omni vision

Upload: nurwnt

Post on 05-Apr-2018

217 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/31/2019 Omni Vision

    1/27

    The Universal PlatformAcquisition Workstation ---

    OmniVision

    2009.Jun.10th

  • 7/31/2019 Omni Vision

    2/27

    Introduction

    Function Explanation

    Daily Workflow

    Image Post Processing -- ANALYTICA

    MMI

    MMI

  • 7/31/2019 Omni Vision

    3/27

    1.Function Explanation

    MMI

    MMI

    There are three ways to start the exam:

    Query patient from DICOM Worklist server and start exam

    Add patient scheduled exam in Local Worklist and start exam

    Emergency

    Click button and immediately start exam without

    insert any patient information and Scheduled Procedure.

  • 7/31/2019 Omni Vision

    4/27

    1.Function Explanation

    MMI

    MMI

    Function of Acquisition

    Acquisition GUI

    Image Preview Area

    1

    2

    3

  • 7/31/2019 Omni Vision

    5/27

    1.Function Explanation

    MMI

    MMI

    Acquisition GUIFunctional Tool Generator interface

    X-ray Technique

    console

    Patient Body Size

    AEC Function

    Dual Detector

    Selection

    Grid On/Off

    Switch ms/mAs

    OmniVision could support the Generator Virtual Console for

    Full System Integration and Retrofit System Integration.

    Current we could support Generator from vendor below :

    - COMED, CPI, EMD, SEDECAL, TOSHIBA

    - DEL MEDICAL --- *Work In Progress

  • 7/31/2019 Omni Vision

    6/27

    1.Function Explanation

    MMI

    MMI

    Image

    Manipulation

    Acquisition GUIFunctional ToolImage Manipulation

    Image Histogram

    Display

    Once image finish Post-Processing, user could apply image

    manipulation on the image.

    Acquired Image Histogram will display as a reference.

  • 7/31/2019 Omni Vision

    7/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Counterclockwise rotation

    Clockwise rotation

    Vertical Flip

    Horizontal Flip

    Invert

    Annotation (Turn On/Off)

    Window Level Adjustment(Right-click could W/L adjustment)

    Apply W/L with selected ROI

    Insert Digital Marker

    Crop Image

    Set Preferred W/L

    Apply Preferred W/L

    Acquisition GUIFunctional ToolImage Manipulation

  • 7/31/2019 Omni Vision

    8/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Detail Patient Information will be shown in this [Information] tab page.

    While switching the different projection series, the image information

    will be updated.

    Image

    Information

    Patient

    Information

    Acquisition GUIFunctional ToolStudy Information

  • 7/31/2019 Omni Vision

    9/27

  • 7/31/2019 Omni Vision

    10/27

    1.Function Explanation

    MMI

    MMI

    Acquisition GUIProcedure View

    Scheduled Procedure View(s)

    Previous / Next

    User could choose the view that they wish to process accordingly.

    After acquired the x-ray image, the thumbnail box will update with the

    processed x-ray image.

  • 7/31/2019 Omni Vision

    11/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Acquisition GUIExam Management

    Acquisition Mode Demo Mode

    Add New View/Projection

    Reject Acquired Images

    (If Rejected Analysis has

    been enable, while click

    this button, operator will

    need to fill up a reason to

    reject image.)

    End Exam to Close Study

    Simulate Hand Switch

    Ready Step.

    Simulate Hand Switch

    Exposure Step.

  • 7/31/2019 Omni Vision

    12/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Review GUIFunction ToolImage Layout

    Double-click on selected study, user could review all the images.

  • 7/31/2019 Omni Vision

    13/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Review GUIFunction ToolImage Layout

    Double-click on selected study, user could review all the images.

    1

    2

    3

  • 7/31/2019 Omni Vision

    14/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Review GUIFunctional ToolImage Layout

    1x1 display layout

    1x2 display layout

    2x1 display layout

    2x2 display layout

  • 7/31/2019 Omni Vision

    15/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Review GUIFunctional ToolManipulation & Measurement

    Manipulation

    & Measureme

    nt Tool

    Image Histogram

    Display

    User could use various Image manipulation tool to review the images.

    For Example, for Orthopeadics doctor, they could apply button, then

    go to [Measurement] page to use to measure the fractured joint

    distance.

  • 7/31/2019 Omni Vision

    16/27

    1.Function Explanation

    MMI

    MMI

    Angle Measurement

    Review GUIFunctional ToolPost Processing

    Zoom In

    Zoom Out

    True Size

    Fit to Screen

    Pan Image

    Magnify Glass

    Regional Enhancement

    Length Measurement

  • 7/31/2019 Omni Vision

    17/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Review GUIFunctional Tool

    Magnify Glass & Regional Enhancement

    Magnify Glass Function

    Regional Enhancement Function

  • 7/31/2019 Omni Vision

    18/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Review GUI Series image

    Series Image

    Previous / Next

    Select the Series Image that wish to review. The Image will shown on the

    Image Review Area.

    Click Left Arrow and Right Arrow to move to Previous or Next Series.

  • 7/31/2019 Omni Vision

    19/27

    1.Function Explanation

    MMI

    MMI Image Preview Area

    Review GUI Exit Review Mode

    Click [Close Study] to exit the Image Review Mode and return to

    Study Management GUI.

  • 7/31/2019 Omni Vision

    20/27

    1.Function Explanation

    MMI

    MMI

    Review GUI Study Management

    Check Sent &

    Print Status o

    f each image

    Manual Delete StudiesManual Send Studies to PACS

    Locked Study will protected

    from Delete action.

  • 7/31/2019 Omni Vision

    21/27

    1.Function Explanation

    MMI

    MMI

    Configuration GUI DICOM Connectivity

    DICOM Connectivity

    DICOM Worklist Server

    DICOM Storage (PACS Server)

    DICOM Print

  • 7/31/2019 Omni Vision

    22/27

    1.Function Explanation

    MMI

    MMI

    Configuration GUI APR Parameter Setting

  • 7/31/2019 Omni Vision

    23/27

    1.Function Explanation

    MMI

    MMI

    Configuration GUI APR Parameter Setting--- 4 - Steps Code Mapping ---

    STEP 1 : Import Hospital Procedure

    Code (CPT Code) in .xls format

    STEP 2 : Confirm all the CPT

    code has been imported

    and displayed in this

    section.

    STEP 3 : Create Procedure

    as similar with

    Code Meaning as

    possible. Then,

    define each pro-

    cedures view.

    STEP 4 : Click Auto Match to

    apply auto-mapping.

    Completed.

  • 7/31/2019 Omni Vision

    24/27

    1.Function Explanation

    MMI

    MMI

    Configuration GUI Reject Analysis Setting

  • 7/31/2019 Omni Vision

    25/27

    2. Daily Workflow

    MMI

    MMI

  • 7/31/2019 Omni Vision

    26/27

    3. Image Processing -- ANALYTICA

    MMI

    MMI

    BY ANALYZING THE DATA FIRST

    ANALYTICA CAN PROVIDE

    Better Visibility and Contrast of Small and Weak Structures

    Better Depiction of Overlapping Structures

    Better Delineation of .. Bone Structures

    Simultaneous Viewing of .. Bone and Soft Tissue

    Reduced Window Width & Window Leveling

    More Robust Dataset Handling

  • 7/31/2019 Omni Vision

    27/27

    3. Image Processing -- ANALYTICA

    MMI

    MMI

    P1 (Parameter 1) : Weighting of structures with a size around 2 pixels.

    - Tuning level of detail features.

    - Value ,detail is clearer and edge is sharper.

    P2 (Parameter 2) : Weighting of structures with a size around 4 pixels.

    - Tuning the sharpness of high-density features.

    - Value , larger structures will be show more detail.

    P3 (Parameter 3) : Weighting of structures with size larger than 4 pixels.

    - Applied to get a proper image contrast and sharpness.

    - Value , image shows more detail but will be smoothed.

    P4 (Parameter 4) : Gamma curve mapping of the image.

    - Adjusted to provide suitable contrast of whole image.

    - Value , image looks more contrast and darker overall.

    P5 (Parameter 5) : Enable will apply Chest Preference LUT Processing;

    Disable will not apply Chest Preference LUT Processing.