Professionally built - Easy to install - Rugged - Compact Design The Model SS-TEG is a sample system designed to present a sample of Natural Gas to the Sensor/Transmier of a moisture analyser when TEG or heavy hydrocarbon condensate may be present, typically downstream of a glycol contactor. The system includes parcle filtering, pressure control and flow control to condion the sample for presentaon to the moisture sensor, which is incorporated into the system. It can either be configured so that the pressure is reduced upstream of the sensor for measuring atmospheric pressure dewpoint or moisture concentraon (SSTEG), or so that the pressure is reduced downstream of the sensor for disposal aſter measuring the dewpoint at sample pressure (SSNGH-P). TEG Natural Gas Dewpoint Sample Systems Model SS-TEG 1 OF 2 © Alpha Moisture Systems Ltd. FEATURES • Incorporates ATEX cerfied heated pressure regulators • Inlet pressure up to 210 barg • Sensor operates at inlet or atmospheric pressure • Process and auxiliary inlet with manual selector valves • Sample inlet and outlet pressure indicaon • Electrically heated pressure regulators for control of sample For all Ex Applicaons the display is installed into the safe area/zone with a Zener Barrier or Ex Isolator. DS1200-AMT-Ex DS2000 Model DS2000 like the AMT-Ex, can be used with the Model SS-TEG dewpoint sample systems. Various ranges are available between -110 0 C to +20 0 C Td. This unit is also perfect for connuous on-line dewpoint measurement in clean, dry Natural Gas. For more details see Data Sheet DS2000-Sensor. Model DS1200-AMT-Ex intrinsically safe dewpoint hygrometer can be used with the Model SS-TEG sample systems. This hygrometer incorporates a 4-20mA transmier with various ranges available between -110 0 C to +20 0 C Td. The AMT-Ex has Auto-Calibraon as standard. For more details see Data Sheet DS1200-AMT-Ex. Recommended Dewpoint Hygrometers Model 6020 Model 6020 Hygrometer is the next level of online dewpoint hygrometers. Based on the well proven, industry leading, ultra-high capacitance aluminium oxide sensor, the Model 6020 is fully loaded with features and funcons to make it a most versale addion to the Alpha Moisture Systems range of trace moisture dewpoint analysers. For more details see Data Sheet Model 6020

Upload: others

Post on 03-Oct-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Model SS‐TEG

Professionally built - Easy to install - Rugged - Compact Design

The Model SS-TEG is a sample system designed to present a sample of Natural Gas to the Sensor/Transmitter of a moisture analyser whenTEG or heavy hydrocarbon condensate may be present, typically downstream of a glycol contactor.

The system includes particle filtering, pressure control and flow control tocondition the sample for presentation to the moisture sensor, which is incorporated into the system. It can either be configured so that the pressure is reduced upstream of the sensor for measuring atmospheric pressure dewpoint or moisture concentration (SSTEG), or so that the pressure is reduced downstream of the sensor for disposal after measuringthe dewpoint at sample pressure (SSNGH-P).

TEG Natural Gas Dewpoint Sample SystemsModel SS-TEG

1 OF 2© Alpha Moisture Systems Ltd.

FEATURES

• Incorporates ATEX certified heated pressure regulators

• Inlet pressure up to 210 barg• Sensor operates at inlet or atmospheric

pressure

• Process and auxiliary inlet with manual selector valves

• Sample inlet and outlet pressure indication

• Electrically heated pressure regulators for control of sample

For all Ex Applications the display is installed into the safe area/zone with a Zener Barrier or Ex Isolator.

DS1200-AMT-Ex

DS2000

Model DS2000 like the AMT-Ex, can be used with the Model SS-TEG dewpointsample systems. Various ranges are available between -1100C to +200C Td. This unit is also perfect for continuous on-line dewpoint measurement in clean, dry Natural Gas. For more details see Data Sheet DS2000-Sensor.

Model DS1200-AMT-Ex intrinsically safe dewpoint hygrometer can be used with theModel SS-TEG sample systems. This hygrometer incorporates a 4-20mA transmitter with various ranges available between -1100C to +200C Td. The AMT-Ex has Auto-Calibration as standard.For more details see Data Sheet DS1200-AMT-Ex.

Recommended Dewpoint Hygrometers

Model 6020

Model 6020 Hygrometer is the next level of online dewpoint hygrometers. Based on the wellproven, industry leading, ultra-high capacitance aluminium oxide sensor, the Model 6020 is fullyloaded with features and functions to make it a most versatile addition to the Alpha MoistureSystems range of trace moisture dewpoint analysers. For more details see Data Sheet Model 6020

Page 2: Model SS‐TEG

800 (31.5�)

838 (33�)

300 (11.8�)

850

(33.

46�)

1000

(39

.37�

)

Alpha Moisture Systems Model SS-TEG Sample Systems

How to Order

SpecificationsWEATHER PROOFING: IP66 ratingPRESSURE REGULATOR: 210bar (3,045 PSI) MaximumSENSOR HOLDER: 316 Stainless SteelPRESSURE GAUGES: Inlet, Sample and Bypass as standardPIPEWORK: All Stainless SteelFITTINGS: All Stainless SteelDIMENSIONS (HxWxD): Approx 1000 x 840 x 300mmWEIGHT: approx 75kg (165 pounds)

RELATED PRODUCTSStandard SS-NGH Sample System or Made to Order

2 OF 2

1888 Model SS-TEG pd130618-Iss-3

Standard or Made to Measure Natural Gas TEG Sample Systems

Model SS-TEG Drawing Schematic Example

Please call us to discuss your exact requirements with our experienced Technical Sales Engineers.Dewpoint measurement requirements within the Natural Gas Sector can be complex and applications differ

greatly from site to site. We offer a “Make to Order” service for all our Sample Systems.

Please Call +44 (0)1274 733 100 or Email: [email protected] to discuss your requirements.

Front Elevation Side Elevation

The Model SS-NGH Dewpoint Sample System has been designed for samplingNatural Gas where there is no risk of contamination by glycol.

The Model SS-NGH is a standard sampling system but can be modified to suit customer’s requirements and complex applications.

The heated regulators incorporated in the system can be installed for 115VAC or 230VAC operation, as specified on order.

© Alpha Moisture Systems Ltd.

Registered in England and Wales No. 3902302VAT Registration No. GB607207563

WEEE Producer Registration No. WEEE/EA0067TX

CERTIFICATE No. FM35600BS EN ISO 9001:2015

Alpha Moisture Systems Limited. Registered Office: Alpha House,

96 City Road, Bradford, BD8 8ES. UK.

Product specification may be subject to change.

Tel +44 (0) 1274 733100

Fax +44 (0) 1274 733200

Email [email protected]

Website www.amsystems.co.uk