SINAPI CHEST DRAINBREAKTHROUGH : 1 WAY VALVE
PneumothoraxHemothoraxGravity Drainage
No ClampingFaster EvacuationPromotes Early Patient MobilitySuperior Infection Control
Advantages
For the evacuation of fluids and air from the mediastinaland pleural cavities in post-operative or trauma settings
HOW TO USE ( at the back)Exclusively Distributed by:
t o m o r r o w ’ s m e d i c a l d e v i c e sWellnessPROTel: (02) 636-3580 • Fax: (02) 636-7542
Mobile: (63) 920-947-7185, (63) 922-823-2282Unit 1006 Richmonde Plaza, San Miguel Avenue cornerLourdes St., Ortigas Center Pasig 1605, Philippines
Email: [email protected]@wellnessproinc.com
www.wellnessproinc.com
DESCRIPTION
The SINAPI Chest Drain is a drain chest drainage unit incorporating a dry seal, blood collection reservoir air leak detector, suction bulb and drainage tap.
INDICATIONS
The SINAPI Chest Drain is used for:
post operative and trauma situations.
cavities.3. Facilitation of lung re-expansion and restoration of normal breathing dynamics.
SET UP
HOW TO CONFIRM CHEST CATHETER POSITION DIRECTLY AFTER INSERTION
but initially re-expand.
This will mean that the catheter in not placed in the pleural space (or pneumothorax is resolved).
tubing.
MOBILE PATIENTS
Shorten the chest catheter ( under aseptic technique ) or position on the
loops) . For Pneumothorax, remove the drainage bag. Fig (1)
BED RIDDEN PATIENTS (Gravity Drainage)
Allow gravity drainage by hanging the device next to bed, lower than the chest.
moving/turning the patients you may place the device on its back on the bed – as long as the liquid in the reservoir remains less than the maximum volume indications. Hang next to the bed when unattended. Fig (2)
MANUALLY INCREASING THE DRAINAGE RATE
Increase the drainage rate by depressing the bulb unit until the bulb stays depress Fig (3) If the bulb re-expands, pinch the tubing above the valve and then depress the bulb. Release both. Repeat the process unit until the bulb stays depresses (up to 30x). This applies suction and increase the
-rax will remain and the bulb will not stay depress- attach external suction if required by physician.
ATTACHING LOW PRESSURE SUCTION
If the bulb does not stay depress attach suction as indicated. Keep the drainage tap closed when suction is applied. The recommended suction pressure is -20cmH20 pressure.
DETECTING BUBBLING/AIR LEAK
Depress the bulb. If it stays depress there is no air leak. If it re-expands, add 25ml clean water into the air vent tubing. Look for bubbling or water level movement. Bubbling or water level movement =air leak. Fig (5)
MONITORING
1. Drainage must be low.
must stay depress and the valve angled ( indicating negative intra-pleural/mediastinal pressure)
X-ray.
MAINTENANCE
- Liquid might accumulate in the tube above the valve. This is normal and an indication of negative pressure above the valve.- During outpatient management, secure the chest catheter with additional plaster closer to the reservoir to prevent dislodgment of the catheter.- Replace the device after 7 days.
DRAINAGE INTO BAG
Securely attach the drainage bag into the
into the bag (clamp LP suction tubing to speed up drainage) Close the tap and close the slide clamp of the bag. Remove the
1
2
3
7
10
9
8
4
5
6
Exclusively Distributed by:
t o m o r r o w ’ s m e d i c a l d e v i c e sWellnessPROTel: (02) 636-3580 • Fax: (02) 636-7542
Mobile: (63) 920-947-7185, (63) 922-823-2282Unit 1006 Richmonde Plaza, San Miguel Avenue cornerLourdes St., Ortigas Center Pasig 1605, Philippines
Email: [email protected]@wellnessproinc.com
www.wellnessproinc.com
ARC-Infruitec North CampusLelie roadIdas Valley
Stellenbosch, 7600South Africa