objectives of canadian infusion pump survey to explore the existing level of infusion pump problems...

14
Objectives of Canadian Infusion Pump Objectives of Canadian Infusion Pump Survey Survey To explore the existing level of infusion pump problems in Canadian Hospitals To obtain data on pump problems experienced in Canadian Hospitals To prioritize infusion pump issues To develop strategies for safer infusion pump use equipment and practice

Upload: maximillian-hicks

Post on 22-Dec-2015

220 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

Objectives of Canadian Infusion Pump Objectives of Canadian Infusion Pump SurveySurvey

To explore the existing level of infusion pump problems in Canadian Hospitals

To obtain data on pump problems experienced in Canadian Hospitals

To prioritize infusion pump issues To develop strategies for safer infusion

pump use equipment and practice

Page 2: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems
Page 3: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

ISMP Canada Infusion Pump Safety Project

Pump Problems Encountered (n=340)

85(25%)

255(75%)

0

50

100

150

200

250

300

No YesResponse

Nu

mb

er o

f R

esp

on

ses

Page 4: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

ISMP Canada Infusion Pump Safety Project

Pump Problems Encountered by Hospital Type (n=340)

3 (50%)6 (14.6%)

34 (22.8%)

42 (29.2%)

102 (70.8%)

115 (77.2%)

35 (85.4%)

3 (50%)

0

20

40

60

80

100

120

140

Community (Non-Teaching)

Community (Teaching) Teaching Unknown

Hospital Types

Nu

mb

er

of

Re

sp

on

se

s

No Problems (Total = 85) With Problems (Total = 255)

Page 5: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

ISMP Canada Infusion Pump Safety ProjectTop 10 Pump Problems (n=340)

91

6965

52 52 50

148 147

99 97

0

20

40

60

80

100

120

140

160

False

Air

in L

ine

Flow R

ate-

Inco

rrect

Rat

e

Flow R

ate-

OD

Tubing p

ositio

n-Fre

e Flo

w

Flow R

ate-

Fre

eflo

w

Stopped

/ Shut O

ff

Syrin

ge In

corre

ctly

Engag

ed

Door Open

Batte

ry F

ailu

re- N

o Ala

rm

No Fre

e Flo

w Mec

hanis

m

Major Problem Types

Nu

mb

er o

f R

esp

on

ses

Page 6: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

ISMP Canada Infusion Pump Safety ProjectPump Problems Reported in Percentages

22.9

42.9

58.2

38.8

43.9

28.6

47.1

57.9

35.6

47.4

0

10

20

30

40

50

60

70

Free Flow Flow Rate Over Dose Air Sensor Other

Major Problem Types

Nu

mb

er o

f R

esp

on

ses

(%)

Total Aggregate (n=340) Ontario Responses (n=98)

Page 7: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

ISMP Canada Infusion Pump Safety ProjectTop 5 Desired Pump Improvements (Total=231)

88

118 115 112

93

0

20

40

60

80

100

120

140

A B C D E

Nu

mb

er

of

Re

sp

on

se

s

Type of ImprovementA. On screen programming instructionsB. Protocol programming C. Ability to program clinical minimum/maximum drug concentrations or dosing parametersD. Clarity of displayE. Ability to program ‘primary’ versus ‘secondary” infusion rates

Page 8: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

Comments ReceivedComments ReceivedFree Flow (26): “set removed from pump without clamping” Tubing:

“if tubing is removed incorrectly during cleaning damage may occur with pieces of tubing not visible remaining inside mechanism. If new tubing inserted freeflow situation can occur”

“tubing stretched out too much” “Free flow mixing with primary and secondary

IV dose occur”

Page 9: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

Comments ReceivedComments ReceivedOverdose (23): “set removed without clamping” “with a CADD pump” Equipment:

“defective part”“dirty detection sensor”“failure of pumping mechanism-too much fluid

went through in a short time, one time” “Pump programmed in mLs/ min instead of mLs/ hr”

Page 10: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

Overdose continued: “Epidural concentration set at peripheral rate” “…when there are too many lines and nurse

confuses the settings.” “tubing not positioned properly in pumping

mechanism” “...syringe loaded incorrectly and pump read

incorrect size of syringe and continued to infuse.”

Comments Received

Page 11: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

Incorrect Flow Rate (34): Battery “suspected source undetermined” “patient initiated” “primary infusing instead of secondary” EMI Equipment

Comments Received

Page 12: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

Incorrect Flow Rate continued: “PCA-wrong end of tubing attached to

medication bag-pump did not alarm” Incorrect input of medication concentration “Lines switched to wrong channel” “Problem with one brand of syringes which

was used in the syringe pumps. Pump would miscalculate syringe size….”

Comments Received

Page 13: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

False Air in Line (16): “have had incidents where the tubing

was completely empty and the alarm didn't go off and other times, the alarm would go off because of a tiny bubble of air hardly visible to the human eye”

“false air in line alarm ,especially when running albumin”

Comments Received

Page 14: Objectives of Canadian Infusion Pump Survey  To explore the existing level of infusion pump problems in Canadian Hospitals  To obtain data on pump problems

Smart PumpsSmart Pumps

Medley by Alaris Medical Systems Colleague CX by Baxter Healthcare

Corporation B. Braun