nicholson 1965

Upload: jubin-kumar

Post on 28-Feb-2018

233 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/25/2019 Nicholson 1965

    1/5

    m i n i m u m i n t h e T C

    v s .

    concentration

    isotherm (21)-correctly describes th e

    da ta obtained. However , i t seems un-

    likely that so many compounds would

    show maxima or minima in nitrogen but

    none of them would do

    so

    in Ar or CO,,

    while they do give positive and negative

    responses in these gases.

    One possible al ternative explanation

    is that a concentration gradient is

    established in the cell as a result of

    thermal diffusion between the cell wall

    and the hotter fi lament. This could

    result in a t ransfer cf thermal energy

    across the gradient, th e so-called Dufo ur

    effect

    7). If

    so, it should be possible

    to el iminate this effect by increasing

    the tu1 bulence in the cell. Preliminary

    investigation with such a cell indicates

    t ha t t h is is

    so.

    T he lack of W peaks and th e low cost

    of CO, suggest that it would be a better

    carrier gas than nit iogen for preparative

    work (and possibly even some analytical

    work). Some workers hav e reported

    collecting samples in CO, by freezing i t

    to dry ice

    (9).

    Also, the wide variation

    in response values (posit ive and

    negative) might be useful for quali tat ive

    analysis of pure compounds. Fur ther

    work w ith CO z and oth er carrier gases of

    low T C is p lanned.

    A CK NO W LE DG M E NT

    The authors thank Floyd Fredr icks

    for obtaining some of th e data .

    LITERATURE CITED

    (1) Bennett,

    L. A . ,

    Vines, R. G., J .

    ( 2 )

    Bohemen. J.. Purnell, J. H., J . A p p l .

    Chem.

    P h y s .

    23

    1587 (1955).

    , .

    . .

    Chem. 8 , 433 (1958).

    (3 ) Dal Piogare,

    S.,

    Juvet,

    R. S. ,

    Jr . ,

    Gas Liquid Chrom atography, p. 192,

    Interscience,

    N e w

    York, 1962;(

    (4 ) Dreisbach, R ., compiler, Physical

    Prop erties of Chemical Com pounds,

    3 Vols., American Chemical Society,

    Washington,

    D .

    C., 1955, 1959, 1961.

    (5) Hansen, R.

    S.,

    Frost ,

    R.

    R., Murphy,

    J. A., J .

    Phys.

    Chem. 68, 2028 (1964).

    ( 6 ) Harvey, D., Norgan, G. O., in

    Vapor Phase Chromatography-Lon-

    don, 1956, p. 74,

    D .

    H . Desty, ed.

    Academic Press, New York, 1957

    ( 7 ) Hirschfelder, J.

    O.,

    Curtiss, C. F.,

    Bird, R. B., Molecular Theory of

    Gases and Liquids, p. 522, Wiley,

    N ew

    York, 1954.

    (8) Hoffmann,

    E.

    G., ANAL. CHE f.34

    1216 (1962).

    (9) Hornstein, I., Croae, P.,

    Ibid. 37

    170 (1965).

    (10) Jamieson, C. R., J .

    Chromatog. 3

    464. 494 11960): 4 . 420 11060): 8.

    ~~ , , -,

    i i i ( i 9 6 2 ) : i S , i 6 0 ( i 964) .

    (11) Keppler, J. G., Dijkstra, G., Schols,

    J. A., i n Tapour Phase Chroma-

    tography-London, 1956, p. 222, D .

    H. Desty, ed., Academic Press, New

    York, 1957.

    ( 1 2 )

    Keulemans, A . I . M., KFantes, A,,

    Rijnders, G. W .

    A.,

    Anal. Chim. Acta

    16 29 (1957).

    (1 3) Littlewood, A. B., Gas Chroma-

    togra phy, pp. 329-33, ilcademic

    Press, hew York, 1962.

    (14) Madison,

    J.

    J.,

    AXAL. CHEY.

    30

    1859 (1958).

    (15) hIessner, A . E., Rosie, D. &I.,

    Argabright, P. A . , Ibid. 31 230 (1959).

    (16) Panson, A. G., Adams, L. lI. .

    Gas Chromatog. 2 , 164 (1964).

    117) Pauschma nn. H..

    2

    Anal. Chem.

    ,

    203. 16 11964).

    (18) Purcell,

    J.

    E., E

    (19) Purnell, H., .Gas Chromatography,

    Chromatog.

    D.

    286, Wilev. New York, 1962.

    Xtre, L. S. ,

    .

    Gas

    3

    69 (1965).

    ( 2 6 )

    Rothman,

    A . J., Bromlep, L. A , ,

    Ind. Ena. Chem. 47. 899 11955).

    (21) Schmiuch,

    L. J.,

    Dinerstein; R.

    A, ,

    (22) Sm ith,

    B .

    I ., Bowden,

    W .

    W.

    (23) SDencer.

    H.

    M. . Am. Chem.

    SOC

    ANAL.

    CHEM.

    32 343 (1960).

    Ibid. 36 82 (1964).

    (24) Stuve,~K. ,n Gas Chromatography-

    1958,

    p.

    178, I). H. Desty, ed.,

    Aca-

    demic Press, X e w Ynrk. l Q 5

    (25) lerzele, AI., J .

    (1964).

    Talanta 10

    937 (1963).

    (26) Williams,

    A. F.,

    Murray,

    W.

    .,

    RECEIVED or review April 7, 1965.

    Accepted July 22, 1965. Pre-ented at

    the Pittsburgh Conference

    o n

    Analytical

    Chemistry and Applied Spectroscopy,

    March 3, 1965.

    Theory and Applica tion

    of

    Cyclic Voltammetry

    f m Measurement of Electrode Reaction Kinetics

    RICHARD

    S.

    NICHOLSON

    Chemistry Department, Michigan State University, East Lansing,

    Mich.

    b

    The theory of cyclic voltammetry

    has been extended to include electron

    transfer reactions which are described

    by the electrochernical absolute rate

    equat ion . Results of theoreti cal calcu-

    lations made it possible to use cyclic

    voltammetry to measure standard rate

    constants fo r electr on trans fer. Thus,

    a

    system which appears reversible

    at one frequency may b e made to

    exhibit kinetic behavior at higher

    frequencies, as indi cated b y increased

    separation of cathodic and anodic

    pe ak potentials. The standard rat e

    constant for electron transfer is de-

    termined from this peak potential

    separation and frequency. The

    method provides an extremely rapid

    and simple way to evaluate electrode

    kinetics. The reduction of cadmium

    i s

    used as an illustration.

    U R I N G

    RECENT years a number of

    D methods have been developed

    fo r t he measu remen t

    of

    electrode reac-

    t ion kinetics. I n one sense, some of

    these determine electrode reversibil i ty

    indi rect ly by measur ing the apparent

    standard rate constant for elect ron

    transfer from only cathodic (or anodic)

    polarization. In a few cases both

    cathodic and anodic polarization give

    consistent results ( I S ) . M an y of the

    relaxation techniques developed for fast

    reactions have the disadvantage tha t

    smal l ampl i tude per turbat ions are used,

    and consequently differences or changes

    in mechanisms are not easily detected.

    A

    method which overcomes this dis-

    advantage an d a t the same t ime gives

    a

    direct e stim ate of reversibility is

    cyclic triangular wave voltammetry.

    Thu s, the presence of homogeneous

    reactions in the mechanism is readily

    detec ted, and in terpre tation of results

    usually is simple. l direct estimate of

    electrode reversibility is provided, be-

    cause the potentials

    at

    which oxidation

    and reduction occur are observed

    directly. For example, at low fre-

    quencies i t may be possible with a given

    system t ha t electrochemical equil ibrium

    always is maintained a t the elect rode

    surface. Under these condit ions the

    sepa ration of catho dic an d anodic peak

    potentials is about 6 0 / n mv. , and t he

    reaction is reversible.

    Clearly for this case no kinetic in-

    formation about the electron transfer

    reaction can be obtain ed. However, if

    frequency is increased sufficiently, a

    point may be reached at which the

    kinetics of ele ctron transfe r become

    competi t ive with the rate of potential

    change. Under these condit ions

    it

    m a y

    be possible to stud y the kinetics of the

    electrode reaction, and the separation of

    peak potentials should be a measure

    of t he stan dard r ate c onsta nt for elec-

    tron transfer. Thu s, a t least in princi-

    ple, one can est imate s tandard rate

    constants simply by observing cyclic

    polarograms on the oscilloscope, and

    then increasing the triangular wave

    freque ncy until the separation of peak

    potentials becomes greater than 60/n

    mv. The s t andard r a t e cons t an t t hen

    should be a calculable function of fre-

    quency at this peak po tential separation.

    Unfortunately, there is no a priori way

    VOL.

    37,

    NO. 1 1 , OCTOBER 1965 0

    1351

  • 7/25/2019 Nicholson 1965

    2/5

    0.4

    n

    0

    I

    120 60 0

    60

    -120

    (E - E h mv.

    Figure 1 . Cyclic polarograms showing

    effect of charge transfer coefficient, a

    iC

    =

    0 . 5 ; CY =

    0.7

    ...... C

    =

    0 . 5 ; a = 0.3

    to mak e thi s correlation because effects

    of con centr ation polarization can be

    treated only by mathematical analysis

    of the mass tran spo rt processes. Such

    an analysis is not available for the

    present case, and consequently applica-

    tions of these ideas using cyclic voltam-

    metry have been limited to simple

    s ta tements tha t a given sys tem appears

    to be reversible or irreversible

    ( 7 ,

    IO).

    I n the st ud y of electrode reaction

    mechanisms, it often is useful to be able

    to obtain experimentally a rapid esti-

    ma te of electron transfer rate s. Be-

    cause this appeared to be possible with

    cyclic vol tamme try in the w ay jus t

    described, we have attempted

    a

    theo-

    retical treatment

    of

    the problem which

    includes the effects of c once ntratio n

    polarization. Our primary objective

    was to determine if cyclic voltammetry

    could be used to provide rapid and

    reasonably accurate determinations of

    s tandard ra te constants for e lectron

    transfer. Thu s, the theoretical correla-

    tions between peak potentials, standard

    rate con stant , and rat e of potential

    scan have been emphasized, although

    other correlations are possible.

    A

    seri-

    ous limitation of this approach would

    result if peak potential separations

    depended also on the charge transfer

    coefficient, C Y . This would require an

    independent measure of C Y and m ake the

    meth od of little use in ter ms of t he

    stated objectives. However, through

    proper selection of conditions, peak

    potential separations become nearly

    independent of

    a .

    Becauqe there is neither theoretical

    nor experimental ad van tage to consider-

    ing more tha n one cycle of th e applied

    triangular wave, theoretical calcula-

    tions havt been limited to this case.

    ConsideratioIL also has been restricted

    to the case of an situ generation of the

    reduced form

    of

    the couple under in-

    vestigation. This actually has two

    advantages. First, it reduces by one

    th e number of indep endent variables in

    th e theoretical calculations. Second, it

    has the experimental advantage

    of

    not

    requiring preparation

    of

    amalgams of

    known concentration for the study of

    reductions at mercu ry electrodes.

    The method described can be applied

    to systems in which homogeneous

    chemical reactions precede or follow

    electron transfer provided such reac-

    tions are either rapid or slow compared

    to the s tandard ra te constant for e lec-

    tron transfer. The cases in which

    these conditions are not met are easily

    detected from the form of th e experi-

    mental current-voltage curves 11) .

    Th e redu ction of ca dmiu m is used as

    an i l lustra t ion of the method developed

    here. Application to other systems is

    in progress and will be reported else-

    where.

    THEORY

    We assume the following mechanism:

    O + n e * R I

    There k , and ka are heterogeneous ra t e

    constants of t he electron transfer, and

    are assumed to be f unctions of potential

    as expressed through the electrochem-

    ical absolute rat e equation. The time

    depend ence of electrod e pote ntial is th e

    form of an isosceles triang le. Syste m I

    is assumed to be initially in equilibrium,

    Both and R are to be soluble either

    in the electrode or solution phase.

    T o account for concentration polariza-

    tion, diffusion to a plane electrode is

    assumed to be th e only source of m ass

    transport . Mathematical formulat ion

    of this problem follows

    kf

    kb

    aco

    v o

    _ -

    a t - D O T a x

    t 2 O , x = O

    Undefined terms have their usual

    significance 11).

    With the e lectrochemical absolute

    rate equat ion, Equat ion 6 can be t rans-

    fo rm ed to

    There

    E

    is the electrode potential; E o

    is the s tandard potent ia l ;

    k ,

    is the

    s tanda rd ra te cons tan t

    at

    E

    = E O

    LY

    is the transfer coefficient; an d the

    remaining terms have their usual

    significance.

    The potent ia l in Equat ion 7 f o r t h e

    first scan (reduc tion) of th e trian gular

    wave takes the form

    (8)

    =

    E ,

    - ut

    and for th e second scan (oxidation)

    E

    =

    E , + u t - 2uX

    9)

    Here, E , is the initial equilibrium poten-

    tial, v is dE/d t , and X is the period of

    the triangular wave.

    If

    Equa t ions 7 , 8, a n d

    9

    are combined

    th e result is

    aco

    ax

    o

    k , ( C o * / C ~ * ) - " [ S x ( t ) ] - *

    x

    CO,

    =

    0 -

    ( C O * / ~ R * ) ~ A ( ~ ) ~ R ,) (10)

    Th e func t ion Sx t) is defined as

    t < X

    (11)

    2ax

    t >

    ( t ) =

    where

    a =

    nFv/RT

    12)

    Equa t ion

    10

    is t he final form of bound-

    ary condit ion

    6 .

    By application of Laplace transf orm

    methods the above boundary value

    problem can be converted to the follow-

    ing dimensionless linear integral equa-

    tion w ith v ariabl e coefficients

    X(Y) [?(CO*/cR*)sd( / )

    =

    4

    T o m a k e E q u a t i on

    13

    dimensionless the

    following substitutions and changes

    of

    variable were made:

    Y

    =

    (Do/DR)"' (14)

    y =

    at

    (15)

    16)

    4 = y 4 s / l / * a D o

    (17)

    x(y) = Do

    aco

    ,/Co*l/?rao,

    For

    large values

    of

    $ (large k ,

    or

    small v Equa t ion 13 becomes independ-

    e n t

    of

    4

    and CY and reduces

    to

    t h e

    1352

    ANALYTICAL CHEMISTRY

  • 7/25/2019 Nicholson 1965

    3/5

    (E

    - E,,h, rnv

    Figure 2. Cyclic pol arograms showing

    effect

    of

    charge transfer parameter, +

    $ =

    7 .0 ; CY = 0.5

    . $ = 0.25; CY =

    0.5

    ..

    6

    80

    100-

    6 ,

    3 120-

    140-

    160-

    corresponding equation for reversible

    (Nernstian) electron transfer

    11).

    Likewise, as

    $

    approaches zero, Equa-

    tion

    13

    approaches the case for tota l ly

    irreversible electron transfer

    (11).

    T h e

    int erm edi ate case, which is of int erest

    here, is described completely by Equa-

    tion 13.

    Solution of Integral Equation.

    E q u a t i o n

    13

    cannot be so lved an-

    a ly t i ca lly . Al though a so lu t ion in

    th e form of an infini te series can be

    obtained, series solut ions for more

    th an one-half cycle becom e hopeless ly

    com pl ica ted . Th e on ly p rac t i ca l ap-

    proach , the re fore , which fo r the

    presen t case can be app l ied wi thou t

    loss of ge nera lity, is numeric al evalu a-

    tion. We hav e used

    a

    method described

    previously (1

    ) .

    Several othe r methods

    also are poss ible , and apparent ly the

    method used by DeTtries and VanDalen

    (6) involves less computational time

    for comparable acculacy ( 5 ) .

    Solutions of E quat ion

    13

    were inde-

    pendent of values of r (C , * /CR*) pro-

    vided r ( Co*/CE*) was greater than e6 5.

    This corresponds to the experimental

    f a c t t h a t

    if

    R is absmt ini t ia l ly and the

    voltage scan is begun a t potentials

    anodic of th e pcllarographic wave,

    current-voltage curves are independent

    of the exact initial potential selected,

    All calculations have been performed

    using

    r (Co*/CR*)

    =

    e65.

    With this s implif iaat ion i t

    was

    possi-

    ble to calculate numerically current-

    voltage curves. These curves depended

    on three vaiiables,

    p

    y, and

    A. It

    has

    been shown previously that

    h

    affects

    only the anodic wave, and that there

    the effect on anodic peak potentials is

    qui te small

    ( 1 1 ) .

    This is particularly

    the case when switching potentials close

    to the cathodic peak are avoided

    [ Ex- El ,Jn> 90 mv.].

    All calcula-

    t ions reported here are based o n

    a

    value

    of (Ex - = 141/n m v. Thus ,

    -

    -

    I I I

    calculated current-voltage curves could

    be interpreted on the basis of only two

    variables,

    \L

    a n d

    a.

    Results of Calculations.

    W h e n

    becomes sufficiently large

    ( k ,

    large,

    or v sm a l l ) , cur ren t -vo l tage curves

    calculated from Equation 13 are inde-

    penden t of the kinet ic param eters

    $

    a n d

    CY. For this case results are identical

    to ones obtained previously where the

    electron transfer was assumed to be

    S erns t i an

    11).

    Thi s limit occurs when

    $ > 7, which is in good agreemen t with

    calculations of Ma tsuda and Ayabe (8).

    W h e n $

    0.5

    th e converse holds.

    The second effect involves

    a

    slight

    displacement

    of

    the waves along the

    potential axis. Fo r example, as

    Y

    decreases, the cathodic peak is displaced

    cathodical ly. At leas t for the data used

    to construct Figure I , th e effect is slight.

    Lloreover, as the cathodic peak shifts

    cathodically with decreases of CY t h e

    anodic peak also shifts cathodically.

    Ther efor e, in ter ms of differences of p eak

    potentials, AEP, changes in

    CY

    t e n d t o

    cancel. Th us, value s of

    AEp

    for the

    present case are nearly independent of

    cy in the range 0.3 < CY lode1 SK2-V operational amplifier

    (G.

    A. Philbrick Researches, Inc.

    Boston, Mass . ) . To improve rise-time

    and increase current capabilities,

    a

    Philbrick Model

    K2-BJ

    booster

    amplifier was used in series with the

    potent io s ta t . T he bandwidth of the

    potent ios ta t was adjus ted for the solu-

    tion used to provide optimum rise-

    t ime and s tabi l i ty .

    To

    do th i s the

    ideas of Booman and Holbrook were

    used

    I , 2 ) . A

    good qualit y commercial

    potent ios ta t (e .g. , Wenking Potent io-

    s ta t , Br inkm ann Ins t rum ents , Wes t -

    bury ,

    3 Y.)

    presumably could be used

    in place of th e poten tios tat just de-

    scribed.

    Two different signal generators were

    used, and both proved adequ ate . One

    was constructed from operat ional ampli-

    fiers and has been described previously

    1 0 ) .

    The o the r was an Exac t Mode l

    255 funct ion generator (Exac t Elec-

    tronics, Inc. , Hillsboro, Ore.) which

    proved useful and v ersatile.

    The d etector was

    a

    Tektronix Model

    536 oscilloscope provided with

    a

    Po-

    laroid camera a t tachment (Tektronix

    T y p e

    C-12).

    The vert ical input

    was

    always

    a

    T y p e

    D

    plug-in preamplifier.

    This could be operated in the differen-

    tial mode and p ermitted use of a floating

    load resistor in the po tentiostatic circuit

    10).

    To measure scan ra tes for the

    function generator constructed from

    operational amplifiers,

    a

    T y p e

    T

    plug-in

    preamplifier was used for the horizontal

    inp ut of th e oscilloscope. W ith the

    Exact funct ion generator, scan ra tes

    could be dialed directly with an

    ac-

    curacy of

    =t2

    and this proved to be

    a

    considerable advantage over th e othe r

    signal generator.

    To

    measure peak

    potential separations

    a

    T y p e

    H

    plug-in

    preamplifier was used for t he horizo ntal

    inp ut of t he oscilloscope. T he hori-

    zontal axis then was driven by the cell

    potential, and peak potential separations

    could be measured with a n accuracy of

    a b o u t 1 2 m v.

    Th e cell assembly has been described

    previously

    10) .

    Materials. S olu t ions were p repa red

    f r o m c a d m i u m s u l f a t e ( B a k e r

    A. R.)

    and anhydro us sod ium su lfa te (Baker

    A . R. ).

    Th e concen t ra t ions were : cad-

    mium sulfate, ea.

    2 x 10-4M1;

    sodium

    sulfate,

    1.0 X ,

    The exact cadmium

    concentra t ion was not determined.

    Xeasurements were made in

    a

    cons tan t

    t em pera tu re room a t am bien t t em pera-

    tures of

    23

    t o

    25 C.

    RESULTS A ND DISCUS SION

    Both one-cycle and multi-cycle

    (steady state) experiments were run

    and as would be expected the differ-

    ences between A E p for the two methods

    were of t he order of ex perim ental erro r.

    Nevertheless , data reported here are

    for one-cycle experiments to conform to

    theoretical calculations. Fo r all experi-

    ments an initial potential of

    -0.370

    volt

    us.

    S.C.E.was used, and the ampli-

    tud e of t he triangular wave always was

    300 mv.

    Results for reduction of 2

    X 10-4M

    cadmium in

    1.022

    sodium sulfate are

    summarized in Table 11. To conver t

    experimentally determined values of

    Table

    II

    Determination with Cyclic

    Voltammetry of k, for Reduction of

    Cadmium

    AEP

    v , x

    n,

    ke,

    volt/sec. mv. 5 cm./sec.

    48.0

    94

    0.70 0 . 2 5

    60.0 98

    0.61 0.25

    90.0 108

    0.48 0.24

    120, o 115 0.41 0.23

    a Determined from Figure

    3;

    see

    b 2 X 10-4M

    cadmium sulfate,

    1.OM

    Equation

    17.

    sodium sulfate.

    to

    k,,

    the

    diffusion coefficients

    of

    Okinaka

    12)

    were used.

    A

    value of

    IL

    equal to

    0.25

    also

    was

    used. Although

    no effort was made to determine a

    accurately, the value of 0.25 was esti-

    mated by comparing with theory the

    symm etry and shape of cathodic and

    anodic waves for cadmium. Th e calcu-

    lated value of

    k ,

    is fairly insensitive to

    variations of

    a

    s already mentioned.

    The values of k , in Table

    I1

    agree

    reason ably &-ellwith some oth er workers

    14).

    The disagreement with others

    (12,

    14)

    is no t entirely explained. Th e

    appar ent ra te constant does change some

    with different supporting electrolytes.

    Fo r exam ple, we find

    k ,

    equal to about

    0.6 cm./sec. in 2 M perchloric acid.

    The value in sulfuric acid is slightly

    higher. In addi t ion, e lectrodes are sub-

    ject to an aging effect noted b y D elahay

    ( 3 , 4 ) .

    We also observe this decrease of

    apparent ra te constants with t ime, but

    the effect in our solutions is no t as g reat

    as

    Delah ay found (about

    a

    facto r of tw o

    change in

    k ,

    for the first

    20

    minutes

    quiescence).

    Th e upper limit of rate constan ts tha t

    can be determined by the above method

    appears to be s e t by

    a

    combination

    of

    factors . F irs t , in the measurement

    of peak potential separations, uncom-

    pensated ohmic potential losses are

    a

    serious source of error. Thus, for the

    rapid scan ra tes required to s tudy fas t

    electrode reactions, peak c urre nts be-

    come fairly large and even relatively

    small solution resistances can introduce

    serious error. Moreover, the effects of

    uncompensated iR drop qual i ta t ively

    are very similar to kinetic effects

    (9).

    This is amply i l lus tra ted by Figure

    3

    where the dashed cu rve is

    a

    plo t of da ta

    calculated previously for the effect of

    iR losses on cyclic voltammetry 9).

    Th e definition of th e horizontal axis ($)

    of Figure

    3

    for this case should b e

    (9)

    :

    =

    l / ( n F / R T ) n F AX

    ( T ~ D O ) ~ C ~ * R ,18)

    Th us, the variation of peak po tential

    separation w ith scan rate is quite similar

    for th e two cases.

    The effect of ohmic potential losses

    can be minimized in three mays. Firs t ,

    1354

    e

    ANALYTICAL CHEMISTRY

  • 7/25/2019 Nicholson 1965

    5/5

    relatively conce nt-rated electroly tes of

    high conductivity can be used to reduce

    the tota l cell resistance. Second, with

    a

    three-electrode potentiostat place-

    me nt of th e Lugg-in capillary probe in

    close proximity with the working elec-

    trode serves to compensate for

    a

    ma-

    jor ity of th e tota l cell resistance [see

    Iiooman and Holbrook for the exact

    relat, ionship

    (2)1.

    Thir d, re la t ively low

    concentra t ions of dectroact ive materia l

    can be used to keep tota l cel l currents

    small. Thi s latte r approach is limited,

    however, by charging curren t . Thu s , a t

    high scan ra tes charging current may

    become an appreciab le fraction of t he

    total cell cuirent, especially when

    lo^

    concentrations of electroactive mat'erial

    are used.

    For th e cadmium sys tem inves t igated

    here

    a

    concentration of

    2

    t o 5 X

    10-4Jl

    appeared to be opt imum for low cel l

    currents without excessive charging

    current contribut ions . hIas imu m peak

    cur ren ts n-ere of the ord er

    of 0.2-0.3

    m a.

    Total solution resistance (ea.

    20

    ohms)

    measured by peak potent ia l shif ts on

    moLing the Luggin probe more than

    10 radii from the working electrode

    agreed with values calculated from the

    conduc tivity of th e solution

    2).

    B y

    careful placement of the Luggin probe

    the uncompensated resistance was esti-

    m a ted to be

    3

    to ohm s

    2) ,

    s o t h a t

    maximum ohmic potential losses were

    of t he o rde r of

    1

    m v.

    The upper l imit

    of

    ra te cons tan t s

    measurable with cycl ic vol tammetry

    therefore depends on several factors.

    The most im portan t of these is proba bly

    conduc tivity of th e solution und er

    investigation . W ith electrolytes of rea-

    sonably high conduct ivi ty, i t should be

    possible to measure rate constants as

    large as

    1

    to

    5

    cm./sec.

    LITERATURE CITED

    (1)

    Booman,

    G. L.,

    Holbrook,

    W.

    B.,

    ( 2 ) Zbid.

    35, 1793 (1963).

    13) Delahav,

    P.,

    Trachtenberg,

    I., J . Am.

    ANAL.C H E M .

    7, 795 (1963).

    Chem.

    S i . '80 094 (1958).-

    (4 ) Delahay,

    P., J . Chim. Pkys .

    54,

    369 (1957

    .

    (5) DeVries,

    W .

    T.,

    Free University,

    Amsterdom. The Xetherlands. orivate

    communication, 1965.

    Electroanal.

    Chem. 8 366 (1964).

    Zbid.

    5 , 17 (1963).

    chem.

    59. 494 11955).

    (6) DeVries,

    W. .,

    VanDalen,

    E., J .

    (7 ) Galus, Z., Lee,

    H. Y.,

    Adams, R. N.,

    (8)

    hlatsuda, H., Ayabe,

    Y.,

    2 Elektro-

    (9) Nicholbon,

    R .

    S. ,

    ANAL.CHEM.37

    (10 ) Nicholson, R .

    S.,

    Shain,

    I., Ibid.

    667 (1965).

    p.

    190.

    (11)

    Zbid.

    36 06 (1964).

    ( 12 )

    Okinaka,

    Y., Talanta

    11,

    203 (1964).

    (13 ) Selis.

    S. >I.. J . Phus. Chem.

    68.

    ~

    2538 (1964).

    (14) Tanaka,

    X.

    amamushi, R.,

    Electro-

    RECEIVED

    or

    review June 28, 1965.

    Accepted August

    3,

    1965. Presented in

    part at the Division of Analytical Chem-

    istry, 150th XTeeting ACS, Atlantic City,

    N . J.,

    September 1965. Work supported

    in part by the United States Army Re-

    search Office-Durham under Con tract

    No .

    DA-31-12PARO-LL308. Other sup-

    port was received

    from

    the National Sci-

    ence Foundation under Grant

    KO.

    GP-

    3830.

    chem.

    -4cta 9,96 3 (1964).

    Pola r o g

    ra

    phy o f Lead(II) i n A q u e o u s H yd r o f u or ic

    Acid 1 to 12M) with

    a

    D r o p p i n g - M e r c u r y

    Electrodle of Tef lon

    HELEN P. RAAEN

    Analytical Chemistry Division, Oak Ridge Nation al Laboratory, Oak Ridge, Tenn.

    b

    The polarographic behavior of

    Pb+2 -0 .03 to 0.8rnM) in aqueous

    HF

    1.00 to 12.00M) was studied with

    a D.M.E. of Teflon. The po lar og rap h-

    ically usable potent ial span for

    12.00M

    HF i s

    about

    +0.4

    to

    -1.0

    volt

    vs.

    S.C.E.; for

    .OOM

    HF, the limit extends

    to about - .2 volts. With in this span

    only one polarographic wave for Pbt2

    exists; it is for the polarographically

    reversible 2-electron-change reduction

    Pb+* -+ PbO.

    The

    E l l 2

    values de-

    termined are

    -0.:380

    for

    1.00M

    HF)

    fa -0.370 for 141.00M

    HF)

    0.003

    volt vs. S.C.E.; the values are not

    affec ted by change in Pb+2 concen-

    tration. The wave is

    of

    excellent

    form and is usable for both quali-

    tati ve and qirantit citive purposes. The

    relation of

    j d ,

    andl also of ( /dt )maz,

    to Pb+2 concentrati on is linear; plots

    of these variables pass through the

    origin. With increase in

    HF

    concen-

    tration from

    1.00

    to 12.00M, no

    change occurs in the number of lead

    waves, shape of the wave, polaro-

    graphic reversibility of the reduction,

    or

    n ;

    the

    i d

    deceas es slightly; and

    the El/* becomes about 10 mv. more

    positive.

    HE

    D R O P P I N G - N E R C U R Y

    electrode

    of

    T

    e f lon (DuP ont )

    has

    been described

    8),

    nd i ts operat ion in noncorroding

    media was inves t igated

    (9,

    1 0 ) . A

    review w as given

    (8)

    of th e polarogra phy

    -done before th e D. 3I .E. of Teflon was

    developed -of substances in liquid hy-

    drofluoric acid and in aqueous acid

    fluoride media.

    The polarography of lead in acid

    fluoride media has been s tudied to

    a

    l im ited e s ten t . Wes t , Dean , and Breda

    ( 2 1 ) obtained a well-defined wave for

    lead in

    0 . 5 X

    Na 2F 2 [qic]-O.Ol gelatin

    solutions of p H 1.1 to 2 . 75 tha t con-

    tained considerable amo unt s of nitric

    acid added to dissolve th e lead f luoride

    precipitate; the solutions therefore con-

    ta ined lead ni t ra te and s l ight ly dis -

    sociated hydrofluoric acid. Th e polarog-

    rap hy of lead, as

    PbF2,

    n l iquid hydro-

    fluoric acid was invest iga ted briefly by

    Clifford

    (2).

    Mesar ic and Hum e

    ( 7 )

    used polarography to s tu dy lead f luoride

    complexes and their solubi l i t ies ; the

    solut ions they s tudied were 1mJf in

    P b + * , u p t o

    0.7M

    in

    F-,

    and of 2M

    ionic s t rength adjus ted by adding

    KaC104. Wi th

    a D.M.E. of

    Teflon,

    Headridge and associates s tudied the

    polarographic behavior

    of

    the ions

    of

    a

    num ber of e lements, including lead, in

    0.1X NH4F-O. lM

    HF

    4 )

    and polaro-

    graphical ly determined molybdenum in

    niobium-base alloys using

    0 . 5 X

    HF-

    0.5M H2S04 as sup port ing medium 5) .

    W i t h t h e D.M.E.of Te flon, the polaro-

    graphic characterist ics of Pb + 2 in

    aqueou s hydrofluoric acid solutions have

    now been s tudied in some detai l . Th e

    data given herein are the f i rs t polaro-

    graph ic da ta repor ted fo r P b+2

    in

    aqueou s solutions of reasonably high

    HF

    concen t rat ion and t ha t con ta in no o the r

    add ed constituents-e.g. , a salt to con-

    tr ibute ionic s t rength or a maximum

    suppressor.

    EXPERIMENTAL

    R e a g e n t s . S t a n d a r d

    Solutions

    of

    P b + 2 n

    Aqueous

    HF.

    A q u a n t i t y

    of

    l ead f luor ide , F bF z ,

    B &

    X purified

    g r a d e ) c a l cu l a te d t o g i ve 1 m X

    solu-

    t i o n w a s a d d e d t o a n a q u e o u s s o lu t i o n

    of HF (1.00, 4.00,

    8.00,

    o r 1 2 . 0 0 M ) .

    Af te r m axim um d is so lu tion the so lu -

    tion was filtered to remove undissolved

    P bF 2. The P b+* concen t rat ion

    of

    the

    filtered solution was determined by

    a

    separate analysis in which

    HF

    was

    removed from tes t port ions by fuming

    VOL.

    37,

    NO.

    1 1 ,

    OCTOBER

    1 9 6 5

    1355