quality assurance activities in coagulation_ippt_97

Upload: kesavanveera

Post on 03-Jun-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    1/56

    Quality Activities inCoagulation Laboratory

    Kiu Ling Ling

    Application Manager

    All Eights (M) SDN.BHD.

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    2/56

    Outline

    1. Establish Control Range with commercial QC

    material

    2. Transference & Validation of reference interval

    3. Calibration of PT-INR system with certified

    plasma

    4. Factor assay: constructing calibration curve

    for accurate result

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    3/56

    Establish Control Range

    with Commercial QCMaterial

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    4/56

    What

    Why

    When

    How

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    5/56

    WHATis Internal Quality Control ?

    Set of procedures undertaken by laboratory

    staffs for the continuous monitoring of operationsand the results of the measurement in order to

    decide whether the results are reliable enough to

    be released.

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    6/56

    IQC

    To monitor the

    accuracy &precision of the

    completeanalyticalprocess

    To detectimmediateanalytical

    errorsDay to Day

    monitoring oftestperformance

    quality

    ( long term

    constancy )

    WHYwe need IQC ?

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    7/56

    Control Material- Important

    CharacteristicsSimilar to test specimens ( Matrix )

    Can be liquid or freeze-dried ( lyophilized )

    known concentration of analyte 2 or 3 levels of control ( normal & abnormal )

    Values cover medical decision points

    Stability up to at least 8 hoursQC material must be tested in the same

    manners as patient sample

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    8/56

    To Establish Control Range

    Collecting Control data

    Perform control repeatedly (minimum 20times) with testing spread over at least 10section

    Calculate Mean , Standard Deviation (SD)

    Determine control limit ( Mean 2SD )

    Plotting Levey-Jennings chart

    - Manually on graph paper

    - MS Excel Software

    - Built in quality software in coagulationanalyzer

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    9/56

    How to Create Levey-Jennings Graph

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    10/56

    APTT normal control data

    1. 23.0 sec

    2. 22.0 sec

    3. 22.5 sec

    4. 23.2 sec

    5. 22.2 sec6. 22.5 sec

    7. 22.8 sec

    8. 23.3 sec

    9. 22.5 sec10. 23.1 sec

    11. 22.4 sec

    12. 22.6 sec

    13. 23.0 sec

    14. 23.0 sec

    15. 22.8 sec

    16. 22.5 sec

    17. 23.0 sec

    18. 23.0 sec

    19. 22.9 sec20. 22.5 sec

    Step 1

    Calculate MEAN

    Mean = Sum ( X1 + X2 +.X20)

    Total Number of Data

    (N)

    MEAN = 454.8 / 20

    = 22.7 sec

    Step 2

    Calculate SD

    SD = 0.4 secSum = 454.8 sec

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    11/56

    Step 3

    Set control limit In Gaussian Distribution :

    68.3% of data is within 1.0 SD

    95.5% of data is within 2.0

    SD

    99.7% of data is within 3.0 SD

    Control limit for APTT

    = Mean 2SD

    = 21.9 23.5 sec

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    12/56

    Step 4: Plotting LJ graph

    MEAN

    +1SD

    +2SD

    -1SD

    -2SD

    -3SD

    +3SD

    22.7

    23.1

    23.5

    23.9

    22.3

    21.9

    21.5

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    13/56

    CLSI C28-A3:Define, Establishing, and

    Verifying ReferenceIntervals in the Clinical

    Laboratory

    3rd

    Edition

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    14/56

    Reference Interval Minimum 120 healthy donors ( CLSI 28-A3)

    Criteria of donor selection

    equal number of sex

    preferred 1650 years old no under any anticoagulant therapy and

    medication

    collect sample over few daysday-to-day variation

    Normal Range = Mean 2SD

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    15/56

    Reference Interval Establishment

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    16/56

    Local PT/INR System Calibration

    Norma

    lRange:Exam

    ple

    ID PT ( sec) APTT (sec) Fib (g/l) TT (Sec)

    1820425 13.4 39.7 3.7 15.8

    1712056 13.2 41.0 3.0 15.0

    1712084 13.8 41.9 2.7 15.9

    1820420 13.5 40.5 3.1 15.9

    1820421 12.8 33.2 3.0 16.2

    1712088 13.4 37.6 3.6 17.2

    1820422 13.0 35.6 3.4 16.3

    1820369 13.5 37.9 3.1 16.5

    1712087 13.5 41.7 3.1 16.6

    1712086 13.0 39.1 3.0 16.6

    1712085 13.1 33.4 3.3 16.1

    1822001 13.0 41.2 3.0 17.5

    D14 12.8 37.8 2.9 15.7

    D15 13.5 35.6 3.2 16.0

    D16 13.1 38.2 2.6 15.8

    D17 12.8 40.8 3.3 17.4

    D18 13.6 41.0 3.4 15.9

    D19 13.0 36.0 2.4 16.4

    D20 12.6 31.7 2.3 17.0

    D21 13.6 35.6 2.7 15.8

    D22 13.0 33.7 2.7 17.4

    D23 13.4 33.5 2.3 17.2

    D24 13.2 36.9 2.6 17.0

    D25 13.3 33.8 3.0 17.3

    D26 12.7 35.0 3.5 15.4

    Mean 13.7 37.2 3.0 16.4

    SD 0.4 3.1 0.4 0.7

    Normal Range

    Mean2SD 12.513.9 sec 30.943.4 sec 2.23.7g/l 15.017.8 sec

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    17/56

    Challenge in Determination of

    Reference IntervalTo determine reliability of reference

    interval (RI) costly & heavy task

    minimum 120 healthy subjects needed

    Difficulty in collecting sufficient number ofqualified reference individuals

    Transfer of reference range from manufactureror other laboratory

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    18/56

    Transference of RI

    If Laboratory has previously established a

    RI for its own population

    simple comparison between two analytical system

    ( new & old ) could be sufficient

    Advantages

    no need to collect sample from referenceindividuals

    can use existing patients samples

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    19/56

    Transference of RI

    If completely comparable ( Excellentcorrelation coefficient , small slope bias, smallintercept, comparable range of value )

    y = 1.004x0.628, r2 = 0.990 Assume the RI for the current method is 50150

    when x = 50, y = 1.004 (50)0.628 = 50

    When x = 150, y = 1.004 (150)0.628 = 150

    New RI still remain 50150

    Values are completely comparable

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    20/56

    Transference of RI

    y = 1.57x0.832, r2 = 0.990 Assume the RI for the current method is 50150

    when x = 50, y = 1.57 (50)0.832 = 78

    When x = 150, y = 1.57 (150)0.832 = 235

    New RI transforms to 78 - 235

    Highly correlated but one showed higher or lowerbias

    result

    Strongly encourages to validate the RI with small

    sample ( n=20 ) of reference individual

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    21/56

    Transference of RI

    same or comparable analytical system

    from another lab or manufacturer package

    insert

    Different population 3 validation method proposed:

    (1) Subjective assessment

    (2) Validation with small number of reference

    individuals ( eg. N = 20 )

    (3) Evaluation with large number of reference

    individuals ( N=60)

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    22/56

    1. Subjective Assessment

    Transfer of RI without validation study

    carefully inspection of all the pertinent

    factors of original appropriate reference

    value study

    reference population demographic/ geographic,pre-analytical & analytical procedure, method of

    determine RI

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    23/56

    2.Validation with Small No. of Reference Individuals ( n = 20 )

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    24/56

    3. Evaluate with Large No. of Reference

    Individuals ( n = 60 )

    Methodology similar to validation with

    n= 20Advantage :

    increasing power for observing statistical

    different between two sets reference specimens

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    25/56

    CLSI H54-A:

    Procedure for Validation ofINR and Local Calibrationof PT/INR Systems

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    26/56

    PT INR System Verification

    PT INR Mathematic conversion of PT result based on

    the responsiveness of the thromboplastine

    reagent used in the assay

    Monitoring of Oral Anticoagulant therapies

    INR = Patient PT ( seconds )

    GMNPT ( Seconds )

    ISI

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    27/56

    PT INR System Verification

    ISI

    Quantitative measure/ mathematic

    indicator of the responsiveness of PT

    system to the defect induced by VKA

    therapy compare to primary WHO

    international Reference Preparation (IRP )

    ISI , responsive to VKA induced defect

    Thromboplastine / instrument specific ISI

    Generic ISI

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    28/56

    PT INR System Verification

    IF reagent / instrument specific ISIis

    used

    ISI verification should be perform

    IF ISI verification fail Local ISI calibration

    IF generic ISIis used

    ISI verification is MANDATORY Local ISI calibration is strongly recommended

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    29/56

    PT INR System Verification Change of reagent & reagent lot number

    Change of instrument, following major repair

    Unexplained major discrepancy of EQA result

    After PT/INR calibration

    At least once per year if not major change Mandatory if generic ISI is used

    When

    Minimum 3 INR certified plasma(INR 1.54.5 ) & 1 NPP

    Commercial plasma calibration set( calibrated against WHO standard )

    Thromboplastine reagent

    What

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    30/56

    PT INR System Verification

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    31/56

    PT INR System Verification

    Reconstitute set of calibration

    plasmas

    ( 4 vialsA : INR 1.0 ; B: INR 2.0C: INR 3.0 ; D: INR 4.0 )

    Aliquot each vial of calibration plasma into 3 microtubes

    for day to day variation assessment

    A1 A2 A3 B1 B2 B3 C1 C2 C3 D1 D2 D3

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    32/56

    A1 A2 A3 B1 B2 B3 C1 C2 C3 D1 D2 D3

    DAY 3

    Frozen at -70 CPerform PT assay

    (Duplicate)

    DAY 2

    Frozen at -70 CPerform PT assay

    (Duplicate)

    DAY 1Perform PT assay

    (Duplicate)

    Calculate the Mean of PT seconds ( 3 days duplicate ) for

    each calibration level ( A, B, C & D)Calculate the INR of each calibration plasma

    ( reagent /instrument specific ISI & local GMNPT )

    Compare INR obtained with INR value from product insert

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    33/56

    Compare INR obtained with INR value from product

    insert

    Different 15 %

    ( one level of calibration

    plasma )

    FAILLocal PT/INR calibrationDifferent certified plasma used

    for verification

    Preferably from different

    manufacturer

    if from same manufacturer,

    different

    lot must be use

    Verification of local ISI

    ( obtained from calibration graph

    )

    Test system with

    different thromboplastin

    local PT/INR system

    calibration with diff set of

    CP

    Contact manufacturer

    for assistance

    FAIL

    AGAIN

    Different < 15 %

    ( each calibration plasma )

    Manufacturer assigned ISI

    is valid for local PT INR

    reporting

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    34/56

    ExampleThromboplastin Certified Calibration

    Plasma

    Name STA Neoplastine Cl Plus AK-Calibrant

    Lot Number 105513 0U03B00

    Expiry Date 2011-10 2011-12-31

    ISI 1.33 -

    AK

    Calibrant

    Day 1 Day 2 Day 3 Mean

    A 13.1 13.3 13.5 13.7 13.3 13.0 13.3

    B 23.5 24.0 25.1 24.6 24.0 23.7 24.2

    C 32.4 33.4 34.4 34.4 34.1 35.1 34.0

    D 45.9 46.7 50.0 49.9 46.7 46.5 47.6

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    35/56

    AK Calibrant INR target

    value

    ( from flyer )

    INR acceptable

    range

    ( 15 % from

    target value )

    Mean PT

    ( seconds )

    Local

    generated INR

    A 1.03 0.881.18 13.3 1.06

    B 2.29 1.952.63 24.2 2.35

    C 3.43 2.92

    3.94 34.0 3.71

    D 5.31 4.516.11 47.6 5.80

    ISI = 1.33

    GMNPT = 12.7 seconds Manufacturer assigned ISI is

    valid for local PT INR

    reporting

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    36/56

    IF ISI Verification FAIL

    Local ISI Calibration is

    MANDATORY!!

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    37/56

    Local ISI Calibration

    Reference method:

    20 normal plasma + 60 stabilized warfarinised

    samples

    Assay for PT

    Plot Log PT ( y-axis ) versus Log assigned INR ( x-

    axis )

    Determine local ISI ( 1/Slope )

    Only if manufacturer do

    not provide local ISI

    calculation service

    i d

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    38/56

    Assigned INR

    ( From flyer )

    Mean PT

    ( from 3 days

    replicate )

    1.03 13.3

    2.29 24.23.43 34.0

    5.31 5.31

    Log INR Log PT

    0.013 1.124

    0.360 1.384

    0.535 1.531

    0.725 1.678

    y = 0.779x + 1.110R = 0.999

    0.600

    0.800

    1.000

    1.200

    1.400

    1.600

    1.800

    0.000 0.100 0.200 0.300 0.400 0.500 0.600 0.700 0.800

    Y = 0.4

    Local MNPT = y intercept at X = 0

    = 1.110

    = anti-log ( 1.110 )

    = 12.9 seconds

    INR cal plasma Method:

    1. measure PT

    2. calculate log10 of PT and INR

    3. Plot the values

    4. Draw line of best fit

    5. Calculate Slope & ISI

    Slope =change in Y/change in X

    = 0.4/0.51

    = 0.784 ( 0.779 frm formula equation)

    Local ISI = 1 / Slope

    = 1 / 0.784 ( 1/0.779 )

    = 1.275 ( 1.28 )

    Log PT

    Example:

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    39/56

    Local MNPT &

    ISI ( frommanufacturer )

    Mean PT from

    3 daysreplicate

    PT INR

    (reagent/

    instrumentspecific ) from

    flyer

    No Manual

    calculation of

    local ISI & MNPT

    Example:

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    40/56

    After Local ISI Calibration..

    Verification should be carry out on local

    calculated ISI

    IF verification pass

    Local calculated ISI is valid for PT INRreporting

    IF verification FAIL again

    Try with different thromboplastine ( reagent )

    Repeat the ISI calibration with another set of

    CP

    Call manufacturer for assisstance

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    41/56

    Factor Assay:

    Constructing CalibrationCurve for accurate result

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    42/56

    Factor Assay: Test Principle

    Factor activities of plasma sample is

    measured by its ability to correct the

    prolonged clotting time of factor deficient

    plasma

    The APTTs or PTs mixtures of diluted test

    plasma and factor deficient plasma areinversely proportional to the concentration

    of the factor in the test plasma mixtures

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    43/56

    Factor Assay Principle

    Diluted sample 100l (calibrator, control or patient )

    Factor deficient plasma 100l

    Test reagent (PT 200l or APTT 100l and CaCl2 100l)

    Measure Clotting of PT or APTT

    Wh d d ?

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    44/56

    What do you need ?

    PT or APTT reagents

    - PT ( Extrinsic : FVII )

    - APTT ( Intrinsic : FVIII, FIX,FXI,

    FXII )

    Factor Deficient Plasma

    - < 1 % of particular coagulation

    factor- >50 % of all other coagulationfactors

    - from individual with severecongenital factor deficiency ( factor

    inhibitor & LA free )- May also prepared by in vitroimmunodepletion

    Reference Plasma

    - At least 20 healthy donors

    - Should have 100 20 % coagulantactivities for all coagulation factors

    - Calibrated against WHO standardor a secondary standard

    - Commercial

    Quality Control

    - Normal control plasma- Abnormal control plasma

    - Should be assayed witheach batch of test

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    45/56

    Preparation of Calibration curve

    Two types of reference curve:

    1. Extended curve To establish the range of linearity for the labs specific

    analytical system

    Change of reagent lot, reference plasma or instrument

    2. Working curve to validate the method of that run

    Used as analytic reference against which test sample aremeasured

    Each run of assay

    Fewer dilutions

    ( limited within linear portion of extended curve )

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    46/56

    Extended Standard Curve

    Reconstitute Reference Plasma ( follow product insert

    )

    Prepare dilution of reference plasma

    Test each dilution standard for PT or

    APTT

    Add Factor deficient plasma to each diluted standard ( 1 :

    1 )

    At least 7 dilution ( 1:5 , 1:10, 1:20, 1:40, 1:80, 1:160, 1:320 )

    Plot PT or APTT results on Y-axis against dilution on X-

    axislo -lo a er

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    47/56

    Reference Plasma Dilution

    Linear range

    ( 1:5 to 1:160 )

    100 %50 %25 %12.5%6.25

    %

    3.13

    %

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    48/56

    Working Standard CurveReconstitute Reference Plasma

    (Same lot as used in extended reference curve )

    Prepare dilution of reference plasma

    Test each dilution standard for PT or

    APTT

    Add Factor deficient plasma to each diluted standard ( 1 :

    1 )

    At least 3 dilutions which fall within the linear range

    Plot PT or APTT results on Y-axis against Factor activity

    on X-axis ( log-log paper )

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    49/56

    79.0

    70.3

    63.5

    57.4

    13 25 50 100

    Concentration %

    Clotting Time, sec

    Log( c) = ( -6.531* Log(t) ) + 13.478

    r = - 0.999

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    50/56

    79.0

    70.3

    63.5

    57.4

    13 25 50 100

    Concentration %

    Clotting Time, sec

    Log( c) = ( -6.531* Log(t) ) + 13.478

    r = - 0.999

    Reference Curve: Validation Criteria

    R > 0.99

    Interpolate Vs Standardvalues < 10 % different

    Within control range

    Duplicate, 10%

    different tolerance

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    51/56

    Raw Data Target Range

    ** Based on 100% unicalibrator value

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    52/56

    How to look at raw data ?

    1.000

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    53/56

    How to look at raw data ?

    Good curve

    Bad curve ( short raw data)Bad curve ( long raw data)

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    54/56

    Optimization of Standard Curve

    Calibrator & reagent preparation

    Fresh dH2O or diluents

    Stabilize time as recommended by

    manufacturer

    Do not use beyond stability calibrator or

    reagent

    Automation

    Perfect washing system

    Syringe is free from bubble

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    55/56

    Optimization of Standard Curve

    Manual

    standard plasma dilution preparation

    Plotting standard curve:

    reference plasma values

    data points plotting

  • 8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97

    56/56

    Thank You !