1 Product certifications
1.1 Safety messages
NoteOther language versions of this Product Certifications Document can befound at Emerson.com/Rosemount.
WARNING
Failure to follow safe installation and servicing guidelines could result indeath or serious injury.
• This document contains information required for compliance with thecertifications.
• Use the level detector only as specified in this document and the productreference manual. Failure to do so may impair the protection providedby the level detector. Refer to the Rosemount 2140 and 2140:SIS LevelDetectors Reference Manual for more instructions.
• The weight of a level detector with a heavy flange and extended forklength may exceed 37 lb. (18 kg). A risk assessment is required beforecarrying, lifting, and installing the level detector.
Explosions could result in death or serious injury.
• Verify that the operating atmosphere of the level detector is consistentwith the appropriate hazardous locations certifications.
• Before connecting a handheld communicator in an explosiveatmosphere, ensure that the instruments in the loop are installed inaccordance with intrinsically safe or non-incendive field wiring practices.
• In explosion-proof/flameproof and non-incendive/type n installations, donot remove the housing covers when power is applied to the leveldetector.
• Both housing covers must be fully engaged to meet flameproof/explosion-proof requirements.
Product Certifications September 2020
2 Rosemount 2140 and 2160:SIS Level Detectors
WARNING
Electrical shock could cause death or serious injury.
• Avoid contact with the leads and terminals. High voltage that may bepresent on leads can cause electrical shock.
• Ensure the power to the level detector is off, and the lines to any otherexternal power source are disconnected or not powered while wiring thelevel detector.
• Ensure the wiring is suitable for the electrical current and the insulation issuitable for the voltage, temperature, and environment.
Process leaks could result in death or serious injury.
• Ensure the level detector is handled carefully. If the process seal isdamaged, gas might escape from the vessel (tank) or pipe.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/ormisconfiguration of end users’ equipment. This could be intentional orunintentional and needs to be protected against.
Physical security is an important part of any security program andfundamental to protecting your system. Restrict physical access byunauthorized personnel to protect end users’ assets. This is true for allsystems used within the facility.
CAUTION
Hot surfacesThe flange and process seal may be hot at high processtemperatures.
Allow to cool before servicing.
CAUTION
Refer to manualThis document contains information required to safely install,connect, commission, operate and maintain, and is required forcertification compliance.
September 2020 Product Certifications
Product Certifications 3
1.2 European Union directive informationA copy of the EU Declaration of Conformity certificate can be found in thesection EU Declaration of Conformity. The most recent revision of thecertificate can be found at Emerson.com/Rosemount.
1.3 Ordinary location certifications1.3.1 U.S.A. Ordinary Location certification
G5
Summary of product certification:
Protection Ordinary location (unclassified, safe area)
Certificate 70098390
Standard UL 61010-1: 2012
Marking Type 4X
The level detector has been examined and tested to determine that thedesign meets basic electrical, mechanical, and fire protection requirementsby CSA, a nationally recognized testing laboratory (NRTL) as accredited bythe Federal Occupational Safety and Health Administration (OSHA).
Special Condition For Use
To be supplied from a Class 2 or Limited Energy Source in accordance withCAN/CSA C22.2 No. 61010-1-12.
1.3.2 Canada Ordinary Location certification
G6
Summary of product certification:
Protection Ordinary location (unclassified, safe area)
Certificate 70098390
Standards CAN/CSA C22.2 No 61010-1-12
ANSI/ISA-12.27.01:2011
Marking Type 4X
Single process seal
The level detector has been examined and tested to determine that thedesign meets basic electrical, mechanical, and fire protection requirementsby CSA, a nationally recognized testing laboratory (NRTL) as accredited bythe Federal Occupational Safety and Health Administration (OSHA).
Product Certifications September 2020
4 Rosemount 2140 and 2160:SIS Level Detectors
Special Condition For Use
To be supplied from a Class 2 or Limited Energy Source in accordance withCAN/CSA C22.2 No. 61010-1-12.
1.4 USA1.4.1 U.S.A. Explosion-proof and Division 2 (XP)
E5
Summary of product certification
Protection Explosion-proof (XP)
Certificate 70098390
Standards FM Class 3600 - 2011
M Class 3615 - 2015
UL 61010-1: 2012
Markings Class I Groups B, C and D, T6...T2
Class I Division 2 Groups A, B, C and D, Type 4X
Class I, Zone 1, AEx db IIC T6…T2 Gb
Temperatures See Flameproof and explosion-proof approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Use (X)
1. The user is to ensure the probe assembly is installed in such a way toprevent any damage due to impact or ignition source due to friction.
2. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
3. Flameproof joints are not intended for repair.
4. To be supplied from a Class 2 or Limited Energy Source in accordancewith CAN/CSA C22.2 No. 61010-1-12.
September 2020 Product Certifications
Product Certifications 5
1.4.2 U.S.A. Intrinsic Safety (IS) and Non-Incendive (NI)
I5
Summary of product certification
Protection Intrinsic Safety (IS) and Non-Incendive (NI)
Certificate 70098390
Standards FM Class 3600:2011
FM Class 3610:2015
FM Class 3611:2004
Markings Class I Groups A, B, C and D, T5…T2
Class I, Division 2, Groups A, B, C, and D, Type 4X
Class I, Zone 0, AEx ia IIC T5…T2 Ga
Controldrawing
71097/1387 (see Installation drawings)
Temperatures See Intrinsically Safe (IS) approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The equipment, when fitted with the transient suppression terminalblock, is not capable of withstanding the 500 volt insulation test. Thismust be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given a protectivepolyurethane paint finish. However, care should be taken to protectit from impact or abrasion if located in Zone 0.
3. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
Product Certifications September 2020
6 Rosemount 2140 and 2160:SIS Level Detectors
1.5 Canada1.5.1 Canada Explosion-proof and Division 2 (XP)
E6
Summary of product certification
Protection Explosion-proof (XP)
Certificate 70098390
Standards ANSI/ISA 12.27.01:2011
CSA Std. C22.2 No. 30 -M1986
CSA Std. C22.2 No.60079-0-15
CSA Std. C22.2. No.60079-1-16
CSA Std. C22.2 No. 61010-1-12
CSA Std. C22.2 No.94-M91
CSA Std. C22.2 No. 213-2016
Markings Class I Groups B, C and D, T6...T2
Class I Division 2 Groups A, B, C and D, Type 4X
Ex db IIC T6…T2 Gb, single seal
Temperatures See Flameproof and explosion-proof approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Use (X)
1. The user is to ensure the probe assembly is installed in such a way toprevent any damage due to impact or ignition source due to friction.
2. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
3. Flameproof joints are not intended for repair.
4. To be supplied from a Class 2 or Limited Energy Source in accordancewith CAN/CSA C22.2 No. 61010-1-12.
September 2020 Product Certifications
Product Certifications 7
1.5.2 Canada Intrinsic Safety (IS) and Non-Incendive (NI)
I6
Summary of product certification
Protection Intrinsic Safety (IS) and Non-Incendive (NI)
Certificate 70098390
Standards ANSI/ISA 12.27.01:2011
CSA Std. C22.2 No. 157 -92
CSA Std. C22.2 No. 60079-0-15
CSA Std. C22.2. No. 60079-11-14
Markings Class I Groups A, B, C and D, T5…T2
Class I, Division 2, Groups A, B, C, and D, Type 4X
Ex ia IIC T5…T2 Ga, single seal
Controldrawing
71097/1387 (see Installation drawings)
Temperatures See Intrinsically Safe (IS) approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The equipment, when fitted with the transient suppression terminalblock, is not capable of withstanding the 500 volt insulation test. Thismust be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given a protectivepolyurethane paint finish. However, care should be taken to protectit from impact or abrasion if located in Zone 0.
3. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
1.5.3 Canadian Registration Number
CRN 0F04227.2C
The requirements of CRN are met when a Rosemount 2140 Level Detector isconfigured with 316/316L stainless steel (1.4401/1.4404) process-wettedparts and either NPT threaded or 2-in. to 8-in. ASME B16.5 flanged processconnections.
Product Certifications September 2020
8 Rosemount 2140 and 2160:SIS Level Detectors
1.6 Europe1.6.1 ATEX Flameproof (FLP) and Dust Proof (DP)
E1
Summary of product certification
Protection Flameproof (FLP)
Certificate Dekra 16ATEX0082X
Standards EN60079-0:2012+A11:2013
EN60079-1:2014
EN60079-26:2015
Markings II 1/2 G
Ex db IIC T6...T2 Ga/Gb
Temperatures See Flameproof and explosion-proof approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The user is to ensure the probe assembly is installed in such a way toprevent any damage due to impact or ignition source due to friction.
2. Non-standard paint options may cause risk from electrostaticdischarge. Avoid installations that could cause electrostatic build-upof painted surfaces, and only clean the painted surfaces with a dampcloth. If paint is ordered through a special option code, contact themanufacturer for more information.
3. Flameproof joints are not intended for repair.
ND
Summary of product certification
Protection Dust Proof (DP)
Certificate Baseefa 16ATEX0137X
Standards IEC EN 60079-0:2018
EN60079-31:2014
Markings II 1 D, Ex ta IIIC (T92 °C…T272°C)
(T200100°C…T200280 °C) Da
Temperatures See Dust approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
September 2020 Product Certifications
Product Certifications 9
Special Conditions For Safe Use (X)
1. Cable entries must maintain the ingress protection rating of theenclosure to at least IP66.
2. Unused cable entries must be filled with suitable blanking plugswhich maintain the ingress protection of the enclosure to at leastIP66.
3. Cable entries and blanking plugs must be suitable for the ambienttemperature of the equipment and be capable of withstanding a 7Jimpact test.
4. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
E8 combines E1 and ND
Temperatures See Flameproof and explosion-proof approval temperatures andDust approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The user is to ensure the probe assembly is installed in such a way toprevent any damage due to impact or ignition source due to friction.
2. Non-standard paint options may cause risk from electrostaticdischarge. Avoid installations that could cause electrostatic build-upof painted surfaces, and only clean the painted surfaces with a dampcloth. If paint is ordered through a special option code, contact themanufacturer for more information.
3. Flameproof joints are not intended for repair.
4. Cable entries must maintain the ingress protection rating of theenclosure to at least IP66.
5. Unused cable entries must be filled with suitable blanking plugswhich maintain the ingress protection of the enclosure to at leastIP66.
6. Cable entries and blanking plugs must be suitable for the ambienttemperature of the equipment and be capable of withstanding a 7Jimpact test.
7. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be taken
Product Certifications September 2020
10 Rosemount 2140 and 2160:SIS Level Detectors
to protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
1.6.2 ATEX Intrinsic Safety (IS)
I1
Summary of product certification
Protection Intrinsic Safety (IS) and Dust (Zone 0, 20)
Certificates Baseefa 16ATEX0136X and Baseefa 16ATEX0137X
Standards IEC EN 60079-0:2018
EN60079-11:2012
EN60079-26:2015
EN60079-31:2014
Markings II 1 G, Ex ia IIC T5...T2 Ga
II 1 D, Ex ta IIIC (T92 °C...T272 °C)
(T200100°C...T200280 °C) Da
Electricalparameters
Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 0.012 μF, and Li = 0 mH
Temperatures See Intrinsically Safe (IS) approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The equipment, when fitted with the transient suppression terminalblock, is not capable of withstanding the 500 volt insulation test. Thismust be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given a protectivepolyurethane paint finish. However, care should be taken to protectit from impact or abrasion if located in Zone 0.
3. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
4. Cable entries must maintain the ingress protection rating of theenclosure to at least IP66.
5. Unused cable entries must be filled with suitable blanking plugswhich maintain the ingress protection of the enclosure to at leastIP66.
September 2020 Product Certifications
Product Certifications 11
6. Cable entries and blanking plugs must be suitable for the ambienttemperature of the equipment and be capable of withstanding a 7Jimpact test.
I8
Summary of product certification
Protection Intrinsic Safety (IS) (Zone 1)
Certificate Baseefa 16ATEX0136X
Standards IEC EN 60079-0:2018
EN60079-11:2012
EN60079-26:2015
Markings II 1/2 G, Ex ib IIC T5...T2 Ga/Gb
Electricalparameters
Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 0.012 μF, and Li = 0 mH
Temperatures See Intrinsically Safe (IS) approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The equipment, when fitted with the transient suppression terminalblock, is not capable of withstanding the 500 volt insulation test. Thismust be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given a protectivepolyurethane paint finish. However, care should be taken to protectit from impact or abrasion if located in Zone 0.
3. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
Product Certifications September 2020
12 Rosemount 2140 and 2160:SIS Level Detectors
1.7 International1.7.1 IECEx Flameproof (FLP) and Dust
E7
(E7 also includes the NK approval)
Summary of product certification
Protection Flameproof and dust-proof
Certificates IECEx DEK 16.0040X and IECEx BAS 16.0106X
Standards IEC60079-0:2017
IEC60079-1:2014
IEC60079-26:2014
IEC60079-31:2013
Markings Ex db IIC T6...T2 Ga/Gb
Ex ta IIIC (T92°C…T272°C) (T200100°C…T200280°C) Da
Temperatures See Flameproof and explosion-proof approval temperatures
See Dust approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The user is to ensure the probe assembly is installed in such a way toprevent any damage due to impact or ignition source due to friction.
2. Non-standard paint options may cause risk from electrostaticdischarge. Avoid installations that could cause electrostatic build-upof painted surfaces, and only clean the painted surfaces with a dampcloth. If paint is ordered through a special option code, contact themanufacturer for more information.
3. Flameproof joints are not intended for repair.
September 2020 Product Certifications
Product Certifications 13
1.7.2 IECEx Intrinsic Safety (IS)
I7
Summary of product certification
Protection Intrinsic Safety (IS)
Certificates IECEx BAS 16.0105X
Standards IEC60079-0:2017
IEC60079-11:2011
Markings Ex ia IIC T5...T2 Ga
Electricalparameters
Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 0.012 μF, and Li = 0 mH
Temperatures See Intrinsically Safe (IS) approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The equipment, when fitted with the transient suppression terminalblock, is not capable of withstanding the 500 volt insulation test. Thismust be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given a protectivepolyurethane paint finish. However, care should be taken to protectit from impact or abrasion if located in Zone 0.
3. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
1.7.3 IECEx Dust
NK
Summary of product certification
Protection Dust
Certificates IECEx BAS 16.0106X
Standards IEC60079-0:2017
IEC60079-31:2013
Markings Ex ta IIIC (T92°C…T272°C) (T200100°C…T200280°C) Da
Temperatures See Dust approval temperatures
Product Certifications September 2020
14 Rosemount 2140 and 2160:SIS Level Detectors
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. Cable entries must maintain the ingress protection rating of theenclosure to at least IP66.
2. Unused cable entries must be filled with suitable blanking plugswhich maintain the ingress protection of the enclosure to at leastIP66.
3. Cable entries and blanking plugs must be suitable for the ambienttemperature of the equipment and be capable of withstanding a 7Jimpact test.
4. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
1.8 Brazil1.8.1 INMETRO Flameproof (FLP)
E2
Summary of product certification
Protection Flameproof
Certificates UL-BR 017.0843X
Standards ANBT NBR IEC 60079-0:2013
ABNT NBR IEC 60079-1:2016
ABNT NBR IEC 60079-26:2016
Markings Ex db IIC T6…T2 Ga/Gb
Temperatures See Flameproof and explosion-proof approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Use (X)
1. Flameproof joints are not intended for repair.
2. The sensor probe shall be handled and installed to avoid friction anddamage from impact.
September 2020 Product Certifications
Product Certifications 15
3. Non-standard paint options may cause risk from electrostaticdischarge. Avoid installations that could cause electrostatic build-upof painted surfaces, and only clean the painted surfaces with a dampcloth. If paint is ordered through a special option code, contact themanufacturer for more information.
1.8.2 INMETRO Intrinsic Safety (IS)
I2
Summary of product certification
Protection Intrinsic Safety (IS)
Certificates UL-BR 17.0837X
Standards ANBT NBR IEC 60079-0:2013
ABNT NBR IEC 60079-11:2013
Markings Ex ia IIC T5…T2 Ga
Electricalparameters
Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 0.012 μF, and Li = 0 mH
Temperatures See Intrinsically Safe (IS) approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The equipment, when fitted with the transient suppression terminalblock, is not capable of withstanding the 500 volt insulation test. Thismust be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given a protectivepolyurethane paint finish. However, care should be taken to protectit from impact or abrasion if located in areas that require EPL Ga(Zone 0).
3. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
Product Certifications September 2020
16 Rosemount 2140 and 2160:SIS Level Detectors
1.9 China1.9.1 NEPSI Flameproof and Dust
E3
Summary of product certification
Protection Flameproof and dust
Certificate GYJ20.1386X (CCC 认证)
Standards GB 3836.1-2010
GB 3836.2-2010
GB 12476.1-2013
GB 12476.5-2013
Markings Ex d IIC T6~T2 Ga/Gb
Ex td A20 T92°C~T272°C) (T200100°C~T200280°C)
IP66
Temperatures See Flameproof and explosion-proof approval temperatures
See Dust approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
See certificate for special conditions.
1.9.2 NEPSI Intrinsic Safety (IS)
I3
Summary of product certification
Protection Intrinsic Safety (IS)
Certificate GYJ20.1385X (CCC 认证)
Standards GB 3836.1-2010
GB 3836.4-2010
Markings Ex ia IIC T5~T2 Ga
Electricalparameters
Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 12 nF, and Li = 0 mH
Temperatures See Intrinsically Safe (IS) approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
September 2020 Product Certifications
Product Certifications 17
Special Conditions For Safe Use (X)
See certificate for special conditions.
1.10 Russia1.10.1 Technical Regulation Customs Union (EAC) Flameproof and Dust
EM
Summary of product certification
Protection Flameproof and dust
Certificate TC RU C-GB.AA87.В.00728
Markings Ex db IIC T6…T2 X
Ex ta IIIC T92°C…T272°C T500100°C…T500280°C Da X
Temperatures See Flameproof and explosion-proof approval temperatures
See Dust approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
See certificate for special conditions.
1.10.2 Technical Regulation Customs Union (EAC) Intrinsic Safety (IS)
IM
Summary of product certification
Protection Intrinsic Safety (IS)
Certificate TC RU C-GB.AA87.В.00728
Markings OEx ia IIC T5…T2 Ga X
Electricalparameters
Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 0.012 μF, and Li = 0 mH
Temperatures See Intrinsically Safe (IS) approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
See certificate for special conditions
Product Certifications September 2020
18 Rosemount 2140 and 2160:SIS Level Detectors
1.11 India1.11.1 CCOE Flameproof
E1
Summary of product certification
Protection Flameproof
Certificate P408160/1
Standards EN60079-0:2012+A11:2013
EN60079-1:2014
EN60079-26:2015
Markings Ex db IIC T6...T2 Ga/Gb
Temperatures See Flameproof and explosion-proof approval temperatures
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The user is to ensure the probe assembly is installed in such a way toprevent any damage due to impact or ignition source due to friction.
2. Non-standard paint options may cause risk from electrostaticdischarge. Avoid installations that could cause electrostatic build-upof painted surfaces, and only clean the painted surfaces with a dampcloth. If paint is ordered through a special option code, contact themanufacturer for more information.
3. Flameproof joints are not intended for repair.
1.11.2 CCOE Intrinsic Safety (IS)
I1
Summary of product certification
Protection Intrinsic Safety (IS)
Certificate P408161/1
Markings EN60079-0:2012+A11:2013
EN60079-11:2012
EN60079-26:2015
Electricalparameters
Ui = 30 V, Ii = 100 mA, Pi = 0.9 W, Ci = 0.012 μF, and Li = 0 mH
Temperatures See Intrinsically Safe (IS) approval temperatures
September 2020 Product Certifications
Product Certifications 19
Safetyinstructions
See Instructions for hazardous area installations
Special Conditions For Safe Use (X)
1. The equipment, when fitted with the transient suppression terminalblock, is not capable of withstanding the 500 volt insulation test. Thismust be taken into account when installing the equipment.
2. The enclosure may be made of aluminum alloy and given a protectivepolyurethane paint finish. However, care should be taken to protectit from impact or abrasion if located in Zone 0.
3. The enclosure may be given a non-standard paint finish which mayconstitute a potential electrostatic ignition risk. Care should be takento protect it from external conditions conducive to the build-up ofelectrostatic charge on such surfaces. The enclosure must not berubbed or cleaned with a dry cloth.
1.12 Combinations of approvals
K1 Combines I1 and E1
K5 Combines I5 and E5
KB Combines I5, I6, E5, and E6
KZ Combines G5 and G6
1.13 Instructions for hazardous area installations
1.13.1 General safety
1. Installation of this equipment shall be carried out by suitably trainedpersonnel, in accordance with the applicable code of practice.
2. Inspection and maintenance of this equipment shall be carried out bysuitably trained personnel, in accordance with the applicable code ofpractice.
3. The user should not repair this equipment.
4. The certification of this equipment relies upon the followingmaterials used in its construction:
Housing and cover Aluminum alloy ASTM B85 A360.0 orstainless steel 316C12
Probe (partitionwall)
Stainless steel 316/316L, or UNS N10276,or UNS N10002, or UNS N30002
Product Certifications September 2020
20 Rosemount 2140 and 2160:SIS Level Detectors
Seals Silicone
5. If equipment is likely to come into contact with aggressivesubstances, it is the user’s responsibility to take suitable precautionsthat prevent it from being adversely affected, thus ensuring the typeof protection is not compromised.Aggressive substances: Acidic liquids or gases that may attackmetals or solvents that may affect polymeric materials.Suitable precautions: Regular checks as part of routine inspectionsor establishing from a material's data sheet that it is resistant tospecific chemicals.
6. It is the responsibility of the user to ensure:a. The voltage and current limits for this equipment are not
exceeded.
b. The joint requirements between the probe and the vesseltank are compatible with the process media.
c. The joint tightness is correct for the joint material used.
d. Safe working practices for the media and process concernedare followed during the installation and maintenance of theequipment.
7. The probe fork is subjected to small vibration stresses as part of itsnormal function. As this provides a partition wall, it is recommendedthat the fork should be inspected every two years for signs of defects.
8. Technical dataa. The operating pressure must not exceed the rating of the
fitted coupling or flange.
b. For material details and pressure ratings, refer to theRosemount 2140 and 2140:SIS Level Detectors Product DataSheet.
c. Year of manufacture is marked on the product label.
1.13.2 ATEX specific instructions
The equipment has not been assessed as a safety related device (as referredto by directive 2014/34/EU Annex II, clause 1.5).
1.13.3 Intrinsic Safety specific instructions
1. The Rosemount 2140 and 2140:SIS may be used in a hazardous areawith flammable gases and vapors with apparatus groups IIA, IIB andIIC, and temperature classes T1 to T6. The temperature class of the
September 2020 Product Certifications
Product Certifications 21
installation will be determined from the highest process or ambienttemperature.
2. The equipment with Product Certifications code “I8” is suitable forinstallation across the boundary between an area that specificallyrequires Equipment Protection Level Ga (Zone 0) and an area thatspecifically requires Equipment Protection Level Gb (Zone 1). Theprobe forks (and extension tube) are only to be installed in Zone 0.
3. When fitted with the Transient Protection Terminal Block (optioncode T1), the unit does not meet the requirements of clause 6.3.13(Isolation of circuits from earth or frame) in EN 60079-11:2012 (IEC60079-11:2011).
4. Electrical parameters: Ui: 30 V, Ii: 100 mA, Pi: 0.9 W, Ci: 0.012 μF, Li: 0mH
1.13.4 Flameproof and explosion-proof specific instructions
1. The Rosemount 2140 and 2140:SIS may be used in a hazardous areawith flammable gases and vapors with apparatus groups IIA, IIB andIIC, and temperature classes T1 to T6. The temperature class of theinstallation will be determined from the highest process or ambienttemperature.
2. The equipment is suitable for installation across the boundarybetween an area that specifically requires Equipment ProtectionLevel Ga (Zone 0) and an area that specifically requires EquipmentProtection Level Gb (Zone 1). The probe forks (and extension tube)are only to be installed in Zone 0.
3. Repair or modification of flame-paths is not permitted.
4. It is the responsibility of the user to ensure:a. The housing anti-rotation grub screw is screwed in fully.
b. The housing cover locking screws are unscrewed tightlyagainst the covers.
c. Only suitably certified cable entry devices will be utilizedwhen connecting this equipment. The cable entrytemperature may exceed 70 °C.
d. That suitable temperature rated cable is used. Forconnections in an ambient in excess of 60 °C, use cable ratedto at least 90 °C.
e. That any unused cable entries are sealed with suitablycertified stopping plugs.
f. The protective earth terminal is connected to an externalprotective earthing system.
Product Certifications September 2020
22 Rosemount 2140 and 2160:SIS Level Detectors
5. Only screws and fittings provided by the manufacturer can beattached to the housing.
6. Electrical parameters: Vmax = 42.4 Vdc, Imax = 23.5 mA
1.13.5 Dust (protection by enclosure) specific instructions
1. The Rosemount 2140 and 2140:SIS may be used in a hazardous areawith flammable gases and vapors with apparatus groups IIA, IIB andIIC, and temperature classes T1 to T5. The temperature class of theinstallation will be determined from the highest process or ambienttemperature.
2. It is the responsibility of the user to ensure:a. The housing anti-rotation grub screw is screwed in fully.
b. The housing cover locking screws are unscrewed tightlyagainst the covers.
c. Only suitably certified cable entry devices will be utilizedwhen connecting this equipment. The cable entrytemperature may exceed 70 °C.
d. That suitable temperature rated cable is used. Forconnections in an ambient in excess of 60 °C, use cable ratedto at least 90 °C.
September 2020 Product Certifications
Product Certifications 23
1.14 Intrinsically Safe (IS) approval temperatures
Figure 1-1: 2140****M* Temperature De-rating (Intrinsically Safe)
6.263"
T4...T1
T5...T1 T3...T1
70
60.7
53
40
0
-60-40 0 9570 130 180
Process Temperature (°C)
Am
bien
t Tem
pera
ture
(°C
)
Figure 1-2: 2140****E* Temperature De-rating (Intrinsically Safe)
T5...T1 T3...T1 T2...T1
70
60
40
66.8 63.4
0
-60-70 0 9570 130 195 260
T4...T1
Process Temperature (°C)
Am
bien
t Tem
pera
ture
(°C
)
Product Certifications September 2020
24 Rosemount 2140 and 2160:SIS Level Detectors
1.15 Flameproof and explosion-proof approval temperatures
Figure 1-3: 2140****M* Temperature De-rating (Flameproof/Explosion-proof)
T5...T1
T6...T1 T4...T1 T3...T1
80
77706560
0
-40-40 0 80 95 130 180
Process Temperature (°C)
Am
bien
t Tem
pera
ture
(°C)
Figure 1-4: 2140****E* Temperature De-rating (Flameproof/Explosion-proof)
T5...T1
T6...T1 T4...T1 T3...T1 T2...T1
8079 77
7073.5
0
65
-40-70 0 80 95 130 195 260
Process Temperature (°C)
Am
bien
t Tem
pera
ture
(°C
)
September 2020 Product Certifications
Product Certifications 25
1.16 Dust approval temperaturesMinimum ambient air temperature (Ta) = -20 °C
Figure 1-5: 2140****M* Maximum Surface Temperature
200
10092
19280
60
8000
-40 180
Ambient
T500***°C
T***°C
Process Temperature (°C)
Surf
ace
Tem
pera
ture
(°C
)
Max
imum
Am
bien
t Tem
pera
ture
(°C
)
Figure 1-6: 2140****E* Maximum Surface Temperature
Ambient
T500***°C
T***°C
280
10092
27280
70
8000-70 260
Process Temperature (°C)
Surf
ace
Tem
pera
ture
(°C
)
Am
bien
t Tem
pera
ture
(°C
)
Product Certifications September 2020
26 Rosemount 2140 and 2160:SIS Level Detectors
1.17 Installation drawings
Figure 1-7: USA and Canada installation drawing 71097/1387 (Page 1)
APPROVAL
TITLE
APPR.DRG. I.S. &
NONINCENDIVE
CONTROL DRAWING FOR
ROSEMOUNT 2140
DOCUMENT NUMBER: -
71097/1387
Page 1 of 5
AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16
REVISION DATE ECO No. NAME APPROVED SEE ECO
L2073 Iss. AA© ROSEMOUNT MEASUREMENT LIMITED 2017
THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS
CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR
PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.
CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.
GENERAL NOTES:
1. ASSOCIATED APPARATUS MANUFACTURER'S INSTALLATION DRAWINGMUST BE FOLLOWED WHEN INSTALLING THIS EQUIPMENT.
2. CONTROL EQUIPMENT CONNECTED TO BARRIER MUST NOTUSE OR GENERATE MORE THAN 250 Vrms or Vdc,
3. RESISTANCE BETWEEN INTRINSICALLY SAFE GROUND ANDEARTH GROUND MUST BE LESS THAN 1 OHM,
4. INSTALLATION SHOULD BE IN ACCORDANCE WITH APPLICABLELAWS/REGULATIONS AND CODE OR PRACTICE. I.E. FOR CANADA,THE CANADIAN ELECTRICAL CODE (CSA C22.1);FOR AMERICA, THE NATIONAL ELECTRICAL CODE (ANSI/NFPA 70)AND ANSI/ISA-RP12.6 “INSTALLATION OF INTRINSICALLY SAFESYSTEMS FOR HAZARDOUS (CLASSIFIED) LOCATIONS”.
5. THE ASSOCIATED APPARATUS, BARRIER OR ISOLATOR MUST BE APPROVED.FOR CANADA, TO CANADIAN STANDARDS BY AN NRTL ACCREDITED BY THESTANDARDS COUNCIL OF CANADA (SCC).FOR AMERICA, TO AMERICAN STANDARDS BY AN NRTL ACCREDITED BYTHE OCCUPATIONAL SAFETY AND HEALTH ADMINISTRATION (OSHA).
6. WARNING -SUBSTITUTION OF COMPONENTS MAY IMPAIRINTRINSIC AND NON-INCENDIVE SAFETY, AVERTISSEMENT: LA SUBSTITUTION DE COMPOSANTS PEUTCOMPROMETTRE LA SÉCURITÉ INTRINSÈQUE ET LA SÉCURITÉNON INCENDIAIRIES
7. ASSOCIATED APPARATUS MUST MEET THE FOLLOWING PARAMETERS:Uo or Voc or Vt LESS THAN or EQUAL TO Ui (Vmax)Io or Isc or It LESS THAN or EQUAL TO Ii (Imax)Po or Pmax LESS THAN or EQUAL TO Pi (Pmax)Ca IS GREATER THAN or EQUAL THE SUM OF ALL Ci's PLUS CcableLa IS GREATER THAN or EQUAL THE SUM OF ALL Li's PLUS Lcable
8. THE ASSOCIATED APPARATUS MUST BE A RESISTIVELY LIMITED SINGLE OR MULTIPLE CHANNEL APPROVED BARRIER HAVING PARAMETERS LESS THAN THOSEQUOTED, AND FOR WHICH THE OUTPUT AND THE COMBINATIONS OF OUTPUTS ISNON-IGNITION CAPABLE FOR THE CLASS, DIVISION AND GROUP OF USE,
9. FIELD WIRING SHOULD BE RATED TO 70°C MINIMUM,
September 2020 Product Certifications
Product Certifications 27
Figure 1-8: USA and Canada installation drawing 71097/1387 (Page 2)
APPROVAL
TITLE
APPR.DRG. I.S. &
NONINCENDIVE
CONTROL DRAWING FOR
ROSEMOUNT 2140
DOCUMENT NUMBER: -
71097/1387
Page 2 of 5
AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16
REVISION DATE ECO No. NAME APPROVED SEE ECO
L2073 Iss. AA
© ROSEMOUNT MEASUREMENT LIMITED 2017
THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS
CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR
PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.
CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.
DIV 1 INSTALLATION OPTIONS
THE ROSEMOUNT TRANSMITTER IS APPROVED AS INTRINSICALLY SAFE WHEN USED IN CIRCUIT WITH APPROVEDBARRIERS WHICH MEET THE ENTITY PARAMETERS LISTED IN THE CLASS I. DIVISION 1 GROUPS INDICATED.ADDITIONALLY, THE ROSEMOUNT 751 FIELD SIGNAL INDICATOR IS APPROVED AS INTRINSICALLY SAFE WHEN CONNECTED IN CIRCUIT WITH ROSEMOUNT TRANSMITTERS AND APPROVED BARRIERS WHICH MEET THE ENTITY PARAMETERS LISTED FOR CLASS I, DIVISION 1, GROUPS INDICATED.
TO ASSURE AN INTRINSICALLY SAFE SYSTEM. THE TRANSMITTER AND BARRIER MUST BE WIRED IN ACCORDANCEWITH THE BARRIER MANUFACTURER'S FIELD WIRING INSTRUCTIONS AND THE APPLICABLE CIRCUIT DIAGRAM.
Product Certifications September 2020
28 Rosemount 2140 and 2160:SIS Level Detectors
Figure 1-9: USA and Canada installation drawing 71097/1387 (Page 3)
APPROVAL
TITLE
APPR.DRG. I.S. &
NONINCENDIVE
CONTROL DRAWING FOR
ROSEMOUNT 2140
DOCUMENT NUMBER: -
71097/1387
Page 3 of 5
AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16
REVISION DATE ECO No. NAME APPROVED SEE ECO
L2073 Iss. AA© ROSEMOUNT MEASUREMENT LIMITED 2017
THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS
CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR
PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.
CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.
ENTITY CONCEPT APPROVALSTHE ENTITY CONCEPT ALLOWS INTERCONNECTION OF INTRINSICALLY SAFE APPARATUS TO ASSOCIATED APPARATUS NOT SPECIFICALLY EXAMINED IN COMBINATION AS A SYSTEM. THE APPROVED VALUES OF MAX. OPEN CIRCUIT VOLTAGE (Voc OR Vt) AND MAX. SHORT CIRCUIT CURRENT (Isc OR It) AND MAX,POWER (Voc X Isc/4) OR (Vt X It/4), FOR THE ASSOCIATED APPARATUS MUST BE LESS THAN OR EQUAL TO THE MAXIMUM SAFE INPUT VOLTAGE (Vmax), MAXIMUM SAFE INPUT CURRENT (Imax), AND MAXIMUM SAFE INPUT POWER (Pmax) OF THE INTRINSICALLY SAFE APPARATUS. IN ADDITION, THE APPROVED MAX. ALLOWABLECONNECTED CAPACITANCE (Ca) OF THE ASSOCIATED APPARATUS MUST BE GREATER THAN THE SUM OF THE INTERCONNECTING CABLE CAPACITANCE AND THE UNPROTECTED INTERNAL CAPACITANCE (Ci) OF THE INTRINSICALLY SAFE APPARATUS, AND THE APPROVED MAX. ALLOWABLE CONNECTED INDUCTANCE (La) OF THE ASSOCIATED APPARATUS MUST BE GREATER THAN THE SUM OF THE INTERCONNECTING CABLE INDUCTANCE AND THE UNPROTECTED INTERNAL INDUCTANCE (Li) OF THE INTRINSICALLY SAFE APPARATUS,
NOTE: ENTITY PARAMETERS LISTED APPLY ONLY TO ASSOCIATED APPARATUS WITH LINEAR OUTPUT.
Vmax= 30V Vt OR Voc IS LESS THAN OR EQUAL TO 30V
Imax = 100mA It OR Isc IS LESS THAN OR EQUAL TO 100mA
Pmax = 0.9 WATT ( ) OR ( ) IS LESS THAN OR EQUAL TO 0.9 WATT
Ci = 0.012µF Ca IS GREATER THAN 0.012µF
Li = 0mH La IS GREATER THAN 0mH
CLASS I, DIV 1, GROUPS A AND B
CLASS I, DIV 1, GROUPS C AND D
Vmax= 30V Vt OR Voc IS LESS THAN OR EQUAL TO 30V
Imax = 100mA It OR Isc IS LESS THAN OR EQUAL TO 100mA
Pmax = 0.9 WATT ( ) OR ( ) IS LESS THAN OR EQUAL TO 0.9 WATT
Ci = 0.012µF Ca IS GREATER THAN 0.012µF
Li = 0mH La IS GREATER THAN 0mH
DIV 2 INSTALLATION OPTIONS
September 2020 Product Certifications
Product Certifications 29
Figure 1-10: USA and Canada installation drawing 71097/1387 (Page 4)
APPROVAL
TITLE
APPR.DRG. I.S. &
NONINCENDIVE
CONTROL DRAWING FOR
ROSEMOUNT 2140
DOCUMENT NUMBER: -
71097/1387
Page 4 of 5
AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16
REVISION DATE ECO No. NAME APPROVED SEE ECO
L2073 Iss. AA
© ROSEMOUNT MEASUREMENT LIMITED 2017
THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS
CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR
PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.
CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.
CLASS I, DIV.2 HAZARDOUS (CLASSIFIED) LOCATION
APPROVED
ASSOCIATED
NONINCENDIVE
APPARATUS [1]
UNSPECIFIED
EQUIPMENT [3]
UNCLASSIFIED
LOCATION
MECHANICALLY
PROTECTED
WIRE PER APPLICABLE
ELECTRICAL CODE
NOTES:
[1] ASSOCIATED NON-INCENDIVE APPARATUS PARAMETERS SHALL BE THE SAME AS THOSE SHOWN ON PAGE 3
[2] MUST BE INSTALLED IN ACCORDANCE WITH THE APPLICABLE ELECTRIC CODE FOR WIRING IN DIVISION 2 HAZARDOUS (CLASSIFIED) LOCATIONS.
[3] SUPPLIED BY A CLASS 2 OR LIMITED ENERGY SOURCE IN ACCORDANCE WITH CAN/CSA-C22.2 No. 61010-1-12
Product Certifications September 2020
30 Rosemount 2140 and 2160:SIS Level Detectors
Figure 1-11: USA and Canada installation drawing 71097/1387 (Page 5)
APPROVAL
TITLE
APPR.DRG. I.S. &
NONINCENDIVE
CONTROL DRAWING FOR
ROSEMOUNT 2140
DOCUMENT NUMBER: -
71097/1387
Page 5 of 5
AB 24/01/17 MBY-05601 GP DRAWN JPA 19/10/16
REVISION DATE ECO No. NAME APPROVED SEE ECO
L2073 Iss. AA© ROSEMOUNT MEASUREMENT LIMITED 2017
THE REPRODUCTION, DISTRIBUTION, & UTILISATION OF THIS DOCUMENT AS WELL AS THE COMMUNICATION OF ITS
CONTENTS TO OTHERS WITHOUT EXPRESS AUTHORISATION IS PROHIBITED. OFFENDERS WILL BE HELD LIABLE FOR
PAYMENT OF DAMAGES. ALL RIGHTS RESERVED IN THE EVENT OF PATENT, UTILITY MODEL OR DESIGN.
CERTIFIED PRODUCT: ALTERATIONS TO THIS DOCUMENT MUST BE APPROVED BEFORE IMPLEMENTATION.
NONINCENDIVE FIELD CIRCUITCLASS I. DIV. 2 LOCATIONS
NON-HAZARDOUS DIVISION 2 HAZARDOUS (CLASSIFIED) LOCATIONLOCATION Vmax1 Vmax2 Vmax3 VmaxN
Ci1 Ci2 Ci3 CiNLi1 Li2 Li3 LiN
Imax1 Imax2 Imax3 ImaxN
APPROVEDNONINCENDIVE
SUPPLY
WIRING PER APPLICABLEELECTRICAL CODE
IN NORMAL OPERATION
DEVICES CONTROL THROUGH-CURRENT
4-20mA/HART Vmax 42.4VMaxImum normal 23.5mAoperatIng current Ca 0.012µF La 0µH
ROSEMOUNT 2140 TRANSMITTERS ARE CURRENT CONTROLLERSON INDIVIDUAL PARALLEL BRANCHES WITH RESPECT TO THEPOWER SUPPLY. IN NONINCENDIVE INSTALLATIONS THE ImaxFOR EACH TRANSMITTER IS NOT RELATED TO THE MAXIMUMCURRENT OF THE POWER SUPPLY (Isc) lN THE SAME MANNERAS FOR TRANSMITTER INSTALLED PER I.S. REQUIREMENTS.BECAUSE NONINCENDIVE REQUIREMENTS INCLUDE ONLYNORMAL OPERATING CONDITIONS.IT SHALL BE SUPPLIED BY A CLASS 2 OR LIMITED ENERGYSOURCE IN ACCORDANCE WITH CAN/CSA22.2 No. 61010-1-12
PARAMETERS DEVICEVoc = MInImum of (Vmax1, Vmax2, ...., VmaxN) Imax1 >= Iq1 + Isignal1 Imax2 >= Iq2 + Isignal2 , , , ImaxN >= IqN + IsignalN
Ca <= Ci1 + Ci2 + ... + CiN + CcableLa <= Li1 + Li2 + ... + LiN + Lcable
Imax for an Individual device = Iq + Isignal
Iq = Quiescent current through devlce (Maximum quiescent current for the device)Isignal = Signaling current through device (Protocol may limit signaling to one device at a time)
OperatIng Imax = Iq1 + Iq2 + ... + IqN + Isignal max
Isignal max = Max. of (lsignal1, Isignal2 ..... IsignalN)
D D D D
MECANICALLY PROTECTEDWIRE PER APPLICABLE
ELECTRICAL CODE
September 2020 Product Certifications
Product Certifications 31
1.18 EU Declaration of Conformity
Figure 1-12: EU Declaration of Conformity (Page 1)
EU Declaration of Conformity No: RMD 1110 Rev. AG
ne 3 fo 1 egaP
We,
Rosemount Tank Radar AB Layoutvägen 1 S-435 33 MÖLNLYCKE Sweden
declare under our sole responsibility that the product,
Rosemount™ 2140 Vibrating Fork Liquid Level Detector
manufactured by,
Rosemount Tank Radar AB Layoutvägen 1 S-435 33 MÖLNLYCKE Sweden
to which this declaration relates, is in conformity with the provisions of the European Union
Directives, including the latest amendments, as shown in the attached schedule.
Assumption of conformity is based on the application of the harmonized standards and, when
applicable or required, a European Union notified body certification, as shown in the attached
schedule.
(signature)
Manager Product Approvals (function)
Dajana Prastalo (name)
24-Jun-20;
(date of issue)
Product Certifications September 2020
32 Rosemount 2140 and 2160:SIS Level Detectors
Figure 1-13: EU Declaration of Conformity (Page 2)
EU Declaration of Conformity No: RMD 1110 Rev. AG
ne 3 fo 2 egaP
EMC Directive (2014/30/EU)
Harmonized Standards: EN 61326-1:2013, EN 61326-2-3 :2013
Other Standards Used: EN 61326-3-1:2008
ATEX Directive (2014/34/EU)
Dekra 16ATEX0082X – (Flameproof)
Equipment Group II Category 1/2 G
Ex db IIC T6...T2 Ga/Gb
Harmonized Standards: EN 60079-0:2012+A11:2013, EN 60079-1:2014,
EN 60079-26:2015
Baseefa 16ATEX0136X – (Intrinsic safety)
Equipment Group II Category 1G
Ex ia IIC T5...T2 Ga
Equipment Group II Category 1/2 G
Ex ib IIC T5...T2 Ga/Gb
Harmonized Standards: IEC EN 60079-0:2018, EN 60079-11:2012, EN
60079-26:2015
Baseefa 16ATEX0137X – (Dust Protection by Enclosure)
Equipment Group II Category 1 D
Ex ta IIIC (T92°C…T272°C) (T200100°C…T200280°C) Da
Harmonized Standards: IEC EN 60079-0:2018, EN 60079-31:2014
RoHS Directive (2011/65/EU)
The Model 2140 is in conformity with Directive 2011/65/EU of the European Parliament
and of the Council on the restriction of the use of certain hazardous substances in electrical
and electronic equipment.
(Minor variations in design to suit the application and/or mounting requirements are identified by alpha/numeric
characters where indicated * above)
September 2020 Product Certifications
Product Certifications 33
Figure 1-14: EU Declaration of Conformity (Page 3)
EU Declaration of Conformity No: RMD 1110 Rev. AG
ne 3 fo 3 egaP
ATEX Directive Notified Body
SGS Fimko Oy [Notified Body Number: 0580]
Takomotie 8
FI-00380, Helsinki
Finland
DEKRA Certification B.V. [Notified Body Number: 0344]
Meander 1051
6825 MJ, Arnhem
Netherlands
ATEX Notified body for Quality Assurance
DNV Nemko Presafe AS [Notified Body Number: 2460]
Veritasveien 1
1322 HØVIK
Norway
Product Certifications September 2020
34 Rosemount 2140 and 2160:SIS Level Detectors
1.19 China RoHS
含含有有China RoHS管管控控物物质质超超过过最最大大浓浓度度限限值值的的部部件件型型号号列列表表 Rosemount 2140List of Rosemount 2140 Parts with China RoHS Concentration above MCVs
部部件件名名称称 Part Name
有有害害物物质质 / Hazardous Substances
铅铅 Lead(Pb)
汞汞 Mercury
(Hg)
镉镉 Cadmium
(Cd)
六六价价铬铬 Hexavalent Chromium
(Cr +6)
多多溴溴联联苯苯 Polybrominated
biphenyls(PBB)
多多溴溴联联苯苯醚醚 Polybrominated diphenyl ethers
(PBDE)
电子组件 Electronics Assembly
O O O O O
壳体组件 Housing
Assembly O O O O O
传感器组件Sensor
AssemblyX O O O O
本表格系依据SJ/T11364的规定而制作. This table is proposed in accordance with the provision of SJ/T11364.
O: 意为该部件的所有均质材料中该有害物质的含量均低于GB/T 26572所规定的限量要求.O: Indicate that said hazardous substance in all of the homogeneous materials for this part is below the limit requirement ofGB/T 26572.
X: 意为在该部件所使用的所有均质材料里,至少有一类均质材料中该有害物质的含量高于GB/T 26572所规定的限量要求.X: Indicate that said hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement of GB/T 26572.
O
O
O
September 2020 Product Certifications
Product Certifications 35
*00825-0200-4140*Product Certifications
00825-0200-4140, Rev. AGSeptember 2020
Emerson Automation Solutions6021 Innovation Blvd.Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888
+1 952 949 7001
North America Regional OfficeEmerson Automation Solutions8200 Market Blvd.Chanhassen, MN 55317, USA
+1 800 999 9307 or +1 952 906 8888
+1 952 949 7001
Latin America Regional OfficeEmerson Automation Solutions1300 Concord Terrace, Suite 400Sunrise, FL 33323, USA
+1 954 846 5030
+1 954 846 5121
Europe Regional OfficeEmerson Automation Solutions EuropeGmbHNeuhofstrasse 19a P.O. Box 1046CH 6340 BaarSwitzerland
+41 (0) 41 768 6111
+41 (0) 41 768 6300
Asia Pacific Regional OfficeEmerson Automation Solutions1 Pandan CrescentSingapore 128461
+65 6777 8211
+65 6777 0947
Middle East and Africa Regional OfficeEmerson Automation SolutionsEmerson FZE P.O. Box 17033Jebel Ali Free Zone - South 2Dubai, United Arab Emirates
+971 4 8118100
+971 4 8865465
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Youtube.com/user/RosemountMeasurement
©2020 Emerson. All rights reserved.
Emerson Terms and Conditions of Sale areavailable upon request. The Emerson logo is atrademark and service mark of Emerson ElectricCo. Rosemount is a mark of one of the Emersonfamily of companies. All other marks are theproperty of their respective owners.