Page 1: PiCCO ... ...Simple – Safe – Speedy - Specific

1

PiCCO ...

...Simple – Safe – Speedy - Specific

PiCCO-Technology

PiCCO plus Standalone Monitor

Page 2: PiCCO ... ...Simple – Safe – Speedy - Specific

2

Contents

1. What is the PiCCO-

Technology?....................................................................................

2. What are the advantages of the PiCCO-

Technology?....................................................

3. How does the PiCCO-Technology work?.......................................................................

4. How to use the PiCCO-Technology?.............................................................................

5. Which disposables do I need for the PiCCO-Technology?............................................

6. References……………………………………………………………………………………..

7. Where can I get what I

need?..........................................................................................

3

4

6

19

22

24

25

Page 3: PiCCO ... ...Simple – Safe – Speedy - Specific

3

1. What is the PiCCO-Technology?

The PiCCO Technology is a combination of 2 techniques for advanced hemodynamic and volumetric management without the necessity of a pulmonary artery catheter in most patients:

a. Transpulmonary thermodilution b. Arterial pulse contour analysis

t

-∆T

t

-∆T

see also page 8 see also page 16

Page 4: PiCCO ... ...Simple – Safe – Speedy - Specific

4

2. What are the advantages of the PiCCO-Technology?

Thermodilution Parameters• Cardiac Output CO• Global Enddiastolic Volume GEDV• Intrathoracic Blood Volume ITBV• Extravascular Lung Water EVLW*

Pulse Contour Parameters• Pulse Continuous Cardiac Output PCCO• Systemic Vascular Resistance SVR• Stroke Volume Variation SVV

* not available in USA

The PiCCO measures the following main parameters:

Page 5: PiCCO ... ...Simple – Safe – Speedy - Specific

5

2. What are the advantages of the PiCCO-Technology?

Less Invasiveness - Only central venous and arterial access required- No pulmonary artery catheter required- Applicable also in small children

Short Set-up Time - Can be installed within minutes

Dynamic, Continuous Measurement - Cardiac Output, Afterload and Volume Responsiveness are measured beat-by-beat

No Chest X-ray - To confirm correct catheter position no x-ray is necessary

Cost Effective - Less expensive than pulmonary artery catheter technique - Arterial PiCCO catheter can be in place for 10 days or more

- Potential to reduce ICU stay and costs

More Specific Parameters - PiCCO parameters are easy to use and interpret even for less experienced clinical staff

Extravascular Lung Water* - Lung edema can be excluded or quantified at the bed-side* not available in USA

Page 6: PiCCO ... ...Simple – Safe – Speedy - Specific

6

3. How does the PiCCO-Technology work?

Most of hemodynamic unstable and/or severely hypoxemic patients are instrumented with:

The PiCCO-Technology uses any standard CV-line and a thermistor-tipped arterial PiCCO-catheter instead of the standard arterial line.

Central venous line (e.g. for vasoactive agents administration…)

Arterial line (accurate monitoring of arterial pressure, blood samples…)

Page 7: PiCCO ... ...Simple – Safe – Speedy - Specific

7

Central venous line (CV)

Thermodilution catheter with lumen for arterial pressure measurement

• Axillary (A)• Brachial (B)• Femoral (F)• Radial (R), long catheter

Arterial pressure transducer

Configuration

CV

A

B

F

R

Page 8: PiCCO ... ...Simple – Safe – Speedy - Specific

8

a. Transpulmonary Thermodilution

Transpulmonary thermodilution measurement simply requires the central venous injection of a cold (< 8°C) or room-tempered (< 24°C) saline bolus…

CV Bolus Injection

Right Heart

Lungs

Left Heart

PiCCO Catheter e.g. in femoral artery

Page 9: PiCCO ... ...Simple – Safe – Speedy - Specific

9

PiCCO Thermodilution Cardiac Output

After central venous injection of the indicator, the thermistor in the tip of the arterial catheter measures the downstream temperature changes

The cardiac output is calculated by analysis of the thermodilution curve using a modified Stewart-Hamilton algorithm:

Injection

t

-Tb

Page 10: PiCCO ... ...Simple – Safe – Speedy - Specific

10

Global Enddiastolic Volume GEDVIntrathoracic Blood Volume ITBVExtravascular Lung Water EVLW*

ln Tb

injection

At

recirculation

MTt t

e -1

DSt

Tb

* not available in USA

PiCCO Volumetric Parameters

These volumetric parameters are obtained by advanced analysis of the thermodilution curve. (Detailed information and formulas available on request.)

Advanced Thermodilution Curve Analysis

Page 11: PiCCO ... ...Simple – Safe – Speedy - Specific

11

Global Enddiastolic Volume

Global Enddiastolic Volume (GEDV) is the volume of blood contained in the 4 chambers of the heart.

Page 12: PiCCO ... ...Simple – Safe – Speedy - Specific

12

Intrathoracic Blood Volume

Intrathoracic Blood Volume (ITBV) is the volume of the 4 chambers of the heart + the blood volume in the pulmonary vessels.

Page 13: PiCCO ... ...Simple – Safe – Speedy - Specific

13

Extravascular Lung Water*

Extravascular Lung Water (EVLW)* is the amount of water content in the lungs. It allows bedside quantification of the degree of pulmonary edema.

* not available in USA

Page 14: PiCCO ... ...Simple – Safe – Speedy - Specific

14

Intrathoracic Blood Volume, ITBV and Global Enddiastolic Volume, GEDV have shown to be far more sensitive and specific to cardiac preload than the standard cardiac filling pressures CVP + PCWP but also than right ventricular enddiastolic volume. 2,3,5,6,8,9,12,13,22

The striking advantage of ITBV and GEDV is that they are not wrongly influenced by mechanical ventilation and give correct information on the preload status under any condition. 2,3,6,7,8,9,12,13, 22

PiCCO Preload Indicators

Page 15: PiCCO ... ...Simple – Safe – Speedy - Specific

15

Extravascular Lung Water, EVLW* assessment by transpulmonary thermodilution has been validated against dye dilution and the reference gravimetric method.11,16,21,23

Extravascular Lung Water*

* not available in USA

Extravascular Lung Water, EVLW* has shown to have a clear correlation to severity of ARDS, length of ventilation days, ICU-Stay and Mortality and to be superior toassessment of lung edema by chest x-ray.7,8,15,20,23,24

Page 16: PiCCO ... ...Simple – Safe – Speedy - Specific

16

Arterial pulse contour analysis provides continuous beat-by-beat parameters obtained from the shape of the arterial pressure wave.The algorithm is capable of computing each single stroke volume (SV) after being calibrated by an initial transpulmonary thermodilution.

b. Arterial Pulse Contour Analysis

t

-∆T

t

-∆T

Calibration

t [s]

P [mm Hg] SV

Page 17: PiCCO ... ...Simple – Safe – Speedy - Specific

17

CO is calculated as stroke volume x heart rate

Cardiac Output and Systemic Vascular Resistances

As pulse contour analysis continuously measures stroke volume and arterial pressure, cardiac output (CO) and systemic vascular resistance (SVR) are computed as follows:

SVR is calculated as (mean arterial pressure - central venous pressure) / CO

Page 18: PiCCO ... ...Simple – Safe – Speedy - Specific

18

SVV reflects the sensitivity of the heart to the cyclic changes in cardiac preload induced by mechanical ventilation.1,14,17,18,19

SVV can predict whether stroke volume will increase with volume expansion.1,14,17,18,19

Stroke Volume Variation (SVV)

In mechanically ventilated patients without arrhythmia,

Page 19: PiCCO ... ...Simple – Safe – Speedy - Specific

19

1. Connect the injectate-temperature sensor housing to the CV line already in place.

2. Insert a PiCCO arterial thermistor catheter into a large artery, preferable femoral artery, but also brachial / axillary artery and radial artery (with long catheter).

3. Connect the injectate sensor, the arterial catheter’s thermistor and pressure line to your PiCCO monitor.

4. For blood pressure transfer to any bedside monitoring system, connect the cable at the back side of the PiCCO monitor.

5. Now the system is ready to work.

6. For information how to handle your PiCCO monitor, please refer to your accompanying PiCCO Operator’s Manual.

4. How to use the PiCCO-Technology?

Page 20: PiCCO ... ...Simple – Safe – Speedy - Specific

20

How to manage my patient with the PiCCO-Technology?

Management of a patients hemodynamic situation is easily possible by following the therapeutic guideline shown below.+

It was developed out of daily clinical practice, has shown to be successful in over a hundred thousand patients and refers to below listed normal values of indices:

Cardiac Index CI 3.0 – 5.0 l/min/m2

Global Enddiastolic Blood Volume Index GEDI 680 – 800 ml/m2

Intrathoracic Blood Volume Index ITBI 850 – 1000 ml/m2

Stroke Volume Variation SVV 10 %Extravascular Lung Water Index* ELWI* 3.0 – 7.0 ml/kg

* not available in USA+without guarantee

Page 21: PiCCO ... ...Simple – Safe – Speedy - Specific

21

Decision tree for hemodynamic / volumetric monitoring**

*not available in USA**without guarantee

*

+

+

CI (l/min/m2)

GEDI (ml/m2 )or ITBI (ml/m2 )

ELWI (ml/kg)(slowly responding)

>3.0<3.0

>700>850

<700<850

>700>850

<700<850

ELWI (ml/kg)

GEDI (ml/m2)or ITBI (ml/m2)

CFI (1/min)or GEF (%)

<10 >10 <10 <10 <10>10 >10 >10

V+ V+! V+!V+Cat Cat

OK!

V-

>700>850

700-800850-1000

>4.5>25

>5.5>30

>4.5>25

700-800850- 1000

Cat

>5.5>30

>700>850

700-800850-1000

700-800850- 1000

10 10 10 10

V-

V+= volume loading (! = cautiously) V- = volume contraction Cat = catecholamine / cardiovascular agentsSVV only applicable in ventilated patients without cardiac arrhythmia

>700>850

<10Optimise to SVV (%) <10 <10 <10

RESULTS

TARGET

THERAPY

1.

2. <10 <10 <10 <10

Page 22: PiCCO ... ...Simple – Safe – Speedy - Specific

22

PULSIOCATH arterial thermodilution catheter

Injectate temperature sensor housing

Any standard central venous catheter

5. Which disposables do I need with the PiCCO-Technology?

• Specifically designed for less invasive volumetric, hemodynamic monitoring• To be placed with Seldinger Technique• Several sizes available for pediatric and adult patients• Can remain within the patient up to 10 days or longer

Page 23: PiCCO ... ...Simple – Safe – Speedy - Specific

23

PULSIOCATH arterial thermodilution catheters product range

Article number PV2013L07 PV2014L08 PV2014L16 PV2015L20 PV2014L50LGWA

Outer diameter 3F(~20G) / 0,9mm

4F(~18G) / 1,4mm

4F(~18G) / 1,4mm

5F(~16G) / 1,7mm

4F(~18G) / 1,4mm

Usable length 7cm 8cm 16cm 20cm 50cm

Common feature Latex free

The catheters are also available as complete kits (e.g. PVPK2015L20-46), including a disposable pressure transducer and the injectate temperature sensor housing. These kits can optionally be ordered with an additional pressure line for intermittent central venous pressure monitoring.Catheters have to be selected depending on patient size, weight and insertion site.

PULSIOCATH arterial thermodilution catheters are specifically designed for less invasive volumetric hemodynamic monitoring with the PiCCO Technology. The catheters are placed with Seldinger Technique. Several versions and sizes are available. They can remain within the patient up to 10 days or even longer.

Technical specifications are subject to change without further notice

Page 24: PiCCO ... ...Simple – Safe – Speedy - Specific

24

1. Berkenstadt H et al., Anesth Analg, 20012. Bindels A et al., Crit Care 4, 20003. Boussat S et al., Int Care Med 20024. Brock H et al., Eur J Anaesth 19 (4), 20025. Della Rocca G et al., Eur J Anaesth 19, 20026. Della Rocca G et al., Anesth Analg 95, 20027. Eisenberg PR et al., Am Rev Respir Dis 136 (3), 19878. Gödje O et al., Chest 118, 20009. Gödje O et al., Eur J of Cardio-thoracic Surgery 13, 199810. Haperlin et al., Chest, 198511. Katzenelson et al., SCCM 2001, San Diego12. Lichtwarck-Aschoff M et al., Journal of Critical Care 11 (4), 199613. Lichtwarck-Aschoff M et al., Intensive Care Med 18, 199214. Michard F et al., Yearbook of Intensive Care Med, 200215. Mitchell JP et al., Am Rev Respir Dis 145 (5), 199216. Neumann et al., Intensive Care Med 199917. Reuter DA et al., Crit Care Med, 200318. Reuter DA et al., Intensive Care Med, 200219. Reuter DA et al., Brit J Anaesth, 200220. Sakka SG et al., Chest 122, 200221. Sakka S et al., Intensive Care Med 200022. Sakka S et al., Journal of Critical Care 14 (2), 199923. Sturm JA, Practical Applications of Fiberoptics in Critical Care Monitoring,

199024. Takeda A et al., J Vet Med Sci 57, 1995

6. References

Page 25: PiCCO ... ...Simple – Safe – Speedy - Specific

25

7. Where can I get what I need?

For further information or ordering, please contact your local PULSION Office: PULSION Medical Systems AG

Stahlgruberring 28D- 81829 München

GermanyTel.: +49 (0)1805 PULSION

+49 (0)89 4599140Fax: +49 (0)89 45991418

[email protected]

PULSION Medical U.K., Ltd.P.O. Box 315Arundel Road

Uxbridge, MiddlesexGB- UB8 2US, England

Great BritainTel.: +44 (0) 1895 455255Fax: +44 (0) 1895 274035

[email protected]

PULSION France sarl6, Place Jeanne d´Arc

F- 13100 Aix en ProvenceFrance

Tél. : +33 (0)4 42 27 67 19Fax : +33 (0)4 42 27 44 90

[email protected]

PULSION Medical Systems Ibérica S.L.

Pol. Ind. Las Nieves, C/ Puerto Canencia 21

E- 28935 Móstoles, MadridSpain

Tel.: +34 91 665 73 12Fax: +34 91 616 94 27

[email protected]

Seda S.p.A.Via Tolstoi, 7/B

I- 20090 Trezzano sul Naviglio (MI)Italy

Tel.: +39 02 48424219Fax: +39 02 48424290

[email protected]

PULSION Benelux nv/saMaaltecenter, Blok G

Derbystraat 341B- 9051 Gent (SDW)

BelgiumTel. : +32 (0)9 242 99 10Fax : +32 (0)9 242 99 11

[email protected]

PULSION Pacific Pty. Ltd.P.O. Box 823

AUS- Randwick, NSW Australia, 2031

AustraliaTel.: +61 (0) 2 9340 5525Fax: +61 (0) 2 9340 5515

[email protected]

If your country is not listed above, please

contact

PULSION Germany

Page 26: PiCCO ... ...Simple – Safe – Speedy - Specific

26

The PiCCO-Technology is also available as a module for Philips IntelliVue / CMS patient monitoring systems.

For more information please refer to your local PULSION representative (page 25) or visit www.pulsion.com

Page 27: PiCCO ... ...Simple – Safe – Speedy - Specific

27

For further information, please visit

www.pulsion.com

There you will find also information on other revolutionary PULSION products:

LiMON - noninvasive bedside liver function monitor ICG-PULSION - indocyanine green dye for diagnostics VoLEF - PiCCO plus add-on monitor for cardiac sub-volumes and RVEF IC-VIEW - laser fluorescence makes tissue perfusion visible

@ PULSION 07/2003

MPI812205