ram 744 - gfg instrumentationgoodforgas.com/wp-content/uploads/2013/12/ram_744... · the ram 744 is...
TRANSCRIPT
• Continuousin-lineCOmonitorwithlowmaintenance,dependable,long-lifeelectrochemicalsensor
• Threestagefiltrationforupto8users
RAM 744 Autocal® Wall Mounted or Fully Portable Compressed Air Monitoring System
RAM 744In-line CO Monitor
1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800)959-0329•(734)769-0573•www.goodforgas.com
Best Safeguard for Quality Breathing AirTheRAM744istheidealmonitorforportableapplicationsandwherespaceislimited.Itismountedwiththefiltrationsystem,allowingforrapidresponseandeliminatingthepossibilityofsamplelinesbecomingblockedorcrimped.Theflowrateisfixedsotherearenotouchyflowmeterstoadjust.
CalibrationandmaintenancearemadesimplewithourinnovativeAutoCal®systemthatallowstheCOsensortobecalibratedwithatestgasconnectionandtheflipofaswitch.Thealarmlevelissetto10ppm.
Thecompact,portable“aircase”enclosesthethree-stage25,50or100CFMfilterpanelwiththeRAM744COmonitorinaruggedcarryingcase*.Allmodelshavequick-connectinletsandoutlets.
AnyGfGInstrumentationRAMinstrumentcanbecustom-configuredtomeettheneedsoftheenduser’sexactspecifications.
Respiratory air specificationsAllGfGInstrumentationRAMinstrumentssatisfytherequirementsofOSHArule29CFR1910.134(d)(2)(ii)forcarbonmonoxidemonitoring.TheywillalsomonitorGradeEair,generallyspecifiedforbreathingairincylinders(SCBA).
OSHA Regulations (Standard - 29 CFR) Respiratory Protection - 1910.134(i)(6)Forcompressorsthatarenotoil-lubricated,theemployer shall ensure that carbon monoxidelevels in the breathing air do not exceed 10ppm.(i)(7)For oil-lubricated compressors, the employershall use a high-temperature or carbonmonoxide alarm, or both, to monitor carbonmonoxide levels. If only high-temperaturealarms are used, the air supply shall bemonitored at intervals sufficient to preventcarbon monoxide in the breathing air fromexceeding10ppm.
7.5"
2.25"
Specifications subject to change without notification
Specifications
Distributedby:
Rev.3(3/18/14)
Tel: (800) 959-0329 or (734) 769-0573 Fax: (734) 769-1888E-mail: [email protected] Website: www.goodforgas.com
Gases
Carbonmonoxide(CO)Detection Range
CO0-99ppm
Detection Principle
COspecificElectrochemicalcell
Display
2digit.5"digitbacklighted*LCD*Externalpowersourcesonly.
Expected Sensor Life
Greaterthan1year
Alarm Point
CO10ppm
Alarms
Audible83dBat1'(30cm)VisualFlashingLEDExternalalarm90dBat1m
Airline Inlet /Outlet
Standardpressure70to150psi1/2"FPT-50to100CFMpanels(wallmount)3/8"-25CFMpanels(wallmount)Dynaquipquickconnects-outlet
Available Relay Contacts and Ratings
N/O,common,N/C0.3Aat125VAC0.1Aat30VDC
Simple Flow Rate
1.0cubicfeetofairperhour(scfh)
Power Source
110VAC(wallpack)12VDCexternal9VDCinternal
Temperature Range
-4to+122ºF/-20to+50ºC
Available Models
8025-744 25CFMwallmountedfilterpanelwithRAM744(2ports)
8050-744 50CFMwallmountedfilterpanelwithRAM744(4ports)
8100-744 100CFMwallmountedfilterpanelwithRAM(8ports)
*9025-744 25CFMportablefilterpanel(2ports)*9050-744 50CFMportablefilterpanel(4ports)*9100-744 100CFMportablefilterpanel(8ports)