Page 1: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

Positector DPM Con Sensor IR

Estandar T Aire

DF-DPMIR1E

Page 2: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

Full Guide v. 3.4

PosiTector®DPMDew Point Meter

Page 3: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

Table of ContentsTable of Contents

Introduction ......................................................... 1Quick Start .......................................................... 1Operation Overview ............................................ 3Calibration ........................................................... 5Why is Measurement Important .......................... 6Probes ................................................................. 7Menu Operation .................................................. 9Alarm Mode ......................................................... 9Setup Menu ......................................................... 11

Reset .............................................................. 11Gage Info ....................................................... 12Wet Bulb Temperature ................................... 12Anemometer .................................................. 12Trend Chart .................................................... 12

Memory Management ......................................... 14Standard Models ........................................... 14Advanced Models .......................................... 14Annotate ......................................................... 16Display ........................................................... 16Summaries ..................................................... 17Screen Capture .............................................. 17

Auto Log Mode .................................................... 17Accessing Stored Measurement Data ................ 19

PosiSoft Desktop ........................................... 19PosiSoft.net .................................................... 19PosiSoft Software .......................................... 19PosiTector App. .............................................. 19

Page 4: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

Connect Menu ..................................................... 20WiFi......... ....................................................... 20Access Point................................................... 20USB.......... ...................................................... 22USB Serial Streaming..................................... 23Sync.net Now.................................................. 24Bluetooth Smart................ ............................. 24Bluetooth......................................................... 25Updates .......................................................... 26

Power Supply / Battery Indicator ........................ 26Troubleshooting .................................................. 27Technical Data .................................................... 28Returning For Service ......................................... 29Warranty .............................................................. 29

Page 5: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

1

IntroductionIntroductionThe PosiTector Dew Point Meter (DPM) is a hand-held,electronic Gage that continually measures, calculates, andrecords climatic conditions quickly and accurately. It consists of abody (Standard or Advanced) and probe (page 7).

The PosiTector DPM powers-up when the center navigationbutton is pressed. Readings are updated automatically eachsecond. To preserve battery life, the Gage powers down afterapproximately 5 minutes of no activity. All settings are retained.

Quick StartQuick Start

1. Remove the protective rubber cap from the surfacetemperature sensor.

2. Power-up Gage by pressing the center navigation button.

USB Port

LCD Display

Probe

PosiTector body

Page 6: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

2

RH - Relative Humidity (measured)Ta - Air Temperature (measured)Ts - Surface Temperature (measured)Td - Dew Point (calculated)Ts-Td - Surface minus Dew Point

Temperature (calculated)Tw - Wet Bulb Temperature (calculated)

(Advanced Models Only) V - Wind Velocity (DPM-A model only)

Typical Screen

Wrist StrapWe recommend attaching andwearing the supplied wrist strap.

1

2

Protective Lens ShieldThe LCD is covered with a thin plastic film for protection againstfingerprints and other marks during shipment. This film, whileusually removed before using the Gage, can be left in place toprotect against paint overspray or debris. Replacements can bepurchased.

Protective CapThe PosiTector DPM is shipped with a protective plastic cap overthe surface temperature sensor. Remove this cap prior to use.Replace it when the Gage is not in use to protect the sensor.

DPM-D probes only display RH, Ta, & TdNOTE:

Page 7: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

3

RH - The relative humidity sensor is located inside the probehousing. The sensor may require a longer period to stabilize in theevent of extreme humidity changes.

Ta - The air temperature sensor is also located inside the probehousing. (Also known as Dry Bulb Temperature)

Ts - A surface temperature thermocouple is located on the tip ofthe probe (built-in probe model). It is spring loaded and designedto quickly self-align on a surface. Press it against a surface andallow a few seconds for the value to stabilize. Do not drag the sen-sor over the surface. Wet the surface for a faster response. Coverwith the included protective plastic cap when not in use.

Does not apply to DPM-D models

Operation OverviewOperation Overview

The PosiTector DPM powers-up when the center navigationbutton is pressed. Readings are updated automatically eachsecond. To preserve battery life, the Gage powers down afterapproximately 5 minutes of no button activity. All settings areretained.

Button Functions- Normal Operation

The Pause/Resume (+) button stops values from automaticallyupdating to allow for closer examination of the relationshipbetween the values or to allow time to manually record the entiredataset. Press (+) again to resume.

Scroll throughTrend Charts(Advanced only)

Access the Menu

Pause / Resume

Restore brightnessafter dimming(Advanced only)

NOTE:

Page 8: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

4

Td - The dew point temperature is a function of air temperatureand relative humidity (see ISO 8502-4 Annex A). It is the temper-ature to which a volume of air must be cooled in order to reachsaturation.

Ts - Td - The surface minus dew point temperature is importantin judging the risk of condensation, or dew, forming on a surface.For example, ISO 8502-4 recommends that, unless otherwiseagreed, the temperature of a steel surface generally should be atleast 3°C (5°F) above the dew point when protective coatings areapplied.

Does not apply to DPM-D models

NOTE:

Tw - (Advanced models only) The wet bulb temperature is thetemperature of adiabatic saturation. This is the temperatureindicated by a moistened thermometer bulb exposed to air flow.The wet bulb temperature is calculated from Ta and RH using anair pressure of 1.0 atmospheres (1013mbar). It approximates thetemperature indicated by a moistened thermometer bulb exposedto airflow. Variations of Tw at other pressures could be greater than±1°C (±2°F).

V - Wind velocity is measured with a hot-wire anemometer locatedinside the aluminum housing (DPM-A models only)

The Gage may require a period of time before the above climaticreadings stabilize. This is most noticeable when the Gage isrelocated from one extreme environment to another (i.e. from awarm location to a cold location). A period of up to 30 minutesshould provide adequate time for stabilization.

For surface temperature measurement on steel over anextended of time, consider using the optional Magnetic SurfaceTemperature probe (pg. 7).

NOTE:

Page 9: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

Smart TrendTM Indicators - identify rising, falling or stablereadings. The Gage compares each measurement with recentvalues to provide the current direction of change.

5

Smart TrendIndicators

While the absence of an arrow does not necessarily signifystability, it can be taken to mean that particular reading isundergoing little or no change. Slow changes could thereforeoccur without being noted by an indicator arrow.

CalibrationCalibration is the controlled and documented process ofmeasuring traceable calibration standards and verifying that theresults are within the stated accuracy of the Gage. Calibrationsare typically performed by the Gage manufacturer or by a certifiedcalibration laboratory in a controlled environment using adocumented process.

The PosiTector DPM is shipped with a Certificate of Calibrationshowing traceability to a national standard. For organizations withre-certification requirements, the PosiTector DPM may bereturned at regular intervals for calibration. DeFelskorecommends that our customers establish Gage calibrationintervals based upon their own experience and work environment.Based on our product knowledge, data and customer feedback, aone year calibration interval from either the date of calibration,date of purchase, or date of receipt is a typical starting point.

Written Calibration Procedures are available online at no chargeat www.defelsko.com.

Page 10: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

6

Why is Measurement Important?Why is Measurement Important?For industrial painters, the primary reason for measuring climaticconditions is to avoid rework and the premature failure ofprotective coatings. Recommendations and requirements arecovered under various internationally recognized standards.

Temperature – It is important to measure both surface and airtemperatures as they are often different. Application at incorrecttemperatures can cause defects including: blistering, pinholing,cratering, dry spray & mud cracking. By measuring surfacetemperature it is possible to avoid potential application problems ifair or surface temperatures become too hot or too cold forsatisfactory film formation. Surface and air temperatures directlyaffect the rate of cure of the applied coating or lining system andare major factors in the formation of surface moisture or dew.ASTM D3276 Standard Guide for Painting Inspectors (MetalSubstrates) states that the minimum air/surface temperature isusually 5ºC (40ºF).

Relative Humidity - the amount of moisture in the air expressedas a percent of overall volume. High relative humidity can retardthe rate of solvent evaporation. For this reason, the maximumrelative humidity at which coatings or linings can be applied andcured is generally set at 85%. This can be set as an alarm in theAdvanced model.

Dew Point Temperature - the temperature at which moisture willbegin to form on a surface. The dew point temperature is a functionof air temperature and the relative humidity. (see ISO 8502-4Annex A). It is the temperature to which a volume of air must becooled in order to reach saturation. A number of major standardsorganizations call for careful dew point control. Moisture on afreshly blasted steel substrate will cause the steel to rust. If a thin,invisible film of moisture is trapped between the applied coatingand the substrate, the system will likely fail prematurely. To ensurethat dew point problems do not exist, ISO 8502-4 and SSPCguidelines state that the surface temperature must be a minimumof 3ºC or 5ºF above the dew point during the 3 critical phases ofcoating preparation, application, and cure.

Page 11: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

7

When powered-up, the PosiTector automatically determineswhich probe is attached and does a self-check.To disconnect a probe from a body, slide the plastic probeconnector horizontally (in the direction of the arrow) away from thebody. Reverse these steps to attach a different probe. Itis not necessary to power-down the Gage whenswitching probes.

The PosiTector body accepts a wide variety of probetypes including magnetic, eddy current and ultrasoniccoating thickness, surface profile, shore hardnessdurometer, salt contamination and ultrasonic wallthickness probes. See www.defelsko.com/probes

ProbesProbes

PosiTector DPM probes:

PosiTector DPM / DPM-IRIntegral Surface Temperature Sensor

PosiTector DPM-S Magnetic Surface Temperature Sensor

PosiTector DPM-A AnemometerIntegral Surface Temperature Sensor

PosiTector DPM-DCabled probe with NPT threads

K-TypeConnector

SurfaceTemperature

WindVelocity

Air Temperature &Relative Humidity

SurfaceTemperature

Probe also accepts other K-Typethermocouple sensors (optional)

1/2” NPT Threads

Sintered Metal Tip (protects sensor)

Air Temperature &Relative Humidity

Air Temperature &Relative Humidity

SurfaceTemperature / IR

Sensor

Page 12: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

8

To ensure optimal performance of your PosiTector DPM, donot obstruct the airflow near the air temperature and humidity sensors. Keepfingers away from the sensor, as body heat can cause incorrect readings.Allow time for the Gage to acclimate and for the readings to stabilize whenmoving the Gage between different environments.When using the surface temperature probe, do not use excessive downwardforce, and do not drag it sideways. When the Gage is not in use, place theplastic cap over the surface temperature sensor to prevent damage.

For DPM-A Probes, do not touch or attempt to clean the hot wireanemometer as damage to the sensor can result.

CAUTION:

The Magnetic Surface Temperature Sensor can be disconnectedfrom the K-Type thermocouple connector (DPM-S model only) andreplaced with one of the following optional sensors:Hand-held Surface Temperature Probe

- for spot checking hard-to-reach surfacesSelf-adhering Surface Temperature Probe

- sticks to solid materials including aluminum, wood, concrete,etc. Ideal for unattended monitoring.

Liquid Temperature Probe- general purpose probe with stainless steel shaft

PosiTector bodies accept all PosiTector DPM, 6000, 200, SPG,RTR, SST, SHD and UTG probes

NOTE:

Page 13: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

9

The PosiTector DPM can automatically alert the user whencurrent climatic conditions exceed pre-set values. When selected,the Alarm icon is displayed at the top of the screen. Accordingto ISO 8502-4 and ASTM D3276, steel surfaces should generallybe 3°C (5°F) above the dew point when painting. The Alarm modecan be used in conjunction with Memory and Auto Log modes.

Alarm ModeAlarm Mode

Menu OperationMenu OperationGage functions are menu controlled. To access the Menu, power-up the Gage, then press the center navigation button.

Below is a sample menu for a PosiTector DPM Advanced Model:

Press the (+) button to switch pages in amultipage menu.

Press the (-) button to return to the previousmenu or page.

Select Exit to exit from any menu.

Some buttons have a tick box to theirright to indicate current status. An emptybox indicates that feature is not active.

Current selection is displayedwith lightened background

ü

To navigate, use the Up and Down buttons to scroll vertically and to SELECT.

Previous Page(Exit)

Up

Down

Next Page

Access the Menu(Center/Select)

Page 14: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

10

Advanced models allow the user to set up custom alarmconditions for each reading in a dataset. Use the Up and Downbuttons to scroll through each value and corresponding operator.Advanced models also allow the user to configure the gage tosend alerts via email.

For each reading, select between XX (off), ≥(greater than or equal to), and ≤ (less than orequal to) using the (-) and (+) buttons. Press thedown button and set the value using the(-) and (+) buttons.

In alarm mode, if an alarm condition is met, therelevant reading is displayed in red, and theAlarm icon is displayed beside it (see imageat right). If the “Sound” checkbox was selectedin the Alarm Mode setup screen, an audio tonewill also be emitted.

To configure alarm notification via email, selectemail Setup from the Alarm screen. Select thetick box to enable and enter a valid emailaddress. An email test can be peformed byselecting Test email. Internet connection viaWiFi or through PosiSoft Desktop is requiredfor email Alarms.

Standard models have one, fixed alarm.When selected, the Gage will alert the userwhen the surface temperature is less than3°C (5°F) above the dew point temperature.Enable by selecting the tick box. The Gagewill alert the user with an audio tone and theLCD screen will reverse (see right) whenconditions may be unsuitable for painting.

Page 15: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

11

It performs the same function as a menu Reset plus:

• Bluetooth Pairing info is cleared.• Alarm values are set to default.• Menu settings are returned to the following:

Setup MenuSetup MenuReset

Reset (soft reset) restores factory settings and returns the Gageto a known condition. The following occurs:

• All batches, stored datasets, images and batch names areerased.• Menu settings are returned to the following:

Memory = OFFAuto Log = OFFTrend Chart = NoneAuto Dim = ON

Bluetooth & Stream = OFFWiFi & Access Point = OFFDisplay = NoneAnemometer = ON (DPM-A only)

NOTE: Date, Time and WiFi are not affected by either Reset.

Units (temperature) = CelsiusFlip Display = NormalAuto Sync .net = ONAlarms = OFFAlarm Sound = OFFSound = ONLanguage = EnglishAnemometer = ON (DPM-A only)

Units (velocity) = m/sBattery Type = AlkalineBacklight = NormalBluetooth Smart = OFFUSB Drive= ONAuto Log Interval = 5 minWet Bulb = OFF

Perform a more thorough Hard Reset as follows:1 Power down the Gage and wait 5 seconds.2 Simultaneously press and hold the (+) and center

buttons until the Reset symbol appears. This returns the Gage to a known, “out-of-the-box” condition.

Converts displayed and printed words to the selected language.Language

Page 16: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

12

Wet Bulb

When selected, the wet bulb temperature (Tw) is displayed on themain screen. It is calculated from Ta and RH using an air pressureof 1.0 atmospheres (1013 mbar). Variations of Tw at otherpressures could be greater than ±1°C (±2°F).

(Advanced models only)

Displays a real-time graph of the readings over the last threeminutes. Use Trend Charts to monitor short-term environmentalchanges and spot trends.

Press the Up button to scroll through the trend chartsfrom the main screen.

Trend Chart

Shortcut:

(Advanced models only)

Gage InfoDisplays the model number & serial number, probe type & serialnumber, PosiSoft.net registration key, the amount of remainingmemory for storage of readings, date and time, and softwarepackages. For security purposes, the registration key is requiredto add the Gage to your free PosiSoft.net account.

Anemometer (PosiTector DPM-A probes only)

When Enabled, thePosiTector DPM-A willdisplay wind velocity (V)in the selected unit ofmeasure (see Unitspg.13). Slide cover down to

expose hot wireanemometer

Hot Wire Anemometer

Surface Temperature

Air Temperature &Relative Humidity

NOTE: Dew point (Td) and surface temperature (Ts) are notdisplayed when the Anemometer is enabled.

Converts the displayed units for temperature and wind velocity.

Temperature (all models): Celsius (°C) or Fahrenheit (°F)Velocity (DPM-A only): m/s, ft/s, km/h, mph, and kn

Units

Page 17: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

13

Selects the type of batteries used in the Gage from a choice of“Alkaline”, “Lithium” or “NiMH” (Nickel-metal hydriderechargeable). If NiMH is selected, the Gage will trickle charge thebatteries while connected via USB to a PC or optional AC charger.The battery state indicator icon is calibrated for the selectedbattery type. No damage will occur if the wrong battery type isselected.

DeFelsko recommends the use of eneloop (NiMH)rechargeable batteries. It may take as long as 12 hours to chargedead batteries.

Battery Type

NOTE:

- Switching units will close any open memory batch.- Alarm values do not have units.

NOTE:

Selects display brightness (Sun, Normal or Night). All settings willdim slightly after a period of no activity to conserve battery life.Press the Down button to brighten the display.

Backlight (Advanced models only)

All batches are date-stamped when created, and all datasets aretime-stamped (24 hour format) when stored into memory. It istherefore important to keep both the date and time current usingthis menu option. Use the Up and Down buttons to scroll, and the(-) and (+) buttons to adjust a value. The Gage’s date and timecan also be viewed in Gage Info (pg. 12).

Set Clock

This option causes the display to rotate 180°. Ideal for right-side-up viewing when the sensors are pointed away from the user.

Flip Display

NOTE: DPM-A models automatically rotate when the instrumentis reorientated .

SoundToggle the Gage sounds ON (checked) or OFF (unchecked).

Page 18: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

14

Memory ManagementMemory Management

Scroll throughdisplay modes(Advanced only)

Delete last dataset

Access the Menu

Store a dataset inmemory

Restore brightnessafter dimming(Advanced only)

The PosiTector DPM has internal memory storage for recordingmeasurement datasets. Stored datasets can be reviewed on-screen or accessed via computers, tablets and smart phones.Datasets are date and time-stamped. Datasets can be storedmanually (using the (+) button), or automatically using the Auto Log mode (pg. 17).

Button functions with Memory ON:

Standard models store up to 2,500 datasets in one batch. Thefollowing appears within the Memory menu:

On: turns memory onOff: turns memory off (stored datasets remain in memory)Clear: removes all datasets from memoryView: lists group statistics and all stored datasets on thedisplay. It will begin by showing statistics based on all readingsin memory. Use the Up and Down buttons to scroll through alldatasets. Press to exit.

Advanced models store 100,000 datasets in up to 1,000batches (groups).

Sets the emissivity ratio for the PosiTector DPM-IR infraredtemperature sensor.

Select from a User Entered Value or one of the following presetoptions:

- Aluminum (rough) = 0.07 - Plastics = 0.94- Zinc (Galvanized) = 0.25 - Steel (Oxidized) = 0.95- Concrete = 0.54 - Wood = 0.95- Paint = 0.96

Emissivity (PosiTector DPM-IR Models Only)

Page 19: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

15

ViewScroll using the Up or Down buttons through information,statistical summary, and a list of each dataset in the currentlyopened batch. Press to exit.

σ - Standard Deviation↓ - Minimum Value

x - Average↑ - Maximum Value

New BatchCloses any currently opened batch andcreates a new batch name using the lowestavailable number. For example, if only Batch1 and Batch 3 exist, then Batch 2 would becreated and made the current batch. The

icon appears. New batch names are datestamped at the time they are created.

DeleteRemoves a batch completely from memory. Use the (+) and (-)buttons to scroll through pages of saved batches. The name isdeleted and all datasets are erased.

CloseStops the recording process, closes the current batch, andremoves batch information from the display.

OpenSelects a previously created batch name to open and makecurrent. Use the (+) and (-) buttons to scroll through pages ofsaved batches. When selected, the batch name and number ofstored datasets will be displayed.

The following appears within the Memory menu:

Page 20: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

16

• When a batch is open, press Up to scroll through the abovedisplay options.

Shortcuts:

Display options for the currently open batch are:Display

Td, Ts, Ta, Ts-Td, %RH, V: A chart of batch readingsImage: An image downloaded from PosiSoft.net duringsynchronization Notes: Instructions or notes downloaded from PosiSoft.netduring synchronization.None: default screen showing current dataset

Sends a statistical summary and individual datasets to theoptional Bluetooth Wireless Printer.

Print

To cancel printing, press and hold the (-) and (+) buttonssimultaneously NOTE:

AnnotateCreate meaningful batch names andenter notes directly on the Gage using afamiliar onscreen QWERTYkeyboard.

Use the Gages navigation and (-)(+)buttons to enter annotations.

Annotations can be synchronized withPosiSoft.net, the PosiTector App andare included in PosiSoft USB Drivereports.

(Advanced models only)

• Remove the last dataset from the current open batch bypressing (-).

Page 21: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

17

Screen CaptureScreen CapturePress both (-)(+) buttons at any time to capture and save an imagecopy of the current display. The last 10 screen captures are storedin memory and can be accessed when connected to a computer(see PosiSoft USB Drive pg. 19).

Summaries (Advanced models only)

Displays a summary of all storedbatches including the name, probetype, number of readings and type.

In the following example, Batch 3(B3) indicates an “F” 6000 coatingthickness probe was used to record atotal of “6” readings in “PA2” mode.

The PosiTector DPM can automatically record datasets intomemory at user selected time intervals. When selected, the AutoLog icon will appear on the display, with a countdown timer untilthe next dataset is stored. Readings on the display are onlyupdated when a dataset is stored.

Auto Log continues recording until the instrument loses power orthe user turns off Auto Log mode. Stored datasets will remain inmemory until full, then the oldest dataset or batch will beoverwritten to allow space for the next dataset.

Auto Log ModeAuto Log Mode

Turn on Gage/Cancel Auto Logmode

Scroll through displaymodes (Advancedmodels only)

It is recommended that fresh alkaline batteries beinstalled prior to an extended Auto Log if not connected tocontinuous USB power. Connect the Gage to a USB powersource for long-term use using the included USB cable.

NOTE:

Page 22: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

18

Advanced models powers down betweenreadings to conserve batteries. At eachinterval, the instrument will wake up, take areading, sync to PosiSoft.net (if configured)and power down. The instrument can recordapproximately 8,000 datasets while in thismode using alkaline batteries, and can beconfigured to Auto Batch at Daily, Weekly,and Monthly intervals.

Datasets can be sent directly to a user-specified FTP site. Thedata is sent in a comma-delimited text files (one file per reading),which can be parsed for embedding in a web page or imported intoa database. Ideal for use with custom developed software.

Select FTP Setup from the Auto Logscreen to setup ftp transfer.

Enable and enter the FTP Site address,username, and password. To test the con-nection select Test Upload.

If connection fails, check FTP Setup settingsand retry.

ü

Standard models remain powered-up while in Auto Log mode.The last recorded dataset will remain on the display until the nextdataset has been stored. The Gage will store datasets for 40-50hours on a set of fresh alkaline batteries depending on the intervalof the measurements.

Remote Monitoring of Environmental ConditionsThe PosiTector DPM is ideal for remotely monitoringenvironmental parameters over time on the jobsite, paint shop ormanufacturing floor.

See our online technical application note at:http://www.defelsko.com/applications/monitoring

Page 23: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

19

Accessing Stored Measurement DataAccessing Stored Measurement DataDeFelsko offers the following free solutions for viewing, analyzingand reporting data:

PosiSoft USB Drive - Connect theGage to a PC/Mac using the suppliedUSB cable. View and print readingsand graphs using universal PC/Macweb browsers or file explorers. Nosoftware or internet connection required. USB Drive must beselected in the Gage’s “Connect > USB” menu (pg. 24).

PosiSoft Desktop - Powerful desktop software (PC/Mac) fordownloading, viewing, printing and storing measurement data.Includes a customizable, templated PDF Report Generator. Nointernet connection required.

PosiSoft.net - Web-based application offering secure, centralizedstorage of measurement data. Access your data from any web-connected device.

PosiTector App - (Advanced models only, s/n 784000+) App forcompatible iOS and Android smart devices. Permits users tocreate, save and share professional PDF reports. Add images andnotes using the smart device’s camera and keyboard.

Learn more at http://www.defelsko.com/posisoft/posisoft.htm

Page 24: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

20

Turns WiFi functionality ON. When selected, the icon will appearon the display. To deactivate WiFi, uncheck the Enable box.

Enable

Connect MenuConnect Menu

ü

Allows connection to your local wireless network or mobile hotspot. Ideal for using your network’s internet connection forsynchronizing stored measurements with PosiSoft.net (pg.19).See www.defelsko.com/wifi

Connect your smart device/computer to a PosiTector Advancedbody wirelessly without the need for a separate network. Allowswireless access to the gages built-in PosiSoft USB drive wheneverWiFi network is not available

To enable, select Access Point from the Connect > WiFi menu.The Access Point icon will display in the upper left of thePosiTector display.

Securing your Access Point - To ensure the PosiTector is only accessible to authorized devices,it is important that you enter a passphrase (password) for theAccess Point. The default Passphrase is password.

In the Connect > WiFi > Setup menu, select AP Passphrase.

Press button to display on-screen keyboard. Enter aPassphrase for the Access Point. The Passphrase will be requiredfor all devices connecting to the PosiTectors Access Point.

WiFi (Advanced models only)

Access Point ü

Page 25: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

21

The PosiTector is now visible to all WiFi enabled devices. Simplyconnect your devices WiFi to the new PosiTector Access point. AllPosiTectors are uniquely identified by their respective gage bodyserial numbers.

AP Channel - Default Channel: 6For most users the default channel will not have to be altered. Ifyou are experiencing poor connection or are unable to connect, tryanother channel.

In the Connect > WiFi > Setup menu, select AP Channel.

Press the UP center navigation button to highlight the channel.Use the (-) or (+) buttons to change the channel. Press the DOWNnavigation button and select OK. Press the button.

Networks: With WiFi Enabled (pg. 20) the Gage will allow the userto add a new network and will automatically check for availablelocal networks. Available networks detected by the Gage are listedon the screen along with any networks that the Gage haspreviously been connected to that are not currently withinconnection range.

Information: Gage displays information about the local WiFinetwork connection including...• SSID: the network’s name• State: displays if the Gage is connected to the network or not• IP Address: the network’s IP AddressSetup: Allows user to setup a WiFi connection• IP Settings: enter the IP information as follows...

IP Type (DHCP or Static), IP Address, Gateway, Netmask, DNS1, DNS2

Page 26: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

22

üWhen USB Drive is checked , the PosiTector gage uses aUSB mass storage device class which provides users with a sim-ple interface to retrieve stored data in a manner similar to USBflash drives, digital cameras and digital audio players.

USB Drive is also required to import stored measurements intoPosiSoft Desktop software (pg.19). Once connected, any com-puter can view measurements stored in memory by navigating avirtual drive labeled "PosiTector” using the included USB cable.

A formatted HTML report is viewed by selecting the "index.html"or “START_HERE.html” file found in the root directory. Optionally,text ".txt" files located in each batch folder provide access tomeasurement values. Stored datasets and graphs can be viewedor copied using universal PC/Mac web browsers or file explorers.

Measurement data is stored in comma-separated text files whichcan easily be imported into supporting applications such asspreadsheets and databases.

Retrieving stored screen captures:The last 10 screen captures stored in memory can be accessedby navigating to the “screen capture” directory within the“PosiTector” virtual drive.

USB

Information (continued):• Server Enable: enables a connection between the network andthe Gage• Gage Name: enter a name for the Gage (up to 14 Characters).AP Channel: The access point channel corresponding to afrequency range. (Default Channel: 6)AP Passphrase: A series of characters, numbers or symbols usedto log on to a WiFi network. (Default Passphrase: password)• WiFi Reset: erases all WiFi settings

Page 27: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

23

• When your PosiTector is first connected to your Windows PC viaa USB cable, an enumeration process is started that installs devicedrivers without re-booting your computer. You may see severalpop-up windows in the taskbar at the bottom right of your screen.Wait for the entire process to be completed before proceeding.

• When connected, power is supplied through the USB cable. Thebatteries are not used and the body will not automatically powerdown. If rechargeable (NiMH) batteries are installed, theinstrument will trickle charge the batteries.

NOTES:

Serial Streaming via USB (Advanced models only, serial numbers 784000 and greater)

Advanced gage bodies have the ability to live stream datasets toa USB connected computer. Using 3rd party data collection orSPC software, datasets can be captured as they are taken.

This following guide provides step-by-step directions for estab-lishing communication between the PosiTector Advanced and aUSB connected computer.

PosiTector Advanced USB Serial Streaming Instructionshttp://www.defelsko.com/usb/PosiTectorAdvanced-USBserial.pdf

Page 28: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

24

When Enabled , allows communication with a smart devicerunning the PosiTector App (pg. 19) via auto-pairing BluetoothSmart (BLE) wireless technology.

Select fbatches to flag them for synchronization to thePosiTector App. New batches created while Bluetooth Smart isenabled are automatically selected.

Bluetooth Smart (Advanced models only, serial numbers 784000 and greater)

Sync .net NowThe below WiFi, USB and Bluetooth menus contain a Sync .netNow option. When selected, the Gage immediately synchronizesstored measurement data via its respective communicationmethod (internet connection required).

Alternatively, select Auto Sync .net from within the USB connectmenu to automatically synchronize upon connection to a PC.Additional measurements added to memory while connected aresynchronized only when the USB cable is disconnected andreconnected, or when the Sync .net Now option is selected.

WiFi connected gages automatically attempt synchronizationupon power-up.

PosiSoft Desktop (pg. 19) is required when using aUSB connection to synchronize measurements with PosiSoft.net.

ü

ü

Sync BatchesWith Bluetooth Smart enabled, select Sync Batches to transferselected batches to the PosiTector App. This is useful whenswitching between smart devices, as only readings and batchesthat have yet to be synchronized with any smart device are syn-chronized automatically.

If Bluetooth Smart is disabled, data from batchesselected in the Sync Batches menu are held in a queue untilcommunication with the PosiTector App is re-established.

NOTE:

NOTE:

Page 29: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

25

Transfers selected batches to the PosiTector App (useful whenswitching between devices).

The Send Batches option is visible in the menu when the Gage isconnected to a smart device running the PosiTector App.

Send Batches

ü

Bluetooth (Advanced models only

Allows individual readings to be sent to a computer, printer orcompatible device as they are taken using Bluetooth wirelesstechnology. See www.defelsko.com/bluetooth

PairingThe instrument and receiving device must be paired beforestored or streamed datasets can be transmitted. For pairinginstructions, see http://www.defelsko.com/bluetooth

InfoLists information about your current Bluetooth connection, includ-ing the currently paired device and MAC address.

StreamWhen checked, the instrument will stream datasets to the pairedBluetooth Device as they are taken. Datasets can be streamedas they are taken to the optional Bluetooth printer or third-partycomputer software.

Powder ProbesDisplays menu options that enable the PosiTector body to com-municate with wireless PosiTector PC powder probes. See http://www.defelsko.com/pc

Page 30: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

26

The battery indicator displays a full bar with fresh alkalineor fully charged batteries installed. As the batteries weaken, thebar will be reduced. When the battery symbol is lowthe Gage can still be used, but the batteries should be changed orrecharged at the earliest opportunity. The Gage will turn offautomatically when batteries are very low, preceded by a LowBattery Warning on the display.

To retain all user settings and stored memory readings, onlyreplace the batteries after the Gage has powered-down.

Power Supply / Battery IndicatorPower Supply / Battery IndicatorPower Source: 3 AAA alkaline, Lithium or optional Nickel-metalhydride (NiMH) rechargeable batteries. For best battery indicatorresults, ensure the appropriate Battery Type is selected in theSetup > Battery Type menu (pg. 13).

Battery performance decreases at low temperatures.

!

Determines if a software update is available for your Gage. If anupdate is available, a prompt will appear allowing the user tochoose to perform the update at this time or not.

To perform an update the Gage must be connected to an internetconnected computer with PosiSoft Desktop (pg. 19), or WiFinetwork (pg. 20).

Updates

Ensure that stored measurements are backed up to a PCor PosiSoft.net. The Gage will perform a Hard Reset (pg. 11) aftercompletion of the update and ALL readings in memory will beerased.

NOTE:

2 / 3

DO NOT unplug the Gage during the update operation.

Currently installingupdate 2 of 3 total

Progress bar

Updates - Sample Screen

!

Page 31: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

27

TroubleshootingTroubleshootingMost conditions can be cleared with a Hard Reset (pg. 11). If not, visit our Service & Support page for information on sendingyour Gage to us for a no charge evaluation. See: http://www.defelsko.com/service

When the surface temperature probe is not on a surface, AirTemperature and Surface Temperature appear to be differentUnder normal conditions with the surface probe in air Ta and Tsshould be within the combined tolerance of each sensor (typically1°C each). It is possible for the difference to be larger if thesurface probe has recently been in contact with surfaces that areat a different temperature than air.

RH value is slow to stabilizeIf the probe is exposed to very low humidity for an extended timeperiod and then placed in a high humidity environment, thedisplayed Relative Humidity (RH) may appear to respond slowly.Allow the probe to stabilize 30 minutes for a change of RH greaterthan 50%. It may take longer if the probe was left at low RH forseveral months. To increase response time wrap the sensor in adamp cloth overnight to recondition the RH sensor.

Surface Temperature reads much higher than expectedThe probe tip may be dirty or damaged. Use care if attempting toclean dirt or overspray from the sensor. If Ts continues to readhigh then return the probe for service.

Surface temperature value does not agree with expectedsurface temperature or surface temperature probe is slow torespondIn some cases the surface being measured is very rough and agood thermal connection from the probe tip to the surface is notpossible. If possible place a small drop of water on the surfaceand then place the probe tip into the water drop. Quick andaccurate results can be obtained in this manner.

Page 32: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

28

This device complies with part 15 of the FCC Rules. Operation is subject to the followingtwo conditions: (1) This device may not cause harmful interference, and (2) this devicemust accept any interference received, including interference that may cause undesiredoperation.

Technical DataTechnical DataConforms to: ISO 8502-4, BS7079-B and ASTM D3276

Range Accuracy Resolution

SurfaceTemp.

-40° to 80° C80° to 190° C-40° to 175° F175° to 375° F

±0.5° C±1.5° C±1° F±3° F

0.1° C0.1° C0.1° F0.1° F

IR SurfaceTemp.**

-70° to 380° C-94° to 716° F

±1° C +(0.01°C/°C)***

±2° F +(0.01°F/°F)***

0.1° C0.1° F

AirTemp.

-40° to 80° C-40° to 175° F

±0.5° C±1° F

0.1° C0.1° F

Humidity 0 to 100% ±3 % 0.1 %

Wind Speed*

0 - 20 m/s0 - 65 ft/s

0 - 3900 fpm,0 - 72 km/h0 - 44 mph0 - 38 kn

±3 % + least

significantdigit

0.1 m/s0.1 ft/s1 fpm,

0.1 km/h0.1 mph0.1 kn

Operating Range: -40° C to +80° C -40° F to +175° F

PosiTector DPM, DPM-S, DPM-A, DPM-D, DPM-IR

**DPM-A Models Only**DPM-IR Models Only***At Ta = 23° C

PosiTector DPM-IR:- Distance to Target Size Ratio (D:S) - 5.7:1- Response Time: <500 milliseconds (95% resposne)- Spectral Response: 8 - 14 um

Page 33: Positector DPM Con Sensor IR Estandar T Aire DF-DPMIR1E · For industrial painters, the primary reason for measuring climatic conditions is to avoid rework and the premature failure

29

Limited Warranty, Sole RemedyLimited Warranty, Sole Remedyand Limited Liabilityand Limited Liability

DeFelsko's sole warranty, remedy, and liability are the expresslimited warranty, remedy, and limited liability that are set forth onits website: www.defelsko.com/terms

www.defelsko.com© DeFelsko Corporation USA 2017

All Rights Reserved

This manual is copyrighted with all rights reserved and may not be reproduced or transmitted, in wholeor part, by any means, without written permission from DeFelsko Corporation.

DeFelsko, PosiTector, and PosiSoft are trademarks of DeFelsko Corporation registered in the U.S. andin other countries. Other brand or product names are trademarks or registered trademarks of theirrespective holders.

Every effort has been made to ensure that the information in this manual is accurate. DeFelsko is notresponsible for printing or clerical errors.

Returning for ServiceReturning for ServiceBefore returning the instrument for service…

1. Install new batteries in the proper alignment as shown within batterycompartment.

2. Examine the surface temperature sensor for dirt or damage.

3. If the humidity sensor is slow to respond or reading erratically, reconditionthe sensor by removing the probe from the gage body and wrapping it ina damp cloth overnight.

4. Perform a Hard Reset. (see pg. 11)

5. If issue is not resolved, Update (pg. 26) your PosiTector gage body andre-attempt measurements.

If you must return the Gage for service, please fill out and include theService Form located at www.defelsko.com/support with the Gage. Besure to also include the probe, your company name, company contact,telephone number and fax number or email address.