qc delta pv ppt student 09
TRANSCRIPT
-
8/14/2019 QC Delta PV Ppt Student 09
1/8
CLS 500 Application and Interpretation of Clinical Laboratory Data
QC, Delta Check and Predictive Value Lecture
Quality Control, Delta Checks
and Predictive Value
CLS 500 Application and Interpretation of ClinicalLaboratory Data
Ricki Otten, MT(ASCP)SC
2
1. Discuss the use of quality control in
ensuring the validity of laboratory testresults
2. Define the acceptable range as it
relates to quality control and how this
range is determined.
Objectives
3
3. Define the terms as they relate to quality
control:
a. Accuracy
b. Precision
c. Reliability
4. Explain the use of delta checks to
assure accurate patient test results
Objectives
4
Objectives
5. Define the following terms:
a. Test sensitivity
b. Test specificity
c. Predictive value
6. Discuss how sensitivity, specificity and
predictive value are used to assess the
diagnostic usefulness of a test
5
Who needs it?
What is it?
Why is it done?
How is it used?
Quality Control (QC)
6
The most basic and fundamental
expectation of a laboratory is
-
8/14/2019 QC Delta PV Ppt Student 09
2/8
CLS 500 Application and Interpretation of Clinical Laboratory Data
QC, Delta Check and Predictive Value Lecture
7
QC is a statistical tool that is used to
assure the reliability of test results
Statistical quality control practices are
critical for assuring test results are correct
Quality Control (QC)
8
QC: What is it?
The value of the quality control specimen isknown
and is represented by an acceptable range of
values that has been statistically calculated
Acceptable range = Mean 2 SD
For each control level for each assay:
1. Control is analyzed repeatedly on separate days
2. Values are collected until 30+ values are obtained
3. Statistics are calculated: mean, SD, %CV
4. Acceptable range is determined
9Acceptable range = Mean 2SD = 6.8 0.6 = 6.2 7.4 g/dl 10
QC statistically validates the accuracy
(correctness) of
Methodology
Testing procedure
Reagents Analytical Run
Entire test system
Instruments
QC: What Does it Do?
11
QC cannot determine if the sample
submitted is from the correct patient!
12
The laboratory wants to report the mostaccurate and precise values
QC samples are analyzed along with thepatient samples
Evaluation of QC results determines thevalidity of the analytical run
If analytical run is valid, then we are confidentthe patient results are also valid
QC: How is it Used?
-
8/14/2019 QC Delta PV Ppt Student 09
3/8
CLS 500 Application and Interpretation of Clinical Laboratory Data
QC, Delta Check and Predictive Value Lecture
13
If QC results are within acceptable limits
then the analytical run is validated as accurate.
This means the patient results are valid and can
be reported
14
15
If QC results are not within acceptable limitsthen the analytical run is not valid.
This indicates an analytical error has occurred that
jeopardizes the patient results.
This means the patient results may not be valid
and cannot be reported.
After troubleshooting the error, all samples must
be reanalyzed and QC must be statistically
re-evaluated before patient test results are reported16
17
What does accuracy and precision mean?
18
-
8/14/2019 QC Delta PV Ppt Student 09
4/8
CLS 500 Application and Interpretation of Clinical Laboratory Data
QC, Delta Check and Predictive Value Lecture
19
Accuracy refers to the closeness of a
measurement to the true value
Precision refers to the reproducibility of a
measurement
Reliability refers to the accuracy and
precision of a measurement
20
Quality Control Exercise
Evaluate the following analytical run in
relation to test validity.
Should the patient test results be released
to the medical record?
Why or why not?
21
69
150
43
132
97
1. Level I control
2. Level II control
3. John Smother
4. Gomer Pyle
5. Jayne Sealess
Result (mg/dl)Analytical Run: plasma glucose
162 198 mg/dl180 +/-9 mg/dlLevel II
62 78 mg/dl70 +/-4 mg/dlLevel I
Acceptable
RangeMean +/- 2SD
Quality Control Acceptable Limits
Critical low: < 45 mg/dl
Critical high: > 450 mg/dl
Fasting plasma glucose: 70-99 mg/dl
Expected Reference Range and Critical
Limits
Is level I control within acceptable limits?Is level II control within acceptable limits? 22
69 OK
150 Not OK
43
132
97
1. Level I control
2. Level II control
3. John Smother
4. Gomer Pyle
5. Jayne Sealess
Result (mg/dl)Analytical Run: plasma glucose
162 198 mg/dl180 +/-9 mg/dlLevel II
62 78 mg/dl70 +/- 4 mg/dlLevel I
Acceptable
RangeMean +/- 2SD
Quality Control Acceptable Limits
Critical low: < 45 mg/dl
Critical high: > 450 mg/dl
Fasting plasma glucose: 70-99 mg/dl
Expected Reference Range and Critical
Limits
Can patient test results be reported? Why/why not?
23
69 OK
150 Not OK
43
132
97
1. Level I control
2. Level II control
3. John Smother
4. Gomer Pyle
5. Jayne Sealess
Result (mg/dl)Analytical Run: plasma glucose
162 198 mg/dl180 +/-9 mg/dlLevel II
62 78 mg/dl70 +/-4 mg/dlLevel I
Acceptable
RangeMean +/- 2SD
Quality Control Acceptable Limits
Critical low: < 45 mg/dl
Critical high: > 450 mg/dl
Fasting plasma glucose: 70-99 mg/dl
Expected Reference Range and CriticalLimits
Can patient test results be reported? Why/why not?
No; the validity of the analytical run is questioned because
control level II is outside of acceptable limits. All testing must
be repeated and when all controls are within acceptable limits,
the patient test results may be reported.24
Delta Check
Another tool used by the laboratory to
assure accurate patient test results for
specific analytes
The most recent test result for a particular
patient is compared to the most previous
test result for that patient
-
8/14/2019 QC Delta PV Ppt Student 09
5/8
CLS 500 Application and Interpretation of Clinical Laboratory Data
QC, Delta Check and Predictive Value Lecture
25
Delta Check
If the test result has changed significantlyfrom the last time the test was done, a
delta check failure occurs
The delta check failure means the changein test results is not physiologicallypossible
The allowable difference is pre-determined by various investigators andclinicians 26
Delta Check
The technologist is now responsible to
determine the cause of the delta check
failure before releasing test results
Test sample may have been compromised
Test interference: hemolyzed, lipemic, icteric
Sample not handled properly
Drawn at incorrect time: trough, peak, tolerance
Drawn in incorrect sample tube: B-R-G-P-G
Treatment?
Is this the correct patient?
Redraw the specimen?
27
Jan.1 Jan.8 Jan.12 Jan. 13
BUN 88 95 105 22*
(mg/dl)
*Delta check failure
What is a likely cause for this failure?
Should test results be reported?28
Jan.1 Jan.2 Jan.3 Jan.4
K+ 3.9 3.6 4.1 6.9*
(mmol/L)
*Delta check failure
What is a likely cause for this failure?
Should test results be reported?
29
Diagnostic Efficacy
Diagnostic sensitivity
Diagnostic specificity
Positive predictive value Negative predictive value
30
Ideal
All persons with disease will test positive
(referred to as diagnostic sensitivity)
All persons without disease will test negative
(referred to as diagnostic specificity)
Test results would be: 100% sensitive
100% specific
-
8/14/2019 QC Delta PV Ppt Student 09
6/8
CLS 500 Application and Interpretation of Clinical Laboratory Data
QC, Delta Check and Predictive Value Lecture
31
Reality
All methods have an inherent amount of
error that will affect test results
No method is able to detect all persons
with disease accurately -and-
No method is able to detect all persons
without disease accurately
32
Possible Test Outcomes
- (TN)- (FN)
Negative testresult
+ (FP)+ (TP)
Positive test
result
Patients withoutdisease
Patients withdisease
33
- (TN)- (FN)
Negative testresult
+ (FP)+ (TP)
Positive test
result
Patients withoutdisease
Patients withdisease
Test sensitivity: TP/(TP + FN) x 100
Test specificity: TN/(TN + FP) x 100
Pos predictive value: TP/(TP + FP) x 100
Neg predictive value: TN/(TN + FN) x 100 34
Hypothetical Example
A total of 3000 patients were assessed for
acute myocardial infarction (AMI) with
Troponin I (cTnI) measurements:
1000 patients tested positive for cTnI and
of those, 980 had diagnostic AMI;
2000 patients tested negative for cTnI and
of those, 30 had diagnostic AMI
35
Determine the sensitivity, specificity,
positive predictive value (+PV) and
negative predictive value (-PV) of the
test used for assessment of AMI
36
Construct 2x2 Table
Neg test
result
Pos test
result
Patientswithout AMI
Patientswith AMI
-
8/14/2019 QC Delta PV Ppt Student 09
7/8
CLS 500 Application and Interpretation of Clinical Laboratory Data
QC, Delta Check and Predictive Value Lecture
37
Calculate Test Sensitivity
Test sensitivity: TP/(TP + FN) x 100
38
Calculate Test Specificity
Test specificity: TN/(TN + FP) x 100
39
Calculate Pos Predictive Value
Pos predictive value: TP/(TP + FP) x 100
40
Calculate Neg Predictive Value
Neg predictive value: TN/(TN + FN) x 100
41
PSA values >4.0 ng/ml
Sensitivity = 79% Specificity = 46%
+PV = 33% -PV = 87%
265 (TN)41 (FN)
Neg test result
(4.0 ng/ml)
Patients withoutprostate cancer
Patients withprostate cancer
42
High Cut-off Value
If a high cutoff value is used,
then the specificity of the test can reach100%
However, the sensitivity will decrease
-
8/14/2019 QC Delta PV Ppt Student 09
8/8
CLS 500 Application and Interpretation of Clinical Laboratory Data
QC, Delta Check and Predictive Value Lecture
43
Low Cut-off Value
If a low cutoff value is used,
then the sensitivity of the test can reach
100%
However, the specificity will decrease
44
Disease PrevalenceUsing PSA cut off value of 4.0 ng/ml:
Sensitivity = 79%; Specificity = 46%; +PV = 33%; -PV = 87%)
59.4%0.5 (50% of pop)
26.8%0.2 (20% of pop)
14.0%0.1 (10% of pop)
1.5%0.01 (1% of pop)
0.1%0.001 (0.1% of population)
Pos Predictive ValuePrevalence of disease