quality assurance activities in coagulation_ippt_97
TRANSCRIPT
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
1/56
Quality Activities inCoagulation Laboratory
Kiu Ling Ling
Application Manager
All Eights (M) SDN.BHD.
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
2/56
Outline
1. Establish Control Range with commercial QC
material
2. Transference & Validation of reference interval
3. Calibration of PT-INR system with certified
plasma
4. Factor assay: constructing calibration curve
for accurate result
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
3/56
Establish Control Range
with Commercial QCMaterial
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
4/56
What
Why
When
How
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
5/56
WHATis Internal Quality Control ?
Set of procedures undertaken by laboratory
staffs for the continuous monitoring of operationsand the results of the measurement in order to
decide whether the results are reliable enough to
be released.
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
6/56
IQC
To monitor the
accuracy &precision of the
completeanalyticalprocess
To detectimmediateanalytical
errorsDay to Day
monitoring oftestperformance
quality
( long term
constancy )
WHYwe need IQC ?
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
7/56
Control Material- Important
CharacteristicsSimilar to test specimens ( Matrix )
Can be liquid or freeze-dried ( lyophilized )
known concentration of analyte 2 or 3 levels of control ( normal & abnormal )
Values cover medical decision points
Stability up to at least 8 hoursQC material must be tested in the same
manners as patient sample
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
8/56
To Establish Control Range
Collecting Control data
Perform control repeatedly (minimum 20times) with testing spread over at least 10section
Calculate Mean , Standard Deviation (SD)
Determine control limit ( Mean 2SD )
Plotting Levey-Jennings chart
- Manually on graph paper
- MS Excel Software
- Built in quality software in coagulationanalyzer
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
9/56
How to Create Levey-Jennings Graph
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
10/56
APTT normal control data
1. 23.0 sec
2. 22.0 sec
3. 22.5 sec
4. 23.2 sec
5. 22.2 sec6. 22.5 sec
7. 22.8 sec
8. 23.3 sec
9. 22.5 sec10. 23.1 sec
11. 22.4 sec
12. 22.6 sec
13. 23.0 sec
14. 23.0 sec
15. 22.8 sec
16. 22.5 sec
17. 23.0 sec
18. 23.0 sec
19. 22.9 sec20. 22.5 sec
Step 1
Calculate MEAN
Mean = Sum ( X1 + X2 +.X20)
Total Number of Data
(N)
MEAN = 454.8 / 20
= 22.7 sec
Step 2
Calculate SD
SD = 0.4 secSum = 454.8 sec
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
11/56
Step 3
Set control limit In Gaussian Distribution :
68.3% of data is within 1.0 SD
95.5% of data is within 2.0
SD
99.7% of data is within 3.0 SD
Control limit for APTT
= Mean 2SD
= 21.9 23.5 sec
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
12/56
Step 4: Plotting LJ graph
MEAN
+1SD
+2SD
-1SD
-2SD
-3SD
+3SD
22.7
23.1
23.5
23.9
22.3
21.9
21.5
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
13/56
CLSI C28-A3:Define, Establishing, and
Verifying ReferenceIntervals in the Clinical
Laboratory
3rd
Edition
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
14/56
Reference Interval Minimum 120 healthy donors ( CLSI 28-A3)
Criteria of donor selection
equal number of sex
preferred 1650 years old no under any anticoagulant therapy and
medication
collect sample over few daysday-to-day variation
Normal Range = Mean 2SD
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
15/56
Reference Interval Establishment
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
16/56
Local PT/INR System Calibration
Norma
lRange:Exam
ple
ID PT ( sec) APTT (sec) Fib (g/l) TT (Sec)
1820425 13.4 39.7 3.7 15.8
1712056 13.2 41.0 3.0 15.0
1712084 13.8 41.9 2.7 15.9
1820420 13.5 40.5 3.1 15.9
1820421 12.8 33.2 3.0 16.2
1712088 13.4 37.6 3.6 17.2
1820422 13.0 35.6 3.4 16.3
1820369 13.5 37.9 3.1 16.5
1712087 13.5 41.7 3.1 16.6
1712086 13.0 39.1 3.0 16.6
1712085 13.1 33.4 3.3 16.1
1822001 13.0 41.2 3.0 17.5
D14 12.8 37.8 2.9 15.7
D15 13.5 35.6 3.2 16.0
D16 13.1 38.2 2.6 15.8
D17 12.8 40.8 3.3 17.4
D18 13.6 41.0 3.4 15.9
D19 13.0 36.0 2.4 16.4
D20 12.6 31.7 2.3 17.0
D21 13.6 35.6 2.7 15.8
D22 13.0 33.7 2.7 17.4
D23 13.4 33.5 2.3 17.2
D24 13.2 36.9 2.6 17.0
D25 13.3 33.8 3.0 17.3
D26 12.7 35.0 3.5 15.4
Mean 13.7 37.2 3.0 16.4
SD 0.4 3.1 0.4 0.7
Normal Range
Mean2SD 12.513.9 sec 30.943.4 sec 2.23.7g/l 15.017.8 sec
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
17/56
Challenge in Determination of
Reference IntervalTo determine reliability of reference
interval (RI) costly & heavy task
minimum 120 healthy subjects needed
Difficulty in collecting sufficient number ofqualified reference individuals
Transfer of reference range from manufactureror other laboratory
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
18/56
Transference of RI
If Laboratory has previously established a
RI for its own population
simple comparison between two analytical system
( new & old ) could be sufficient
Advantages
no need to collect sample from referenceindividuals
can use existing patients samples
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
19/56
Transference of RI
If completely comparable ( Excellentcorrelation coefficient , small slope bias, smallintercept, comparable range of value )
y = 1.004x0.628, r2 = 0.990 Assume the RI for the current method is 50150
when x = 50, y = 1.004 (50)0.628 = 50
When x = 150, y = 1.004 (150)0.628 = 150
New RI still remain 50150
Values are completely comparable
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
20/56
Transference of RI
y = 1.57x0.832, r2 = 0.990 Assume the RI for the current method is 50150
when x = 50, y = 1.57 (50)0.832 = 78
When x = 150, y = 1.57 (150)0.832 = 235
New RI transforms to 78 - 235
Highly correlated but one showed higher or lowerbias
result
Strongly encourages to validate the RI with small
sample ( n=20 ) of reference individual
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
21/56
Transference of RI
same or comparable analytical system
from another lab or manufacturer package
insert
Different population 3 validation method proposed:
(1) Subjective assessment
(2) Validation with small number of reference
individuals ( eg. N = 20 )
(3) Evaluation with large number of reference
individuals ( N=60)
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
22/56
1. Subjective Assessment
Transfer of RI without validation study
carefully inspection of all the pertinent
factors of original appropriate reference
value study
reference population demographic/ geographic,pre-analytical & analytical procedure, method of
determine RI
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
23/56
2.Validation with Small No. of Reference Individuals ( n = 20 )
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
24/56
3. Evaluate with Large No. of Reference
Individuals ( n = 60 )
Methodology similar to validation with
n= 20Advantage :
increasing power for observing statistical
different between two sets reference specimens
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
25/56
CLSI H54-A:
Procedure for Validation ofINR and Local Calibrationof PT/INR Systems
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
26/56
PT INR System Verification
PT INR Mathematic conversion of PT result based on
the responsiveness of the thromboplastine
reagent used in the assay
Monitoring of Oral Anticoagulant therapies
INR = Patient PT ( seconds )
GMNPT ( Seconds )
ISI
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
27/56
PT INR System Verification
ISI
Quantitative measure/ mathematic
indicator of the responsiveness of PT
system to the defect induced by VKA
therapy compare to primary WHO
international Reference Preparation (IRP )
ISI , responsive to VKA induced defect
Thromboplastine / instrument specific ISI
Generic ISI
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
28/56
PT INR System Verification
IF reagent / instrument specific ISIis
used
ISI verification should be perform
IF ISI verification fail Local ISI calibration
IF generic ISIis used
ISI verification is MANDATORY Local ISI calibration is strongly recommended
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
29/56
PT INR System Verification Change of reagent & reagent lot number
Change of instrument, following major repair
Unexplained major discrepancy of EQA result
After PT/INR calibration
At least once per year if not major change Mandatory if generic ISI is used
When
Minimum 3 INR certified plasma(INR 1.54.5 ) & 1 NPP
Commercial plasma calibration set( calibrated against WHO standard )
Thromboplastine reagent
What
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
30/56
PT INR System Verification
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
31/56
PT INR System Verification
Reconstitute set of calibration
plasmas
( 4 vialsA : INR 1.0 ; B: INR 2.0C: INR 3.0 ; D: INR 4.0 )
Aliquot each vial of calibration plasma into 3 microtubes
for day to day variation assessment
A1 A2 A3 B1 B2 B3 C1 C2 C3 D1 D2 D3
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
32/56
A1 A2 A3 B1 B2 B3 C1 C2 C3 D1 D2 D3
DAY 3
Frozen at -70 CPerform PT assay
(Duplicate)
DAY 2
Frozen at -70 CPerform PT assay
(Duplicate)
DAY 1Perform PT assay
(Duplicate)
Calculate the Mean of PT seconds ( 3 days duplicate ) for
each calibration level ( A, B, C & D)Calculate the INR of each calibration plasma
( reagent /instrument specific ISI & local GMNPT )
Compare INR obtained with INR value from product insert
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
33/56
Compare INR obtained with INR value from product
insert
Different 15 %
( one level of calibration
plasma )
FAILLocal PT/INR calibrationDifferent certified plasma used
for verification
Preferably from different
manufacturer
if from same manufacturer,
different
lot must be use
Verification of local ISI
( obtained from calibration graph
)
Test system with
different thromboplastin
local PT/INR system
calibration with diff set of
CP
Contact manufacturer
for assistance
FAIL
AGAIN
Different < 15 %
( each calibration plasma )
Manufacturer assigned ISI
is valid for local PT INR
reporting
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
34/56
ExampleThromboplastin Certified Calibration
Plasma
Name STA Neoplastine Cl Plus AK-Calibrant
Lot Number 105513 0U03B00
Expiry Date 2011-10 2011-12-31
ISI 1.33 -
AK
Calibrant
Day 1 Day 2 Day 3 Mean
A 13.1 13.3 13.5 13.7 13.3 13.0 13.3
B 23.5 24.0 25.1 24.6 24.0 23.7 24.2
C 32.4 33.4 34.4 34.4 34.1 35.1 34.0
D 45.9 46.7 50.0 49.9 46.7 46.5 47.6
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
35/56
AK Calibrant INR target
value
( from flyer )
INR acceptable
range
( 15 % from
target value )
Mean PT
( seconds )
Local
generated INR
A 1.03 0.881.18 13.3 1.06
B 2.29 1.952.63 24.2 2.35
C 3.43 2.92
3.94 34.0 3.71
D 5.31 4.516.11 47.6 5.80
ISI = 1.33
GMNPT = 12.7 seconds Manufacturer assigned ISI is
valid for local PT INR
reporting
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
36/56
IF ISI Verification FAIL
Local ISI Calibration is
MANDATORY!!
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
37/56
Local ISI Calibration
Reference method:
20 normal plasma + 60 stabilized warfarinised
samples
Assay for PT
Plot Log PT ( y-axis ) versus Log assigned INR ( x-
axis )
Determine local ISI ( 1/Slope )
Only if manufacturer do
not provide local ISI
calculation service
i d
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
38/56
Assigned INR
( From flyer )
Mean PT
( from 3 days
replicate )
1.03 13.3
2.29 24.23.43 34.0
5.31 5.31
Log INR Log PT
0.013 1.124
0.360 1.384
0.535 1.531
0.725 1.678
y = 0.779x + 1.110R = 0.999
0.600
0.800
1.000
1.200
1.400
1.600
1.800
0.000 0.100 0.200 0.300 0.400 0.500 0.600 0.700 0.800
Y = 0.4
Local MNPT = y intercept at X = 0
= 1.110
= anti-log ( 1.110 )
= 12.9 seconds
INR cal plasma Method:
1. measure PT
2. calculate log10 of PT and INR
3. Plot the values
4. Draw line of best fit
5. Calculate Slope & ISI
Slope =change in Y/change in X
= 0.4/0.51
= 0.784 ( 0.779 frm formula equation)
Local ISI = 1 / Slope
= 1 / 0.784 ( 1/0.779 )
= 1.275 ( 1.28 )
Log PT
Example:
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
39/56
Local MNPT &
ISI ( frommanufacturer )
Mean PT from
3 daysreplicate
PT INR
(reagent/
instrumentspecific ) from
flyer
No Manual
calculation of
local ISI & MNPT
Example:
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
40/56
After Local ISI Calibration..
Verification should be carry out on local
calculated ISI
IF verification pass
Local calculated ISI is valid for PT INRreporting
IF verification FAIL again
Try with different thromboplastine ( reagent )
Repeat the ISI calibration with another set of
CP
Call manufacturer for assisstance
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
41/56
Factor Assay:
Constructing CalibrationCurve for accurate result
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
42/56
Factor Assay: Test Principle
Factor activities of plasma sample is
measured by its ability to correct the
prolonged clotting time of factor deficient
plasma
The APTTs or PTs mixtures of diluted test
plasma and factor deficient plasma areinversely proportional to the concentration
of the factor in the test plasma mixtures
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
43/56
Factor Assay Principle
Diluted sample 100l (calibrator, control or patient )
Factor deficient plasma 100l
Test reagent (PT 200l or APTT 100l and CaCl2 100l)
Measure Clotting of PT or APTT
Wh d d ?
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
44/56
What do you need ?
PT or APTT reagents
- PT ( Extrinsic : FVII )
- APTT ( Intrinsic : FVIII, FIX,FXI,
FXII )
Factor Deficient Plasma
- < 1 % of particular coagulation
factor- >50 % of all other coagulationfactors
- from individual with severecongenital factor deficiency ( factor
inhibitor & LA free )- May also prepared by in vitroimmunodepletion
Reference Plasma
- At least 20 healthy donors
- Should have 100 20 % coagulantactivities for all coagulation factors
- Calibrated against WHO standardor a secondary standard
- Commercial
Quality Control
- Normal control plasma- Abnormal control plasma
- Should be assayed witheach batch of test
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
45/56
Preparation of Calibration curve
Two types of reference curve:
1. Extended curve To establish the range of linearity for the labs specific
analytical system
Change of reagent lot, reference plasma or instrument
2. Working curve to validate the method of that run
Used as analytic reference against which test sample aremeasured
Each run of assay
Fewer dilutions
( limited within linear portion of extended curve )
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
46/56
Extended Standard Curve
Reconstitute Reference Plasma ( follow product insert
)
Prepare dilution of reference plasma
Test each dilution standard for PT or
APTT
Add Factor deficient plasma to each diluted standard ( 1 :
1 )
At least 7 dilution ( 1:5 , 1:10, 1:20, 1:40, 1:80, 1:160, 1:320 )
Plot PT or APTT results on Y-axis against dilution on X-
axislo -lo a er
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
47/56
Reference Plasma Dilution
Linear range
( 1:5 to 1:160 )
100 %50 %25 %12.5%6.25
%
3.13
%
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
48/56
Working Standard CurveReconstitute Reference Plasma
(Same lot as used in extended reference curve )
Prepare dilution of reference plasma
Test each dilution standard for PT or
APTT
Add Factor deficient plasma to each diluted standard ( 1 :
1 )
At least 3 dilutions which fall within the linear range
Plot PT or APTT results on Y-axis against Factor activity
on X-axis ( log-log paper )
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
49/56
79.0
70.3
63.5
57.4
13 25 50 100
Concentration %
Clotting Time, sec
Log( c) = ( -6.531* Log(t) ) + 13.478
r = - 0.999
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
50/56
79.0
70.3
63.5
57.4
13 25 50 100
Concentration %
Clotting Time, sec
Log( c) = ( -6.531* Log(t) ) + 13.478
r = - 0.999
Reference Curve: Validation Criteria
R > 0.99
Interpolate Vs Standardvalues < 10 % different
Within control range
Duplicate, 10%
different tolerance
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
51/56
Raw Data Target Range
** Based on 100% unicalibrator value
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
52/56
How to look at raw data ?
1.000
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
53/56
How to look at raw data ?
Good curve
Bad curve ( short raw data)Bad curve ( long raw data)
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
54/56
Optimization of Standard Curve
Calibrator & reagent preparation
Fresh dH2O or diluents
Stabilize time as recommended by
manufacturer
Do not use beyond stability calibrator or
reagent
Automation
Perfect washing system
Syringe is free from bubble
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
55/56
Optimization of Standard Curve
Manual
standard plasma dilution preparation
Plotting standard curve:
reference plasma values
data points plotting
-
8/12/2019 Quality Assurance Activities in Coagulation_IPPT_97
56/56
Thank You !