Safety instructionsVEGACONNECTCONNECT.CXA4PTB 07 ATEX 2013 X

II (1)GD [Ex ia] IIC II 1G, II 2G Ex ia IIC T6

Document ID: 33002

Page 2: Safety instructions VEGACONNECT CONNECT

2 VEGACONNECT CONNECT.CXA4

33002-EN-180307

Contents1 Area of applicability ................................................................................................................. 4

2 General information ................................................................................................................. 4

3 Technical data .......................................................................................................................... 43.1 Use as appropriate instrument ......................................................................................... 43.2 Use as intrinsically safe instrument .................................................................................. 6

4 Protection against static electricity ....................................................................................... 6

5 Mounting/installation of the communication versions ........................................................ 65.2 Use as appropriate instrument integrated in the connection box, communication via I²C-

Bus (CONNECT.CXA4) .................................................................................................... 65.3 Use as appropriate instrument integrated in a VEGA sensor, communication via I²C-Bus

(CONNECT.CXX4) ........................................................................................................... 85.4 Use as appropriate instrument integrated in the junction box, communication via HART

(CONNECT.CXA4) ........................................................................................................... 85.5 Use as intrinsically safe instrument, PLICSCOM integrated in the junction box, applica-

tion as handheld (CONNECT.CXAP) ................................................................................ 9

6 USB connection cable ........................................................................................................... 10

7 Certificationsofthesensorssuitableforconnection ........................................................ 10

Please note:These safety instructions are part of the documentation:32628 - VEGACONNECT 433005 - EC type approval certificate PTB 07 ATEX 2013 XEditing status: 2018-03-06

Page 3: Safety instructions VEGACONNECT CONNECT

3VEGACONNECT CONNECT.CXA4

3300

2-EN

-180

307

DE Sicherheitshinweisefür den Einsatz in explosionsgefährdeten Bereichen

EN Safety instructionsfor the use in hazardous areas

FR Consignes de sécuritépour une application en atmosphères explosibles

IT Normative di sicurezzaper l'impiego in luoghi con pericolo di esplosione

ES Instrucciones de seguridadpara el empleo en áreas con riesgo de explosión

PT Normas de segurançapara utilização em zonas sujeitas a explosão

NL Veiligheidsaanwijzingenvoor gebruik op plaatsen waar ontploffingsgevaar kan heersen

SV Säkerhetsanvisningarför användning i explosiionsfarliga områden

DA Sikkerhedsforskriftertil anvendelse i explosionsfarlig atmosfare

FI Turvallisuusohjeeträjähdysvaarallisisssa tiloissa käyttöä varten

EL Υποδείξεις ασΦαλείαςγια τη χρησιμοποίηση σε περιοχές που υπάρχει κίνδυνος έκρηξης

DE Die vorliegenden Sicherheitshinweise sind im Download unter www.vega.com standard-mäßig in den Sprachen deutsch, englisch, französisch und spanisch verfügbar. Weitere EU-Landessprachen stellt VEGA nach Anforderungen zur Verfügung.

EN These safety instructions are available as a standard feature in the download area under www.vega.com in the languages German, English, French and Spanish. Further EU languag-es will be made available by VEGA upon request.

FR Les présentes consignes de sécurité sont disponibles au téléchargement sous www.vega.com en standard en allemand, en anglais, en francais et en espagnol. VEGA met à disposition d'autres langues de l'Union Européenne selon les exigences.

ES Las indicaciones de seguridad presentes están disponibles en la zona de descarga de www.vega.com de forma estándar en los idiomas inglés, francés y español. VEGA pone a dis-posición otros idiomas de la UE cuando son requeridos.

Page 4: Safety instructions VEGACONNECT CONNECT

4 VEGACONNECT CONNECT.CXA4

33002-EN-180307

1 Area of applicabilityThese safety instructions apply to the interface adapter VEGACONNECT CONNECT.CX** accord-ing to the EC type approval certificate PTB 07 ATEX 2013 X (certification number on the type label).

2 General informationThe VEGACONNECT is an interface converter for connection of communication capable VEGA instruments to the USB interface of a Windows PC. Alternatively it can also be used as a universal HART modem for sensors of other manufacturers. The signals and protocols of the USB interface are converted into the industrial HART or I²C protocol. With an additional indicating and adjustment module it can also be used as a handheld.The VEGACONNECT consists of a junction box into which the communication module VEGACONNECT 4 is installed. Here, it works either as HART or I²C interface converter depending on the connected sensor.Alternatively the communication module VEGACONNECT 4 can be directly used on intrinsically safe circuits in VEGA instruments of plics® series. In this version it operates as I²C interface con-verter.In addition the junction box can also be used as display and adjustment module PLICSCOM. This intrinsically safe version enables parameter adjustment of a plics® sensor via the intrinsically safe I²C interface and is designated as handheld.The following versions are certified for VEGACONNECT:

Interface converter in junction box (CONNECT.CXA4)As appropriate instrument it is used as HART or I²C interface converter. The communication module VEGACONNECT 4 is inserted in the junction box.

Interface converter in plics® instrument (CONNECT.CXX4)As appropriate instrument it is used as I²C interface converter. The communication module VEGACONNECT 4 is directly inserted in a VEGA-plics® instrument.

Handheld (CONNECT.CXAP)As intrinsically safe instrument and in conjunction with a PLICSCOM the connection box is used for remote indication and adjustment of a plics® sensor with respectively certified, intrinsically safe I²C interface. No additional instruments or software is necessary.If VEGACONNECT 4 is used as appropriate instrument, it should only be operated outside the haz-ardous area. If there is no explosive atmosphere, the use within this hazardous area is also allowedIf VEGACONNECT is used as intrinsically safe instrument (handheld), it can also be operated in explosive areas of all combustible materials of gas groups IIA, IIB and IIC requiring category 1G or 2G instruments.If the instrument is installed and operated as accessory or intrinsically safe instrument, the general Ex mounting instructions EN 60079-14 as well as these safety instructions must be observed.The operating instructions as well as the installation regulations and standards that apply for explo-sion protection of electrical systems must always be observed.The installation of explosion-protected systems must always be carried out by qualified personnel.

3 Technical data

3.1 Use as appropriate instrumentThe following data are valid for the versions CONNECT.CXA4 and CONNECT.CXX4 in the function as interface converter. The VEGACONNECT contains a non intrinsically safe (USB) and two intrinsi-

Page 5: Safety instructions VEGACONNECT CONNECT

5VEGACONNECT CONNECT.CXA4

3300

2-EN

-180

307

cally safe circuits (I²C/HART). The I²C and HART circuit are galvanically connected and should not be connected simultaneously.Voltage supply via USB interface (non intrinsically safe circuit)Max. operating voltage 6 V DCReference voltage Um 16 V AC/DC

Signal circuit I²C/HART (intrinsically safe circuits) Ex ia IICCommunication

Ʋ I²C interface Connection via I²C plug or sliding contacts Ʋ HART interface Connection via 2 mm plug

Max. terminal voltage Uo 6 VMax. terminal voltage Ui

Ʋ I²C interface 6 V Ʋ HART interface 30 V

Max. current Io Ʋ I²C interface 198 mA Ʋ HART interface 3.7 mA

Max. power Po Ʋ I²C interface 327 mW Ʋ HART interface 5.6 mW

Max. power Pi Ʋ I²C interface 360 mW Ʋ HART interface -

Effective internal capacitance Ci Ʋ I²C interface 0 nF Ʋ HART interface 1.2 nF

Effective internal inductance Li Ʋ I²C interface 0 mH Ʋ HART interface 0 mH

Permissible external capacity Co Ʋ I²C interface 40 µF Ʋ HART interface 40 µF

Permissible external inductance Lo Ʋ I²C interface 0.8 mH Ʋ HART interface 1000 mH

Separation of intrinsically safe - non intrinsically safe circuit Ʋ Peak value of nominal voltage 375 V

Ambient conditionsAmbient temperature1) -20 … +60 °C (-4 … +140 °F)

1) If VEGACONNECT 4 is directly mounted into a sensor, please also observe the temperature class table of the respective sensors.

Page 6: Safety instructions VEGACONNECT CONNECT

6 VEGACONNECT CONNECT.CXA4

33002-EN-180307

Ambient temperature with PLICSCOM Ʋ T6 -20 … +46 °C (-4 … +115 °F) Ʋ T5, T4, T3, T2, T1 -20 … +60 °C (-4 … +140 °F)

Electrical protective measuresProtection rating IP 40

3.2 Use as intrinsically safe instrumentVEGACONNECT as handheldFor applications with integrated PLICSCOM (version CONNECT.CXX*), please take note of the temperature classification table of the VEGA sensor on which the handheld is operated. All certified VEGA-plics® sensors can be used (please compare list of certifications).

4 Protection against static electricityWhen used as intrinsically safe instrument in the version "Handheld" the housing is provided with a warning label referring to the safety instructions in case of electrostatic charge during operation.

Caution: Plastic parts! Danger of electrostatic charging!

• Avoid friction• No dry cleaning

5 Mounting/installation of the communication versionsFor applications with USB communication the VEGACONNECT 4 integrated in the junction box (CONNECT.CXA4) or in a VEGA sensor (CONNECT.CXX4), must be operated on an intrinsically safe circuit only for service purposes. Here, the VEGACONNECT must be operated outside the explosive area, or it must be guaranteed that there is no explosive atmosphere during operation.The I²C- and HART connection cable must not be wired simultaneously.

5.2 Use as appropriate instrument integrated in the connection box, communication via I²C-Bus (CONNECT.CXA4)

Application variety 1The VEGACONNECT with connected notebook, PLC or modem must only be operated within the safe area outside the explosive area.

Page 7: Safety instructions VEGACONNECT CONNECT

7VEGACONNECT CONNECT.CXA4

3300

2-EN

-180

307

3

1 2

4

5

OPEN

TWIST

USB

LOCK

Fig. 2: Use as appropriate instrument in the safe area1 Safe area2 Explosive area (zone 0 or 1)3 VEGACONNECT 44 Junction box5 I²C cable

Application variety 2The VEGACONNECT with connected notebook, PLC or modem must only be operated in the explo-sive area of zone 1, if an explosive atmosphere can be excluded.

3

1 2

4

5

OPEN

TWIST

USB

LOCK

Fig. 3: Use as appropriate instrument in the explosive area (zone 1) without explosive atmosphere1 Explosive area (zone 1) without explosive atmosphere2 Explosive area (zone 0 or 1)3 VEGACONNECT 44 Junction box5 I²C cable

Page 8: Safety instructions VEGACONNECT CONNECT

8 VEGACONNECT CONNECT.CXA4

33002-EN-180307

5.3 Use as appropriate instrument integrated in a VEGA sensor, com-munication via I²C-Bus (CONNECT.CXX4)

The VEGACONNECT with connected notebook, PLC or modem must only be operated in the explo-sive area of zone 1, if an explosive atmosphere can be excluded.

3

1 2

Fig. 4: Use as appropriate instrument in the explosive area (zone 1) without explosive atmosphere1 Explosive area (zone 1) without explosive atmosphere2 Explosive area (zone 0 or 1)3 VEGACONNECT 4 integrated in VEGA sensor

5.4 Use as appropriate instrument integrated in the junction box, com-munication via HART (CONNECT.CXA4)

The VEGACONNECT with connected notebook, PLC or modem must only be operated in the safe area outside the explosive area. The connection is carried out via HART communication on the intrinsically safe signal and power supply.

Page 9: Safety instructions VEGACONNECT CONNECT

9VEGACONNECT CONNECT.CXA4

3300

2-EN

-180

307

3

1 2

4

5

6 7

OPEN

TWIST

USB

LOCK

VEGAMET 381

on

1

2

1 2

%

Fig. 5: Use as appropriate instrument in the safe area1 Safe area2 Explosive area (zone 0 or 1)3 VEGACONNECT 44 Junction box5 HART cable6 Power supply unit e.g. VEGAMET, VEGATRENN, PLC7 Intrinsically safe signal and power supply circuit

5.5 Use as intrinsically safe instrument, PLICSCOM integrated in the junction box, application as handheld (CONNECT.CXAP)

The junction box with integrated indication and adjustment module PLICSCOM may be operated within the explosive area of zone 0 or 1. Please observe the temperature classification table of VEGACONNECT and the respective sensor.

Page 10: Safety instructions VEGACONNECT CONNECT

10 VEGACONNECT CONNECT.CXA4

33002-EN-180307

31 2

4

5

OPEN

TWIST

LOCK

Fig. 6: Use as intrinsically safe instrument in the explosive area (zone 0 or 1)1 Safe area2 Explosive area (zone 0 or 1)3 Display and adjustment module PLICSCOM4 Junction box5 I²C cable

6 USB connection cableIf the supplied USB connection cable is replaced by another one, make sure that the insulation thickness of the cable is over 0.65 mm.

7 CertificationsofthesensorssuitableforconnectionVEGACONNECT may be operated via I²C interface on the following sensors with the respective certificate:

CertifiedsensorsforgasExCertificate VEGA sensor

PTB 03 ATEX 2035 X VEGABAR BR50/60.CX******

PTB 03 ATEX 2069 X VEGABAR BR50/60.DX******

PTB 99 ATEX 2176 VEGASON SON 5*KEX.*C/D****

PTB 00 ATEX 2142 X VEGASON SON 5*KEX0.*C/D****

PTB 99 ATEX 2181 X VEGASON SON 5*PEX0.*G****

PTB 03 ATEX 2213 X VEGASON SN6*.C*/C_**H***

PTB 03 ATEX 2214 X VEGASON SN6*.C*/C_**P/F***

PTB 99 ATEX 2088 X VEGAFLEX FLEX 5*KEX0.*D******

PTB 99 ATEX 2098 X VEGAFLEX FLEX 5*KEX0.*B/D******

Page 11: Safety instructions VEGACONNECT CONNECT

11VEGACONNECT CONNECT.CXA4

3300

2-EN

-180

307

Certificate VEGA sensor

PTB 99 ATEX 2183 X VEGAFLEX FLEX 5*PEX0.*G******

PTB 01 ATEX 2155 X VEGAFLEX FLEX 5*K.C*D******

PTB 01 ATEX 2183 X VEGAFLEX FLEX 5*P.C*G******

PTB 01 ATEX 2196 X VEGAFLEX FLEX 5*K.D*D***

PTB 02 ATEX 2142 X VEGAFLEX FX6*.CX/C_***H***

PTB 03 ATEX 2057 X VEGAFLEX FX6*.CX/C_***P/F***

PTB 02 ATEX 2169 X VEGAFLEX FX6*.DX/D_***H/V***

PTB 99 ATEX 2055 X VEGAPULS PULS5*KEX0. *D*****

PTB 99 ATEX 2016 VEGAPULS PULS5*KEX. *D*****

PTB 99 ATEX 2066 X VEGAPULS PULS5*KEX0. *D/B*****

PTB 02 ATEX 2034 X VEGAPULS PULS5*KEX0.EB*****

PTB 99 ATEX 2022 X VEGAPULS PULS5*KEX. *D*****

PTB 99 ATEX 2166 X VEGAPULS PULS5*PEX0. *G*****

PTB 99 ATEX 2167 VEGAPULS PULS5*PEX. *G*****

PTB 00 ATEX 2045 VEGAPULS PS42. A*D******

PTB 00 ATEX 2044X VEGAPULS PS4*. C*D******

PTB 00 ATEX 2151 X VEGAPULS PS45. C*D******

PTB 01 ATEX 2159 X VEGAPULS PS4*. C*D******

PTB 00 ATEX 2095 X VEGAPULS PS4*. A*E******

PTB 00 ATEX 2094 X VEGAPULS PS4*. C*E******

PTB 00 ATEX 2148 X VEGAPULS PS45. C*E******

PTB 00 ATEX 2106 VEGAPULS PS4*. A*G***

PTB 00 ATEX 2105 X VEGAPULS PS4*. C*G******

PTB 00 ATEX 2150 X VEGAPULS PS45. C*G****

PTB 02 ATEX 2002 X VEGAPULS PS4*. C*G******

PTB 00 ATEX 2055X VEGAPULS PS4*. B*D******

PTB 00 ATEX 2054X VEGAPULS PS4*. D*B, D******

PTB 01 ATEX 2199X VEGAPULS PS4*. D*B, D******

PTB 00 ATEX 2149 X VEGAPULS PS45. D*B, D******

PTB 03 ATEX 2060 X VEGAPULS PS6*.CX***H***

PTB 04 ATEX 2040 X VEGAPULS PS68C_***H***

PTB 03 ATEX 2089 X VEGAPULS PS6*.CX/C***K/L/P/F***

PTB 04 ATEX 2042 X VEGAPULS PS68C_***P/F***

PTB 03 ATEX 2163 X VEGAPULS PS6*.DX/D_***D/E/H/V***

PTB 04 ATEX 2041 X VEGAPULS PS68D_***H/V***

PTB 06 ATEX 2022 X VEGAPULS PS6*.E_D/E/H/V***

TÜV 05 ATEX 2808 X VEGACAL CL6*.C_***P/F***

Page 12: Safety instructions VEGACONNECT CONNECT

12 VEGACONNECT CONNECT.CXA4

33002-EN-180307

Certificate VEGA sensor

TÜV 05 ATEX 2767 X VEGACAL CL6*.C_***H***

TÜV 05 ATEX 2799 X VEGACAL CL6*.D_***H***

TÜV 05 ATEX 2827 X VEGACAL CL6*.D_***P/F***

TÜV 03 ATEX 2040 X VEGACAP E32(H)EX

TÜV 03 ATEX 2041 X VEGACAP E34PAEX

KEMA 06 ATEX 0232 X VEGASON SONS6*.E*E*

CertifiedsensorsfordustEx

DMT 02 ATEX E 233 VEGAFLEX FX6*.**** CK PTB 02 ATEX 2142XPTB 03 ATEX 2057X

BVS 04 ATEX E 080X VEGAPULS PS6*.**** CK PTB 03 ATEX 2060XPTB 03 ATEX 2089X

DK PTB 03 ATEX 2060XPTB 03 ATEX 2163X

BVS 03 ATEX E 361 VEGABAR BR5/6*.**** CK PTB 03 ATEX 2053X

BVS 05 ATEX E 159 VEGACAL CL6*.**** CK TÜV 05 ATEX 2767XTÜV 05 ATEX 2808X

DK TÜV 05 ATEX 2799XTÜV 05 ATEX 2827X

Page 13: Safety instructions VEGACONNECT CONNECT

13

Notes

VEGACONNECT CONNECT.CXA4

3300

2-EN

-180

307

Page 14: Safety instructions VEGACONNECT CONNECT

14

Notes

VEGACONNECT CONNECT.CXA4

33002-EN-180307

Page 15: Safety instructions VEGACONNECT CONNECT

15

Notes

VEGACONNECT CONNECT.CXA4

3300

2-EN

-180

307

Page 16: Safety instructions VEGACONNECT CONNECT

Printing date:

VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermany

3300

2-E

N-1

8030

7

All statements concerning scope of delivery, application, practical use and operat-ing conditions of the sensors and processing systems correspond to the information available at the time of printing.Subject to change without prior notice

© VEGA Grieshaber KG, Schiltach/Germany 2018

Phone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected]