mac 600 resting ecg analysis system - ge … › - › jssmedia › 7c5b85c8518347b7...proprietary...

14
MAC 600 Resting ECG Analysis System Education Services Clinical Development Quick Reference Guide QRG MAC 600 R3 062210.indd 1 6/24/10 10:17 AM

Upload: others

Post on 06-Jul-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

MAC™600 Resting ECG Analysis System

Education ServicesC l i n i c a l D e v e l o p m e n t

Quick Reference Guide

QRG MAC 600 R3 062210.indd 1 6/24/10 10:17 AM

Page 2: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

QRG MAC 600 R3 062210.indd 2 6/24/10 10:17 AM

Page 3: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

NoticeThe materials contained in this document are intended for educational purposes only. This document does not establish specifications, operating procedures or maintenance methods for any of the products referenced. Always refer to the official written materials (labeling) provided with the product for specifications, operating procedures and maintenance requirements.

Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to agents and employees of GE Healthcare or other parties expressly licensed by GE. Unlicensed use is strictly prohibited.

Skin Preparation and Lead Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Acquire an ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Print Continuous Rhythm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

File Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Contents

QRG MAC 600 R3 062210.indd 3 6/24/10 10:17 AM

Page 4: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

Skin

Pre

para

tion

and

Lead

Pla

cem

ent

1

NRLAlternative

C2IECAHA

V2

C3AHA

V3IEC

C4AHA

V4IEC

C5IECAHA

V5

C6IECAHA

V6

FIECAHA

LL

FIECAHA

LL

Alternative

Alternative

LIECAHA

LA

LIECAHA

LAAlternative RIECAHA

RA

V1

AHA

C1

IEC

RIECAHA

RA

IECAHA

NRLIECAHA

QRG MAC 600 R3 062210.indd 4 6/24/10 10:17 AM

Page 5: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

V1 red C1 red Fourth intercostal space at the right sternal border.

V2 yellow C2 yellow Fourth intercostal space at the left sternal border.

V3 green C3 green Midway between C2/V2 and C4/V4.

V4 blue C4 brown Mid-clavicular line in the fifth intercostal space.

V5 orange C5 black Anterior auxiliary line on the same horizontal level as C4/V4.

V6 purple C6 purple Mid-auxiliary line on the same horizontal level as C4/V4 and C5/V5.

LA black L yellow Above left wrist (alternate placement: left deltoid).

LL red F green Above left ankle (alternate placement: upper leg close to torso).

RL green N black Above right ankle (alternate placement: upper leg close to torso).

RA white R red Above right wrist (alternate placement: right deltoid).

AHA lead IEC lead Electrode Placement

Skin

Pre

para

tion

and

Lead

Pla

cem

ent

2

Skin Preparation and Lead Placement 1. Shave any hair from each electrode site and degrease each electrode site with alcohol.

2. Dry skin completely.

3. Apply the leads on the prepared area.

4. Verify the leads are connected and working properly.

Table 1 AHA and IEC Lead Placement

QRG MAC 600 R3 062210.indd 5 6/24/10 10:17 AM

Page 6: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

Leads

ECG

F4

MAC 600

- +Press ‘ECG’ (Green) key to record an ECG.

0:29:40 5-Aug-2009

77 BPM

Patient Data 25 mm/s 10 mm/mV More

I

II

III

aVR V1 V4

aVL V2 V5

aVF V3 V6

F3F2F1

1 2 3

4 5 6

7 8 9

esc 0

Acqu

ire a

n EC

G

3

6

7

3a

1

43b

QRG MAC 600 R3 062210.indd 6 6/24/10 10:17 AM

Page 7: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

Acquire an ECG

Power the system on.

2. Prepare the patient. Refer to section Skin Preparation and Lead Placement for details.

3. Verify waveform quality by:

• Checking Hook-up Advisor. Green (always right most circle) indicates a good quality waveform.

• Press the Leads key to scroll through the leads.

Press F1 to open the Patient Data window. Enter the patient ID, DOB and gender using the numbered keyboard and arrow pad.

• Press F1 to insert a dash while entering the patient ID if needed.

• Press F2 to insert a + symbol while entering the patient ID if needed.

Acqu

ire a

n EC

G4

0:29:40 5-Aug-2009 77 BPM

I

II

III

aVR V1 V4

aVL V2 V5

aVF V3 V6

- +

Enter Patient Data.

ID numberDate of birthGenderReturn

MAC600

0:29:40 5-Aug-2009 77 BPM

I

II

III

aVR V1 V4

aVL V2 V5

aVF V3 V6

- +

Enter Patient Data.

ID numberDate of birthGenderReturn

MAC600

4

Color Display is a purchasable option.

3a

3b

5. Select Return, and press the center of the arrow pad to close the Patient Data window.

Press the ECG key on the keypad to record an ECG.

Note! Refer to the Operator’s Manual for instruction on how to use Secondary ID, age and alfa-numeric input feature.

7

Use the F keys to make any necessary adjustments to:

• Speed (F2)

• Gain (F3)

•More (F4) > Filter (F1) to adjust Waveform Filter. Press More (F4) to return ECG Screen.

1 6

QRG MAC 600 R3 062210.indd 7 6/24/10 10:17 AM

Page 8: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

Acqu

ire a

n EC

G

5

F4F3F2F1

1 2 3

4 5 6

7 8 9

esc 0 6

- +

Press ‘ECG’ (Green) key to record an ECG. 0:29:40

5-Aug-2009 77 BPM

Continue Rhythm Store Cancel

I

II

aVR V1 V4

aVL V2 V5

9 108. Press Analysis (F2) to view the Measurement and Interpretation information. Once you have pressed Analysis you see the F keys as displayed in the above illustration.

Note! Analysis and Measurement are purchasable options.

If after reviewing the ECG and Analysis if you wish to stop and acquire another ECG, press Cancel (F4) and press the ECG key on the keypad to record another ECG. Otherwise continue with step 10.

Other options from the Analysis window include:

Press Continue (F1) to continue with the acquired 12 lead ECG.

Press Store (F3) to store the ECG to the SD Card.

Note! Storing to SD card is a purchasable option.

b

c Press Rhythm (F2) to view the waveform.

ca 9b

10

a

QRG MAC 600 R3 062210.indd 8 6/24/10 10:17 AM

Page 9: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

2

F4F3F2F1

1 2 3

4 5 6

7 8 9

esc 0

- +

Directory of fi les in SD card

Select Select All Report Setup More

Directory of f i les in SD cardFile Manager

ID number Date/ Time Sent0000001 01- Aug-09 23:35:48 No0000002 30-Jul -09 23:35:48 No0000003 29-Jul -09 03:17:10 No0000004 29-Jul -09 03:15:36 No0000005 29-Jul -09 03:16:14 No

Total f i les: 20 Selected f i les: 1

1

3 Prin

t Con

tinou

s Rh

ythm

6

Print Continous Rhythm

Press the Leads key to select the desired 3-lead group for rhythm.

Press the Rhythm key on the Keypad to start the printing of continuous rhythm.

Note! If your display is showing the 12-lead or 6-lead presentation, pressing the Rhythm key will result in the display of 3 leads from the group configured as the autorhythm and will begin rhythm printing with those 3 leads.

Press the Stop Writer key on the keypad to stop the printing of the continuous rhythm.

1

2

3

QRG MAC 600 R3 062210.indd 9 6/24/10 10:17 AM

Page 10: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

File

Man

aage

r (Pu

rcha

sabl

e op

tion)

7

F4F3F2F1

1 2 3

4 5 6

7 8 9

esc 0

- +

Directory of fi les in SD card

Select Select All Report Setup More

Directory of f i les in SD cardFile Manager

ID number Date/ Time Sent0000001 01- Aug-09 23:35:48 No0000002 30-Jul -09 23:35:48 No0000003 29-Jul -09 03:17:10 No0000004 29-Jul -09 03:15:36 No0000005 29-Jul -09 03:16:14 No

Total f i les: 20 Selected f i les: 1

File Manager (Option)

1. From the startup screen press More (F4).

2. Press File Manager (F3).

Press Report Setup (F3) to select desired format for your printed reports.

a. Press Report Format (F1) to open the reports selection screen.

b. Use the arrow pad to go to the desired report.

c. Enter 1 for each report you wish printed. You may enter a higher number if you wish more copies.

d. Select Return to save and exit screen.

e. Press More (F4).

f. Press Return (F2) to get back to the main File Manager Screen.

Press Select (F1) to select individual reports. Use the arrow pad to navigate through the list and press the Enter key of the arrow pad to highlight the desired report. Use the same procedure to highlight multiple reports if desired.

If you wish to highlight all reports press Select All (F2).

34

5

354

Arrow key

Enter key

QRG MAC 600 R3 062210.indd 10 6/24/10 10:17 AM

Page 11: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

File

Man

ager

8

F4F3F2F1

1 2 3

4 5 6

7 8 9

esc 0 6

- +Directory of fi les in SD card

Print Display Transmit More

Directory of f i les in SD cardFile Manager

ID number Date/ Time Sent0000001 01- Aug-09 23:35:48 No0000002 30-Jul -09 23:35:48 No0000003 29-Jul -09 03:17:10 No0000004 29-Jul -09 03:15:36 No0000005 29-Jul -09 03:16:14 No

Total f i les: 20 Selected f i les: 1

Once you have selected a report(s) you have the following options:

Print (F1): prints the ECG of the selected report(s)

Display (F2): allows you to view the Medians (F1), Analysis (F2) or Print (F3) of the selected record(s).

Transmit (F3): allows you to transmit to a MUSE system or XML output via Serial Line.

Note! Serial Transmission is a purchasable option.

Select More (F4) then Return (F2) to get back to File Manager main screen.

Selected report

1

2

3

4

1 2 43

Note! Your F key order may vary slightly from what is described based on your purchased options.

QRG MAC 600 R3 062210.indd 11 6/24/10 10:17 AM

Page 12: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

File

Man

ager

9

F4F3F2F1

1 2 3

4 5 6

7 8 9

esc 0 6

- +

Directory of fi les in SD card

Delete Save XML Save PDF More

Directory of f i les in SD cardFile Manager

ID number Date/ Time Sent0000001 01- Aug-09 23:35:48 No0000002 30-Jul -09 23:35:48 No0000003 29-Jul -09 03:17:10 No0000004 29-Jul -09 03:15:36 No0000005 29-Jul -09 03:16:14 No

Total f i les: 20 Selected f i les: 1

5. More (F4): Gives more selections including:

Delete (F1): press to delete record. User must enter password to delete record(s) from system

Save XML (F2): saves record to SD card in XML format

Note! Save to XML is purchasable option.

Save PDF (F3): saves record to SD card in PDF format

Note! Save to PDF is purchasable option.

More (F4): returns previous screen. Press Return (F1) to return to File Manager main screen.

6. From the main File Manager screen press More (F4) then press Resting ECG (F3) to return to the Resting ECG screen.

a

b

c

d

a b c d

QRG MAC 600 R3 062210.indd 12 6/24/10 10:17 AM

Page 13: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

QRG MAC 600 R3 062210.indd 13 6/24/10 10:17 AM

Page 14: MAC 600 Resting ECG Analysis System - GE … › - › jssmedia › 7c5b85c8518347b7...Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to

© 2010 General Electric Company – All rights reserved.

General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation.

GE, GE Monogram, Marquette, 12SL, MAC and MUSE are trademarks of General Electric Company.

GE Medical Systems Information Technology Inc., a General Electric company, doing business as GE Healthcare.

Notice: The materials contained in this document are intended for educational purposes only. This document does not establish specifications, operating procedures or maintenance methods for any of the products referenced. Always refer to the official written materials (labeling) provided with the product for specifications, operating procedures and maintenance requirements.

Proprietary Training Material, property of GE Healthcare. Use of these materials is limited to agents and employees of

GE Healthcare or other parties expressly licensed by GE Healthcare. Unlicensed use is strictly prohibited.

PN: 2047426-003 DPrinted in India

GE Healthcare9900 Innovation DriveWauwatosa, WI 53223USA

www.gehealthcare.com

0459

QRG MAC 600 R3 062210.indd 14 6/24/10 10:17 AM