metal micro-detector timepix imaging synchrotron radiation beams

14
Metal micro-detector TimePix imaging synchrotron radiation beams at the ESRF Bio-Medical Beamline ID17 Andrii Chaus Institute for Nuclear Research National Academy of Sciences Kiev, Ukraine RD51 mini week June 14, 2012

Upload: anana

Post on 08-Jan-2016

53 views

Category:

Documents


9 download

DESCRIPTION

Metal micro-detector TimePix imaging synchrotron radiation beams at the ESRF Bio-Medical Beamline ID17. Andrii Chaus Institute for Nuclear Research National Academy of Sciences Kiev, Ukraine. RD51 mini week June 14, 2012. Introduction. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Metal micro-detector TimePix imaging synchrotron radiation beams

Metal micro-detector TimePix imaging synchrotron radiation beams

at the ESRF Bio-Medical Beamline ID17

Andrii Chaus

Institute for Nuclear ResearchNational Academy of Sciences

Kiev, Ukraine

RD51 mini weekJune 14, 2012

Page 2: Metal micro-detector TimePix imaging synchrotron radiation beams

2

Introduction

A.Chaus, June 14, 2011

The main goal of radiotherapy is to deposit a high dose of ionizing radiation in a tumor while keeping the absorbed dose in the surrounding healthy tissue at a tolerant level.

For measurement of such high dose radiation proof detector is needed. Metal detector is actually the radiation hard detector.

Page 3: Metal micro-detector TimePix imaging synchrotron radiation beams

The high intensity X-ray mini-beams (50- 600 keV, Intensity: 2,7×109 photons/(s×mm2) ) with sizes of 600  µm and 1200  µm center-to-center separation have to be delivered with a Multi Slit Collimator (MSC), inserted into the beam.

A.Chaus, June 14, 2011

Page 4: Metal micro-detector TimePix imaging synchrotron radiation beams

PRINCIPLE OF OPERATION Incident particles on the Incident particles on the strips initiate secondary strips initiate secondary electron emission as they electron emission as they pass through the nearly pass through the nearly transparent medium. When transparent medium. When this happens, a positive this happens, a positive charge appears at the charge appears at the integrator end is measured.integrator end is measured.

To improve the extraction To improve the extraction of secondary electrons an of secondary electrons an accelerating electric field is accelerating electric field is applied around the strip.applied around the strip.

This technology works with This technology works with x-rays, protons and ion x-rays, protons and ion beams. Additionally, the beams. Additionally, the strips are nearly strips are nearly transparent to beams, transparent to beams, significantly reducing significantly reducing degradation that is degradation that is experienced by absorbing experienced by absorbing detectors.detectors.

Page 5: Metal micro-detector TimePix imaging synchrotron radiation beams

Structure of Medipix detector

5A.Chaus, June 14, 2011

Page 6: Metal micro-detector TimePix imaging synchrotron radiation beams

TimePix – Metal Detector

Photo of the individual pixels of the TimePIx chip (55 x 55 μm2, 256 x 256 pixels) CERN, MEDIPIX Collaboration.

Diameter of the area for bonding ~15 μm.

A positive voltage has been appliedto a mesh over the chip area – to collect Secondary electrons emitted by metal surface of the chip

6A.Chaus, June 14, 2011

Page 7: Metal micro-detector TimePix imaging synchrotron radiation beams

Metal TimePix detector performance at the Bio-medical beamline ID17

(ESRF, Grenoble, November 2010).

•Rotation of the chopper and vertical movement of the gonio-meter table with a TimePix detector were synchronized:• SR beam shaped by the collimating system was impinging the detector area 10 times, shifted every time by 1200  µm.

7A.Chaus, June 14, 2011

Page 8: Metal micro-detector TimePix imaging synchrotron radiation beams

Metal TimePix detector performance at the Bio-medical beamline ID17

(ESRF, Grenoble, November 2010).

8A.Chaus, June 14, 2011

Page 9: Metal micro-detector TimePix imaging synchrotron radiation beams

Very important value is the peak-to-valley dose ratio (PVDR). When we used all data from detector (including cloud) PVDR is ~ 5. But when we used only data in peaks and between peaks our one dimension figure changes. Value of cloud is blue value minus yellow value.

One dimension presentation of 8 beams.

Blue – with cloud. Yellow – without.

Page 10: Metal micro-detector TimePix imaging synchrotron radiation beams

-5

0

5

10

15

20

25

30

35

40

45

50

0 2000 4000 6000 8000 10000 12000 14000

Position,

Do

se, G

r

TimePix

Gafromic f ilm

TimePix (Metal) and Gafchromic films measuring dose distribution

at the Bio-medical beamline ID17 (ESRF, Grenoble, November 2010).

Nearly perfect correspondence of the two measurements WITH A DIFFERENCE:

TimePix provides IMAGING IMMEDIATELY! Conventional dose distribution measurement (gafchromic films) using scanning microscope technique takes up to 24 hours.

10

Page 11: Metal micro-detector TimePix imaging synchrotron radiation beams

Metal TimePix detector performance at the Bio-medical beamline ID17 (ESRF, Grenoble, November 2010)

Irradiation impact – ‘memory’ of the beams

11A.Chaus, June 14, 2011

Page 12: Metal micro-detector TimePix imaging synchrotron radiation beams

CONCLUSIONS

The TimePix metal detector is a suitable tool for the Minibeam Radiation therapy applications. The results obtained for high intensity synchrotron radiation illustrate an excellent performance of the TimePix providing in real time 2D image of the dose distribution over many beams with a spatial resolution of about 10 μm. Peak-to-Valley Dose Ratios measured by the Time-Pix and gafhromic films agree well. In our measurements we show that we have some clouds close to peaks and value of this clouds must be considered when cure will be created for the patients.

12A.Chaus, June 14, 2011

Page 13: Metal micro-detector TimePix imaging synchrotron radiation beams

13

Acknowledgements• These studies were carried out with coauthors:• M. Campbell, L. Tlustos, X.Llopart , CERN• S. Pospisil, IEAP, Prague• Y. Prezado, M. Renier (ESRF).• V. Pugatch, O.Kovalchuk, O. Okhrimenko, D.

Storozhik, INR NASU, Kiev

• We would like to thank MEDIPIX-2 Collaboration

• This work was funded partially by the CNCP (UK) project “Microstrip Metal Detectors”.

You can find the PDF file of published article here:http://authors.elsevier.com/offprints/NIMA54351/

f9f5587dc38bb7e192ad9a053fd6ea9c

A.Chaus, June 14, 2011

Page 14: Metal micro-detector TimePix imaging synchrotron radiation beams

14

THANK YOU FOR YOUR ATTENTION