Jawaharlal Nehru Engineering College, Aurangabad.

Electrical Engineering Department

Laboratory Manual

ELECTRICAL MEASUREMENT & INSTRUMENTATION

For Second Year (EEP) Students

Manual made by-

Prof. B.A.Patil

Page 2: Jawaharlal Nehru Engineering College, Aurangabad

FORWARD

It is my great pleasure to present this laboratory manual for second year engineering students for

the subject of Electrical Machine & Instrumentation. Keeping in view the vast coverage required

for visualization of concepts of Electrical Machines & Instrumentation components. With simple

language.

As a student, many of you may be wondering with some of the questions in your mind regarding

the subject and exactly what has been tried is to answer through this manual.

Faculty members are also advised that covering these aspects in initial stage itself, will greatly

relived them in future as much of the load will be taken care by the enthusiasm energies of the

students once they are conceptually clear.

Prof. Dr. H.H. Shinde Principal

Page 3: Jawaharlal Nehru Engineering College, Aurangabad

LABORATORY MANNUAL CONTENTS

This manual is intended for the second year students of Electrical Electronics & Power

Engineering branch in the subject of Electrical Measurement & Techniques. This manual

typically contains practical/Lab Sessions related with electrical measurement & techniques

covering various aspects of the subject to enhanced understanding.

Although, as per the syllabus, only descriptive treatment is prescribed, we have made the efforts

to cover various aspects of electrical measurement & techniques subject covering types of

different electrical instruments, their operating principals, their characteristics and different

bridges, Industrial panel meters & oscilloscope will be complete in itself to make it meaningful,

elaborative understandable concepts and conceptual visualization.

Students are advised to thoroughly go through this manual rather than only topics mentioned in

the syllabus as practical aspects are the key to understanding and conceptual visualization of

theoretical aspects covered in the books.

Good Luck for your Enjoyable Laboratory Sessions.

Prof. B.A.Patil

Page 4: Jawaharlal Nehru Engineering College, Aurangabad

SUBJECT INDEX

1. Do’s and Don’ts

2. Lab exercise:-

1) Earth Resistance Measurement

2) To study CRO

3) To study measurement of capacitance

4) Kelvin’s Double Bridge

5) Measurement of Power in three phase circuit using two wattmeter method

6) To study ammeter and voltmeter

7) To study transducer

8) LVDT Characteristics

9) To study measurement of self inductance (Anderson Bridge)

10) Temperature Measurement using Thermocouple

11) Calibration of Voltmeter and Ammeter by Potentiometer

12) Determination of unknown inductance using Maxwell’s inductance bridge method

Page 5: Jawaharlal Nehru Engineering College, Aurangabad

DOs and DONDOs and DONDOs and DONDOs and DON·T DOs in Laboratory: T DOs in Laboratory: T DOs in Laboratory: T DOs in Laboratory:

1. Understand the equipment to be tested and apparatus to be used.

2. Select proper type (i.e. A. c. or D. C.) and range of meters.

3. Do not touch the live terminals.

4. Use suitable wires (type and size).

5. All the connection should be tight.

1. Do not leave loose wires (i.e. wires not connected).

2. Get the connection checked before switching ‘ON· the supply.

3. Never exceed the permissible values of current, voltage, and / or speed of any machine,

apparatus, wire, load, etc.

4. Switch ON or OFF the load gradually and not suddenly.

5. Strictly observe the instructions given by the teacher/Lab Instructor

Instruction for Laboratory Teachers: Instruction for Laboratory Teachers: Instruction for Laboratory Teachers: Instruction for Laboratory Teachers:

1. Submission related to whatever lab work has been completed should be done during the next

lab session. The immediate arrangements for printouts related to submission on the day of

practical assignments.

2. Students should be taught for taking the observations /readings of different measuring

instruments under the able observation of lab teacher.

3. The promptness of submission should be encouraged by way of marking and evaluation

patterns that will benefit the sincere students.

Page 6: Jawaharlal Nehru Engineering College, Aurangabad

2. Lab Exercises:

Exercise No1: (2 Hours) – 1 Practical

Earth Resistance Measurement

Aim:- To study earthing in electrical power system and to measure earth resistance using the

Earth Tester.

Experimental setup: - 1. Earth Tester (megger) -1 no.

2. Spikes --2 No’s

Theory:-

1. Meaning of Earthing ?

2. Importance of Earthing.

3. Earth Electrode.

4. Environmental effects on Earth resistance.

5. Earth tester.

Circuit Diagram:-

Page 7: Jawaharlal Nehru Engineering College, Aurangabad

Procedure:-

1) Put the two spikes acting as current & potential electrode in to the ground at a distance of 25

m & 12.5 m from earth electrode under test.

1) Connect the two spikes to C2 & P2 terminals respectively.

2) Short the P1 & C1 terminals of motor & connect it to the earth electrode under test.

3) Place the megger on horizontal firm stud.

4) Turn the handle of megger to speed slightly higher than rated speed & note down the

deflection of the needle.

5) Take down the 3 to 4 readings by keeping the distance same and placing the electrodes at the

other positions.

6) Take the average of these readings which is equal to earth resistance.

Conclusion:-

The value of earth resistance is ------------- Ω.

Page 8: Jawaharlal Nehru Engineering College, Aurangabad

Exercise No 2: (2 Hours) – 1 Practical

Study of CRO

Aim:- Measurement of voltage, frequency and phase using CRO, Measurement of frequency by

Lissijous method.

Experiment setup: - CRO, function generator, probes.

Circuit diagram:-

Fig. 1: Functional block of a simple CRO

Typical Specifications:-

VERTICAL DEFLECTION:-

Bandwidth (-3dB): d.c. to 20MHz ( 2Hz to 20KHz on a.c.)

Sensitivity: 2mV/cm to 10V/cm

Accuracy: ±3%

Input Impedance: 1M /28pf approx.

Input Coupling: DC-GND-AC

Page 9: Jawaharlal Nehru Engineering College, Aurangabad

Input Protection: 400V d.c. or pk a.c.

HORIZONTAL DEFLECTION:-

Time base: 0.5us/cm to 0.2us/cm, 18 ranges

Accuracy: ±3%

ADDITIONAL FACILITIES:-

Calibrator: 1V, 2% square wave at approx. 1 KHz.

Ramp Output: Approx. ± 3.5V ramp from 5K .

SUPPLY:-

220/240V ± 10%

45 TO 65 Hz approx. 40VA.

Fig No. 2: - Simple CRO

Page 10: Jawaharlal Nehru Engineering College, Aurangabad

Theory:- Applications of Oscilloscope

I. Measurement of Voltage:-

The most direct voltage measurement made with the help of an oscilloscope is the peak to

peak (p-p) value. The rms value of the voltage can then be easily calculated from the p-p value.

To measure the voltage from the CRT display, one must observe the setting of the vertical

attenuator expressed in V/div and the peak to peak deflection of beam, i.e. the number of

divisions.

The peak value of voltage is then computed as follows.

Vp-p = (volts/div) ⋅ (no. Of div)

Fig No. 3:- Sine Waveform

II. Period and Frequency Measurement:-

The period and frequency of periodic signals are easily measured with an oscilloscope. The

waveform must be displayed such that a complete cycle is displayed on the CRT screen.

Accuracy is generally improved if a signal cycle displayed fills as much of the horizontal

distance across the screen as possible.

The period is calculated as follows.

T = (time/div) ⋅ (No. of div/cycle)

The frequency is then calculated as f = 1/T

III. Measurement of Frequency by Lissajous Method:-

Page 11: Jawaharlal Nehru Engineering College, Aurangabad

This particular pattern results when sine waves are applied simultaneously to both pairs of the

deflection plates. If one frequency is an integral multiple (harmonic) of the other, the pattern will

be stationary, and is called a lissajous figure.

In this method of measurement a standard frequency is applied to one set of deflection plates of

the CRT tube while the unknown frequency (of approximately the same amplitude) is

simultaneously applied to the other set of plates. However, the unknown frequency is presented

to the vertical plates and the known frequency (standard) to the horizontal plates. The resulting

patterns depend on the integral and phase relationship between the two frequencies. (The

horizontal signal is designated as fh and the vertical signal as fv.

Measurement Procedure:-

Set up the oscilloscope and switch off the internal sweep (change to Ext). Switch off sync

control. Connect the signal source as given in Fig.3. Set the horizontal and vertical gain control

for the desired width and height of the pattern. Keep frequency fv constant and vary frequency

fh, noting that the pattern spins in alternate directions and changes shape. The pattern stands still

whenever fv and fh are in an integral ratio (either even or odd). The fv = fh pattern stands still

and is a single circle or ellipse. When fv = 2fh, a two loop horizontal pattern is obtained as

shown in Fig. 5.

To determine the frequency from any Lissajous figure, count the number of horizontal loops in

the pattern, divide it by the number of vertical loops and multiply this quantity by fh , (known or

standard frequency).

In Fig.4 (g), there is one horizontal loop and 3 vertical loops, giving a fraction of 1/3. The

unknown frequency fv is therefore 1/3 fh. An accurately calibrated, variable frequency oscillator

will supply the horizontal search frequency for frequency measurement. For the case where the

two frequencies are equal and in phase, the pattern appears as a straight line at an angle of 45°

with the horizontal. As the phase between the two alternating signals changes, the pattern

changes cyclically, i.e. an ellipse (at 45° with the horizontal) when the phase difference is π/4, a

circle when the phase difference is π/2 and an ellipse (at 135° with horizontal) when the phase

difference is 3π/4, and a straight line pattern (at 135° with the horizontal) when the phase

difference is π radians.

Page 12: Jawaharlal Nehru Engineering College, Aurangabad

Fig No. 4: - Lissajous Pattern for Integral Frequencies

As the phase angle between the two signals changes from π to 2π radians, the pattern changes

correspondingly through the ellipse-circle-ellipse cycle to a straight line. Hence the two

frequencies, as well as the phase displacement can be compared using Lissajous figures

techniques.

When the two frequencies being compared are not equal, but are fractionally related, a more

complex stationary pattern results, whose form is dependent on the frequency ratio and the

relative phase between the two signals as in fig 5.

Page 13: Jawaharlal Nehru Engineering College, Aurangabad

Fig No.5: - Lissajous Pattern for Non-Integral Frequencies

The fractional relationship between the two frequencies is determined by counting the number of

cycles in the vertical and horizontal.

fv = (fraction) × fh

or fv = number of horizontal tangencies

fh = number of vertical tangencies

Observations:-

1. Peak to peak voltage Vp-p = (volts/div) × (no. Of div) =

2. Time period T = (time/div) × (No. of div/cycle)=

The frequency is then calculated as f = 1/T=

Page 14: Jawaharlal Nehru Engineering College, Aurangabad

3. Measurement of phase=

4. Measurement of unknown frequency by Z-Modulation=

Conclusion:-

Thus we have measured voltage, frequency and phase using CRO.

Page 15: Jawaharlal Nehru Engineering College, Aurangabad

Exercise No.3: (2 Hours) – 1 Practical

Measurement of capacitance(Schering’s Bridge).

Aim: - To measure unknown capacitance using Schering bridge.

Experiment setup:- AC source, resistor, capacitor, detector, connecting leads etc.

Theory:-

A Schering Bridge is a bridge circuit used for measuring an unknown electrical capacitance and

its dissipation factor. The dissipation factor of a capacitor is the ratio of its resistance to its

capacitive reactance. The Schering Bridge is basically a four-arm alternating-current (AC) bridge

circuit whose measurement depends on balancing the loads on its arms. Figure below shows a

diagram of the Schering Bridge.

Fig: Schering Bridge

In the Schering Bridge above, the resistance values of resistors R1 and R2 are known, while the

resistance value of resistor R3 is unknown. The capacitance values of C1 and C2 are also known,

while the capacitance of C3 is the value being measured. To measure R3 and C3, the values of

C2 and R2 are fixed, while the values of R1 and C1 are adjusted until the current through the

ammeter between points A and B becomes zero. This happens when the voltages at points A and

B are equal, in which case the bridge is said to be 'balanced'.

When the bridge is balanced, Z1/C2 = R2/Z3, where Z1 is the impedance of R1 in parallel with

C1 and Z3 is the impedance of R3 in series with C3. In an AC circuit that has a capacitor, the

Page 16: Jawaharlal Nehru Engineering College, Aurangabad

capacitor contributes a capacitive reactance to the impedance. The capacitive reactance of a

capacitor C is 1/2πfC.

As such, Z1 = R1/[2πfC1((1/2πfC1) + R1)] = R1/(1 + 2πfC1R1) while Z3 = 1/2πfC3 + R3.

Thus, when the bridge is balanced:

2πfC2R1/(1+2πfC1R1) = R2/(1/2πfC3 + R3); or

2πfC2(1/2πfC3 + R3) = (R2/R1)(1+2πfC1R1); or

C2/C3 + 2πfC2R3 = R2/R1 + 2πfC1R2.

When the bridge is balanced, the negative and positive reactive components are equal and cancel

out, so

2πfC2R3 = 2πfC1R2 or

R3 = C1R2 / C2.

Similarly, when the bridge is balanced, the purely resistive components are equal, so

C2/C3 = R2/R1 or

C3 = R1C2 / R2.

Note that the balancing of a Schering Bridge is independent of frequency.

Procedure:-

1. Connect Cx to be measured at point C.

2. Connect proper range std. capacitance Cs at point C.

3. Connect points B and D at Vin of audio detector.

4. Make power ON the unit.

5. Alternatively adjust R4 and R1 to get null condition or as minimum sound as possible.

6. Make power off

7. Measure the values of R1 and R2

By using formula calculate the value of Cx and Rx.

Cx = C3 R4/R1

Rx = R1 C4/C3

8. Dissipation factor D = wR4C4

Note: It may be observed that if Cx/Cs ratio is more, then balance condition is noted for more

than one set of R, C values.

In that case repeat steps with two different C’s values and take average reading of two.

Page 17: Jawaharlal Nehru Engineering College, Aurangabad

Observations:-

Resistance R1 =

Resistance R2 =

The value of capacitance Cx = C3 R4/R1 =

The value of resistance Rx = R1 C4/C3 =

Conclusion:-

Thus we have measured the value of unknown capacitance using Schering bridge.

Page 18: Jawaharlal Nehru Engineering College, Aurangabad

Exercise No.4: (2 Hours) – 1 Practical

Measurement of Resistance using Kelvin’s Double Bridge

Aim:- To measure unknown low resistance using Kelvin’s double bridge.

Apparatus:- Kelvin’s Bridge board……1

Null detector……………...1

Supply voltage……………1

Unknown Resistance……..1

Connecting wires…………set

Theory:-

The problem involved in the measurement of low resistance with an ordinary wheat stone

bridge can be eliminated using Kelvin’s double bridge. This bridge uses two ratio ARMS &a

four terminal low resistance standard resistor (S)

The outer ratio Arms consists of P&Q inner ratio Arms P, 2 both the arms are connected to

the potential terminals so as to eliminate the effect of leads

Under balanced condition no current flow through the galvanometer and hence the potential

drop across the resistor of an outer Arms is equal to sum of drop across the low resistance and

inner arm resistance.

Circuit Diagram:-

Page 19: Jawaharlal Nehru Engineering College, Aurangabad

Procedure:-

1) Connect the circuit as per circuit diagram.

2) Connect the unknown resistance “R”.

3) Switch on the power supply.

4) By varying the standard resistance “S” obtains the balance condition (Null deflection).

5) Calculate the unknown resistance by using the Formula.

R = × s

Precautions:-

1) Avoid loose connections.

2) Avoid parallax errors.

Observation table:-

Conclusion:-

The unknown resistance is calculating by using Kelvin’s double bridge.

Page 20: Jawaharlal Nehru Engineering College, Aurangabad

Exercise No.8: (2 Hours) – 1 Practical

LVDT CHARACTERISTICS

Aim:- To study the characteristics of LVDT using ANSHUMAN LVDT trainer kit.

Apparatus:- ANSHUMAN LVDT trainer kit,

Theory:-

The basic structure of LVDT is a movable core of a permeable material & three coils as

shown in fig. The inner core is a primary, which provides magnetic flux through its excitation by

some AC source. The two secondary coils have voltages induced do to flux linkage with the

primary. When the core is centrally located the voltage induced in each secondary is same &

when the core is displaced, the change in flux linkage causes one secondary voltage to increase

& other to decrease. The secondary windings are generally connected in series opposite so that

the voltage induced in each are out of phase with the other. In this case, as shown in the fig.2, the

output voltage is zero when the core is centrally located & increases as the core is moved in

either direction in or out.

The voltage amplitude is linear with the core displacement over some range of core travel.

Furthermore there is a phase shift as the core moves both to & from the central location.

Block diagram:-

Page 21: Jawaharlal Nehru Engineering College, Aurangabad

Procedure:-

1. Connect the PIN- D-PIN type male connector of LVDT box to the PIN D- type female

connector of LVDT signal conditioning circuit.

2. Connect the main cords of the LVDT signal conditioning board in the main supply

socket.

3. Rotate the micrometer screw such that the reading in micrometer is exactly 10 mm.

Because this is our null or zero position. The reading below the 10 mm is considered as

+ve displacement reading above the 10 mm is considered as –ve displacement.

4. For e.g. , If the reading is 7 mm then, Displacement =10 – 7= 3mm & If the reading is 14

mm, then, displacement =10 – 14= -4 mm.

5. Make the power ON.

6. Now display should show 00.00, If not,

7. If the error is above +/- 00.10 mm, then adjust the coil such that the display matches

exactly with the reading i. e. 00.00.

8. If the error is below +/- 00.10 mm, If the error is above +/- 00.10 mm, then adjust the coil

such that the display matches exactly with the reading i.e. 00.00.

9. Now rotate the micrometer such that the reading on the micrometer is displacement =

10 – 0.0 = 10.00 mm.

10. Now the reading on display must be 10.00, if not Adjust the SPAN POT of signal

conditioning circuit such that the reading in display must be 10.00. the display actually

shows the core displacement.

11. Now again rotate the micrometer screws such that the reading on micrometer is 10 mm.

Now the display must show 00.00 mm.

12. If not, repeat the steps 5 to 9 again.

13. Now again rotate the micrometer so that the reading on micrometer is 20.00 mm.

14. Displacement = (10.00 - 20.00) = - 10.00 mm.

15. Now check the display reading. In this case the reading on the display is around – 10.00

mm.

16. Now LVDT trainer kit is ready for the experiment.

17. Now again rotate the micrometer screw such that the displacement of core is zero.

Page 22: Jawaharlal Nehru Engineering College, Aurangabad

18. Now move the core of the LVDT IN THE +VE & -ve direction with respect to the null

position & observe the readings on the display & enter your readings in the observation

table.

Observation Table:-

Least count of micrometer = value of smallest division on main scale

Total no of divisions

= 0.01 mm

Sr.

No.

Core

displacement

Display

reading

Secondary

Voltage

Graph:- 1. Core displacement Vs display reading

2. Core Displacement Vs Secondary Voltage

Conclusion:- Distance on display depends upon core position in +ve & -ve direction.

Page 23: Jawaharlal Nehru Engineering College, Aurangabad

Exercise No.10: (2 Hours) – 1 Practical

Temperature Measurement Using Thermocouple.

Aim: - To Plot the characteristic of Thermocouple.

Experimental Setup: -

Thermocouple:

i. Copper – Constantan

ii. Iron - Constantan

Electric Heater

Water Container

Multimeter

Thermometer

Temp. Measurement Trainer.

Theory: -

• Thermocouples.

• Thermoelectric Effect.

• Seeback Effect and its circuit diagram.

• Practical Thermocouple and its circuit diagram.

• List Standard Thermocouples and its comparison.

Circuit Diagram:-

Page 24: Jawaharlal Nehru Engineering College, Aurangabad

Procedure: -

1. Connect the Thermocouple supplied to you, at the input terminals. If Copper constantan

thermocouple is used copper wire must be connected to positive terminal and constantan

wire must be connected to negative terminal.

2. Deep the junction in water container.

3. Hold the thermometer in water container.

4. Switch ON the electric water heater.

5. Note down the temp. on thermometer, temp. of the measuring instrument and note down

the voltage output at the end terminal of both metal wire using multimeter.

6. Plot the Temperature vs. output characteristic.

Observation table:-

Sr.no. Thermometer

reading (°c)

Instrument reading

( °c)

Output voltage

Precaution: - While connecting the thermocouple to the input terminals, observe the polarity.

Graph: - Output voltage Vs Temperature(°c).

Conclusion: -

Page 25: Jawaharlal Nehru Engineering College, Aurangabad

Exercise No.5: (2 Hours) – 1 Practical

Measurement of power in 3-phase A.C. circuit by two wattmeter’s method.

Aim:-

1. To Measure the active reactive power in 3 phase circuit.

2 .To Measure the power factor.

Apparatus Required:-

1. 3-phase Auto transformer 20 Amp. 440v, 50 Hz.

2. Wattmeter dynamometer type 2 No. 250v, 5A

3. Ammeter moving Iron type : 1 no(10A)

4. Voltmeter Moving Iron type 1 No.(600V)

5. 3 phase Load or 3phase induction motor (415V, 5H.P.)

6. Connective leads.

Theory:-

Two wattmeter method can be employed to measure power in a 3- phase, 3 wire star or delta

connected balance or unbalanced load. In this method, the current coils of the wattmeter are

connected in any two lines say R and Y and potential coil of each wattmeter is joined across the

same line and third line i.e. B.

Then the sum of the power measured by two wattmeter W1 and W2 is equal to the power

absorbed By the 3 phase load

Total power P = √3VLILCOS = (W1+W2 watts)* M.F.

Power factor COS = (W1+W2) *M.F.

√3 VLIL

= P/ √ 3 VLIL

And reactive power of load= Q=√3 (W1+W2)* M.F.

Page 26: Jawaharlal Nehru Engineering College, Aurangabad

Circuit diagram:-

Procedure:-

1. Connect the Voltmeter, Ammeter and Wattcmeters to the load through 3 phase

Autotransformer as shown fig and set up the Autotransformer to Zero position.

2. Switch on the 3 phase A.C. supply and adjust the autotransformer till a suitable voltage. Note

down the readings of wattcmeters, voltmeter& ammeter

3. Vary the voltage by Autotransformer and note down the various readings.

4. Now after the observation switch off and disconnect all the Equipment or remove the lead

wire.

Page 27: Jawaharlal Nehru Engineering College, Aurangabad

OBSERVATION TABLE:

Multiplying factor of the wattmeter is…………….

Calculations:

Total power = (W1+W2) *Multiplying factor

tan =√3 (W2-W1)

W1+W2

= tan-1√3 (W2-W1)

W1+W2

Power Factor = cos = (W1+W2) M.F.

√3 VLIL

Reactive power = √3 (W1-W2)* M.F.

IR =IY = IB for Balance Load

Precautions & Sources of Error:-

1. Proper currents and voltage range must be selected before putting the instruments in the

circuit.

2. If any Wattmeter reads backward, reverse its pressure coil connection and the reading as

negative.

3. As the supply voltage Fluctuates it is not possible to observe the readings correctly.

Conclusion: - The power measured in the circuit and there corresponding power factors in

observation table.

Page 28: Jawaharlal Nehru Engineering College, Aurangabad

Exercise No.9: (2 Hours) – 1 Practical

Anderson’s bridge

Aim: - To measure the self - inductance of a given coil by Anderson’s bridge method.

Apparatus: - Inductor, standard capacitor, resistors (fixed resistances and variable pots as given

in the circuit) signal generator, head phones and connecting terminals.

Theory: -

Anderson’s bridge is the most accurate bridge used for the measurement of self – inductance

over a wide range of values, from a few micro-Henries to several Henries. In this method the

unknown self-inductance is measured in terms of known capacitance and resistances, by

comparison. It is a modification of Maxwell’s L – C Bridge. In this bridge, double balance is

obtained by the variation of resistances only, the value of capacitance being fixed.

Circuit Diagram:-

Procedure:-

The circuit diagram of the bridge is as shown in the figure. The coil whose self-inductance is to

be determined, is connected in the arm AB, in series with a variable non-inductive resistor R1.

Arms BC, CD and DA contain fixed and non – inductive resistors R2, R3 and R4 respectively.

Another non - inductive resistor R5 is connected in series with a standard capacitor C and this

Page 29: Jawaharlal Nehru Engineering College, Aurangabad

combination is put in parallel with the arm CD. The head - phones are connected between B and

E. The signal generator is connected between A and C junctions.

Select one capacitor and one inductor and connect them in appropriate places using patch chords.

The signal generator frequency is adjusted to audible range. A perfect balance is obtained by

adjusting R1 and R5 alternatively till the head – phones indicate a minimum sound. The values

of R1 and R5 are measured with a multi-meter (While measuring the R1 and R5 values, they

should be in open circuit).In the balance condition the self – inductance value of the coil is

calculated by using the above formula. The experiment is repeated with different values of C.

Observation Table:-

Sr.No. Capacity

( C )

μ F

Resistance

( R1 )

Ω

Resistance

( R5 )

Ω

Calculated

value (L)

C [ ( R1+ R2 )

R5 +R2R4]

mH

Standard

value of L

mH

Calculations: - Inductance of given coil L = C [(R1+ R2) R5 + R2R4] mH

Where C = Capacity of the standard capacitor (μ F)

R2, R3, R4 = Known, fixed and non – inductive resistances (KΩ)

R1, R5 = Variable resistances (KΩ)

Precautions: - 1) The product (CR2R4) must always be less than L.

2) R1 and R5 are adjusted until a minimum sound is heard in head – phones.

Conclusion:-

Page 30: Jawaharlal Nehru Engineering College, Aurangabad

Exercise No.:11 (2 Hours) – 1 Practical

Aim :Calibration of Voltmeter and Ammeter by Potentiometer

• Calibration of voltmeter using DC potentiometer

Items Required

1. Calibration of voltmeters and ammeter by Potentiometer

2. Potentiometer

3. Sliding jockey

4. Mains cord

5. Patch cords

Procedure

1. Connect the mains cord to Nvis 6032 Trainer and switch On Mains Supply.

2. Switch on the trainer by Power switch.

3. Note the output of standard DC supply (Vdc) by connecting terminal 32 to digital

voltmeter V1’s positive terminal and ground terminal 6 to negative of V1.

4. Once voltage is noted from V1, disconnect them and connect the negative terminal

of galvanometer G1 to positive terminal 32 of DC supply.

5. Connect positive terminal of G1 to jokey.

6. Connect terminal 3 and 4 to digital ammeter A1 polarity wise.

7. Connect DC potentiometer between terminal 5 and 6. Connect 5 to X and 6 to Z

terminal.

8. Vary VR2 knob to set the current in A1 (say 30 mA).

9. Touch jokey to X and then to Z terminals of potentiometer and see the reading of

galvanometer. Compare both reading of galvanometer (if terminal X gives positive reading than

terminal Z will give negative reading). If it is not the case then change the value of current in A1

(step 8) and check again.

10. Now slide the jokey on potentiometer wire and the find null point i.e., the point where

galvanometer G1 shows zero reading.

11. Now measure distance L (in cm) moved from terminal Z to null point. Note: Count the

number of wire ‘n’ from the Z terminal, for odd line of wire take

Page 31: Jawaharlal Nehru Engineering College, Aurangabad

reading from lower scale and for even line wire take reading from upper scale (r) Now put the

value of r and n in below formula to calculate Distance (L) L = [(n-1)*100 + r] cm

For example, if the null point is at the 7th wire then read the lower scale and if reading is 68 cm

then the distance is given by L = (7-1)*100 + 68 cm = 668cm

12. Divide output of DC Supply Vdc to the distance L. This will give a constant C which is

voltage drop per cm. L V C = dc

13. Now disconnect DC supply from Galvanometer G1.

14. Connect terminal 1 to 7 and terminal 2 to 8.

15. Connect analog DC Voltmeter (V) between terminals 9 & 10 (i.e 33 to 9 and 34 to 10).

16. Connect the voltage divider input and output terminals (i.e 11 to 27, 12 to 28, 13 to 29 and 14

to 30).

17. Connect terminal 15 to negative terminal of Galvanometer G1.

18. Connect positive terminal of Galvanometer G1 to jokey.

19. Set the voltage in analog DC Voltmeter (V) to some value (say 1 V) with the help of VR1

knob.

20. Touch jokey to X and then to Z terminals of potentiometer and see the reading of

galvanometer. Compare both reading of galvanometer (if terminal X gives positive reading than

terminal Z will give negative reading). If the two readings are of same sign than use the voltage

divider knob to change the voltage ratio ‘M’ and check again.

21. Now slide the jokey on potentiometer wire and the find null point i.e., the point where

galvanometer G1 shows zero reading.

22. Now measure distance D (in cm) moved from terminal Z to null point.

23. Note the value of DC voltmeter (V), voltage ratio M and null point position D in the

Observation Table-1.

24. Now change the value of V by VR1 knob and repeat the above steps. Record all the readings

in Observation Table-1.

25. Also plot the graph between V v/s Vp to obtain calibration curve for the analog voltmeter.

Page 32: Jawaharlal Nehru Engineering College, Aurangabad

Observation Table-1

Value of standard DC Supply Vdc = ………

Value of Calibration constant C = Vdc/L = ………

Page 33: Jawaharlal Nehru Engineering College, Aurangabad

Objective: Calibration of Ammeter using DC potentiometer

Items Required

1. Calibration of voltmeters and ammeter by Potentiometer Trainer

2. Potentiometer

3. Sliding jockey

4. Mains cord

5. Patch cords

Procedure:

1. Connect the mains cord to Nvis 6032 Trainer and switch on Mains Supply.

2. Switch on the trainer by Power switch.

3. Note the output of DC supply (Vdc) by connecting terminal 32 to digital voltmeter V1’s

positive terminal and ground terminal 6 to negative of V1.

4. Once voltage is noted from V1, disconnect them and connect the negative terminal of

galvanometer G1 to positive terminal 32 of DC supply.

5. Connect positive terminal of G1 to jokey.

6. Connect terminal 3 and 4 to digital ammeter A1 polarity wise.

7. Connect DC potentiometer between terminal 5 and 6. Connect 5 to X and 6 to Z terminal.

8. Vary VR2 knob to set the current in A1 (say 35 mA).

9. Touch jokey to X and then to Z terminals of potentiometer and see the reading of

galvanometer. Compare both reading of galvanometer (if terminal X gives positive reading than

terminal Z will give negative reading). If it is not the case then change the value of current in A1

(step 8) and check again.

10. Now slide the jokey on potentiometer wire and the find null point i.e., the point

where galvanometer G1 shows zero reading.

11. Now measure distance L (in cm) moved from terminal Z to null point.

Note: Count the number of wire ‘n’ from the Z terminal, for odd line of wire take

reading from lower scale and for even line wire take reading from upper scale (r)

Now put the value of r and n in below formula to calculate Distance (L)

L = [(n-1)*100 + r] cm

For example, if the null point is at the 7th wire then read the lower scale and if

reading is 68 cm then the distance is given by L = (7-1)*100 + 68 cm = 668cm

Page 34: Jawaharlal Nehru Engineering College, Aurangabad

12. Divide output of DC Supply Vdc to the distance L. This will give a constant C

which is voltage drop per cm.

L

C = Vdc

13. Now disconnect DC supply from Galvanometer G1.

14. Connect terminal 1 to 17 and terminal 2 to 18.

15. Connect analog DC Ammeter (A) between terminals 18 & 20 (i.e 36 to 18 and 35

to 20).

16. Now make all the rotary knob of Variable Resistance to their highest value.

17. Connect Variable Resistance between terminal 17 & 19 (i.e 17 to 37 and 19 to 38).

18. Connect the voltage divider input and output terminals (i.e 21 to 27, 22 to 28, 23 to

29 and 24 to 30).

19. Connect terminal 25 to negative terminal of Galvanometer G1.

20. Connect positive terminal of Galvanometer G1 to jokey.

21. Set the value of variable resistance to any arbitrary value and then set the current in

analog DC Ammeter (A) to some value (say 0.1 A) with the help of VR1 knob. If

the current can not be set to the desired value then change the value of variable

resistance to get the desired current.

22. Touch jokey to X and then to Z terminals of potentiometer and see the reading of

galvanometer. Compare both reading of galvanometer (if terminal X gives positive

reading than terminal Z will give negative reading). If the two readings are of same

sign then use the voltage divider knob to change the voltage ratio ‘M’ and check

again.

23. Now slide the jokey on potentiometer wire and the find null point i.e., the point

where galvanometer G1 shows zero reading.

24. Now measure distance D (in cm) moved from terminal Z to null point.

25. Note the value of DC Ammeter (A), voltage ratio M and null point position D in

the Observation Table -2

26. Change the value of current in interval of 0.1 A and record in Observation Table-2

27. Also plot the graph between I v/s Ip to obtain calibration curve for the analog

Ammeter

Page 35: Jawaharlal Nehru Engineering College, Aurangabad

Conclusion:

Page 36: Jawaharlal Nehru Engineering College, Aurangabad

Exercise no:-12

Aim: Determination of unknown inductance using Maxwell’s inductance bridge method

Equipments Needed:

1. Nvis 6533

2. 2 mm patch cords

3. Digital multimeter

4. Oscilloscope

Circuit diagram: Mimic illustration with connections for Maxwell‟s inductance bridge method for

measurement of unknown inductance is shown in the below figure.

Circuit Diagram:

Page 37: Jawaharlal Nehru Engineering College, Aurangabad

Procedure:

1. Connect a patch cord between socket „13‟ of Vin terminals of Maxwell‟s inductance bridge

and socket „29‟ of Vout terminals of the 1 KHz sine wave generator

2. Connect a patch cord between socket „14‟ of Vin terminals of Maxwell‟s inductance bridge

and socket „30‟ of Vout of the 1 KHz sine wave generator.

3. Connect a patch cord between sockets „1‟ and „2‟ and connect another patch cord between

sockets „8‟ and „9‟ to determine the value of Lx1 and Rx1.

4. Connect patch cords between sockets „15‟ and „17‟ and sockets „16‟ and „18‟for the purpose

of null detection.

5. Set the potentiometer R2 in counter clockwise direction.

6. Switch „On‟ the power supply.

7. Set the amplitude of 1 KHz sine wave generator maximum i.e. rotate amplitude control knob

fully clockwise.

8. Rotate the potentiometer R2 towards clockwise direction very precisely to find a condition

where null (Zero or minimum reading) is displayed on DPM.

9. Switch „Off‟ the power supply.

10. Remove all the patch cords connected to bridge circuit for accurate reading of R2.

11. Now measure the resistance R2 between test-points „5‟ and „6‟ using a digital multimeter.

12. Calculate the value of inductance Lx1 and resistance Rx using the formula

Lx = L1.R2/ R4

Where, Lx=Lx1, L1= 12H, R4=100 ohm

13. Calculate the value of unknown resistance using the formula

Rx = R2.R3 /R4. Where, R3=120, R4=100ohm

Page 38: Jawaharlal Nehru Engineering College, Aurangabad

14. Now again connect a patch cord between sockets „1‟ and „3‟ and another patch cord

between sockets „8‟ and „10‟ as shown in below figure to determine the value of Lx2 and Rx2.

Page 39: Jawaharlal Nehru Engineering College, Aurangabad

15. Repeat the above procedure from step 6 to 9.

16. Remove all the patch cords connected to bridge circuit for accurate reading of R2.

17. Now measure the value of resistance R2 between test-points „5‟ and „6‟ using a digital

multimeter.

18. Calculate the value of inductance Lx2 and resistance Rx using the formula

Lx = L1.R2/ R4 Where, Lx=Lx2, L1= 12microH, R4=100ohm

19. Calculate the value of unknown resistance using the formula

Rx = R2.R3/ R4. Where, R3=237ohm, R4=100ohm

Page 40: Jawaharlal Nehru Engineering College, Aurangabad
Page 41: Jawaharlal Nehru Engineering College, Aurangabad

Note:

1. The actual values of inductors Lx1, Lx2 and Lx3 are56μH, 24μH and 12μH.

2. The actual value of resistor Rx is 470 Ω.

3. Calculated RX is (Rx + r), where r is internal resistance of corresponding inductor.

4. Small amount of error would be there due to component tolerance and also due to human

error.

Conclusion:-