model dc-3300 - frank's hospital...

50
NIDEK LASER DIODE PHOTOCOAGULATOR Model DC-3300 SERVICE MANUAL December, 2004 XDC7**RDA001E/E Total 50 pages

Upload: buiquynh

Post on 25-Mar-2018

226 views

Category:

Documents


5 download

TRANSCRIPT

Page 1: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

NIDEK

LASER DIODE PHOTOCOAGULATOR

Model DC-3300

SERVICE MANUAL

December, 2004XDC7**RDA001E/E

Total 50 pages

Page 2: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

NIDEK CO., LTD. : 34-14, Maehama, Hiroishi-cho, Gamagori, Aichi 443-0038, Japan(Manufacturer) Telephone: (0533) 67-6611

Facsimile: (0533) 67-6610NIDEK CO., LTD : 6th Floor, Takahashi Bldg., No.2, 3-chome, Kanda-jinboucho(Tokyo Office) Chiyoda-ku, Tokyo 101-0051, Japan

Telephone: (03) 3288-0571Facsimile: (03) 3288-0570Telex: 2226647 NIDEK J

NIDEK INCORPORATED : 47651 Westinghouse Drive, Fremont, California 94539, U. S. A.(United States Agent) Telephone: (510) 226-5700

Facsimile: (510) 226-5750NIDEK SOCIETE ANONYME : Europarc 13, rue Auguste Perret, 94042 CRETEIL, France(Authorized Representative) Telephone: (01) 49 80 97 97

Facsimile: (01) 49 80 32 08

Page 3: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

Page

Table of Contents

§1 CAUTIONS FOR MAINTENANCE ......................................................................... 1-1

§2 TROUBLESHOOTING .............................................................................................. 2-1

§3 SUB-TROUBLESHOOTING ..................................................................................... 3-13.1 The system does not start at all ................................................................................ 3-13.2 Fan does not work .................................................................................................... 3-33.3 Indication on the front panel is not correct ............................................................... 3-33.4 Err. 20 appears .......................................................................................................... 3-43.5 Setting of coagulation power output cannot be changed correctly ........................... 3-53.6 Setting of coagulation time cannot be changed correctly ......................................... 3-53.7 Repeat mode cannot be set correctly ........................................................................ 3-53.8 Brightness of aiming beam cannot be changed correctly ......................................... 3-53.9 Aiming beam cannot be seen .................................................................................... 3-53.10 Err. 3 appears .......................................................................................................... 3-63.11 READY indication does not light up by pressing the STATUS switch .................. 3-73.12 STAND BY indication does not light up by pressing the STATUS switch ............ 3-73.13 Counter cannot be reset .......................................................................................... 3-73.14 Laser beam cannot be emitted ................................................................................ 3-73.15 Err. 1 appears .......................................................................................................... 3-83.16 Err. 10 appears ........................................................................................................ 3-93.17 Err. 12 appears ........................................................................................................ 3-93.18 Err. 13 appears ........................................................................................................ 3-93.19 Err. 15 appears ........................................................................................................ 3-93.20 Err. 53 appears ...................................................................................................... 3-103.21 Err. 54 appears ...................................................................................................... 3-103.22 INTLK2.7 appears ................................................................................................ 3-113.23 INTLK4.7 appears ................................................................................................ 3-123.24 Actual power output is low................................................................................... 3-133.25 Actual power output is high ................................................................................. 3-143.26 EMERGENCY OFF button does not work .......................................................... 3-143.27 Err. 4 appears ........................................................................................................ 3-14

§4 REPLACEMENT ......................................................................................................... 4-14.1 Replacement of fuses ............................................................................................... 4-14.2 Replacement of upper cover ..................................................................................... 4-14.3 Replacement of lower cover ..................................................................................... 4-24.4 Replacement of front cover ...................................................................................... 4-24.5 Replacement of PS1 ................................................................................................. 4-34.6 Replacement of BA05 (SW BOARD)...................................................................... 4-34.7 Replacement of BA02 (LASER CONTROL BOARD) ........................................... 4-3

Page 4: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

Page4.8 Replacement of BA01 (MASTER BOARD) ........................................................... 4-34.9 Replacement of Key control switch ......................................................................... 4-44.10 Replacement of EMERGENCY OFF button ......................................................... 4-44.11 Replacement of fan ................................................................................................. 4-54.12 Replacement of BA03 (DISPLAY BOARD) ......................................................... 4-54.13 Replacement of BA09 (SENSOR BOARD) of safety shutter ................................ 4-64.14 Replacement of BA09 (SENSOR BOARD) of coagulation shutter ...................... 4-64.15 Replacement of microswitch of CONDENSOR ASSY ......................................... 4-7

§5 ADJUSTMENT ............................................................................................................ 5-15.1 Optical adjustment ................................................................................................... 5-1

5.1.1 Adjustment of optical axis of aiming beam .................................................... 5-15.1.2 Adjustment of fiber alignment ........................................................................ 5-25.1.3 Operation check of coagulation shutter sensor ............................................... 5-35.1.4 Operation check of safety shutter sensor ......................................................... 5-3

5.2 Electrical adjustment ................................................................................................ 5-45.2.1 Setting of LASER CONTROL BOARD (BA02) ............................................ 5-45.2.2 Setting of DISPLAY BOARD (BA03) ............................................................ 5-45.2.3 Input voltage check ......................................................................................... 5-45.2.4 Adjustment of T.E.COOLER .......................................................................... 5-55.2.5 COAG STANDBY check ................................................................................ 5-55.2.6 D/A converter signal check ............................................................................. 5-55.2.7 Current limit check .......................................................................................... 5-5

5.3 General adjustment ................................................................................................... 5-65.3.1Current limit adjustment .................................................................................. 5-65.3.2 Adjustment of 1/4 wavelength constant position ............................................ 5-65.3.3 Overshoot check .............................................................................................. 5-65.3.4 Light control adjustment ................................................................................. 5-65.3.5 Adjustment of each monitor ............................................................................ 5-75.3.6 Adjustment of overpower ................................................................................ 5-75.3.7 Adjustment of each delivery unit .................................................................... 5-85.3.8 Adjustment of aiming beam ............................................................................ 5-95.3.9 Adjustment of COAG TIME ......................................................................... 5-105.3.10 Adjustment of REPEAT TIME .................................................................... 5-10

§6 REFERENCES ............................................................................................................. 6-16.1 Wiring diagram (CABLE ASSY) ............................................................................. 6-16.2 Tools ......................................................................................................................... 6-26.3 Error code list ........................................................................................................... 6-36.4 Interlock code list ..................................................................................................... 6-36.5 DIP SW function list ................................................................................................ 6-46.6 System parameter ..................................................................................................... 6-4

6.6.1 Change/Setting method of value of parameter ................................................ 6-46.6.2 System parameter list ...................................................................................... 6-5

Page 5: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

§1 CAUTIONS FOR MAINTENANCE

• Be sure to use this manual after reading and understanding the operator’s manual for the DC-3300 thoroughly.

• Only properly trained service technicians can repair the system.

• During the repair, any indications are necessary to prohibit the unauthorized persons fromentering the working area and to prevent the accidents such as emitting laser to personnel.

• Wear protective goggles when the laser is emitted.

• Never look directory into the laser diode beam even if wearing goggles.

• Never point the probe tip or the laser aperture to personnel.

• Apply the lock-tight to the screws when they are loosened and tightened again.

• For maintenance, turn OFF the power and disconnect the socket from the wall-outlet unlessthis manual specifies “Turn ON the power”.

• In case that the abnormal smell and/or noise is produced from DC-3300, turn the key switch tothe OFF position immediately or press the EMERGENCY OFF button.

• When performing a maintenance with the covers (upper, lower or front cover) removed, takethe following measures in order to protect the laser diodes from static electricity.

• Spread a conductive mat on the workbench, then connect the earth wire from the mat to aground.

• Field service representative should put a wrist strap on his/her wrist, and connect the earthwire from it.

• Earth the measuring instruments such as oscilloscope and laser power meter as well.• Use a soldering iron provided with an earth wire.• Before using a tool, always touch the metal part to the conductive mat.• Connect an earth wire to the terminal on the rear panel as well.

• Specifications and design are subject to change without notice for improvement. Since theTechnical Bulletin will be released every time significant changes are made to the system, referto the Technical Bulletin together with this manual.

Page 6: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

1 - 2

Page 7: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

§2 TROUBLESHOOTING

Turn ON the key control switch. Does thesystem start correctly?

Can the setting of coagulation poweroutput be changed correctly?

Can the setting of coagulation time bechanged correctly?

Can the brightness of aiming beam bechanged correctly?

YES

YES

YES

3.5 Setting of coagulation power outputcannot be changed correctly

3.6 Setting of coagulation time cannot bechanged correctly

3.8 Brightness of aiming beam cannot bechanged correctly

3.9 Aiming beam cannot be seen3.10 Err. 3 appears

Does READY light up by pressing theSTATUS switch?

YES

3.11 READY indication does not light upby pressing the STATUS switch

Is the system put in the initial conditioncorrectly after count down?

YES

NO3.4 Err. 20 appears

Can the repeat mode be set correctly? 3.7 Repeat mode cannot be reset correctlyNO

YES

NO

3.1 The system does not start at all3.2 Fan does not work3.3 Indication on the front panel is not correct3.27 Err. 4 appears

NO

YES

NO

NO

NO

Page 8: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

2 - 2

Does STAND BY light up by pressing theSTATUS switch?

Is the laser diode beam emitted correctlyby pressing the foot switch?

Does the system stop by pressing theEMERGENCY OFF button?

YES

YES

3.14 Laser beam cannot be emitted3.15 Err. 1 appears3.16 Err. 10 appears3.17 Err. 12 appears3.18 Err. 13 appears3.19 Err. 15 appears3.20 Err. 53 appears3.21 Err. 54 appears3.22 INTLK2.7 appears3.23 INTLK4.7 appears3.24 Actual power output is low3.25 Actual power output is high

3.26 EMERGENCY OFF button does notwork

Is the Application COUNTER reset bypressing the Counter Reset switch?

YES

NO3.13 Counter cannot be reset

YES

NO

3.12 STAND BY does not light up bypressing the STATUS switch

NO

END

NO

Page 9: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

§3 SUB-TROUBLESHOOTING

3.1 The system does not start at all

Is the voltage between 1st and 2nd pins ofthe connector P505, and, P501 and P503 onBA05, +12V?

Is the voltage of the outlet proper?

Are connectors P901, P902 and P903 onPS1 (switching power supply) connectedcorrectly?

Are connectors P501, P502, P503, P504and P505 on BA05 connected correctly?

Is the voltage between 1st pin of theconnector J902 and 1st pin of J903 on PS1(switching power supply) +12V?Is the voltage between 2nd pin of J902 and2nd pin of J903 on PS1 (switching powersupply) +12V?

YES

YES

YES

Correct the connection.

Correct the connection.

Failure of PS1 (switching power supply).Replace it with the new one (see 4.5).

YES

Failure of EA21.Repair or replace it with the new one.

Are the fuses normal?

YES

NO Replace the fuses with the new ones (see4.1).

Is the voltage between 2nd and 3rd pins ofthe connector J901 on PS1 (switchingpower supply) proper?

Failure of EA20.Repair or replace it with the new one.

NO

YES

NO

Obtain the proper line voltage.NO

YES

NO

NO

NO

Page 10: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 2

Is the EMERGENCY OFF button iscorrect?

Is the voltage between connectors P502and P504 on BA05 +12V?

Is the voltage between 1st and 2nd pins ofthe connector J201 on BA02 +12V?

Is the voltage between 1st and 2nd pins ofthe connector J209 on BA02 +12V?Is the voltage between 3rd and 4th pins ofthe connector J209 on BA02 +5V?

Is the key control switch correct?

YES

YES

YES

Failure of EA22.Repair or replace it with the new one.

Failure of BA02.Replace it with the new one (see 4.7).

Failure of the key control switch.Replace it with the new one (see 4.9).

YES

Failure of the EMERGENCY OFF button.Replace it with the new one (see 4.10).

Is the connector P201 on BA02 connectedcorrectly?

YES

NOCorrect the connection.

Is the voltage between 1st and 3rd pins ofthe connector J113 on BA01 +12V?Is the voltage between 1st and 2nd pins ofthe connector J110 on BA01 +5V?

Failure of EA13.Repair or replace it with the new one.

NO

YES

NO

Failure of BA05.Replace it with the new one (see 4.6).

NO

YES

NO

Failure of BA01.Replace it with the new one (see 4.8).

NO

NO

Page 11: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 3

3.2 Fan does not workIs the connector P207 on BA02 connectedcorrectly?

Failure of the fan.Replace it with the new one (see 4.11).

YES

Is the voltage of connector P207 on BA02correct?

Connector P2071-2: DC12V

YES

NO Failure of BA02.Replace it with the new one (see 4.7).

Correct the connection.NO

3.3 Indication on the front panel is not correct

Is the flexible flat cable between theconnector J301 on BA03 and the connectorJ106 on BA01 connected correctly?

Failure of BA03.Replace it with the new one (see 4.12).

NO Failure of BA01.Replace it with the new one (see 4.8).

NOCorrect the connection.

Is the indication on the front panel correct?

YES

YES

Page 12: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 4

3.4 Err. 20 appears

Adjust T.E.COOLER.

Is the connector P901 on PS1 connectedcorrectly?

Is the voltage output to TP9 on BA02 otherthan 0V?

Is the voltage output to TP9 on BA02 otherthan 0V?

Is the trouble resolved?

YES

YES

YES

Failure of the cable EA08.Replace it with the new one.

Failure of the laser head.Send the system back to NIDEK.

Failure of BA01.Replace it with the new one (see 4.8).

YES

Is the connector P201 on BA02 connectedcorrectly?

YES

NOCorrect the connection.

Is the trouble resolved? Failure of BA02.Replace it with the new one (see 4.7).

NO

YES

NO

Correct the connection.NO

NO

NO

Page 13: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 5

3.5 Setting of coagulation power output cannot be changedcorrectly

Perform as described in 3.3.

3.6 Setting of coagulation time cannot be changed correctlyPerform as described in 3.3.

3.7 Repeat mode cannot be set correctlyPerform as described in 3.3.

3.8 Brightness of aiming beam cannot be changed correctlyPerform as described in 3.3.

3.9 Aiming beam cannot be seenIs the connector P114 on BA01 connectedcorrectly?

Does the safety shutter ASSY block theaiming laser optical path?

Can the aiming beam be seen by turningVR1 on BA01 clockwise?

YES

YES

Adjust the safety shutter.

Failure of BA01.Replace it with the new one (see 4.8).

Is the voltage between TP12 and A. GNDon BA01 +1.2V or less?Is the aiming beam emitted from theAIMING ASSY?

YES

NO Failure of aiming laser.Replace EA04 with the new one.

Misalignment of the optical axis of theaiming beam. Send the system back toNIDEK.

YES

NO

Correct the connection.NO

NO

Page 14: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 6

3.10 Err. 3 appears

Release the interlock and error.

Is the voltage between 1st-2nd, 3rd-5th and6th-5th pins of the connector J103 onBA01 +12V?

Is the voltage between 4th and 5th pins ofthe connector J103 on BA01 +5V, whenthe shutter plate blocks the laser opticalpath?Is the voltage between these points 0V,when the shutter plate does not block thelaser optical path?

YES

Failure of BA09 of EA02.Replace it with the new one (see 4.13).

Adjust the operation of the safety shutter.

Is the connector P103 on BA01 connectedcorrectly?

YES

NO Failure of BA01.Replace it with the new one (see 4.8).

Is the safety shutter placed on the sensordetection unit?

NO

Correct the connection.NO

Failure of BA01.Replace it with the new one (see 4.8).

YES

NO

YES

Page 15: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 7

3.11 READY indication does not light up by pressing theSTATUS switch

Perform as described in 3.5.

3.12STAND BY indication does not light up by pressingthe STATUS switch

Perform as described in 3.5.

3.13Counter cannot be resetPerform as described in 3.5.

3.14Laser beam cannot be emittedIs the connector P905 on the foot switchconnected correctly?

Failure of BA01.Repair it with the new one (see 4.8).

YES

YES

YES

Failure of BA01.Replace it with the new one (see 4.8).

Repair the broken part of EA13 or replaceit with the new one.

Is the voltage between 7th and 10th pins ofthe connector J113 on BA01 +12V?

YES

NO

Is the voltage between 8th-10th and 9th-10th pins of the connector J113 on BA01+5V when the foot switch is pressed?

Failure of the foot switch.Replace it with the new one.

NO

Correct the connection.NO

NO

Is there continuity between these points onEA13?

7th pin of P113 - 1st pin of J9058th pin of P113 - 2nd pin of J9059th pin of P113 - 3rd pin of J90510th pin of P113 - 4th pin of J905

Page 16: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 8

3.15 Err. 1 appears

Release the interlock and error.

Is the voltage between 1st and 2nd pins ofthe connector J102 on BA01 +12V?

Is the voltage between 4th and 5th pins ofthe connector J102 on BA01 +5V, whenthe shutter plate blocks the laser opticalpath?Is the voltage between these points 0V,when the shutter plate does not block thelaser optical path?

YES

Failure of BA09 of EA03.Replace it with the new one (see 4.14).

Adjust the operation of the safety shutter.

Is the connector P102 on BA01 connectedcorrectly?

YES

NO Failure of BA01.Replace it with the new one (see 4.8).

Is the safety shutter placed on the sensordetection unit?

NO

Correct the connection.NO

Failure of BA01.Replace it with the new one (see 4.8).

YES

Is the voltage between 3rd and 5th, and 6thand 5th pins of the connector J102 onBA01 +12V?

NO Failure of BA01.Replace it with the new one (see 4.8).

YES

NO

YES

Page 17: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 9

3.16 Err. 10 appears

Failure of the laser head.Return the system back to NIDEK.

3.17 Err. 12 appears

Failure of the laser head.Return the system back to NIDEK.

3.18 Err. 13 appears

Failure of BA02.Return the system back to NIDEK.

3.19 Err. 15 appears

Release the interlock and error.

Adjust the overpower.

Check the actual power output at the endof the laser aperture of the main body orthe delivery system. Do they meet thespecification below?

90 mW or less: ±20%100 mW or less: ±15%

YES

Adjust the actual power output, at the endof laser aperture of the main body or thedelivery system, and coagulation poweroutput setting to meet the specification.

NO

Page 18: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 10

3.20 Err. 53 appears

3.21 Err. 54 appears

Is the fiber connected correctly?

YES

YES

NOConnect it correctly.

Failure of BA01.Replace it with the new one (see 4.8).

NO

Is the voltage between 2nd and 3rd, and5th and 6th pins of the connector J101 onBA01 +0V when the fiber is connected?

Is the voltage between 2nd and 3rd, and5th and 6th pins of the connector J101 onBA01 +5V when the fiber is disconnected?

YES

Failure of BA01.Replace it with the new one (see 4.8).

NO

Failure of the microswitch or the wire ofEA01.Repair EA01 or replace it with the new one(see 4.15).

Is the voltages between 8th and 10th, and9th and 10th pins of the connector J113 onBA01 +5V when the foot switch ispressed?

YES

YES

NO Failure of the foot switch.Replace it with the new one.

Repair the broken part of EA13 or replaceit with the new one.

NO

Failure of BA01.Replace it with the new one.

Is there continuity between these points onEA13?

7th pin of P113 and 1st pin of J9058th pin of P113 and 2nd pin of J9059th pin of P113 and 3rd pin of J90510th pin of P113 and 4th pin of J905

Page 19: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 11

3.22 INTLK2.7 appears

Release the interlock and error.

Adjust the overpower.

Check the actual power output at the endof the laser aperture of the main body orthe delivery system. Do they meet thespecification below?

90 mW or less : ±20%100 mW or less: ±15%

YES

Adjust the actual power output at the endof the laser aperture of the main body orthe delivery system and coagulation poweroutput setting to meet the specification.

NO

Page 20: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 12

3.23 INTLK4.7 appears

Release the interlock and error.

Are there any objects blocking the air inletand outlet of the system?

Is the room temperature within thespecified range for use (between 10°Cand 30°C)?

Adjust the room temperature within thespecified range for use.

Leave the system for more than 30 minutesand use the system after it is cooled down.

Does the fan work?

YES

NO Move the objects away from the air inlet ormove the system.

Can 2000 shots of the laser beamcontinuously be emitted by keeping onpressing the foot switch in the followingcondition?Coagulation power output: 500 mWCoagulation time: 0.2 sec.Interval time: 0.2 sec.

YES

NO

See 3.2 Fan does not work

The system is normal. INTLK4.7 appearsdue to the temporary temperature-riseinside the system, however, the systemcan be used because the temperatureinside the system is lowered.

YES

NO

NO

YES

Page 21: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 13

3.24 Actual power output is lowDisconnect the fiber-optic cable from thedelivery system.

Is the fiber alignment performed?

Change the setting of system parameter sothat the setting power coincide with theactual power output.

YES

Replace the fiber-optic cable with the newone.

Failure of the laser head or misalignmentof the laser head optical axis.Return the system back to NIDEK.

Is the end face of the fiber-optic cablecleaned?

YES

NOPerform the fiber alignment.

Does the setting power coincide with theactual power output?

NO

Clean the end face of the fiber-optic cable.NO

Adjust the over power.

YES

Is the transmittancy of the fiber-optic cablemore than 70%?

YES

NO

Page 22: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

3 - 14

3.25 Actual power output is high

3.26 EMERGENCY OFF button does not work

Disconnect the connector of fiber-opticcable from the delivery system.

Adjust the overpower.

Change the setting of system parameter sothat the setting power output coincide withthe actual power output.

Is the connector P508 on BA05 connectedcorrectly?

Failure of BA05.Replace it with the new one (see 4.6).

YES

Is the EMERGENCY OFF button normal?

YES

Connect it correctly.NO

Failure of the EMERGENCY OFF button.Replace it with the new one (see 4.10).

NO

3.27 Err. 4 appears

Check that EEPROM (IC8) was replaced when BA01 was replaced (see 4.8).If EEPROM is not replaced, please replace it. If it was replaced, EEPROM is faulty.Replace EEPROM. Then contact to NIDEK for the value of the system parameter and reset it.(see 5.2 and 6.6.2).

Page 23: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

§4 REPLACEMENT

4.1 Replacement of fuses

1) Unscrew PC3×4.2) Remove the stopper.3) Disconnect the power cord plug from the

outlet.4) Remove the fuse holder.5) Replace the fuses with the new one.6) Attach the removed parts in the reverse

order.

4.2 Replacement of upper cover

1) Unscrew PC3×4Cr (n=4).2) Replace the upper cover with the new one.3) Attach the removed parts in the reverse order.

Fuse

Fuse holder

Upper cover

PC3×4Cr

Fuseholder

PC3×4Stopper

Page 24: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

4 - 2

4.3 Replacement of lower cover

1) Unscrew PC3×4Cr (n=4).2) Replace the lower cover with the new one.3) Attach the removed parts in the reverse

order.

4.4 Replacement of front cover

1) Remove the cap (n=2).2) Unscrew PC3×4Cr (n=2).3) Unscrew PC3×6Cr (n=2).4) Replace the front cover with the new one.5) Attach the removed parts in the reverse order.

Lower cover

PC3×4Cr

Cap

Front cover

PC3×4Cr

PC3×6Cr

Page 25: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

4 - 3

4.5 Replacement of PS1

1) Remove the upper cover (see 4.2).2) Unscrew SB4×12 (n=3).3) Disconnect the connector from PS1.4) Replace PS1 with the new one.5) Attach the removed parts in the reverse

order.

4.6 Replacement of BA05 (SW BOARD)

1) Remove the upper cover (see 4.2).2) Unscrew PC3×4 (n=4).3) Disconnect the connector from BA05 (SW

BOARD).4) Replace BA05 with the new one.5) Attach the removed parts in the reverse

order.

4.7 Replacement of BA02 (LASER CONTROL BOARD)

1) Remove the lower cover (see 4.3).2) Unscrew PC3×4 (n=4).3) Disconnect the connector from BA02

(LASER CONTROL BOARD).4) Replace BA02 with the new one.5) Attach the removed parts in the reverse

order.

4.8 Replacement of BA01 (MASTER BOARD)

1) Remove the lower cover (see 4.3).2) Unscrew PC3×4 (n=4).3) Disconnect the connector from the

BA01(MASTER BOARD).4) Replace BA01 with the new one.5) Take CPU(IC1) and EEPROM(IC8) out from

the removed board and mount these on the newboard.

6) Attach the removed parts in the reverse order.

BA05

SB4×12

BA01

PC3×4

PC3×4

PS1

BA02

PC3×4CPU

EEPROM

Page 26: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

4 - 4

4.9 Replacement of Key control switch

1) Remove the front cover (see 4.4).2) Disconnect the connector P506 from BA05

(SW BOARD).4) Replace the Key control switch with the new

one.5) Attach the removed parts in the reverse

order.

4.10 Replacement of EMERGENCY OFF button

1) Remove the front cover.2) Disconnect the connector P508 from BA05

(SW BOARD).3) Replace the EMERGENCY OFF button

with the new one.4) Attach the removed parts in the reverse order.

Key control switch

EMERGENCY OFF button

Page 27: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

4 - 5

4.11 Replacement of fan

1) Remove the upper cover (see 4.2).2) Remove the lower cover (see 4.3).3) Disconnect the connector P207 from BA02

(LASER CONTROL BOARD).4) Unscrew TC3×4Cr (n=4).5) Remove the fan together with frame D.6) Unscrew 3N4 (n=4).7) Replace the fan with the new one.8) Attach the removed parts in the reverse

order.

4.12 Replacement of BA03 (DISPLAY BOARD)

1) Remove the upper cover (see 4.2).2) Remove the lower cover (see 4.3).3) Remove the front cover (see 4.4).4) Disconnect the all connectors from the front

panel AS.5) Unscrew PC3×4 (n=4).6) Remove the front panel.7) Unscrew N3 (n=4).8) Replace BA03 (DISPLAY BOARD) with

the new one.9) Attach the removed parts in the reverse order.

PC3×4

3N4

Front panel

BA03

TC3×4Cr

Frame D

Fan

N3

Page 28: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

4 - 6

4.13 Replacement of BA09 (SENSOR BOARD) of safetyshutter

1) Remove the upper cover (see 4.2).2) Remove the lower cover (see 4.3).3) Disconnect the connector P103 from the

BA01 (MASTER BOARD).4) Unscrew CK2×3 (n=2).5) Replace BA09 (SENSOR BOARD) of the

safety shutter with the new one.6) Attach the removed parts in the reverse

order.

4.14 Replacement of BA09 (SENSOR BOARD) ofcoagulation shutter

1) Remove the upper cover (see 4.2).2) Remove the lower cover (see 4.3).3) Disconnect the connector P102 from BA01

(MASTER BOARD).4) Unscrew CK2×3 (n=2).5) Replace BA09 (SENSOR BOARD) of the

coagulation shutter with the new one.6) Attach the removed parts in the reverse

order.

BA09

BA09

CK2×3

CK2×3

Page 29: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

4 - 7

4.15 Replacement of microswitch of CONDENSOR ASSY

1) Remove the upper cover (see 4.2).2) Remove the lower cover (see 4.3).3) Remove the front cover (see 4.4).4) Remove the front panel (see 4.12).5) Disconnect the connector P101 from BA01

(MASTER BOARD).6) Unscrew FK2×5 (n=2).7) Remove the cover (M054).8) Unscrew CK2×4 (n=2).9) Replace the microswitch together with the

sensor base (M052).10)Unscrew 82053-CC06A (n=2).11)Replace the microswitch with the new one.12)Attach the removed parts in the reverse

order.

82053-CC06A

CK2×4FK2×5

Microswitch

Sensor base

Cover

Page 30: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

4 - 8

Page 31: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

§5 ADJUSTMENT

5.1 Optical adjustment

5.1.1 Adjustment of optical axis of aiming beam

Use the DC-2 optical axis adjustment jig and adjust the inclination of the dichroic mirror AS andthe aiming mirror AS so that the position of aiming beam image coincides with the position ofLD (Laser Diode) image at the center of scale.

1. EA04 of the aiming base AS is rotatedso that the aiming beam image turns horizontally.Check it with a paper placed in front of the aiming base AS.

2. The focus of the aiming beam image is adjusted byback-and-forth position adjustment of EA04and loosening of screw G007 of EA04.

3. Adjust by the dichroic mirror AS so that the position of aiming beam image coincideswith the position of LD image. (The aiming beam image reflected on the scale turnsvertically.)

4. Remove the object lens for DC-2 optical axis adjustment jig and adjust as follows forcoinciding with the position of reflected aiming image.

1) Vertical position adjustment:Move the aiming mirror AS to the reverse direction from the displaceddirection.

2) Horizontal position adjustment:Move the aiming mirror AS to the same direction from the displaceddirection.

3) Put the object lens and adjust with the dichroic mirror AS so that theposition of aiming beam image coincides with the center of the scale.

4) Repeat steps 1 through 3, and check that it becomes as shown in the figureon the right regardless of the existence of the object lens.

Aiming laser

Dichroic mirror AS

Aiming mirror AS

Screw

G007

Rotate EA04to adjust the focus

Aiming image

Page 32: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

5 - 2

5.1.2 Adjustment of fiber alignment

1. Connect the fiber and install a power meter to the place of laser output.

2. Perform the following adjustments so that the power of the fiber output becomes maximum. • Loosen the condenser lens back-and forth position setscrew and rotate 17740-M021 of the

condenser lens ASSY (1130) (back-and-forth position adjustment of the condenser lens). • Loosen the condenser lens rotation position setscrew and rotate 17621-M020 of the condenser

lens ASSY (1130) (rotation position adjustment of the condenser lens). • Loosen the alignment adjustment setscrew and rotate 17621-M117 of the collector optics

ASSY (1300) (alignment adjustment). * Repeat position adjustment, rotation adjustment, and alignment adjustment so that the fiber

output becomes maximum. * After adjustments, tighten each setscrew and check the output again.

M021

17621-M020

1130 1300

13611-M117

Condenser lens back-and-forthposition setscrew

Condenser lens rotation po-sition setscrew

Alignment adjustmentsetscrews (n=2)

Page 33: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

5 - 3

5.1.3 Operation check of coagulation shutter sensor

1. Connect an oscilloscope as follows:ch.1→ pin 2 of J102 on MASTER BOARDch.2 → pin 5 of J102 on MASTER BOARD

2. Press the TIME switch to set the TIME indication on the display to 1.00. Press theINTERVAL switch to set the INTERVAL indication on the display to 1.0.

3. Emit the laser beam and check OPEN TIME with the waveform of the oscilloscope.4. Connect ch.2 of the oscilloscope to pin 6 of J102 on MASTER BOARD.5. Emit the laser beam and check CLOSE TIME with the waveform of the oscilloscope.6. Check that OPEN TIME and CLOSE TIME are within the following range.

Delay time (open)(msec) + delay time (close) (msec) + parameter of address 54 ≤≤≤≤≤ 93Parameter of address 54 50 to 55 (default: 50)

* If they are not within the above range, adjust the horizontal position of 17211-M028 (seethe figure below).

5.1.4 Operation check of safety shutter sensor

1. Connect an oscilloscope as follows:ch.1→ pin 2 of J103 on MASTER BOARDch.2 → pin 4 of J103 on MASTER BOARD

2. Press the TIME switch to set the TIME indication on the display to 1.00. Press theINTERVAL switch to set the INTERVAL indication on the display to 1.0.

3. Press the AIMING switch to turn on or off the aiming beam and check that OPEN TIMEis 40 msec or less with the waveform of the oscilloscope. * If it is not within the above range, adjust the horizontal position of 17211-M028 (see the

above figure).4. Connect ch.2 of the oscilloscope to pin 6 of J102 on MASTER BOARD.5. Press the AIMING switch to turn on or off the aiming beam and check that CLOSE

TIME is 40 msec or less with the waveform of the oscilloscope. * If it is not within the above range, adjust the horizontal position of 17211-M028 (see the

above figure).

ch.1

ch.2

OPEN TIME CLOSE TIME

17211-M028 17211-M028

Page 34: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

5 - 4

5.2 Electrical adjustment

5.2.1 Setting of LASER CONTROL BOARD (BA02)

1. Set VR4, VR5, VR9, VR11, VR13, and VR15 to MID.2. Turn VR3 fully to CW direction (MAX).3. Turn VR8, VR7 and VR12 fully to CCW direction (MIN).4. Remove the jumper pin, if it is attached to JP1, JP2, JP3, and JP6.5. Check that the jumper pin is attached to JP4 and JP5.

5.2.2 Setting of DISPLAY BOARD (BA03)

1. Leave No.1, No.2, and No.4 of the DIP switch as they are and set the others to the ONposition.

5.2.3 Input voltage check

Connect a multimeter to the system and check following input voltages.

1. The 1st and 2nd pins of the connector P209 on LASER CONTROL BOARD (BA02):DC12V

2. The 3rd and 4th pins of the connector P209 on LASER CONTROL BOARD (BA02):DC5V

VR5 VR1VR1

VR2

VR1

VR4

VR3

VR7 VR1VR1

VR6 VR8

VR9

VR1

TP7 TP1

TP6

TP2 TP1

TP8

TP4

TP1

TP9

TP1

TP1

TP5

TP3

Page 35: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

5 - 5

5.2.4 Adjustment of T.E.COOLER

1. Adjust VR11 so that the voltage of TP10 of BA02 becomes 2.92 V.2. Place the system in the Service Mode by setting No.8 of the DIP switch to the OFF position.3. Check the voltage of TP9 of BA02 and find the corresponding temperature among the

table below.4. Adjust VR15 of BA02 so that the S.SIZE indication shows the temperature of step 3.

5.2.5 COAG STANDBY check

1. Attach the foot switch.2. Remove the dummy laser.3. Connect a digital multimeter to pin 3 of P202 on LASER CONTROL BOARD (BA02)

and check that the voltage is changed to 0V and 12V when turning ON/OFF the TRIG.SW.

5.2.6 D/A converter signal check

Turn the power from MIN. to MAX. and check that the voltage of TP4 on LASER CONTROLBOARD (BA02) during coagulation is the same grade as the display.

5.2.7 Current limit check

* Attach the dummy laser.2.Measure the voltage of pin 4 of the connector P210 of BA02 by multimeter, and checkthat the dummy laser carries 2A when turning VR7 of BA02. Check the voltage of TP7 ofBA02 becomes the following at that time.Expression: TP7(V)=P210(A) (Declination must be within 3%.)

Temperature(℃) TP9 voltage Temperature(℃) TP9 voltage Temperature(℃) TP9 voltage0 1.5040 17 2.4427 34 3.40831 1.5515 18 2.5032 35 3.45812 1.6002 19 2.5636 36 3.50723 1.6500 20 2.6239 37 3.55564 1.7010 21 2.6841 38 3.60335 1.7531 22 2.7439 39 3.65076 1.8062 23 2.8032 40 3.69767 1.8604 24 2.8621 41 3.74458 1.9155 25 2.9204 42 3.79139 1.9715 26 2.9780 43 3.8383

10 2.0283 27 3.0348 44 3.885811 2.0859 28 3.0908 45 3.934012 2.1442 29 3.1460 46 3.983113 2.2031 30 3.2003 47 4.033514 2.2625 31 3.2537 48 4.085515 2.3223 32 3.3061 49 4.139416 2.3824 33 3.3576 50 4.1956

Page 36: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

5 - 6

5.3 General adjustment

5.3.1 Current limit adjustment

Measure the voltage of TP7 of BA02 as emitting the laser beam, and turn VR7 of BA02 to be thesame value as the limit current (limit voltage).* Check that the connector of P205 and P206 is unhooked. (Turn the monitor off.)

5.3.2 Adjustment of 1/4 wavelength constant position

Set the current mode. (DIP SW4: ON)Measure the voltage of TP2 of BA02, and turn M014 and1/4 wavelength constant to be minimum voltage.After adjustment, fix M014 by HH3X3 (see the figure onthe right).

5.3.3 Overshoot check

1. Connect the oscilloscope to TP7 of BA02, and irradiate the laser power of 400mW. Checkthat the falling edge part of TP7 waveform does not overshoot. * The oscilloscope is set to DC500 mV/DIV and measured by 5.00 msec/DIV. * Check that the connectors of P205 and P206 are unhooked. (Turn the monitor off)

2. If it overshoots, adjust VR10 of BA02 until it stops.

5.3.4 Light control adjustment

1. Attach the connector P205 and P206 to BA02. Connect the endo photo probe to the mainbody and turn ON the KEY control switch.

2. Turn ON (light mode) DIP SW4 on DISPLAY BOARD BA03.3. Adjust the POWER indication and actual power output of the main body at 50 mW with

VR6 (offset) of BA02 and those at 500mW with VR3 (gain) of BA02. Repeat the adjustmentto be the same value of the indication and the actual measurement value.

4. Change the POWER indication and check that the indication is proper comparing withthe actual power output.

HH3X321/4 wavelength constant

1130M014

Overshoot

Page 37: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

5 - 7

5.3.5 Adjustment of each monitor

1. Adjust the voltage of monitor by VR1 and VR2 on the LASER CONTROL BOARD(BA02).-Offset adjustment

Adjust VR1 (offset) on BA02 so that the voltage of TP2 on BA02 becomes to 100 mV whenthe laser of 50 mW output setting is irradiated.

-Gain adjustmentAdjust VR2 (gain) on BA02 so that the voltage of TP2 on BA02 becomes to 2000 mV whenthe laser of 1000 mW output setting is irradiated.

2. Repeat the above step 1 until the voltage of TP2 on BA01 becomes stable at 50 mW to 100mV and at 1000 mW to 2000 mV.

5.3.6 Adjustment of overpower

1. Connect the delivery unit. Set the power output to 1000mW and the exposure time to 3.00sec.

2. Set No.1 and No.2 of the DIP switch on the DISPLAY BOARD BA03 to the OFF position.3. Perform adjustment by the following procedure for the actual power of 1200 mw or 800

mW.1) Emit the laser beam and turn VR3 on BA02 so the actual power

becomes as shown in the table on the right.2) If the intermittent beep sounds are heard at random intervals,

increase the corresponding parameter. If the intermittent beepsounds are not heard, decrease the parameter (see 6.6.1).

3) Emit the laser beam and change the parameter to the point where the intermittent beepsounds are heard or not heard (see 6.6.1).

4. Emit the laser beam. Turn VR3 of BA02 so that the actual power becomes 1000mW(returned to the original state).

5. Set the power output to 50 mW and the exposure time to 3.00 sec.6. Perform adjustment by the following procedure for the actual power of 75mw or 25mW.

1) Emit the laser beam and turn VR6 of BA02 so the actualpower becomes as shown in the table on the right.

2) If the intermittent beep sounds are heard at random intervals,increase the corresponding parameter. If the intermittentbeep sounds are not heard, decrease the parameter (see6.6.1).

3) Emit the laser beam and change the parameter to the point where the intermittent beepsounds are heard or not heard (see 6.6.1).

7. Emit the laser beam. Turn VR6 of BA02 so that the actual power becomes 50 mW (returnedto the original state).

8. Set No.1 and No.2 of the DIP switch back and perform the operation check includingrecheck of the actual power. For the MP delivery, emit the laser beam of 500 mW in 60sec. and check that no error occurs.

Actual power Parameter1200 mW Address 55800 mW Address 58

Actual power Parameter75 mW Address 5625 mW Address 59

Page 38: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

5 - 8

5.3.7 Adjustment of each delivery unit

1. Wear laser goggles for DC laser.2. Connect the shot fiber to the main body.3. Measure the actual power output every 100 mW from 100 mW to 2000 mW (indication).4. Calculate: actual power output (mW) / indication (mW), and input those values every

100mW from the parameter address 30 to 49 in order.5. Connect each delivery to the main body, change the following parameters, and adjust the

power output to be within specifications.

Specifications: 50 - 90 mW → ±20%, 100 - Max mW → ±15%

S.SIZE 75 100 150 200 250 260~990

Address 100 101 102 103 104 21

S.SIZE 75 100 120 150 160 or more

Address 120 121 122 123 27

S.SIZE 60 80 100 120 130 or more

Address 130 131 132 133 27

S.SIZE 75 100 150 200 250 260~990

Address 140 141 142 143 144 28

Address 24

Address 26

S.SIZE 300 500 700 900 950~4000

Address 110 111 112 113 22

S.SIZE 100 150 200 250 300 400 410 or more

Address 150 151 152 153 154 155 20

S.SIZE 100 150 200 250 300 400 410 or more

Address 160 161 162 163 164 165 25

S.SIZE 100 150 200 250 300 400 410 or more

Address 170 171 172 173 174 175 23

S.SIZE 300 400 500 600 700 250 350 450 550 560 or more

Address 180 181 182 183 184 185 186 187 188 29

YC-Combo

BIO

SL-1600 /30SL

HAAG

Long

Short

END PHOTO

Q(hHAAG revision)

R(SL-1600 revision)

MP-1/2

MP-4

MP-5

Page 39: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

5 - 9

5.3.8 Adjustment of aiming beam

1. Connect the End Photo to the main body and set a power meter of the aiming beam.2. Set the aiming indication of the front panel to MAX.3. Adjust VR1 on BA01 so that the aiming beam comes to be between 0.4 and 0.6 mW.4. Measure the voltage between TP12 and GND on BA01. (Verify that it is not 1.5 V or

more.)5. Set the aiming indication of the front panel to MIN.6. Adjust the parameter of address 19 so that the aiming beam becomes 1/8 of MAX or less.7. Set the aiming indication on the front panel to MAX again and check whether the aiming

beam is between 0.4 and 0.6 mW. *Turn OFF (to input parameters) DIP SW 6 on DISPLAY BOARD BA03 and return it to ON

after input.8. Set the deliveries except End Photo as the following conditions: MAX: 0.4 - 0.6 mW, MIN

1/8 of MAX or less, with correction at the parameters of the table below because thepower output of those deliveries changes.

Delivery Address

HAAG 73

ENDO 74

SL-1600/30SL 75

BIO 76

YC-COMBO 77

R (SL-1600 revision) 78

MP-4 (SL-130) 80

MP-1/2 (SL-1600/30SL) 81

MP-5 (HAAG) 82

Q (HAAG revision) 83

Page 40: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

5 - 10

5.3.9 Adjustment of COAG TIME

1. Wear laser goggles for DC laser.2. Monitor the laser beam by an analyzing recorder.3. Connect TP5 and A. GND on BA02 to an oscilloscope.4. Set COAG TIME indication on the front panel to 0.2 sec. (Power is about 600 mW.)5. Adjust the parameter of address 54 so that the leading edge of the laser becomes as the

picture below.

6. Adjust the parameter of address 52 so that the GOAG TIME indication becomes 0.2 sec.(Specification: indicated value ±10%)

7. Change the COAG TIME indication on the front panel to be between 0.02 and 3.00 sec.and check whether the COAG TIME meets the specification. (Specification: indicatedvalue ±20% in the case of 0.02 to 0.09 sec., indicated value ±10% in the case of 0.10 to5.00 sec.)

8. Check the COAG TIME of the front panel, changing it from 0.02 to 5.00 sec.If a waveform is not shown by setting 0.02 sec., set the parameter of address 52 to 20. * Turn OFF (to input parameters) DIP SW 6 on DISPLAY BOARD BA03 and return it to ON

after input.

5.3.10 Adjustment of REPEAT TIME

1. Set the REPEAT TIME indication on the front panel to 0.1 sec.2. Adjust the parameter of address 53 so that the REPEAT TIME indication becomes 0.1

sec. (Specification: ±10% of indicated value)3. Change the REPEAT TIME indication on the front panel to be between 0.1 and 1.0 sec.

and check whether the REPEAT TIME meets the specification. (Specification: indicatedvalue ±10%) * Set No.6 of the DIP switch on the DISPLAY BOARD BA03 (to input parameters) to the

OFF position and set it back to the ON position after inputting.

NG example Good example

Page 41: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

§6 REFERENCES

6.1 Wiring diagram (CABLE ASSY)

Page 42: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

6 - 2

6.2 Tools

• Phillips screwdriver set• Screwdriver set• Precision screwdriver set• Hexagonal wrench set• Long-nose pliers• Nippers• Tweezers• Soldering iron• Soldering removal• Solder• Blower• Power meter (COHERENT MODEL 201)• Power meter for aiming (COHERENT MODEL202)• Alcohol mixture (Methanol 50% + ether 50%)• Kimwipe (Wipers S-200 Jyujyo Kimberly)• DC2 optical axis adjustment jig

Page 43: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

6 - 3

6.3 Error code list

6.4 Interlock code list

ErrorNO. Symptom Cause

1Operation failure of coagulation shutter

Operation failure of coagulation shutter or sensor failure

3 Operation failure of safety shutter Operation failure of safety shutter or sensor failure

4 Data failure of EEPROM Mismatch data of EEPROM

10 Failure of LD Failure of LD

11 Excessively big or small output of detector Failure of LD or detector signal breaking

12 Failure of LD Failure of LD

13 Overcurrent of LD When LD current exceeds the set value

15 Shortage output of LD Excessively small output of LD

20 Failure of temperature control of LD When LD temperature exceeds the set value

53 Failure of trigger signal circuit Failure of trigger switch or cable breaking

54 Failure of fiber signal circuit Failure of fiber switch or cable breaking

90Failure of program data

When the contents of ROM in CPU is rewrittenMismatch SUM

91Failure of EEPROM data

When the contents of EEPROM is rewrittenMismatch SUM

93 Failure of memory data Mismatch data of RAM in CPU

94 Key Lock error Operation failure of program

InterlockNO. Symptom Cause

2.7 Over power Excessibly big output of LD

4.7 Heating of device Heating of internal device

6.3 Remote interlock: OFFRemote interlock connector, or Remote interlockswitch: OPEN

Page 44: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

6 - 4

6.5 DIP SW function list

A DIP SW (SW1) list and its operation are shown below. DIP SW is usually set to the ONposition. When it is set to the OFF position, the following operation will be performed. DIP SWis on the display board.Note: Set all switches back to the ON position after the operation is finished. Even if one switchis set to the OFF position, D.P. (Decimal Point) is displayed on “COUNTER”.“ON” = 1 2 3 4 “OFF” = 1. 2. 3. 4.

DIP SW Operation/ Function Content

#1 Cancel InterlockCancel the detection of Interlock. However, laser irradiation isstopped by the hardware at the time of interlock.

#2 Cancel Error Cancel the detection of Error.

#3 Not in use

#4 Current/Light Mode switching"ON" : Light control mode"OFF" : Current control mode

#5 Continuous irradiation mode Continuous irradiation setting is available.

#6 Parameter setting mode Parameter setting is available.

#7 Not in use

#8 Service mode Laser output, and display of current, etc.

6.6 System parameter

6.6.1 Change/Setting method of value of parameter

1. Set No. 6 of the DIP switch (SW1) on the display board to the OFF position.When it is set to the OFF position, data will be displayed on the following indications.Address: TIME indication of the display panelDisplay character: COUNTER indication and S.SIZE indication on the display panelInitial value or value of parameter: POWER indication on the display panel

2. Press the switch or of TIME on the display panel to display the address to be changed.3. Change the parameter value by turning the POWER knob.4. Check the beeps by pressing the RESET switch (Completion of change and setting).5. Set No. 6 of the DIP switch (SW1) of the display board back to the ON position.

Note1)Except correction of laser output and change of the opening-and-closing timing of each shutter,the reclosing of KEY SW is required.

Note2)If the power is turned off with the DIP switch (SW1) No.6 set to the OFF position (with theparameter display after parameter change), Err91 (data failure of EEPROM) is displayed atthe time of next power-on. Be sure to turn off the power after setting No.6 to the ON position.If Err91 is displayed, set No.2 of the DIP switch (SW1) to the OFF position. After restartingthe system, reset the parameter with the procedure of 1 to 5.

Page 45: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

6 - 5

Address Displaycharacter

Initialvalue

Content

1 Adj 0 Control of adjustment mode 0: Normal mode 1: Adjustment mode

2 AL on 0 Control of laser firing sound 0: No firing sound 1: Firing sound

3 n-A 0 Not in use

4 cnt dn 10 Count down timer

5 tot ctr 0 Total counter operation stop 0: Count 1: Stop

6 n-A 0 Not in use

7 rdy oFF 2 Auto-off (min) of "READY" 0: does not become OFF

8 ctr rst 0 Memory of counter value 0: Memory 1: Clear

9 LAS RST 0 Memory of aiming value 0: Memory 1: Clear

10 idL UP 1 Offset of idle current of LD

11 Spot t 10 Waiting time of spot size operation detection during coagulation X 10 (msec)

12 Spot SnS 2 Spot size detection sensitivity (bit)

13 Err ti 0 Waiting time of re-detection of Err and INTLK (msec)

14 Err ctr 0 The number of re-detection of Err and INTLK

15 tr2 0 Input SW of trigger SW 0: 2 circuit 1: 1 circuit

16 2dl 0 Waiting time after "READY" X 10 (msec)

17 A Ld Adj 100 Soft-gain of aiming

18 EnE oFS 100 Soft-offset of POWER 100=±0

19 A Ld oFS 100 Soft-offset of aiming

20 P-1.2 Pr 100 MP-1/2(SL-1600/30SL) POWER correction factor

21 SL16 Pr 129 SL-1600 POWER correction factor

22 r Pr 119 R(SL-1600 revision) POWER correction factor

23 P-5 Pr 100 MP-5(HAAG) POWER correction factor

24 Endo Pr 100 End photo delivery POWER correction factor

25 P-4 Pr 100 MP-4(SL-130) POWER correction factor

26 bio Pr 122 Binocular indirect ophthalmoscope delivery POWER correction factor

27 HAAG Pr 130 HAAG delivery POWER correction factor

28 yccb Pr 115 Combo delivery POWER correction factor

29 q Pr 100 Q(HAAG revision) POWER correction factor

30 100 Fib 100 100mW fiber correction factor

31 200 Fib 100 200mW fiber correction factor

32 300 Fib 100 300mW fiber correction factor

33 400 Fib 100 400mW fiber correction factor

34 500 Fib 100 500mW fiber correction factor

35 600 Fib 100 600mW fiber correction factor

6.6.2 System parameter list

Page 46: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

6 - 6

36 700 Fib 100 700mW fiber correction factor

37 800 Fib 100 800mW fiber correction factor

38 900 Fib 100 900mW fiber correction factor

39 1000 Fib 100 1000mW fiber correction factor

40 1100 Fib 100 1100mW fiber correction factor

41 1200 Fib 100 1200mW fiber correction factor

42 1300 Fib 100 1300mW fiber correction factor

43 1400 Fib 100 1400mW fiber correction factor

44 1500 Fib 100 1500mW fiber correction factor

45 1600 Fib 100 1600mW fiber correction factor

46 1700 Fib 100 1700mW fiber correction factor

47 1800 Fib 100 1800mW fiber correction factor

48 1900 Fib 100 1900mW fiber correction factor

49 2000 Fib 100 2000mW fiber correction factor

50 ti Lo 2 Coagulation time minimum value X 10 (msec)

51 ti Hi 500 Coagulation time maximum value X 10 (msec)

52 coAG ti 8 Coagulation time correction value (+msec)

53 rEP ti 30 Repeat time correction value (+msec)

54 SS oP 50 Laser output overshoot waiting time (from laser-on to shutter-open) (msec)

55 in27 GAin 16 Adjustment of INTLK2.7 (Gain +%)

56 in27 oFS 7 Adjustment of INTLK2.7 (Offset +mW)

57 in27 ti 50 Start time of over power check (msec)

58 Er15 GAin 17 Adjustment of Err15 (Gain -%)

59 Er15 oFS 25 Adjustment of Err15 (Offset -mW)

60 P Lo 5 Coagulation output minimum value X 10 (mW)

61 P Hi 200 Coagulation output maximum value X 10 (mW)

62 Ld P St 150 Maximum output of LD

63 Ld c St 130 Maximum current of LD

64 Ld t Sb 10 Setting value of t.CAL (℃)

65 Ld t St 25 Temperature of LD (℃)

66 Ld t Lo 77 Adjustment of Err20 Minimum value of LD temperature (77 : 0℃)

67 Ld t Hi 200 Adjustment of Err20 Maximum value of LD temperature (200 : 45℃)

68 ti H ttt 600 TTT coagulation time maximum value (sec)

69 PrE LSr 0 Waiting time before laser firing (msec)

70 SL Hi P 150 Slit lamp delivery maximum output X 10 (mW)

Page 47: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

6 - 7

71 ShuT oP 30 Errl detection waiting time (coagulation shutter open) (msec)

72 ShuT cL 30 Errl detection waiting time (coagulation shutter close) (msec)

73 HAAG A Pr 225 HAAG delivery aiming output correction factor

74 Endo A Pr 100 Endo photo delivery aiming output correction factor

75 SL16 A Pr 110 SL-1600 aiming output correction factor

76 bio A Pr 200 Binocular indirect ophthalmoscope delivery aiming output correction factor

77 yccb A Pr 250 Combo delivery aiming output correction factor

78 r A Pr 100 R (SL-1600 revision) aiming output correction factor

79 Ld oFF 0 Current detection limit when LD is off

80 P-4 A Pr 100 MP-4 (SL-130) aiming output correction factor

81 P-1.2 A Pr 100 MP-1/2 (SL-1600/30SL) aiming output correction factor

82 P-5 A Pr 100 MP-5 (HAAG) aiming output correction factor

83 q A Pr 100 Q(HAAG revision) aiming output correction factor

84 n-A 100 Not in use

85 n-A 100 Not in use

86 n-A 100 Not in use

87 n-A 100 Not in use

88 n-A 100 Not in use

89 n-A 100 Not in use

90 Spot Lt 200 MP delivery spot size correction setting (um)

91 Spot oFS 40 MP delivery offset output correction

92 3rd Pr 0 MP delivery power monitor 0: Effective 1: Ineffective

93 ovEr GAin 30 MP delivery over power adjustment (Gain +%)

94 ovEr oFS 50 MP delivery over power adjustment (Offset +mW)

95 undr GAin 30 MP delivery under power adjustment (Gain -%)

96 undr oFS 50 MP delivery under power adjustment (Offset -mW)

97 3rd ti 70 MP delivery power monitor detection waiting time (msec)

98 n-A 100 Not in use

99 Ai ON-OF 0 External ON/OFF function of aiming beam 0: Ineffective 1: MP effective

100 SL16 75u 185 SL-1600/30SL coagulation output correction factor (spot size = 75um)

101 SL16 100u 140 SL-1600/30SL coagulation output correction factor (spot size = 100um)

102 SL16 150u 117 SL-1600/30SL coagulation output correction factor (spot size = 150um)

103 SL16 200u 104 SL-1600/30SL coagulation output correction factor (spot size = 200um)

104 SL16 250u 100 SL-1600/30SL coagulation output correction factor (spot size = 250um)

105 n-A 100 Not in use

Page 48: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

6 - 8

106 n-A 100 Not in use

107 n-A 100 Not in use

108 n-A 100 Not in use

109 n-A 100 Not in use

110 r 300u 200 R(SL-1600 revision) coagulation output correction factor (spot size = 300um)

111 r 500u 135 R(SL-1600 revision) coagulation output correction factor (spot size = 500um)

112 r 700u 120 R(SL-1600 revision) coagulation output correction factor (spot size = 700um)

113 r 900u 105 R(SL-1600 revision) coagulation output correction factor (spot size = 900um)

114 n-A 100 Not in use

115 n-A 100 Not in use

116 n-A 100 Not in use

117 n-A 100 Not in use

118 n-A 100 Not in use

119 n-A 100 Not in use

120 G LG 75u 120 HAAG(Long) coagulation output correction factor (spot size = 75um)

121 G LG 100u 105 HAAG(Long) coagulation output correction factor (spot size = 100um)

122 G LG 120u 100 HAAG(Long) coagulation output correction factor (spot size = 120um)

123 G LG 150u 100 HAAG(Long) coagulation output correction factor (spot size = 150um)

124 n-A 100 Not in use

125 n-A 100 Not in use

126 n-A 100 Not in use

127 n-A 100 Not in use

128 n-A 100 Not in use

129 n-A 100 Not in use

130 G St 60u 120 HAAG(Short) coagulation output correction factor (spot size = 60um)

131 G St 80u 105 HAAG(Short) coagulation output correction factor (spot size = 80um)

132 G St 100u 100 HAAG(Short) coagulation output correction factor (spot size = 100um)

133 G St 120u 100 HAAG(Short) coagulation output correction factor (spot size = 120um)

134 n-A 100 Not in use

135 n-A 100 Not in use

136 n-A 100 Not in use

137 n-A 100 Not in use

138 n-A 100 Not in use

139 n-A 100 Not in use

140 yccb 75u 150 COMBO coagulation output correction factor (spot size = 75um)

Page 49: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

6 - 9

141 yccb 100u 125 COMBO coagulation output correction factor (spot size = 100um)

142 yccb 150u 110 COMBO coagulation output correction factor (spot size = 150um)

143 yccb 200u 105 COMBO coagulation output correction factor (spot size = 200um)

144 yccb 250u 100 COMBO coagulation output correction factor (spot size = 250um)

145 yccb 300u 100 COMBO coagulation output correction factor (spot size = 300um)

146 yccb 400u 100 COMBO coagulation output correction factor (spot size = 400um)

147 yccb 500u 100 COMBO coagulation output correction factor (spot size = 500um)

148 n-A 100 Not in use

149 n-A 100 Not in use

150 P-1.2 100u 100 MP-1.2 coagulation output correction factor (spot size = 100um)

151 P-1.2 150u 100 MP-1.2 coagulation output correction factor (spot size = 150um)

152 P-1.2 200u 100 MP-1.2 coagulation output correction factor (spot size = 200um)

153 P-1.2 250u 100 MP-1.2 coagulation output correction factor (spot size = 250um)

154 P-1.2 300u 100 MP-1.2 coagulation output correction factor (spot size = 300um)

155 P-1.2 400u 100 MP-1.2 coagulation output correction factor (spot size = 400um)

156 n-A 100 Not in use

157 n-A 100 Not in use

158 n-A 100 Not in use

159 n-A 100 Not in use

160 P-4 100u 100 MP-4(SL-130) coagulation output correction factor (spot size = 100um)

161 P-4 150u 100 MP-4(SL-130) coagulation output correction factor (spot size = 150um)

162 P-4 200u 100 MP-4(SL-130) coagulation output correction factor (spot size = 200um)

163 P-4 250u 100 MP-4(SL-130) coagulation output correction factor (spot size = 250um)

164 P-4 300u 100 MP-4(SL-130) coagulation output correction factor (spot size = 300um)

165 P-4 400u 100 MP-4(SL-130) coagulation output correction factor (spot size = 400um)

166 n-A 100 Not in use

167 n-A 100 Not in use

168 n-A 100 Not in use

169 n-A 100 Not in use

170 P-5 100u 100 MP-5(HAAG) coagulation output correction factor (spot size = 100um)

171 P-5 150u 100 MP-5(HAAG) coagulation output correction factor (spot size = 150um)

172 P-5 200u 100 MP-5(HAAG) coagulation output correction factor (spot size = 200um)

173 P-5 250u 100 MP-5(HAAG) coagulation output correction factor (spot size = 250um)

174 P-5 300u 100 MP-5(HAAG) coagulation output correction factor (spot size = 300um)

175 P-5 400u 100 MP-5(HAAG) coagulation output correction factor (spot size = 400um)

176 n-A 100 Not in use

177 n-A 100 Not in use

Page 50: Model DC-3300 - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/ophthalmology... · Model DC-3300 SERVICE MANUAL December, ... 6.3 Error code list ... •

6 - 10

178 n-A 100 Not in use

179 n-A 100 Not in use

180 q LG 300u 100 Q(HAAG revision) Long coagulation output correction factor (spot size = 300um)

181 q LG 400u 100 Q(HAAG revision) Long coagulation output correction factor (spot size = 400um)

182 q LG 500u 100 Q(HAAG revision) Long coagulation output correction factor (spot size = 500um)

183 q LG 600u 100 Q(HAAG revision) Long coagulation output correction factor (spot size = 600um)

184 q LG 700u 100 Q(HAAG revision) Long coagulation output correction factor (spot size = 700um)

185 q St 250u 100 Q(HAAG revision) Short coagulation output correction factor (spot size = 250um)

186 q St 350u 100 Q(HAAG revision) Short coagulation output correction factor (spot size = 350um)

187 q St 450u 100 Q(HAAG revision) Short coagulation output correction factor (spot size = 450um)

188 q St 550u 100 Q(HAAG revision) Short coagulation output correction factor (spot size = 550um)

189 n-A 100 Not in use

190 n-A 0 Not in use

191 n-A 0 Not in use

192 n-A 0 Not in use

193 n-A 0 Not in use

194 n-A 0 Not in use

195 n-A 0 Not in use

196 n-A 0 Not in use

197 n-A 0 Not in use

198 n-A 0 Not in use

199 EEPr in 0 Initial setting of EEPROM 99: Initial setting