performance of rs41 in the utls: improved time-lag and ...€¦ · “the sensor is heated for the...

18
Yann Poltera 1 , Beiping P. Luo 1 , Frank G. Wienhold 1 , Simone Brunamonti 1 , Teresa Jorge 1 , Peter Oelsner 2 , Susanne Meier 2 , Ruud Dirksen 2 , Gonzague Romanens 3 , Manish Naja 4 , Suvarna Fadnavis 5 , Thomas Peter 1 1 Institute for Atmospheric and Climate Science, ETH Zurich, Zurich, Switzerland, 2 Deutscher Wetterdienst (DWD) / GCOS Reference Upper Air Network (GRUAN), Lindenberg, Germany 3 Federal Office of Meteorology and Climatology, MeteoSwiss, Payerne, Switzerland, 4 Aryabhatta Research Institute of Observational Sciences (ARIES), Nainital, India, 5 Indian Institute of Tropical Meteorology (IITM), Pune, India Performance of RS41 in the UTLS: Improved time-lag and bias correction for humidity https://www.nasa.gov/sites/default/files/images/652244main_Thunderstorm-anvil_672px.jpg GRUAN ICM-11 Meeting Singapore, 23.05.2019

Upload: others

Post on 11-Apr-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Yann Poltera1, Beiping P. Luo1, Frank G. Wienhold1, Simone Brunamonti1, Teresa Jorge1, Peter Oelsner2, Susanne Meier2, Ruud Dirksen2, Gonzague Romanens3, Manish Naja4, Suvarna Fadnavis5, Thomas Peter1

1Institute for Atmospheric and Climate Science, ETH Zurich, Zurich, Switzerland, 2Deutscher Wetterdienst (DWD) / GCOS Reference Upper Air Network (GRUAN), Lindenberg, Germany 3Federal Office of Meteorology and Climatology, MeteoSwiss, Payerne, Switzerland, 4Aryabhatta Research Institute of Observational Sciences (ARIES), Nainital, India, 5Indian Institute of Tropical Meteorology (IITM), Pune, India

Performance of RS41 in the UTLS: Improved time-lag and bias correction for humidity

https://www.nasa.gov/sites/default/files/images/652244main_Thunderstorm-anvil_672px.jpg

GRUAN ICM-11 Meeting Singapore, 23.05.2019

Page 2: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

● StratoClim 2016-2017 campaigns: unique opportunity to test RS41 in the field against a reference sensor under “tropical” conditions

● Nainital (India), July-August & Nov 2016

● Dhulikhel (Nepal), August 2017

● In total, 42 balloon soundings with RS41 and CFH

● 31 of these with high quality CFH measurements

Balloon soundings in the Asian Monsoon UTLS

CFH

COBALD

RS41

ECC

RS41

ECC

COBALD CFH

RS92

Bamboo stick

Balloon Diameter at launch: ~ 2 m Diameter at burst: ~ 8 m

Parachute

Unwinder (60 m)

CFH preparation Refrigerant liquid (CHF3) at -80°C → Dry ice or liquid N2

Mirror calibration

ECC preparation Electrode solutions Conditioning Flow rate …

Brunamonti et al., ACP, 2018

Page 3: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Thin-film capacitor technology Measures RH (indirectly) Heated to about +5°C warmer than ambient air

temperature, humidity T measured + : efficient de-icing, shorter response time, increased sensitivity around 100% RH + : eliminates need of solar radiation correction – : decreased sensitivity at low RH

Combined uncertainty 4% RH (Vaisala datasheet)

RS41 Humidity Measurement

Photos: Gonzague Romanens

Page 4: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Chilled-mirror technology, with cryogenic refrigerant (R23) to cool the mirror.

Measures frost point (directly)

State-of-the art detector for stratospheric H2O measurements (Fahey et al., 2014).

< 10% uncertainty in mixing ratio up to ~28 km (Vömel et al., 2016).

Subject to contamination, e.g. when flying through mixed phase clouds (collection of cloud droplets in the inlet tube) quality checks needed after each flight.

CFH: state-of-the art hygrometer for the UTLS

www.en-sci.com

Page 5: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

RS41 significant improvement over RS92 Some time-lag and bias issues remain

Absolute differences to CFH [% RH] +- 1 std Relative differences to CFH [%] +- 1 std

RS92 RS41 RS92 RS41

~0.25 %RH too moist in LS time-lag

signature

Page 6: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Derivation of the time-lag correction: (i) Basics Diffusive processes in polymers make capacitive sensors relax exponentially in time to a change in RH (Miloshevich et al., 2004):

𝑈𝑈𝑚𝑚 𝑡𝑡 − 𝑈𝑈𝑎𝑎 = (𝑈𝑈𝑚𝑚 𝑡𝑡0 − 𝑈𝑈𝑎𝑎) 𝑒𝑒−𝜅𝜅 𝑡𝑡−𝑡𝑡0

Um = measured humidity Ua = ambient (“real” / ”assumed”) humidity 1/κ = τ = f (T ) = T-dependent time constant • Dirksen et al. (2014) derived 𝑈𝑈𝑎𝑎 for the RS92

applying lab-based κ and using an iterative approach.

• Here, we change the approach and determine • κVaisala from Um=URS41-raw such that Ua→URS41 • κthis-work from Um=URS41-raw such that Ua→UCFH

• This StratoClim data is “feature-rich”, permitting to understand the time-lag correction applied by Vaisala and to derive the parametrization that complies with CFH.

Example: Nainital 06-Aug-2016, launch 17:56:40

50% mixing ratio difference

10 %RH difference

Page 7: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Fitting the τ-vs-T relationship to the CFH reference measurements

Exponential time constant τ apparently applied by Vendor to

correct the time lag of RS41

Derivation of the time-lag correction: (ii) Using all StratoClim data

τRS41 applied empirically by Vendor is similar to τRS92

τGRUAN-41 (measured, T>210K) is

similar

τRS41 found in this work is up to a factor of 2-3 smaller than τRS92 and

fitting a physical Arrhenius relationship

Page 8: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Bias in the LS: “Zero humidity reference” – not a true zero?

10 20 30 40 50 Ambient temperature [°C]

100

80

60

40

20

Ambi

ent R

H [%

]

2.5

2.0

1.5

1.0

0.5

0

Error [%] RS41 Ground Check error @ 0 RH and 150 °C

Thanks to Gonzague Romanens’ support

We measured the T-dependence of the ground check procedure.

Two heated plateaus at 150° and 90°C

“The sensor is heated for the zero humidity reference measurement and corresponding calibration fine-tuning” (Jauhiainen et al., WMO TECO, 2014).

If 150°C is indeed interpreted as “zero-humidity”, then a bias might be introduced.

In the lower and middle troposphere, this is no problem.

In the tropopause region and stratosphere, when interested in mixing ratios, this may cause large biases.

Page 9: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Correction for the “150°C-bias”? All Lindenberg Nainital/Dhulikhel

• Results for the LS (all: 30-50 hPa, -57.6 °C +- 4.7 °C, 0.63 %RH +- 0.27 %RH) • Magnitude of bias correction agrees with a “160°C-bias” assumption. • However, on average we have to correct a moist bias, not a dry bias • Is there an unknown negative sign?

Prior Posteriori Prior Posteriori Prior Posteriori

∆R

H [%

RH

]

∆R

H [%

RH

]

∆R

H [%

RH

] 0.2 0.4

0.0 -0.2 -0.4

0.4

0.0 -0.2 -0.4

0.4

0.0 -0.2 -0.4

0.2 0.2

RS41 - CFH RS41 - CFH RS41 - CFH

Page 10: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

MW41 Protocol This work

Time-lag and bias-corrected results: absolute differences to CFH

Page 11: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Validation: Lindenberg 2016-2018 (32 CFH/RS41 tandems)

Derivation: Nainital 2016 + Dhulikhel 2017

(31 CFH/RS41 tandems)

Time-lag and bias-corrected results: relative differences to CFH

After correcting for time-lag and bias, the majority of profiles is within ± 10 % of CFH from the ground to 4 km above the tropopause.

However, there are individual flights, where RS41 is either too dry or too moist, which cannot be due to time-lag and only to some degree due to bias.

This may indicate some sensor failure (RS41 or CFH).

Validation with Lindenberg data looks good, besides a remaining, unexplained dry bias of ~ 5 %.

Examples of systematic dry bias systematic moist bias

Page 12: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

To conclude… • StratoClim balloon soundings provide feature-rich humidity profiles, offering a “real-world

laboratory”, which allows to derive time-lag and bias corrections for RS41.

• Vaisala’s time-lag parametrization for RS41 is similar to the previous RS92 sensor, although the sensors are different in size and in design. GRUAN’s measurements (down to 210 K) match however well with this parametrization, and the reason for the lower CFH-based time constant estimates in the warmer range of temperatures is yet unknown.

• We derive an estimate of the RS41 time constant based on CFH data down to 190 K, providing more reliable data up to 4 km above the tropopause.

• Unrelated to time-lag, similar to RS92 (Miloshevisch et al., 2009, Dirksen et al., 2014), it is necessary to correct for biases at low temperatures, calling for a better understanding of the sensor model formulation and ground check procedure.

• While the new formulation yields on average tangible improvements of RS41 in the UTLS, individual soundings still show unexplained relative errors w.r.t. CFH of > 20 %.

Thank you! Questions?

Page 13: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Back up slides T Stratoclim T Lindenberg

Page 14: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”
Page 15: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”
Page 16: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”

Examples of corrections

Page 17: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”
Page 18: Performance of RS41 in the UTLS: Improved time-lag and ...€¦ · “The sensor is heated for the zero humidity reference measurement and corresponding calibration fine- tuning”