Introduction to CT Automatic Exposure Control (AEC) CARE Dose4D CT Automatic Exposure Control System: Physics Principles and Practical Hints Shawna L. Rego, Lifeng Yu, Michael R. Bruesewitz, Thomas J. Vrieze, James M. Kofler, Cynthia H. McCollough CT Clinical Innovation Center, Department of Radiology, Mayo Clinic, Rochester, MN CARE Dose4D CT Automatic Exposure Control System: Physics Principles and Practical Hints Shawna L. Rego, Lifeng Yu, Michael R. Bruesewitz, Thomas J. Vrieze, James M. Kofler, Cynthia H. McCollough CT Clinical Innovation Center, Department of Radiology, Mayo Clinic, Rochester, MN How CARE Dose4D Works Topogram The topogram is used by CARE Dose4D to determine the appropriate tube current. If a topogram is unavailable, CARE Dose4D cannot be used. If the scan length extends beyond the range included in the topogram, CARE Dose4D will use the value of the effective mAs for the closest region included in the topogram. Ideally, the kV used to acquire the scan should match that used to acquire the topogram. If the scan is acquired at a different kV than the topogram, the dose optimization may be slightly affected but the overall quality of the exam will not be jeopardized. There is no need to repeat a topogram that was acquired at 120 kV if the scan requires 140 kV, or visa versa. Positioning the patient at isocenter is essential! Otherwise, the attenuation calculated based upon an off-centered topogram will not be accurate. If a patient is centered too low and the tube is positioned under the table (Fig. 5a), the patient appears wider on the topogram than he/she actually is (Fig. 5b), thus misinforming the CARE Dose4D algorithm regarding the patient width. Similarly, centering too high will misinform CARE Dose4D as to the patient size (Fig. 5c). Note: Images in Figure 5 were simulated for demonstration purposes. Base protocol: Each body part has its own reference attenuation value and default modulation type. The base protocol provides information to the scanner regarding the expected body part to be imaged. From this information, the reference attenuation level and default modulation type (angular, longitudinal, or both) are prescribed by the scanner. Adult protocols are based upon a standard 75 kg adult and pediatric protocols are based upon a 20 kg child (typical of a 5-year-old). For pediatric body exams, select a pediatric base protocol for those patients weighing 55kg or less. An adult base protocol should be used for patients weighing greater than 55kg. Neuro exams are based on the age of the patient. Select the pediatric base protocol for patients age six and under. Use the adult base protocol for patients older than six. CARE Dose4D will decrease the tube current appropriately from the adult reference values, according to the patient size. It is essential to design site-specific protocols using the correct base protocol. The selected base protocol not only affects dose utilization, but can also affect the availability of other scan parameters, such as kernel and kV options. When the patient size falls in between adult and pediatric sizes, the base protocol (adult or child) is not critical—the image quality resulting from the tube current modulation at the transitional sizes is very similar as long as the QRM of the pediatric or adult protocol is chosen appropriately. QRM and Tube Capacity: The maximum tube current achieved during modulation is limited by the tube and generator capacity. As patient size or attenuation increases, the scanner will increase the tube current and may reach the limit of the system (Figure 6). If the estimated required tube current for an acquisition exceeds the system limit at any location within the scan length, a warning message will be presented to the user (Figure 7). This message must be addressed before the user can proceed to acquire the scan. Important considerations in the Clinical Use of CARE Dose4D CARE Dose4D terminology Topogram: The name used on Siemens’ CT equipment for the scanned projection radiograph. This low dose, projection image is used to localize the start and end locations of a scan. In CARE Dose4D, the topogram is also used to determine the attenuation levels of the patient. Only one topogram is required (either AP or lateral); the perpendicular view is estimated using a mathematical model. Effective mAs: This technique parameter equals the actual tube current–time product (mAs) divided by the spiral pitch value. In CARE Dose4D, this value is automatically determined by the system according to the selected Quality Reference mAs (QRM), the topogram, and the selected scan protocol. After the topogram has been acquired, the estimated average effective mAs that will be used during the scan is displayed, which may be larger or smaller than the prescribed QRM. After the scan, this value is updated to the actual average effective mAs used for the scan acquisition, which may vary slightly from the average effective mAs displayed before the scan because CARE Dose4D updates the pre-scan estimated values with more accurate values determined during the actual scan acquisition. Quality Reference mAs (QRM): The technique parameter entered by the user to determine the desired level of image quality. The QRM is expressed in terms of effective mAs. The QRM is set equal to the effective mAs that produces the desired image quality on a standard-sized patient. During the scan, the tube current (and therefore the effective mAs) is automatically adjusted to compensate for variations in patient size and attenuation, thereby providing image quality consistent with that obtained using the QRM level for a standard- sized patient, but appropriate to both smaller or larger patients. Base Protocol: A scan protocol available on the scanner that was created by the manufacturer. In addition to the user-selectable scan parameters, a base protocol contains information relating to the anatomical structures within the expected scan range and their corresponding reference attenuation values. This information cannot be viewed or edited by the user. Different base protocols are available for adult and pediatric scans. All user scan protocols are created by modifying a base protocol. CARE Dose4D principles CARE Dose4D performs automatic tube-current modulation in the angular and longitudinal directions and also adapts to different anatomical regions and patient sizes. The first step in using CARE Dose4D is to acquire a topogram of the patient, which the software will use to determine the required tube current for each projection of the entire scan. Based upon the topogram data, the attenuation for one projection (e.g. PA) is measured and the attenuation for the complementary projection (e.g. lateral) is calculated with consideration given to the patient girth and anatomic region (head, thorax, pelvis, etc.). Next, the user checks the CARE Dose4D box in the selected scan of the protocol. Marking this box disables the manual effective mAs entry and allows a value to be entered into the QRM field. The CARE Dose4D software then calculates the estimated optimal tube current for every angular projection. The generator/tube load and system limits are also considered to ensure appropriate tube capacity through out the scan. Finally, during the scan, CARE Dose4D will perform on-line attenuation-based tube current-modulation. This fine-tunes the predicted optimal tube current values (estimated from the topogram data) according to the actual attenuation measured 180 degrees earlier in the tube rotation. This allows the system to adapt to temporally varying levels of attenuation, such as from the injection of contrast media or due to patient motion between the topogram and actual scan acquisition, as well as to optimally modulate tube current in extremely elliptical regions, such as the shoulders. The on-line feedback system reads the transmission values at a given angle and uses that information to predict the optimal tube current for the projection that will occur 180 degrees later in the tube rotation. The system lowers the tube current through areas of the body that have a decreased attenuation relative to the “reference attenuation level” of a standard-sized patient and raises the tube current where the anatomy is more attenuating. Figure 2 shows the measured attenuation and calculated tube current for a scan of the chest, abdomen and pelvis. The algorithm for calculating the required tube current is graphically described in Figure 3 and the configuration menu for altering the strength of the tube current modulation is shown in Figure 4. Figure 2: On-line attenuation-based tube current modulation. The topogram is initially used for estimating the attenuation levels along the PA and lateral directions of the patient and predicting the required tube current for each table position. During each tube rotation of the acquisition, real-time feedback determines higher-order corrections to the estimated tube current for each subsequent rotation. The blue curve represents the attenuation level measured from the on-line feedback system for each table position. The red curve represents the tube current for each table position calculated based upon the attenuation level at that projection angle. The yellow curve represents the average tube current over 360 degrees, which is used to calculate the effective mAs that is displayed on each reconstructed image slice after the acquisition is complete. Figure 3: Graphical description of the algorithm used to calculate the required tube current. The horizontal axis represents the attenuation level relative to a reference attenuation level (1.0). The vertical axis represents the modulated tube current relative to a reference tube current (1.0, corresponding to the QRM prescribed by the operator). The four blue curves represent four modulation strengths. The straight line yields constant image noise, which reaches the maximum allowable tube current for attenuation values that are only moderately larger that the reference — this method is not used by CARE Dose4D. The other three curves represent the three different strengths of tube current modulation (strong, average, weak) allowed by CARE Dose4D. The strength is an operator configurable parameter on the scanner. Figure 4: Modulation strength setting Each scanner is globally configurable to a particular modulation strength (strong, average, or weak) for both slim patients/thin regions and obese patients/thick regions. This setting has a global effect on the operation of CARE Dose4D for the scanner and is not protocol specific. Clinical Tips and Tricks Large patients Large patients require a significant increase in tube current, which may exceed the maximum value available for a given configurations of kV, rotation time and pitch. In this situation, a warning dialog box (Fig. 7) will appear indicating that at some point in the scan, the system will need more tube current than is allowed. The user has the option of continuing with the current settings, which may or may not produce an optimal image, or to cancel the acquisition. Cancelling this message allows the user to modify the scan parameters such that the scan can proceed without any potential compromise in image quality. This warning message only appears when CARE Dose4D is on. Options for increasing the dose in spite of the system tube current limitation include: Decrease the pitch. – This will increase the number of photons per slice without affecting the reconstructed slice thickness, at the expense of increased scan time. Switch to a larger total collimation and then decrease pitch. – This will provide a larger longitudinal coverage for each rotation, decreasing the total scan time. The pitch can then be decreased to increase the number of photons per slice (see above). However, using a larger collimation will increase the thinnest reconstructed slice thickness that is available. Change from 120 kV to 140 kV. – This improves x-ray penetration and tube output. Image contrast will be lowered somewhat and the QRM will need to be adjusted (downward) to compensate for the increase in kV. Increase the rotation time. – This will allow more photons per rotation. Scan times will be increased. Patients with metallic implants Metallic implants, which can cause strong artifacts in the form of dark or bright streaks (Fig. 8), are not uncommon in patients receiving CT examinations. They occur when an insufficient number of photons reach the detector, having been highly attenuated by the dense metal. For larger implants, CARE Dose4D will slightly increase the tube current for the affected projections. As shown in Fig. 8, with an implant (Fig. 8a), the average effective tube current/rotation (blue curve, Fig. 8c) increases in the region over the metal compared to when the metal has been removed (Figure 8b), where the average effective tube current/rotation (pink curve, Fig. 8c) is slowly reduced over the same region. This demonstrates that CARE Dose4D attempts to reduce the streaking artifact caused by the metal by increasing the tube current over the implant. To avoid an excessive increase in dose, which will not fully resolve the metal artifacts, the tube current is not maximized over the entire implant, increasing only a moderate amount. The tube current is not increased for small metal objects, which cause minimal streak artifacts. The following actions may help to improve image quality in patients with metallic implants: Use 140 kV. – Higher kV images may have less metal artifacts due to reduced beam hardening and decreased attenuation through the implant. Use a longer rotation time and smaller spiral pitch, consistent with the scan indication. – This will provide more dose through the implant when CARE Dose4D increases the tube current to the system limit. Appropriately increase the QRM. – A larger tube current will, in some cases, reduce the streaking artifacts to a more acceptable level. It is important to note that increasing the tube current will not completely eliminate the streaking artifact and will always increase patient radiation dose. Check with the manufacture to see if metal artifact reduction software is available. Figure 7: Warning dialogue box that is presented to the user when any desired tube current exceeds the system limit. Figure 8: CT images of a cadaver pelvis (a) with a metallic artificial hip and (b) after removal of the implant. (c) Lateral topogram through a cadaver with a metallic hip implant. The blue curve shows the average effective mAs/rotation with the implant in place. The pink curve shows the average effective mAs/rotation after removal of the implant. The yellow line represents the maximum effective mAs that could be obtained. Figure 5a: Patient is centered too low in the scan field, making the patient appear wider than he/she actually is. The resultant exam may use more dose than is necessary. Figure 5b: Patient is properly centered in the scan field. Note that the tube is positioned under the table in Figures 5a–c. Figure 5c: Patient is centered too high in the scan field, making the patient appear narrower than he/she actually is. The resultant images may be too noisy. Why use AEC in CT? The use of ionizing radiation for medically appropriate imaging procedures far outweighs the small increase in risk. However, prudence demands that users and manufacturers take all reasonable steps to decrease the radiation dose to the patient, especially considering the continued increase in the use of medical imaging. One of the most important technologies used to reduce dose in CT is AEC, which aims to automatically modulate the tube current to compensate for variations in patient attenuation, both between different patients and within any given patient. Within any given patient, there can be strong variations in x-ray attenuation as the tube rotates around the patient, particularly in anatomic regions that are very elliptical in shape, such as the thorax and pelvis. The use of a constant tube current causes the x-ray projections through the most attenuating view angles to primarily determine the noise level in the reconstructed cross sectional image. The projections through less attenuated view angles deliver radiation that contributes to the overall radiation dose, but does not further improve the quality of the final image. Angular tube current modulation allows the scanner to modify the current in real-time, based upon the attenuation at any given projection angle, such that similar noise can be maintained regardless of the projection angle. Angular tube current modulation can provide a dose reduction of 40-50% without compromising image quality. The same concept can be used to adapt the tube current to different anatomic regions and to varied patient sizes in order to produce consistent image quality at the lowest achievable dose. For smaller patients, less tube current, and therefore less dose, is needed to obtain the desired image quality. Larger patients require an increase in tube current to achieve diagnostic image quality, with a necessary increase in the delivered radiation dose. AEC has become a standard dose reduction and image quality tool and is offered in various forms by all major CT manufacturers. Effective use of this important dose management tool requires an understanding of the AEC system implemented by the manufacturer. The four major AEC schemes are summarized in Table I, according to manufacturer. Table I: Summary of the four most common AEC strategies AEC Basic Principles AEC is a generic name for any technique aimed at optimizing dose utilization by adjusting the tube current in real-time to accommodate differences in attenuation due to patient anatomy, shape, and size. The tube current may be modulated as a function of projection angle (Fig. 1a), longitudinal location along the patients (Fig. 1b), or both (Fig. 1c). Purpose The purpose of this exhibit is to extend this basic description of AEC, providing an explanation of the principles of operation for one specific system (CARE Dose4D, Siemens Medical Solutions). The objective is to equip CT operators with the information needed to optimally use this AEC system effectively in both routine and challenging clinical situations. Manufacturer AEC Trade name Image Quality Reference Goal General Electric Auto mA, Smart mA Noise Index Constant image noise regardless of attenuation level, using tube currents within prescribed minimum and maximum values. Toshiba SureExposure Target Image Quality Level Siemens CARE Dose4D Quality Reference Effective mAs Constant image quality regardless of attenuation level, with reference to a target mAs level for a standard-sized patient. Philips DoseRight Reference Image Keep the same image quality as in the reference image, regardless of attenuation level. Figure 6: (a) Tube current modulation as a function of angle and z position. All desired values fall below the maximum tube current of the system. (b) If the desired tube current exceeds the system limit at any location, the tube current will be restricted to the maximum allowable limit. The user will be informed that the image quality may be decreased locally (i.e. in some regions, see Fig. 7). The operator must decide if such limitation is acceptable or else alter the scan parameters such that the system can accommodate all desired tube current values. Angular tube current modulation Longitudinal tube current modulation Longitudinal and Angular tube current modulation mA during 1 rotation Tube Angle 3 2 1 4 1 2 3 4 Angular Modulation Average mA per rotation Scan Distance (z) Longitudinal (z) Modulation (Low Attentuation) Thorax (Moderate Attentuation) Liver (High Attentuation) Shoulder Scan Distance (z) Longitudinal (z) and Angular Modulation mA Scan Distance (z) (Low Attentuation) Thorax (Moderate Attentuation) Liver (High Attentuation) Shoulder FIGURE 1b FIGURE 1c FIGURE 1a 2.0 1.8 1.6 1.4 1.2 1.0 0.8 0.6 0.4 0.2 0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 constant image noise strong average weak constant dose Relative Tube Current Relative Attentuation reference attenuation reference tube current Help Cancel Default Settings Apply OK Breathhold time Breathing time Scan protocol list length mAs adaptation to patient size Decrease for slim patients / regions Increase for obese patients / regions Average Weak Average Strong Examination Configuration 6 15 s s 50 Patient Topogram Application Workflow Messages Contrast FIGURE 4 FIGURE 3 FIGURE 2 Patient X-ray Tube Detectors X-ray Tube Detectors Patient Patient X-ray Tube Detectors FIGURE 5c FIGURE 5b FIGURE 5a Upper System Limit (max tube current) Tube Current Scan Range (z) Angle modulated tube current Tube Current Scan Range (z) Upper System Limit (max tube current) Angle modulated tube current Mean tube current per rotation FIGURE 6 Scan Controller Patient Load Cancel Peak exposure demand exceeds system limit. Locally increased image noise may result. You may - increase the kV - increase the Rotation time - decrease Pitch FIGURE 7 300 250 200 150 100 50 0 0 Effective mAs 100 200 300 400 500 Slice Location (mm) FIGURE 8 a b c

Upload: others

Post on 05-Dec-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Introduction to CT Automatic Exposure Control (AEC)

CARE Dose4D CT Automatic Exposure Control System: Physics Principles and Practical HintsShawna L. Rego, Lifeng Yu, Michael R. Bruesewitz, Thomas J. Vrieze, James M. Kofler, Cynthia H. McCollough

CT Clinical Innovation Center, Department of Radiology, Mayo Clinic, Rochester, MN

CARE Dose4D CT Automatic Exposure Control System: Physics Principles and Practical HintsShawna L. Rego, Lifeng Yu, Michael R. Bruesewitz, Thomas J. Vrieze, James M. Kofler, Cynthia H. McCollough

CT Clinical Innovation Center, Department of Radiology, Mayo Clinic, Rochester, MN

How CARE Dose4D Works

Topogram• The topogram is used by CARE Dose4D to

determine the appropriate tube current.

• If a topogram is unavailable, CARE Dose4Dcannot be used.

• If the scan length extends beyond therange included in the topogram, CAREDose4D will use the value of the effectivemAs for the closest region included in thetopogram.

• Ideally, the kV used to acquire the scanshould match that used to acquire thetopogram. If the scan is acquired at adifferent kV than the topogram, the doseoptimization may be slightly affected butthe overall quality of the exam will not bejeopardized. There is no need to repeat atopogram that was acquired at 120 kV ifthe scan requires 140 kV, or visa versa.

• Positioning the patient at isocenter isessential! Otherwise, the attenuationcalculated based upon an off-centeredtopogram will not be accurate. If a patientis centered too low and the tube ispositioned under the table (Fig. 5a), thepatient appears wider on the topogramthan he/she actually is (Fig. 5b), thusmisinforming the CARE Dose4D algorithmregarding the patient width. Similarly,centering too high will misinform CAREDose4D as to the patient size (Fig. 5c).

Note: Images in Figure 5 were simulated fordemonstration purposes.

Base protocol:• Each body part has its own reference

attenuation value and default modulationtype.

• The base protocol provides information tothe scanner regarding the expected bodypart to be imaged.

• From this information, the referenceattenuation level and default modulation type(angular, longitudinal, or both) are prescribedby the scanner. Adult protocols are based upon a standard 75 kg adult and pediatric protocols are based upon a 20 kg child(typical of a 5-year-old). For pediatric body exams, select a pediatric base protocol for those patients weighing 55kg or less. Anadult base protocol should be used for patients weighing greater than 55kg. Neuro exams are based on the age of the patient.Select the pediatric base protocol for patients age six and under. Use the adult base protocol for patients older than six. CAREDose4D will decrease the tube current appropriately from the adult reference values, according to the patient size.

• It is essential to design site-specific protocols using the correct base protocol. The selected base protocol not only affectsdose utilization, but can also affect the availability of other scan parameters, such as kernel and kV options.

• When the patient size falls in between adult and pediatric sizes, the base protocol (adult or child) is not critical—the imagequality resulting from the tube current modulation at the transitional sizes is very similar as long as the QRM of the pediatricor adult protocol is chosen appropriately.

QRM and Tube Capacity: The maximum tube current achieved during modulation is limited by the tube and generator capacity. As patient size orattenuation increases, the scanner will increase the tube current and may reach the limit of the system (Figure 6). If theestimated required tube current for an acquisition exceeds the system limit at any location within the scan length, a warningmessage will be presented to the user (Figure 7). This message must be addressed before the user can proceed to acquirethe scan.

Important considerations in the Clinical Use of CARE Dose4D

CARE Dose4D terminologyTopogram: The name used on Siemens’ CT equipment for the scanned projection radiograph. This low dose,projection image is used to localize the start and end locations of a scan. In CARE Dose4D, the topogram isalso used to determine the attenuation levels of the patient. Only one topogram is required (either AP or lateral);the perpendicular view is estimated using a mathematical model.

Effective mAs: This technique parameter equals the actual tube current–time product (mAs) divided by the spiralpitch value. In CARE Dose4D, this value is automatically determined by the system according to the selected QualityReference mAs (QRM), the topogram, and the selected scan protocol. After the topogram has been acquired, theestimated average effective mAs that will be used during the scan is displayed, which may be larger or smaller thanthe prescribed QRM. After the scan, this value is updated to the actual average effective mAs used for the scanacquisition, which may vary slightly from the average effective mAs displayed before the scan because CARE Dose4Dupdates the pre-scan estimated values with more accurate values determined during the actual scan acquisition.

Quality Reference mAs (QRM): The technique parameter entered by the user to determine the desired levelof image quality. The QRM is expressed in terms of effective mAs. The QRM is set equal to the effective mAsthat produces the desired image quality on a standard-sized patient. During the scan, the tube current (andtherefore the effective mAs) is automatically adjusted to compensate for variations in patient size andattenuation, thereby providing image quality consistent with that obtained using the QRM level for a standard-sized patient, but appropriate to both smaller or larger patients.

Base Protocol: A scan protocol available on the scanner that was created by the manufacturer. In addition tothe user-selectable scan parameters, a base protocol contains information relating to the anatomical structureswithin the expected scan range and their corresponding reference attenuation values. This information cannotbe viewed or edited by the user. Different base protocols are available for adult and pediatric scans. All userscan protocols are created by modifying a base protocol.

CARE Dose4D principlesCARE Dose4D performs automatic tube-current modulation in the angular and longitudinal directions and alsoadapts to different anatomical regions and patient sizes.

The first step in using CARE Dose4D is to acquire a topogram of the patient, which the software will use todetermine the required tube current for each projection of the entire scan. Based upon the topogram data, theattenuation for one projection (e.g. PA) is measured and the attenuation for the complementary projection (e.g.lateral) is calculated with consideration given to the patient girth and anatomic region (head, thorax, pelvis, etc.).

Next, the user checks the CARE Dose4D box in the selected scan of the protocol. Marking this box disables themanual effective mAs entry and allows a value to be entered into the QRM field. The CARE Dose4D softwarethen calculates the estimated optimal tube current for every angular projection. The generator/tube load andsystem limits are also considered to ensure appropriate tube capacity through out the scan.

Finally, during the scan, CARE Dose4D will perform on-line attenuation-based tube current-modulation. This fine-tunesthe predicted optimal tube current values (estimated from the topogram data) according to the actual attenuationmeasured 180 degrees earlier in the tube rotation. This allows the system to adapt to temporally varying levels ofattenuation, such as from the injection of contrast media or due to patient motion between the topogram and actualscan acquisition, as well as to optimally modulate tube current in extremely elliptical regions, such as the shoulders.The on-line feedback system reads the transmission values at a given angle and uses that information to predict theoptimal tube current for the projection that will occur 180 degrees later in the tube rotation. The system lowers thetube current through areas of the body that have a decreased attenuation relative to the “reference attenuation level”of a standard-sized patient and raises the tube current where the anatomy is more attenuating. Figure 2 shows themeasured attenuation and calculated tube current for a scan of the chest, abdomen and pelvis. The algorithm forcalculating the required tube current is graphically described in Figure 3 and the configuration menu for altering thestrength of the tube current modulation is shown in Figure 4.

Figure 2: On-line attenuation-based tube current modulation.The topogram is initially used for estimating the attenuation levels along thePA and lateral directions of the patient and predicting the required tubecurrent for each table position. During each tube rotation of the acquisition,real-time feedback determines higher-order corrections to the estimatedtube current for each subsequent rotation. The blue curve represents theattenuation level measured from the on-line feedback system for each tableposition. The red curve represents the tube current for each table positioncalculated based upon the attenuation level at that projection angle. Theyellow curve represents the average tube current over 360 degrees, whichis used to calculate the effective mAs that is displayed on eachreconstructed image slice after the acquisition is complete.

Figure 3: Graphical description of the algorithm used tocalculate the required tube current.The horizontal axis represents the attenuation level relative to a referenceattenuation level (1.0). The vertical axis represents the modulated tubecurrent relative to a reference tube current (1.0, corresponding to the QRMprescribed by the operator). The four blue curves represent four modulationstrengths. The straight line yields constant image noise, which reaches themaximum allowable tube current for attenuation values that are onlymoderately larger that the reference — this method is not used by CAREDose4D. The other three curves represent the three different strengths oftube current modulation (strong, average, weak) allowed by CARE Dose4D.The strength is an operator configurable parameter on the scanner.

Figure 4: Modulation strength settingEach scanner is globally configurable to a particular modulation strength(strong, average, or weak) for both slim patients/thin regions and obesepatients/thick regions. This setting has a global effect on the operation ofCARE Dose4D for the scanner and is not protocol specific.

Clinical Tips and Tricks

Large patientsLarge patients require a significant increase in tube current, which may exceed the maximum value availablefor a given configurations of kV, rotation time and pitch. In this situation, a warning dialog box (Fig. 7) willappear indicating that at some point in the scan, the system will need more tube current than is allowed.The user has the option of continuing with the current settings, which may or may not produce an optimalimage, or to cancel the acquisition. Cancelling this message allows the user to modify the scan parameterssuch that the scan can proceed without any potential compromise in image quality. This warning messageonly appears when CARE Dose4D is on.

Options for increasing the dose in spite of the system tube current limitation include:• Decrease the pitch.

– This will increase the number of photons per slice without affecting the reconstructed slicethickness, at the expense of increased scan time.

• Switch to a larger total collimation and then decrease pitch.– This will provide a larger longitudinal coverage for each rotation, decreasing the total scan time.

The pitch can then be decreased to increase the number of photons per slice (see above). However,using a larger collimation will increase the thinnest reconstructed slice thickness that is available.

• Change from 120 kV to 140 kV.– This improves x-ray penetration and tube output. Image contrast will be lowered somewhat and the

QRM will need to be adjusted (downward) to compensate for the increase in kV.

• Increase the rotation time.– This will allow more photons per rotation. Scan times will be increased.

Patients with metallic implantsMetallic implants, which can cause strong artifacts in the form of dark or bright streaks (Fig. 8), are not uncommonin patients receiving CT examinations. They occur when an insufficient number of photons reach the detector,having been highly attenuated by the dense metal. For larger implants, CARE Dose4D will slightly increase the tubecurrent for the affected projections. As shown in Fig. 8, with an implant (Fig. 8a), the average effective tubecurrent/rotation (blue curve, Fig. 8c) increases in the region over the metal compared to when the metal has beenremoved (Figure 8b), where the average effective tube current/rotation (pink curve, Fig. 8c) is slowly reduced overthe same region. This demonstrates that CARE Dose4D attempts to reduce the streaking artifact caused by themetal by increasing the tube current over the implant. To avoid an excessive increase in dose, which will not fullyresolve the metal artifacts, the tube current is not maximized over the entire implant, increasing only a moderateamount. The tube current is not increased for small metal objects, which cause minimal streak artifacts.

The following actions may help to improve image quality in patients with metallic implants:• Use 140 kV.

– Higher kV images may have less metal artifacts due to reduced beam hardening and decreasedattenuation through the implant.

• Use a longer rotation time and smaller spiral pitch, consistent with the scan indication.– This will provide more dose through the implant when CARE Dose4D increases the tube current to

the system limit.

• Appropriately increase the QRM.– A larger tube current will, in some cases, reduce the streaking artifacts to a more acceptable level.

It is important to note that increasing the tube current will not completely eliminate the streakingartifact and will always increase patient radiation dose.

• Check with the manufacture to see if metal artifact reduction software is available.

Figure 7: Warning dialogue box that is presented to the user when anydesired tube current exceeds the system limit.

Figure 8: CT images of a cadaver pelvis (a) with a metallicartificial hip and (b) after removal of the implant. (c) Lateraltopogram through a cadaver with a metallic hip implant. Theblue curve shows the average effective mAs/rotation with theimplant in place. The pink curve shows the average effectivemAs/rotation after removal of the implant. The yellow linerepresents the maximum effective mAs that could be obtained.

Figure 5a: Patient is centered too low in the scan field, making the patient appear widerthan he/she actually is. The resultant exam may use more dose than is necessary.

Figure 5b: Patient is properly centered in the scan field. Note that the tube is positionedunder the table in Figures 5a–c.

Figure 5c: Patient is centered too high in the scan field, making the patient appear narrowerthan he/she actually is. The resultant images may be too noisy.

Why use AEC in CT?The use of ionizing radiation for medically appropriate imaging procedures far outweighs the smallincrease in risk. However, prudence demands that users and manufacturers take all reasonable steps todecrease the radiation dose to the patient, especially considering the continued increase in the use ofmedical imaging. One of the most important technologies used to reduce dose in CT is AEC, which aimsto automatically modulate the tube current to compensate for variations in patient attenuation, bothbetween different patients and within any given patient.

Within any given patient, there can be strong variations in x-ray attenuation as the tube rotates aroundthe patient, particularly in anatomic regions that are very elliptical in shape, such as the thorax and pelvis.The use of a constant tube current causes the x-ray projections through the most attenuating view anglesto primarily determine the noise level in the reconstructed cross sectional image. The projections throughless attenuated view angles deliver radiation that contributes to the overall radiation dose, but does notfurther improve the quality of the final image. Angular tube current modulation allows the scanner tomodify the current in real-time, based upon the attenuation at any given projection angle, such that similarnoise can be maintained regardless of the projection angle. Angular tube current modulation can providea dose reduction of 40-50% without compromising image quality.

The same concept can be used to adapt the tube current to different anatomic regions and to variedpatient sizes in order to produce consistent image quality at the lowest achievable dose. For smallerpatients, less tube current, and therefore less dose, is needed to obtain the desired image quality. Largerpatients require an increase in tube current to achieve diagnostic image quality, with a necessary increasein the delivered radiation dose. AEC has become a standard dose reduction and image quality tool andis offered in various forms by all major CT manufacturers.

Effective use of this important dose management tool requires an understanding of the AEC systemimplemented by the manufacturer. The four major AEC schemes are summarized in Table I, according tomanufacturer.

Table I: Summary of the four most common AEC strategies

AEC Basic PrinciplesAEC is a generic name for any technique aimed at optimizing dose utilization by adjusting the tube currentin real-time to accommodate differences in attenuation due to patient anatomy, shape, and size. The tubecurrent may be modulated as a function of projection angle (Fig. 1a), longitudinal location along thepatients (Fig. 1b), or both (Fig. 1c).

PurposeThe purpose of this exhibit is to extend this basicdescription of AEC, providing an explanation of theprinciples of operation for one specific system (CAREDose4D, Siemens Medical Solutions). The objective isto equip CT operators with the information needed tooptimally use this AEC system effectively in both routineand challenging clinical situations.

Manufacturer AEC Trade name Image Quality Reference Goal

General Electric Auto mA, Smart mA Noise Index Constant image noise regardless ofattenuation level, using tube currentswithin prescribed minimum and maximumvalues.Toshiba SureExposure Target Image Quality Level

Siemens CARE Dose4D Quality Reference EffectivemAs

Constant image quality regardless ofattenuation level, with reference to atarget mAs level for a standard-sizedpatient.

Philips DoseRight Reference Image Keep the same image quality as in thereference image, regardless ofattenuation level.

Figure 6: (a) Tube current modulation as a function of angle and z position. All desired values fall below the maximum tube current of the system.(b) If the desired tube current exceeds the system limit at any location, the tube current will be restricted to the maximum allowable limit. The userwill be informed that the image quality may be decreased locally (i.e. in some regions, see Fig. 7). The operator must decide if such limitation isacceptable or else alter the scan parameters such that the system can accommodate all desired tube current values.

Angular tube current modulation

Longitudinal tube current modulation

Longitudinal and Angular tube currentmodulation

mA during1 rotation

Tube Angle

3

2

1

4

1

2

3

4 Angular Modulation

Average mAper rotation

Scan Distance (z)

Longitudinal (z) Modulation

(LowAttentuation)

Thorax(Moderate

Attentuation)

Liver(High

Attentuation)

Shoulder

Scan Distance (z)

Longitudinal (z) and Angular Modulation

mA

Scan Distance (z)

(LowAttentuation)

Thorax(Moderate

Attentuation)

Liver(High

Attentuation)

Shoulder

FIGURE 1b

FIGURE 1c

FIGURE 1a

2.0

1.8

1.6

1.4

1.2

1.0

0.8

0.6

0.4

0.2

00.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0

cons

tant

imag

e no

ise

strong

average

weak

constant dose

Rel

ativ

e Tu

be

Cu

rren

t

Relative Attentuation

refe

ren

ce a

tten

uat

ion

reference tube current

HelpCancelDefault SettingsApplyOK

Breathhold time

Breathing time

Scan protocol list length

mAs adaptation to patient size

Decrease for slim patients / regions

Increase for obese patients / regions

Average

WeakAverageStrong

Examination Configuration

6

15

s

s

50

Patient TopogramApplicationWorkflow Messages Contrast

FIGURE 4

FIGURE 3

FIGURE 2

Patient

X-ray Tube

Detectors

X-ray Tube

Detectors

Patient

Patient

X-ray Tube

Detectors

FIGURE 5c

FIGURE 5b

FIGURE 5a

Upper System Limit (max tube current)

TubeCurrent

Scan Range (z)

Anglemodulatedtube current

TubeCurrent

Scan Range (z)

Upper System Limit (max tube current)

Anglemodulatedtube current

Meantube currentper rotation

FIGURE 6

Scan Controller

Patient

Load Cancel

Peak exposure demand exceeds system limit.Locally increased image noise may result.

You may- increase the kV- increase the Rotation time- decrease Pitch

FIGURE 7

300

250

200

150

100

50

00

EffectivemAs

100 200 300 400 500

Slice Location (mm)

FIGURE 8

a b

c

Rego-1288249.qxd 11/19/07 3:35 PM Page 1