Next Steps in AAV Characterization via Light Scattering Measuring AAV Critical Quality Attributes Anatolii Purchel, Ph.D., Michelle Chen, Ph.D., John Champagne, Ph.D., Wyatt Technology Corporation Introduction Adeno-associated virus (AAV). AAV has mild immune response, able to deliver its payload to a wide range of host cells, and is becoming a prominent delivery vehicle in gene therapy. Multi-Angle Light Scattering (MALS) determines absolute molar mass and size directly, in solution, independent of elution time and conformation. This technique, combined with multiple concentration detectors allows independent protein and nucleic acid payload characterization. AAV quality attributes. SEC-MALS method is presented for measuring total capsid particle concentration, relative capsid content, and percentage of monomer or aggregates. FFF-MALS and DLS. Field-flow fractionation and batch DLS were employed to further characterize amount of aggregates and particle concentration. Conformation-independent molar mass analysis that does not rely on standards Triple-detection system allows independent quantification of AAV capsid protein and nucleic acid payload No a priori knowledge about the AAV structure or con-tent is required Batch DLS can be used as high-throughput assays prior to preferred analytical (high resolution) approach afforded by SEC- or AF4-MALS FFF-MALS determines if aggregation occurred after AAV synthesis or processing The SEC-UV-MALS-dRI measures AAV particle concentration, relative capsid content, and aggregation all in a single run SEC-MALS: Determines molar masses independently of elution time, and does not require calibration against reference molecules. Protein Conjugate Analysis Calculating Number of Particles DLS FFF-MALS FFF-MALS Conclusions SEC-MALS Setup ( ) 2 scattered dn I Mc dc θ SEC-MALS typical setup consists of HPLC system, DAWN ® and Optilab ® . For AAV characterization, two distinct concentration detectors are necessary, e.g. RI and UV280 FFF-MALS Typical setup consists of HPLC system, Eclipse TM AF4 TM , DAWN ® , and Optilab ® . Separation happens in a channel on a semi-permeable membrane Method development: appropriate column to assure 100 % mass recovery, peak symmetry, system validation with more than one standard Chromatograms of Empty and Full AAVs Two injections of each sample are overlaid Calculations: AAV protein and DNA mass (m capsid and m DNA ) as well as AAV protein and DNA molar masses (M capsid and M DNA ) are obtained from protein conjugate analysis. C Full = m DNA x N A / (M DNA x v) – full AAV particle concentration (v – injection volume) C AAV = m capsid x N A / (M capsid x v) total AAV particle concentration (v – injection volume) Calculated CQA’s: C Full , C Empty , V g /C p (V g, C p – genome and AAV particle titer) Step by step guide: 1. Validate the system with two or more standards 2. Validate the system with empty and filled AAVs 3. Process the data using Protein Conjugate Analysis C g /V p ratio measured using protein conjugate analysis and compared to the theoretical value based on mixing ratio Batch DLS/SLS provides size and MW as well as concentration very quickly and easily. Use as high- throughput assays prior to preferred analytical (high resolution) approach afforded by SEC- or AF4-MALS. Empty Full Empty AAV: R h = 14.7 ± 0.7 nm M w = 3.67 ± 0.01 MDa Full AAV: R h = 15.6 ± 0.1 nm M w = 6.78 ± 0.03 MDa FFF is able to separate large AAV aggregates and monomers A subsequent blank injection is useful for determination of sample retention on the membrane AF4 allows large aggregates characterization HMW aggregates visible by AF4-MALS may be removed by SEC column Empty AAV Full AAV Protein Conjugate Analysis: provides independent values for protein mass, DNA mass, and respective molar masses Protein conjugate analysis determines molar mass of the protein and DNA components independently DNA Adenovirus AFM/TEM FFF-MALS % Difference Total Particle Count 2.9 x 10 10 2.8 x 10 10 2 Average Radius 43.0 45.0 5 Relative Scale Time, min 5.0 10.0 15.0 20.0 25.0 30.0 0.0 0.5 1.0 UV Subsequent blank injection Relative Scale 5.0 10.0 15.0 0.0 0.5 1.0 UV SEC Column fragment oligomers Large aggregates? Volume, mL Relative Scale 10.0 15.0 20.0 25.0 30.0 0.0 0.5 1.0 UV Large aggregates fragment oligomers Time, min AAV Monomer Aggregates SEC-MALS App Note DLS App Note About FFF

Upload: others

Post on 22-May-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Next Steps in AAV Characterization via Light Scattering ... · SEC- or AF4-MALS • FFF-MALS determines if aggregation occurred after AAV synthesis or processing The SEC-UV-MALS-dRI

Next Steps in AAV Characterization via Light Scattering Measuring AAV Critical Quality Attributes

Anatolii Purchel, Ph.D., Michelle Chen, Ph.D., John Champagne, Ph.D., Wyatt Technology Corporation

Introduction

Adeno-associated virus (AAV). AAV has mild immune response, able to deliver its payload to a wide range of host cells, and is becoming a prominent delivery vehicle in gene therapy.

Multi-Angle Light Scattering (MALS) determines absolute molar mass and size directly, in solution, independent of elution time and conformation. This technique, combined with multiple concentration detectors allows independent protein and nucleic acid payload characterization.

AAV quality attributes. SEC-MALS method is presented for measuring total capsid particle concentration, relative capsid content, and percentage of monomer or aggregates.

FFF-MALS and DLS. Field-flow fractionation and batch DLS were employed to further characterize amount of aggregates and particle concentration.

• Conformation-independent molar mass analysis that does not rely on standards

• Triple-detection system allows independent quantification of AAV capsid protein and nucleic acid payload

• No a priori knowledge about the AAV structure or con-tent is required

• Batch DLS can be used as high-throughput assays prior to preferred analytical (high resolution) approach afforded by SEC- or AF4-MALS

• FFF-MALS determines if aggregation occurred after AAV synthesis or processing

The SEC-UV-MALS-dRI measures AAV particle concentration, relative capsid

content, and aggregation all in a single run

SEC-MALS: Determines molar masses independently of elution time, and does not require calibration against reference molecules.

Protein Conjugate Analysis Calculating Number of Particles

DLS FFF-MALS FFF-MALS Conclusions

SEC-MALS Setup

( )2

scattered

dnI Mcdc

θ ∝

SEC-MALS typical setup consists of HPLC system, DAWN® and Optilab®. For AAV characterization, two distinct concentration detectors are necessary, e.g. RI and UV280

FFF-MALS Typical setup consists of HPLC system, EclipseTM AF4TM, DAWN®, and Optilab®. Separation happens in a channel on a semi-permeable membrane

Method development: appropriate column to assure 100 % mass recovery, peak symmetry, system validation with more than one standard

Chromatograms of Empty and Full AAVsTwo injections of each sample are overlaid

Calculations: AAV protein and DNA mass (mcapsid and mDNA) as well as AAV protein and DNA molar masses (Mcapsid and MDNA) are obtained from protein conjugate analysis. CFull = mDNA x NA / (MDNA x v) – full AAV particle concentration (v – injection volume)CAAV = mcapsid x NA / (Mcapsid x v) – total AAV particle concentration (v – injection volume)Calculated CQA’s: CFull, CEmpty, Vg/Cp (Vg, Cp – genome and AAV particle titer)Step by step guide:1. Validate the system with two or more standards2. Validate the system with empty and filled AAVs3. Process the data using Protein Conjugate Analysis

Cg/Vp ratio measured using protein conjugate analysis and compared to the theoretical value based on mixing ratio

Batch DLS/SLS provides size and MW as well as concentration very quickly and easily. Use as high-throughput assays prior to preferred analytical (high resolution) approach afforded by SEC- or AF4-MALS.

Empty

Full

Empty AAV:Rh = 14.7 ± 0.7 nm

Mw = 3.67 ± 0.01 MDa

Full AAV: Rh = 15.6 ± 0.1 nm

Mw = 6.78 ± 0.03 MDa

FFF is able to separate large AAV aggregates and monomersA subsequent blank injection is useful for determination of sample retention on the membrane

AF4 allows large aggregates characterization

HMW aggregates visible by AF4-MALS may be removed by SEC column

Empty AAVFull AAV

Protein Conjugate Analysis: provides independent values for protein mass, DNA mass, and respective molar masses

Protein conjugate analysis determines molar mass of the protein and DNA components independently

DNA

Adenovirus

AFM/TEM FFF-MALS %Difference

Total Particle Count 2.9 x 1010 2.8 x 1010 2

Average Radius 43.0 45.0 5

Rel

ativ

e Sc

ale

Time, min 5.0 10.0 15.0 20.0 25.0 30.0

0.0

0.5

1.0

UV

Subsequent blank injection

Rel

ativ

e Sc

ale

5.0 10.0 15.0

0.0

0.5

1.0UV

SEC Column

fragmentoligomersLarge aggregates?

Volume, mL

Rel

ativ

e Sc

ale

10.0 15.0 20.0 25.0 30.0

0.0

0.5

1.0

UV

Large aggregatesfragment oligomers

Time, min

AAV Monomer

Aggregates

SEC-MALS App Note DLS App Note About FFF