MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
DEPARTMENT: SPG NO: DATE ISSUED: SUPERSEDES:
Institutional Laboratory Management Office
ILMO-05-02
July 1, 2013 ILMO-05-01
EECE-EELAB-02-01
SUBJECT:
Policies and Guidelines on Repair, Preventive Maintenance, and Calibration of Laboratory Tools, Instruments, and Equipment
PREPARED BY: APPROVED BY: EFFECTIVITY:
ENGR. ROEL JOHN C. JUDILLA
ILMO Director
BONIFACIO T. DOMA, JR., PH.D.
EVP-Academic Affairs
Effective Immediately
I. PURPOSE
This Standard Practice Guide (SPG) provides the policies and guidelines on repair, preventive
maintenance, and calibration (RPM/C) of laboratory tools, instruments, and equipment. The schedule of calibration of laboratory tools, instruments, and equipment is included in this SPG.
II. SCOPE
This SPG is applicable to all laboratory tools, instruments, and equipment managed by the
Institutional Laboratory Management Office (ILMO). III. POLICIES
1. The Institutional Laboratory Management Office is responsible for the repairs and preventive maintenance of laboratory tools, instruments, and equipment. In cases where RPM cannot be done in-house, ILMO requests for the appropriate service
provider.
2. ILMO ensures that laboratory tools, instruments, and equipment requiring calibration are calibrated by authorized calibration service providers.
IV. GUIDELINES
1. The laboratory assistants are responsible to keep all tools, instruments, and equipment in good condition. Incidents of loss, damage, or breakage must be reported immediately to the Operations and Maintenance Engineer.
2. In all laboratories, except in the ChE-Chm laboratories where a stricter preventive
maintenance program is implemented to ensure the safety of students, faculty members, and laboratory staff during laboratory classes and research experiments, the
SPG No: ILMO-05-02
Page 2 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
schedule of repair and preventive maintenance procedures, as follows, must be followed: a. Inspection and testing – at least once a week b. Cleaning – every after use; end of each term c. Repair works – end of each term, or as the need arises based on the assessment of
the laboratory assistant and the Operations and Maintenance Engineer
3. All laboratory assistants must ensure that the tools, instruments, and equipment in their respective laboratories that require calibration are calibrated in accordance with the
manufacturer’s specifications, or whenever the tools, instruments, and equipment were subjected to some form of abuse that may affect performance and accuracy of
measurements, test results or outputs.
4. The laboratory assistants and the Operations and Maintenance Engineer must ensure that RPM procedures and calibration services are budgeted. Items that need to be replaced must also be budgeted.
5. If RPM/C of a tool, instrument, or equipment is done in-house (by ILMO, Campus
Development and Management Office, or Development Office for Information Technology), the laboratory assistants must make sure that the serviced tool,
instrument, or equipment is tagged with the ILMO RPM/C certification sticker.
6. For tools, instruments, and equipment that need RPM/C from an authorized service provider, the laboratory assistants must submit the RPM/C service request slip to ILMO one month before the RPM/C schedule. The laboratory assistants must make sure that the authorized service provider tags the serviced tool, instrument, or equipment with their official RPM/C certification sticker, in addition to the Certificate of RPM/C they provide. If the authorized service provider has no official RPM/C certification sticker, the ILMO RPM/C certification sticker will be used.
7. The end-of-term laboratory inventory reports of laboratory assistants must include the RPM/C status of each tool, instrument, or equipment in their respective laboratories .
V. ATTACHMENTS
A. Schedule of Calibration of Tools, Instruments, and Equipment B. Preventive Maintenance Program for ChE-Chm Laboratories C. ILMO RPM/C Certification Sticker D. RPM/C Service Request Slip
SPG No: ILMO-05-02
Page 3 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
VI.DISTRIBUTION LIST
Office/Department Copy
Holder # ILMO 0
OEVPAA 1 DOIT 2
CDMO 3
Makati Laboratory Administrator 4 CEGE Laboratory 5
ChE-Chm Laboratory 6 EE-ECE-COE Laboratory 7
EMSE Laboratory 8
IE-EMG Laboratory 9
AR/MAS Laboratory 10
ME/Shop Laboratory 11 Physics Laboratory 12
Psychology Laboratory 13
Water Refilling Station 14
CQIO 15
SPG No: ILMO-05-02
Page 4 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
DEPARTMENT: SPG NO: DATE ISSUED: SUPERSEDES:
Institutional Laboratory Management Office
ILMO-05-02
July 1, 2013 ILMO-05-01
EECE-EELAB-02-01
SUBJECT:
Policies and Guidelines on Repair, Preventive Maintenance, and Calibration of Laboratory Tools, Instruments, and Equipment
PREPARED BY: APPROVED BY: EFFECTIVITY:
ENGR. ROEL JOHN C. JUDILLA
ILMO Director
BONIFACIO T. DOMA, JR., PH.D.
EVP-Academic Affairs
Effective Immediately
Attachment A: SCHEDULE OF CALIBRATION OF TOOLS, INSTRUMENTS, AND EQUIPMENT
Survey Equipment Method of Calibration Schedule
Engineer's Level (Dumpy)
NAK 2 Collimator
Annually
Engineer's Level (Precise)
Theodolite (Optical) Three-Prism Assembly
Engineer's Transit
Total Station
Mechanical Instruments Method of Calibration Schedule
Pressure Gauges Dead Weight Tester Annually
Temperature Gauges Comparison Method
Weighing Scales Method of Calibration Schedule
Digital Weighing Scale
Comparison Method Annually
Solution Balance
Platform Scale
Spring Balance
Analytical Balance
Ergonomics Equipment Method of Calibration Schedule
Sound Level Meter Self-calibration Before Use
Light meter
Comparison Method Annually
Radiometer
Noise Dosimeter
Psychrometer
Barometer
SPG No: ILMO-05-02
Page 5 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
Biometrics
Blood Pressure Monitor Every 2 years
Electronics Test Equipment Method of Calibration Schedule
Oscilloscope Self-Calibration Before Use
Digital Multimeter Comparison Method Annually
Function Generator
Electrical Measuring Equipment Method of Calibration Schedule
DC Ammeter and Mill iammeter
Comparison Method Annually AC Ammeter and Mill iammeter
Wattmeter
Equipment for Destructive Testing Method of Calibration Schedule
Universal Testing Machine Comparison Method Annually
Rockwell Hardness Tester
Thermal Analysis Equipment Method of Calibration Schedule
TMA Quartz Disc Indium Calibration Semi-Annually
TGA Sealed Pans Weight Calibration Semi-Annually
Digital Scanning Calorimeter Indium Test Semi-Annually
SPG No: ILMO-05-02
Page 6 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
DEPARTMENT: SPG NO: DATE ISSUED: SUPERSEDES:
Institutional Laboratory Management Office
ILMO-05-02
July 1, 2013 ILMO-05-01
EECE-EELAB-02-01
SUBJECT:
Policies and Guidelines on Repair, Preventive Maintenance, and Calibration of Laboratory Tools, Instruments, and Equipment
PREPARED BY: APPROVED BY: EFFECTIVITY:
ENGR. ROEL JOHN C. JUDILLA
ILMO Director
BONIFACIO T. DOMA, JR., PH.D.
EVP-Academic Affairs
Effective Immediately
Attachment B: PREVENTIVE MAINTENANCE PROGRAM FOR CHE-CHM LABORATORIES
PREVENTIVE MAINTENANCE PROCEDURES: ANALYTICAL BALANCE
ACTIVITY FREQUENCY
1. Check the condition of the following parts:
Housing and display glass
Weighing pan and pan support
Operating keys
Leveling feet and level
Protective cover
Draft shield
AC adapter, power cable and fuse
Measuring cell and balance main board
Before use
2. Clean pan and compartment After every use
3. Calibrate with standard weights Quarterly
4. Manufacturer cleaning and calibration Annually
AUTOCLAVE
ACTIVITY FREQUENCY
1. Drain exhaust bottle When water is above the HIGH level
2. Drain chamber When it will not be used for a long period
3. Clean chamber and body Monthly
SPG No: ILMO-05-02
Page 7 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
5. Check steril ization indicator tape Before use
6. Check gaskets Weekly
7. Manufacturer maintenance Annually
ATOMIC ABSORPTION SPECTROPHOTOMETER
ACTIVITY FREQUENCY
1. Clean burner head After every use
2. Check tubings, pump and lamps Before use
3. Manufacturer preventive maintenance The following areas should be considered during the maintenance: Alignment and wavelength calibration of the optical system.
Checking and cleaning of burner/sample systems.
Adjustment and calibration of the electronics.
Checking and cleaning of the solvent delivery system.
General Cleaning of the Instrument.
Instrument performance validation.
Semi-Annually
BOMB CALORIMETER
ACTIVITY FREQUENCY
1. Clean the unit Every after use
2 Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check the electronic circuit and its performance.
Calibrate the unit.
Check and test for any leakages and loose connections.
Clean the external and internal rubber tubings.
Annually
FISHER CENTRIFUGE MODEL 225
ACTIVITY FREQUENCY
1. Clean exposed surfaces, rotor, and protective bowl Quarterly
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check motor
Check motor brush for replacement.
Adjust speed control.
Annually
SPG No: ILMO-05-02
Page 8 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
CHAIN BALANCE
ACTIVITY FREQUENCY
1. Check the condition of the following parts: Metal base
Glass cylinder
Hard rolled aluminum alloy beam
Roller weight
Plated chain
Thermometer plummet
Stirrup
Slotted glass basket
Before use
2. Manufacturer preventive maintenance Annually
CONDUCTIVITY METER
ACTIVITY FREQUENCY
1. Clean conductivity probe Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check the electronic circuit and its performance.
Check and clean conductivity probe for any deposits
Check voltages current from probe to cable and AC cable.
Calibrate conductivity a 1314ms prepared by laboratory technician.
Annually
DISSOLVED OXYGEN METER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check membrane for deterioration.
Check battery level and electronics.
Every two years
DISTILLING APPARATUS
ACTIVITY FREQUENCY
1. Drain to prevent accumulation of boiler scale Weekly
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check the electronic circuit and its performance.
Clean the internal and external body.
Annually
SPG No: ILMO-05-02
Page 9 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
ELECTROANALIZER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check electronic circuit and two (2) potentiometers current controls.
Check and calibrate analogue meters.
Check motor wirings and motor belt.
Check electrical connections for any leakages and short circuit. Check fuses and
solenoid valves. Clean the main unit.
Annually
FUME HOODS
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Remove all corrosion, spot prime and paint.
Inspect electrical connections, tighten as required.
Inspect motor control contacts for wear or pitting, replace as needed.
Tighten all terminal lugs.
Replace the drive belt as needed.
Check the hood performance.
Bi-Annual
FURNACE
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the ma intenance
Check electronic circuit and its performance.
Check temperature controller circuit program.
Check thermocouple probe.
Check thermometer temperature against certified thermometer or equivalent.
Clean the internal and external body of the unit.
Annually
GAS CHROMATOGRAPHY
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance Semi-annually
SPG No: ILMO-05-02
Page 10 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
The following areas should be considered during the maintenance
Laboratory Readiness Check and verify that the power outlets are of the proper configuration
for the plugs on the gas chromatography and integrator. Measure that the power is within specification.
Electrical/ Interface Inspection
Check power supply requirements against actual supply.
Check all electrical and terminal connections.
Perform cleaning on internal parts and electronic boards.
Gas Lines Check and test the gas l ine tubing.
Purge all the gases for few minutes
Check for leaks in all gas l ines.
Instrument Set-up Verify the following gas connections, injector, detector and power source.
Check and verify the connection to Integrator.
Check FID connector cable
Perform cleaning on FID
Instrument Start-up Connect the GC to the main power outlet
Turn on the system
Check GC operation and program
Inject solvent (hexane) to verify the function of detector
HOTPLATE
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Clean the bimetal control points.
Check the following for possible replacement:
a. Insulator on adjusting shaft. b. Heater coil c. Cycle l ight
d. Control unit e. Heating element
Calibrate.
Bi-Annual
INCUBATOR
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Clean the interior body.
Bi-Annual
SPG No: ILMO-05-02
Page 11 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
Check thermometer temperature against certified thermometer or
equivalent.
ION SELECTIVE ELECTRODE METER
ACTIVITY FREQUENCY
1. Clean electrode Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check probe for cracks and power levels of fi l l ing solution.
Check reference junction. Clean electrode
Check response time
Check electronics for loose connections and cracked leads.
Annually
MICROSCOPE
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check and clean the eyepiece.
Clean and check the observation alignment
Check and clean the objective
Check and clean the calibration of the mech stage.
Check and clean the condenser i l lumination
Check and clean the electrical stage
Bi-Annual
OVEN
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check electronic circuit and performance.
Check temperature controller circuit program.
Check electronic timers.
Check thermometer temperature against certified thermometer or
equivalent Clean the internal and external body of the unit.
Annually
SPG No: ILMO-05-02
Page 12 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
SARTORIUS BASIC pH METER PB-20
ACTIVITY FREQUENCY
1. Clean electrode Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check probe for cracks and power levels of fi l l ing solution.
Check reference junction.
Clean electrode
Check response time
Check electronics for loose connections and cracked l eads.
Annually
REFRACTOMETER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check the electronic circuit and its performance
Calibrate using a standard liquid
Check and clean the lens.
Test, check lamp and voltages
Annually
REFRIGERATOR
ACTIVITY FREQUENCY
1. Clean external and internal body Quarterly
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Clean the interior body
Check thermometer temperature against certified thermometer or
equivalent.
Every two years
SOLUTION CALORIMETER
ACTIVITY FREQUENCY
1. Clean external and internal body Every after use
2. Manufacturer preventive maintenance
Check the rotating cell for leaks.
Clean the motor and oil the motor bearing.
Check and clean blower fi lters.
Check the electronic circuit and its performance.
Check parameters
Annual ly
SPG No: ILMO-05-02
Page 13 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
SOXHLET FAT EXTRACTOR
ACTIVITY FREQUENCY
1. Clean external and internal body Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check electrical circuit main fuse.
Check on/off switches.
Check heaters connection, fuses and any leakages or short circuit.
Check electrical connections for any leakages and short circuit.
Clean the unit.
Bi-Annual
TENSIOMETER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Align beam string.
Calibrate the unit
Annually
TRANSFERENCE NUMBER APPARATUS
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check electronic circuit and potentiometers current controls.
Check and calibrate analogue meter using another meter (AC/DC mA)
Check tubings and connections.
Check electrical connections for any leakages and short circuit.
Check fuses and solenoid valves.
Clean the main unit.
Annually
TRIPLE BEAM BALANC E
ACTIVITY FREQUENCY
1. Clean pan and compartment Every after use
2. Check with a standard weights Quarterly
SPG No: ILMO-05-02
Page 14 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
UV-VIS SPECTROPHOTOMETER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Alignment and wavelength calibration of the optical system.
Check and clean burner/sample systems (additional to routine customer
maintenance) Adjustment and calibration of the electronics.
Check and clean the instrument pneumatics.
Check and clean the solvent delivery system.
Clean the instrument.
Validate the instrument performance.
Annually
VISCOMETER - BROOKFIELD
ACTIVITY FREQUENCY
1. Clean Every after use
2. Check the condition of the following parts: Base, Leveling screws, Upright rod, Jam Nut, and Clamp Assembly.
Before use
2. Manufacturer check-up and calibration Annually
AUTOCLAVE
ACTIVITY FREQUENCY
1. Clean the unit Every after use
3 Calibration and maintenance
The following areas should be considered during the maintenance:
Check the electronic circuit and its performance.
Calibrate the unit.
Check and test for any leakages and loose connections.
Clean the external and internal rubber tubing.
Annually
COD REACTOR and PHOTOMETER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check the electronic circuit and performance
Calibrate the unit
Annually
SPG No: ILMO-05-02
Page 15 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
COLONY COUNTER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check the electronic circuit and performance
Calibrate the unit
Annually
ELECTRONIC DENSIMETER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check the electronic circuit and performance Calibrate the unit
Annually
INCUBATOR WITH SHAKER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Clean the interior body
Check thermometer temperature against certified thermometer or
equivalent.
Bi-Annually
LAMINAR FLOW
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Remove all corrosion, spot prime and paint.
Inspect electrical connections, tighten as required.
Inspect motor control contacts for wear or pitting, replace as needed.
Tighten all terminal lugs.
Replace the drive belt as needed.
Check the hood performance
Bi-Annually
SPG No: ILMO-05-02
Page 16 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
PIPETOR
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Calibrate the unit
Annually
SONICATOR
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check the electronic circuit and performance
Calibrate the unit
Annually
SPECTROPHOTOMETER
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Alignment and wavelength calibration of the optical system.
Check and clean burner/sample systems (additional to routine c ustomer
maintenance) Adjustment and calibration of the electronics.
Check and clean the instrument pneumatics.
Check and clean the solvent delivery system.
Clean the instrument.
Validate the instrument performance.
Annually
UV TRANSILLUMINATOR
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Alignment and calibration of the optical system.
Adjustment and calibration of the electronics.
Validate the instrument performance.
Annually
SPG No: ILMO-05-02
Page 17 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
WATERBATH
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Check electronic circuit and performance.
Check electronic timers.
Check thermometer temperature against certified thermometer or
equivalent Clean the internal and external body of the unit.
Annually
FOURIER TRANSFORM INFRARED
ACTIVITY FREQUENCY
1. Clean Every after use
2. Manufacturer preventive maintenance
The following areas should be considered during the maintenance
Purging Nitrogen gas to expel unwanted gas that will damage optics.
Replace dessicant.
Run background testing
Perform instrument validation testing.
Reset maintenance warnings.
Clean the internal and external body of the unit.
Bi-Annually
SPG No: ILMO-05-02
Page 18 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
DEPARTMENT: SPG NO: DATE ISSUED: SUPERSEDES:
Institutional Laboratory Management Office
ILMO-05-02
July 1, 2013 ILMO-05-01
EECE-EELAB-02-01
SUBJECT:
Policies and Guidelines on Repair, Preventive Maintenance, and Calibration of Laboratory Tools, Instruments, and Equipment
PREPARED BY: APPROVED BY: EFFECTIVITY:
ENGR. ROEL JOHN C. JUDILLA
ILMO Director
BONIFACIO T. DOMA, JR., PH.D.
EVP-Academic Affairs
Effective Immediately
Attachment C: ILMO RPM/C CERTIFICATION STICKER
SPG No: ILMO-05-02
Page 19 of 19
MAPÚA INSTITUTE OF TECHNOLOGY STANDARD PRACTICE GUIDE
DEPARTMENT: SPG NO: DATE ISSUED: SUPERSEDES:
Institutional Laboratory Management Office
ILMO-05-02
July 1, 2013 ILMO-05-01
EECE-EELAB-02-01
SUBJECT:
Policies and Guidelines on Repair, Preventive Maintenance, and Calibration of Laboratory Tools, Instruments, and Equipment
PREPARED BY: APPROVED BY: EFFECTIVITY:
ENGR. ROEL JOHN C. JUDILLA
ILMO Director
BONIFACIO T. DOMA, JR., PH.D.
EVP-Academic Affairs
Effective Immediately
Attachment D: RPM/C SERVICE REQUEST SLIP