shots, shot and more shots on omega allows diagnostics and

23
Shots, shot and more shots on OMEGA allows diagnostics and targets development, opening new areas of science Presentation to OMEGA Laser User Group April 27, 2011 Joe Kilkenny NIF Diagnostic Leader & General Atomics V. P.

Upload: others

Post on 16-Apr-2022

10 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Shots, shot and more shots on OMEGA allows diagnostics and

Shots, shot and more shots on OMEGA allows diagnostics and targets development, opening new

areas of science

Presentation to

OMEGA Laser User Group April 27, 2011

Joe Kilkenny

NIF Diagnostic Leader &

General Atomics V. P.

Page 2: Shots, shot and more shots on OMEGA allows diagnostics and

This is an exciting time to be doing High Energy Density Physics: US has 3 forefront facilities

As well an important role for OMEGA is to develop techniques / diagnostics for NIF

Page 3: Shots, shot and more shots on OMEGA allows diagnostics and

Drivers

Advances in High Energy Density Physics Require:

Given a driver, innovation in targets and diagnostics drive new science

Theory

Targets

Diagnostics

Page 4: Shots, shot and more shots on OMEGA allows diagnostics and

Laser

Most of the > 40 NIF diagnostics were developed on Nova, OMEGA or NTS

4

FABS31

on NIF

Energetics

Energetics diagnostics. Full Aperture Backscatter

Gated

X-ray

Detector

Capsule

Capsule diagnostics,

developed on OMEGA,

Ignition

UGT experience, development

on OMEGA with neutrons

NIF, OMEGA and Z are generational machines

-diagnostics will continuously evolve

Page 5: Shots, shot and more shots on OMEGA allows diagnostics and

NIF’s ignition diagnostics measuring: Yield, Areal density ( R), Ion Temperature (Tion), Bang time, burn duration and Shape developed on OMEGA

Diagnostic Primary purpose Acronym

Zirconium Nuclear Activation Yield Well NADS

Copper Nuclear Activation Yield NAD 20

Magnetic Recoil Spectrometer Yield, R, Tion MRS

Gamma Reaction History N bang time, burn duration

GRH

Neutron Time of Flight

20m – Equator Tion , R NTOF_20E

20m - Alcove Tion , R NTOF_20A

Neutron Bangtime N bang time NTOF_BT

DT – high yield Yield > 5e13 NTOF4_DTHI

DT – Low yield Yield < 5e13 NTOF4_DTLO

Hardened Gated X-ray Imager (hGXI)

X-ray Shape, X-ray bang time

hGXI

Neutron Imaging (NIS) Neutron shape NIS

Why measure attributes different ways?

Page 6: Shots, shot and more shots on OMEGA allows diagnostics and

MacKinnon—NIC Technical Review, February 23-25, 2011 6 NIF-02211-21072s2.ppt

Complementary redundant diagnostics are essential for the scientific method

• Karl Popper, Hutchinson Press, 1959: cannot prove a theory (measurement) is right, can prove it is wrong - by test ( comparison with other diagnostics

NIF-1209-17998.ppt Kilkenny, NIC Review, December 9–11, 2009 6

• Down Scatter Fraction (DSF): measure by NToF and Magnetic Recoil Spectrometer • R: measure with NToF, MRS, Radchem and by ARC • Hot spot T measured by x-rays and neutron emission

Falsifiability and Testability : Hypotheses are nets: only he who casts will catch

Page 7: Shots, shot and more shots on OMEGA allows diagnostics and

To make a contribution to NIF diagnostics you don’t have to work at a large national Laboratory

MacKinnon—NIC Technical Review, February 23-25, 2011 7 NIF-02211-21072s2.ppt

LLNL – FABS,NBI – Dante I & II – VISAR – DISC – RAGS – hGXI – ARIANE – DIXI – NAD

AWE – FFLEX

LANL – GXD – burn – Rad. Chem. II – n imaging

SNL – SPIDER – DISC – NAD (Cu) – NToF – RAGS

Universities develop valuable niches: MIT- CR39, U of M targets, Imperial College- high speed electronics.

NSTec – Calibration – NToF

LBNL – Calibration – Rad. Chem.

CEA – N imaging

Duke – Calib.

MIT – MRS – WRF – PTOF

LLE – NToF – 4 fidu system – MRS, NADS – SPBT – HEXRS – PSBT

U of M. – HEMPI

Page 8: Shots, shot and more shots on OMEGA allows diagnostics and

Pat McKenty’s polar direct drive exploding pushers verify the performance of the ignition diagnostics

MacKinnon—NIC Technical Review, February 23-25, 2011 8

NIF-02211-21072s2.ppt

Yn: Magnetic Recoil Spectrometer (MRS)

Yn: Neutron Activation (NAD, Zr)

Zr

Yn: Neutron Activation (NAD, Cu)

Cu

Exploding pushers:Direct drive – 125kJIsotropic Yield ~ 2e14Low R < 20mg/cm2

Polar direct drive platform designed, developed at LLE

1.5mm diam. Capsule w/ 10 atm 50:50 DT fill

Low R provides ideal, low background for yield and downscatter diagnostics

Page 9: Shots, shot and more shots on OMEGA allows diagnostics and

Yield: Zr Cu and Zr activation measures yield for DT shots to absolute accuracy of ± 7% : BUT how do we know it is right?

MacKinnon—NIC Technical Review, February 23-25, 2011 9 NIF-02211-21072s2.ppt

Exp.Pusher: Raw Data, 8.7mm sample

Co

un

ts

24 hours counting

89Zr

1.0 3.5

8.7

3 samples with

different thickness

Placed in a well 4.5m from TCC

Neutron energy (MeV)

0

0.2

0.4

0.6

0.8

1

1.2

0 5 10 15 20 25 30

In-115(n,n')In-115m

Nb-93(n,2n)Nb-92m

Zr-90(n,2n)Zr-89C-12(n,2n)C-11

Cu-63(n,2n)Cu-62

90Zr(n,2n) 89Zr (t1/2=3.3 d)

63Cu(n,2n) 62Cu (t1/2=9.74 m)

Cro

ss s

ecti

on

(b

)

Activation Foil cross sections 14 MeV n

10 MeV n

Y = 2.3 x 1014 ± 7 %

Shot N

101030

Page 10: Shots, shot and more shots on OMEGA allows diagnostics and

Yield: Copper activation has an absolute accuracy of ±10%- maybe

MacKinnon—NIC Technical Review, February 23-25, 2011 10 NIF-02211-21072s2.ppt

Y = 2.5 x 1014 ± 10 %

Copper Decay curve N101030 – Exp pusher

0

100

200

300

400

500

600

0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0

Co

un

ts/M

inu

teTime (minutes)

9.5 mm Cu at 29 m

1.0 mm Cu at 19 m

Copper sampling yield on equator and in neutron alcove

Uncertainty Estimates Error±1

Detector efficiency & Gamma-ray self-shielding

8.6 %

Cu cross section uncertainty 1.3 %Position uncertainty (awaiting measurements) 0.1 %

Scatter/absorption from Al and W attenuators 5.0 %

TOTAL SYSTEMATIC 10.0 %

Statistical 2.5 %

1mm Cu

1mm Cu

9.5mm Cu

Page 11: Shots, shot and more shots on OMEGA allows diagnostics and

MacKinnon—NIC Technical Review, February 23-25, 2011 11 NIF-02211-21072s2.ppt

The NIF magnetic recoil spectrometer (MRS) was developed on OMEGA by MIT and LLE

Kilkenny, NIC Review, December 9–11, 2009 11NIF-1209-17998.ppt

• MRS has measured rr on OMEGA

• Action items were focused on transfer

of instrument technology to NIF

Page 12: Shots, shot and more shots on OMEGA allows diagnostics and

Yield: The MRS has been designed and implemented for simultaneous measurements of R, Y1n and Tion

MacKinnon—NIC Technical Review, February 23-25, 2011 12 NIF-02211-21072s2.ppt

Low-Res tf = 275 μm

Med-Res tf = 125 μm

CR-39 “ironing board”

Rf = 26 cm

Ra = 570 cm

Aa = 20 cm2

Accuracy requirement for the MRS absolute yield measurement < 10% for Y1n > 1014

n

d+

Is the CR-39 “ironing board” radioactive?

Page 13: Shots, shot and more shots on OMEGA allows diagnostics and

MacKinnon—NIC Technical Review, February 23-25, 2011 13 NIF-02211-21072s2.ppt

NIF WRF proton spectrometer used in 2009, improvements for 2010

CR39 used in Wedge Range Filter ( WRF) after major OMEGA/MIT development on MIT accelerator

NIF WRF proton spectrometer used in 2009, improvements for 2010

East /West summit on Good Friday 2011

Page 14: Shots, shot and more shots on OMEGA allows diagnostics and

Yield: MRS provides absolute primary yield to ± 5% accuracy: nd cross section and foil area uncertainty are largest components

MacKinnon—NIC Technical Review, February 23-25, 2011 14 NIF-02211-21072s2.ppt

MRS has different systematic errors from nuclear activation ( NAD)

MRS Spectrum for Pusher shot

Yield = 2.2e14 ± 5%

μ

Yield Error Budget

N101030

zero downscatter

Page 15: Shots, shot and more shots on OMEGA allows diagnostics and

Three independent Yn diagnostics ( Zr, Cu & MRS) gives 4% error in the weighted mean yield

MacKinnon—NIC Technical Review, February 23-25, 2011 15 NIF-02211-21072s2.ppt

So what?: surprisingly (at first) we found the OMEGA nTOF calibration did not transfer to NIF

MRS

Well NADS Zr

NADS 19m Cu

Neutron Yield

2.1 2.2 2.3 2.4 2.5 2.6 2.7 Yn x1014

N101030-0

02-9

99

Weighted mean yield = 2.3e14 ± 8.9e12 ( ± 3.9%) 2/ (n-1) = 0.9

Weighted

mean

Page 16: Shots, shot and more shots on OMEGA allows diagnostics and

R: Neutron downscatter measured from three different directions to minimize sampling errors due to directionality of scattered neutrons

MacKinnon—NIC Technical Review, February 23-25, 2011 16 NIF-02211-21072s2.ppt

20 m

dN

/dt

TOF [ns]

23ns

“Wilson

2E14”

Downscatter ratio (DSR*) technique

MRS - measure tracks in DSR region

MRS+ 2x NTOF20’s on 3 lines of sight

• LLE cal. • SNL, NsTEC cal.

• LLNL installation

NTOF – scattered neutrons in tail

Primaries 12-10MeV

DSN THD

* DSR = (10-12MeV) / (13-15MeV)

Page 17: Shots, shot and more shots on OMEGA allows diagnostics and

0.00

0.02

0.04

0.06

0.08

0.10

0.12

410 420 430 440 450 460 470 480 490 500 510 520

neu

tro

ns/(

Yie

ld n

s)

Time (ns)

R: Down Scatter Ratio (DSR) by MRS and NToF20 are self consistent within the measurement error

MacKinnon—NIC Technical Review, February 23-25, 2011 17 NIF-02211-21072s2.ppt

• Background level from exploding pushers is reproducible • Shorter cables planned for NTOF20 will reduce uncertainty

Lower cable induced background levels by ~3x

NToF Signal clearly shows down scatter signal on all THD shots

MRS: DSR error < 10% for Yn >1e14 and .R ~ 1g/cm2

THD-3: DSR = 0.027 ± 0.005 THD-3: DSR = 0.024 ± 0.003

(N110201)

N110201

Exp pushers

50x signal

13 - 15 MeV 10 - 12 MeV

THD

Neutr

on/s

Area normalized

Page 18: Shots, shot and more shots on OMEGA allows diagnostics and

Bangtime: Gamma Reaction History (GRH) routinely measuring Bang Time (w/in 30 ps) and Burn Width (w/in 15 ps)

MacKinnon—NIC Technical Review, February 23-25, 2011 18 NIF-02211-21072s2.ppt

GRH-6m at (64,20) DT exploding pusher Data

No CO2

Exploding Pushr (N101212)

Deco

nvo

lved

Gau

ssia

n

Tim

ing

Sh

ot

Fit

(N

101029)

Bang time +/- 30 ps

Burn width +/- 15 ps

DT

0

Yie

ld R

ate

n/n

s)

0

5 1014

1 2 3 -1

Time (ns)

Bangtime agrees with NToFBT data to within 100 ps

3-10MeV GRH will be fielded on DT implosions for yield and 4.4MeV carbon (ablator density measurement)

• 4 Gas Cherenkov cells • Detectors installed with Ultra-fast MCP PMTs and optical Mach Zehnders

Page 19: Shots, shot and more shots on OMEGA allows diagnostics and

A new Particle Time of flight (pTOF) detector has been implemented for shock-bang and compression-bang time using D3He-protons and DD-neutrons, respectively

MacKinnon—NIC Technical Review, February 23-25, 2011 19 NIF-02211-21072s2.ppt

DIM (0,0) WRFs DIM (90,78) WRF + CVD

• Relative timing of shock and compression feature is constraint to implosion models

• Project led by MIT• Requires Exploding pusher to qualify

MIT

Particle bang time data DHe3 SYMCAP 1208

Proton Spectrum from WRF

Energy (MeV)

Pro

ton

s/ M

eV

• Compression BT = 21.2 ± 0.15 ns

Page 20: Shots, shot and more shots on OMEGA allows diagnostics and

NIF’s $20M Neutron Imaging System has begun performance qualification using McKenty’s exploding pushers

MacKinnon—NIC Technical Review, February 23-25, 2011 20 NIF-02211-21072s2.ppt

Signal: 7500 counts

pusher

Reconstructed Neutron Image Raw Image

NIS Pinhole array

Exploding pusher DATA

150 μm

Yield = 2e14

Compare to gated x-ray

Image

150 μm

Page 21: Shots, shot and more shots on OMEGA allows diagnostics and

MacKinnon—NIC Technical Review, February 23-25, 2011 21 NIF-02211-21072s2.ppt 4/27/11 MR 01-12-11 21

P. Volegov – NIS analysisR. Tommasini – hGXI analysis

Consistent with neutron hot spot inside x-ray emitting region ?

Remnant of Si pusher emits strongly in 6keV x-ray region

NIF s owl is skeptical without image registration

Page 22: Shots, shot and more shots on OMEGA allows diagnostics and

The best of times in HED!

22

• Innovative diagnostics allow advances in HED • New diagnostics require a significant number of development shots

• OMEGA users are playing a major role in development of NIF diagnostics

But what next ? Decadal innovation in diagnostics by new minds will allow us to fully exploit OMEGA, Z and NIF.

Think clever, think big!

Page 23: Shots, shot and more shots on OMEGA allows diagnostics and