QUALITY ASSURANCE PROCEDURE FOR

SRS AND SRT TREATMENTSMEIJERS A., PLAUDE S., UTEHINA O., POPOVS S.

RIGA EAST UNIVERSITY HOSPITAL

Thursday, October 13, 11

SRS / SRT TREATMENTS

INTRODUCED IN 2009 AT RIGA EAST UNIVERSITY HOSPITAL.

EQUIPMENT:

NOVALIS TX (VARIAN), EXACTRAC (BRAINLAB)

VARIOUS LOCALIZATIONS:

HEAD, LIVER ETC.

Thursday, October 13, 11

CHALLENGES

FRACTIONATED TREATMENT

1 - 5 GY PER FRACTION10 - 39 FRACTIONS

Thursday, October 13, 11

CHALLENGES

FRACTIONATED TREATMENT

Thursday, October 13, 11

CHALLENGES

FRACTIONATED TREATMENT

Thursday, October 13, 11

CHALLENGES

FRACTIONATED TREATMENT

Thursday, October 13, 11

CHALLENGES

FRACTIONATED TREATMENT

MISTREATMENT OF A SINGLE FRACTION DOES NOT HAVE SIGNIFICANT IMPACT ON DELIVERED DOSE

Thursday, October 13, 11

CHALLENGES

RADIO-SURGERY

4 - 20 GY PER FRACTION1 - 3 FRACTIONS

Thursday, October 13, 11

CHALLENGES

RADIO-SURGERY

Thursday, October 13, 11

CHALLENGES

RADIO-SURGERY

MISTREATMENT IN A SINGLE FRACTION HAS SIGNIFICANT IMPACT ON ALL THE TREATMENT

Thursday, October 13, 11

TREATMENT ACCURACY

SRSSRT

FRACTIONATEDTHERAPY

SRS / SRT REQUIRES HIGHER QUALITY IN TERMS OF TREATMENT DELIVERY

Thursday, October 13, 11

TREATMENT ACCURACY

SRS / SRT REQUIRES EXTENDEDQUALITY ASSURANCE PROCEDURE

Thursday, October 13, 11

CAUSES OF INACCURACY

PATIENT POSITIONING

MISCONFIGURATION OF TPS

SOFTWARE BUGS

MECHANICAL FAULTS OF TREATMENT DELIVERY SYSTEM

MIS-CALIBRATION OF LINAC

IONIZATION CHAMBERS

MECHANICAL INDICATORS

Thursday, October 13, 11

POSITION UNCERTAINTIES

EXACTRAC (BRAINLAB)

Thursday, October 13, 11

QUALITY ASSURANCE

HARDWARERELATED

PATIENTRELATED

Thursday, October 13, 11

QUALITY ASSURANCE

HARDWARERELATED

PATIENTRELATED

Thursday, October 13, 11

PATIENT-RELATED QA FOR SRS/SRT

DOSE PLANE MEASUREMENT

ABSOLUTE POINT-DOSE MEASUREMENT

WINSTON-LUTZ TEST

Thursday, October 13, 11

DOSE PLANE MEASUREMENT

1. CALCULATION OF TREATMENT PLAN ON QA PHANTOM

2. MEASUREMENT WITH RADIOCHROMIC FILM

3. COMPARISON OF MEASURED DOSE VS CALCULATED

Thursday, October 13, 11

FILM DOSIMETRY

GAFCHROMIC EBT 2 (ISP) FILMS

DOSE RANGE UP TO 40 GY

CALIBRATION: TEST FILM WITH KNOWN DOSE

METHOD: COMPOSITE ANALYSIS

0

7.5

15.0

22.5

30.0

1 10 100 1000

Dose

, Gy

Tone Index

Thursday, October 13, 11

FILM DOSIMETRY

ANALYSIS:

GAMMA INDEX - 1MM / 5% / SUPPRESS < 5%

$GPLQLVWUDWLYH�'DWD,QVWLWXWLRQ 5$.86�/2&�750).3K\VLFLVW 6DQGLMD�3ODXGH3DWLHQW,' ������������3DWLHQW�1DPH 7HUHVNRYLFD��1LQD&RPPHQW SDNDXVLVBK)57

'&25�BBBBBBBBBB

*DPPD��'���3DUDPHWHUV��� PP�'LVWDQFH��7R��$JUHHPHQW��� ��'RVH�'LIIHUHQFH�ZLWK�UHI��WR 0D[��GRVH�RI�PHDVXUHG�GDWD�VHW8VH�LQFUHDVHG�WROHUDQFH�RI ��� ��'RVH�'LII��IRU�YDOXHV�EHORZ ��� *\��RU�$8�6XSSUHVV�GRVHV�EHORZ ��� ��RI�PD[��GRVH�RI�PHDVXUHG�GDWD�VHW

6WDWLVWLFV1XPEHU�RI�'RVH�3RLQWV ����(YDOXDWHG�'RVH�3RLQWV ���� �� ���� ���3DVVHG ���� �� ���� ���)DLOHG �� �� ��� ���5HVXOW ���� �����*UHHQ�

*DPPD��'$ULWKPHWLF�0HDQ �����0LQ �� /5 � ����� PP�� 7* � ���� PP� �����0D[ �� /5 � ��� PP�� 7* � ���� PP� �����0HGLDQ �����

$EVROXWH�'LIIHUHQFH$ULWKPHWLF�0HDQ ����� *\0LQ �� /5 � ����� PP�� 7* � ���� PP� ����� *\0D[ �� /5 � ���� PP�� 7* � ����� PP� ����� *\0HGLDQ ����� *\

6HWWLQJV3DVVLQJ�FULWHULD *DPPD������*UHHQ ���� ��� WR �������<HOORZ ���� ��� WR ���� ��5HG ������ WR ���� ��

,�KHUE\�FRQILUP�WKDW�,�KDYH�FKHFNHG�DOO�GDWD�PHQWLRQHG�DERYH�DQG�WKDW�WKH�UDGLRORJLF�WUHDWPHQW�DFFRUGLQJ�WKLV�GDWD�VKDOO���VKDOO�QRWSURFHHG�

'DWH 1DPH�DQG�)XQFWLRQ 6LJQDWXUH

37:���9HUL6RIW���� 3DWLHQW,'�������������� 3DWLHQW�1DPH��7HUHVNRYLFD��1LQD 'DWH��������������������� �

$GPLQLVWUDWLYH�'DWD,QVWLWXWLRQ 5$.86�/2&�750).3K\VLFLVW 6DQGLMD�3ODXGH3DWLHQW,' ������������3DWLHQW�1DPH 7HUHVNRYLFD��1LQD&RPPHQW SDNDXVLVBK)57

'&25�BBBBBBBBBB

*DPPD��'���3DUDPHWHUV��� PP�'LVWDQFH��7R��$JUHHPHQW��� ��'RVH�'LIIHUHQFH�ZLWK�UHI��WR 0D[��GRVH�RI�PHDVXUHG�GDWD�VHW8VH�LQFUHDVHG�WROHUDQFH�RI ��� ��'RVH�'LII��IRU�YDOXHV�EHORZ ��� *\��RU�$8�6XSSUHVV�GRVHV�EHORZ ��� ��RI�PD[��GRVH�RI�PHDVXUHG�GDWD�VHW

6WDWLVWLFV1XPEHU�RI�'RVH�3RLQWV ����(YDOXDWHG�'RVH�3RLQWV ���� �� ���� ���3DVVHG ���� �� ���� ���)DLOHG �� �� ��� ���5HVXOW ���� �����*UHHQ�

*DPPD��'$ULWKPHWLF�0HDQ �����0LQ �� /5 � ����� PP�� 7* � ���� PP� �����0D[ �� /5 � ��� PP�� 7* � ���� PP� �����0HGLDQ �����

$EVROXWH�'LIIHUHQFH$ULWKPHWLF�0HDQ ����� *\0LQ �� /5 � ����� PP�� 7* � ���� PP� ����� *\0D[ �� /5 � ���� PP�� 7* � ����� PP� ����� *\0HGLDQ ����� *\

6HWWLQJV3DVVLQJ�FULWHULD *DPPD������*UHHQ ���� ��� WR �������<HOORZ ���� ��� WR ���� ��5HG ������ WR ���� ��

,�KHUE\�FRQILUP�WKDW�,�KDYH�FKHFNHG�DOO�GDWD�PHQWLRQHG�DERYH�DQG�WKDW�WKH�UDGLRORJLF�WUHDWPHQW�DFFRUGLQJ�WKLV�GDWD�VKDOO���VKDOO�QRWSURFHHG�

'DWH 1DPH�DQG�)XQFWLRQ 6LJQDWXUH

37:���9HUL6RIW���� 3DWLHQW,'�������������� 3DWLHQW�1DPH��7HUHVNRYLFD��1LQD 'DWH��������������������� �

ACCEPTABLE:GAMMA PASS > 90%

UNACCEPTABLE:GAMMA PASS <85%

MEAN 96.4 ± 2.6 %

Thursday, October 13, 11

PATIENT-RELATED QA FOR SRS/SRT

DOSE PLANE MEASUREMENT

ABSOLUTE POINT-DOSE MEASUREMENT

WINSTON-LUTZ TEST

Thursday, October 13, 11

POINT DOSE MEASUREMENT

1. CALCULATION OF TREATMENT PLAN ON QA PHANTOM

2. MEASUREMENT WITH IONIZATION CHAMBER

3. COMPARISON OF MEASURED DOSE VS CALCULATED

Thursday, October 13, 11

POINT DOSE MEASUREMENT

HAND-MADE PHANTOMPINPOINT CHAMBER (PTW)

0.016 CM3

ANALYSIS:LOCAL DIFFERENCE IN %

ACCEPTABLE < 3%UNACCEPTABLE > 5%

MEAN -1.2 ± 1.6 %

Thursday, October 13, 11

PATIENT-RELATED QA FOR SRS/SRT

DOSE PLANE MEASUREMENT

ABSOLUTE POINT-DOSE MEASUREMENT

WINSTON-LUTZ TEST

Thursday, October 13, 11

WINSTON-LUTZ TEST

1. ALIGN POINTER TO ISOCENTRE

2. TAKE PORTAL IMAGES FROM NECESSARY ANGLES

3. DETERMINE CENTRE OF FIELD

4. DETERMINE CENTRE OF POINTER SPHERE

5. DETERMINE THE DISTANCE BETWEEN CENTERS

Thursday, October 13, 11

WINSTON-LUTZ TEST

ACCEPTABLE < 1MMUNACCEPTABLE > 1MM

0.50

0.75

1.00

Sphe

re s

ize, m

m

Thursday, October 13, 11

PATIENT-RELATED QA FOR SRS/SRT

DOSE PLANE MEASUREMENT

ABSOLUTE POINT-DOSE MEASUREMENT

WINSTON-LUTZ TEST

Thursday, October 13, 11

CONCLUSIONS

THE DEVELOPED PATIENT RELATED QA PROCEDURE IS SUFFICIENT TO VERIFY ACCURACY OF TREATMENT PLANNING SYSTEM CALCULATIONS, PRECISION OF TREATMENT DELIVERY MECHANICS AND SYSTEM IN GENERAL.

ADDITIONAL MACHINE RELATED VERIFICATIONS ARE NECESSARY.

QA PROGRAM, BASED ON DESCRIBED PRINCIPLES, HAS BEEN IMPLEMENTED IN RIGA EAST UNIVERSITY HOSPITAL.