lc/ms of intact therapeutic monoclonal antibodies …...suresh babu c.v. agilent technologies, inc....
TRANSCRIPT
LC/MS of Intact TherapeuticMonoclonal Antibodies UsingAgilent AdvanceBio RP-mAb
Author
Suresh Babu C.V.
Agilent Technologies, Inc.
Application Note
Biologics and Biosimilars
Introduction
Therapeutic proteins such as monoclonal antibodies (mAbs) are gaining muchattention in the biopharmaceutical industry. Post translational modifications duringthe mAb manufacturing process commonly result in microheterogeneity. Hence, it isimportant to monitor and analyze these changes to ensure drug efficacy.Reversed-phase LC/MS is the routine method for the analysis of mAbs at intact orfragment level, which provides accurate molecular mass. The correct choice ofLC column and method is critical to achieve fast analysis times and reproduciblehigh-resolution separations.
We used Agilent AdvanceBio RP-mAb columns for intact mAb analysis todemonstrate the fast analysis time and highly reproducible chromatographicseparation provided by these columns. AdvanceBio RP-mAb columns with C4,SB-C8, and Diphenyl chemistries are based on Agilent Poroshell technology withsuperficially porous particles (3.5 µm) with wide pores (450Å) [1]. These columnsdeliver higher resolution and faster run times compared to columns packed withfully porous particles of the same size. In this work, multiple therapeutic mAbs,including an antibody drug conjugate (ADC), were analyzed using AdvanceBioRP-mAb columns in an LC/MS method. This approach delivered fast analysis timesand accurate mass determination of intact mAbs.
2
Experimental
SamplesTherapeutic monoclonal antibodies mAb1, mAb2, andantibody drug conjugate mAb3 were purchased from a localpharmacy and stored according to the manufacturers’instructions. mAb1, mAb2, and mAb3 (lys-conjugated ADC)were diluted to 1 µg/µL using 0.1% formic acid in 3% ACN,and 1 µL was injected.
InstrumentationLC: Agilent 1290 Infinity LC
MS: Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight(Q-TOF) with Agilent Jet Stream ion source
Results and DiscussionDetermination of therapeutic protein molecular weights andglycosylation profiles is best established by LC/MS. Achievinggood chromatographic peak and high signal-to-noise ratio m/zmass spectra is a tradeoff when working with high molecularweight proteins such as mAbs [2]. In addition, for high-throughput mAb screening in a batch-to-batch variation study,fast and reproducible LC/MS methods are required.AdvanceBio RP-mAb columns are packed with superficiallyporous particles. This particle design reduces diffusiondistances, which allows high linear velocities to be usedwhile still maintaining high chromatographic efficiency. Thisapproach delivers high-speed and high-resolution separations.To improve resolution, different C4, SB-C8, and Diphenylbonded phases are available for flexible method development.
Figures 1 and 2 show the LC/MS data for the three intactmAbs with AdvanceBio RP-mAb C4 and Diphenyl columns,respectively. Both columns provided excellent total ionchromatogram peak shapes with narrow peak width (0.5 minat base) using an isopropanol:acetonitrile:water mobile phase(Figures 1A and 2A). The charge-state envelope ranged from2,000 to 4,000 m/z with a Gaussian distribution (Figure 1Band 2B). The raw mass spectra were converted to zero-chargemass spectra using the maximum entropy and peak modeling(pMod) deconvolution algorithms in Agilent MassHunterBioConfirm software. The deconvoluted spectra are shown inFigures 1C and 2C. Five major glycoforms are seen in themAb1 deconvoluted spectrum, while four major glycoformsare evident in the mAb2 spectrum. Due to the strongheterogeneity of mAb3, split peaks were observed with bothC4 and Diphenyl columns (Figure 2A). These split peakscorresponded to the mAb with different payloads.Deconvoluted mass spectra for mAb3 (Figure 2C) showedincreasing payload trend in steps of one drug load with eightmajor drug conjugations (D0, D1, D2...D8).
ConditionsParameter Agilent 1290 Infinity LC
Column AdvanceBio RP-mAb C4, 2.1 × 50 mm, 3.5 µm(p/n 799775-904 )AdvanceBio RP-mAb Diphenyl, 2.1 × 75 mm, 3.5 µm(p/n 799775-944)
Inj vol 1 µL
Sample thermostat 5 °C
Mobile phase A 0.1% Formic acid in water
Mobile phase B 80% IPA:10% ACN: 9.9% water with 0.1% formic acid
Gradient At 0 min & 20% BAt 4 min & 20% BAt 5 min & 40% BAt 10 min & 70% BAt 11 min & 90% BAt 11.1 min & 20% B
Stop time 11.1 min
Post time 4 min
Column temp 80 °C
Flow rate 0.6 mL/min
Parameter Agilent 6530 Accurate-Mass Q-TOF LC/MS
Ion mode Positive ion mode, dual AJS ESI (profile)
Drying gas temp 350 °C
Drying gas flow 8 L/min
Sheath gas temp 400 °C
Sheath gas flow 11 L/min
Nebulizer 35 psi
Capillary voltage 5,500 V
Fragmentor voltage 380 V
Skimmer voltage 65 V
Oct RF Vpp 750 V
Acquisition parameters MS mode Data acquired at 1 GHz, MS only mode, mass range
2,000 to 6,000 m/z
Data analysis Data from LC/MS were analyzed usingAgilent MassHunter Qualitative Analysis software andAgilent MassHunter BioConfirm software. Max entropyand pMod deconvolution algorithms were used toobtain zero-charge spectra of the drug conjugatedantibody (mAb3) and mAb1/mAb2, respectively.
3
Figure 1. Intact mAb mass analysis on an Agilent AdvanceBio RP-mAb C4, 2.1 × 50 mm, 3.5 µm column. A) Total ionchromatogram, (B) mass spectrum, (C) deconvoluted spectrum.
Figure 2. Intact mAb mass analysis on an Agilent AdvanceBio RP-mAb Diphenyl, 2.1 × 50 mm, 3.5 µm column. A) Total ionchromatogram, (B) mass spectrum, (C) deconvoluted spectrum.
×107
×107
×107
×104
×104
×103
×105
×105
×103
0123456
1 2 3 4 5 6 7 8 9 10
0123456
1 2 3 4 5 6 7 8 9 10
012345
Acquisition time (min)1 2 3 4 5 6 7 8 9 10
00.40.81.21.62.02.4
2,200 2,600 3,000 3,400 3,800
00.20.40.60.81.01.21.41.6
2,200 2,600 3,000 3,400 3,800
00.40.81.21.6
22.42.8
Mass-to-charge (m/z)2,200 2,600 3,000 3,400 3,800
00.40.81.21.6
22.4 148,225.23
148,063.92 148,389.73
148,545.04147,917.97
147,800 148,000 148,200 148,400 148,600
00.20.40.60.81.01.21.4 147,244.07 147,405.32
147,079.63 147,564.57
Deconvoluted mass (amu)
147,000 147,200 147,400 147,600
00.40.81.21.62.02.42.8 153,018.05151,101.04 152,058.74
149,183.80150,143.59 153,970.66
154,928.31
156,046.34148,225.52
148,000 152,000 156,000
G0/G0F
G0F/G0F
G0F/G1F
G1F/G1F
G1F/G2F
D0
D1D2
D3 D4D5
D6D7
D8
G0F/G0F
G0F/G1F G0F/G2F
G1F/G2F
A B C
mAb1
mAb3 (antibody drug conjugated)
mAb 2
Coun
ts
Coun
ts
Coun
ts
Coun
ts
Coun
ts
Coun
ts
Coun
ts %
Co
unts
%
Coun
ts %
0
1
2
3
4
5
1 2 3 4 5 6 7 8 9 10
0
1
2
3
4
5
1 2 3 4 5 6 7 8 9 10
01
234
5
1 2 3 4 5 6 7 8 9 10
00.40.81.21.62.02.4
2,200 2,600 3,000 3,400 3,800
0
0.4
0.8
1.2
1.6
2.0
2,400 2,800 3,200 3,600 4,000
00.40.81.21.62.02.42.83.2
2,200 2,600 3,000 3,400 3,800 4,200
00.20.40.60.81.01.21.41.6 148,222.07
148,385.74148,062.43
148,549.88147,916.85
147,800 148,000 148,200 148,400 148,600
00.20.40.60.81.01.21.41.61.82.0 147,241.47
147,402.69
147,081.28 147,565.13
147,000 147,200 147,400 147,600 147,800
00.81.62.43.24.0
152,058.30153,014.75
151,106.88150,142.95 153,974.62
149,187.21154,928.52
156,044.28148,227.09
148,000 150,000 152,000 154,000 156,000 158,000
D0
D1D2
D3D4 D5
D6
D7
D8
G0F/G0F
G0F/G1FG0F/G2F
G1F/G2F
G0/G0F
G0F/G0F
G0F/G1F
G1F/G1F
G1F/G2F
×107
×107
×107
×104
×104
×103
×105
×105
×103
Acquisition time (min) Mass-to-charge (m/z) Deconvoluted mass (amu)
A B C
Counts
Counts
Counts
Counts
Counts
Counts
Counts %
Counts %
Counts %
mAb1
mAb3 (antibody drug conjugated)
mAb 2
www.agilent.com/chem
For Research Use Only. Not for use in diagnostic procedures.
Agilent shall not be liable for errors contained herein or for incidental or consequentialdamages in connection with the furnishing, performance, or use of this material.
Information, descriptions, and specifications in this publication are subject to changewithout notice.
© Agilent Technologies, Inc., 2015, 2017Printed in the USANovember 2, 20175991-6296EN
Conclusions
We successfully analyzed therapeutic mAbs usingAgilent AdvanceBio RP-mAb columns coupled to anAgilent 6530 Accurate-Mass Q-TOF LC/MS. The C4 andDiphenyl AdvanceBio RP-mAb columns, with anMS-compatible method, delivered fast and high-resolutionanalysis for intact and ADC mAbs.
References
1. Gritti, F. Chromatog. Today June 2012, 4-11.
2. Gudihal, R.; Suresh Babu C. V.; Tang, N. Analysis ofMonoclonal Antibody (mAb) Using Agilent 1290 InfinityLC System Coupled to Agilent 6530 Accurate-MassQuadrupole Time-of-Flight (Q-TOF); Application note,Agilent Technologies, Inc. Publication number 5991-4266EN, 2014.
For More Information
These data represent typical results. For more information onour products and services, visit our Web site atwww.agilent.com/chem.