettes
were tested to examine what amount of force was needed to actuate both levels
of the plunger.
54
TRANSCAT Pipette Calibration Project 05417
PipetteVolumetric
Setting Level 1 Level 2
Fisherbrand
(0-200) 20 uL 830g 2.4kg
50 uL 900g 2.4kg
100 uL 900g 2.5kg
150 uL 900g 2.4kg
200 uL 900g 2.4kg
Ranin (0-
20) 5 uL 900g 3.8kg
10 uL 700g 4kg
15 uL 700g 4kg
Oxford
(0-100) 10 uL 700g 2.8kg
50 uL 700g 2.5kg
100 uL 700g 2.6kg
Table 13 Force Requirements
The level one actuation is the amount of force used to inject the tip with
the appropriate amount of fluid and then eject the fluid. The level two is the clean
out stroke that is done at the time when all the substance is out of the tip and this
is just to ensure that all of the fluid is out of the tip. When the force was converted
to pounds of force it was calculated that approximately two pounds of force would
be required to actuate level one and to make sure that the entire clean out stroke
was actuated a force of eleven pounds will be used. The forces in Table 13 were
found using calibrated weights that were set on top of the pipette plunger and
with gravity constant after the suitable weight was on the pipette to actuate it; that
weight was the required force. Since the pipette requires two different forces to
run the calibration it would be a difficult process to accomplish with something
like a stepper motor. The first problem with using a stepper motor is that not only
would you need to have an output back to the motor to relay the force applied;
but as the volume of the pipette changes so does the range of motion of the first
stage actuation. This concludes that the best solution for the system is a
pneumatic one because not only can the pneumatics take care of the two
55
TRANSCAT Pipette Calibration Project 05417
different forces but it can also handle the different ranges of travel. The next
objective was to figure out what size pneumatic cylinder would be needed to
operate the pipette. With the assumption that the air supply would be no less
then 50 pounds per square inch, the equation below was used to find that a ¾”
bore cylinder would be appropriate for this application.
Clippard catalog series UDR-12 double acting cylinders will be appropriate
for both pneumatic cylinders. For the cylinder that is going to actuate the plunger
(cylinder #1) the range of actuation needs to be maximum distance of 1.14”
therefore the designed cylinder will have an actuation length of 2”. The cylinder
that will control the vertical movement of the system (cylinder #2) will have an
actuation length of 6” this will account for clearance of the lid of the calibration
device and capable of insertion of pipette tip into the test dye. According to
Clippard, the ¾” bore cylinders have a ¼” rod, which, also according to Clippard,
will handle a load of 190 pounds at a length of 5” until buckling occurs. The rod of
cylinder 2 will actuate to a length of 6” the buckling load is still far greater then
the applied load that it will experience. The maximum weight experienced by the
rods is nowhere in the range of 190 pounds.
To control the pneumatic cylinders there will be pneumatic valves with
regulators and speed control on each valve or inline with each valve. The speed
control is important to minimize vibration with control of the vertical motion and
most important to the actuation speed of the plunger. For this application the
system will use adjustable flow control valves that will be inline meter out needle
valves. The MFC series from Clippard is a good choice for the needle control
valves. There will be three different air pressures required in the system make
this possible there will have to be three air regulators that control each of the
pressures to the valves. The three regulators that are needed are all at different
ranges: 0 – 50 psi for the cylinder #2, 0 – 10 psi for cylinder #1 (first actuation),
and 0 – 30 psi for cylinder #1 (second actuation). The MAR series from the
Clippard catalog will work properly for all of these applications. The valves that
56
TRANSCAT Pipette Calibration Project 05417
will be electrically controlled are yet to be determined because the team may
have the valves donated to them, so the rest of the design may be completed
around these valves. For some specifications the valves will be controlled by 0 -
24 VDC and will have a manual actuation so it can be operated without the
electrical system.
5.1.6 Quality Review
After reading through this, I found that the report contains all of the facts
that we had covered over the quarter in a concise manner. Still, this contains
mostly information pertaining to the plunger actuation piston. If one were to think
of the piston that acts as a stand, one would have to take into effect the weight of
the setup, or:
You could then apply the force equation again, this time on the larger piston in
order to see if the pressure would be enough. I know from my experience with
the representatives from Roessler that the piston should be able to support up to
6 pounds if the cylinder rises 4-8 inches, which is enough to clear the colorimetric
calibrator. Speaking of which, we might want to add in the fact that the
colorimetric calibrator is 5 inches high by itself and 8 inches high with an opened
lid. Besides this, the analysis seems to be okay from a mechanical standpoint.
57
TRANSCAT Pipette Calibration Project 05417
5.2 Electrical Analysis5.2.1 Design Specifications:
The objective of this project is to fulfill all the requirements set forth by
Transcat. The design is to reduce variability and error induced by human
interaction with the pipettes and the calibration process.
5.2.2 Micro-controller Specifications: One RS232 serial port for PC communication (GUI)
At least 8 I/O ports
Low memory requirement (RAM size: 32 Bytes, EEPROM Size: 2K
Bytes)
Ease of programming (BASIC)
5.2.3 Sensor Specifications: 5V Logic
Piston sensors must be able to accurately sense piston position
Base sensors must be able to sense rotational position
Sensor for cover must be able to determine if cover is open or
closed
5.2.4 Power Supply Specifications: Must supply 5VDC and 24VDC
Must use 120VAC 60Hz single phase power
Must be able to handle a maximum current draw of 1A as valves runs on 1W
power.
5.2.5 Micro-controller5.2.5.1 Selection of Micro-controller
The Basic Stamp was chosen for this project due to its ease of use and
cost. The main constraint of the micro-controller is that we need RS232 serial
communication to a PC as well as at least eight input/output lines.
58
TRANSCAT Pipette Calibration Project 05417
Basic stamp 2 Module specs are:
24 pin DIP
Microcontroller: PIC16C57
Processor speed: 20 MHz
Program Execution Speed: 4000 instructions/sec.
RAM size: 32 Bytes (6I/O, 26 Variable)
EEPROM (Program) Size: 2K Bytes, 500 instructions
No. Of I/O pins: 16+ 2 Dedicated Serial
Voltage Requirements: 5-15 V
Current Draw @ 5V: 3mA Run /50uA Sleep
Source/Sink Current per I/0:20mA/25mA
Source/Sink Current per unit: 40mA/50mA per 8I/O pins
PBASIC Commands: 42
PC Programming Interface: Serial Port (9600 baud)
Windows Text Editor: Stampw.exe (v1.04 and up)
5.2.5.2 Integration with Electrical System
The electrical system currently has five inputs and three outputs. The
inputs are in the form of sensors. The three outputs are solenoid valves to
control air flow. For the micro-controller to work in this situation, we will need to
make a simple interface to make sure the sensors are on or off, with simply a 5V
or 0V digital input to the micro-controller. A different interface is required for the
output, as the solenoids valves require 24VDC and approximately 100mA,
sometimes up to 200mA when they actively switch. Since the micro-controller
can only provide 5VDC at around 20mA to an output pin, we will need to have a
power FET or a driver to provide the necessary power to the valves. Also
needed is a simple interface for RS232 communication with the PC. This serial
communication will provide the initialization for the micro-controller as well as
feedback and instructions to the user. The Stamp 2 Board has a built in RS232
59
TRANSCAT Pipette Calibration Project 05417
interface, as well as room for an EEPROM, if additional memory is required as
the project progresses.
5.2.5.3 Inputs: Piston Up
Piston Down
Dye Position Sensor
Calibrator Position Sensor
Calibrator Cover Sensor
5.2.5.4 Outputs: Piston Up/Down
Stage 1 Pipette Activation
Stage 2 Pipette Activation
5.2.5.5 Serial: Receive Command to Begin Run from user
Send feedback messages to user
(i.e. “Please rotate the pipette to the Dye Position”)
Query user after each run if the wish to run again or end run
60
TRANSCAT Pipette Calibration Project 05417
6 Bill of Materials
Part QtyUnit Price
Total Price
DIOS Micro-controller 40 Pin 1 $24.95 $24.95DIOS Carrier Board #4 1 $39.95 $39.95DIOS Compiler 1 $0.00 $0.00TI SN754410 Solenoid Driver 1 $4.49 $4.49Power Supply Radio Shack 1 $17.99 $17.99Misc. Electronics 1 $10.00 $10.00Magnetic Proximity Sensor 1 $2.00 $2.00Hall Effect Sensor 2 $4.00 $8.00Optical Interrupter Sensor 1 $8.00 $8.00Adjustable Air Regulator 3 $5.00 $15.00Adjustable Flow Control 4 $32.30 $129.20Pneumatic Cylinder 7/8" Bore 1 $50.00 $50.00Pneumatic Cylinder 1 $30.00 $30.00Solenoid Valve 2-way 2 $18.00 $36.00Solenoid Valve 3-way 1 $25.00 $25.00Raw MaterialsShipping Costs 1 $30.00 $30.00Total $420.58
61
TRANSCAT Pipette Calibration Project 05417
Appendix A: System Schematic
62
TRANSCAT Pipette Calibration Project 05417
Appendix B: Electrical System
63
TRANSCAT Pipette Calibration Project 05417
Appendix C: Flow Chart of System
64
TRANSCAT Pipette Calibration Project 05417
65
TRANSCAT Pipette Calibration Project 05417
66
TRANSCAT Pipette Calibration Project 05417
Appendix D: Beer-Lambert’s Law
An unknown concentration can be found using a known absorbance of the
liquid and applying Beer's law. The Beer-Lambert law (or Beer's law) is the linear
relationship between absorbance and concentration of an absorbing species.
The general Beer-Lambert law is usually written as:
c = A/(a(lambda) * b)
Where a(lambda) is a wavelength-dependent absorptivity coefficient, b is
the path length, and c is the liquid concentration. There is also a blank of known
concentration and volume in the colorimeter. The system then uses the following
equation to find the unknown volume:
Concentration (test) x Volume (test) = Concentration (blank) x Volume (blank)
Appendix E: Hall Effect
The sensors used to detect the vertical position of the cylinder are called
Hall Effect sensors. The Hall effect is a phenomena of a voltage created
transversely to the current applied in a conductor, when there is a magnetic field
perpendicular to the current flow. The sensors respond to a presence or
interruption of a magnetic field by producing either an analog or digital output
proportional to the strength of the field.
67
TRANSCAT Pipette Calibration Project 05417
Appendix F: STAMP 2 MODULE
68
TRANSCAT Pipette Calibration Project 05417
Appendix E: Fishbone Diagram
69
Error in Calibration Reading
Other Equipments
Environment
Calibration Method
Other Factors
The operator Pipette
Liquid viscosity
Tips Calibration with different
angles
Not smooth motion
Calibration with variable speed
Direct contact of sunlight
Pipette micrometer
setting
Pre – Wetting / Pre – Rinsing Tips
Release of Plunger
Immersion Depth
Position
Humidity
Barometric Pressure
Temperature