operation manual - cole-parmer · pdf fileoperation manual cole-parmer® mpr16 multipurpose...

22
Operation Manual Cole-Parmer ® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE 1065z76_MAN_Rev.1 ©2016 Cole-Parmer Instrument Company, LLC

Upload: hoangminh

Post on 21-Mar-2018

232 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

Operation Manual

Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges

Models 17406-20, -25, -50, -55

CE

1065z76_MAN_Rev.1

©2016 Cole-Parmer Instrument Company, LLC

Page 2: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

Table of Contents

Preface 1

Introduction 1

Unpacking 1

Precautions 1

Setup Conditions 2

Preparation 2

Hazardous Substances 2

Operation & Symbols 3

Features & Advantages 4

Technical Specifications – Centrifuge 5

Technical Specifications – Rotors, Adapters, and Reducers 6

Transport and Set Up 7

Making Sure the Rotor is Balanced 9

Installation/Fixation for a Fixed-Angle Rotor 10

Installation/Fixation for a Swing-Out Rotor 11

Programming & Operation 12

Maintenance & Care 14

Troubleshooting 18

Contact Information 20

Page 3: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

1

Preface

Before starting to use the Cole-Parmer series of multipurpose centrifuges, read through this instruction manual carefully and follow the instructions. The information contained in this instruction manual is the property of Cole-Parmer Instrument Company, LLC; it is forbidden to copy or pass on this information without explicit approval. Failure to follow the instructions and safety information in this instruction manual will result in the expiration of the warranty.

This centrifuge is a laboratory product used to separate substance mixtures of different densities. This centrifuge should be operated by trained personnel only.

Introduction

Cole-Parmer series of benchtop multipurpose centrifuge (MPR16 refrigerated models 17406-20 and -25; MP16 ambient models 17406-50 and -55) offers rugged construction without sacrificing reliable performance. Extra-thick stainless steel chamber and zinc-coated alloy and steel frame increase durability for inorganic separations requiring strong solvents. Industrial-grade inverter drive (VFD) and brushless motor provide a long service life. User-accessible service sections allow onsite maintenance.

Refrigerated models feature a highly accurate PID controller which allows better overall temperature control, achieving the desired set temperature. In addition, the large, efficient CFC-free compressor uses less power and lasts longer.

LED displays show temperature, rpm, and time. Programmable memory supports 10 programs including acceleration and deceleration rates, allowing you to quickly perform routine tasks.

The included fixed-angle rotor (model 17406-09) accommodates eight 50-mL tubes—ideal for civil, environmental, and water supply engineering applications. Increase versatility by using optional rotors; call for details.

Unpacking

Check individual parts against the list of items below. If anything is missing or damaged, please contact Cole-Parmer immediately.

1. Multipurpose centrifuge

2. Fixed-angle rotor (model 17406-09)

3. Rotor installation kit

4. Power cord

5. User manual

Precautions

Observe the safety instructions. Not following these instructions can cause damage. The centrifuge is to be used for its intended use only. Improper use can cause damages, contamination, and injuries with fatal consequences.

The centrifuge should be operated by trained personnel only.

In order to ensure safe operation of the Cole-Parmer centrifuge, general safety regulations must be followed: refer to the regulations in your country.

Page 4: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

2

Setup Conditions

Plug the centrifuge only into sockets which have been properly grounded.

Turn off the centrifuge at the main switch. The mains plug must be freely accessible at all times. Press the STOP key to shut the centrifuge down.

Pull out the power supply plug or disconnect the power supply in an emergency.

Maintain a radius of at least 12 in. (30 cm) around the centrifuge.

Do not place any dangerous substances within the security zone.

Set up in a well-ventilated environment, place on a level and rigid surface with adequate load-bearing capacity.

Preparation

It is the general obligation of the operator to make sure that the proper protective clothing is used. Refer to the "Laboratory Biosafety Manual" of the World Health Organization (WHO) and the regulations in your country.

Do not touch the mechanical components of the rotor and do not make any changes to the mechanical components.

Do not touch the electronic components of the centrifuge or alter any electronic or mechanical components.

Use only with rotors which have been properly installed. Follow the instructions on pages 10-11.

Do not use rotors which show any signs of corrosion and/or cracks. Contact your supplier if the centrifuge and rotors are in any conspicuous condition, e.g. showing signs of corrosion and/or cracks.

Use only with rotors which have been loaded properly.

Never overload the rotor.

Always balance the samples.

Use only rotors and accessories for this centrifuge which have been approved by the supplier. Exceptions to this rule are commercially available glass or plastic centrifuge tubes, provided they have been approved for the speed or the RCF value of the rotor.

Make sure the rotor is locked properly into place before operating the centrifuge.

Implement of special measures which ensure that no one can approach the centrifuge for longer than absolutely necessary while it is running.

If used for foodstuffs machinery, for cosmetics or pharmaceutical products, only use closed or aerosol-tight containers for centrifugation.

Hazardous Substances

Especially when working with corrosive samples (salt solutions, acids, bases), the accessory parts and vessel have to be cleaned carefully.

Do not centrifuge explosive or flammable materials or substances which could react violently with one another.

The centrifuge is neither inert nor protected against explosion. Never use the centrifuge in an explosion-prone environment.

Do not centrifuge inflammable substances.

Page 5: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

3

&

Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms without suitable safety precautions.

If centrifuging any hazardous materials, mind the “Laboratory Biosafety Manual” of the WHO. When centrifuging microbiological samples from the Risk Group II, aerosol-tight biological seals must be used.

For materials in a higher risk group, extra safety measures have to be taken. If toxins or pathogenic substances have gotten into the centrifuge or its parts, appropriate disinfection measures have to be taken. See "Disinfection" on page 15. Highly corrosive substances which can cause material damage and impair the mechanical stability of the rotor should only be centrifuged in corresponding protective tubes. If a hazardous situation occurs, turn off the power supply to the centrifuge and leave the area immediately.

Operation & Symbols

Never use the centrifuge if parts of its cover panels are damaged or missing.

Never start the centrifuge when the door is open.

Do not move the centrifuge while it is running.

Do not lean on the centrifuge.

Do not place anything on top of the centrifuge during a run. This applies also for the rotor landing pad.

Never open the door until the rotor has come to a complete stop and this has been confirmed in the display.

Emergency door release may be used in emergencies only to recover samples from the centrifuge, e.g. during a power failure (see section "Mechanical Emergency Door Release” on page 18).

Do not open the centrifuge, while it is running.

In any case of mechanical breakdown rotor failure, like burst of the rotor, breaking bottles or shaft crack, the centrifuge is not aerosol-tight.

In case of rotor failure, the centrifuge can be damaged. Leave the room immediately.

Symbols used on the centrifuge:

This symbol refers to general hazards.

CAUTION! means that material damage could occur.

WARNING! means that injuries or material damage or contamination could occur.

This symbol refers to biological hazards.

Observe the information contained in the instruction manual to keep yourself and your environment safe.

This symbol refers to information on hazards, described within this manual.

Page 6: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

4

Features & Advantages

Built to last, strong construction yet offering a sleek modern design that will fit into any modern laboratory. No compromise on quality—only the best components are used in the manufacturing process.

User Features Advantages

Large, bright blue LED display Clarity and ease of use

10 acceleration rates, 10 deceleration rates Sample assistance

10 memory programs Multiple users

Timer 0 to 99 minutes and Hold in 10-sec increments Total flexibility

Pulse short run Fast pelleting

Run in speed (rpm) or RCF (G) in 10-rpm increments Accuracy

Sound <60 dB Quiet

Orientation acceleration rate Prevents initial sample side deposits

Refrigerated models PID controlled Accuracy

Mechanical Features Advantages

Extra-thick stainless steel bowl Easy to clean and rust free

Port to lid Tachometer speed certification

Alloy and steel frame with composites Strong light and quiet

World-leading industrial-grade Inverter Reliability and strength

Quality European brushless motor Quiet, cool, and reliable

High technology airflow (ambient models) Cooler running

Safety Features Advantages

Multipoint lid locking Lid safety

Emergency lid-release Power failure release

Lid gas strut Lid dropping safety

Lid lock detection Lid must be shut in order to operate

Imbalance detection Easy “eye” volume loading

Overspeed sensor Safe detection of speed

Set inverter values Electronic safety of speed

Barrier ring Extra metal protection of chamber

Motor overheat sensor Safe motor protection

Note: All centrifuges have user-accessible service sections to access all safety parameters

Page 7: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

5

Technical Specifications – Centrifuge

Model 17406-20 (display indicative only)

Speed: 500 to 6000 rpm, in 10 rpm steps

RCF max: 4800 RCF x g

Max capacity: 1 L

Temperature (refrigerated models): –9 to 40°C (15.8 to 104°F), PID controlled to ±1°C

Acceleration rates: 10 programs

Deceleration rates: 10 programs

Memory: 10 programs

Timer: 0 to 99 minutes and Hold, in 30 second steps

Dimensions (W x H x D): Refrigerated models: 17.7 x 12.4 x 25 in. (45 x 31.5 x 63.5 cm) Ambient models: 15.7 x 12.2 x 19.7 in. (40 x 31 x 50 cm)

Weight without rotor: Refrigerated models: 138 lb (62.2 kg) Ambient models: 70 lb (31.6 kg)

Power: Refrigerated: 110 VAC, 60 Hz (model 17406-20); 230 VAC, 50 Hz (model 17406-25) Ambient: 110 VAC, 60 Hz (model 17406-50); 230 VAC, 50 Hz (model 17406-55)

Rotor included: Fixed-angle model 17406-09. Additional rotors are available upon request.

Page 8: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

6

Technical Specifications – Rotors, Adapters, and Reducers

Rotors

Catalog number 17406-09 17406-11 17406-07 (included) (not included) (not included)

Rotor model BRK5308 BRK5510 BRK5100

Description Fixed angle, polypropylene, Swing out, aluminum Fixed angle, polypropylene 8 x 50 mL 4 x 250 mL 6 x 100 mL

Buckets (set of 4) No Included No

Sealed lid No Available No

Tube size max 30 x 120 mm 45 x 110 mm 85 x 130 mm

Minimum speed 500 rpm 500 rpm 500 rpm

Maximum speed 6000 rpm 4000 rpm 6000 rpm

Maximum RCF x g 4800 2600 4800

Radius max (cm) 12 14.6 12

Sample tube angle 30 0 30

Acceleration time (sec) 35 30 35

Deceleration time (sec) 35 30 35

Autoclaveable (frequency) 121°C (20) 121°C (20) 121°C (20)

Minimum temperature 4°C (20°C max) 4°C (20°C max) 4°C (20°C max) for refrigerated models

Tube Adapters

Catalog number 17406-62 (set of 4)

Model AN0650

Tube type Nalgene™/ Oakridge, round bottom, 50 mL, 29 x 107 mm

Reducers

Catalog number 17406-34 (set of 4)

Model RM15 [15ml]

Description Max tube size 17 x 120 mm

Catalog number 17406-36 (set of 4)

Model RM25 [25ml]

Description Max tube size 25 x 100 mm

Page 9: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

7

Transport and Set Up

Location

The centrifuge should only be operated indoors. The set-up location must fulfill the following requirements:

• A safety zone of at least 12 in. (30 cm) must be maintained around the centrifuge.

• People and hazardous substances must be kept out of the safety zone while centrifuging.

• The supporting structure must be stable and free of resonance.

• The supporting structure must be suitable for horizontal setup of the centrifuge.

• The centrifuge should not be exposed to heat and strong sunlight.

• The set-up location must be well-ventilated at all times.

WARNING!

UV rays reduce the stability of plastics. Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.

Transporting the Centrifuge

• Due to its weight, the centrifuge should be carried by several people.

• Always remove the rotor before moving the centrifuge.

• Always lift the centrifuge on both sides. Never lift the centrifuge by its front or back panel.

• Transport the centrifuge upright and if at all possible in its packaging.

NOTICE

Store the original centrifuge packaging. Contact a shipping company for the transport or inform the customer service.

Always remove the rotor before moving the centrifuge. If you do not remove the rotor, you might damage the centrifuge drive or centrifuge spindle.

Aligning the Centrifuge

The horizontal alignment of the centrifuge must be checked every time after moving it to a different location. The supporting structure must be suitable for horizontal setup of the centrifuge.

CAUTION!

If the centrifuge isn't leveled, imbalances can occur and the centrifuge can be damaged. Do not place anything under the centrifuge feet to level the centrifuge.

Page 10: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

8

Mains Connection

1. Turn off the power supply switch located on the left side of the centrifuge (press "0").

Power supply switch

2. Plug the centrifuge into grounded electrical sockets only.

3. Check whether the cable complies with the safety standards of your country.

4. Make sure that the voltage and frequency correspond to the figures on the rating plate.

5. Establish the connection to the power supply with the connecting cable.

Storage

WARNING!

When removing the centrifuge and accessories from use you have to clean and if necessary disinfect or decontaminate the entire system. If in doubt, contact the supplier.

• Before storing the centrifuge and accessories, it must be cleaned and if necessary disinfected and decontaminated.

• Store the centrifuge in a clean, dust-free location.

• Be sure to place the centrifuge on its feet.

• Avoid direct sunlight.

Shipping the Centrifuge

Before shipping the centrifuge, please bear the following in mind:

• The centrifuge must be clean and decontaminated.

• You must provide proper documentation confirming the decontamination.

• If in doubt, contact the Cole-Parmer.

Page 11: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

9

Making Sure the Rotor is Balanced

In centrifugation, it is vitally important that rotors are balanced correctly. Any slight imbalance when weighing or placing your sample in the rotor could be exaggerated once the centrifuge is spinning.

Please note the examples shown below of correct loading of samples:

Incorrect Incorrect Correct

Page 12: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

10

Installation/Fixation for a Fixed-Angle Rotor

Installation kit includes:

M6 washer

Rotor fixing screw, M6 x 20 mm

Allen key

Step 1. Place the rotor on the Step 2. Place the washer over

shaft of the motor. the central hole on the rotor.

Step 3. Place the rotor fixing screw Step 4. Hand tighten the rotor into the central hole of the rotor. fixing screw with the Allen key.

CAUTION!

Please ensure that the fixing installation is checked and hand-tightened on a regular basis.

Failure to follow these instructions could cause serious damage to the centrifuge and possible injury to persons nearby and would make the warranty invalid.

Page 13: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

11

Installation/Fixation for a Swing-Out Rotor

Installation kit includes:

M6 washer

Rotor fixing screw, M6 x 20 mm

Allen key

Step 1. Place the rotor on the shaft of the motor. Step 2. Place the M6 washer over the central

hole on the rotor.

Step 3. Place the rotor fixing screw into Step 4. Hand tighten with Allen key.

the central hole of the rotor.

Step 5. Ensure buckets are fully inserted

before use

CAUTION!

Please ensure that the fixing installation is checked and hand-tightened on a regular basis.

Failure to follow these instructions could cause serious damage to the centrifuge and possible injury to persons nearby and would make the warranty invalid.

Page 14: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

12

Programming & Operation

Please make sure the lid is kept closed at all times, otherwise temperature control cannot be held. Allow at least 20 minutes for below ambient temperatures to be reached and stabilized. Wipe the bowl regularly to remove moisture that can collect due to the dew point effect.

Note: For above ambient temperatures, please lower the temperature to 4°C first (to work the internal

gases). Once this temperature has been reached, then raise 20°C to 40°C as required.

How to Program the Centrifuge

To set temperature: Press up or down arrows below temp display.

To set speed in rpm: Press up or down arrows below speed display.

To set time: Press up or down arrows below time display.

To set RCF x g: Press and hold the Function button, and press Lid Open button. Note the direction of blue dot in the speed display: it moves to top left for RCF program, bottom right for rpm.

To set acceleration or deceleration rates: Press and hold the Function button, then press Speed Down arrow for acceleration or press Speed Up arrow for deceleration (brake). Both show 0 to 9 inclusive, with 9 being the highest rate.

To set memory and store (first set speed or RCF and time): Press and hold Function button,

then press Pulse button to select numbers (0 to 9 inclusive). Press Time Down arrow to store.

To recall memory (first select memory number, see above): Press and hold Function button,

then press Time Up arrow to recall memory.

Display showing programmed run of 4°C, 1500 rpm, 10 minutes.

Display showing actual bowl temperature (19°C) and programmed run.

Two dots indicate preconditioning run in progress which will run at half the programmed speed.

Page 15: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

13

Before Operation

If the centrifuge has been in operation for some time, over an hour, at the same temperature, then equilibration is not necessary. However, if the centrifuge has just been turned on then we recommend the following for the best and most accurate results.

Below ambient or 18°C (4 to 18°C / 39 to 64°F) – Refrigerated Models Only

Turn on the unit. Select your temperature, make sure the rotor is in position and firmly fixed. Close the lid. Wait at least 20 minutes.

Whilst waiting place your samples, with tubes in refrigerated unit at the desired temperature.

After your set temperature has been reached, use the precondition mode by pressing down the Function button and Start/Stop button together. Two dots will appear on the display showing that precondition mode has been activated.

Precondition mode runs the centrifuge at half programmed speed to ensure full mixing of the chamber air and therefore equilibrates all areas of the rotor and parts for maximum efficiency.

Once this has finished, lift the lid and load samples in position. See page 9 for correct loading procedures.

How to Operate the Centrifuge

Load samples, fit lid (if applicable). Set Speed or RCF x g, then designated time.

Close the centrifuge lid. Press down firmly on both left and right sides.

Press Start button (bottom left), the motor will now accelerate to the designated setting.

When time has counted down (time display right), the rotor will decelerate (brake).

According to your programmed setting (9 being the fastest), the rotor will stop and “OpEn” will appear on the display.

Press Lid open button and retrieve your samples.

You can stop the run at any time by pressing the Start/Stop button.

How to Use the Pulse (short run) Feature

This method allows very short runs mainly for “pelleting” purposes.

Set your desired speed or RCF x g. Time is not relevant.

Load your samples. Fit lid (If applicable).

Press Pulse button (bottom right) keep held down. You will see the Time display counting up in seconds. As long as you keep the button held down, the time will increase.

On releasing the Pulse button, the centrifuge will brake.

Press Lid open button when “OPEN” appears on the display and retrieve your samples.

CAUTION!

Follow all safety instructions on previous pages. Always load samples evenly.

Page 16: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

14

Maintenance & Care

Cleaning Intervals

For the sake of personal, environmental, and material protection, it is your duty to clean and if necessary disinfect the centrifuge on a regular basis. See recommended cleaning intervals:

Rotor chamber .......... Daily or when polluted

Rotor .......................... Daily or when polluted

Accessories ............... Daily or when polluted

Cabinet ...................... Once per month

Ventilation holes ........ Every six months

CAUTION!

Use only approved cleansers. Refrain from using any other cleaning or decontamination procedure than those recommended here. If you are not entirely sure that the intended procedure or cleanser is safe for the equipment, contact the supplier.

Cleaning

Clean centrifuge and accessories as follows:

• Turn off the centrifuge and remove power supply plug.

• Open the centrifuge lid.

• Grab the rotor handle with both hands and lift it vertically from the centrifuge spindle.

• Remove the centrifuge tubes and adapters.

• Use a neutral cleaning agent with a pH value between 6 and 8 for cleaning. Never use caustic cleaning agents such as soap suds, phosphoric acid, bleaching solutions, or scrubbing powder.

• Rinse the cavities out thoroughly.

• Use a soft brush without metal bristles to remove stubborn residue.

• Afterwards rinse with distilled water.

• Dry all of the rotors and accessories after cleaning with a soft cloth or in a warm air cabinet at a maximum temperature of 50°C. Higher temperatures could damage the material and shorten the lifetime of the parts.

• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil. Also treat the cavities with oil.

• Treat the bolt of the swing out rotor with bolt grease.

• Store the aluminum parts at room temperature or in a cold-storage room with the cavities pointing down

CAUTION!

When cleaning, do not allow liquids, especially organic solvents, to get on the drive shaft, the bearings, or the locks. Organic solvents break down the grease in the motor bearing; the drive shaft could lock up.

Page 17: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

15

Disinfection

Disinfect the centrifuge immediately whenever infectious material has spilled during centrifugation. Decontaminate the affected parts immediately. Take other precautions if need be.

The rotor chamber and the rotor should be treated preferably with a neutral disinfectant. If unsure, contact the supplier for questions regarding the use of other disinfectants.

Observe the safety precautions and handling instructions for the cleaning agents used.

Disinfect the rotor and accessories as follows:

• Turn off the centrifuge and remove power supply plug.

• Open the centrifuge lid.

• Grab the rotor handle with both hands and press against the Auto-Lock button. At the same time, pull the rotor directly upwards it from the centrifuge spindle.

• Remove the centrifuge tubes and adapters and dispose of them or disinfect them.

• Treat the rotor and accessories according to the instructions for the disinfectant. Adhere strictly to the given application times.

• Be sure the disinfectant can drain off the rotor.

• Rinse the rotor and rotor li d thoroughly with water and then rub down.

• Dispose of the disinfectant according to the applicable guidelines.

• Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a maximum temperature of 50°C.

• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil. Also treat the cavities with oil.

• Treat the bolt of the swing out rotor with bolt grease.

WARNING!

Infectious material can get into the centrifuge when a tube breaks or as a result of spills. Keep in mind the risk of infection when touching the rotor and take all necessary precautions. In case of contamination, make sure that others are not put at risk.

Page 18: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

16

Decontamination

Decontaminate the centrifuge immediately whenever radioactive material has spilled during centrifugation. In case of contamination, make sure that others are not put at risk. Decontaminate all affected parts immediately. Take other precautions if need be.

Before using any cleaning or decontamination methods except those recommended by the supplier, users should check with the supplier that the proposed method will not damage the equipment.

For general radioactive decontamination, use a solution of equal parts of 70% ethanol, 10% SDS and water.

Disinfect the rotor and accessories as follows:

• Turn off the centrifuge and remove power supply plug.

• Open the centrifuge lid.

• Grab the rotor handle with both hands and press against the Auto-Lock button. At the same time, pull the rotor directly upwards it from the centrifuge spindle.

• Remove the centrifuge tubes and adapters and dispose of them or disinfect them.

• Rinse the rotor first with ethanol and then with deionized water. Adhere strictly to the given application times.

• Be sure the decontamination solution can drain off the rotor. Rinse the rotor and accessories thoroughly with water.

• Dispose of the decontamination solution according to the applicable guidelines.

• Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a maximum temperature of 50°C.

• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil. Also treat the cavities with oil.

• Treat the bolt of the swing out rotor with bolt grease.

WARNING!

Radioactive material can get into the centrifuge when a tube breaks or as a result of spills. Keep in mind the risk of infection when touching the rotor and take all necessary precautions.

Autoclaving

• Clean the rotor before autoclaving and rinse thoroughly with distilled water.

• Remove all accessories (tubes, adapters) from the rotor. Place the rotor on a flat surface.

• Rotors and adapter can be autoclaved at 121°C (250°C) with a maximum permissible autoclave cycle is 20 minutes.

CAUTION!

No chemical additives are permitted in the steam. Never exceed the permitted temperature and duration when autoclaving.

Page 19: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

17

Servicing

We recommend having the centrifuge and accessories serviced once a year by an authorized service technician; contact the supplier for details. The service technician checks the following:

• Electrical equipment

• Suitability of the set-up site

• Door lock and safety system

• Rotor

• Installation/fixation of the rotor and the centrifuge spindle

Shipping and Disposing

For proper disposal of the centrifuge, refer to the regulations in your country. If in doubt, contact the supplier. Also refer to "Transporting the Centrifuge" and "Shipping the Centrifuge" on pages 7-8.

For the countries of the European Union the disposal is regulated by the European Union's Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC.

WARNING!

When removing the centrifuge and accessories from use for disposal, you must clean and if necessary disinfect or decontaminate the entire system. If in doubt, contact Cole-Parmer.

Performance Recommendations

• Always keep buckets and adapters clean and free from moisture at all times, this will help prevent corrosion and keep them in good, safe operating condition.

• Use only supplier-approved accessories with your product.

• Always ensure samples are distributed in a diametric pattern to ensure the unit runs balanced.

• Rotor pins should be kept clean and lubricated on a regular basis.

• It is recommended that your centrifuge is switched off and the lid left open overnight to ensure that any moisture build up can evaporate.

• Always ensure that lids and accessories are correctly tightened before running the unit.

• Ensure that all buckets are present in the rotor when running your centrifuge, as using the unit with an unbalanced rotor can cause distortion and excessive wear.

• Do not use any kind of scourers or abrasive material to clean the buckets and ensure that a pH neutral cleaner is always used.

• It is recommended that regular maintenance and cleaning of the unit and accessories is carried out.

• Be sure to make note of the rotor expiration date which can be found on the underside of your rotor.

Page 20: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

18

Troubleshooting

Mechanical Emergency Door Release

During a power failure, you will not be able to open the centrifuge lid with the regular electric lid release. A mechanical override is provided to allow sample recovery in the case of an emergency. However, this should be used only in emergencies and after the rotor has come to a complete stop

WARNING!

The rotor can still be spinning at high speed. If touched, it can cause serious injuries. Always wait until the rotor has come to a complete stop without braking. The brake does not work when there is no current. The braking process lasts much longer than usual.

Proceed as follows:

1. Make sure the rotor has stopped (view port in the lid).

WARNING!

Never use your hands or other tools to brake the rotor.

2. Pull out the power supply plug.

3. On the bottom of the housing is one black plastic plug which you can pry out of the side plate with a screwdriver. It is located below the front side of the centrifuge. It is reachable with a screwdriver. Pull the release cord attached to it to trigger the mechanical door release. The door will open and the samples can be removed. Open the front door.

Below front side Mechanical Emergency Door Release on the downside

4. Place the centrifuge so that you have access to the downside of the front panel.

CAUTION!

Due to its weight, the centrifuge should be carried by several people. Always lift the centrifuge from both sides.

5. Push the cord back into the centrifuge and mount the plug.

6. Reconnect the centrifuge once the power has been restored. Switch on the centrifuge.

7. Press the OPEN key to have the door locks operative again.

Page 21: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

19

Codes Shown on the LED Display

OPE n – You may now open the lid. WAI t – Wait 5 seconds, press Lid Open again.

L Id? – Close the lid. T Ach o? – Check that your tubes are not too long

and fouling the lid or if anything else is stopping the rotation of the rotor.

bAL? – An imbalance has occurred. Please check the tubes are diagonally opposite and of the same weight. Check if the tubes are broken; if so use better quality, or if glass reduce the speed to the manufacturer’s recommended G force.

NOTE: If drucH or drLuE (drive) show, please contact Cole-Parmer.

Test Routine (all LED models)

This allows you to turn the buzzer on or off and to check the sensors for Tachometer, Rotor recognition, Lid open, and Imbalance detector. Display shows 0 or 1 as you change settings or perform diagnostic checks.

Step 1. Centrifuge lid must be in open position. Power must be off at rear switch.

Step 2. Press Function button (and keep held down) while turning on at rear switch. When display reads tESt, release Function button.

Step 3. Press Time Up arrow to scroll through test functions. As you scroll, digits in the time segment of the display will change. Use Speed Up arrow to change settings or enter sensor diagnostic menu.

Buzzer off/on (0/1)….……………………………….................................. Time = 34

Speed in 10/100 (0/1) rpm increments………...………….……….…...… Time = 35

EEPROM check…………………………………………….……………..... Time = 36

LED display check. Speed Up arrow enters sensor diagnostics……..... Display = 8888 88

Sensor diagnostics……...……………………………………………...…… Time = 38

A. Rotor recognition/Tachometer sensors – turn rotor until 0 turns to 1 for first two digits of speed display.

B. Imbalance detector – push rotor to one side until 0 turns to 1 in the last digit of speed display.

C. Lid open detector – close lid. Third digit of speed display will turn from 0 to 1.

Buzzer test............................................................................................. Display = Bu2t St

Watchdog test…………………………………………….……………….... Display = Do9t St

“Hello” should display last, then unit automatically exits Test Routine…Display = HELLO

Page 22: Operation Manual - Cole-Parmer · PDF fileOperation Manual Cole-Parmer® MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models 17406-20, -25, -50, -55 CE

20

Error Messages and Troubleshooting

tAChO shown on the display

Note this does not necessarily mean a Tachometer or board problem.

First check that the sample tubes are not too tall and preventing the rotor to move when the lid is closed

Check the rotor and motor for movement; they must rotate smoothly

Check the tachometer on the underside for dust or intrusions to the sensor

If none of the above, contact Cole-Parmer.

bAL shown on the display

An imbalance has occurred.

Check the sample tubes for equal and opposite positioning in rotor.

See “Programming & Operation” section for correct loading and instructions.

Turn off the centrifuge at the power inlet and wait 2 minutes. If bAL still shows on display the detector is damaged.

If none of the above, contact Cole-Parmer.

Speed will not go beyond 2000 rpm

Electronics have set the rotor into safety mode

Check the sensor on top of the motor, check the magnets are clean (wipe with a dry cloth) on underside of rotor.

If none of the above, contact Cole-Parmer.

Noisy

Balance the samples in the rotor, check the lid seals are in place and in good order.

Lid support is weak

Gas strut failure, contact Cole-Parmer.

For Information, Contact:

Toll-Free: 1-800-323-4340

Phone: 1-847-549-7600

Fax: 1-847-247-2929

ColeParmer.com

Note: If you need to contact Cole-Parmer for troubleshooting or repair, please provide the

model number and serial number of your centrifuge.