MONALISAan Update

David Urner

Paul Coe

MatthewWarden

Armin Reichold

Monitoring, Alignment & Stabilisation with high Accuracy

Nanobeam'08 D. Urner 2

MONALISA

• Is an interferometric metrology system for continuous monitoring of position critical accelerator components

• Consists of a fixed network of evacuated interferometric distance meters

Nanobeam'08 D. Urner 3

Concepts

Nanobeam'08 D. Urner 4

QD0

Shint

ake

Mon

itor

Table

Compact Straightness MonitorVery Schematic View of ATF2 Setup

10cm

Nanobeam'08 D. Urner 5

Compact Straightness Monitor

• 6D position transferred from left to right– breaking of symmetries is important

• Preliminary simulation results of CSM Resolution: – y:10nm – distance meter resolution: 1nm = Resolution in z-direction– Positional change of optics components with respect to each other: 1nm. That’s the challenge!

10cm

Nanobeam'08 D. Urner 6

Measurement lines

We measure distances along measurement lines using two techniques:•Absolute distance interferometry <m resolutions•Displacement interferometry nm resolutions

Each line is the same, and is capable of performing both types of measurement.

Absolute distanceDisplacement

Nanobeam'08 D. Urner 7

Interferometer operation

Intensity

Nanobeam'08 D. Urner 8

Interferometer operation

Phase = 2π (Optical Path Distance) / Wavelength

Φ = 2π D / λ = 2π D (ν / c)

D = (c/ 2π) (ΔΦ/Δnu)

R = (c/ 2π) (Δθ/Δnu)D = R (ΔΦ/Δθ)

ΔD = (c/2π ν) ΔΦ

Fixed Frequency Interferometry

Frequency Scanning Interferometry

frequeny scanning

Nanobeam'08 D. Urner 9

Distance meter

• Measurement Frequencies:– FFI: up to 10kHz– FSI: up to 1Hz

• Long term stability determines low frequency behaviour– Minutes possible– Lot of work needed to extend to hours or days.

• Advantage of interferometric measurement system is fairly low cost per line.– Use of telecom frequency allows use of cheap commercial

hardware– Cheap amplification of light– Current estimate: as low as £800 per distance metre

Nanobeam'08 D. Urner 10

Current Status

Nanobeam'08 D. Urner 11

Vacuum System

8 way fibre ribbon

Tapered hole

Vacuum vessel wall

Nanobeam'08 D. Urner 12

First Measurements

in Vacuum

• Unexpected mechanical behaviour

• Tensioning of a drum – more mechanical

vibrations in vacuum– Hard to separate these

from resolution effects

• Need for vibration isolation and possibly damping

Cha

nge

in L

engt

hM

easu

red

Fre

quen

cy

Pressure1 Atm vac

Vacuum features

air

Atm Vac

Nanobeam'08 D. Urner 13

Atm vac

Changing Pressure

• Good agreement between FSI and FFI

• Decent correlation between pressure and measured OPD

Displacement (microns)

-0.2 0 0.2

Calculated overall change: 69m

Nanobeam'08 D. Urner 14

FSI• Dominated by actual vibrations!

– Need damping

– Need better launch head

Nanobeam'08 D. Urner 15

Compact Launch Head

• Launch Head needs following features:– Compact– Stable/rugged

• Temperature• Vibrations

– Cheap• Prototype:

– Adjustable– Currently being built

• Future: Jig to position– Glue components onto

substrate

Titanium

Nanobeam'08 D. Urner 16

Fixed Frequency Interferometry• Improvement of amplifier to reduce temperature

dependence• Test of Amplifier for White Noise behavior:

– Linear with laser power– With high laser power translates to 1pm/√Hz

• Spectrum above 1kHz seems to be white noise dominated

• Spectrum below 1kHz clearly vibration dominated– Vibration Isolation– Stable launch head

• Eventually see effects from– Air turbulences– Laser frequency instabilities

0 1 2 3 4 5 Frequency [kHz]

Nanobeam'08 D. Urner 17

Frequency Stabilisation• Lock laser to spectral feature of rubidium• Use a frequency doubling crystal to reach this

frequency

Nanobeam'08 D. Urner 18

Frequency Stabilisation• All parts are at Oxford• Mechanical assembly in progress

Nanobeam'08 D. Urner 19

Old design

Vacuum System at KEK

Nanobeam'08 D. Urner 20

Double bellow system: - Advantage:

- smaller frame - Disadvantage:

- active feedback- slightly less good compensation of forces

New Design

Vacuum System at ATF2

Nanobeam'08 D. Urner 21

Double bellow system:

OpticsVacuum

Compensation Chamber2 Atmosphere pressure

Active feedback: - measure pressure in inner and outer bellow chamber - stabilize sum of pressures

+

Valve to air

Valve to pressured air

<< >>

Nanobeam'08 D. Urner 22

Summary

• Very first measurements in vacuum

• Compact Launch Head

• Frequency Stabilization

• Vacuum System for ATF2