RENASYS Softport Quick reference guide - Remove paper tape around the canister tubing to release tubing to the full length - Open both of the orange clips - Align the canister so that the volume marks are facing forward - Push the canister gently over the inlet port of the device - Engage both orange clips (clips will click when they are properly engaged) 1. Installing the canister 2. Start up of device 3. Select vacuum setting 4. Start therapy Use “Up” and “Downto adjust therapy level. Press “Select” to start therapy. Press and hold the “Power” button for 2 seconds until start up message appears. Look for “raisin-like” appearance Listen for a “hissing” sound, indicating a leak Feel the dressing, which should be hard to the touch 5. Check dressing for seal Possible sites of leak Where the Soft Port attaches to the film Wrinkles or crease Skin fold or crevice Intervention Pinch paste around drain (if applicable) Secure area where tube exits with blue tape (if applicable) Seal film edges with blue tape or additional transparent film If hole found in transparent film, patch with additional transparent film Welcome Starting V X X Symbols Power button Turns the device on and off Battery indication Shows the status of battery life. Flashes when the battery life reaches levels that require user intervention Up selector Allows the pressure setting to be increased and scroll through menu options Down selector Allows the pressure setting to be decreased and scroll through menu options Keypad lock Locks the keypad to restrict accidental adjustment of therapy. When activated the light will illuminate Audio pause/alarm suppress Silences the alarm for approximately 2-3 minutes. Start therapy/select Allows therapy to be started or paused. It is also used to confirm settings within therapy Alarm messages The battery has up to 3 hours therapy time remaining The battery has up to 1 hour therapy time remaining The battery has only 2 minutes of therapy time remaining After 2 minutes in the extremely low state the device will power off The vacuum level is low or there is a leak in the system for longer than 30 seconds (see User Manual for details for resolution) The system has encountered a high vacuum condition (see User Manual for details for resolution) The system has encountered an excessively high vacuum (of greater than 235mmHg) (see User Manual for details for resolution) The system detects the canister is full or there is a blockage in the system (see User Manual for resolution) The system has detected a significant leak for more than 1 minute (see User Manual for resolution) The device has an unrecoverable error (see User Manual for resolution) The device has been left in standby for more than 15 minutes (see User Manual for resolution) ! Attention ! Low Battery !! RECHARGE !! V. Low Battery !! RECHARGE !! EX. LOW BATTERY !! RECHARGE NOW !! BATTERY FAIL !! WARNING !! LOW VACUUM !! WARNING !! HIGH VACUUM !! THERAPY STOP !! OVER VACUUM !! WARNING !! BLOCKAGE/FULL !! WARNING !! LEAK !!! DEVICE FAILED !!! Please Return ! Attention ! INACTIVE

Upload: others

Post on 17-Sep-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: RENASYS Softport · 2018. 1. 4. · 3 hours therapy time remaining The battery has up to 1 hour therapy time remaining The battery has only 2 minutes of therapy time remaining After

RENASYS Softport Quick reference guide

- Remove paper tape around the canister tubing to release tubing to the full length- Open both of the orange clips- Align the canister so that the volume marks are facing forward- Push the canister gently over the inlet port of the device- Engage both orange clips (clips will click when they are properly engaged)

1. Installing the canister

2. Start up of device 3. Select vacuum setting 4. Start therapy

Use “Up” and “Down” to adjust therapy level.

Press “Select” to start therapy.

Press and hold the “Power” button for 2 seconds until start up message appears.

Look for “raisin-like” appearance

Listen for a “hissing” sound, indicating a leak

Feel the dressing, which should be hard to the touch

5. Check dressing for seal

Possible sites of leak • Where the Soft Port

attaches to the film• Wrinkles or crease• Skin fold or crevice

Intervention Pinch paste around drain (if applicable)Secure area where tube exits with blue tape (if applicable)Seal film edges with blue tape or additional transparent filmIf hole found in transparent film, patch with additional transparent film

WelcomeStarting V X X

Symbols Power button Turns the device on and off

Battery indication Shows the status of battery life. Flashes when the battery life reaches levels that require user intervention

Up selector Allows the pressure setting to be increased and scroll through menu options

Down selector Allows the pressure setting to be decreased and scroll through menu options

Keypad lock Locks the keypad to restrict accidental adjustment of therapy. When activated the light will illuminate

Audio pause/alarm suppress Silences the alarm for approximately 2-3 minutes.

Start therapy/select Allows therapy to be started or paused. It is also used to confirm settings within therapy

Alarm messages

The battery has up to 3 hours therapy time remaining

The battery has up to 1 hour therapy time remaining

The battery has only 2 minutes of therapy time remaining

After 2 minutes in the extremely low state the device will power off

The vacuum level is low or there is a leak in the system for longer than 30 seconds (see User Manual for details for resolution)

The system has encountered a high vacuum condition (see User Manual for details for resolution)

The system has encountered an excessively high vacuum (of greater than 235mmHg) (see User Manual for details for resolution)

The system detects the canister is full or there is a blockage in the system (see User Manual for resolution)

The system has detected a significant leak for more than 1 minute (see User Manual for resolution)

The device has an unrecoverable error (see User Manual for resolution)

The device has been left in standby for more than 15 minutes (see User Manual for resolution)

! Attention! Low Battery

!! RECHARGE!! V. Low Battery

!! RECHARGE!! EX. LOW BATTERY

!! RECHARGE NOW!! BATTERY FAIL

!! WARNING!! LOW VACUUM

!! WARNING!! HIGH VACUUM

!! THERAPY STOP!! OVER VACUUM

!! WARNING!! BLOCKAGE/FULL

!! WARNING!! LEAK

!!! DEVICE FAILED!!! Please Return

! Attention! INACTIVE

Page 2: RENASYS Softport · 2018. 1. 4. · 3 hours therapy time remaining The battery has up to 1 hour therapy time remaining The battery has only 2 minutes of therapy time remaining After

4

5

7

9

10

11

12

14

13

8

1

2

3

6

Debride any devitalized or necrotic eschar tissue. Cleanse wound bed and pat dry.

If desired, apply skin sealant to peri-wound skin to protect from exposure to moisture and adhesive.

If desired, a non-adherent dressing may be applied. Trim a single layer of non-adherent gauze and lay across the wound bed.

Apply a layer of saline-moistened antimicrobial gauze to wound bed. Continue to apply in layers, until the gauze loosely fills the entire wound cavity.

Precaution: If multiple pieces of wound filler are needed to fill the wound cavity, count and record how many pieces are present to ensure all wound filler pieces are removed at a dressing change.

While holding transparent film, expose one side of the adhesive backing by removing a single panel, and apply over the wound.

Cover wound filler with transparent film, removing adhesive panels to seal, then top stabilization panel.

Film should extend at least 5cm beyond the wound margin and be securely anchored to periwound area to maintain a good seal. When using multiple pieces of film overlap the edges by a minimum of 7.5cm.

Cut a small opening (no less than 2cm) in the centre of the film, over the wound filler. Remove any loose transparent film and dispose of away from the wound.

Remove the adhesive panel from the Soft Port dressing, and align the opening directly over the cut hole in the transparent film.

Use gentle pressure to anchor the Soft Port to the transparent film.

Smooth the dressing down while removing the Soft Port’s top stabilization frame.

Secure the Soft Port to the patient according to local protocol. Ensure the aeration disc, located near the quick click connector is not covered.

Connect the RENASYS Soft Port to the canister tubing by pushing the quick click connectors together.

Activate the RENASYS device and adjust to the prescribed therapy level. Finished dressing should be hard to the touch and leak-free.

Dressing the wound: RENASYS™-G with Soft Port Dressing Dressing the wound: RENASYS -F with Soft Port Dressing

6

4

5

7

8

9

10

11

12

13

14

Dressing changesRENASYS™-G (gauze) dressings should be changed 48 hours after the initial application of therapy. If no leak is present and the patient is comfortable, dressing changes should occur 2-3 times per week.RENASYS-F (foam) dressings should be changed every 48 to 72 hours after the initial application of therapy. If no leak is present and the patient is comfortable, dressing changes should occur no less than 3 times per week.These are recommended guidelines only as many factors influence dressing change frequency, such as heavy drainage, drainage with sediment or infected wounds. Check dressings regularly and monitor the wound to check for signs of infection. Infected wounds may require more frequent dressing changes. If there any signs of systemic infection or advancing infection at the wound site contact the treating clinician immediately.

If RENASYS-F with Soft Port dressing adheres to the woundAfter powering the device down, apply normal saline into the wound dressing and let it soak for 15-30 minutes before gently removing the foam. Dispose of the dressing in accordance with local guidelines.

Canister selectionThe RENASYS GO device can only be used with 300ml (66800914) or Large (66800916) canister kit. Always use the smallest canister possible. Canister kits are single patient use. Do not reuse. The canister kit should be changed when the canister is full or at least once a week. If the canister shows any sign of damage, discard and use a new canister.

Capping off tubing

Connecting the tubing Disconnect the tubing Cap the tubing

Modes of operationThere are 2 modes of operation – Continuous and Intermittent.

Note: Continuous mode is recommended for NPWT. In the Continuous mode, the device will maintain the selected vacuum level without stopping until switched off.In the Intermittent mode, the device cycles on and off in increments of 5 minutes “ON” (active vacuum) and 2 minutes “OFF” (no vacuum).

To change the therapy between Continuous and Intermittent mode:1. Turn the device off.

2. Simultaneously press the Down + Select + Power buttons for 2 seconds.

3. Press the Up or Down buttons to move between

Continuous and Intermittent mode and press the Select button to confirm.

4. Once therapy is started the display will show which mode of therapyhas been selected.

Supporting healthcare professionals for over 150 years

Questions?For 24/7 NPWT clinical support, call 1-800-463-7439

Smith & Nephew Inc. www.smith-nephew.com/canada2280 Argentia RoadMississauga, ON L5N 6H8

T 1 800 463 7439 F 1 800 671 9140

™Trademark of Smith & Nephew© Smith & Nephew September 2015

3

1

Debride any devitalized or necrotic eschar tissue. Cleanse wound bed and pat dry.

If desired, apply skin sealant to peri-wound skin to protect from exposure to moisture and adhesive.

If desired, a non-adherent dressing may be applied. Trim a single layer of non-adherent gauze and lay across the wound bed.

Cut the foam dressing to fit the size and shape of the wound and place the cut foam into the wound. Avoid overpacking. Foam should completely fill the wound cavity. It may be necessary to stack pieces of foam in deep wounds.

Precaution: If multiple pieces of wound filler are needed to fill the wound cavity, count and record how many pieces are present to ensure all wound filler pieces are removed at dressing change.

While holding transparent film, expose one side of the adhesive backing by removing a single panel, and apply over the wound.

Cover wound filler with transparent film, removing adhesive panels to seal, then top stabilization panel.

Film should extend at least 5cm beyond the wound margin and be securely anchored to periwound area to maintain a good seal. When using multiple pieces of film overlap the edges by a minimum of 7.5cm.

Cut a small opening (no less than 2cm) in the centre of the film, over the wound filler. Remove any loose transparent film and dispose of away from the wound.

Remove the adhesive panel from the Soft Port dressing, and align the opening directly over the cut hole in the transparent film.

Use gentle pressure to anchor the Soft Port to the transparent film.

Smooth the dressing down while removing the Soft Port’s top stabilization frame.

Secure the Soft Port to the patient according to local protocol. Ensure the aeration disc, located near the quick click connector is not covered.

Connect the RENASYS Soft Port tothe canister tubing by pushing the quick click connectors together.

Activate the RENASYS device and adjust to the prescribed therapy level. Finished dressing should be hard to the touch and leak-free.

2

Note: Continuous mode is recommended for NPWT.

RENA-AG-1509EN (CA99145)