Special features
− 100% quality control of thechecking process
− Powerful algorithms foracquiring and assessingswitching points and lockingoperations
− Universal twin-channel amplifierfor many of the usual marketsensors with TEDS sensordetection
− Memory function for results,curves and device settings
− Convenient integration into theautomation system by means ofstandardized fieldbus interfaces
MP85A−SMP85ADP−SEASYswitch
μPA
D
Intelligent signal conditioning andpowerful algorithms for acquiringand assessing switching points
and locking operations
Keyboard and display
Carrier-frequency excitationvoltage 4.8 kHz
Switching testinputs, controlinputs andcontrol outputs
24 VVoltage supply,electrically isolated
Profibus DPinterface(for MP85ADP−S)
AD
MMC memorycard
Power supply for active transducers5 V or 24 V
SSItrans-ducer
�10 V
CANopeninterface
MP85A−S / MP85ADP−S
FAST Ethernetinterface
EEPROM
B2359-2.0 en
Dat
a S
heet
![Page 2: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/2.jpg)
DescriptionSwitch checking device for production and laboratoryWith EASYswitch, the switching characteristics of the switching elements and locking systems are recorded bymeans of force/displacement or torque/angle of rotation sensors. Tactile feedback when operating push buttonsand rotary actuators, as well as the switching characteristics of single pole or multipole voltage switches can beobjectified and documented with EASYswitch.With the MP85A−S EASYswitch, quality assurance can be integrated into the checking process. Operationalreliability, economic efficiency and product liability are of paramount importance.EASYswitch includes all the functions necessary for 100% checking, together with user-friendly, flexible softwarethat allows the user to configure variable test sequences without in-depth programming knowledge.
Performance features and advantages:• Easy configuration and starting up with free parameterization and visualization software (PME Assistant)
Download from www.hbm.com − Support − Software• Precise process analysis to meet requirements in compliance with standards• Flexible system for monitoring different workpieces, 1000 different device settings and 1000 different
workpieces/processes• Storage of results, curves and statistics, as well as the device settings in the device itself, on a memory card
or on an external PC• Continuous traceability thanks to integrated process control and the statistics functions of the stored processes• Integration via digital inputs/outputs or integrated fieldbus interfaces to primary control systems, such as a PLC
control or process control systems• Flexible application, tailored to use at manual workstations• Expansion of existing machines, easy to install
A) Haptic test of a locking system by means of force/displacement monitoring
Evaluation criteria:
1. Force at mechanical switching peak Fa
2. Drop in force after mechanical switching peak Fa−Fr
3. Click ratio (Fa−Fr) / Fa x 100%
4. Displacement differential Sr−Sa
5. Force/displacement ratio (Fa−Fr) / (Sr−Sa)
F /
N
S / mmSa Sr
Fq
Fr
Fa
Fr
1.
2.
4.
5.
B2359-2.0 enHBM 2
![Page 3: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/3.jpg)
B) Switch test of electrical switching systems
Solutions for test functionsAs many as 5 switching events can be electrically verified simultaneously or consecutively, via the integrated digitalinputs:
− Is the switching point or the switching process in the switching test window ?− Is the direction of actuation (break/make) correct ?− Is the actuating force / actuating moment correct?− Have there been switching repetitions?
Measurement curves and results are recorded and evaluated online. If one or both conditions are not met, an NOKwindow is activated.
OK / NOK testing controls production and helps to minimize downtime. Warning limits monitor the productionprocess and make it possible to protect the machines. Control signals control start/stop and limit value cutoff eithervia digital inputs, the Ethernet interface or the Profibus-DPV1 interface (optional).All the individual and cumulative results, switching curve curves and switching events (break/make function) arestored.
MP85A(DP)−S function blocks
Setting Measuring
TEDStransducerand amplifier
Evaluationcriteria
Loading/Storing
Data storage Specialfunctions
Measuring/Visualization
Immediate utilization of evaluation criteria• Sensor data does not need to be set manually when using sensors with TEDS technology• The TEDS data is read via the sense leads from the sensor (instead of an additional cable). The cable and the
connector can be deployed as usual.• The MP85A−S is ready for use in a matter of seconds
B2359-2.0 en HBM3
![Page 4: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/4.jpg)
Process analysis with tolerance windows / measurement and visualization
The following windows are used to evaluate the production process:
1 alarm window Limits at which an alarm is triggered. This window is used to protectthe machine.
1 range window Defines the range in which measured values are stored, within which all the other tolerance windows are positioned.
1...9 tolerance windows for analyzing the switching and locking actions. In switch check mode, the switching contacts are evaluated via the digital inputs. Haptic windows monitor the switching characteristic.With standard tolerance windows, there is a free choice of incoming and outgoing sides;the windows can be evaluated in real time. Mean value windows are also possible. All window types can be used as required and can overlap.
Thread-in window:Start window operates in real time (to protect the machines and the system)
Process window:Monitors the force/displacement (or torque/angle of rotation) curve of the switch or locking system
Thresholds:Thresholds monitor the process curve in defined ranges.
Switch check window:Monitors the function and position of the electrical contact of a switch. As many as 5 switchingcontacts can be checked in one procedure.
Block window:The end window monitors the block force.
Haptic window:Monitors the switching characteristic of the switch or the locking system.
B2359-2.0 enHBM 4
![Page 5: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/5.jpg)
Systems of coordinates:Tolerance window coordinates can be defined absolutely or relatively (dynamically). Use the relative system ofcoordinates if the absolute position of the workpieces/test specimens is not always the same. It is possible to mixthese two windows.
Application:
Analysis with tolerance windowsWith relative x coordinates, all that is measured is the relative movement as from the start or end position, relativeto the x axis of the two test pieces.
Start and stop conditionsThe start and stop conditions are used to synchronize measurement control to the checking process. The signalingoptions are via: CANopen, PROFIBUS-DPV1, digital input or internal trigger.Start and stop conditions are available for a wide variety of applications, such as:
• external start and stop signal• target value y and overshoot time• target value x and overshoot time• target value x and target value y• standstill recognition• Stop on reverse motion for channel x
Standard operator panels (IPCs) can be used to visualize the process in situ. The device can also be integratedinto existing systems later on. Connection is made via the CAN Profibus-DPV1 or FASTEthernet interface of theMP85A-S devices.
B2359-2.0 en HBM5
![Page 6: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/6.jpg)
Quality control / statistics
Quality and tool wear for the checking process areassessed using statistics functions.
Statistics graphics can be called to clearly displaythe OK and NOK processes.
Global statistics with a process counter aregrouped by parameter sets.
The tolerance window result can be read at aglance for each parameter set.
This allows
− tool wear
− component tolerances or
− damage to a machine
to be detected early.
The graphic windows in Counting can be used toanalyze the distribution of the OK and NOKprocesses individually for each window.
Counting automatically calculates the distributionof the minima and maxima with accompanyingstandard deviation.
Statistics data are stored in the device.
Data management / loading and storing
With the MP85DP−S, it is possible to store andclearly visualize results, curves, statistics and thedevice settings. This allows processes to beanalyzed later on and ensures 100% traceability.
You can choose whether to store the data on yourPC or on the multimedia card in the device.
Storage on an MMC card can be set up as a circularbuffer for the last 1000 curves.
In both cases, curves and/or results can be storedin ASCII or Qdas format:
− NOK processes only or
− OK processes only or
− all processes
The data on the MMC card can then be transcribedto the PC.
A protocol providing all information in detail can beprinted out for every process, if needed.
B2359-2.0 enHBM 6
![Page 7: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/7.jpg)
Management of production data
Ethernet
MP85ADP−Sproductionmonitoring
The curve and result data can be storedwith a component part number. This al-lows easy and unambiguous archiving.
Directory/componentpart number
Internal MMC
Directory/componentpart number
Utilizing the existing infrastructure
Using the standard Ethernet interface enables networ-ked systems with remote maintenance to be built up.
SpecificationsBasic device MP85A / MP85A DPAccuracy class 0.1
Supply voltage VDC 24; electrical isolation from the measurement system (typically500 VDC)
Permissible supply voltage range V 18...30
Power consumptionMP85A, typicallyMP85ADP, typicallyMP85A / MP85ADP, max.
WWW
7910
Behavior in the event of a supply voltage failure Automatic data retention after power failure
Typical backup battery life (CR2032) for the realtime clock years 5
Evaluation unit specificationsMax. number of measured value triplets
(channel x), (channel y), (time)4000 (automatic data reduction)
Sampling rate Hz 2400
Start conditions Internal start signalExternal start signal
Target value xTarget value y
Target value x + target value y
Stop conditions Internal stop signalExternal stop signal
Target value y + overshoot timeTarget value x + overshoot time
Target value x + target value y + overshoot timeStandstill recognition
Return channel x
Process-end conditions External signalSimultaneous with end of start condition
Target value xTarget value y
Target value x and target value y
Number of parameter sets / measurement programs in the device 31 plus factory setting
Number of parameter sets on the optional MMC 31 in XML format
1000 in binary formatTypical switching between parameter sets ms 200
B2359-2.0 en HBM7
![Page 8: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/8.jpg)
Evaluation
Tolerance band 64 interpolation points, adjusted as required
Envelope curve 4 tolerance ranges, 64 interpolation points, adjusted as required
Tolerance windows, maximum number 9
Type of window oblique or straight
Evaluation methods per window Switch check
Haptic test
Real time evaluation (online for machine protection)
Analyzing the course of the curve in the window (min/max)
Analyzing the mean x or y value in the window
Analyzing vertical or horizontal thresholds (online)
x coordinates for the tolerance window absolute or relative to the start position, orrelative to the end position
y coordinates for the tolerance window absolute or relative to Fmin of tolerance window 2,relative to Fmax of tolerance window 2 orrelative to Fmean of tolerance window 2
Typical duration of offline evaluation, end window ms 6
Typical duration of offline evaluation, straight window ms 5 + 0.1/measurement pair in window
Typical duration of offline evaluation, oblique window ms 10 + 0.3/measurement pair in window
Statistics (separately for each parameter set)
Maximum number of test cycles 4 x 109
Number of histogram classes for 2 values(xmax, xmin, ymax, ymin)
9 per tolerance window
Transducer and amplifierCarrier frequency kHz 4.8 �1 %
Bridge excitation voltage Veff 2.5 �5 %
Transducers that can be connectedStrain gages, half and full bridgesInductive half and full bridges, LVDTsPotentiometric transducers
ΩmHΩ
170 ... 20004 ... 160
170 ... 2000
Input sensitivities Measuring range (mV/V)
4 100 1000
mV/V 0.2 ... 4 3.5 ... 100 50 ... 1000
Length of transducer cable, max. m 500
Scaling range, max. digits 999999, at 10% of the input measuring range
Scaling range, min. digits 100, at 100% of the input measuring range
Permissible common-mode voltage, max. V �5.5
Common-mode rejection0...60 Hz0...1000 Hz0...4800 Hz
dBdBdB
�120�96�50
Linearity error % � 0.03
Noise voltage, typically Measuring range (mV/V)
4 100 1000
0...1 Hz0...10 Hz0...100 Hz0...1000 Hz
μV/VppμV/VppμV/VppμV/Vpp
0.10.25
12
2.562550
2560250500
Sampling rate, max. 1/s 2400
B2359-2.0 enHBM 8
![Page 9: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/9.jpg)
Measurement frequency range, adjustable
4th order low-pass with Bessel characteristic
Nominal(rated)
value fc(Hz)
−1dB(Hz)
−3dB(Hz)
Phasedelay(ms)
Rise time(ms)
Overshoot(%)
1000 980 1400 0.550 0.260 4
500 440 690 0.860 0.510 1.5200 190 320 1.6 1.11 1.5100 100 160 2.9 2.13 1.350 51 83 4.6 4.24 1
20 25 41 8.2 8.36 110 13 21 15.5 16.8 05 6.1 10.3 30.2 33.4 0
2 3.1 5.2 60 67 01 1.6 2.6 119 137 0
0.5 0.79 1.30 240 272 0
0.2 0.19 0.32 950 1070 00.1 0.09 0.16 2500 2170 00.05 0.049 0.081 3750 4280 0
Shunt calibration mV/V 1�3%
Effect of operating voltageon zero pointon sensitivity
% f.s.% f.s.
� 0.01� 0.01
Effect of 10 K variation in ambient temperature Measuring range (mV/V)4 100 1000
on zero point full bridgeon zero point half bridgeon sensitivity
μV/VμV/V
%
110
0.05
2040
0.05
2002000.05
Long-term drift over 48 h(measuring range 4 mV/V; 0.5 h after activation)
μV/V 2
DC-voltage transducersTransducers that can be connected DC-voltage transducers, voltage sourcesNominal (rated) measuring range V �10Input signal range V �10.5Scaling range, max. digits 999999, at 10% of the input measuring rangeScaling range, min. digits 100, at 100% of the input measuring rangeInternal resistance of the signal source kΩ �1Permissible common-mode voltage, max. V 2Measurement frequency range, adjustable (−1 dB) Hz 0.05 ... 1000Filter characteristics Bessel, 4th orderLinearity error % �0.03Sampling rate, max. 1/s 2400
Incremental transducersTransducers that can be connected Incremental transducers (up/down counter with zero index
signal)
Voltage supply 5 V, max. 150 mA or 24 V, max. 300 mA
2-channel mode Time-division multiplex methodInputs (F1 (�), F2 (�), Ix (�)) Differential inputs (RS422)
Input levelsLow levelHigh levelEach line to measurement ground, max.
VVV
� 0.8�2�14
Hysteresis V 0.07
Permissible common-mode voltage, max. V −7 / +12
Input impedance, typically kΩ 12Detection of direction of rotation via �90o phase-shifted signal F2Pulse counting input range pulses 0 ... 999999Maximum pulse rate pulses/
s1 000 000
Interval between 2 successive edges F1(�), F2(�)
ns �400
Scaling range, max. digits 20 at 1 pulseScaling range, min. digits 1 at 10000 pulsesMeasurement frequency range, adjustable (−1 dB) Hz 0.05 ... 1000
Sampling rate, max. 1/s 2400
B2359-2.0 en HBM9
![Page 10: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/10.jpg)
SSI transducersTransducers that can be connected Displacement and angle transducers with SSI interfaceVoltage supply 5 V, max. 150 mA or 24 V, max. 300 mA2-channel mode Time-division multiplex methodData input D(�) Differential input (RS422)Input levels, data input D (�)
Low levelHigh levelEach line to measurement ground, max.
VVV
�0.8�2�14
Hysteresis V 0.07
Permissible common-mode voltage, max. V −7 ... +12
Clock output Cl (�)Differential output voltage Cl(�), without load, max.Differential output voltage Cl(�), RL = 50 ohms, min.
VV
Differential output (RS422)5.82
Common-mode voltage at Cl (�), max. V 3
Short-circuit current, clock output Cl (�), typically mA 100
Resolution, single turn bits 12, 13Resolution, multi-turn bits 24, 25Scaling range, max. digits 20 at 1 pulseScaling range, min. digits 1 at 10000 pulsesMeasurement frequency range, adjustable (−1 dB) Hz 0.05 ... 1000Sampling rate, max. 1/s 1200Baud rates kBaud 100, 200, 500, 1000
Coding Gray code
Potentiometric displacement transducer see transducer carrier frequency
General specificationsLimit value switches
Number 4 per channel
Reference level gross
Hysteresis % 1 ... 100
Adjustment accuracy digit 1
Response time, typically (fc=1000 Hz) ms �2
Control outputsNumber 4 (MP85ADP(−S)) / 8 (MP85A(−S)), electrically isolated
Function Process OK/NOK, process started/running, processfinished/valid, limit values 1−4, transducer test result, result
tolerance window, status MMC, status channel x/y
Nominal (rated) voltage, external power supply VDC 24
Permissible supply voltage range V 10 ... 30
Maximum output current per output A 0.5
Short-circuit current, typically (Uext. = 24 V, RL� 0.1 ohm) A 0.8
Short-circuit period unlimited
Isolation voltage, typ. VDC 500
Control inputsNumber 1 (MP85ADP) / 5 (MP85A), electrically isolated
Function Switch test (up to 5 with MP85A−S), zero balance, shuntcalibration, parameter set selection, start/stop process,
transducer test, store/clear statistics
Input voltage range LOW V 0 ... 5
Input voltage range HIGH V 10 ... 30
Input current, typically, (High level = 24 V) mA 12
Isolation voltage, typ. VDC 500
Ethernet interfaceTransmission protocol MBits/s TCP/IP, can be networked per IEEE802
Transfer rate, max. MBits/s 10 and 100 (automatic selection)
Topology (twisted pairs) 2LED display for reviversLED display for revivers
Transmitter (RxD/TxD) and link 2Line length, maximum m 100
Cable type UPT category 5 or shielded twisted pair (STP)
Connecting socket RJ−45
B2359-2.0 enHBM 10
![Page 11: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/11.jpg)
CAN interfaceProtocol CAN 2.0B; CANopen compatible
Hardware bus link to ISO 11898
Baud ratesMaximum line lengths
kBits/s
m
1000 500 250 125 100 50 20 10
25 250 500 1000 600 1000 1000 1000
Termination resistor connectable by switch
Connection terminals
Multi Media Card (memory card on Flash basis)
Function Storage of: parameter sets, curves and results, statistics,circular buffer of last 1000 curves, file transfer to PC
Usable types MBytes 8, 16, 32, 64, 128, 256, 512, 1024
Data transmission rate, typically kBytes/s
2−8
File system DOS, FAT format
Display
Type 2-line, 8-character alphanumeric, LCD
Keyboard Touch-sensitive keypad with three keys, pressure-sensitive
Temperature rangeNominal (rated) temperature range oC 0 ...50Operating temperature range oC −20 ... +50Storage temperature range oC −20 ... +70
Degree of protection IP20
Dimensions (W x H x D) mm 55 x 146 x 156
Weight, approx. g 800
Profibus DP interface (MP85ADP only)
Protocol Profibus DP Slave, per DIN19245−3
Baud rate, max. MBaud 12
Slave address 3−123, set via the keyboard
Profibus ID number Hex 699
Configuration data bytes 5
Parameter data , max. bytes 6 (+7DP standard)
Function Access to and parameterization of all MP85A functions (remotecontrol)
Parameterization (asynchronous) per DPV1 standard
Input data , max. bytes 142
Output data, max. bytes 40
Update rate, inputs ms 1 (for 4 measured values)
Update rate, outputs ms �10, for zero setting, limit values
Diagnostic data bytes 48
Profibus connection 9-pin sub-D (DIN19245−3), electrically isolated from powersupply and measurement ground
B2359-2.0 en HBM11
![Page 12: MP85A S MP85ADP S Data Sheet - hottinger.skhottinger.sk/Prospekty/Elektronika/spol/PME/MP85/p_MP85a_s_e.pdf · −100% quality control of the ... Force at mechanical switching peak](https://reader035.vdocument.in/reader035/viewer/2022062908/5ac9365a7f8b9aa1298ce913/html5/thumbnails/12.jpg)
Dimensions of the PME modules:
59 152
61
150
Scope of supply
4 plug-in screw terminals, coded Phoenix order number:1x voltage supply and CAN, 6-pin MV STBW 2.5/6−ST−5.08 GY
2 x transducers, 8-pin MCVW 1.5/8−ST−3.81 GY
1x In/Out digital, 8-pin MC 1.5/8−ST−3.5 GY
FASTpress Suite system CD with:
free PME Assistant setup software
Online Help with Tricks&Tips
Quick Reference Guide for beginners
PME Assistant PLUS (demo version) with EASYsetup: User administration and EASYteach: statistical process analysis and report generation
MP85A Toolkit:
Function module kit for creating separate interfaces on operator panels via Ethernet under Windows XP, Windows CE and WindowsMobile
MULTImonitor (demo version of the production software for simultaneous display of up to 4 checking processes)EASYmonitor CE (demo version of the production software for operation at a terminal with Windows CE operatingsystem)EASYmonitor Mobil (application for operation with a PDA or pocket PC)
Accessories (not included in the scope of supply):DT85: 1/4 VGA b/w display for visualization and parameterization of several MP85As via the CAN interface
Multimedia card (MMC); e.g. Sandisk (www.sandisk.com)
Standard flat ribbon cable, 10-pin, 1.27 mm pitch