Part 1. General Instrumentation Concepts

Reading Assignment: Chapter 1 in our textbook.

Page 2: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

What is an Instrument?

• Instruments: Devices that can be used to make a measurement and give quantitative (or sometimes qualitative) results

• Biomedical Instruments:Devices that can be used to make measurements of biological or medical quantities and give quantitative (or sometimes qualitative) results

Page 3: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Examples of FamiliarBiomedical Instrumentation

Clinical Thermometer Stethoscope

Page 4: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Instruments abound in a Neonatal Intensive Care Unit!

Page 5: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

General Instrument Block Diagram

Page 6: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Classification of Instrument Operating Modes

1. Direct vs. Indirect Measurement Methods

Example #1. Direct vs. Indirect temperature measurement

Page 7: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Direct Temperature MeasurementClinical Mercury Thermometer

Bulb is brought directly into contact with the body part or material whose temperature is to be measured, and as the mercury in the bulb exponentially changes to match this temperature, the mercury expands or contracts, pushing the very thin column of mercury up or down along a calibrated scale.

Temperature measurement range: 35 – 42 degrees C

Page 8: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Indirect Temperature Measurement via “non-contact” Infrared (IR) Pyrometer)

Laser beam indicates middle of temperature measurement “spot”. Radius of spot size is given by S = D/10.

Page 9: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

IR Blackbody Radiation: Planck’s Law

• Planck's law describes the radiation spectral density “I” at all wavelengths emitted from a black body at temperature T. As a function of frequency ν, Planck's law is written as:

Note: h is Planck’s constant, k is Boltzmann’s constant, c is the speed of light

1

12),(

2

3

kT

h

ec

hTI

c

Page 10: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Planck’s Law (Radiation Spectral Density) Plotted for Several Values of Temperature

2

1

),(

dTIR

Total Radiation R in Watts/m2

Between band of wavelengths

Page 11: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Typical IR Pyrometer Features:

• Infrared IR Thermometer with Laser Pointer• Non Contact Temperature Measurements in

degree F and degree C• Temperature Range: -25 to 716 °F (-32 to 380

°C)• Accuracy: 2%, Battery: 9V included• Dimensions: 6" x 5" x 1.5"

Page 12: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Direct vs. Indirect Measuring Instrument Example #2:

Direct vs. Indirect Intraocular Eye Fluid Pressure Measurement

Page 13: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Intraocular Pressure (IOP)

• Intraocular pressure (IOP) is a measure of the fluid pressure inside the eyeball (cornea).

• High IOP indicates glaucoma, which is a silent, symptomless debilitating illness that results in tunnel vision and eventually in blindness if it goes untreated with pressure reducing eye drops.

• We should all have our IOP checked regularly, since this is the first indication of glaucoma!

Page 14: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

• Applanation tonometry measures IOP by directly contacting a probe with the cornea of the eye. It measures the the force required to flatten (applanate) a constant area of the cornea.

• Goldmann Applanation Tonometry is considered to be the gold standard in tonometry as it is the most widely accepted method of determining approximate IOP.

• The Goldmann tonometer principle is based on the ”Imbert-Fick Law”: “the pressure in a sphere filled with liquid and surrounded by an infinitely thin membrane is measured by the counterpressure which just flattens the membrane.“

Page 15: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

The Imbert-Fick Law appears in textbooks of optometry but you will NOT find it in Physics textbooks! WHY???

• This is because this pseudo-scientific “so-called” “law” was actually concocted by Hans Goldmann himself to justify the operation of his tonometer!

• This law is really just Newton’s 3rd Law “you press on a stone with a certain force, it presses back on you with the same force.”

• The “applanation” requirement is necessary to ensure that you are measuring only the force due to the fluid pressure inside the eye and not also the additional lateral forces exerted by the tension on the spherical corneal membrane.

Page 16: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Goldmann Applanation Tonometer

Page 17: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Goldmann Tonometer in use

Page 18: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Conversion from probe force in grams to mm Hg• Pressure is equal to force per unit area but one must factor in the density

of mercury (Hg) which is 13.6 gm/cm3.• The probe diameter for a Goldmann tonometer is 3.06 mm. Why was this

strange diameter value chosen by Hans Goldmann? THERE WAS METHOD IN HIS MADNESS, as we see below!The cross-sectional area of the probe is given by Aprobe = π(3.06E-3/2)2 = 0.7354 mm2

• Find the pressure in mmHg if the applanation force on the probe is 1 gmPressure = Force/AreaPressure = 1 gm / .073541714 cm2 x 13.6 gm/cm3 of HgPressure = 1 gm x cm3 of Hg / .07354cm2 x 13.6 gmPressure = 1 cm of Hg / 1Pressure = 1 cm of Hg = 10 mmHg

• Therefore, one gram of force is needed to flatten an area equal to .073541714 cm2 when the eye contains a pressure of 10 mmHg.=> Each large number on the micrometer is in grams, where 1 gram of probe force equals 10 mmHg of IOP

Page 19: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Goldmann Tonometer Operation• A special disinfected prism is mounted on the tonometer head

and then placed against the cornea. • The examiner then uses a cobalt blue filter in front of a slit lamp

(emits a thin sheet of light) to view two green semi circles. • The force applied to the tonometer head is then adjusted using

the dial until the inner edges of these green semicircles meet. • Because the probe makes contact with the cornea, a

topical anesthetic is introduced onto the surface of the eye in the form of an eye drops. Patient must not rub eye until anesthetic wears off to avoid hurting eye, since there is no feeling in eye!

Page 20: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Eyeball is applanated when semicircles (viewed through a slit lamp) are brought into alignment so that their edges touch. That is when IOP = Fplunger/(plunger area)

Page 21: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Patients do not like having their corneas anesthetized!

• Might there be a better way that does not require contact with the cornea?

• Enter the indirect non-contact “Air-Puff Applanation Tonometer”

• Not as accurate as the Goldman contact tonometer, but useful for screening patients to find which ones need to be checked more carefully with the Goldmann tonometer.

Page 22: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Indirect Measurement of IOP: Non-Contact “Air Puff” TonometryAir stream velocity steadily increased until cornea applanates at which time all light rays are collected by the lens and are focused onto the detector, causing detector output voltage to rise above a predetermined threshold, turning off the airstream. IOP is related to airstream velocity at time of applanation. All of this must happen quite literally within “The blink of an eye”; the air puff measurement must be complete before the opthalmic reflex kicks in and causes the patient’s eye to involuntarily close.

Page 23: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Problems with the Air Puff• Abandon the Air Puff

Let’s face it, most patients dread the air-puff tonometer. (The cartoon at left from our colleague, Scott Lee, O.D., in his entertaining book, Sight Gags: Cartoons for Eye Doctors and Their Patients, tells the tale.)

• Although the air-puff has served as a reasonable IOP “screening device” for many years, it is not at all patient-friendly, and patients anticipate the air puff and often close their eyes even before the air puff arrives!

• We have waited nearly 30 years to see a technology evolve that makes the air-puff tonometer obsolete.

Page 24: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Back to Direct Measurement of IOP: Rebound Technology

• The Rebound Tonometer was developed in Finland circa 2006 http://www.icaretonometer.com

• Required a decade of development by Finnish MD Antti Kontiola. “Icare” Company.

• Determines IOP by bouncing a small plastic tipped metal probe against the cornea. Probe is DISPOSABLE, so this eliminates the need for disinfecting the probe.

• The device uses an induction coil to magnetize the probe and fire it against the cornea.

• As the probe bounces against the cornea and back in to the device it creates an induced voltage waveform from which the intraocular pressure is calculated.

Page 25: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Rebound Technology, Cont’d• Software analyzes the probe deceleration, contact time

and other parameters of the probe while it touches the cornea. The deceleration and other rebound parameters of the probe change as a function of IOP. In simple terms, the higher the IOP, the faster the probe decelerates and the shorter the contact time.

• The device is simple, cheap, portable, and easy to use. • Anesthesia is not needed since the touch of the probe is

so gentle; the measurement is barely noticed by the patient!

• It is particularly suitable for children and non-compliant (old and confused) patients.

Page 26: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

The Icare Rebound Tonometer

Page 27: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Other Instrument Classification Categories

Digital

Page 28: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

6. Null-mode vs. Deflection ModeNull-mode Instrument

• The purpose of any null mode instrument is to act like a laboratory balance scale, indicating when the two quantities are equal. The laboratory scale balance beam doesn't actually weight anything; rather, it simply indicates equality between the unknown mass and a pile of standard (calibrated) known masses.

Balance beam acts as a “null detector”, its scale need not be accurately calibrated, but it must accurately indicate the “null” or balance condition.

Page 29: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Null-Mode Voltmeter

The voltage across R2 is to be measured in the “high resistance” circuit consisting of the battery, R1 and R2. Adjust voltage source until no click is heard in headphones when switch is operated. At this point the adjustable source voltage is equal to the voltage to be measured (the instrument is “nulled”. The audio transformer increases the input impedance of the headphones, but note that this audio transformer DOES NOT load down (alter the voltage in) the circuit being measured when no click is heard, since at that point there is 0V across the transformer, and so NO CURRENT flows through it, no matter how low its impedance!

Page 30: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Null detector can be any kind of voltage/current sensing device. It need NOT read voltage accurately or linearly, but it must indicate when the voltage across it goes to 0.If possible, the sensitivity of the null detector should be increased as a null is approached to get the most accurate null possible.

Notes:1.Impedance of null detector element need NOT be high, since it does not load down the circuit whose voltage is being measured when it reads “0” (nulled condition). The null detector need not read accurately.2.Likewise, the voltmeter used to measure the adjustable voltage source value need NOT be of high impedance, since it is across a voltage source, not across the high resistance in the circuit. However, it must be very accurate.

Page 31: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Deflection Mode Instrument

• While a null-mode instrument is as accurate as its known standard value that the unknown quantity is balanced against, it is an iterative process that can take time to complete.

• The deflection-mode instrument is faster but less accurate. The best example of a deflection mode instrument is a spring-loaded scale that measures weight.

Page 32: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Deflection Mode Instrument Example #1: Hanging Spring Scale

Spring scale operation is based upon Hooke’s Law for a spring:

Fspring = kx

Where k is the “spring constant” (N/m), and x is the deflection of free end of the spring from its 0-force (equilibrium) position.

Approximate deflection is read along analog scale

Page 33: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Deflection Mode Instrument Example #2: D’Arsonval Ammeter Movement

Developed in 1882 Jacques-Arsène d'Arsonval with a stationary permanent magnet and a moving coil of wire, suspended by coiled hair springs. The concentrated magnetic field and delicate suspension made these instruments sensitive and they could be mounted in any position. By 1888 Edward Weston had brought out a commercial form of this instrument, which became a standard component in electrical equipment. This design is almost universally used in moving-vane meters today. PERHAPS NO OTHER INSTRUMENT THAT WAS DESIGNED A CENTURY AGO IS STILL USED IN ITS SAME ORIGINAL FORM!

Page 34: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

D’Arsonval Ammeter Internal Structure

Page 35: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 36: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 37: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Because the D’Arsonval meter movement consists of a spring and a mass, its governing equations for needle movement are described by a 2nd order differential equation. The meter movement is inherently underdamped without adding some sort of damping mechanism. Thus the step response of the meter can be oscillatory about the final needle resting place; when the meter is suddenly connected across a dc voltage level, the needle will overshoot the final value, then undershoot, and gradually die down to read the desired value, which is very annoying for the user of the meter. To sufficiently damp the system so that the needle does not take long to settle, friction must be introduced. However this slows down the meter response. A better solution is apply more friction when the needle moves fast and less friction when the needle moves slow. This can be done using “Dynamic Aluminum Vane Damping” shown on next slide.

Page 38: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Dynamic Needle Damping: Aluminum damping vaneAs aluminum closed loop moves in the dc magnetic field, an

induced current flows in the loop, creating an opposing torque that is proportional to velocity of the needle. This serves to

damp the speed of the needle when it is moving fast, but not damp it so much when it is moving slow. This reduces the

chance of needle oscillation about its equilibrium point.

Page 39: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 40: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

+Vo/2

-

/+Vo/2-2-

+Vo/2

-

Page 41: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 42: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Interfering and Modifying Inputs, Cont’d

Page 43: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 44: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 45: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Temperature, Age, etc.

Page 46: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 47: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Practical Signal Notch (Band Reject) Filtering Example

• 60 Hz ac power-line interference is always present in indoor biopotential measurements.

• In special situations, this kind of interference can be neglected, but this is not a general rule.

• In laboratory experiments and clinical analysis, it is hard (and expensive) to isolate the subject of measurement from electrical fields produced by a power line.

• In human biopotential recordings, it is common practice to apply a 50/60 Hz notch filter to reduce this kind of interference.

• In such cases, there is no considerable distortion observed on the recorded signal

Page 48: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Bootstrapped High Q Twin-”T” 60 Hz Notch Filter

See Bootstrapped Twin “T” Frequency Response. (Disconnecting wire from output of op amp to junction of C3and R3 and grounding junction of C3 and R3 yields much broader (Low Q) Twin “T” Response.

Page 49: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 50: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Noise-Cancelling Stereo Headphones

J1 is microphone jack that is directed out Into the noisy environment – it picks up Noise only, not the desired audio signal.J2 is audio input jack and J3 is the audio output jack.

R23 is audio volume control, R14 is noise cancelling adjustment,

S1 selects between noise cancelling mode where external environment cannot be heard, and noise amplifying mode, where external environment CAN be heard.

Page 51: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 52: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 53: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Static Characteristics of Instruments (DC response)

• To measure static characteristics of an instrument, user must wait for as long as necessary for the instrument reading to settle before the measurement is read.

Page 54: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Static Characteristics of Instruments• Accuracy

– Accuracy = (True Value – Measured)/(True Value)– True value is usually traceable back to a value archived at the NIST (National Institute of Standards and

Technology) at Boulder Colorado.– Example: Time and Frequency Standards Division of NIST

WWVH (Hawaii) WWVB (Fort Collins, Colorado)

Page 55: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

NIST Time and Frequency Division

• The Time and Frequency Division, part of the NIST Physics Laboratory, maintains the standard for frequency and time interval for the United States, provides official time to the United States, and carries out a broad program of research and service activities in time and frequency metrology.

Page 56: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

• NIST Shortwave radio station WWVH broadcasts time and frequency information 24 hours per day, 7 days per week to listeners worldwide. The station is located on the Island of Kauai, Hawaii on a 12 hectare (30 acre) site near Kekaha at Kokole Point. The information broadcast by WWVH includes time announcements, standard time intervals, standard frequencies, UT1 time corrections, a BCD time code, geophysical alerts, marine storm warnings, and Global Positioning System (GPS) status reports. WWVH operates in the high frequency (HF) portion of the radio spectrum and radiates 10,000 W on 5, 10, and 15 MHz, and 5000 W on 2.5 MHz. Each frequency is broadcast from a separate transmitter. Although each frequency carries the same information, multiple frequencies are used because the quality of HF reception depends on many factors such as location, time of year, time of day, the frequency being used, and atmospheric and ionospheric propagation conditions. The variety of frequencies makes it likely that at least one frequency will be usable at all times.

• NIST VLF (Very Low Frequency) radio station WWVB is located on the same site as WWV near Fort Collins, Colorado. The WWVB broadcasts are used by millions of people throughout North America to synchronize consumer electronic products like wall clocks, clock radios, and wristwatches. In addition, WWVB is used for high level applications such as network time synchronization and frequency calibrations. WWVB continuously broadcasts time and frequency signals at 60 kHz. The carrier frequency provides a stable frequency reference traceable to the national standard. There are no voice announcements on the station, but a time code is synchronized with the 60 kHz carrier and is broadcast continuously at a rate of 1 bit per second using pulse width modulation. The carrier power is reduced and restored to produce the time code bits. The carrier power is reduced by 17 dB at the start of each second, so that the leading edge of every negative going pulse is on time. Full power is restored 0.2 s later for a binary “0”, 0.5 s later for a binary “1”, or 0.8 s later to convey a position marker. The binary coded decimal (BCD) format is used so that binary digits are combined to represent decimal numbers.

Page 57: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

NIST Development: Miniature Atomic Clock

• NIST researchers have demonstrated a minuscule atomic clock with inner workings about the size of a grain of rice and potential applications in atomically precise timekeeping in portable, battery-powered devices for secure wireless communications, more precise navigation, and other applications.

Page 58: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

• Precision --- # of distinct alternatives from which the result is selected.– Example: 3 digit vs 4 digit digital scale

153153.7

• Resolution --- Smallest change that can be measured between two nearly equal quantities. – Example: 2 pounds for 3-digit scale

0.2 pounds for 4-digit scale• Repeatability --- Variation in measured values

when reading of the same measurand is repeated again

and again. – Example: Repeatablility might be specified as

+ or – 0.1 pound for 4-digit scale.

Page 59: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

• Statistical Error Control– Guaranteed error bounds on errors that are both systematic and random.– Systemic errors can be controlled by calibration and correction factors– Random errors can be controlled by averaging repeated measurements

Systemic errors are associated with a particular instrument or experimental technique (e.g., a faulty ruler, inaccurate instruments, misaligned gun sights, etc.). And these errors always bias the results in one direction, either too high or too low.Random errors on the other hand are unbiased as they come from unpredictable sources (e.g., quiver while holding the device, lab technician getting tired, inconsistent gunpowder quality, etc.). The results obtained are EQUALLY likely to be too high or too low.

Page 60: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Static Sensitivity–If just one input is varied, with other inputs held constant (desired inputs, modifying inputs, and interfering inputs), the static sensitivity of the output (the measured quantity) to that particular input is defined as:

int_ PoQInput

OutputySensitivit

--Evaluation about a Q point is important when the output vs. input variation is not linear. If it is linear, any Q point will yield the same sensitivity.

Page 61: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Linear Least Squared Error Curve Fitting (Linear Regression) => Finds the “best fit” of the transducer’s output vs. input behavior to a straight line, y = mx+b. In a linear instrument, b must = 0. Thus the bias, or “Zero Offset” (b) that is read at transducer output may be removed by subtraction, and the slope “m “ is the static sensitivity “k”

Page 62: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Q-Point must be specified for a nonlinear transducer

NOTE: Sensitivity at Q-Point #1 > Sensitivity at Q-Point #2

QQ-Point#1 Point

#1 Q-Point#2#2

Page 63: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Zero Drift and Sensitivity Drift

• Zero Drift – Change in modifying inputs such as change in dc offset at electrodes

• Sensitivity Drift - Change in gain of instrument (transducer or amplifier) due to temperature or pressure changes, DC Power supply change, etc.

Page 64: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Linearity

• Note: the behavior of a linear transducer is characterized by y = mx + b

where m = slope = static sensitivity (gain) b = zero offset = measurement “bias”.

Note that this bias must be made = 0 via subtraction (calibration) in order to make the instrument linear according to the definition

above.

Page 65: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Three Linearity Specification Methods

• “±B% of full scale” (pipe) accounts best for linearity deviations near zero input• “±A% of reading” (cone) accounts best for linearity deviations near full scale• “±A% of reading or ±B% of full scale, whichever is greater” (funnel) accounts best

for all linearity deviations.

Page 66: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 67: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Transducer Impedance• Indicates degree to which instrument “loads”

(decreases) quantity being measured.• For any transducer input quantity xd1, there is a related input

quantity xd2, such that xd1*xd2 = Power delivered to transducer

• Call xd1 the “Effort Variable” (quantities such as voltage, pressure, force)

• Call xd2 the “Flow Variable” (quantities such as current, fluid flow rate, velocity)

• We define transducer impedance as Zx = xd1/xd2 = effort variable/flow variable

• The power absorbed by the transducer is P = xd1*xd2 => P = (xd22)*Zx => P = xd12 / Zx

• We desire to minimize the power absorbed by the transducer so the transducer does not disturb the quantity being measured. Therefore to measure effort variables (xd1) we want Zx to be relatively high; to measure flow variables (xd2) we want Zx to be relatively low.

Page 68: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Typical Nonlinear Static Transducer Behaviors

Page 69: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 70: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 71: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 72: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 73: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 74: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 75: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Two kinds of potentiometers (pots): linear slider and rotational (Carbon and Wirewound)

Page 76: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

R1

R2

L

EK

x

yySensitivitStatic

L

xE

RR

REy

RRL

xR

RRL

xR

_

21

2

)21)(1(1

)21(2

Linear (fader) potentiometer used as a displacement measuring transducer

Page 77: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 78: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 79: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 80: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 81: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 82: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 83: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 84: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 85: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 86: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 87: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 88: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 89: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 90: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 91: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 92: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 93: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 94: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 95: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Eq 1.33 in text

Page 96: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 97: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 98: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Eqn 1.35 in Textbook

Page 99: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Using MAPLE to solve this 2nd – order Differential Equation

Solution Equivalent to Text Equation 1.33

Page 100: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 101: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 102: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 103: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

τd > 0

Page 104: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 105: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Periodic Signal Bandwidth Determination

Any periodic signal can be expressed as a sum of harmonically related sinusoids. This is known as Fourier Series Decomposition. If f(t) is a periodic function with period T, then

Page 106: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 107: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 108: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 109: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 110: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Maple Fourier Series Analysis of Square Wave

Page 111: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 112: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

General List of Instrument Specifications (Consider this when designing a new instrument!!!)

Page 113: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 114: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook
Page 115: Part 1. General Instrumentation Concepts Reading Assignment: Chapter 1 in our textbook

Biomedical

Instrument

Design

Cycle