Precision in Quantitative Imaging: Trial Development and Quality Assurance Susanna I Lee MD, PhD Thanks to: Mitchell Schnall, Mark Rosen. Dan Sullivan, Patrick Bossuyt

Upload: dangphuc

Post on 29-Aug-2019

218 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Precision in Quantitative Imaging:

Trial Development and

Quality Assurance

Susanna I Lee MD, PhD

Thanks to:

Mitchell Schnall, Mark Rosen. Dan Sullivan, Patrick Bossuyt

Page 2: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Imaging Chain: Patient Data

Raw data

Image

analysis123……………

2346…………..

65789…………

6578…………..

Data output

Image

reconstruction

Image

processing

Data analysis

Page 3: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Clinical Trials: Imaging is an “Assay”

♦ Disease Detection

• Screening

♦ Characterize Disease

• Diagnosis, eligibility or prognosis

• Anatomic distribution (e.g. tumor staging)

• Higher level features (e.g. heterogeneity, vascularity, etc.)

♦ Monitor Therapeutic Response

• Change in features with therapy

• Anatomic (e.g. RECIST) or functional (e.g. SUV)

Page 4: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

REFERENCE STANDARD OF

PATIENT OUTCOME

• OVERALL SURVIVAL (OS)

• PROGRESSION FREE SURVIVAL

(PFS) OR DISEASE FREE

SURVIVAL (DFS)

• TREATMENT RESPONSE

REFERENCE STANDARD OF

DISEASE STATE

• PATHOLOGY

• CONFIRMATORY TEST

• FOLLOWUP

MANAGEMENT

• STANDARD OF CARE

• STUDY DEFINED

INDEX TEST

• IMAGING EXAM

• IMAGE-GUIDED

PROCEDURE

Schema

ENROLLED

PARTICIPANT

Diagnostic accuracy

Biomarker

Page 5: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Cancer Biomarkers

GeneticGenome

Expression (e.g.

RNA or protein)

Serum

(CA125,

PSA, AFP,

CA19-9)

Tissue(e.g. hormone

receptors,

cytokeratins)

PATIENT OUTCOME

1. Overall survival (OS)

2. Progression free

survival (PFS)

3. Clinical response

Imaging

Page 6: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

What is a good biomarker?

♦Stable technology

♦Available widely

♦Standardized image acquisition

♦Reproducible

♦Range of normal defined

Sargent DJ, Rubinstein L, Schwartz L et al. Eur J Cancer 2009; 45: 290.

Balance “state of the art” with

“generalizability”

Page 7: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Pre-treatment

Post-treatment

Range of

test-retest

Lankester KJ, Taylor NJ, Stirling JJ et al. Br J Cancer. 2005.93:979.

Variability: Test – Retest

Same patient, day, scanning protocol but separate imaging sessions

Conclusion

Index test variability precludes detecting pre- vs. post-treatment change.

Page 8: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Signal Requires Data Quality

Page 9: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Precision vs. Bias

precise

accurate

not precise

not accurate

precise

not accurate

Page 10: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Multiple Sclerosis MRI

♦ Image acquisition

• T1

• T2

• Post-gadolinium T1

♦ Image analysis

• Number of new or enlarging lesions

• Number of enhancing lesions

Page 11: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

• MRI endpoint in MS treatment studies

• 157 publications from 1995 to 2006

Page 12: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Imaging Chain: Patient Data

Raw data

Image

analysis123……………

2346…………..

65789…………

6578…………..

Data output

Image

reconstruction

Image

processing

Data analysis

Page 13: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Imaging Manual

♦ Hardware and software

♦ Scanner calibration

♦ Patient preparation

♦ Scanning protocol

♦ Post-processing

Image acquisition manual with a step by step description is part

of any prospective study design.

Page 14: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

What determines resolution?

♦Physics of acquisition (i.e. modality)

♦Sampling (e.g. matrix, detector size)

♦Filtering and other contributions inherent in

the reconstruction

Page 15: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Point Spread Function

Patient Image

Page 16: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Image ““““edges”””” approximate anatomy

Structure

Real Edge

Image Edge

Page 17: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Resolution and Sampling

160 x 160 matrix 320 x 224 matrix

Page 18: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Filtering

4 2

222

2

2 2 2

1

1

1

1

1111

1

1

1

1 1 1 1 1

Page 19: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Filtering: MRI

Vendor 1 Vendor 2

Page 20: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Filtering: X-ray

Vendor 1 Vendor 2

Page 21: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Partial Volume Effects

Completely

in scan planePartially in

scan plane

Page 22: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Partial Volume Effects: CT

Completely in scan plane Partially in scan plane

HU = 0 HU = 30-60

Page 23: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Partial Volume Effects: PET

lesion

lesion

Blurred margins

Lower intensity

Page 24: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield
Page 25: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Quantitative Imaging

Biomarkers Alliance (QIBA)

♦ Started by RSNA 2007

♦ Mission: Improve the value and practicality of

quantitative imaging biomarkers by reducing

variability across devices, patients and time

♦ “Build imaging devices that are also measuring

devices”

♦ “Industrialize imaging biomarkers”

https://www.rsna.org/QIBA/

Page 26: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

QIBA Approach

1. Identify the sources of error and variability

2. Specify potential solutions in the form of profiles

3. Test these solutions

4. Promulgate profiles to vendors and users

• Advise vendors what must be implemented in their product

• Communicate the necessary procedures to users

Purpose of profiles:

Page 27: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

QIBA Profile Activity Diagram

Equipment

Assessment

Subject

Preparation

Image

Reconstruction

Image

AnalysisInterpretation

Image

Acquisition

Manufacturer specification (pre-delivery)

Installation specification

Maintenance Quality

Assurance

Page 28: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

ACR Core Lab QA

♦Site qualification

• Instrument performance

• Training

♦Monitoring of image acquisition

• Scan header for protocol compliance

• On-line technologist qualitative review

• Periodic radiologist review

♦Centralized image analysis

• Post-processing

• Reader study

Page 29: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield
Page 30: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Require Site Protocol Compliance

Type T1 weighted GRE

Orientation Sagittal

Pulse Sequence Dynamic 3D

Field Of View (FOV) 16-18 cm

Slice Thickness 64 slices of thickness ≤ 2.5

mm

Skip

Matrix min. 256 x 192

Frequency A/P

NEX 2

Phase Wrap NO

Fat-Saturation YES

TR ≤ 20 ms

Effective TE 4.5 ms

Scan Duration Between 4.5 and 5 minutes

Flip angle <= 45 degrees

Correct

Submitted

ACRIN 6657 (I-SPY 1 trial), Nola Hylton, PI

Page 31: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Imaging Chain: Patient Data

Raw data

Image

analysis123……………

2346…………..

65789…………

6578…………..

Data output

Image

reconstruction

Image

processing

Data analysis

Page 32: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Image analysis: Turning image into data

♦User extracted features

♦Semi automated

♦Automated

Feature 1

Feature 2

Feature 3

.

.

.

Page 33: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Radiomics: Deep Learning

Untrained neural network

Page 34: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Radiomics: Deep Learning

Trained neural network

Page 35: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Radiomics: Automated Image Analysis

Improve diagnostic accuracy

Page 36: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Radiomics: Automated Image Analysis

Triage clinical workflow

Page 37: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Semi-automated: Manual Segmentation

Quantitative

feature

extraction

Large image

datasets

segmented for

tumor

Integrate with

genomic &

clinical data

for machine

learning

Model

predictive

indices

Aerts HJ et al. Nat Commun. 2014 ;5:4006.

Page 38: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Reader Extracted Features

•Density• Fluid, soft tissue, calcified

•Shape• Round, oval, irregular

•Size• Linear, volume

•Margin• Sharp, blurred, spiculated

•Intensityhigh/medium/low/minimal

•Summary assementBIRADS level

Page 39: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

0.0 0.2 0.4 0.6 0.8 1.0

FP

0.0

0.2

0.4

0.6

0.8

1.0

TP

Beam, Layde, Sullivan Arch Intern Med 1996; 156:209-213

ROC operating points of 108 radiologists

reading same mammograms

Skill

Value

judgments

Page 40: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

• Increases and decreases of <10% can be a result of

inherent variability.

Reader Variability: Size

Oxnard GR et al. J Clin Oncol. 2011;29:3114-9.

progression

Page 41: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

SUV=4.0 SUV=5.6

Variability Introduced by ROI Selection

Slice 483 Slice 479 Slice 479

SUV=6.6

Page 42: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Variability Introduced by ROI Selection

• All ROI protocols show excellent inter-observer agreement (ICC 0.94)

• Different ROI protocols yield different ADC values

Priola AM et al. Eur Radiol. 2016 Aug 11.

Page 43: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Effect of Windowing

Soft tissue window Liver window

Page 44: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Effect of Windowing

• Significant measurement differences between window settings (p<0.001).

• No significant differences in measurement variability between the lung and

mediastinal window settings (p>0.05).

Lung window Mediastinal window

Kim H et al. PLoS One. 2016;11:e0148853.

Page 45: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Reader Study

♦ Reader blinded to reference standard

♦ Multiple readers

♦ Independent rather than consensus reads

♦ Rules for image interpretation

• Clinical information available to reader

• Image selection, windowing, order, etc.

• Choosing index lesions

• Selecting region of interest (ROI)

• Definition of positive vs. negative test

♦ Washout period between read sessions of paired imaging exams

♦ Digital data forms and screen to document reader study

A manual defining reader rules and training cases are part of any

prospective study design.

Page 46: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

♦Overall trial framework

♦Hypothesis and specific aims

♦ Participants

♦ Index test

♦ Reference standard

♦ Data analysis plan (statistics)

♦ Conclusions and implications

♦ Funding and compliancehttp://www.equator-network.org/reporting-guidelines/stard/

Page 47: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Index Test (Imaging Exam)

♦ STARD 10 - Index test, in sufficient detail to allow replication

♦ STARD 12 - Definition of and rationale for test positivity cut-offs or

result categories of the index test, distinguishing pre-specified from

exploratory

♦ STARD 13 - Whether clinical information and reference standard

results were available to the performers/readers of the index test

♦ STARD 25 - Any adverse events from performing the index test

Page 48: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield

Steps toward precision:

♦ Define image acquisition

• Equipment, patient preparation, protocol

• Balance “state of the art” with “generalizability”

♦ Define image analysis

• Read rules, training and testing

♦ Validate the system

• Test-retest, reader agreement measurements

♦ Build in procedures for ongoing QA

Page 49: Precision in Quantitative Imaging: Trial Development and ... PDFs... · • All ROI protocols show excellent inter-observer agreement (ICC 0.94) • Different ROI protocols yield