BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

A-101ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
RESOURCE LIST (SMS USE ONLY)
AutoCAD SHX Text
DESIGNATION
AutoCAD SHX Text
PG NUMBER
AutoCAD SHX Text
DATE
AutoCAD SHX Text
BIOGRAPH mCT-S/X/EXCEL/FLOW
AutoCAD SHX Text
MI19-000.891.01.06.02
AutoCAD SHX Text
02/17
AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
THE MINIMUM ROOM VOLUME REQUIRED FOR THE WATER-AIR SPLIT HEAT EXCHANGER IS 424 CUBIC FEET.
AutoCAD SHX Text
6'-5" x 2'-7"
AutoCAD SHX Text
1
AutoCAD SHX Text
2
AutoCAD SHX Text
3
AutoCAD SHX Text
4
AutoCAD SHX Text
5
AutoCAD SHX Text
6
AutoCAD SHX Text
8
AutoCAD SHX Text
7
AutoCAD SHX Text
9
AutoCAD SHX Text
10
AutoCAD SHX Text
11
AutoCAD SHX Text
12
AutoCAD SHX Text
14
AutoCAD SHX Text
13
AutoCAD SHX Text
15
AutoCAD SHX Text
17
AutoCAD SHX Text
16
AutoCAD SHX Text
18
AutoCAD SHX Text
20
AutoCAD SHX Text
22
AutoCAD SHX Text
25
AutoCAD SHX Text
28
AutoCAD SHX Text
26
AutoCAD SHX Text
30
AutoCAD SHX Text
29
AutoCAD SHX Text
ORIENTATION POINT
AutoCAD SHX Text
SIEMENS HIGHLY RECOMMENDS THE CUSTOMER PROVIDES THE FOLLOWING AREAS, ALL WHICH SHOULD BE IN CLOSE PROXIMITY TO THE SCANNER ROOM: PATIENT UP-TAKE ROOM, PATIENT DRESSING ROOM & EQUIPMENT UTILITY ROOM.
AutoCAD SHX Text
SCANNER ROOM
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
AS AN ALTERNATIVE, A SEPARATE SOURCE OF CHILLED WATER CAN BE PROVIDED BY THE CUSTOMER/CONTRACTOR. SEE CHILLED WATER REQUIREMENTS, SHEET M-101-A.
AutoCAD SHX Text
23
AutoCAD SHX Text
HEIGHT OF WINDOW TO BE COORDINATED WITH COUNTERTOP/DESK HEIGHT.
AutoCAD SHX Text
IT IS THE RESPONSIBILITY OF THE CUSTOMER/CONTRACTOR TO PROVIDE A MEANS OF MOUNTING THE PC TOWER(S) OFF OF FINISHED FLOOR FOR DAMAGE PROTECTION AGAINST TIP-OVER, FLUIDS, IMPACT ETC. ADDITIONAL MOUNTING NOT NECESSARY IF SIEMENS CONTAINER IS UTILIZED.
AutoCAD SHX Text
IT IS THE RESPONSIBILITY OF THE CUSTOMER/CONTRACTOR TO PROVIDE A MEANS OF MOUNTING THE PC TOWER(S) OFF OF FINISHED FLOOR FOR DAMAGE PROTECTION AGAINST TIP-OVER, FLUIDS, IMPACT ETC.
AutoCAD SHX Text
SIEMENS SUPPLIED WALL MOUNT HANGER AND RADIATION THERAPY PALLET. CUSTOMER/CONTRACTOR INSTALLS AND SUPPLIES MOUNTING HARDWARE.
AutoCAD SHX Text
THE UPS IS SUPPLIED, LOCATED AND INSTALLED BY SIEMENS.
AutoCAD SHX Text
19
AutoCAD SHX Text
24
AutoCAD SHX Text
THE OUTDOOR COOLING UNIT IS SUPPLIED BY SIEMENS, LOCATED AND INSTALLED BY THE CUSTOMER/ CONTRACTOR. THE INDOOR HEAT EXCHANGER CABINET IS SUPPLIED AND LOCATED BY SIEMENS, INSTALLED BY THE CUSTOMER/CONTRACTOR. IF THE SYSTEM IS TO BE RELOCATED, THE OUTDOOR COOLING UNIT IS TO BE DEINSTALLED, RELOCATED AND REINSTALLED BY THE CUSTOMER/CONTRACTOR, IF NECESSARY. THE INDOOR HEAT EXCHANGER CABINET IS RELOCATED BY SIEMENS, INSTALLED BY THE CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
21
AutoCAD SHX Text
27
AutoCAD SHX Text
HEIGHT DIFFERENCE BETWEEN HEAT EXCHANGER AND OUTDOOR UNIT MUST NOT EXCEED ±65'-5".
AutoCAD SHX Text
SYNGO MULTIMODALITY WORKPLACE LOCATED AT CUSTOMER'S DISCRETION.
AutoCAD SHX Text
HEIGHT DIFFERENCE BETWEEN HEAT EXCHANGER AND GANTRY MUST NOT EXCEED ±32'-6".
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
ARCHITECTURAL EQUIPMENT PLAN
AutoCAD SHX Text
SCALE: 1/4" = 1'-0"
AutoCAD SHX Text
EQUIPMENT LEGEND
AutoCAD SHX Text
NO
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
SMS
AutoCAD SHX Text
WEIGHT
AutoCAD SHX Text
BTU/HR
AutoCAD SHX Text
DIMENSIONS (INCHES)
AutoCAD SHX Text
REMARKS
AutoCAD SHX Text
SYM
AutoCAD SHX Text
(LBS)
AutoCAD SHX Text
TO AIR
AutoCAD SHX Text
W
AutoCAD SHX Text
D
AutoCAD SHX Text
H
AutoCAD SHX Text
1
AutoCAD SHX Text
-
AutoCAD SHX Text
OPERATING CONTROL CONSOLE W/KEYBOARD AND CONTROL BOX
AutoCAD SHX Text
91
AutoCAD SHX Text
-
AutoCAD SHX Text
47 1/4
AutoCAD SHX Text
31 1/2
AutoCAD SHX Text
28 3/4
AutoCAD SHX Text
DESIGN FROM YEAR 09/2006
AutoCAD SHX Text
2
AutoCAD SHX Text
-
AutoCAD SHX Text
19" FLAT SCREEN DUAL CONTROL MONITORS
AutoCAD SHX Text
29
AutoCAD SHX Text
<225
AutoCAD SHX Text
16 1/2
AutoCAD SHX Text
8 3/8
AutoCAD SHX Text
16 1/8
AutoCAD SHX Text
ON CUSTOMER'S COUNTER
AutoCAD SHX Text
3
AutoCAD SHX Text
PIT
AutoCAD SHX Text
POWER CONNECTION TERMINAL - ICS
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
WALL MOUNTED
AutoCAD SHX Text
4
AutoCAD SHX Text
ICS
AutoCAD SHX Text
SYNGO ACQUISITION WORKPLACE
AutoCAD SHX Text
<30
AutoCAD SHX Text
<1706
AutoCAD SHX Text
9 13/16
AutoCAD SHX Text
29 1/2
AutoCAD SHX Text
18 1/2
AutoCAD SHX Text
OFF FLOOR/IN CONTAINER
AutoCAD SHX Text
5
AutoCAD SHX Text
-
AutoCAD SHX Text
IMAGE EVALUATION KEYBOARD - BIOGRAPH ADVANCED WORKFLOW (OPTION)
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
ON CUSTOMER'S COUNTER
AutoCAD SHX Text
6
AutoCAD SHX Text
-
AutoCAD SHX Text
19" FLAT SCREEN DUAL MONITORS FOR IES - BIOGRAPH ADVANCED WORKFLOW (OPTION)
AutoCAD SHX Text
29
AutoCAD SHX Text
<225
AutoCAD SHX Text
16 1/2
AutoCAD SHX Text
8 3/8
AutoCAD SHX Text
16 1/8
AutoCAD SHX Text
ON CUSTOMER'S COUNTER
AutoCAD SHX Text
7
AutoCAD SHX Text
IES
AutoCAD SHX Text
SYNGO EVALUATION WORKPLACE - BIOGRAPH ADVANCED WORKFLOW (OPTION)
AutoCAD SHX Text
<30
AutoCAD SHX Text
<1706
AutoCAD SHX Text
9 13/16
AutoCAD SHX Text
29 1/2
AutoCAD SHX Text
18 1/2
AutoCAD SHX Text
OFF FLOOR/IN CONTAINER
AutoCAD SHX Text
8
AutoCAD SHX Text
PIT2
AutoCAD SHX Text
POWER CONNECTION TERMINAL - BIOGRAPH ADVANCED WORKFLOW (OPTION)
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
WALL MOUNTED
AutoCAD SHX Text
9
AutoCAD SHX Text
IMS
AutoCAD SHX Text
CONTAINER & CONTAINER TABLE FOR ICS/IES (OPTION)
AutoCAD SHX Text
67
AutoCAD SHX Text
-
AutoCAD SHX Text
31 1/2
AutoCAD SHX Text
31 1/2
AutoCAD SHX Text
28 3/8
AutoCAD SHX Text
HOUSING FOR ICS/IES
AutoCAD SHX Text
10
AutoCAD SHX Text
B
AutoCAD SHX Text
BIOGRAPH mCT FLOW 20 40 64 EDGE GANTRY
AutoCAD SHX Text
7,184
AutoCAD SHX Text
5,123*
AutoCAD SHX Text
92 5/16
AutoCAD SHX Text
53 1/2
AutoCAD SHX Text
80 1/4
AutoCAD SHX Text
*ADDITIONAL HEAT DISSIPATED TO WATER 61,473 BTU/HR
AutoCAD SHX Text
10
AutoCAD SHX Text
B
AutoCAD SHX Text
BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128 GANTRY
AutoCAD SHX Text
7,184
AutoCAD SHX Text
5,123*
AutoCAD SHX Text
92 5/16
AutoCAD SHX Text
53 1/2
AutoCAD SHX Text
80 1/4
AutoCAD SHX Text
*ADDITIONAL HEAT DISSIPATED TO WATER 61,473 BTU/HR
AutoCAD SHX Text
11
AutoCAD SHX Text
-
AutoCAD SHX Text
PATIENT TABLE
AutoCAD SHX Text
1,665
AutoCAD SHX Text
1,708
AutoCAD SHX Text
17
AutoCAD SHX Text
148 7/8
AutoCAD SHX Text
48
AutoCAD SHX Text
12
AutoCAD SHX Text
PDC
AutoCAD SHX Text
POWER DISTRIBUTION CABINET & UPS
AutoCAD SHX Text
<1,374
AutoCAD SHX Text
6,824
AutoCAD SHX Text
35 1/2
AutoCAD SHX Text
27 1/2
AutoCAD SHX Text
76 7/8
AutoCAD SHX Text
UPS LOCATED INSIDE OF PDC
AutoCAD SHX Text
13
AutoCAD SHX Text
ES
AutoCAD SHX Text
ETHERNET SWITCH FOR ICS, IRS, GANTRY & PDC'S
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
LOCATED INSIDE PDC
AutoCAD SHX Text
14
AutoCAD SHX Text
ES2
AutoCAD SHX Text
ETHERNET SWITCH FOR IES, DICOM CAMERAS & NETWORK PRINTER
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
LOCATED INSIDE PDC
AutoCAD SHX Text
15
AutoCAD SHX Text
IRS
AutoCAD SHX Text
IMAGE RECONSTRUCTION SYSTEM
AutoCAD SHX Text
114
AutoCAD SHX Text
2,491
AutoCAD SHX Text
23 1/2
AutoCAD SHX Text
31 1/2
AutoCAD SHX Text
27
AutoCAD SHX Text
OFF FLOOR
AutoCAD SHX Text
16
AutoCAD SHX Text
ACS
AutoCAD SHX Text
PET ACQUISITION SYSTEM
AutoCAD SHX Text
<60
AutoCAD SHX Text
1,706
AutoCAD SHX Text
8
AutoCAD SHX Text
21
AutoCAD SHX Text
17 13/16
AutoCAD SHX Text
OFF FLOOR
AutoCAD SHX Text
17
AutoCAD SHX Text
PRS
AutoCAD SHX Text
PET RECONSTRUCTION SYSTEM
AutoCAD SHX Text
<60
AutoCAD SHX Text
1,706
AutoCAD SHX Text
8
AutoCAD SHX Text
21
AutoCAD SHX Text
17 13/16
AutoCAD SHX Text
OFF FLOOR
AutoCAD SHX Text
18
AutoCAD SHX Text
APD
AutoCAD SHX Text
POWER CONNECTION TERMINAL - ACS/PRS
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
13 9/16
AutoCAD SHX Text
2 15/16
AutoCAD SHX Text
5 11/16
AutoCAD SHX Text
WALL MOUNTED
AutoCAD SHX Text
19
AutoCAD SHX Text
PDU
AutoCAD SHX Text
POWER DISTRIBUTION UNIT & COMP UPS - PDU
AutoCAD SHX Text
792
AutoCAD SHX Text
3,800
AutoCAD SHX Text
27 3/4
AutoCAD SHX Text
30
AutoCAD SHX Text
61 1/4
AutoCAD SHX Text
ON FLOOR
AutoCAD SHX Text
20
AutoCAD SHX Text
PPS
AutoCAD SHX Text
PET GANTRY UPS (OPTION)
AutoCAD SHX Text
106
AutoCAD SHX Text
1,150
AutoCAD SHX Text
17 1/2
AutoCAD SHX Text
28 1/2
AutoCAD SHX Text
6
AutoCAD SHX Text
LOCATED INSIDE PDU
AutoCAD SHX Text
21
AutoCAD SHX Text
-
AutoCAD SHX Text
LEAD PIG
AutoCAD SHX Text
392
AutoCAD SHX Text
-
AutoCAD SHX Text
17
AutoCAD SHX Text
17
AutoCAD SHX Text
20 1/2
AutoCAD SHX Text
RADIOACTIVE SOURCE STORAGE
AutoCAD SHX Text
22
AutoCAD SHX Text
S
AutoCAD SHX Text
HEAT EXCHANGER CABINET - WATER/AIR SPLIT (OPTION)
AutoCAD SHX Text
911
AutoCAD SHX Text
3,412
AutoCAD SHX Text
39 3/8
AutoCAD SHX Text
27 1/4
AutoCAD SHX Text
77
AutoCAD SHX Text
ON FLOOR
AutoCAD SHX Text
23
AutoCAD SHX Text
S2
AutoCAD SHX Text
OUTDOOR UNIT - WATER/AIR SPLIT (OPTION)
AutoCAD SHX Text
410
AutoCAD SHX Text
102,364
AutoCAD SHX Text
100 3/4
AutoCAD SHX Text
42 1/8
AutoCAD SHX Text
44 5/16
AutoCAD SHX Text
24
AutoCAD SHX Text
-
AutoCAD SHX Text
RTP PALLET (OPTION)
AutoCAD SHX Text
23.5
AutoCAD SHX Text
-
AutoCAD SHX Text
21
AutoCAD SHX Text
10
AutoCAD SHX Text
80 1/2
AutoCAD SHX Text
SIEMENS/CUSTOMER SUPPLIED WALL MOUNTED.
AutoCAD SHX Text
25
AutoCAD SHX Text
F
AutoCAD SHX Text
CARE VISION SINGLE MONITOR (INTERVENTION) (OPTION)
AutoCAD SHX Text
79
AutoCAD SHX Text
256
AutoCAD SHX Text
29 1/8
AutoCAD SHX Text
26
AutoCAD SHX Text
56 11/16
AutoCAD SHX Text
MOBILE MONITOR CART
AutoCAD SHX Text
26
AutoCAD SHX Text
IN1
AutoCAD SHX Text
MEDRAD DISPLAY CONTROL UNIT (OPTION)
AutoCAD SHX Text
8
AutoCAD SHX Text
-
AutoCAD SHX Text
12 1/2
AutoCAD SHX Text
9
AutoCAD SHX Text
13 1/2
AutoCAD SHX Text
HEIGHT WITH SCREEN UP
AutoCAD SHX Text
27
AutoCAD SHX Text
IN2
AutoCAD SHX Text
MEDRAD BASE UNIT (OPTION)
AutoCAD SHX Text
14
AutoCAD SHX Text
-
AutoCAD SHX Text
11
AutoCAD SHX Text
8 3/4
AutoCAD SHX Text
11 1/2
AutoCAD SHX Text
UNDER COUNTER ON SHELF
AutoCAD SHX Text
28
AutoCAD SHX Text
IN3
AutoCAD SHX Text
CEILING MOUNTED MEDRAD INJECTOR (OPTION)
AutoCAD SHX Text
106
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
SEE MFG SPECIFICATIONS
AutoCAD SHX Text
29
AutoCAD SHX Text
MMW
AutoCAD SHX Text
SYNGO MULTIMODALITY WORKPLACE COMPUTER (OPTION)
AutoCAD SHX Text
55
AutoCAD SHX Text
-
AutoCAD SHX Text
19 3/4
AutoCAD SHX Text
10
AutoCAD SHX Text
23 5/8
AutoCAD SHX Text
ON CUSTOMER'S COUNTER
AutoCAD SHX Text
30
AutoCAD SHX Text
-
AutoCAD SHX Text
SYNGO MULTIMODALITY WORKPLACE KEYBOARD AND MONITOR (OPTION)
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
-
AutoCAD SHX Text
ON CUSTOMER'S COUNTER
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
1 11
AutoCAD SHX Text
FINISHED ROOM HEIGHT
AutoCAD SHX Text
FOR BIOGRAPH mCT GANTRY ONLY
AutoCAD SHX Text
CAREVISION MONITOR/CEILING MOUNT
AutoCAD SHX Text
RECOMMEND 8'-0" MINIMUM 7'-5"
AutoCAD SHX Text
MIN. 8'-6 3/8" MAX. 12'-1 11/16"
AutoCAD SHX Text
CEILING MOUNT INJECTOR
AutoCAD SHX Text
MINIMUM 8'-9 1/2"
AutoCAD SHX Text
ARCHITECTURAL NOTES
AutoCAD SHX Text
1) ALL PRELIMINARY EQUIPMENT LAYOUTS SUBMITTED BY SIEMENS HEALTHCARE ARE BASED ON THE RECOMMENDED SPACE NECESSARY FOR THE OPERATION AND SERVICEABILITY OF THE EQUIPMENT BEING PROPOSED. SIEMENS WILL NOT SUBMIT AN EQUIPMENT LAYOUT THAT IS NOT IN THE BEST INTEREST OF BOTH THE CUSTOMER AND SIEMENS. ALL EQUIPMENT LAYOUTS ARE BASED EITHER ON AN ACTUAL SITE SURVEY OR ARCHITECTURAL DRAWINGS SUPPLIED TO SIEMENS. SIEMENS WILL NOT BE RESPONSIBLE FOR ANY ALTERATIONS THAT ENCROACH WITHIN DESIGNATED SAFETY AND SERVICE CLEARANCE ZONES AS INDICATED ON DRAWINGS (I.E.. PIPE CHASES, VENTILATION DUCTS, CASEWORK, AND SOFFITS, ETC.) MADE BY THE CUSTOMER OR REQUIRED BY A CUSTOMER'S ARCHITECTURAL FIRM ONCE PRELIMINARY DRAWINGS HAVE BEEN SUBMITTED AND APPROVED. DO NOT ALTER ANY SPECIFICATIONS AND/OR DIMENSIONS WITHOUT CONTACTING AND RECEIVING WRITTEN CONFIRMATION FROM SIEMENS PROJECT MANAGER. 2) SIEMENS HEALTHCARE IS NOT AN ARCHITECTURAL OR ENGINEERING FIRM. DRAWINGS SUPPLIED BY SIEMENS ARE NOT CONSTRUCTION DRAWINGS. THEREFORE, THESE DRAWINGS ARE TO BE USED ONLY FOR INFORMATION TO COMPLEMENT ACTUAL CONSTRUCTION DRAWINGS AVAILABLE FROM A CUSTOMER APPOINTED ARCHITECTURAL REPRESENTATIVE OR A CUSTOMER'S ENGINEERING DESIGN GROUP. THE CUSTOMER'S ARCHITECT AND GENERAL CONTRACTOR SHALL BE ULTIMATELY RESPONSIBLE FOR COMPLIANCE WITH ALL APPLICABLE CODES AND PROFESSIONAL DESIGN REQUIREMENTS INCLUDING OSHA/NEC SAFETY CLEARANCE REQUIRMENTS IN ADDITION TO SIEMENS-REQUIRED SAFETY/SERVICE CLEARANCES SHOWN. 3) THE CUSTOMER IS RESPONSIBLE FOR ALL ROOM AND AREA PREPARATION COSTS, PROFESSIONAL FEES, PERMITS, REPORTS, AND INSPECTION FEES. 4) EQUIPMENT WARRANTIES, EXPRESSED OR IMPLIED ON THE PART OF SIEMENS SHALL BE CONTINGENT UPON STRICT COMPLIANCE WITH THE ARCHITECTURAL, STRUCTURAL, ELECTRICAL, MECHANICAL AND RECOMMENDATIONS AND REQUIREMENTS CONTAINED IN THESE DRAWINGS, UNLESS SPECIFIED OTHERWISE. 5) ALL DIMENSIONS SHOWN ARE FROM FINISHED SURFACES UNLESS SPECIFIED OTHERWISE. 6) THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION PHYSICIST. ACTUAL PROTECTION REQUIREMENTS SHALL BE SPECIFIED BY A REGISTERED RADIATION PHYSICIST AT CUSTOMER'S ENGAGEMENT AND EXPENSE. RESPONSIBILITY FOR ALL INFORMATION AS TO THE ROOM LOCATION, USE, AND NUMBER OF ANTICIPATED EXAMINATIONS TO BE PERFORMED PER TIME PERIOD SHALL BE PROVIDED TO THE PHYSICIST BY THE CUSTOMER. THE CUSTOMER SHALL FURTHER TAKE ALL RESPONSIBILITY IN THE COMMUNICATION AND COORDINATION OF ACTIVITIES OF THE RADIATION PHYSICIST AND THE ARCHITECTURAL REPRESENTATIVE. 7) SIEMENS HEALTHCARE SHALL BE RESPONSIBLE FOR SIEMENS EQUIPMENT INSTALLATION, CALIBRATION, CONNECTION AND INSTALLATION OF SIEMENS PROVIDED CABLES. THE CUSTOMER/ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR TERMINATIONS OF CUSTOMER/ELECTRICAL CONTRACTOR-SUPPLIED CABLES TO SIEMENS EQUIPMENT. IN THE EVENT THAT SPECIFIC TRADE RULES OR LICENSE REQUIREMENTS PROHIBIT THIS, THE CUSTOMER SHALL INITIATE THE SERVICES OF APPROVED OTHER CONTRACTORS AND PAY FOR SELECTED, APPROVED PARTIES TO PERFORM THIS WORK WITH SUPERVISION PROVIDED BY SIEMENS. CALIBRATION WHEN ACCOMPLISHED OUTSIDE OF NORMAL INSTALLATION SEQUENCES DUE TO CONTRACTOR OR TRADE RULE ACTIONS OR REQUIREMENTS SHALL BE SUPPORTED BY, CHARGED TO, AND ACCEPTED BY THE CUSTOMER AS AN ADDITIONAL INSTALLATION EXPENSE. 8) THE CUSTOMER SHALL COORDINATE WITH SIEMENS PROJECT MANAGER THE LOCATIONS AND TRAVEL OF ALL ANCILLARY EQUIPMENT TO BE CEILING OR WALL MOUNTED (I.E.: O.R. LIGHTS, MEDICAL GAS COLUMNS, PHYSIOLOGICAL MONITORING INJECTORS, CRT PLATFORMS, SPRINKLER HEADS, SMOKE DETECTORS, ELECTRICAL OUTLETS, HVAC GRILLES, SPEAKERS, AND GENERAL ROOM LIGHTING, ETC.). 9) THE GENERAL CONTRACTOR/CUSTOMER SHALL BE RESPONSIBLE FOR ALL FINAL PAINT, TOUCH-UP AND ANY COSMETIC OR TRIM WORK WHICH NEEDS TO BE OR IS REQUIRED TO BE COMPLETED AFTER THE INSTALLATION OF THE SIEMENS EQUIPMENT AND ANY ASSOCIATED SUPPORT APPARATUS.
AutoCAD SHX Text
CUSTOMER FROM THE APPROPRIATE STATE AGENCY, IF APPLICABLE.
AutoCAD SHX Text
DIAGNOSTIC OR THERAPEUTIC PURPOSES, MUST BE OBTAINED BY THE
AutoCAD SHX Text
CONSTRUCTION OR STRUCTURAL MODIFICATIONS UTILIZING X-RAY FOR
AutoCAD SHX Text
PRIOR TO SIEMENS EQUIPMENT INSTALLATION, APPROVAL OF
AutoCAD SHX Text
STATE AGENCY REVIEW
AutoCAD SHX Text
INSTALLATION & OPERATION OF THE SIEMENS EQUIPMENT.
AutoCAD SHX Text
HEREWITH, AS THEY ARE ESSENTIAL FOR THE SUCCESSFUL
AutoCAD SHX Text
CONTRACTOR. FOLLOW DESIGN RECOMMENDATIONS INCLUDED
AutoCAD SHX Text
DETAILED, FURNISHED AND INSTALLED BY THE CUSTOMER AND/OR
AutoCAD SHX Text
1) ALL CASEWORK IS EITHER EXISTING OR IS TO BE DESIGNED,
AutoCAD SHX Text
CASEWORK & ACCESSORY NOTES
AutoCAD SHX Text
BE PROVIDED BY THE CUSTOMER.
AutoCAD SHX Text
2) ALL FURNITURE (CHAIRS, ETC.) FOR THE CONTROL ROOM ARE TO
AutoCAD SHX Text
CLIMATE CONTROL MUST BE PROVIDED 24 HOURS A DAY, 7 DAYS A WEEK. TEMPERATURE SETBACKS ARE NOT ALLOWED. PLEASE SEE EQUIPMENT LEGEND FOR SITE SPECIFIC HEAT DISSIPATION. SCANNER ROOM: THE SCANNER ROOM SHOULD MAINTAIN BETWEEN 68°F-79°F (± 2.7°F PER HR.) AND A RELATIVE HUMIDITY OF 20-75% WITH DewPoint BELOW 63°F. A BAROMETRIC PRESSURE: 10.9 TO 15.4 PSI. AIR CONDITIONING MUST BE PROVIDED 24 HRS A DAY, 7 DAYS A WEEK. EQUIPMENT ROOM: THE EQUIPMENT ROOM SHOULD MAINTAIN BETWEEN 64°F-86°F AND A RELATIVE HUMIDITY OF 20-75%. A BAROMETRIC PRESSURE: 10.9 TO 15.4 PSI. AIR CONDITIONING MUST BE PROVIDED 24 HRS A DAY, 7 DAYS A WEEK. CONTROL ROOM: THE CONTROL ROOM SHOULD MAINTAIN BETWEEN 68°F-79°F AND A RELATIVE HUMIDITY OF 20-75%. A BAROMETRIC PRESSURE: 10.9 TO 15.4 PSI. EXTERIOR AIR VENTS SHOULD BE EQUIPPED WITH A FILTRATION SYSTEM OF THE FILTER CLASS EU3 TO EU4 TO FILTER DUST PARTICLES UP TO >10µm. µm. m. THE ROOM AIR SHOULD BE PROTECTED AGAINST CONTAMINATION BY HYDROGEN SULPHIDE, EVEN IN SMALL AMOUNTS. THE MOST COMMON SOURCES FOR HYDROGEN SULFIDES ARE: EXHAUST FUMES AND WASTE WATER OF FILM PROCESSORS. EXPOSED SEWER DRAINAGE - NON SYPHON INCLUDED. SEWER PIPE OR IN FLOOR DRAIN. EXHAUST FUMES FROM DIESEL POWER UNITS EMERGENCY POWER, ETC. IF A DANGER OF SUCH CONTAMINATION EXISTS, CORRECTIVE ACTIONS IS REQUIRED E.G., EXTRACTOR FANS SIPHON MODIFICATION OF VENTILATION INTAKE, ETC.
AutoCAD SHX Text
ENVIRONMENTAL REQUIREMENTS
AutoCAD SHX Text
EMERGENCY POWER OFF (EPO) BUTTONS REQUIRED IN CONTROL ROOM AREA, SCANNER ROOM AREA AND EQUIPMENT AREA.
AutoCAD SHX Text
DOOR (SAFETY) SWITCH REQUIRED ON ALL DOORS ACCESSING THE EXAMINATION ROOM IN ACCORDANCE WITH LOCAL CODES.
AutoCAD SHX Text
PLANNING REQUIREMENTS
AutoCAD SHX Text
SYSTEM COMPONENT
AutoCAD SHX Text
DECIBEL LEVEL
AutoCAD SHX Text
PDC CABINET
AutoCAD SHX Text
(AT 3'-3" DISTANCE)
AutoCAD SHX Text
5555
AutoCAD SHX Text
<55 (1)
AutoCAD SHX Text
IRS TOWER
AutoCAD SHX Text
HEAT EXCHANGER - WATER/AIR SPLIT
AutoCAD SHX Text
<6060
AutoCAD SHX Text
1) NOISE DEPENDS ON THE ROOM TEMPERATURE AND THE PROCESSOR LOAD.
AutoCAD SHX Text
PDU CABINET
AutoCAD SHX Text
5555
AutoCAD SHX Text
ACS / PRS TOWER PC
AutoCAD SHX Text
<5050
AutoCAD SHX Text
PET/CT GANTRY
AutoCAD SHX Text
<6868
AutoCAD SHX Text
NOISE LEVEL
AutoCAD SHX Text
PHS
AutoCAD SHX Text
<6060
AutoCAD SHX Text
LEAD OR EQUIVALENT SHIELDING MAY BE REQUIRED IN THE WALLS OF THE SCANNER ROOM, HOTLAB AND/OR PATIENT PREPARATION AREAS. IT IS THE RESPONSIBILITY OF THE CUSTOMER TO VERIFY WITH THE SITE'S RADIATION SAFETY OFFICER THAT RADIATION DOSE RATES FROM THE PET PATIENT AND/OR ISOTOPE WILL NOT EXCEED LOCAL RADIATION SAFETY GUIDELINES IN THE ROOM ADJACENT TO SCANNER, HOTLAB, AND/OR PATIENT PREPARATION AREAS. IMPROPER SHIELDING MAY AFFECT CAMERA'S PERFORMANCE.
AutoCAD SHX Text
RADIATION SAFETY
AutoCAD SHX Text
RAM LICENSE NEEDS TO BE APPLIED THROUGH GOVERNMENT AGENCY AS EARLY AS POSSIBLE. PLEASE ADDRESS WITH YOUR RSO (RADIATION SAFETY OFFICER). RAM LICENSE MUST BE OBTAINED NO LATER THAN 4 WEEKS AHEAD OF SCHEDULED DELIVERY. DELAY OF INSTALLATION MAY OCCUR IF SITE HAS NOT OBTAINED RAM LICENSE AT THIS TIME. RADIOACTIVE SOURCES NEEDED TO COMPLETE CALIBRATION OF EQUIPMENT WILL NOT BE SHIPPED TO SITE WITHOUT VALID RAM LICENSE.
AutoCAD SHX Text
RAM LICENSE
AutoCAD SHX Text
THE CT PRODUCES RADIATION WHILE PERFORMING BIOGRAPH mCT FLOW OR BIOGRAPH mCT SCANS. RADIATION CONCERNS FOR PET LIE IN THE USE OF RADIOACTIVE ISOTOPES FOR CLINICAL SCANNING OR SERVICE SCANS. THE RADIOACTIVE SOURCES REQUIRED FOR INSTALLING AND SERVICING THE PET GANTRY, INCLUDE PHANTOM CYLINDERS OF SEVERAL mCi ASSAY AND TWO APPROXIMATE 1.0 - 1.5 mCi ROD SOURCES AS NEEDED FOR THE GANTRY OFFSET CALIBRATION. SOURCES MUST BE AVAILABLE UPON SYSTEM INSTALLATION. A STORAGE AREA SHOULD BE DESIGNATED FOR SOURCES UNTIL INSTALLATION TO LIMIT EXPOSURE. ADDITIONAL RADIATION CONSIDERATIONS: STATIC MAGNETIC FIELD: B< 100µT < 100µT 100µT µT T MAGNET FIELD VARIATION: B < 25µT B < 25µT < 25µT 25µT µT T BACKGROUND RADIATION: <10 µSv/h (1mR/h) 6 FT FROM <10 µSv/h (1mR/h) 6 FT FROM µSv/h (1mR/h) 6 FT FROM µSv/h (1mR/h) 6 FT FROM Sv/h (1mR/h) 6 FT FROM CENTER OF FIELD OF VIEW
AutoCAD SHX Text
eff
AutoCAD SHX Text
RADIATION AND STORAGE CONSIDERATIONS

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

A-102ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
ORIENTATION POINT
AutoCAD SHX Text
SERVICE CLEARANCE
AutoCAD SHX Text
SAFETY CLEARANCE
AutoCAD SHX Text
PHS MOVEMENT
AutoCAD SHX Text
SERVICE CLEARANCE
AutoCAD SHX Text
SERVICE CLEARANCE
AutoCAD SHX Text
SAFETY CLEARANCE
AutoCAD SHX Text
SERVICE CLEARANCE
AutoCAD SHX Text
GANTRY IN SERVICE POSITION
AutoCAD SHX Text
SAFETY CLEARANCE
AutoCAD SHX Text
SCANNER ROOM
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
SERVICE CLEARANCE
AutoCAD SHX Text
FOOT EXTENSION
AutoCAD SHX Text
SAFETY/SERVICE CLEARANCE PLAN
AutoCAD SHX Text
SCALE: 1/4" = 1'-0"
AutoCAD SHX Text
IF THE SAFETY DISTANCES ARE NOT OBSERVED, SAFETY MEASURES IN ACCORDANCE WITH LOCAL CODES SHOULD BE UTILIZED (FOR EXAMPLE BARRIERS, WARNING SIGNS, AND SAFETY MATS).
AutoCAD SHX Text
SAFETY CLEARANCE NOTE
AutoCAD SHX Text
.0519
AutoCAD SHX Text
.0794
AutoCAD SHX Text
.0159
AutoCAD SHX Text
NA
AutoCAD SHX Text
.0054
AutoCAD SHX Text
.0425
AutoCAD SHX Text
.0519
AutoCAD SHX Text
.0666
AutoCAD SHX Text
.01309
AutoCAD SHX Text
.0649
AutoCAD SHX Text
.0676
AutoCAD SHX Text
.0673
AutoCAD SHX Text
.0620
AutoCAD SHX Text
.1112
AutoCAD SHX Text
.2910
AutoCAD SHX Text
NA
AutoCAD SHX Text
NA
AutoCAD SHX Text
39.4
AutoCAD SHX Text
19.7
AutoCAD SHX Text
0
AutoCAD SHX Text
-19.7
AutoCAD SHX Text
-39.4
AutoCAD SHX Text
-78.7
AutoCAD SHX Text
-59.1
AutoCAD SHX Text
-39.4
AutoCAD SHX Text
-19.7
AutoCAD SHX Text
0
AutoCAD SHX Text
19.7
AutoCAD SHX Text
39.4
AutoCAD SHX Text
59.1
AutoCAD SHX Text
78.7
AutoCAD SHX Text
.1578
AutoCAD SHX Text
MEASUREMENT IN uGy/mAs
AutoCAD SHX Text
.0381
AutoCAD SHX Text
.0424
AutoCAD SHX Text
.0432
AutoCAD SHX Text
.0673
AutoCAD SHX Text
NA
AutoCAD SHX Text
.0337
AutoCAD SHX Text
.0390
AutoCAD SHX Text
.0386
AutoCAD SHX Text
NA
AutoCAD SHX Text
NA
AutoCAD SHX Text
.0236
AutoCAD SHX Text
.0258
AutoCAD SHX Text
.0210
AutoCAD SHX Text
NA
AutoCAD SHX Text
NA
AutoCAD SHX Text
.0177
AutoCAD SHX Text
.0183
AutoCAD SHX Text
.0153
AutoCAD SHX Text
.0023
AutoCAD SHX Text
NA
AutoCAD SHX Text
.0145
AutoCAD SHX Text
.0143
AutoCAD SHX Text
.0121
AutoCAD SHX Text
.0019
AutoCAD SHX Text
NA
AutoCAD SHX Text
.3293
AutoCAD SHX Text
.1316
AutoCAD SHX Text
.5814
AutoCAD SHX Text
.1514
AutoCAD SHX Text
98.4
AutoCAD SHX Text
118.1
AutoCAD SHX Text
137.8
AutoCAD SHX Text
59.1
AutoCAD SHX Text
.0365
AutoCAD SHX Text
.0241
AutoCAD SHX Text
.0133
AutoCAD SHX Text
.0036
AutoCAD SHX Text
.0038
AutoCAD SHX Text
.0103
AutoCAD SHX Text
.0230
AutoCAD SHX Text
.0253
AutoCAD SHX Text
.0282
AutoCAD SHX Text
.0215
AutoCAD SHX Text
.0170
AutoCAD SHX Text
.0138
AutoCAD SHX Text
SCANNING WAS PERFORMED USING A MAXIMUM SLICE THICKNESS OF 64 x 0.6 mm (38.4 mm) AT 140 kV THROUGH THE SYSTEM AXIS IN THE VERTICAL PLANE. PHANTOM USED: CYLINDRICAL PMMA PHANTOM, 32 cm IN DIAMETER, 16 CM LONG. THE PHANTOM WAS CENTERED IN THE TOMOGRAPHIC PLANE.
AutoCAD SHX Text
VERTICAL LOCAL DOSE DISTRIBUTION
AutoCAD SHX Text
INCHES
AutoCAD SHX Text
INCHES
AutoCAD SHX Text
SCALE 1/4"=1'-0"
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
RADIATION SCATTER
AutoCAD SHX Text
.0237
AutoCAD SHX Text
.0491
AutoCAD SHX Text
.0639
AutoCAD SHX Text
.0673
AutoCAD SHX Text
.0616
AutoCAD SHX Text
.0495
AutoCAD SHX Text
.0200
AutoCAD SHX Text
.0079
AutoCAD SHX Text
.0427
AutoCAD SHX Text
.1243
AutoCAD SHX Text
.1578
AutoCAD SHX Text
.1189
AutoCAD SHX Text
.0496
AutoCAD SHX Text
.0098
AutoCAD SHX Text
.0034
AutoCAD SHX Text
.6021
AutoCAD SHX Text
.0029
AutoCAD SHX Text
.0028
AutoCAD SHX Text
.0047
AutoCAD SHX Text
.0124
AutoCAD SHX Text
.0525
AutoCAD SHX Text
.3332
AutoCAD SHX Text
.5814
AutoCAD SHX Text
.3525
AutoCAD SHX Text
.0485
AutoCAD SHX Text
.0091
AutoCAD SHX Text
.0286
AutoCAD SHX Text
.0755
AutoCAD SHX Text
.1373
AutoCAD SHX Text
.1514
AutoCAD SHX Text
.1228
AutoCAD SHX Text
.0837
AutoCAD SHX Text
.0301
AutoCAD SHX Text
.0364
AutoCAD SHX Text
.0508
AutoCAD SHX Text
.0626
AutoCAD SHX Text
.0676
AutoCAD SHX Text
.0653
AutoCAD SHX Text
.0489
AutoCAD SHX Text
.0333
AutoCAD SHX Text
.0256
AutoCAD SHX Text
.0322
AutoCAD SHX Text
.0379
AutoCAD SHX Text
.0386
AutoCAD SHX Text
.0357
AutoCAD SHX Text
.0329
AutoCAD SHX Text
.0274
AutoCAD SHX Text
.0206
AutoCAD SHX Text
.0226
AutoCAD SHX Text
.0235
AutoCAD SHX Text
.0210
AutoCAD SHX Text
.0244
AutoCAD SHX Text
.0221
AutoCAD SHX Text
.0190
AutoCAD SHX Text
.0142
AutoCAD SHX Text
.0161
AutoCAD SHX Text
.0176
AutoCAD SHX Text
.0153
AutoCAD SHX Text
.0167
AutoCAD SHX Text
.0166
AutoCAD SHX Text
.0155
AutoCAD SHX Text
59.1
AutoCAD SHX Text
39.4
AutoCAD SHX Text
19.7
AutoCAD SHX Text
0
AutoCAD SHX Text
-19.7
AutoCAD SHX Text
-39.4
AutoCAD SHX Text
-59.1
AutoCAD SHX Text
-78.7
AutoCAD SHX Text
-98.4
AutoCAD SHX Text
-118.1
AutoCAD SHX Text
-78.7
AutoCAD SHX Text
-59.1
AutoCAD SHX Text
-39.4
AutoCAD SHX Text
-19.7
AutoCAD SHX Text
0
AutoCAD SHX Text
19.7
AutoCAD SHX Text
39.4
AutoCAD SHX Text
59.1
AutoCAD SHX Text
78.7
AutoCAD SHX Text
78.7
AutoCAD SHX Text
-137.8
AutoCAD SHX Text
.0254
AutoCAD SHX Text
.0374
AutoCAD SHX Text
.0418
AutoCAD SHX Text
.0432
AutoCAD SHX Text
.0425
AutoCAD SHX Text
.0363
AutoCAD SHX Text
.0250
AutoCAD SHX Text
.0121
AutoCAD SHX Text
.0127
AutoCAD SHX Text
.0130
AutoCAD SHX Text
.0121
AutoCAD SHX Text
.0130
AutoCAD SHX Text
.0128
AutoCAD SHX Text
.0118
AutoCAD SHX Text
MEASUREMENT IN uGy/mAs SCALE 1/4"=1'-0"
AutoCAD SHX Text
SCANNING WAS PERFORMED USING A MAXIMUM SLICE THICKNESS OF 64 x 0.6 mm (38.4 mm) AT 140 kV THROUGH THE SYSTEM AXIS IN THE HORIZONTAL PLANE. PHANTOM USED: CYLINDRICAL PMMA PHANTOM, 32 cm IN DIAMETER, 16 CM LONG. THE PHANTOM WAS CENTERED IN THE TOMOGRAPHIC PLANE.
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
HORIZONTAL LOCAL DOSE DISTRIBUTION
AutoCAD SHX Text
INCHES
AutoCAD SHX Text
INCHES
AutoCAD SHX Text
RADIATION SCATTER
AutoCAD SHX Text
THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT GANTRY IS SHIPPED WITH THE DETECTORS INSTALLED. TO AVOID PERMANENT DAMAGE TO THE CRYSTALS DURING SHIPMENT, THE TEMPERATURE TOLERANCE MUST REMAIN BETWEEN 50°F-86°FAND RELATIVE HUMIDITY OF 20% TO 75%. A BAROMETRIC PRESSURE: 10.9 TO 15.4 psi. THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT MAXIMUM STORAGE TIME IS SIX MONTHS. BIOGRAPH mCT FLOW OR BIOGRAPH mCT COMPONENTS HAVE TO BE STORED IN THE RESPECTIVE TRANSPORT PACKAGING. IF THE ORIGINAL PACKAGING PROVIDED IS NOT AVAILABLE, EQUIVALENT PACKAGING WILL NEED TO BE USED FOR INTERMEDIATE STORAGE. WHEN MOVING THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT GANTRY FROM A COLD ENVIRONMENT INTO A WARM ROOM, ALLOW THE SYSTEM TO STABILIZE AT ROOM TEMPERATURE FOR 24 HOURS BEFORE OPENING.
AutoCAD SHX Text
STORAGE CONSIDERATIONS
AutoCAD SHX Text
SIEMENS REMOTE SERVICE
AutoCAD SHX Text
TO ENSURE THE UPTIME OF YOUR SYSTEM DURING THE WARRANTY PERIOD (AND BEYOND WITH A SERVICE AGREEMENT), SIEMENS REMOTE SERVICES (SRS) REQUIRES REMOTE LOCAL AREA NETWORK ACCESS TO SIEMENS SYSTEMS.
AutoCAD SHX Text
THE PREFERRED CONNECTION METHOD IS (VPN) VIRTUAL PRIVATE NETWORK (WHERE THE CUSTOMER HAS AVAILABLE A VPN CAPABLE FIREWALL OR OTHER VPN APPLIANCE). THIS METHOD PROVIDES THE POSSIBILITY FOR REMOTE SYSTEM DIAGNOSTICS WITHOUT ADDITIONAL HARDWARE. PLEASE CONTACT SIEMENS REMOTE SERVICES (800-888-SIEM) TO DETERMINE IF THIS METHOD IS SUITABLE FOR YOUR SITE.
AutoCAD SHX Text
CUSTOMER VPN SERVER
AutoCAD SHX Text
CUSTOMER DMZ
AutoCAD SHX Text
IMAGING DEVICE SINGLE HOST IP OR IP SUBNET LIST
AutoCAD SHX Text
SIEMENS REMOTE SERVICES DMZ
AutoCAD SHX Text
PUBLIC INTERNET
AutoCAD SHX Text
FIREWALL
AutoCAD SHX Text
SRS ACCESS SERVER
AutoCAD SHX Text
CISCO 3745 ROUTER
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
2 11
AutoCAD SHX Text
FINISHED ROOM HEIGHT
AutoCAD SHX Text
FOR BIOGRAPH mCT GANTRY ONLY
AutoCAD SHX Text
CAREVISION MONITOR/CEILING MOUNT
AutoCAD SHX Text
RECOMMEND 8'-0" MINIMUM 7'-5"
AutoCAD SHX Text
MIN. 8'-6 3/8" MAX. 12'-1 11/16"
AutoCAD SHX Text
CEILING MOUNT INJECTOR
AutoCAD SHX Text
MINIMUM 8'-9 1/2"
AutoCAD SHX Text
TRANSPORT AND DELIVERY
AutoCAD SHX Text
TOTAL CT GANTRY TRANSPORT DEVICE WEIGHT: 5,267 LBS.
AutoCAD SHX Text
CT GANTRY WITHOUT COVERS AND TRANSPORT DEVICE: 5,026 LBS. CT TRANSPORT DEVICE: 417 LBS.
AutoCAD SHX Text
NORMAL TRANSPORT REQUIREMENTS:
AutoCAD SHX Text
NARROW SPACE TRANSPORT REQUIREMENTS:
AutoCAD SHX Text
DURING THE MOVEMENT OF THE GANTRY THROUGH CORRIDORS THE TRANSPORT CASTERS ARE SWIVELED OUT FOR STABILITY. SEE MAXIMUM WIDTH AND MINIMUM LENGTH ABOVE FOR TRANSPORT CASTERS SWIVELED OUT.
AutoCAD SHX Text
WHEN TRANSPORTING THE GANTRY THROUGH A NARROW SPACE OR DOORWAY THE TRANSPORT CASTERS ARE SWIVELED IN AS SHOWN IN THIS SKETCH. AS SOON AS THE SYSTEM PASSES THROUGH THE NARROW SPACE THE TRANSPORT CASTER MUST BE SWIVELED OUT TO AVOID TIPPING HAZARD.
AutoCAD SHX Text
FLOOR LOAD DURING TRANSPORT FOR BOTH GANTRIES:
AutoCAD SHX Text
ACCESS FLOORS HAVE TO BE DESIGNED FOR A WEIGHT CAPACITY OF A MINIMUM OF 882 LBS. PER SLAB/PLATE. DURING TRANSPORT OF THE GANTRIES, THE LOAD MAY BE HIGHER AT CERTAIN INDIVIDUAL POINTS DUE TO UNEVEN FLOORING. IF REQUIRED, COVER THE TRANSPORT ROUTE WITH METAL SHEETS FOR LOAD DISTRIBUTION.
AutoCAD SHX Text
TOTAL PET GANTRY TRANSPORTING DEVICE AND STABILIZER WEIGHT: 2,830 LBS.
AutoCAD SHX Text
PET GANTRY TRANSPORT DEVICE: 532 LBS. PET GANTRY STABILIZER: 100 LBS.
AutoCAD SHX Text
LIFTING GANTRIES WITH CRANE:
AutoCAD SHX Text
IF GANTRIES NEED TO BE LIFTED CRANING BASKET MUST BE USED.
AutoCAD SHX Text
NORMAL TRANSPORT REQUIREMENTS:
AutoCAD SHX Text
WHEN TRANSPORTING THE GANTRY THROUGH A NARROW SPACE OR DOORWAY THE TRANSPORT CASTERS ARE SWIVELED IN AS SHOWN IN THIS SKETCH.

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

S-101ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
ORIENTATION POINT
AutoCAD SHX Text
SCANNER ROOM
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
S-101
AutoCAD SHX Text
1
AutoCAD SHX Text
S-101
AutoCAD SHX Text
3
AutoCAD SHX Text
S-101
AutoCAD SHX Text
2
AutoCAD SHX Text
S-101
AutoCAD SHX Text
4
AutoCAD SHX Text
NOTE: FOR THE WEIGHTS OF ALL SIEMENS EQUIPMENT SHOWN ON THIS PLAN, SEE THE "EQUIPMENT LEGEND" ON SHEET A-101.
AutoCAD SHX Text
S-101
AutoCAD SHX Text
5
AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
STRUCTURAL FLOOR PLAN
AutoCAD SHX Text
SCALE: 1/4" = 1'-0"
AutoCAD SHX Text
D
AutoCAD SHX Text
C
AutoCAD SHX Text
B
AutoCAD SHX Text
A
AutoCAD SHX Text
63
AutoCAD SHX Text
90
AutoCAD SHX Text
116
AutoCAD SHX Text
89
AutoCAD SHX Text
CT MEASUREMENT RESULTS:
AutoCAD SHX Text
MEASUREMENT POINTS (ADJUSTABLE FOOT)
AutoCAD SHX Text
LOAD (STATIC) POUNDS
AutoCAD SHX Text
AMPLITUDE (DYNAMIC) POUNDS
AutoCAD SHX Text
TOTAL
AutoCAD SHX Text
1060
AutoCAD SHX Text
1518
AutoCAD SHX Text
1053
AutoCAD SHX Text
1052
AutoCAD SHX Text
STATIC FLOOR LOADING DUE TO GANTRY'S OWN WEIGHT
AutoCAD SHX Text
DIFFERENCE BETWEEN MINIMUM AND MAXIMUM FLOOR LOADING DURING GANTRY ROTATION
AutoCAD SHX Text
STATIC LOAD
AutoCAD SHX Text
AMPLITUDE
AutoCAD SHX Text
DESCRIPTION:
AutoCAD SHX Text
4883
AutoCAD SHX Text
H
AutoCAD SHX Text
G
AutoCAD SHX Text
F
AutoCAD SHX Text
E
AutoCAD SHX Text
PET MEASUREMENT RESULTS:
AutoCAD SHX Text
MEASUREMENT POINTS (ADJUSTABLE FOOT)
AutoCAD SHX Text
TOTAL
AutoCAD SHX Text
658
AutoCAD SHX Text
504
AutoCAD SHX Text
636
AutoCAD SHX Text
488
AutoCAD SHX Text
2286
AutoCAD SHX Text
A
AutoCAD SHX Text
G
AutoCAD SHX Text
H
AutoCAD SHX Text
D
AutoCAD SHX Text
B
AutoCAD SHX Text
E
AutoCAD SHX Text
F
AutoCAD SHX Text
C
AutoCAD SHX Text
NOTES:
AutoCAD SHX Text
1) THE VALUES PROVIDED FOR FLOOR LOADING APPLY ONLY IF THE GANTRY IS SATISFACTORILY LEVELED.
AutoCAD SHX Text
2) THE FLOOR STRUCTURE MUST BE CAPABLE OF WITHSTANDING THE OCCUPIED WEIGHT OF THE GANTRY AND THE INDIVIDUAL CONTACT AREA LOADING.
AutoCAD SHX Text
FLOOR LOADING
AutoCAD SHX Text
LOAD (STATIC) POUNDS
AutoCAD SHX Text
3" DIAMETER OPENINGS IN THE BACK OF THE CABINET FOR CABLE INLET.
AutoCAD SHX Text
BOLTING THE POWER DISTRIBUTION UNIT TO THE FLOOR IS ONLY NECESSARY WHEN LOCAL OR NATIONAL REGULATIONS REQUIRE IT. (EXAMPLE: EARTHQUAKE ZONES). BOLT THE CABINET TO THE FLOOR USING THE L-BRACKETS THAT ARE INCLUDED IN WITH THE PDU AND ANCHOR THROUGH THE DRILL HOLES IN THE FLOOR PLATE. MATERIALS FOR BOLTING MUST BE SUPPLIED ON-SITE. CUSTOMER/CONTRACTOR TO PROVIDE (4) EYE BOLTS AND (2) RATCHET STRAPS FOR MOUNTING THE ACS/PRS CPU'S ON TOP OF THE THE PDU CABINET ONLY IF SPACE IS LIMITED. SIEMENS INSTALLERS TO MOUNT THE ACS/PRS ON TOP OF PDU CABINET.
AutoCAD SHX Text
BOTTOM
AutoCAD SHX Text
(4) PDU FEET
AutoCAD SHX Text
POWER DISTRIBUTION UNIT (PDU)
AutoCAD SHX Text
(4) PDU ROLLERS
AutoCAD SHX Text
FRONT VIEW
AutoCAD SHX Text
REAR VIEW
AutoCAD SHX Text
TOP VIEW
AutoCAD SHX Text
BOTTOM VIEW
AutoCAD SHX Text
(6) 5/8" HOLES
AutoCAD SHX Text
SCALE: 1/2"=1'-0"
AutoCAD SHX Text
PDU CABINET
AutoCAD SHX Text
2
AutoCAD SHX Text
HEAT EXCHANGER CABINET (S) (WATER/AIR SPLIT COOLING SYSTEM) BOLTING THE HEAT EXCHANGER CABINET TO THE FLOOR IS ONLY NECESSARY WHEN LOCAL OR NATIONAL REGULATIONS REQUIRE IT. (EXAMPLE: EARTHQUAKE ZONES). BOLT THE CABINET TO THE FLOOR USING ANCHORS THROUGH THE DRILL HOLES IN THE FLOOR PLATE. MATERIALS FOR BOLTING MUST BE SUPPLIED ON-SITE.
AutoCAD SHX Text
SCALE: 1/2"=1'-0"
AutoCAD SHX Text
HE CABINET BASE
AutoCAD SHX Text
FRONT
AutoCAD SHX Text
(4) 5/8" HOLES
AutoCAD SHX Text
OPENINGS IN CABINET BASE FOR CABLE/HOSE INLET
AutoCAD SHX Text
4
AutoCAD SHX Text
BOLTING THE POWER DISTRIBUTION CABINET TO THE FLOOR IS ONLY NECESSARY WHEN LOCAL OR NATIONAL REGULATIONS REQUIRE IT. (EXAMPLE: EARTHQUAKE ZONES). BOLT THE CABINET TO THE FLOOR USING ANCHORS THROUGH THE DRILL HOLES IN THE FLOOR PLATE. MATERIALS FOR BOLTING MUST BE SUPPLIED ON-SITE.
AutoCAD SHX Text
POWER DISTRIBUTION CABINET (PDC)
AutoCAD SHX Text
(4) 9/16" HOLES
AutoCAD SHX Text
OPENING IN CABINET BASE FOR CABLE INLET
AutoCAD SHX Text
SCALE: 1/2"=1'-0"
AutoCAD SHX Text
PDC CABINET BASE
AutoCAD SHX Text
FRONT
AutoCAD SHX Text
CABLE STORAGE
AutoCAD SHX Text
3
AutoCAD SHX Text
STRUCTURAL NOTES
AutoCAD SHX Text
1) THE CUSTOMER/CONTRACTOR SHALL FURNISH AND INSTALL ALL STRUCTURAL SUPPORT MEMBERS AND NEEDED HARDWARE FOR THE INSTALLATION OF THE SIEMENS EQUIPMENT. 2) THE OVERHEAD STRUCTURAL SUPPORT SYSTEM SHALL BE FIXED, RIGID AND BRACED FOR SWAY. 3) ALL STRUCTURAL SUPPORT MEMBERS SHALL BE TRUE, SQUARE, LEVEL, PARALLEL AND COPLANAR WITH RESPECT TO EACH OTHER, WITH A HORIZONTAL STRUCTURAL SUPPORT MEMBER TO BE LOCATED AND SET WITH A TRANSIT. 4) ALL STRUCTURAL SUPPORT DETAILS SHOWN ARE SAMPLE DETAILS BASED UPON TYPICAL AND STANDARD BUILDING PRACTICES AND ARE NOT INTENDED AS ACTUAL CONSTRUCTION DETAILS. ALL CONSTRUCTION DETAILS AND SUPPORT CALCULATIONS SHALL BE PREPARED BY A PROFESSIONAL STRUCTURAL ENGINEER AT THE CUSTOMER'S EXPENSE. IN THE EVENT AN EXISTING SUPPORT SYSTEM IS TO BE USED, IT WILL BE THE CUSTOMER'S RESPONSIBILITY TO VERIFY THE INTEGRITY OF THAT SYSTEM. 5) MOUNTING PLATES, FRAMES, AND HARDWARE SUPPLIED BY SIEMENS AS DETAILED IN THIS DRAWING SET ARE INSTALLED BY SIEMENS UNLESS OTHERWISE REQUIRED. ANY DEVIATION FROM THE PROVIDED MATERIALS OR MOUNTING METHODS MUST BE DESIGNED AND DOCUMENTED BY THE STRUCTURAL ENGINEER OF RECORD. ALTERNATE MOUNTING MATERIALS (I.E. ANCHORS, THREADED ROD, BACKING PLATES, ETC.) MUST BE SUPPLIED BY THE CUSTOMER/CONTRACTOR. SIEMENS MAY REQUIRE ASSISTANCE FROM THE CUSTOMER/CONTRACTOR WITH INSTALLATION WHEN UTILIZING ALTERNATE MOUNTING MATERIALS. 6) ALL CEILING FIXTURES (I.E. AIR SUPPLY GRILLES, AIR RETURN GRILLES, EXHAUST GRILLES, SPRINKLER HEADS, INCANDESCENT AND FLUORESCENT LIGHT FIXTURES, INTERCOM SPEAKERS, MEDICAL GAS COLUMNS, ETC.) SHALL BE INSTALLED FLUSH MOUNTED WITH THE FINISHED CEILING TO PROVIDE FREE AND UNRESTRICTED TRAVEL OF THE SMS CEILING MOUNTED EQUIPMENT. 7) THE BOTTOM SIDE OF THE UNISTRUT CEILING GRID AND ANY CEILING MOUNTED SUPPORT PLATES ARE TO BE INSTALLED FLUSH WITH THE FINISHED CEILING. THE CUSTOMER/CONTRACTOR SHALL ALSO PROVIDE COVERSTRIPS FOR THE UNISTRUT. 8) THE STRUCTURAL PLANNING AS SHOWN ON THE 1/4" STRUCTURAL PLAN HAS BEEN COORDINATED WITH THE EQUIPMENT LOCATION AS SHOWN ON THE 1/4" EQUIPMENT LAYOUT PLAN. FOR THIS REASON, ANY DEVIATIONS FROM THE STRUCTURAL PLANNING AS SHOWN MUST BE APPROVED BY SMS PLANNING DEPARTMENT. 9) THE STRUCTURAL ENGINEER OF RECORD SHALL BE RESPONSIBLE FOR THE DESIGN AND DETAIL OF FLOOR, WALL, AND CEILING STRUCTURES IN ACCORDANCE WITH THE STRUCTURAL INFORMATION SHOWN, AND LOCAL GOVERNING BUILDING CODES.
AutoCAD SHX Text
THE GANTRY AND PATIENT TABLE (PHS) WILL BE SECURELY MOUNTED TO THE FLOOR. REFER TO THE FLOOR REQUIREMENT NOTE LOCATED ON THE S-501 SHEET. IT IS THE CUSTOMER'S RESPONSIBILITY TO MEET THE FLOOR REQUIREMENTS SPECIFICATIONS. THE WEIGHT CAPACITY OF THE FLOOR MUST BE EVALUATED BY A STRUCTURAL ENGINEER. THE GANTRY AND PATIENT TABLE (PHS) IS FASTENED TO THE FLOOR AND/OR MACHINE BASE PAD (EPOGROUT 758 OR MASTERFLOW 648 CP) WITH GRADE 5, 1/2-13 UNC-2A THREADED JACKSCREWS SUPPLIED BY SIEMENS, USING POWERS AC100+ GOLD ANCHOR ADHESIVE. MINIMUM ALLOWABLE TENSION LOAD PER SCREW TO BE 1000 LBS. SUPPLIED BY SIEMENS. POSITIONING AND LEVELING THE CT GANTRY HEIGHT IS ADJUSTED WITH 4 ADJUSTABLE FEET FASTENED TO 4 MOUNTING BLOCK. THE PET GANTRY HEIGHT IS SECURELY FASTENED TO THE PET RAILS, WHICH ARE SECURELY FASTENED TO 2 MOUNTING BLOCKS. THE PATIENT TABLE (PHS) IS ADJUSTED THROUGH THE 34 ADJUSTABLE JACKSCREWS. SECURING THE PATIENT TABLE (PHS) TO THE FLOOR IS MANDATORY. DIMENSIONS/MOUNTING HOLES LOCATIONS THE SIEMENS PROJECT MANAGER SHOULD REFER TO THE BIOGRAPH mCT LOAD DRAWINGS FOR MOUNTING HOLE LOCATIONS FOR THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT. SIEMENS INSTALLERS ARE RESPONSIBLE FOR DRILLING THE GANTRY AND PATIENT TABLE (PHS) HOLES PER THE DRILL TEMPLATE THAT IS INCLUDED WITH THE DELIVERY MATERIALS.
AutoCAD SHX Text
THE CUSTOMER/CONTRACTOR WHEN CORE DRILLING FOR PIPES AND CONDUITS, IT IS IMPORTANT TO AVOID THE AREAS UNDER THE CT GANTRY, PET GANTRY AND PATIENT TABLE (PHS) WHERE THERE ARE BOLT HOLES AS OUTLINE IN THIS DETAIL. CORE DRILLING IN THE NO CORE ZONES WILL WEAKEN THE STRUCTURAL INTEGRITY OF THE MOUNT POINT WITHIN THE ZONES. A 7.5" RADIUS IS REQUIRED TO BE HELD FROM THE EDGE OF ANY FLOOR BOLT HOLE.
AutoCAD SHX Text
THE GANTRY AND PATIENT TABLE ARE SUPPLIED AND INSTALLED BY SIEMENS.
AutoCAD SHX Text
NO CORE DRILL ZONE
AutoCAD SHX Text
NO CORE DRILL ZONE
AutoCAD SHX Text
PET GANTRY - (2) MOUNTING BLOCKS - 5/8"
AutoCAD SHX Text
MAXIMUM EXTRACTION FORCE FOR THE POINTS WHERE THE PATIENT TABLE (PHS) IS ATTACHED AS WELL AS FOR THE EXISTING MOUNTING FRAME AND RAISED OR DOUBLE FLOORING. ACCORDING TO THE IEC-60601-1 A SAFETY FACTOR OF 4 HAS TO BE OBSERVED.
AutoCAD SHX Text
1
AutoCAD SHX Text
PET RAILS - FASTENED TO THE (2) PET GANTRY MOUNTING BLOCKS
AutoCAD SHX Text
2
AutoCAD SHX Text
CT GANTRY - (4) ADJUSTABLE FEET FASTENED TO (4) MOUNTING BLOCKS - 5/8"
AutoCAD SHX Text
3
AutoCAD SHX Text
PHS - ADJUSTED THROUGH (34) JACKSCREWS - 5/8"
AutoCAD SHX Text
4
AutoCAD SHX Text
NO CORE DRILL ZONES
AutoCAD SHX Text
5
AutoCAD SHX Text
OUTLINE OF MACHINE BASE PAD (EPOGROUT 758)
AutoCAD SHX Text
SIEMENS PROJECT MANAGERS CAN ORDER A FULL SIZE TEMPLATE DISPLAYING: SCANNER OUTLINE ANCHOR LOCATIONS CABLE ENTRANCES MACHINE BASE PAD OUTLINE (EPOGOUT - IF NEEDED) THIS TEMPLATE CAN BE ORDERED BY THE TYPICAL DRAWING NUMBER FOR THIS SYSTEM.
AutoCAD SHX Text
OUTLINE OF GANTRY
AutoCAD SHX Text
OUTLINE OF PATIENT TABLE (PHS)
AutoCAD SHX Text
1
AutoCAD SHX Text
1
AutoCAD SHX Text
2
AutoCAD SHX Text
2
AutoCAD SHX Text
3
AutoCAD SHX Text
3
AutoCAD SHX Text
3
AutoCAD SHX Text
3
AutoCAD SHX Text
5
AutoCAD SHX Text
5
AutoCAD SHX Text
4
AutoCAD SHX Text
5
AutoCAD SHX Text
5
AutoCAD SHX Text
GANTRY AND PATIENT TABLE (PHS) EQUIPMENT MOUNTING SPECIFICATIONS:
AutoCAD SHX Text
BOLTING REQUIREMENTS:
AutoCAD SHX Text
NO CORE DRILL ZONE REQUIREMENTS:
AutoCAD SHX Text
SCALE: 1/2"=1'-0"
AutoCAD SHX Text
GANTRY AND PATIENT TABLE (PHS) MOUNTING DETAIL
AutoCAD SHX Text
NO CORE DRILL ZONE
AutoCAD SHX Text
NO CORE DRILL ZONE
AutoCAD SHX Text
NO CORE DRILL ZONE
AutoCAD SHX Text
NO CORE DRILL ZONE
AutoCAD SHX Text
1
AutoCAD SHX Text
RTP PALLET WALL MOUNT HANGER
AutoCAD SHX Text
(4) 1/4" HOLES
AutoCAD SHX Text
SCALE: 1/2"=1'-0"
AutoCAD SHX Text
RTP PALLET WALL MOUNT HANGER
AutoCAD SHX Text
SIEMENS SUPPLIED WALL MOUNT HANGER AND RADIATION THERAPY PALLET (RTP PALLET). THE RTP PALLET WALL MOUNT HANGER REQUIRES TO BE MOUNTED TO WALL. CUSTOMER/CONTRACTOR INSTALLS AND SUPPLIES THE MOUNTING HARDWARE. RTP PALLET WALL MOUNT HANGER SHOULD BE MOUNTED APPROXIMATELY 50" A.F.F.
AutoCAD SHX Text
5
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
3 11
AutoCAD SHX Text
FINISHED ROOM HEIGHT
AutoCAD SHX Text
FOR BIOGRAPH mCT GANTRY ONLY
AutoCAD SHX Text
CAREVISION MONITOR/CEILING MOUNT
AutoCAD SHX Text
RECOMMEND 8'-0" MINIMUM 7'-5"
AutoCAD SHX Text
MIN. 8'-6 3/8" MAX. 12'-1 11/16"
AutoCAD SHX Text
CEILING MOUNT INJECTOR
AutoCAD SHX Text
MINIMUM 8'-9 1/2"

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

S-102ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
ORIENTATION POINT
AutoCAD SHX Text
SCANNER ROOM
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
S-102
AutoCAD SHX Text
3
AutoCAD SHX Text
S-102
AutoCAD SHX Text
2
AutoCAD SHX Text
S-102
AutoCAD SHX Text
1
AutoCAD SHX Text
NOTE: FOR THE WEIGHTS OF ALL SIEMENS EQUIPMENT SHOWN ON THIS PLAN, SEE THE "EQUIPMENT LEGEND" ON SHEET A-101.
AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
STRUCTURAL CEILING PLAN
AutoCAD SHX Text
SCALE: 1/4" = 1'-0"
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
4 11
AutoCAD SHX Text
FINISHED ROOM HEIGHT
AutoCAD SHX Text
FOR BIOGRAPH mCT GANTRY ONLY
AutoCAD SHX Text
CAREVISION MONITOR/CEILING MOUNT
AutoCAD SHX Text
RECOMMEND 8'-0" MINIMUM 7'-5"
AutoCAD SHX Text
MIN. 8'-6 3/8" MAX. 12'-1 11/16"
AutoCAD SHX Text
CEILING MOUNT INJECTOR
AutoCAD SHX Text
MINIMUM 8'-9 1/2"
AutoCAD SHX Text
CARE VISION OVERHEAD SUPPORT SIDE ELEVATION
AutoCAD SHX Text
NONE
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REV 0
AutoCAD SHX Text
FROM AFF TO UNFINISHED CEILING
AutoCAD SHX Text
8'-6 3/8" TO 10'-2"
AutoCAD SHX Text
ROOM HEIGHT WITH SHORT FLANGE PIPE
AutoCAD SHX Text
UNFINISHED CEILING
AutoCAD SHX Text
INTERMEDIATE CEILING
AutoCAD SHX Text
FINISHED FLOOR
AutoCAD SHX Text
1) IF THE ROOM HEIGHT DOES NOT FALL WITHIN THE RECOMMENDED RANGE, THE MONITOR VIEW ZONE DOES NOT COVER THE ENTIRE RANGE OF THE PREFERRED 4'-7" TO 5'-7" RANGE. IN THIS CASE, WE RECOMMEND USING THE MONITOR TROLLEY. 2) A FREE SPACE OF AT LEAST 2" BETWEEN THE HIGHEST GANTRY POINT (EX. GANTRY TILT, OPENED SERVICE LID, ETC) AND THE LOWEST MCS CONTACT POINT MUST BE MAINTAINED. OTHERWISE ONLY THE MONITOR TROLLEY IS PERMITTED. 3) SHORTENING OF THE FLANGE PIPES IS NOT PERMITTED.
AutoCAD SHX Text
10'-2" TO 12'-1 11/16"
AutoCAD SHX Text
ROOM HEIGHT WITH LONG FLANGE PIPE
AutoCAD SHX Text
RECOMMENDED RANGE NOTE 1)
AutoCAD SHX Text
2'-7 3/8"
AutoCAD SHX Text
LONG FLANGE
AutoCAD SHX Text
SHORT FLANGE
AutoCAD SHX Text
LONG FLANGE
AutoCAD SHX Text
SHORT FLANGE
AutoCAD SHX Text
2'-9 3/4"
AutoCAD SHX Text
HIGHEST POSITION
AutoCAD SHX Text
LOWEST POSITION
AutoCAD SHX Text
LONG FLANGE
AutoCAD SHX Text
SHORT FLANGE
AutoCAD SHX Text
10 3/4"
AutoCAD SHX Text
6'-7"
AutoCAD SHX Text
LONG FLANGE
AutoCAD SHX Text
SHORT FLANGE
AutoCAD SHX Text
NOTE 2)
AutoCAD SHX Text
3
AutoCAD SHX Text
MEDRAD INJECTOR (CL)
AutoCAD SHX Text
CEILING MOUNTING PLATE DETAIL
AutoCAD SHX Text
NONE
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
METAL FRAMEWORK
AutoCAD SHX Text
DECK ABOVE
AutoCAD SHX Text
CEILING
AutoCAD SHX Text
UNISTRUT
AutoCAD SHX Text
1
AutoCAD SHX Text
FOUR 3/8" DIA. A325 BOLTS (SUPPLIED BY MEDRAD) FOR MOUNTING THE COUNTERPOISE MOUNTING FLANGE TO THE MOUNTING PLATE
AutoCAD SHX Text
FOUR 1/2" DIA. A307 BOLTS FOR ANCHORING THE MOUNTING PLATE TO UNISTRUT OR SUPPORT FRAME, SUPPLIED AND INSTALLED BY CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
3/8" THICK STEEL MOUNTING PLATE, SUPPLIED BY MEDRAD AND INSTALLED BY CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
TOTAL HUNG WEIGHT OF THE COUNTERPOISE SYSTEM AND INJECTOR: 106 LBS.
AutoCAD SHX Text
MOUNTING PLATE, SUPPLIED BY MEDRAD, INSTALLED BY CUSTOMER/CONTRACTOR. LOCATED EVEN WITH CEILING IF STANDARD MOUNT IS USED.
AutoCAD SHX Text
THE MOUNTING PLATE MUST BE INSTALLED ABOVE THE FINISHED CEILING. THERE SHOULD BE SPACE ABOVE THE PLATE TO ACCESS NUTS AND LOCK WASHERS DURING INSTALLATION.
AutoCAD SHX Text
MINIMUM
AutoCAD SHX Text
33 1/2" MEDIUM MOUNT (OPTION)
AutoCAD SHX Text
39 3/8" LONG MOUNT (OPTION)
AutoCAD SHX Text
GANTRY HEIGHT +9" MIN. ABOVE FINISHED FLOOR RECOMMENDED
AutoCAD SHX Text
REV 1
AutoCAD SHX Text
SUPPORT STRUCTURE FOR CEILING MOUNTING PLATE MUST BE DESIGNED BY AN ENGINEER, BRACED FOR SWAY, AND INSTALLED BY A QUALIFIED CONTRACTOR.
AutoCAD SHX Text
PULL BOX "IN3"
AutoCAD SHX Text
CEILING MOUNT ASSEMBLY
AutoCAD SHX Text
HORIZONTAL SUPPORT ARM
AutoCAD SHX Text
FINISHED FLOOR
AutoCAD SHX Text
MEDRAD INJECTOR
AutoCAD SHX Text
STANDARD MOUNT
AutoCAD SHX Text
MOUNTING FLANGE
AutoCAD SHX Text
2
AutoCAD SHX Text
CARE VISION MONITOR CEILING MOUNT DETAIL
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
1" = 1'-0"
AutoCAD SHX Text
THE WEIGHT CAPACITY OF THE CEILING MUST BE EVALUATED BY A STRUCTURAL ENGINEER. 1) SIX SPACERS WITH THREADED BORES FOR MOUNTING THE FLANGE PIPE CEILING PLATE. 2) FOUR THREADED BORES FOR THE THREADED RODS. 3) HOLES FOR MOUNTING TO UNFINISHED CEILING. 4) HOLES FOR THREADED RODS. 5) THREADED RODS, SUPPLIED BY SIEMENS AND INSTALLED BY CUSTOMER/CONTRACTOR. 6) CANOPY "SMALL" - 1'-1 3/8" "BIG" - 1'-7 11/16""BIG" CANOPY IS MANDATORY FOR UNFINISHED CEILING INSTALLTION. 7) FLANGE PIPE IS ALREADY ASSEMBLED TO THE SUPPORT ARM WHEN DELIVERED. "SHORT" FLANGE PIPE: LENGTH= 11 3/16", AVAILABLE FOR ROOMS WITH A CEILING HEIGHT UP TO 10'-2". "LONG" FLANGE PIPE: LENGTH= 2'-7 1/2", AVAILABLE FOR ROOMS WITH A CEILING HEIGHT ABOVE 10'-2". 8) SPACER SLEEVES CONTAINING THREADED RODS BY SIEMENS AND INSTALLED BY THE CUSTOMER/CONTRACTOR. THREADED RODS ARE SUPPLIED IN 3'-3 3/8" LENGTHS AND MAY BE SHORTENED TO ACCOMMODATE CEILING HEIGHT. 9) SPACER PLATE: ALWAYS IN THE CENTER OF THE SUBCONSTRUCTION. NOT NEEDED IF TOTAL SUBCONSTRUCTION LENGTH IS <1'-7 11/16". 10) HOLES FOR MOUNTING CEILING PLATE TO THE INTERFACE PLATE. 11) THREADED BORE FOR M16 FOR TAP BOLT (HORIZONTAL ADJUSTMENT) 12) THREADED BORE M6 FOR GROUNDING CONNECTION. THE SYSTEM CABLES ARE ROUTED THROUGH THE FLANGE PIPE. A DRILLING TEMPLATE IS INCLUDED IN THE DELIVERY VOLUME FOR THE ANCHOR PLATE. 6 HILTI PASS-THROUGH ANCHORS ARE INCLUDED IN THE DELIVERY VOLUME, HSL-3-G M12/50, SUPPLIED BY SIEMENS, EXTRACTION FORCE IS MIN. 1,349 LBS. AT EACH CEILING MOUNTING POINT.
AutoCAD SHX Text
UNFINISHED CEILING INSTALLATION
AutoCAD SHX Text
INTERMEDIATE CEILING INSTALLATION
AutoCAD SHX Text
INTERFACE PLATE, 1/2" THICK, SUPPLIED BY SIEMENS AND INSTALLED BY THE CUSTOMER/ CONTRACTOR.
AutoCAD SHX Text
CEILING PLATE, 1/2" THICK, SUPPLIED AND INSTALLED BY SIEMENS.
AutoCAD SHX Text
ANCHOR PLATE, 1/2" THICK, SUPPLIED BY SIEMENS AND INSTALLED BY THE CUSTOMER/ CONTRACTOR.
AutoCAD SHX Text
1)
AutoCAD SHX Text
2)
AutoCAD SHX Text
5/8" (4)
AutoCAD SHX Text
1 1/2" (6)
AutoCAD SHX Text
INTERFACE PLATE
AutoCAD SHX Text
5)
AutoCAD SHX Text
7)
AutoCAD SHX Text
6)
AutoCAD SHX Text
UNFINISHED CONCRETE CEILING OR UNISTRUT
AutoCAD SHX Text
3)
AutoCAD SHX Text
4)
AutoCAD SHX Text
1 3/16" (6)
AutoCAD SHX Text
M16 (6)
AutoCAD SHX Text
M12 (4)
AutoCAD SHX Text
SYSTEM CABLES
AutoCAD SHX Text
9)
AutoCAD SHX Text
CEILING PLATE, 1/2" THICK, SUPPLIED AND INSTALLED BY SIEMENS. MOUNTED FLUSH WITH THE INTERMEDIATE CEILING.
AutoCAD SHX Text
INTERFACE PLATE
AutoCAD SHX Text
SYSTEM CABLES
AutoCAD SHX Text
DROP CEILING
AutoCAD SHX Text
8)
AutoCAD SHX Text
6)
AutoCAD SHX Text
7)
AutoCAD SHX Text
MIN. 4" TO MAX. 3'-3 9/16"
AutoCAD SHX Text
IF L1 IS <1'-7 11/16": L2 = L1 - 3/16" L3 = L1 - 3 3/8" IF L1 IS >1'-7 11/16": L2 = L1 - 3/16" L3 = L1 - 5" L1 - 5" 2
AutoCAD SHX Text
L1: CEILING DISTANCE (DISTANCE BETWEEN CONCRETE CEILING AND INTERMEDIATE CEILING = TOTAL LENGTH OF SUBCONSTRUCTION. L2: REQUIRED LENGTH OF THREADED RODS L3: REQUIRED LENGTH OF SPACER SLEEVES
AutoCAD SHX Text
13/16" (6)
AutoCAD SHX Text
M16 (6)
AutoCAD SHX Text
10)
AutoCAD SHX Text
11)
AutoCAD SHX Text
12)
AutoCAD SHX Text
5)
AutoCAD SHX Text
REV 2

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

S-501ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
5 11
AutoCAD SHX Text
FREQUENCY (Hz)
AutoCAD SHX Text
1
AutoCAD SHX Text
CONTINUOUS VIBRATIONS
AutoCAD SHX Text
ALLOWABLE ACCELERATION rms-VALUE m/s²
AutoCAD SHX Text
10
AutoCAD SHX Text
100
AutoCAD SHX Text
0.01
AutoCAD SHX Text
0.1
AutoCAD SHX Text
TRANSIENT VIBRATIONS (SHOCKS)
AutoCAD SHX Text
ANY TRANSIENT VIBRATION HAS TO BE LESS THAN 0.2 m/s² PEAK-TO-PEAK IN THE TIME DOMAIN. THE VIBRATIONS HAVE TO BE MEASURED WITH A SAMPLING RATE OF 1000Hz.
AutoCAD SHX Text
0.001
AutoCAD SHX Text
FLOOR AND BUILDING VIBRATION CAN REDUCE IMAGE QUALITY FOR THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT SYSTEM IN THE THREE SPATIAL DIRECTIONS, ACCELERATION IN VIBRATIONS AT THE MOUNTING POINTS OF THE GANTRY AND THE PATIENT TABLE (PHS) MUST NOT EXCEED THE THRESHOLDS AS DESCRIBED HERE. THE THRESHOLD IS DEFINED AS ACCELERATION rms VALUE (ROOT MEAN SQUARE) IN m/s² OF AN FFT SPECTRUM DERIVED WITH A FREQUENCY RESOLUTION OF 1 Hz AND USING A HANNING-WINDOW. THE VIBRATIONS HAVE TO BE MEASURED WITH A SAMPLING RATE OF 1000Hz USING AN ANTI-ALIASING-FILTER WITH A LIMIT FREQUENCY OF 250Hz. THE THRESHOLD IS VALID FOR VIBRATIONS AT THE INSTALLATION LOCATION WITH A BIOGRAPH mCT FLOW OR BIOGRAPH mCT IN POSITION. MEASUREMENTS MUST BE TAKEN PRIOR TO THE INSTALLATION OF THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT, THEREFORE CHANGES IN THE EIGEN FREQUENCY OF THE SLAB CAUSED BY THE ADDITIONAL MASS OF THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT MUST BE CONSIDERED WHEN COMPARING THE FREQUENCY SPECTRUM WITH THE THRESHOLD. VALUES OF THE THRESHOLD ARE SHOWN BELOW.
AutoCAD SHX Text
0.0001
AutoCAD SHX Text
FLOOR AND BUILDING VIBRATIONS
AutoCAD SHX Text
BIOGRAPH STRUCTURAL BORNE NOISE
AutoCAD SHX Text
THE INHERENT VIBRATION SIGNATURE IMPINGED ON THE BUILDING STRUCTURE BY THE SYSTEM DUE TO CT DRUM ROTATION, ETC...IS DEFINED IN 12660-MER-01S-01-P41-101. THE CT IS PRIMARY DRIVER OF STRUCTURE BORNE NOISE AND IS NOT DAMPED BY THE PET IN ANY WAY AS THE GANTRIES ARE MECHANICALLY DECOUPLED. THE PET AND PHS ARE NOT A SIGNIFICANT SOURCE OF VIBRATION.
AutoCAD SHX Text
THE ENGINEER OF RECORD OF THE BUILDING SHALL PROVIDE SUPPORT STRUCTURE DESIGNED TO SUPPORT ALL WEIGHTS AND FORCES. THE ENGINEER OF RECORD FOR THE BUILDING AND SIEMENS ENGINEERING SHALL JOINTLY REVIEW DEVIATIONS FROM THE FOLLOWING REQUIREMENTS. IT IS THE CUSTOMER'S RESPONSIBILITY TO CONTRACT A QUALIFIED SPECIALIST TO IMPLEMENT SITE MODIFICATIONS THAT MEET THESE SPECIFIC LIMITS AND TO DESIGN STRUCTURAL SOLUTIONS IN CASE OF DEVIATIONS. 1) THE MINIMUM ALLOWABLE CONCRETE THICKNESS FOR NONSEISMIC REGIONS OF THE SCANNER ROOM FLOOR IS 4.5". 2) THE CONDITIONS OF FLOORING, VIBRATION-FREE LOCATION, AND/OR INSTALLATION OF THE GANTRY AND PATIENT TABLE ONLY ON: CONCRETE FLOORING CONCRETE CLASS C20/25 TO C50/60 ACCORDING TO DIN 1045-1. DIN 1045-2 ACCEPTABLE FLOOR STRUCTURAL MATERIALS FOR LOAD BEARING AREAS OF THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT GANTRY AND PHS ARE RESTRICTED TO CONCRETE, STEEL, OR HIGH AGGREGATE EPOXY GROUTS: SUCH AS EPOGROUT 758 OR MASTERFLOW 648. THE FLOOR STRUCTURE FOR THE LOAD BEARING AREAS SHALL NOT CONTAIN COMPLIANT MATERIALS THAT ARE SUBJECT TO MOVEMENT WITH THE PASSAGE OF TIME; SUCH AS LEAD, WOOD OR SAND/MORTAR MIXES. 3) THE CONCRETE PROPERTIES: COMPRESSIVE STRENGTHS: RECOMMENDED CONCRETE IS > 20 MPa > 20 MPa (2,900 psi). COMPRESSIVE MODULUS OF ELASTICITY: CONCRETE SHALL BE > 20684 MPa (3,000,000 psi). > 20684 MPa (3,000,000 psi). . FLEXURAL MODULUS OF ELASTICITY: CONCRETE SHALL BE > 20684 MPa (3,000,000 psi). > 20684 MPa (3,000,000 psi). 20684 MPa (3,000,000 psi). 20684 MPa (3,000,000 psi). CONCRETE MUST BE CURED AT LEAST 28 DAYS PRIOR TO MACHINE INSTALLATION. CONCRETE FLOORING TO BE TESTED BY A STRUCTURAL ENGINEER. 4) THE EVENNESS AND LEVELNESS OF THE FLOOR: LEVELNESS: VARIATION OF THE FLOOR LEVELNESS IN THE GANTRY AND PHS AREAS SHOULD NOT EXCEED .5 INCHES OVER THE ENTIRE FOOTPRINT OF THE SYSTEM. VARIATION IS TO BE MEASURED AT THE GANTRY AND PHS MOUNTING POINTS. 5) THE FLOOR COVERING REQUIREMENTS: FLOOR COVERING: VINYL FLOOR COVERING WITH A MINIMUM STATIC LOAD LIMIT RATING OF 5.2 MPa (750.0 psi REFERENCE ASTM F 970) IS RECOMMENDED UNDER THE GANTRY AND PHS. SOFT FLOOR COVERING ARE SUBJECT TO GRADUAL MOVEMENT (CREEP). UNDER GANTRY: REMOVE ALL FLOOR COVERING IN THE LOAD BEARING AREAS. INSTALLATION OF THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT ON A FLOATING FLOOR WITHOUT SUB-CONSTRUCTIONS IS PROHIBITED. 6) THE MACHINE BASE PAD (OPTIONAL) TO BE CREATED AND ADHERE TO THE CONCRETE FLOOR WITH THE APPROVAL FROM THE SITE ENGINEER OF RECORD USING GROUTING MATERIAL EPOGROUT 758 (L&M CONSTRUCTION CHEMICALS, INC 1-800-362-3331) OR MASTERFLOW 648 CP (BASF 1-800-243-6739) TO ADDRESS THE FOLLOWING CONDITIONS: TO INCREASE THE FLOOR THICKNESS. TO CORRECT LEVELNESS OF THE FLOOR. EXCESSIVE INTERFERENCE BETWEEN MOUNTING ANCHORS AND REBAR. REPAIR CONCRETE HOLES. REQUIRED 3" MACHINE BASE PAD WHEN UTILIZING SURFACE MOUNT DUCT UNDER THE GANTRY. THE MACHINE BASE SHOULD BE MINIMUM OF 1" AND A MAXIMUM OF 2 1/2" THICK, UNLESS UTILIZING SURFACE MOUNT DUCT UNDER THE GANTRY 3" THICK REQUIRED AND APPLIED DIRECTLY TO A CLEAN CONCRETE SURFACE WITH NO INTERMEDIATE MATERIALS BETWEEN THE CONCRETE AND THE EPOGROUT. IF THE MACHINE BASE PAD IS USED, SOLID ALUMINUM SPACERS (SUPPLIED BY THE CUSTOMER) WITH NO INTERMEDIATE MATERIALS WILL BE NEEDED BENEATH THE PET GANTRY SERVICE RAILS. EACH SPACERS MUST HAVE A HEIGHT MATCHING THE HEIGHT OF THE MACHINE BASE PAD AND MUST BE 31" LONG AND 4" WIDE. SITE MANAGER IS RESPONSIBLE FOR HAVING THE SPACERS CUSTOM BUILT TO THESE DIMENSIONS. 7) THE ANCHOR PROPERTIES: TENSION CAPABILITY: ALLOWABLE TENSION LOAD CAPABILITY FOR EMBEDDED CONCRETE ANCHORS SHALL BE GREATER THAT 1000.0 LB. ANCHOR DEPTH: ANCHOR EMBEDMENT DEPTH AND CONCRETE THICKNESS SHALL COMPLY WITH ICBO GUIDELINES FOR THE ANCHOR. 4" DRILL DEPTH TYPICAL FOR ADHESIVE ANCHORING. BIOGRAPH mCT FLOW OR BIOGRAPH mCT SHALL BE FASTENED TO THE FLOOR AND/OR MACHINE BASE PAD WITH GRADE 5, 1/2-13 UNC-2A THREADED FASTENERS AS SUPPLIED BY SIEMENS. MINIMUM EXTRACTION FORCE ACCORDING TO THE IEC 60601-1 SAFETY FACTOR OF 4 HAS TO BE OBSERVED.
AutoCAD SHX Text
FLOOR REQUIREMENTS

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

E-101ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
MMW
AutoCAD SHX Text
MMW
AutoCAD SHX Text
ACS1
AutoCAD SHX Text
PRS
AutoCAD SHX Text
PDC
AutoCAD SHX Text
A
AutoCAD SHX Text
ES
AutoCAD SHX Text
ES2
AutoCAD SHX Text
APD
AutoCAD SHX Text
A1
AutoCAD SHX Text
PDU
AutoCAD SHX Text
HD1
AutoCAD SHX Text
B
AutoCAD SHX Text
HD2
AutoCAD SHX Text
PIT
AutoCAD SHX Text
ICS
AutoCAD SHX Text
IES
AutoCAD SHX Text
PIT2
AutoCAD SHX Text
IN1
AutoCAD SHX Text
IN2
AutoCAD SHX Text
IN3
AutoCAD SHX Text
F1
AutoCAD SHX Text
T/H
AutoCAD SHX Text
AutoCAD SHX Text
H
AutoCAD SHX Text
EPO
AutoCAD SHX Text
DS
AutoCAD SHX Text
EPO
AutoCAD SHX Text
DOOR SWITCHES ARE OPTIONAL (PER LOCAL CODE REQUIREMENTS).
AutoCAD SHX Text
ORIENTATION POINT
AutoCAD SHX Text
EPO
AutoCAD SHX Text
SCANNER ROOM
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
B1
AutoCAD SHX Text
S2
AutoCAD SHX Text
S1
AutoCAD SHX Text
S
AutoCAD SHX Text
A2
AutoCAD SHX Text
DC
AutoCAD SHX Text
DC
AutoCAD SHX Text
1
AutoCAD SHX Text
"GROUND ELECTRODE" FOR ISOLATION TRANSFORMER INSIDE PDU.
AutoCAD SHX Text
2
AutoCAD SHX Text
CONDENSATION HOSE ROUTED ACCORDING TO LOCAL CODE TO SELECTED DRAIN.
AutoCAD SHX Text
M-501
AutoCAD SHX Text
3
AutoCAD SHX Text
26
AutoCAD SHX Text
25
AutoCAD SHX Text
23
AutoCAD SHX Text
24
AutoCAD SHX Text
PPS
AutoCAD SHX Text
E-501
AutoCAD SHX Text
2
AutoCAD SHX Text
E-501
AutoCAD SHX Text
2
AutoCAD SHX Text
DEDICATED POWER SOURCE, SUPPLIED AND INSTALLED BY CUSTOMER/CONTRACTOR. SEE "POWER SCHEDULE".
AutoCAD SHX Text
3
AutoCAD SHX Text
4
AutoCAD SHX Text
4
AutoCAD SHX Text
7
AutoCAD SHX Text
5
AutoCAD SHX Text
8
AutoCAD SHX Text
9
AutoCAD SHX Text
11
AutoCAD SHX Text
12
AutoCAD SHX Text
14
AutoCAD SHX Text
16
AutoCAD SHX Text
21
AutoCAD SHX Text
27
AutoCAD SHX Text
28
AutoCAD SHX Text
29
AutoCAD SHX Text
29
AutoCAD SHX Text
30
AutoCAD SHX Text
31
AutoCAD SHX Text
32
AutoCAD SHX Text
6
AutoCAD SHX Text
10
AutoCAD SHX Text
A4
AutoCAD SHX Text
IRS
AutoCAD SHX Text
IRS1
AutoCAD SHX Text
B3
AutoCAD SHX Text
ICS1
AutoCAD SHX Text
PDC1
AutoCAD SHX Text
PDU1
AutoCAD SHX Text
13
AutoCAD SHX Text
15
AutoCAD SHX Text
17
AutoCAD SHX Text
18
AutoCAD SHX Text
19
AutoCAD SHX Text
20
AutoCAD SHX Text
22
AutoCAD SHX Text
TRANSFORMER INSIDE THE HEAT EXCHANGER IS SUPPLIED BY SIEMENS, INSTALLED BY THE CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
ELECTRICAL RACEWAY PLAN
AutoCAD SHX Text
SCALE: 1/4" = 1'-0"
AutoCAD SHX Text
CONDUIT LENGTH CALCULATIONS
AutoCAD SHX Text
IF SITE SPECIFIC CONDITIONS EXCEED THE FOLLOWING ASSUMED VALUES THEN ADDITIONAL LENGTH MUST BE SUBTRACTED BY THE ELECTRICAL CONTRACTOR FROM THE MAXIMUM CONDUIT LENGTHS LISTED.
AutoCAD SHX Text
IF DUCT LOCATIONS ARE ALTERED FROM THE SHOWN LAYOUT IT IS THE ELECTRICAL CONTRACTORS RESPONSIBILITY TO RECALCULATE THE MAXIMUM CONDUIT LENGTHS.
AutoCAD SHX Text
ASSUMED VALUES USED IN CALCULATING STATED MAXIMUM CONDUIT LENGTHS: VERTICAL DUCTS - 10'-0" FLOOR PENETRATIONS - 3'-0"
AutoCAD SHX Text
ELECTRICAL LEGEND
AutoCAD SHX Text
SYM
AutoCAD SHX Text
SIZE
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
REMARKS
AutoCAD SHX Text
SUPPLIED AND INSTALLED BY CUSTOMER/CONTRACTOR
AutoCAD SHX Text
A
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED FLUSH WITH FINISHED WALL AT FLOOR LINE IN SHOWN LOCATION.
AutoCAD SHX Text
ANCILLARY WIRING
AutoCAD SHX Text
A1
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED FLUSH WITH FINISHED WALL AT FLOOR LINE IN SHOWN LOCATION.
AutoCAD SHX Text
PDU WIRING
AutoCAD SHX Text
A2
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED FLUSH WITH FINISHED WALL AT FLOOR LINE IN SHOWN LOCATION.
AutoCAD SHX Text
HEAT EXCHANGER CABINET-WATER/AIR SPLIT
AutoCAD SHX Text
A4
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED FLUSH WITH FINISHED WALL AT FLOOR LINE IN SHOWN LOCATION. CONNECT 6 FOOT LONG FLEXIBLE WHIP OR WHIPS TO PULL BOX FOR TERMINATION AT THE POWER DISTRIBUTION UNIT (PDU). REFER TO THE POWER SCHEDULE.
AutoCAD SHX Text
DIRECT POWER FEED TO THE POWER DISTRIBUTION UNIT (PDU)
AutoCAD SHX Text
ACS1
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH 6" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
PET ACQUISITION SYSTEM POWER
AutoCAD SHX Text
APD
AutoCAD SHX Text
---
AutoCAD SHX Text
POWER CONNECTION TERMINAL FOR ACS/PRS WITH (1) 208/220/240 L6-20R RECEPTACLE SUPPLIED BY SIEMENS WITH THE ACS/PRS.
AutoCAD SHX Text
POWER DISTRIBUTION FOR ACS/PRS.
AutoCAD SHX Text
B
AutoCAD SHX Text
12" x 12"
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH A 5" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
GANTRY CABLE ACCESS SEE SHEET E-501
AutoCAD SHX Text
B1
AutoCAD SHX Text
8" x 8"
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH A 6" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
WATER HOSES ACCESS SEE SHEET E-501 & M-101
AutoCAD SHX Text
B3
AutoCAD SHX Text
6" x 6"
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH A 3" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
GANTRY CABLE ACCESS DATA/COMMUNICATION SEE SHEET E-501
AutoCAD SHX Text
DC
AutoCAD SHX Text
---
AutoCAD SHX Text
DISCONNECT SWITCH MOUNTED NEAR OUTDOOR UNIT TO DISCONNECT POWER SUPPLY FROM INDOOR UNIT. SUPPLIED BY SIEMENS, INSTALLED BY CUSTOMER/CONTRACTOR IN ACCORDANCE WITH LOCAL CODES.
AutoCAD SHX Text
OUTDOOR UNIT-WATER/AIR SPLIT
AutoCAD SHX Text
EPO
AutoCAD SHX Text
---
AutoCAD SHX Text
EMERGENCY POWER OFF BUTTON THAT PREVENTS RESETTING OF CIRCUIT BREAKER WHEN IN THE OFF POSITION WITH PROTECTIVE COVER, MOUNTED ON WALL AT 5'-0" ABOVE FINISH FLOOR. THERE SHALL BE AN EPO IN EACH ROOM OF THE SUITE WHERE SIEMENS EQUIPMENT IS LOCATED. EXACT LOCATIONS TO BE DETERMINED BY CUSTOMER/CONTRACTOR. SUPPLIED BY CUSTOMER/CONTACTOR.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
ES
AutoCAD SHX Text
---
AutoCAD SHX Text
ETHERNET SWITCH FOR ICS, IRS, GANTRY & PDC SUPPLIED BY SIEMENS. LOCATED INSIDE PDC CABINET.
AutoCAD SHX Text
ES2
AutoCAD SHX Text
---
AutoCAD SHX Text
ETHERNET SWITCH FOR IES, DICOM CAMERAS & NETWORK PRINTER SUPPLIED BY SIEMENS. LOCATED INSIDE PDC CABINET.
AutoCAD SHX Text
F1
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED ABOVE FINISHED CEILING, EXACT LOCATION TO BE COORDINATED WITH SIEMENS PROJECT MANAGER.
AutoCAD SHX Text
CARE VISION MONITOR CEILING MOUNT
AutoCAD SHX Text
ICS
AutoCAD SHX Text
PIT
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH TWO 3" SLEEVES RUNNING THROUGH FLOORSLAB ENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATIONS.
AutoCAD SHX Text
IMAGE CONSTRUCTION SYS. POWER
AutoCAD SHX Text
ICS1
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH 3" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATIONS.
AutoCAD SHX Text
IMAGE CONSTRUCTION SYS. DATA/COMMUNICATION
AutoCAD SHX Text
IES
AutoCAD SHX Text
PIT2
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX OR OPENING IN DUCT LOCATED ADJACENT TO REMOTE IES.
AutoCAD SHX Text
BIOGRAPH ADVANCED WORKFLOW - REMOTE IMAGE EVALUATION SYSTEM
AutoCAD SHX Text
IN1
AutoCAD SHX Text
IN2
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED FLUSH WITH FINISHED WALL IN CONTROL AREA, EXACT LOCATION TO BE COORDINATED WITH SIEMENS PROJECT MANAGER.
AutoCAD SHX Text
INJECTOR ELECTRONICS
AutoCAD SHX Text
IN3
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED ABOVE FINISHED CEILING IN SHOWN LOCATION.
AutoCAD SHX Text
CEILING MTD. INJECTOR
AutoCAD SHX Text
IRS
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH 4" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
IMAGE RECONSTRUCTION CAB. POWER
AutoCAD SHX Text
IRS1
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH 3" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
IMAGE RECONSTRUCTION CAB. DATA/COMMUNICATION
AutoCAD SHX Text
MMW
AutoCAD SHX Text
---
AutoCAD SHX Text
ETHERNET CONNECTION TO HOSPITAL NETWORK, EXACT LOCATION TO BE COORDINATED WITH SIEMENS PROJECT MANAGER.
AutoCAD SHX Text
SYNGO MULTIMODALITY WORKPLACE.
AutoCAD SHX Text
PDC
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH 6" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
POWER DISTRIBUTION CAB. POWER
AutoCAD SHX Text
PDC1
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH 3" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
POWER DISTRIBUTION CAB. DATA/COMMUNICATION
AutoCAD SHX Text
PDU
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH 6" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
POWER DISTRIBUTION UNIT POWER
AutoCAD SHX Text
PDU1
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED BELOW FLOOR SLAB WITH 3" SLEEVE RUNNING THROUGH FLOOR SLABENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATION.
AutoCAD SHX Text
POWER DISTRIBUTION UNIT DATA/COMMUNICATION
AutoCAD SHX Text
PRS
AutoCAD SHX Text
---
AutoCAD SHX Text
FIX POINT DESIGNATION, SAME PULL BOX/OPENING AS ACS1.
AutoCAD SHX Text
PET RECONSTRUCTION SYSTEM
AutoCAD SHX Text
PPS
AutoCAD SHX Text
---
AutoCAD SHX Text
FIX POINT DESIGNATION, SAME PULL BOX/OPENING AS PDU.
AutoCAD SHX Text
TRIPPLITE UPS (PET UPS)
AutoCAD SHX Text
S
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
TWO PULL BOXES MOUNTED BELOW FLOOR SLAB WITH TWO 6" SLEEVES RUNNING THROUGHFLOOR SLAB ENDING FLUSH WITH FINISHED FLOOR IN SHOWN LOCATIONS.
AutoCAD SHX Text
HEAT EXCHANGER CABINET-WATER/AIR SPLIT
AutoCAD SHX Text
S1
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX THAT IS PROVIDED ON THE OUTDOOR COOLING UNIT.
AutoCAD SHX Text
OUTDOOR COOLING UNIT-WATER/AIR SPLIT
AutoCAD SHX Text
S2
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
PULL BOX MOUNTED ADJACENT TO OUTDOOR COOLING UNIT PROVIDED WITH FLEX-TITE CONDUIT FROM PULL BOX TO WATER HOSE CONNECTIONS ON OUTDOOR COOLING UNIT.
AutoCAD SHX Text
OUTDOOR COOLING UNIT- WATER/AIR SPLIT
AutoCAD SHX Text
T/H
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
TEMPERATURE/HUMIDITY SENSOR SUPPLIED BY SIEMENS WITH POWER DISTRIBUTION UNIT TO BE LOCATED 60" ABOVE FINISHED FLOOR IN SHOWN LOCATION OR PER LOCAL CODE.
AutoCAD SHX Text
TEMPERATURE/HUMIDITY SENSOR
AutoCAD SHX Text
HD1
AutoCAD SHX Text
10" x 3 1/2"
AutoCAD SHX Text
ELECTRICAL DUCT TO RUN HORIZONTALLY ON THE WALL AT THE FLOOR LINE AND SURFACE MOUNTED ON FINISHED WALL AS SHOWN. DUCT TO BE DIVIDED INTO THREE SECTIONS WITH METAL DIVIDERS.
AutoCAD SHX Text
RACEWAY
AutoCAD SHX Text
HD2
AutoCAD SHX Text
10" x 3 1/2"
AutoCAD SHX Text
ELECTRICAL DUCT TO RUN HORIZONTALLY ON THE WALL AT THE FLOOR LINE AND SURFACE MOUNTED ON FINISHED WALL AS SHOWN. DUCT TO BE DIVIDED INTO THREE SECTIONS WITH METAL DIVIDERS. OPENING IN FACE OF RACEWAY TO BE DETERMINED AT THE TIME OF INSTALL.
AutoCAD SHX Text
RACEWAY
AutoCAD SHX Text
CONTRACTOR SUPPLIED CABLES
AutoCAD SHX Text
FROM
AutoCAD SHX Text
VIA
AutoCAD SHX Text
TO
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
REMARKS
AutoCAD SHX Text
POWER SOURCE
AutoCAD SHX Text
1,A4
AutoCAD SHX Text
PDU
AutoCAD SHX Text
3-PHASE CONDUCTORS, 1 NEUTRAL AND PLUS GROUND ALL TO BE THE SAME SIZE. SIZED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
PDU
AutoCAD SHX Text
2
AutoCAD SHX Text
GROUND ELECTRODE
AutoCAD SHX Text
1-#8 TO BUILDING GROUND. DETERMINED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
PDU
AutoCAD SHX Text
A1,3
AutoCAD SHX Text
EPO
AutoCAD SHX Text
DETERMINED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
EPO
AutoCAD SHX Text
4
AutoCAD SHX Text
EPO
AutoCAD SHX Text
DETERMINED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
COMP UPS/PDU
AutoCAD SHX Text
A1,5
AutoCAD SHX Text
EPO
AutoCAD SHX Text
DETERMINED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
PPS/PDU
AutoCAD SHX Text
A1,6
AutoCAD SHX Text
EPO
AutoCAD SHX Text
DETERMINED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
EPO
AutoCAD SHX Text
7,A
AutoCAD SHX Text
PDC/UPS
AutoCAD SHX Text
DETERMINED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
PDC
AutoCAD SHX Text
A,8
AutoCAD SHX Text
WARNING LIGHT
AutoCAD SHX Text
DETERMINED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE DETAIL E-501
AutoCAD SHX Text
PDC
AutoCAD SHX Text
A,9
AutoCAD SHX Text
DOOR SAFETY SWITCH
AutoCAD SHX Text
DETERMINED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE DETAIL E-501
AutoCAD SHX Text
PDU
AutoCAD SHX Text
10
AutoCAD SHX Text
S
AutoCAD SHX Text
DETERMINED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
W/A SPLIT COOLING SYSTEM SEE POWER SCHEDULE
AutoCAD SHX Text
SIEMENS SUPPLIED CABLES
AutoCAD SHX Text
FROM
AutoCAD SHX Text
VIA
AutoCAD SHX Text
TO
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
REMARKS
AutoCAD SHX Text
PDU
AutoCAD SHX Text
A1,11
AutoCAD SHX Text
T/H
AutoCAD SHX Text
COMMUNICATION CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 98'-0".
AutoCAD SHX Text
PDC
AutoCAD SHX Text
12
AutoCAD SHX Text
ICS/PIT
AutoCAD SHX Text
POWER CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
PDC1
AutoCAD SHX Text
13
AutoCAD SHX Text
ICS1/PIT
AutoCAD SHX Text
DATA/COMMUNICATION CABLES.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
PDC
AutoCAD SHX Text
14
AutoCAD SHX Text
IRS
AutoCAD SHX Text
POWER CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
PDC1
AutoCAD SHX Text
15
AutoCAD SHX Text
IRS1
AutoCAD SHX Text
DATA/COMMUNICATION CABLES.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
PDC
AutoCAD SHX Text
HD1
AutoCAD SHX Text
PDU
AutoCAD SHX Text
POWER AND DATA/COMMUNICATION CABLES.
AutoCAD SHX Text
MAXIMUM LENGTH 22'-0".
AutoCAD SHX Text
PDC
AutoCAD SHX Text
HD1
AutoCAD SHX Text
ACS
AutoCAD SHX Text
DATA CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 22'-0".
AutoCAD SHX Text
PDC
AutoCAD SHX Text
HD1
AutoCAD SHX Text
PRS
AutoCAD SHX Text
DATA CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 22'-0".
AutoCAD SHX Text
PDC
AutoCAD SHX Text
16
AutoCAD SHX Text
B
AutoCAD SHX Text
POWER CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
PDC1
AutoCAD SHX Text
17
AutoCAD SHX Text
B3
AutoCAD SHX Text
DATA/COMMUNICATION CABLES.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
B3
AutoCAD SHX Text
18
AutoCAD SHX Text
ICS1/PIT
AutoCAD SHX Text
DATA CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
B3
AutoCAD SHX Text
19
AutoCAD SHX Text
IRS1
AutoCAD SHX Text
DATA CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
B3
AutoCAD SHX Text
20
AutoCAD SHX Text
ACS1
AutoCAD SHX Text
DATA CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
PDU
AutoCAD SHX Text
21
AutoCAD SHX Text
B
AutoCAD SHX Text
POWER CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 98'-0".
AutoCAD SHX Text
PDU1
AutoCAD SHX Text
22
AutoCAD SHX Text
B3
AutoCAD SHX Text
DATA/COMMUNICATION CABLES.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
PDU
AutoCAD SHX Text
HD1
AutoCAD SHX Text
PDC
AutoCAD SHX Text
POWER CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 22'-0".
AutoCAD SHX Text
PDU
AutoCAD SHX Text
HD1
AutoCAD SHX Text
APD
AutoCAD SHX Text
POWER CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 22'-0".
AutoCAD SHX Text
APD
AutoCAD SHX Text
LOOSE
AutoCAD SHX Text
ACS
AutoCAD SHX Text
POWER CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 3'-0".
AutoCAD SHX Text
APD
AutoCAD SHX Text
LOOSE
AutoCAD SHX Text
PRS
AutoCAD SHX Text
DATA CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 3'-0".
AutoCAD SHX Text
S
AutoCAD SHX Text
23
AutoCAD SHX Text
B1
AutoCAD SHX Text
WATER HOSE.
AutoCAD SHX Text
MAXIMUM LENGTH 65'-0".
AutoCAD SHX Text
B1
AutoCAD SHX Text
24
AutoCAD SHX Text
S
AutoCAD SHX Text
WATER HOSE.
AutoCAD SHX Text
MAXIMUM LENGTH 65'-0".
AutoCAD SHX Text
DRAIN HOLE UNDER GANTRY
AutoCAD SHX Text
25
AutoCAD SHX Text
ON-SITE DRAIN
AutoCAD SHX Text
CONDENSATION HOSE.
AutoCAD SHX Text
MAXIMUM LENGTH 32'-9".
AutoCAD SHX Text
IN1
AutoCAD SHX Text
LOOSE
AutoCAD SHX Text
IN2
AutoCAD SHX Text
INJECTOR CABLE.
AutoCAD SHX Text
IN2
AutoCAD SHX Text
26
AutoCAD SHX Text
IN3
AutoCAD SHX Text
INJECTOR CABLE.
AutoCAD SHX Text
ICS
AutoCAD SHX Text
27
AutoCAD SHX Text
F1
AutoCAD SHX Text
CONTROL CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 98'-0".
AutoCAD SHX Text
PDC
AutoCAD SHX Text
A,28
AutoCAD SHX Text
F1
AutoCAD SHX Text
POWER CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 78'-0".
AutoCAD SHX Text
S
AutoCAD SHX Text
A2,29,DC,29
AutoCAD SHX Text
S1
AutoCAD SHX Text
POWER CABLE AND CONTROL CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 131'-0".
AutoCAD SHX Text
S
AutoCAD SHX Text
30
AutoCAD SHX Text
B
AutoCAD SHX Text
DATA CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 98'-0".
AutoCAD SHX Text
S
AutoCAD SHX Text
31
AutoCAD SHX Text
S2
AutoCAD SHX Text
WATER HOSES.
AutoCAD SHX Text
MAXIMUM LENGTH 119'-0".
AutoCAD SHX Text
PDC
AutoCAD SHX Text
32
AutoCAD SHX Text
IES/PIT2
AutoCAD SHX Text
DATA CABLE.
AutoCAD SHX Text
MAXIMUM LENGTH 82'-0".
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
6 11
AutoCAD SHX Text
FINISHED ROOM HEIGHT
AutoCAD SHX Text
FOR BIOGRAPH mCT GANTRY ONLY
AutoCAD SHX Text
CAREVISION MONITOR/CEILING MOUNT
AutoCAD SHX Text
RECOMMEND 8'-0" MINIMUM 7'-5"
AutoCAD SHX Text
MIN. 8'-6 3/8" MAX. 12'-1 11/16"
AutoCAD SHX Text
CEILING MOUNT INJECTOR
AutoCAD SHX Text
MINIMUM 8'-9 1/2"

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

E-102ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
ORIENTATION POINT
AutoCAD SHX Text
ACS1
AutoCAD SHX Text
PRS
AutoCAD SHX Text
IRS1
AutoCAD SHX Text
PDC
AutoCAD SHX Text
A
AutoCAD SHX Text
ES
AutoCAD SHX Text
ES2
AutoCAD SHX Text
APD
AutoCAD SHX Text
A1
AutoCAD SHX Text
PDU
AutoCAD SHX Text
B
AutoCAD SHX Text
HD1
AutoCAD SHX Text
HD2
AutoCAD SHX Text
PIT
AutoCAD SHX Text
ICS
AutoCAD SHX Text
IES
AutoCAD SHX Text
PIT2
AutoCAD SHX Text
IN1
AutoCAD SHX Text
IN2
AutoCAD SHX Text
IN3
AutoCAD SHX Text
F1
AutoCAD SHX Text
IRS
AutoCAD SHX Text
SCANNER ROOM
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
A2
AutoCAD SHX Text
S
AutoCAD SHX Text
B1
AutoCAD SHX Text
T/H
AutoCAD SHX Text
PPS
AutoCAD SHX Text
E-501
AutoCAD SHX Text
1
AutoCAD SHX Text
A4
AutoCAD SHX Text
B3
AutoCAD SHX Text
ICS1
AutoCAD SHX Text
PDC1
AutoCAD SHX Text
PDU1
AutoCAD SHX Text
ELECTRICAL DIMENSION PLAN
AutoCAD SHX Text
SCALE: 1/4" = 1'-0"
AutoCAD SHX Text
EQUIPMENT GROUND CONDUCTOR TO COMPLY WITH THE FOLLOWING: 1) SIZED EQUIVALENT TO THE PHASE CONDUCTORS (FULL SIZED GROUND). 2) DERIVED FROM THE ELECTRICAL SERVICE, TRANSFORMER OR MAIN DISTRIBUTION PANEL FEEDING THE SIEMENS EQUIPMENT. 3) RUN IN THE SAME CONDUIT, TROUGH OR RACEWAY AS THE PHASE CONDUCTORS. 4) CONTINUOUS, WITH NO BREAKS OR USE OF CONDUIT, CHASSIS OR EARTH AS THE SOLE GROUNDING PATH. 5) BONDED TO CHASSIS AND/OR CONDUIT IN ACCORDANCE WITH THE NEC REQUIREMENTS. 6) MINIMIZE CONNECTIONS OR TERMINALS TO ENSURE CONTINUITY OVER THE LIFE OF THE INSTALLATION. 7) AS A NORM, THERE SHOULD NOT BE ANY CURRENT PRESENCE ON THE GROUND CONDUCTOR, BUT IT IS ACCEPTABLE TO HAVE <500mA DURING OPERATION OF THE <500mA DURING OPERATION OF THE 500mA DURING OPERATION OF THE DURING OPERATION OF THE IMAGING EQUIPMENT.
AutoCAD SHX Text
GROUNDING NOTES
AutoCAD SHX Text
POOR POWER WILL ALTER EQUIPMENT PERFORMANCE
AutoCAD SHX Text
POWER QUALITY
AutoCAD SHX Text
IT IS IN THE CUSTOMER'S INTEREST THAT THE ELECTRICAL CONTRACTOR BE RESPONSIBLE FOR TESTING AND VERIFYING THAT THE EQUIPMENT POWER SUPPLY COMPLIES WITH THE SIEMENS SPECIFICATIONS.
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
7 11
AutoCAD SHX Text
FINISHED ROOM HEIGHT
AutoCAD SHX Text
FOR BIOGRAPH mCT GANTRY ONLY
AutoCAD SHX Text
CAREVISION MONITOR/CEILING MOUNT
AutoCAD SHX Text
RECOMMEND 8'-0" MINIMUM 7'-5"
AutoCAD SHX Text
MIN. 8'-6 3/8" MAX. 12'-1 11/16"
AutoCAD SHX Text
CEILING MOUNT INJECTOR
AutoCAD SHX Text
MINIMUM 8'-9 1/2"
AutoCAD SHX Text
ELECTRICAL NOTES
AutoCAD SHX Text
1) COMPLIANCE: ELECTRICAL WORK SHALL BE IN COMPLIANCE WITH THE NATIONAL ELECTRICAL CODE (NFPA-70), O.S.H.A. REGULATIONS, AS WELL AS APPLICABLE REGULATIONS OF CITY, COUNTY, STATE AND FEDERAL AGENCIES. PROVIDE MATERIALS AND EQUIPMENT THAT COMPLY WITH ANSI, IEEE AND NEMA STANDARDS AND ARE U.L. LISTED AND LABELED. THE CUSTOMER'S/CONTRACTOR'S WORK AND ALL EQUIPMENT INSTALLED SHALL COMPLY WITH THE CURRENT EDITION OF THE NATIONAL ELECTRICAL CODE ADOPTED/ENFORCED BY THE AUTHORITY HAVING JURISDICTION. 2) QUALITY ASSURANCE: THE CONTRACTOR SHALL VERIFY EXISTING CONDITIONS IN THE FIELD TO INSURE THAT THE NEW WORK WILL FIT INTO THE EXISTING STRUCTURE AS SHOWN ON THE DRAWINGS. SHOULD ANY CONDITIONS EXIST OR BE DISCOVERED THAT PREVENT THE INSTALLATION OF WORK AS SHOWN, THE CONTRACTOR SHALL NOTIFY THE OWNER'S REPRESENTATIVE PRIOR TO FABRICATION OF EQUIPMENT, OR THE PERFORMANCE OF ANY WORK THAT MAY BE AFFECTED. DO NOT ALTER DRAWINGS, DIMENSIONS, OR SPECIFICATIONS IN ANY WAY WITHOUT CONTACTING AND RECEIVING WRITTEN CONFIRMATION FROM SIEMENS PROJECT MANAGER. ALL DIMENSIONS ARE FROM FINISHED SURFACES. CONDUIT AND PULL BOXES TO BE INSTALLED BY THE CUSTOMER/CONTRACTOR WITH LOCATIONS BEING FIELD VERIFIED BY THE SIEMENS PROJECT MANAGER. 3) POWER SUPPLY SOURCE: POWER SUPPLIES FOR SIEMENS HEALTHCARE EQUIPMENT SHALL BE FROM A MEDICAL IMAGING PANEL OR BUILDING SERVICE EQUIPMENT THAT IS A GROUNDED 3 OR 4-WIRE 'WYE' SOURCE PER THE SPECIFIC EQUIPMENT OPERATION REQUIREMENTS. A DEDICATED CIRCUIT SHALL BE PROVIDED THAT IS KEPT ENTIRELY FREE AND INDEPENDENT OF ALL OTHER BUILDING WIRING. NO ELEVATORS, GENERATORS, PUMPS, HVAC OR SIMILAR EQUIPMENT SHALL BE CONNECTED TO THE SAME CIRCUIT OR MEDICAL IMAGING PANEL THAT SERVES THE SIEMENS HEALTHCARE EQUIPMENT. IF THE POWER SUPPLY SOURCE DOES NOT MEET THE SPECIFIC SIEMENS EQUIPMENT POWER REQUIREMENTS, THE CONTRACTOR SHALL PROVIDE THE NECESSARY EQUIPMENT REQUIRED TO ESTABLISH THE POWER SUPPLY IN ACCORDANCE WITH THE REQUIRED POWER SUPPLY PARAMETERS OF THE SIEMENS EQUIPMENT. THE CONTRACTOR SHALL COORDINATE THIS WORK WITH THE CUSTOMER AND/OR UTILITY COMPANY FIELD REPRESENTATIVE. 4) WORK FURNISHED BY CUSTOMER/CONTRACTOR: WORK NOT PROVIDED BY SIEMENS HEALTHCARE BUT SHOWN ON DRAWINGS TO BE FURNISHED AND INSTALLED BY CUSTOMER/CONTRACTOR INCLUDES, BUT IS NOT LIMITED TO, THE FOLLOWING, UNLESS NOTED OTHERWISE: ELECTRICAL RACEWAYS AND DUCTS, WIRING TROUGHS, PULL BOXES, CONDUITS, CIRCUIT BREAKERS, ACCESS PANELS, EMERGENCY OFF BUTTONS, DOOR SWITCHES, WARNING LIGHTS, WIRING, WIRING DEVICES, CONNECTORS, LIGHTING EQUIPMENT AND GROUNDING. 5) RACEWAY AND CONDUIT NOTES: ALL CONDUITS SHALL BE INSTALLED IN COMPLIANCE WITH THE CURRENT ENFORCED EDITION OF THE NATIONAL ELECTRICAL CODE. CONDUIT BODIES SHALL NOT BE USED. WHERE A CONDUIT ENTERS A BOX, FITTING, OR OTHER ENCLOSURE, AN INSULATED THROAT CONNECTOR SHALL BE PROVIDED TO PROTECT THE WIRE FROM ABRASION. ALL CONNECTORS FOR EMT SHALL BE COMPRESSION OR DOUBLE SET SCREW TYPE. KEEP RACEWAYS AT LEAST 6 INCHES AWAY FROM PARALLEL RUNS OF FLUES OR STEAM AND HOT WATER PIPES. INSTALL RACEWAY RUNS ABOVE WATER AND STEAM PIPES PROVIDED THAT CABLE RUN DISTANCES ARE MAINTAINED. USE TEMPORARY CLOSURES TO PREVENT FOREIGN MATTER FROM ENTERING RACEWAY. CONDUIT RUNS ARE SHOWN SCHEMATICALLY. INSTALL CONDUIT WITH A MINIMUM OF BENDS IN THE SHORTEST PRACTICAL DISTANCE CONSIDERING THE BUILDING CONSTRUCTION AND OBSTRUCTIONS, EXCEPT AS OTHERWISE INDICATED. THE CONTRACTOR SHALL MAKE CERTAIN THAT ANY CONDUIT/RACEWAY RUNS CONTAINING SIEMENS HEALTHCARE CABLES DO NOT EXCEED THE SPECIFIED MAXIMUM DISTANCES AS SHOWN ON THE ELECTRICAL DETAILS. LISTED CONDUIT SIZES FOR SIEMENS-SUPPLIED CABLES MUST BE MAINTAINED IN ORDER TO ENABLE THE TOTAL CABLE BUNDLE INCLUDING CONNECTORS TO BE PULLED THROUGH WITHOUT DAMAGE. PROVIDE ENCLOSED METAL WIRE DUCT RACEWAY SYSTEM WHERE SHOWN ON DRAWINGS WITH DIVIDERS TO SEPARATE THE DUCT INTO TWO OR THREE SEPARATE COMPARTMENTS AS SHOWN ON THE SIEMENS PLANS (FOR POWER AND SIEMENS HEALTHCARE CABLING). DIVIDERS AND CROSSOVER PIECES TO BE PROVIDED AS NECESSARY. THE CABLE TO CABLE AS WELL AS THE CIRCUIT TO CIRCUIT SEPARATION REQUIREMENT WAS EVALUATED DURING THE UL SYSTEM CERTIFICATION OF THE EQUIPMENT. ADDITIONAL SEPARATION OF THE SYSTEM CABLE ASSEMBLIES INTO SEPARATE OR PARTITIONED RACEWAYS, UNLESS OTHERWISE NOTED, IS NOT NECESSARY TO INSURE SEPARATION OF CIRCUITS. PROVIDE WIRE DUCT/RACEWAY WITH ACCESSIBLE REMOVABLE COVERS. LOCATIONS OF BUILDING MATERIAL OPENINGS (I.E. ACCESS PANELS) TO BE CUT IN FIELD ARE TO BE COORDINATED WITH THE DRAWING REQUIRMENTS AND BUILDING STRCTURE. THOSE THAT ARE NOT INDICATED OR INTERFER WITH BUILDING ELEMENTS SHALL BE COORDINATED WITH SIEMENS PROJECT MANAGER. ELECTRICAL PULL BOXES AND RACEWAY COVERS SHALL BE INSTALLED IN A MANNER TO ALLOW ACCESSIBILITY FOR INSTALLATION AND MAINTENANCE. CONTRACTORS MUST PROVIDE PULL STRINGS FOR ALL CONDUIT AND WIRE DUCT/RACEWAY. IN-FLOOR TRENCH DUCT AND FLUSH FLOOR BOXES SHALL BE PROVIDED WITH FULLY GASKETED REMOVABLE COVERS. WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED HIGHER THAN 14 FEET ABOVE FINISHED FLOOR, THE ELECTRICAL CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP THE SIEMENS INSTALLERS PULL SIEMENS SUPPLIED CABLES AT CUSTOMER'S EXPENSE. WHEN JUNCTION WHEN JUNCTION BOXES AND WIRE DUCT/RACEWAY ARE MOUNTED ABOVE A HARD CEILING (I.E. SHEET ROCK), A 24" x 24" ACCESS PANEL IS REQUIRED AT EACH JUNCTION BOX AND WITHIN 2 FEET OF EACH RACEWAY TRANSITION (SUCH AS A 90 DEGREE ELBOW OR TEE) IN DUCT/RACEWAY. THERE MUST BE FREE AND CLEAR ACCESS TO JUNCTION BOXES AND WIRE DUCT/RACEWAY. WHEN ACCESS PANELS ARE LOCATED MORE THAN 3 FEET FROM JUNCTION BOXES AND WIRE DUCT/RACEWAY THE ELECTRICAL CONTRACTOR SHALL PROVIDE TWO ELECTRICIANS TO HELP SIEMENS INSTALLERS PULL SIEMENS SUPPLIED CABLES AT CUSTOMER'S EXPENSE. 6) WIRING: ALL WIRING INSTALLED SHALL BE 600 VOLT CLASS, STRANDED TYPE THHN/THWN-2, SINGLE CONDUCTOR ANNEALED COPPER FOR A MAXIMUM OPERATING TEMPERATURE OF 90° C (194° F), SIZED AS INDICATED, INSTALLED IN METAL RACEWAYS. THE CUSTOMER/CONTRACTOR SHALL LEAVE A MINIMUM 10 FEET OF WIRE TAILS AT ALL OUTLET POINTS WITH WIRE IDENTIFICATION TAGGED AT BOTH ENDS FOR FINAL CONNECTION BY THE CUSTOMER/ELECTRICAL CONTRACTOR. 7) SHORT CIRCUIT REQUIREMENTS: ALL CIRCUIT BREAKERS SUPPLIED FOR THE SIEMENS EQUIPMENT REQUIREMENTS SHALL BE RATED HIGHER THAN THE SHORT CIRCUIT AVAILABLE AT THE TERMINALS OF THE ELECTRICAL EQUIPMENT AS DETERMINED BY THE ENGINEER OF RECORD, BUT NOT LESS THAN 35,000A RMS SYMMETRICAL AT 480V, 3-PHASE, 60 HERTZ. THE CONTRACTOR SHALL OBTAIN THE CORRECT SHORT CIRCUIT CURRENT RATING OF ALL THE NEW EQUIPMENT FOR INSTALLATION FROM THE ENGINEER OF RECORD.
AutoCAD SHX Text
SYSTEMS NETWORK (VERIFY WITH SMS PROJECT MANAGER).
AutoCAD SHX Text
ETHERNET CONNECTION TO CUSTOMER'S INFORMATION
AutoCAD SHX Text
ALL MAY NOT APPLY
AutoCAD SHX Text
UNDER FLOOR DUCT
AutoCAD SHX Text
CEILING DUCT
AutoCAD SHX Text
TRENCHDUCT
AutoCAD SHX Text
SYMBOLS
AutoCAD SHX Text
OPENING IN ACCESS FLOORING
AutoCAD SHX Text
(EPO) EMERGENCY POWER OFF BUTTON
AutoCAD SHX Text
DOOR SAFETY SWITCH
AutoCAD SHX Text
WARNING LIGHT (X-RAY ON)
AutoCAD SHX Text
VERTICAL DUCT
AutoCAD SHX Text
PULLBOX IN (FLOOR/WALL/CEILING)
AutoCAD SHX Text
OPENING IN RACEWAY OR TRENCHDUCT
AutoCAD SHX Text
MAIN PANEL OR ENCLOSURE BY CUSTOMER/CONTRACTOR
AutoCAD SHX Text
DS
AutoCAD SHX Text
110 VOLT, 20 AMP, HOSPITAL GRADE DUPLEX OUTLET UNLESS OTHERWISE STATED.
AutoCAD SHX Text
SURFACE DUCT
AutoCAD SHX Text
SPECIAL PURPOSE RECEPTACLE
AutoCAD SHX Text
L6-30R
AutoCAD SHX Text
110 VOLT, 20 AMP, HOSPITAL GRADE QUAD OUTLET
AutoCAD SHX Text
SYSTEM
AutoCAD SHX Text
POWER REQUIREMENTS
AutoCAD SHX Text
POWER CONSUMPTION WITH WATER/AIR SPLIT CONNECTED THROUGH THE POWER DISTRIBUTION UNIT (PDU) (SIEMENS COOLING SYSTEM WATER/AIR) OPERATION - 150 kVA MOMENTARY < 5 SEC. < 5 SEC. 5 SEC. SYSTEM ON (STAND-BY) - 29 kVA (6 FOR PET, 4 FOR CT, 19 FOR HEAT EXCHANGER) SYSTEM ON (COMP ON) - 27.5 kVA (6 FOR PET, 2.5 FOR CT, 19 FOR HEAT EXCHANGER) SYSTEM OFF - 25 kVA (4.3 FOR PET, 1.7 FOR CT, 19 FOR HEAT EXCHANGER) IF AN ON-SITE TRANSFORMER IS REQUIRED TO OBTAIN BIOGRAPH mCT FLOW AND BIOGRAPH mCT OPERATING VOLTAGE, IT MUST BE OF SUFFICIENT CAPACITY AND CHARACTERISTICS TO MAINTAIN SUPPLY VOLTAGE AND IMPEDANCE REQUIREMENTS (TRANSFORMER AND CONDUCTORS). DO NOT CONNECT ANY EXTERNAL UNITS TO THE BIOGRAPH BIOGRAPH mCT FLOW AND BIOGRAPH mCT POWER LINES.
AutoCAD SHX Text
BIOGRAPH mCT FLOW AND BIOGRAPH mCT.S/X 100kW
AutoCAD SHX Text
SEE BELOW
AutoCAD SHX Text
125125
AutoCAD SHX Text
200
AutoCAD SHX Text
THE EXAMINATION ROOM SHOULD BE EQUIPPED WITH AT LEAST ONE EMERGENCY POWER OFF (PANIC) BUTTON.
AutoCAD SHX Text
SYSTEM VOLTAGE (VOLTS)
AutoCAD SHX Text
POWER CONSUMPTION (kVA)
AutoCAD SHX Text
SYSTEM IMPEDANCE (mΩ)
AutoCAD SHX Text
MINIMUM SYSTEM SUPPLY (A)
AutoCAD SHX Text
3 480/277Y ±10%
AutoCAD SHX Text
ITEM
AutoCAD SHX Text
QTY
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
EPO
AutoCAD SHX Text
VARIES
AutoCAD SHX Text
UNLESS OTHERWISE NOTED, ALL ITEMS LISTED IN THIS SCHEDULE SHALL BE SUPPLIED AND INSTALLED BY CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
EPO
AutoCAD SHX Text
NOTE 1 - EPO CIRCUIT #1 EPO CONTACTS TO BE NORMALLY CLOSED, WIRED IN NORMALLY CLOSED, WIRED IN , WIRED IN SERIES, CONNECTED TO THE POWER DISTRIBUTION UNIT CABINET (PDU) INTERNAL MAIN BREAKER. ALL EPO'S TO HAVE MECHANICAL LATCHING MECHANISM. EPO MUST BE MECHANICAL LATCHING MECHANISM. EPO MUST BE . EPO MUST BE RESET BEFORE MAIN BREAKER CAN RESUME OPERATION. WIRING SIZE TO BE DETERMINED BY ELECTRICAL ENGINEER OF RECORD. NOTE 2 - EPO CIRCUIT #2 EPO CONTACTS TO BE NORMALLY CLOSED, WIRED IN NORMALLY CLOSED, WIRED IN , WIRED IN SERIES, CONNECTED TO THE COMP. UPS INSIDE THE POWER DISTRIBUTION UNIT CABINET (PDU). NOTE 3 - EPO CIRCUIT #3 EPO CONTACTS TO BE NORMALLY CLOSED, WIRED IN NORMALLY CLOSED, WIRED IN , WIRED IN SERIES, CONNECTED TO PET UPS OPTION (PPS) INSIDE THE POWER DISTRIBUTION UNIT CABINET (PDU). NOTE 4 - EPO CIRCUIT #4 EPO CONTACTS TO BE NORMALLY CLOSED, WIRED IN NORMALLY CLOSED, WIRED IN , WIRED IN SERIES, CONNECTED TO COMP UPS INSIDE THE POWER DISTRIBUTION CABINET (PDC). THE EPOs MUST BE INSTALLED BY A QUALIFIED ELECTRICAL CONTRACTOR ACCORDING TO NATIONAL ELECTRICAL CODE, STATE AND LOCAL REGULATIONS. THE CUSTOMER IS SOLELY RESPONSIBLE FOR THE IMPLEMENTATION OF THE EPOs AND THEIR ASSOCIATED CIRCUITS AND MUST MAKE THE FINAL DETERMINATION CONSIDERING ALL SITE CONDITIONS AND REGULATORY FACTORS.
AutoCAD SHX Text
POWER SCHEDULE
AutoCAD SHX Text
NC
AutoCAD SHX Text
NC
AutoCAD SHX Text
NC
AutoCAD SHX Text
EPO
AutoCAD SHX Text
EPO
AutoCAD SHX Text
VOLTS
AutoCAD SHX Text
480/277Y
AutoCAD SHX Text
3
AutoCAD SHX Text
PHASES
AutoCAD SHX Text
NEUTRAL
AutoCAD SHX Text
GROUND
AutoCAD SHX Text
TOTAL WIRES
AutoCAD SHX Text
1
AutoCAD SHX Text
1
AutoCAD SHX Text
PDU
AutoCAD SHX Text
1
AutoCAD SHX Text
POWER DISTRIBUTION UNIT (SIEMENS SUPPLIED)
AutoCAD SHX Text
PET UPS (OPTION)
AutoCAD SHX Text
COMP UPS
AutoCAD SHX Text
NC
AutoCAD SHX Text
NC
AutoCAD SHX Text
NC
AutoCAD SHX Text
NC
AutoCAD SHX Text
NC
AutoCAD SHX Text
NC
AutoCAD SHX Text
POWER DISTRIBUTION CABINET (SIEMENS SUPPLIED)
AutoCAD SHX Text
UPS
AutoCAD SHX Text
NOTE #4
AutoCAD SHX Text
#8 TO BUILDING GROUND
AutoCAD SHX Text
ALL CONDUITS AND WIRES SIZES MUST BE DETERMINED BY THE ELECTRICAL ENGINEER OF RECORD PER N.E.C AND TO MAINTAIN SIEMENS IMPEDANCE REQUIREMENTS.
AutoCAD SHX Text
POWER DISTRIBUTION UNIT (PDU) PROVIDED AND INSTALLED BY SIEMENS.
AutoCAD SHX Text
NC
AutoCAD SHX Text
NC
AutoCAD SHX Text
NC
AutoCAD SHX Text
NOTE #1
AutoCAD SHX Text
NOTE #2
AutoCAD SHX Text
NOTE #3
AutoCAD SHX Text
CABLE FURNISHED BY SIEMENS. REFER TO ELECTRICAL LEGEND FOR CONDUIT SIZE
AutoCAD SHX Text
HEAT EXCHANGER CABINET (SIEMENS SUPPLIED)
AutoCAD SHX Text
PPS
AutoCAD SHX Text
MAXIMUM WIRE SIZE FOR TERMINAL LUGS IS #4/0 AWG.
AutoCAD SHX Text
480/277Y VOLTS 3 PHASE, 4 WIRE, PLUS GROUND
AutoCAD SHX Text
MCB
AutoCAD SHX Text
PDU INTERNAL MAIN CIRCUIT BREAKER AMPS (MCB): 225
AutoCAD SHX Text
PDU
AutoCAD SHX Text
PDC
AutoCAD SHX Text
S
AutoCAD SHX Text
PHASE AND NEUTRAL WIRES TO BE THE SAME SIZE. GROUND PER NEC. UNLESS OTHERWISE NOTED, ALL BREAKERS WILL BE 80% RATED
AutoCAD SHX Text
3#8, #10 GROUND
AutoCAD SHX Text
5

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

E-501ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
RELAY WITH A 24 VDC, SUPPLIED BY ELECTRICAL CONTRACTOR. ALL ITEMS EXTERNAL TO SIEMENS POWER DISTRIBUTION CABINET ARE TO BE SUPPLIED AND INSTALL BY CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
"X-RAY ON" WARNING LIGHT ABOVE ROOM ENTRY DOOR OUTSIDE SCAN ROOM
AutoCAD SHX Text
RELAY COIL
AutoCAD SHX Text
X-RAY LIGHT RELAY BOX
AutoCAD SHX Text
LIGHTING PANEL (POWER PANEL)
AutoCAD SHX Text
-
AutoCAD SHX Text
+
AutoCAD SHX Text
24VDC
AutoCAD SHX Text
JUMPER
AutoCAD SHX Text
DS
AutoCAD SHX Text
SUPPLIED 24 VOLTS (FOR DOOR SWITCH ONLY)
AutoCAD SHX Text
FUSE 1A MAX
AutoCAD SHX Text
CUSTOMER/ CONTRACTOR SUPPLIED
AutoCAD SHX Text
LIGHTING PANEL TO BE PROVIDED WITH SWITCH TO TURN OFF WARNING LIGHTS FOR SERVICE PURPOSES.
AutoCAD SHX Text
X-RAY WARNING LIGHTS AND DOOR SWITCH SCHEMATIC
AutoCAD SHX Text
BY CUSTOMER/CONTRACTOR
AutoCAD SHX Text
SIEMENS POWER DISTRIBUTION CABINET (PDC)
AutoCAD SHX Text
SCALE: NONE
AutoCAD SHX Text
AUXILIARY WIRING
AutoCAD SHX Text
CUSTOMER/ CONTRACTOR SUPPLIED
AutoCAD SHX Text
2
AutoCAD SHX Text
NO CORE DRILLINGS PERMITTED WITHIN THESE DEFINED AREA.
AutoCAD SHX Text
NO CORE DRILLINGS PERMITTED WITHIN THESE DEFINED AREA.
AutoCAD SHX Text
ORIENTATION POINT
AutoCAD SHX Text
PLAN VIEW OF GANTRY CABLE INLETS
AutoCAD SHX Text
IMPORTANT ALL CONDUITS MUST HAVE STANDARD SWEEP ELBOWS
AutoCAD SHX Text
OF CORE DRILL
AutoCAD SHX Text
12" X 12" PULL BOX MOUNTED BELOW FLOOR SLAB.
AutoCAD SHX Text
5" CORE DRILL THROUGHFLOOR SLAB FOR POWER CABLES.
AutoCAD SHX Text
OF CORE DRILL
AutoCAD SHX Text
OF CORE DRILL
AutoCAD SHX Text
8" X 8" X 8" PULL BOX MOUNTED BELOW FLOOR SLAB.
AutoCAD SHX Text
6" CORE DRILL THROUGHFLOOR SLAB FOR WATER HOSES.
AutoCAD SHX Text
SCALE: 1"=1'-0"
AutoCAD SHX Text
GANTRY OPENINGS
AutoCAD SHX Text
OF CORE DRILL
AutoCAD SHX Text
3" CORE DRILL THROUGHFLOOR SLAB FOR DATA CABLES.
AutoCAD SHX Text
6" X 6" PULL BOX MOUNTED BELOW FLOOR SLAB.
AutoCAD SHX Text
OF CORE DRILL
AutoCAD SHX Text
OF CORE DRILL
AutoCAD SHX Text
1
AutoCAD SHX Text
CONDUIT LEGEND
AutoCAD SHX Text
SYM
AutoCAD SHX Text
SIZE
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
REMARKS
AutoCAD SHX Text
SUPPLIED AND INSTALLED BY CUSTOMER/CONTRACTOR
AutoCAD SHX Text
1
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUIT FROM POWER SOURCE TO "A4" (PDU) SIZED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
2
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUITS FROM "PDU" TO "GROUND ELECTRODE" FOR ISOLATION TRANSFORMER INSIDE PDU SIZED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
3
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUITS FROM "A1" (PDU) TO "EPO" SIZED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
4
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUIT FROM "EPO" TO "EPO" SIZED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
5
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUIT FROM "A1" (PDU/COMP UPS INSIDE PDU) TO "EPO" SIZED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
6
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUIT FROM "A1" (PDU/PPS UPS INSIDE PDU) TO "EPO" SIZED BY ELECTRICAL CONTRACTOR.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
7
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUIT FROM "EPO" TO "A" (PDC/UPS) SIZED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
SEE POWER SCHEDULE
AutoCAD SHX Text
8
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUIT FROM "A" (PDC) TO "WARNING LIGHT" (X-RAY ON).
AutoCAD SHX Text
SEE SHEET E-501
AutoCAD SHX Text
9
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUIT FROM "A" (PDC) TO "DOOR SAFETY SWITCH".
AutoCAD SHX Text
SEE SHEET E-501
AutoCAD SHX Text
10
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
CONDUIT FROM "PDU" TO "S" SIZED BY ELECTRICAL ENGINEER OF RECORD.
AutoCAD SHX Text
W/A SPLIT COOLING SYSTEM SEE POWER SCHEDULE
AutoCAD SHX Text
11
AutoCAD SHX Text
(1) 3/4"%%c
AutoCAD SHX Text
CONDUIT FROM "A1" (PDU) TO "T/H".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 92'-0"
AutoCAD SHX Text
12
AutoCAD SHX Text
(1) 2 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "PDC" TO "ICS/PIT".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
13
AutoCAD SHX Text
(1) 1 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "PDC1" TO "ICS1/PIT".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
14
AutoCAD SHX Text
(1) 1 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "PDC" TO "IRS".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
15
AutoCAD SHX Text
(1) 3/4"%%c
AutoCAD SHX Text
CONDUIT FROM "PDC1" TO "IRS1".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
16
AutoCAD SHX Text
(2) 3"%%c
AutoCAD SHX Text
CONDUITS FROM "PDC" TO "B" WITH A MINIMUM 6" BENDING RADIUS.
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
17
AutoCAD SHX Text
(1) 3"%%c
AutoCAD SHX Text
CONDUIT FROM "PDC1" TO "B3".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
18
AutoCAD SHX Text
(1) 3"%%c
AutoCAD SHX Text
CONDUIT FROM "B3" TO "ICS1/PIT".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
19
AutoCAD SHX Text
(1) 1 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "B3" TO "IRS1".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
20
AutoCAD SHX Text
(1) 1 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "B3" TO "ACS1".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 69'-0"
AutoCAD SHX Text
21
AutoCAD SHX Text
(1) 3"%%c
AutoCAD SHX Text
CONDUIT FROM "PDU" TO "B".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 92'-0"
AutoCAD SHX Text
22
AutoCAD SHX Text
(1) 1 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "PDU1" TO "B3".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
23
AutoCAD SHX Text
(1) 3"%%C
AutoCAD SHX Text
CONDUIT FROM "S" TO "B1". TO CONTAIN SIEMENS WATER HOSES (SUPPLY) WITH A MINIMUM 9" BENDING RADIUS.
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 59'-0" W/A SPLIT COOLING SYSTEM SEE SHEET M-101
AutoCAD SHX Text
24
AutoCAD SHX Text
(1) 3"%%C
AutoCAD SHX Text
CONDUIT FROM "B1" TO "S". TO CONTAIN SIEMENS WATER HOSES (RETURN) WITH A MINIMUM 9" BENDING RADIUS.
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 59'-0" W/A SPLIT COOLING SYSTEM SEE SHEET M-101
AutoCAD SHX Text
25
AutoCAD SHX Text
(1) 1"%%C
AutoCAD SHX Text
CONDUIT, IF REQUIRED PER LOCAL CODE, FOR CONDENSATION HOSE TO SELECTED DRAIN TYPE. THE MINIMUM BENDING RADIUS IS 1 3/16". MAXIMUM LENGTH 32'-9".
AutoCAD SHX Text
CONDENSATION HOSE SEE SHEET M-101
AutoCAD SHX Text
26
AutoCAD SHX Text
(1) 2 1/2"%%C
AutoCAD SHX Text
CONDUIT FROM "IN2" TO "IN3", VERIFY LENGTH RESTRICTIONS WITH MANUFACTURER.
AutoCAD SHX Text
27
AutoCAD SHX Text
(1) 2 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "ICS" TO "F1".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 85'-0"
AutoCAD SHX Text
28
AutoCAD SHX Text
(1) 2 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "A" (PDC) TO "F1".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 68'-0"
AutoCAD SHX Text
29
AutoCAD SHX Text
(1) 1 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "A2" (S) TO "DC" AND "DC" TO "S1".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 115'-0"
AutoCAD SHX Text
30
AutoCAD SHX Text
(1) 1 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "S" TO "B".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 92'-0"
AutoCAD SHX Text
31
AutoCAD SHX Text
(2) 3"%%C
AutoCAD SHX Text
CONDUITS, IF REQUIRED PER LOCAL CODE, FROM "S" TO "S2". THE MINIMUM BENDING RADIUS IS 12.5".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 112'-0"
AutoCAD SHX Text
32
AutoCAD SHX Text
(1) 2 1/2"%%c
AutoCAD SHX Text
CONDUIT FROM "PDC" TO "IES/PIT2".
AutoCAD SHX Text
MAXIMUM CONDUIT LENGTH 76'-0"
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
8 11

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

M-101ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
ORIENTATION POINT
AutoCAD SHX Text
SCANNER ROOM
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
SIEMENS SUPPLIED AND INSTALLED CONDENSATION HOSE ROUTED ACCORDING TO LOCAL CODE TO SELECTED ON-SITE DRAIN.
AutoCAD SHX Text
SIEMENS SUPPLIED WATER HOSES FROM THE HEAT EXCHANGER TO THE OUTDOOR UNIT TO BE INSTALLED BY THE CUSTOMER/CONTRACTOR. THE HOSES SHALL BE RUN IN (2) 3" CONDUITS ORACCORDING TO CODE.
AutoCAD SHX Text
SIEMENS SUPPLIED AND INSTALLED WATER HOSE FROM THE mCT GANTRY TO THE HEAT EXCHANGER (RETURN). THE HOSE SHALL BE RUN IN (1) 3" CONDUIT ORACCORDING TO LOCAL CODE.
AutoCAD SHX Text
THE OUTDOOR COOLING UNIT IS SUPPLIED BY SIEMENS, LOCATED AND INSTALLED BY THE CUSTOMER/ CONTRACTOR. THE INDOOR HEAT EXCHANGER CABINET IS SUPPLIED AND LOCATED BY SIEMENS, INSTALLED BY THE CUSTOMER/CONTRACTOR. IF THE SYSTEM IS TO BE RELOCATED, THE OUTDOOR COOLING UNIT IS TO BE DEINSTALLED, RELOCATED AND REINSTALLED BY THE CUSTOMER/CONTRACTOR, IF NECESSARY. THE INDOOR HEAT EXCHANGER CABINET IS RELOCATED BY SIEMENS, INSTALLED BY THE CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
M-501
AutoCAD SHX Text
3
AutoCAD SHX Text
M-501
AutoCAD SHX Text
2
AutoCAD SHX Text
M-101
AutoCAD SHX Text
1
AutoCAD SHX Text
M-101
AutoCAD SHX Text
2
AutoCAD SHX Text
M-101
AutoCAD SHX Text
3
AutoCAD SHX Text
M-101
AutoCAD SHX Text
4
AutoCAD SHX Text
M-101
AutoCAD SHX Text
4
AutoCAD SHX Text
M-501
AutoCAD SHX Text
4
AutoCAD SHX Text
NOTE: FOR THE HEAT OUTPUT (BTU/HR) OF ALL SIEMENS EQUIPMENT SHOWN ON THIS PLAN, SEE THE "EQUIPMENT LEGEND" ON SHEET A-101.
AutoCAD SHX Text
MECHANICAL PLAN
AutoCAD SHX Text
SCALE: 1/4" = 1'-0"
AutoCAD SHX Text
SCALE: 3/4"=1'-0"
AutoCAD SHX Text
HEAT EXCHANGER
AutoCAD SHX Text
TOP VIEW
AutoCAD SHX Text
FRONT VIEW
AutoCAD SHX Text
CONNECTION DETAILS
AutoCAD SHX Text
WATER INLET FROM OUTDOOR UNIT (PRIMARY CIRCUIT)
AutoCAD SHX Text
WATER INLET TO GANTRY (SECONDARY COOLING CIRCUIT)
AutoCAD SHX Text
WATER RETURN FROM GANTRY (SECONDARY COOLING CIRCUIT)
AutoCAD SHX Text
WATER RETURN TO OUTDOOR UNIT (PRIMARY CIRCUIT)
AutoCAD SHX Text
OPENINGS IN CABINET BASE FOR ROUTING HOSES AND CABLES
AutoCAD SHX Text
1
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
9 11
AutoCAD SHX Text
FINISHED ROOM HEIGHT
AutoCAD SHX Text
FOR BIOGRAPH mCT GANTRY ONLY
AutoCAD SHX Text
CAREVISION MONITOR/CEILING MOUNT
AutoCAD SHX Text
RECOMMEND 8'-0" MINIMUM 7'-5"
AutoCAD SHX Text
MIN. 8'-6 3/8" MAX. 12'-1 11/16"
AutoCAD SHX Text
CEILING MOUNT INJECTOR
AutoCAD SHX Text
MINIMUM 8'-9 1/2"
AutoCAD SHX Text
THE WATER-AIR SPLIT COOLING SYSTEM IS DELIVERED AS A PRE-INSTALLATION KIT. IT MUST BE INSTALLED PRIOR TO DELIVERY OF THE SYSTEM BY THE CUSTOMER/CONTRACTOR. THE FOLLOWING IS INCLUDED IN THE PRE-INSTALLATION KIT: OUTDOOR COOLING UNIT, INDOOR HEAT EXCHANGER CABINET, HOSES AND SYSTEM CABLES, 26.4 GALLONS ANTIFREEZE (GLYKOSOL-N, ADDITIONAL CAN BE ORDERED THROUGH SIEMENS LOCAL SERVICE), CONNECTION KIT AND CONNECTION SCHEMATIC.
AutoCAD SHX Text
SIEMENS OUTDOOR UNIT (FRONT)
AutoCAD SHX Text
SIEMENS HEAT EXCHANGER INDOOR UNIT (FRONT)
AutoCAD SHX Text
SIEMENS GANTRY (BACK SIDE)
AutoCAD SHX Text
SCALE: NONE
AutoCAD SHX Text
WATER/AIR SPLIT COOLING DIAGRAM
AutoCAD SHX Text
A 3/4" HOSE BIB AND A WATER DRAIN (EX. SINK) MUST BE AVAILABLE IN THE EQUIPMENT ROOM OR IN THE SCAN ROOM NEAR THE HEAT EXCHANGER FOR FILLING THE COOLING CIRCUITS.
AutoCAD SHX Text
THE COOLING CIRCUIT BETWEEN THE OUTDOOR UNIT AND THE INDOOR UNIT IS THE PRIMARY COOLING CIRCUIT. THE COOLING CIRCUIT BETWEEN THE GANTRY AND THE INDOOR COOLING UNIT IS THE SECONDARY COOLING CIRCUIT. FILLING OF THE PRIMARY COOLING CIRCUIT MUST BE PERFORMED BY THE CUSTOMER/CONTRACTOR LOCAL SPECIALIST FOR WATER INSTALLATIONS (A PLUMBER). THE STATIC FILL PRESSURE MUST BE SET TO 29 PSI ±3 PSI AFTER FILLING. THE PUMP IN THE PRIMARY COOLING CIRCUIT GENERATES AN OPERATING PRESSURE OF 29-73 PSI AT THE PUMP OUTLET. THE PUMP IN THE SECONDARY COOLING CIRCUIT GENERATES AN OPERATING PRESSURE OF 58-73 PSI AT THE PUMP OUTLET. THE WATER THE WATER VOLUME IN THE PRIMARY COOLING CIRCUIT IS APPROX. 20 GALLONS. THE WATER VOLUME IN THE SECONDARY COOLING CIRCUIT IS APPROX. 13.2 GALLONS, NOT INCLUDING ANY HOSE EXTENSIONS. ANTIFREEZE (GLYKOSOL-N) WILL BE UTILIZED BY GLYKOSOL-N) WILL BE UTILIZED BY ) WILL BE UTILIZED BY THE TECHNICIAN WHO WILL START THE SYSTEM. THE PERMITTED . THE PERMITTED MIXING RATIO IS 40:60 (GLYCOL:WATER) FOR TEMPERATURE RANGE -22° TO 122°. 26.4 GALLONS OF ANTIFREEZE ANTIFREEZE (GLYKOSOL-N) IS INCLUDED IN DELIVERY. WATER CANNOT BE GLYKOSOL-N) IS INCLUDED IN DELIVERY. WATER CANNOT BE ) IS INCLUDED IN DELIVERY. WATER CANNOT BE . WATER CANNOT BE ADDED TO THIS MIXTURE.
AutoCAD SHX Text
2
AutoCAD SHX Text
DUE TO THE HIGH NOISE (<60dB), IT IS RECOMMENDED TO INSTALL THE HEAT EXCHANGER IN THE EQUIPMENT ROOM. THE HEAT EXCHANGER MAY BE INSTALLED TO THE LEFT OR RIGHT OF THE PDC OR INDEPENDENT OF THE PDC CABINET.
AutoCAD SHX Text
SIEMENS SUPPLIES THE 2 - 1" HIGH PRESSURE WATER SUPPLY AND WATER RETURN HOSES WITH 1/2" INSULATION FOR THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT THAT TERMINATE INTO THE SUPPLY/RETURN MANIFOLDS THAT ARE SUPPLIED AND INSTALLED BY SIEMENS. THE MAXIMUM LENGTH OF THE WATER SUPPLY HOSE IS 65'-0" AND WATER RETURN HOSES IS 65'-0". ACCEPTABLE BENDING RADIUS OF THE HOSES IS 9". IT IS VERY IMPORTANT TO RUN THE SUPPLY WATER LINES ON THE LEFT AND THE RETURN LINES ON THE RIGHT (WHEN THE GANTRY IS VIEWED FROM THE FRONT). IT IS RECOMMENDED THAT THE PLUMBING INSTALLER DESIGNATE THE SUPPLY LINE. THE SUPPLY/RETURN MANIFOLDS SUPPLIED AND INSTALLED BY SIEMENS MUST BE DIRECTED TO THE RIGHT (WHEN THE GANTRY IS VIEWED FROM THE FRONT). THE COOLING WATER SUPPLY TO THE GANTRY, INCLUDING ALL FITTINGS, PRESSURE REGULATORS AND SITE CONTROL VALVES, ARE TO BE DESIGNED, CONSTRUCTED AND TESTED BY THE CUSTOMER. THE SITE CONTROL VALVES MUST BE READILY AVAILABLE AND EASILY ACCESSIBLE FOR SHUTOFF AND SHOULD BE LOCATED IN THE EQUIPMENT ROOM.
AutoCAD SHX Text
SIEMENS HEAT EXCHANGER INDOOR UNIT (FRONT)
AutoCAD SHX Text
SIEMENS GANTRY (BACK SIDE)
AutoCAD SHX Text
SCALE: NONE
AutoCAD SHX Text
SECONDARY CIRCUIT
AutoCAD SHX Text
ADAPTER 1" DN25 (2) SUPPLIED BY SIEMENS.
AutoCAD SHX Text
HEAT EXCHANGER:
AutoCAD SHX Text
CHILLED WATER HOSES (SECONDARY COOLING CIRCUIT):
AutoCAD SHX Text
SUPPLY/RETURN MANIFOLDS SUPPLY AND INSTALL BY SIEMENS.
AutoCAD SHX Text
(2) 1" HIGH PRESSURE WATER HOSES SUPPLIED BY SIEMENS. MAX. LENGTH 65'-0".
AutoCAD SHX Text
4
AutoCAD SHX Text
THE CHILLED WATER HOSES ARE SUPPLIED IN THE FOLLOWING LENGTH ONLY: 131 FEET. A 1 1/4" SUPPLY AND RETURN COPPER PIPES CAN BE USED INSTEAD OF THE SIEMENS WATER HOSES, HOWEVER THE SUPPLIED WATER HOSES MUST BE CUT AND USED TO CONNECT THE SIEMENS COMPONENTS TO THE COPPER PIPE. THE LENGTH MAY BE EXTENDED FROM 131 FEET UP TO 196 FEET WITH CUSTOMER/CONTRACTOR SUPPLIED 1 1/4" INNER COPPER PIPE USED WITH SIEMENS WATER HOSES OR FROM 131 FEET UP TO 328 FEET WITH CUSTOMER/CONTRACTOR SUPPLIED 2" INNER COPPER PIPE USED WITH SIEMENS WATER HOSES. 20 ELBOW ADAPTERS (90 DEGREES) ARE PERMITTED WITHIN THE TOTAL RUN. CUSTOMER/CONTRACTOR TO PROVIDE BLEED VALVES INSTALLED AT THE HIGHEST POINT IN THE PIPING RUN AND CUT OFF VALVES AT END OF THE SUPPLY AND RETURN. IF REQUIRED. CUSTOMER/CONTRACTOR TO SUPPLY AND INSTALL FITTINGS TO MATCH THE SELECTED PIPE SIZE. CUSTOMER/CONTRACTOR TO INSULATE WITH 1/2" ARMAFLEX INSULATION, INCLUDING FLEX HOSES. IF THE SIEMENS WATER HOSES ARE EXTENDED (>131 FEET), THE LINE VOLTAGE/CONTROL CABLE THAT RUNS FROM THE HEAT EXCHANGER CABINET TO THE OUTDOOR UNIT MUST ALSO BE EXTENDED. THESE EXTENDED CABLES ARE AVAILABLE THROUGH SIEMENS. ADDITIONAL ANTIFREEZE (GLYKOSOL-N, CAN BE ORDERED THROUGH SIEMENS LOCAL SERVICE) MUST BE SUPPLIED BY THE CUSTOMER/CONTRACTOR FOR HOSE EXTENSION.
AutoCAD SHX Text
SIEMENS OUTDOOR UNIT (FRONT)
AutoCAD SHX Text
SIEMENS HEAT EXCHANGER INDOOR UNIT (FRONT)
AutoCAD SHX Text
(1 1/4" FEMALE NPT SUPPLIED AND INSTALLED BY CUSTOMER/ CONTRACTOR) ONLY IF COPPER PIPE IS USED WITH WATER HOSES.
AutoCAD SHX Text
SIEMENS SUPPLIED DN32 WATER HOSE WITH DN25 ADAPTER CUT TO SUIT OR 1 1/4" COPPER PIPES REQUIRE1 1/4" NPT FEMALE SUPPLIED AND INSTALLED BY CUSTOMER/ CONTRACTOR.
AutoCAD SHX Text
SIEMENS FLEXIBLE WATER HOSES
AutoCAD SHX Text
SIEMENS FLEXIBLE WATER HOSES
AutoCAD SHX Text
SCALE: NONE
AutoCAD SHX Text
PRIMARY CIRCUIT
AutoCAD SHX Text
PRIMARY CIRCUIT
AutoCAD SHX Text
WATER TEMPERATURE RANGE:
AutoCAD SHX Text
AIR INTAKE TEMPERATURE RANGE:
AutoCAD SHX Text
RELATIVE HUMIDITY:
AutoCAD SHX Text
-13° TO 136°
AutoCAD SHX Text
-22° TO 122°
AutoCAD SHX Text
0-100%
AutoCAD SHX Text
CHILLED WATER HOSES (PRIMARY LOOP)
AutoCAD SHX Text
BTU/HR TO AIR (EXHAUST):
AutoCAD SHX Text
119,425 MAXIMUM
AutoCAD SHX Text
CUSTOMER/ CONTRACTOR TO PROVIDE BLEED VALVES IF COPPER PIPE IS USE. REFER TO THE BELOW.
AutoCAD SHX Text
3
AutoCAD SHX Text
-13°F TO 136°FMAXIMUM
AutoCAD SHX Text
ANTIFREEZE (GLYKOSOL-N) ANTIFREEZE (GLYKOSOL-N) MAY BE ADDED TO THE PRIMARY COOLING CIRCUIT ONLY, BY THE TECHNICIAN WHO WILL START THE SYSTEM. THE PERMITTED MIXING RATIO IS 40:60 (GLYCOL:WATER) FOR TEMPERATURE RANGE -22° TO 122°F. 26.4 GALLONS OF ANTIFREEZE (GLYKOSOL-N) IS INCLUDED IN DELIVERY AND ADDITIONAL CAN BE ORDERED THROUGH SIEMENS LOCAL SERVICE. WATER CANNOT BE ADDED TO THIS MIXTURE. ANTIFREEZE (GLYKOSOL-N) IS NOT PERMITTED ON THE SECONDARY CIRCUIT FOR THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT COOLING CIRCUIT.
AutoCAD SHX Text
TEMPERATURE RANGE OF WATER
AutoCAD SHX Text
PRIMARY COOLING CIRCUIT
AutoCAD SHX Text
SECONDARY COOLING CIRCUIT
AutoCAD SHX Text
39°F TO 54°F
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TEMPERATURE GRADIENT OF WATER
AutoCAD SHX Text
6 K/MINUTE
AutoCAD SHX Text
NOMINAL OPERATING PRESSURE
AutoCAD SHX Text
29 TO 73 PSI AT OUTPUT OF THE PUMP
AutoCAD SHX Text
58 PSI TO 73 PSI AFTER FILLING PROCEDURE
AutoCAD SHX Text
WATER QUALITY POTABLE WATER QUALITY FOR BOTH THE PRIMARY AND SECONDARY CIRCUIT. OPERATION WITHOUT WATER WILL DESTROY THE PUMP IN THE COOLING SYSTEM. FOR THIS REASON, THE PUMP OF THE SECONDARY COOLING CIRCUIT CAN BE OPERATED ONLY WHEN FILLED ACCORDINGLY.
AutoCAD SHX Text
FILLING PRESSURE (STATIC)
AutoCAD SHX Text
29 PSI ± 2.9PSIAFTER FILLING PROCEDURE
AutoCAD SHX Text
29 PSI ± 3 PSIAFTER FILLING PROCEDURE
AutoCAD SHX Text
0% TO 100%
AutoCAD SHX Text
HEAT DISSIPATION
AutoCAD SHX Text
15% TO 85%
AutoCAD SHX Text
20 GALLONS WITHOUT HOSE EXTENSION
AutoCAD SHX Text
FILLING VOLUME
AutoCAD SHX Text
13.2 GALLONS WITHOUT HOSE EXTENSION
AutoCAD SHX Text
RELATIVE HUMIDITY
AutoCAD SHX Text
119,425 BTU/HR MAXIMUM TO THE AIR EXHAUST
AutoCAD SHX Text
3,412 BTU/HR (1.0 kW) TO ROOM AIR 61,473 BTU/HR 18.0 kW) TO WATER
AutoCAD SHX Text
BAROMETRIC PRESSURE
AutoCAD SHX Text
10.2 PSI TO 15.3 PSI
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TECHNICAL DATA
AutoCAD SHX Text
AMBIENT AIR TEMPERATURE RANGE (AIR INTAKE)
AutoCAD SHX Text
-22°F TO 122°FMAXIMUM
AutoCAD SHX Text
41°F TO 104°F

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

M-101-AATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
ORIENTATION POINT
AutoCAD SHX Text
SCANNER ROOM
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
SIEMENS SUPPLIED AND INSTALLED CONDENSATION HOSE ROUTED ACCORDING TO LOCAL CODE TO SELECTED ON-SITE DRAIN.
AutoCAD SHX Text
SIEMENS SUPPLIED AND INSTALLED WATER HOSE (SUPPLY) ROUTED IN (1) 3" CONDUIT, OR ACCORDING TO CODE, TO ON-SITE CHILLED WATER CONNECTION.
AutoCAD SHX Text
SIEMENS SUPPLIED WATER HOSE FROM THE FILTER ASSEMBLY TO THE GANTRY (SUPPLY). THE HOSE SHALL BE RUN IN (1) 3" CONDUIT ORACCORDING TO CODE.
AutoCAD SHX Text
SIEMENS SUPPLIED AND INSTALLED WATER HOSE (RETURN) ROUTED IN (1) 3" CONDUIT, OR ACCORDING TO CODE, TO ON-SITE CHILLED WATER CONNECTION.
AutoCAD SHX Text
FACILITY SUPPLIED CHILLED WATER.
AutoCAD SHX Text
M-101-A
AutoCAD SHX Text
1
AutoCAD SHX Text
M-501
AutoCAD SHX Text
3
AutoCAD SHX Text
M-501
AutoCAD SHX Text
2
AutoCAD SHX Text
M-501
AutoCAD SHX Text
1
AutoCAD SHX Text
M-101-A
AutoCAD SHX Text
1
AutoCAD SHX Text
M-101-A
AutoCAD SHX Text
1
AutoCAD SHX Text
NOTE: FOR THE HEAT OUTPUT (BTU/HR) OF ALL SIEMENS EQUIPMENT SHOWN ON THIS PLAN, SEE THE "EQUIPMENT LEGEND" ON SHEET A-101.
AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
MECHANICAL PLAN
AutoCAD SHX Text
SCALE: 1/4" = 1'-0"
AutoCAD SHX Text
WATER INPUT TEMPERATURE
AutoCAD SHX Text
SCALE: 1/2"=1'-0"
AutoCAD SHX Text
REQUIRED TEST FOR VERIFYING WATER PRESSURE DIFFERENTIAL
AutoCAD SHX Text
V
AutoCAD SHX Text
T
AutoCAD SHX Text
P
AutoCAD SHX Text
1-1/4" RETURN TO SITE
AutoCAD SHX Text
TESTING THE SITE WATER PUMP CAPACITY: BEFORE INSTALLING THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT SCANNER, IT IS IMPORTANT TO HAVE THE SITE CONFIRM THAT THE SITE WATER PUMP IS CAPABLE OF MAINTAINING MINIMUM PRESSURE DIFFERENTIAL WHEN OPERATING AT THE GIVEN WATER TEMPERATURE AND FLOW. IT IS THE RESPONSIBILITY OF THE CUSTOMER/CONTACTOR TO VALIDATE THE SITE WATER PUMP CAPACITY BY MEASURING THE PRESSURE DIFFERENTIAL AND BY VERIFYING THAT THE PRESSURE DIFFERENTIAL IS ADEQUATE AS OUTLINED BELOW. MEASURING THE DIFFERENTIAL PRESSURE: THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT DIFFERENTIAL PRESSURE TEST COMPONENTS OUTLINED BELOW SHOULD BE INSTALLED AT THE SITE PERMANENTLY. THE TEMPERATURE GAUGES SHOULD BE PLACED ON THE SUPPLY AND RETURN LINES AND THE FLOW METER CAN BE PLACED ON THE SUPPLY AND RETURN LINE. IT IS IMPORTANT TO INSTALL ONE PRESSURE GAUGE ON THE SUPPLY AND ONE PRESSURE GAUGE ON THE RETURN AS SHOWN BELOW. THE HOSE OR PIPE LABELED "TEMPORARY CONNECTION OF V AND V FOR DP TEST" IS ONLY INSTALLED FOR THE DIFFERENTIAL PRESSURE TEST AND WILL BE REMOVED BEFORE INSTALLATION OF THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT FILTER ASSEMBLY, SUPPLY/RETURN MANIFOLDS AND THE GANTRIES ARE INSTALLED.
AutoCAD SHX Text
P=PRESSURE GAUGE (O TO 150 PSI) F=20 GPM (4542 I/H) FLOW METER V =GLOBE VALVE V =GLOBE VALVE T=TEMPERATURE GAUGE ABOVE PROVIDE BY THE CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
PRESSURE DIFFERENTIAL TEST PROCEDURE
AutoCAD SHX Text
PRESSURE DIFFERENTIAL PSI
AutoCAD SHX Text
TOTAL FLOW GPM
AutoCAD SHX Text
7.9
AutoCAD SHX Text
6.7
AutoCAD SHX Text
9.0
AutoCAD SHX Text
7.3
AutoCAD SHX Text
10.1
AutoCAD SHX Text
8.0
AutoCAD SHX Text
11.4
AutoCAD SHX Text
8.9
AutoCAD SHX Text
13.0
AutoCAD SHX Text
10.0
AutoCAD SHX Text
15.1
AutoCAD SHX Text
11.2
AutoCAD SHX Text
18.0
AutoCAD SHX Text
12.9
AutoCAD SHX Text
22.2
AutoCAD SHX Text
15.2
AutoCAD SHX Text
28.6
AutoCAD SHX Text
18.8
AutoCAD SHX Text
39.2°F
AutoCAD SHX Text
41.0°F
AutoCAD SHX Text
42.8°F
AutoCAD SHX Text
44.6°F
AutoCAD SHX Text
48.2°F
AutoCAD SHX Text
50.0°F
AutoCAD SHX Text
51.8°F
AutoCAD SHX Text
46.4°F
AutoCAD SHX Text
53.6°F
AutoCAD SHX Text
1-1/4" SUPPLY FROM SITE
AutoCAD SHX Text
NOTE: IF THE PRESSURE DIFFERENTIAL (FOR THE TEMPERATURE AND FLOW LISTED IN THE TABLE ABOVE), IS LOWER THAN THE MINIMUM (AS DEFINED IN THE TABLE ABOVE), THE CUSTOMRE/CONTACTOR MUST PROVIDED AND INSTALL A TURBINE BOOSTER PUMP. THE DIFFERENTIAL PRESSURE MUST MEET OR EXCEED THE MINIMUM BEFORE THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT CAN BE INSTALLED. THE VALUES IN THE PRESSURE DIFFERENTIAL IN THE TABLE ABOVE CAN NOT BE USED DURING THE PET/CT WATER BALANCE PROCEDURE DESCRIBED IN THE PET/CT INSTALLATION INSTRUCTIONS. THE FLOW, PRESSURE AND TEMPERATURE GAUGES USED FOR THE PUMP CAPACITY TESTS CAN BE PERMANENTLY INSTALLED NEAR THE FILTER. IT IS RECOMMENDED THAT THE FLOW, PRESSURE AND TEMPERATURE GAUGES USED FOR THE PUMP CAPACITY TESTS BE LEFT IN PLACE FOR FUTURE REFERENCE.
AutoCAD SHX Text
V
AutoCAD SHX Text
T
AutoCAD SHX Text
P
AutoCAD SHX Text
F
AutoCAD SHX Text
TEMPORARY CONNECTION OF V AND V FOR DP TEST
AutoCAD SHX Text
FUTURE INSTALLATION OF THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT GANTRIES
AutoCAD SHX Text
RUNNING THE DIFFERENTIAL PRESSURE TEST: OPEN THE RETURN VALVE V ENTIRELY RECORD THE WATER INPUT TEMPERATURE FOR T (TEMPERATURE SOURCE) OPEN THE SUPPLY VALVE V UNTIL TOTAL FLOW (AS READ FROM THE FLOW METER) MATCH THE TOTAL FLOW REQUIRED, CORRESPONDING TO THE WATER INPUT TEMPERATURE AS SHOWN IN THE TABLE BELOW:
AutoCAD SHX Text
S
AutoCAD SHX Text
R
AutoCAD SHX Text
S
AutoCAD SHX Text
R
AutoCAD SHX Text
S
AutoCAD SHX Text
R
AutoCAD SHX Text
S
AutoCAD SHX Text
CALCULATE DIFFERENTIAL PRESSURE BY SUBTRACTING THE RETURN PRESSURE (P ) FROM THE SUPPLY PRESSURE (P ). CONFIRM THE DIFFERENTIAL PRESSURE MEETS THE MINIMUM SHOWN IN THE TABLE ABOVE FOR THE GIVEN WATER TEMPERATURE AND FLOW.
AutoCAD SHX Text
R
AutoCAD SHX Text
S
AutoCAD SHX Text
S
AutoCAD SHX Text
R
AutoCAD SHX Text
S
AutoCAD SHX Text
R
AutoCAD SHX Text
-
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
10 11
AutoCAD SHX Text
FINISHED ROOM HEIGHT
AutoCAD SHX Text
FOR BIOGRAPH mCT GANTRY ONLY
AutoCAD SHX Text
CAREVISION MONITOR/CEILING MOUNT
AutoCAD SHX Text
RECOMMEND 8'-0" MINIMUM 7'-5"
AutoCAD SHX Text
MIN. 8'-6 3/8" MAX. 12'-1 11/16"
AutoCAD SHX Text
CEILING MOUNT INJECTOR
AutoCAD SHX Text
MINIMUM 8'-9 1/2"
AutoCAD SHX Text
MINIMUM 39°F TO 54°F MAXIMUMTHE WATER TEMPERATURE OF THE ON-SITE COOLING CIRCUIT CANNOT BE LESS TAN 39°F.
AutoCAD SHX Text
OUTPUT TO ON-SITE COOLING WATER
AutoCAD SHX Text
61,473 BTU/HR (18.0 kW)
AutoCAD SHX Text
TEMPERATURE RANGE OF WATER
AutoCAD SHX Text
29 TO 87 PSI (145 PSI MAXIMUM)
AutoCAD SHX Text
NOMINAL OPERATING PRESSURE
AutoCAD SHX Text
MAXIMUM 1 K/MINUTE
AutoCAD SHX Text
TEMPERATURE GRADIENT OF WATER
AutoCAD SHX Text
POTABLE WATER QUALITY WITH PARTICLE SIZE < 0.25mm (250 MICRON). < 0.25mm (250 MICRON). 0.25mm (250 MICRON). 0.25mm (250 MICRON). THE pH REQUIREMENT OF CHILLED WATER MUST BE 8.0-9.5. INSTALL FILTER ASSEMBLY PROVIDED BY SIEMENS AND CUSTOMER/CONTACTOR. SITE PLUMBING TO BE PROVIDED BY SITE PLUMBER. SEE THE FILTER ASSEMBLY DETAIL FOR DESCRIPTION OF THE COMPONENTS PROVIDE BY SIEMENS AND CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
WATER QUALITY
AutoCAD SHX Text
TECHNICAL DATA
AutoCAD SHX Text
REFER TO THE REQUIRED TEST FOR VERIFYING WATER PRESSURE DIFFERENTIAL DETAIL
AutoCAD SHX Text
DIFFERENTIAL PRESSURE
AutoCAD SHX Text
DIFFERENTIAL PRESSURE DROPS SHOWN IN THE WATER PRESSURE DIFFERENTIAL DETAIL APPLY TO THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT SCANNER ONLY. CUSTOMER/CONTRACTOR SHALL MUST INSTALL A TURBINE BOOSTER PUMP. THE DIFFERENTIAL PRESSURE MUST MEET OR EXCEED THE MINIMUM BEFORE THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT CAN BE INSTALLED.
AutoCAD SHX Text
THE SHUT-OFF VALVES AND PRESSURE GAUGES THAT ARE PART OF THE FILTER ASSEMBLY IS PROVIDE BY SIEMENS. PRESSURE GAUGES, FLOW METER, ETC. THAT ARE MOUNTED ON THE SCANNER AND ARE PROVIDED BY SIEMENS. ADDITIONAL GAUGES (FLOW, TEMPERATURE, PRESSURE) ARE RECOMMENDED FOR THE EQUIPMENT ROOM AND ARE TO BE SUPPLIED BY THE CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
SERVICE
AutoCAD SHX Text
OUTPUT TO AIR ON-SITE COOLING WATER
AutoCAD SHX Text
683 BTU/HR (2.0 kW)
AutoCAD SHX Text
HOSPITAL SUPPLIED CHILLED WATER
AutoCAD SHX Text
FLOOR LINE
AutoCAD SHX Text
RETURN
AutoCAD SHX Text
SUPPLY
AutoCAD SHX Text
SIEMENS GANTRY (BACK SIDE)
AutoCAD SHX Text
EX. CABLE DUCT
AutoCAD SHX Text
PIPING AND FIXTURES TO BE SPECIFIED BY THE MECHANICAL ENGINEER OF RECORD AND TO BE SUPPLIED AND INSTALLED BY THE MECHANICAL CONTRACTOR. THE PIPE SIZE FOR TOTAL LENGTH OF PIPING LOOP IS 1 1/4". THE ON-SITE PIPES ARE TO BE INSULATED WITH 1/2" INSULATION BY THE CUSTOMER/CONTRACTOR. THE CHILLED WATER ON THE SUPPLY SIDE WILL TERMINATE WITH A 1 1/4" VALVE SUPPLIED AND INSTALL BY THE CUSTOMER/ CONTRACTOR (TO BE DETERMINE BY THE CUSTOMER/CONTRACTOR PER LOCAL CODE) AND WILL FEED DIRECTLY INTO A SIEMENS SUPPLIED AND INSTALLED FILTER ASSEMBLY. REFER TO THE FILER ASSEMBLY DETAIL. THE CHILLED WATER CONNECTIONS SITE VALVES (WHERE THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT WATER CONNECTS TO THE CUSTOMER SITE) ARE TO BE DESIGNED, CONSTRUCTED AND TESTED BY THE CUSTOMER/CONTRACTOR. THE SITE CONTROL VALVES MUST BE READILY AVAILABLE AND EASILY ACCESSIBLE FOR SHUT-OFF. IT IS RECOMMENDED THAT THE CONTROL VALVES BE LOCATED IN THE EQUIPMENT ROOM. THE CHILLED WATER DELIVERY SYSTEM MUST BE INSTALLED BY CUSTOMER/CONTRACTOR PRIOR TO EQUIPMENT DELIVERY. THE SYSTEM WILL BE CHARGED FOLLOWING THE INSTALLATION OF THE FILTER ASSEMBLY AND CONNECTION TO THE GANTRY.
AutoCAD SHX Text
(2) 1" HIGH PRESSURE WATER HOSES 1" HIGH PRESSURE WATER HOSES HIGH PRESSURE WATER HOSES SUPPLIED BY SIEMENS. MAX. LENGTH 65'-0".
AutoCAD SHX Text
FILTER ASSEMBLY MOUNT ON THE WALL. SIEMENS SUPPLIED AND INSTALLED. REFER TO FILTER ASSEMBLY DETAIL.
AutoCAD SHX Text
FILTER ASSEMBLY (REFER TO THE FILTER ASSEMBLY DETAIL).
AutoCAD SHX Text
PIPING SCHEMATIC CLOSED LOOP CHILLED WATER:
AutoCAD SHX Text
REQUIRED FIXTURE LEGEND
AutoCAD SHX Text
1" BALL VALVE (REFER TO THE FILTER ASSEMBLY DETAIL).
AutoCAD SHX Text
FACILITY CHILLED WATER COOLING CIRCUIT
AutoCAD SHX Text
1" HIGH PRESSURE HOSE SUPPLIED HIGH PRESSURE HOSE SUPPLIED HOSE SUPPLIED BY SIEMENS. MAX. LENGTH 16'-0".
AutoCAD SHX Text
SUPPLY/RETURN MANIFOLDS SUPPLY AND INSTALL BY SIEMENS.
AutoCAD SHX Text
(2) ADAPTER - 1 1/4" NPT SUPPLIED BY SIEMENS
AutoCAD SHX Text
(2) 1" BALL VALVE SUPPLIED BY SIEMENS.
AutoCAD SHX Text
1 1/4" SHUT OFF VALVE (1- SUPPLY) SUPPLIED BY CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
1 1/4" SHUT OFF VALVE (1- RETURN) SUPPLIED BY CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
1 1/4" SHUT OFF VALVE (1- SUPPLY, 1- RETURN) SUPPLIED BY CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
THE COOLING SYSTEM IS INTEGRATED WITHIN THE GANTRY. THE ON-SITE CHILLED WATER SUPPLY AND RETURN ARE CONNECTED DIRECTLY TO THE GANTRY.
AutoCAD SHX Text
ON-SITE CHILLED WATER REQUIREMENTS:
AutoCAD SHX Text
WATER TEMPERATURE: MINIMUM 39°F TO 54°F MAXIMUMNOMINAL OPERATING PRESSURE: 29 TO 87 PSI (145 PSI MAXIMUM) FLOW RATE: DEPENDS ON WATER TEMPERATURE DIFFERENTIAL PRESSURE: AS RELATES TO WATER CIRCULATION
AutoCAD SHX Text
CHILLED WATER HOSES:
AutoCAD SHX Text
A WATER DRAIN (EX. SINK) MUST BE AVAILABLE NEAR THE GANTRY FOR FILLING THE COOLING SYSTEM.
AutoCAD SHX Text
SIEMENS SUPPLIES THE 1" WATER SUPPLY AND WATER RETURN HOSE WITH 1/2" INSULATION FOR THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT THAT TERMINATE INTO THE SUPPLY/RETURN MANIFOLDS THAT ARE SUPPLIED AND INSTALLED BY SIEMENS. THE MAXIMUM LENGTH OF THE WATER SUPPLY AND WATER RETURN HOSES IS 65'-0". ACCEPTABLE BENDING RADIUS OF THE HOSES IS 9". THE OUTER DIAMETER OF THESE HOSES WITH THE INSULATION IS 1 1/2". RECOMMENDED CONDUIT SIZE IS 3 1/2" FOR A SINGLE WATER HOSE RUN PER CONDUIT. IF CONDUIT IS USED FOR THE SUPPLY/RETURN HOSES, 90°ELBOWS MUST BE AVOIDED, DUE TO THE HOSE BEND RADIUS AND THE NUMBER OF HOSES PROVIDED FOR THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT.
AutoCAD SHX Text
ON-SITE CHILLED WATER CONNECTION TO HOSPITAL COOLING CIRCUIT. CUSTOMER/CONTRACTOR TO SUPPLY 1 1/4" SHUT OFF VALVES (1-SUPPLY, 1- RETURN)
AutoCAD SHX Text
SIEMENS GANTRY (BACK SIDE)
AutoCAD SHX Text
WALL
AutoCAD SHX Text
EX. CABLE DUCT
AutoCAD SHX Text
FILTER ASSEMBLY MOUNT ON THE WALL. SIEMENS AND CUSTOMER/CONTRACTPR SUPPLIED. REFER TO FILTER ASSEMBLY DETAIL.
AutoCAD SHX Text
SCALE: NONE
AutoCAD SHX Text
FACILITY CHILLED WATER
AutoCAD SHX Text
(2) 1" HIGH PRESSURE 1" HIGH PRESSURE HIGH PRESSURE WATER HOSES SUPPLIED BY SIEMENS. MAX. LENGTH 65'-0".
AutoCAD SHX Text
1" WATER HOSE SUPPLIED BY SIEMENS. MAX. LENGTH 16'-0".
AutoCAD SHX Text
SUPPLY/RETURN MANIFOLDS SUPPLY AND INSTALL BY SIEMENS.
AutoCAD SHX Text
1

BIOGRAPH mCT FLOW 20 40 64 EDGE OR BIOGRAPH mCT-20 EXCEL S 40 S 64 X 128

13037

M-501ATTENTION:

RE

FE

RE

NC

E D

OC

UM

EN

T - N

OT F

OR

C

ON

STR

UC

TIO

N

AutoCAD SHX Text
TYPICAL FINAL DRAWING SET
AutoCAD SHX Text
AS NOTED
AutoCAD SHX Text
C. MC DONNELL
AutoCAD SHX Text
THE USE OR REPRODUCTION OF
AutoCAD SHX Text
THIS TITLE BLOCK WITHOUT
AutoCAD SHX Text
PROJECT #:
AutoCAD SHX Text
SHEET:
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
REF. #:
AutoCAD SHX Text
FULL EXTENT OF THE LAW.
AutoCAD SHX Text
SIEMENS AUTHORIZATION WILL
AutoCAD SHX Text
RESULT IN PROSECUTION UNDER
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DATE:
AutoCAD SHX Text
SHEET OF
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
DRAWN BY:
AutoCAD SHX Text
N/A
AutoCAD SHX Text
TYPICAL REV 13
AutoCAD SHX Text
-ISSUE BLOCK-
AutoCAD SHX Text
DESCRIPTION
AutoCAD SHX Text
DATE
AutoCAD SHX Text
SYM
AutoCAD SHX Text
BIOGRAPH mCT FLOW/BIOGRAPH mCT
AutoCAD SHX Text
REV 11
AutoCAD SHX Text
FILTER IS SUPPLIED AND INSTALLED BY SIEMENS USING THE LAG SCREWS THAT ARE PROVIDED WITH THIS EQUIPMENT. THE FILTER WILL BE MAINTAINED AND REPLACED BY SIEMENS SERVICE. THE SHUT-OFF VALVES, FILTER MUST BE READILY AVAILABLE AND EASILY ACCESSIBLE FOR SHUTOFF. IT IS RECOMMEND THAT THE SHUT-OFF VALUES BE LOCATED IN THE EQUIPMENT ROOM. SUPPLY SOURCE SHALL BE WITHIN 5 METERS (16'-4 7/8") OF THE FILTER PLACEMENT AND RETURN SOURCE SHALL BE WITHIN 20 METER (65'-7 3/8") OF THE GANTRY MANIFOLD CONNECTION. CUSTOMER/CONTRACTOR SUPPLIED ITEMS ARE LISTED ABOVE.
AutoCAD SHX Text
FILTER ASSEMBLY (WALL MOUNTED)
AutoCAD SHX Text
MOUNT FILTER TO WALL SUPPLIED BY CUSTOMER/CONTRACTOR. THE UNI-STRUT AND HANGERS SUPPLIED BY SIEMENS.
AutoCAD SHX Text
EXISTING WALL
AutoCAD SHX Text
MINIMUM
AutoCAD SHX Text
MINIMUM
AutoCAD SHX Text
BEND RADIUS 9" MAXIMUM TYPICAL
AutoCAD SHX Text
OUT TO SIEMENS SUPPLIED MANIFOLD IN GANTRY
AutoCAD SHX Text
SITE TO PROVIDE DN32 OR 1-1/4" PIPE FOR SUPPLY WATER. THE PIPE SHALL END WITH 1 1/4" NPT FEMALE OR 1 1/4" BSPP FEMALE FITTING WITHIN 5 METERS (16'-4 7/8") OF FILTER PLACEMENT. SIEMENS SHALL PROVIDE MALE ADAPTERS TO ACCOMMODATE 1 1/4" BSPP FEMALE FITTINGS.
AutoCAD SHX Text
FRONT VIEW
AutoCAD SHX Text
SIDE VIEW
AutoCAD SHX Text
SCALE: 1/2"=1'-0"
AutoCAD SHX Text
FILTER ASSEMBLY
AutoCAD SHX Text
(1) 1" HIGH PRESSURE WATER HOSE (16'-4 7/8") LONG
AutoCAD SHX Text
1" QUICK DISCONNECT FITTING
AutoCAD SHX Text
SUPPLY
AutoCAD SHX Text
FILTER AND FITTINGS
AutoCAD SHX Text
DIFFERENTIAL PRESSURE GAUGE
AutoCAD SHX Text
BEND RADIUS 9" MAXIMUM TYPICAL
AutoCAD SHX Text
(1) 1" HIGH PRESSURE WATER HOSE 20 METER (65'-7 3/8") LONG
AutoCAD SHX Text
PROVIDE BY SIEMENS
AutoCAD SHX Text
DRAIN PAN AND LAG SCREWS
AutoCAD SHX Text
UNI-STRUT MOUNTING HARDWARE
AutoCAD SHX Text
1" QUICK DISCONNECT FITTING OUT TO SIEMENS SUPPLIED MANIFOLD IN GANTRY
AutoCAD SHX Text
1" BALL VALVE
AutoCAD SHX Text
MINIMUM
AutoCAD SHX Text
1
AutoCAD SHX Text
A WATER DRAIN IS REQUIRED TO BE SUPPLIED AND INSTALLED BY THE CUSTOMER/CONTRACTOR TO DRAIN CONDENSED WATER FROM THE GANTRY. THE DRAIN MUST BE WITHIN 32'-9" OF THE GANTRY. THE HOSE MAY BE EXTENDED ON-SITE UP TO 65'-7" BY THE CUSTOMER/CONTRACTOR. BELOW ARE SOME EXAMPLES OF POSSIBLE WATER DRAINS. THE MECHANICAL CONTRACTOR IS RESPONSIBLE FOR DETERMINING WHICH DRAIN TYPE TO USE AND FOR SPECIFYING, SUPPLYING AND INSTALLING THE PLUMBING FIXTURES NECESSARY TO CONNECT THE SIEMENS WATER HOSE TO THE SELECTED DRAIN.
AutoCAD SHX Text
THE MAX. FLOW RATE OF THE CONDENSED WATER IS 0.53 GALLONS/HR AT 75% RELATIVE HUMIDITY.
AutoCAD SHX Text
1/2" DRAIN HOLE TO BELOCATED ON THE RIGHT SIDE OF THE GANTRY (WHEN VIEWED FROM THE FRONT). MIN. 4" FLOOR DRAIN AS REQUIRED PER LOCAL CODE. PROVIDE SURFACE MOUNTED PULL BOX OR SURFACE DUCT AS REQUIRED.
AutoCAD SHX Text
SINK DRAIN
AutoCAD SHX Text
FLOOR DRAIN
AutoCAD SHX Text
GANTRY (TOP VIEW FACING FRONT OF THE GANTRY)
AutoCAD SHX Text
GANTRY (BACK SIDE)
AutoCAD SHX Text
WALL
AutoCAD SHX Text
SINK
AutoCAD SHX Text
SIEMENS CONDENSATION WATER HOSE (LENGTH: 32'-9", OUTSIDE DIA. 3/8"). MAX. BENDING RADIUS 1 3/16".
AutoCAD SHX Text
WATER DRAIN
AutoCAD SHX Text
HOSE OUTLET
AutoCAD SHX Text
EX. CABLE DUCT
AutoCAD SHX Text
CONDENSATION WATER DRAIN
AutoCAD SHX Text
HEIGHT A.F.F.
AutoCAD SHX Text
3
AutoCAD SHX Text
IT IS RECOMMENDED THAT THE SIEMENS DRAWINGS BE INCORPORATED WITH THE CONSTRUCTION
AutoCAD SHX Text
THIS DRAWING IS DESIGNED TO CONFORM TO FEATURES AND EQUIPMENT REQUIREMENTS PRESENTED
AutoCAD SHX Text
AT THE TIME OF THEIR PREPARATION. SINCE BOTH THESE FACTORS ARE SUBJECT TO DESIGN
AutoCAD SHX Text
MODIFICATION, THEY ARE NOT TO BE USED FOR CONSTRUCTION PURPOSES.
AutoCAD SHX Text
DOCUMENTS FOR REFERENCE.
AutoCAD SHX Text
THIS DRAWING DOES NOT PROVIDE RADIATION SHIELDING REQUIREMENTS FOR X-RAY AND ASSOCIATED
AutoCAD SHX Text
EQUIPMENT. THE CUSTOMER IS RESPONSIBLE FOR CONSULTING WITH A REGISTERED RADIATION
AutoCAD SHX Text
PHYSICIST TO SPECIFY RADIATION PROTECTION.
AutoCAD SHX Text
ALL DIMENSIONS SHOWN ON THIS DRAWING ARE FROM FINISHED SURFACES.
AutoCAD SHX Text
THIS SET OF PLANS REPRESENTS A COMPLETE SET OF DETAILS AND SHOULD NOT BE SEPARATED.
AutoCAD SHX Text
ALL RIGHTS ARE RESERVED.
AutoCAD SHX Text
11 11
AutoCAD SHX Text
1/2"=1'-0"
AutoCAD SHX Text
SCALE:
AutoCAD SHX Text
OUTDOOR COOLING UNIT MOUNTING DETAIL
AutoCAD SHX Text
SIDE VIEW
AutoCAD SHX Text
TOP VIEW
AutoCAD SHX Text
WEATHER PROTECTION
AutoCAD SHX Text
AIR INTAKE
AutoCAD SHX Text
CABLE CONNECTION BOX
AutoCAD SHX Text
MIN. SERVICE CLEARANCE
AutoCAD SHX Text
THE OUTDOOR UNIT MUST BE LOCATED AND INSTALLED BY THE CUSTOMER/CONTRACTOR. THIS UNIT IS PART OF A PRE-INSTALLATION KIT THAT MUST BE INSTALLED PRIOR TO THE DELIVERY OF THE SYSTEM. IT MAY BE LOCATED OUTSIDE ON THE ROOF OR GROUND, OR IN AN AIR CONDITIONED ROOM IF AN OUTSIDE INSTALLATION IS NOT POSSIBLE. THE MOUNTING SURFACE MUST BE LEVEL. IT IS RECOMMENDED THAT THE UNIT BE PROTECTED AGAINST RAIN, SNOW, ETC. THE OUTDOOR UNIT MUST ALWAYS BE BOLTED TO THE FLOOR THROUGH THE DRILL HOLES IN THE FRAME. MATERIALS FOR BOLTING MUST BE SUPPLIED BY THE CUSTOMER/CONTRACTOR. MOUNTING TO A WALL IS ALSO POSSIBLE. MATERIALS FOR MOUNTING AND STRUCTURAL SUPPORT MUST BE DESIGNED, SUPPLIED AND INSTALLED BY THE CUSTOMER/CONTRACTOR.
AutoCAD SHX Text
FRONT VIEW
AutoCAD SHX Text
WALL MOUNTING - NO SCALE
AutoCAD SHX Text
OUTDOOR COOLING UNIT
AutoCAD SHX Text
ON-SITE SUPPORT
AutoCAD SHX Text
WALL
AutoCAD SHX Text
ON-SITE WEATHER PROTECTION (RECOMMENDED)
AutoCAD SHX Text
NOTE: SPACE FOR SERVICE MUST ALSO BE MAINTAINED
AutoCAD SHX Text
WATER INLET & RETURN
AutoCAD SHX Text
MIN. SERVICE CLEARANCE
AutoCAD SHX Text
WATER INLET & RETURN
AutoCAD SHX Text
COOLING FAN (EXHAUST)
AutoCAD SHX Text
MIN. FOR EXHAUST
AutoCAD SHX Text
WATER INLET & RETURN
AutoCAD SHX Text
4
AutoCAD SHX Text
DETAIL "A"
AutoCAD SHX Text
WATER RETURN HOSE CONNECTION
AutoCAD SHX Text
GANTRY TOP VIEW
AutoCAD SHX Text
GANTRY FRONT VIEW
AutoCAD SHX Text
2-1" HIGH PRESSURE WATER HOSES WITH 1/2" INSULATION FOR CONDENSATION TO BE USED FOR THE WATER SUPPLY AND RETURN SUPPLIED BY SIEMENS. WATER CONNECTION TO THE BIOGRAPH mCT FLOW OR BIOGRAPH mCT IS ACCOMPLISHED THROUGH SUPPLY/RETURN MANIFOLDS SHOWN IN DETAIL "A". THE SUPPLY/RETURN MANIFOLDS ARE SUPPLIED AND INSTALLED BY SIEMENS. THE PIPE FITTINGS MUST BE ORIENTED AS SHOWN IN THE FIGURE ABOVE, WITH THE LARGER FEMALE CONNECTOR POINTING TO THE RIGHT OF THE GANTRY (WHEN VIEWED FROM THE FRONT). IT IS RECOMMENDED THAT THE PLUMBING INSTALLER DESIGNATES THE SUPPLY LINE (THE ONE ON THE LEFT). THE COOLING WATER SUPPLY TO THE GANTRY (INCLUDING ALL FITTINGS, PRESSURE REGULATORS AND CONTROL VALVES) ARE TO BE DESIGNED, CONSTRUCTED AND TESTED BY THE CUSTOMER/CONTRACTOR. THE FILTER ASSEMBLY IS DESIGNED, DELIVERED AND INSTALLED BY SIEMENS AND ONLY REQUIRED WHEN USING FACILITY CHILLED WATER.
AutoCAD SHX Text
WATER SUPPLY/WATER RETURN HOSES ORIENT TO THIS SIDE OF GANTRY
AutoCAD SHX Text
DETAIL "A"
AutoCAD SHX Text
WATER SUPPLY HOSE CONNECTION
AutoCAD SHX Text
GANTRY SIDE VIEW
AutoCAD SHX Text
FRONT OF GANTRY
AutoCAD SHX Text
REAR OF GANTRY
AutoCAD SHX Text
WATER SUPPLY/RETURN MANIFOLD SIEMENS SUPPLIED AND INSTALLED
AutoCAD SHX Text
SUPPLY AND RETURN WATER CORE LOCATION
AutoCAD SHX Text
WATER SUPPLY/RETURN MANIFOLD SIEMENS SUPPLIED AND INSTALLED
AutoCAD SHX Text
SUPPLY/RETURN MANIFOLD CONNECTIONS
AutoCAD SHX Text
SCALE: NONE
AutoCAD SHX Text
WATER SUPPLY HOSE CONNECTION
AutoCAD SHX Text
1