single- and two-phase flow through a globe valve ... · pdf file0 single- and two-phase flow...
TRANSCRIPT
0
Single- and Two-phase Flow through a Globe Valve: experiments and numerical simulations
M.H.M.Lemmens Master Internship Project, reportnumber: WPC 2007.03 Roma, 1 september – 18 december 2006 Supervisors: Università di Roma “La Sapienza”: Prof. C. Alimonti Eindhoven University of Technology: Dr. C.W.M. van der Geld Numerical Simulations: Prof. A. Corsini, Bruno Perugini
1
Table of Contents
Table of Contents .....................................................................................................................................1 Nomenclature ...........................................................................................................................................3 1. Introduction ..........................................................................................................................................5 2. Valve and set-up design and practicalities............................................................................................6
2.1 Valve design ...................................................................................................................................6 2.2 Set-up design ..................................................................................................................................7 2.3 Use of the set-up.............................................................................................................................8
2.3.1 Start set-up...............................................................................................................................8 2.3.2 Measurements..........................................................................................................................8
3. Experiment strategy..............................................................................................................................9 3.1 Operation points .............................................................................................................................9 3.2 Data processing ..............................................................................................................................9
4. Single-phase Experiment Results ....................................................................................................... 10 4.1 Pressure in the set-up.................................................................................................................... 10 4.2. Pressure drop over the valve........................................................................................................ 11
4.2.1 DpValve................................................................................................................................. 11 4.2.2 DpASK92 .............................................................................................................................. 11
4.2 Pitot pressure drop........................................................................................................................ 12 4.3 Fluid velocity in the globe valve .................................................................................................. 13
5. Two-phase flow .................................................................................................................................. 14 5.1 Introduction .................................................................................................................................. 14 5.2 Theory .......................................................................................................................................... 14
5.2.1 Pressure drop ......................................................................................................................... 14 5.2.2 Choking ................................................................................................................................. 15
5.3 Experiment Strategy ......................................................................................................................... 16 5.3.1 Operation points ........................................................................................................................ 16 5.3.2 Tylan calibration........................................................................................................................ 17
6. Two-phase Experiment Results .......................................................................................................... 18 6.1 Pressure drop over the valve......................................................................................................... 18
6.2 DpValve.................................................................................................................................... 18 6.3 DpASK92 ................................................................................................................................. 19
6.4 Pitot pressure drop........................................................................................................................ 19 6.5 Fluid velocity in the globe valve .................................................................................................. 21
7. Two-phase Multiplicator .................................................................................................................... 23 7.1 Theory .......................................................................................................................................... 23 7.2 Experiments .................................................................................................................................. 23
8. Numerical analysis ............................................................................................................................. 25 8.1 XENIOS ....................................................................................................................................... 25 8.2 Grid generation............................................................................................................................. 26 8.3 Boundary conditions..................................................................................................................... 27
8.3.1 Velocity ................................................................................................................................. 27 8.3.2 Turbulent kinetic energy, viscous dissipation rate................................................................. 28
8.4 Pressure and Velocity normalization ............................................................................................ 29 8.3 Flow rate and Valve closings........................................................................................................ 29 8.4 Results .......................................................................................................................................... 29
8.4.1 Valve Pressure drop and Pitot pressure drop ......................................................................... 30 8.4.2 Flow field............................................................................................................................... 30 8.4.3 Pressure contour in valve....................................................................................................... 31 8.4.4 Velocity contour in valve ...................................................................................................... 33
9. Conclusions ........................................................................................................................................ 35 10. Recommendations ............................................................................................................................ 36 11. Bibliography..................................................................................................................................... 37 12. Appendix .......................................................................................................................................... 38
12.1. Single-phase old measurements................................................................................................. 38 12.1.1. Pa ........................................................................................................................................ 38 12.1.2. DpValve.............................................................................................................................. 38 12.1.3. DpPitot................................................................................................................................ 39
2
12.2 Two-phase old measurements..................................................................................................... 40 12.2.1 DpValve............................................................................................................................... 40 12.2.2 Pitot Pressure ....................................................................................................................... 40 12.2.3 Two-phase multiplicator...................................................................................................... 40
3
Nomenclature
Letter Quantity Unit (SI)
Chapter 4
p pressure N/m2
v velocity m/s
Chapter 5
A flow area m2
cfric friction factor -
Dt tube diameter m
D0 momentum flux density -
G mass flux kg/ (m2s)
GVF homogeneous void fraction -
i index -
Mtp two-phase Mach-number -
Q flow rate m3/s
u fluid velocity m/s
x mass fraction -
Chapter 7
i index -
j index -
Chapter 8
C XENIOS numerical simulation
constant
-
cµ model constant -
Cε1 model constant -
Cε2 model constant -
fε1 model constant -
fε2 model constant -
k turbulent kinetic energy m2/s
2
l*
turbulent length scale m
Lref reference length m
m velocity profile parameter -
r radial coordinate m
R radius inlet tube m
Re Reynolds number -
Sij mean strain rate tensor s-1
u*
turbulent velocity scale m/s
Symbol Quantity Unit (SI)
Chapter 5
α void fraction -
ΦL liquid multiplication factor -
ρ density kg/m3
µ dynamic viscosity kg/(ms)
4
Chapter 7
∆p pressure drop N/m2
ΦLo liquid only multiplication factor -
Chapter 8
ε viscous dissipation rate m2/s3
ε~ modified viscous dissipation rate m2/s
3
σk effective Prandtl number for diffusion m
σε effective Prandtl number for dissipation m
νt eddy viscosity m2/s
θ tangential coordinate -
τ turbulent time scale s
Subscript Quantity
Chapter 4
dyn dynamic
pitot pitot tube
stat static
tot total
Chapter 5
a air
acc acceleration
g gas
grav gravitation
fric friction
m mixture
w water
0 reference position 0
Chapter 6
mix mixture of water and air
tot Total of water flow rate and air flow rate
Chapter 7
exp experiment
s single-phase
t two-phase
Chapter 8
max maximum
mean mean
XENIOS as measured in XENIOS
5
1. Introduction
Two-phase flows have a wide range of applications in both science and engineering. The problem that
is often faced when studying two-phase flow is the lack of knowledge of its behavior. Especially in
process technical equipment, in contrary to single-phase flow, predicting pressure drop in two-phase
flow is very complicated. Most relations that exist for predicting two-phase flow pressure drop are
empirical. To get familiar with the behavior of two-phase flows in process technical equipment,
experiments are thus necessary. The topic of this report is the behavior of a two-phase flow in a Globe
Valve designed by the Dutch company ‘Mokveld’. The final goal is the development of a model of the
valve for two-phase flow. This valve is e.g. used in the oil- and gas industry for pressure and flow
control. Two-phase flow, during this study, consists out of water and air. The two-phase flow through
the valve will be studied under several flow conditions. Variable conditions are the flow rate, the
closing of the valve and the void fraction. The experimental data enables to develop a model for two-
phase flow through such a globe valve.
In accordance with the before said, experiments have to be done to get insight in the behavior of the
two-phase flow through the globe valve. The experiments are carried out in the laboratory of the
engineering faculty of ‘Università "La Sapienza" di Roma’. An experimental set-up is build around the
globe valve to be able to do measurements on the valve. Before a two-phase flow is applied to the test
set-up, experiments are done with only one phase: water. This is done to test the sensors and to get
insight in the behavior of the test set-up as a whole. Practicalities of these experiments are discussed in
chapter 2 and 3; the results are presented in chapter 4. For different flow conditions, different
measurements are done. Experiments with almost the same set-up are already done before. Only a
small difference in the set-up is made. Since the change made in the set-up is very small, the
differences between the measurements are expected to be negligible. The old measurements are thus a
good reference for the new ones.
After the single-phase flow experiments are done, experiments with the same water flow rates and
valve closings are done with two-phase flow. Air is added to the water flow. Theory and practicalities
involved with the two-phase flow experiments are discussed in chapter 5. Thereafter, in chapter 6, the
results of these two-phase flow experiments are presented and analyzed. Here extra attention will be
paid to the influence of the void fraction, since it is the most important parameter in two-phase flows.
The results from the experiments are also compared to those of the old set-up. Moreover single-phase
flow experiments and two-phase experimental results will be compared (chapter 7). The comparison of
the two-phase flow data with the single-phase flow data is a first start for the development of a model
for the valve.
The last topic treated in this report is a numerical model (chapter 8) of the globe valve created in
XENIOS. A finite element model is made of the valve. With this model the flow field in the valve can
be studied for different flow conditions. In this study only a model is made for the single-phase flow.
The flow field can give insight in the experimental results. The experimental data functions as a
reference for the data obtained by the numerical model. The experimental data together with the
numerical model is a starting point for the final development of a model for the valve in two-phase
flow.
First the design of the valve and the set-up will be discussed in chapter 2.
6
2. Valve and set-up design and practicalities
2.1 Valve design
The valve is of the RZD-RQX type, a Globe Valve made by Mokveld. ‘Globe’ refers to the spherical
shape of the valve. This valve is used for pressure and flow control for liquid, gas and two-phase flows.
Therefore also is referred to the valve as ‘Control Valve’. The typical design of the closing enables to
set the pressure drop and thus the pressure in the system. The inlet of the valve has a diameter of 3”
(76.2 mm). In figure 2.1, two cross-sections of the valve can be seen. The axial axis will be indicated as
the x-axis, the vertical direction as the y-axis. The direction perpendicular to the x-y plane is the z-axis.
The postion (0,0,0) is at the center of the valve , at the intersection point of the horizontal and vertical
dashed lines. The left picture in figure 2.1 shows a cross-section at the x-y plane through the point
(0,0,0). The right picture shows a cross-section at the y-z plane through the point (0,0,0).
Figure 2.1 Cross Section RZD-RQX Mokveld Valve
The flow enters the valve at the left and flows around the body in the center. After the flow has passed
the body it flows through multiple holes (indicated in figure 2.1 as ‘standard cage’). The flow exits the
the valve on the right. By turning the vertical shaft on top of the valve, the cylinder in the middle of the
body can be moved in horizontal direction. When moving to the right, it closes the holes of the cage.
When the cylinder is at the most right position, the flow is blocked. When the cylinder is at the most
left position all holes are opened. The number of closed holes determines the pressure drop over the
valve. The pressure drop over the valve can be set by its closing. Evidentially the fluid and the flow
rate also determine the pressure drop.
7
2.2 Set-up design
To be able to make measurements on the valve, a set-up is build around the valve. A scheme of the set-
up is given in figure 2.2. This figure gives only the connections between the different elements in the
set-up. The black lines represent the connections. The drawing is not on scale and the locations are not
like in reality. The flow directions are 1-2-3-4-6-11-10-9-8-5-7-1 and 10-9-11-10. In table 2.1 the
different elements of figure 2.2 are explained.
Figure 2.2 Lay-out of the set-up
Number Explanation
1 Diverging tube to bridge the flow from the 2” to the 3” part.
2 Straight tube to develop the flow after the diverging section.
3 RZD-RQX Mokveld Globe Valve.
4 Converging tube to bridge the flow from the 3” to the 2” part.
5 Pressure tap to measure the static pressure at the pressure side of the pump in the set-up.
6 Separator to separate the air from the two-phase flow.
7 Valve to control the flow rate through the part of the set-up with the Globe Valve.
8 Electronic flow sensor.
9 Valve to divide the flow between the part of the set-up with the Globe Valve and the start-
up cycle (10-9-11-10).
10 Rotary Pump.
11 Water tank. The pump takes the water from the bottom of the tank. At the top of the tank
water flows in, coming from valve 9 and the separator.
12 The air supply which delivers the gas for the two-phase flow.
13 Pressure tap at the inlet of the Globe Valve.
14 Pressure tap at the outlet of the Globe Valve.
15 Pitot pressure tap, to determine dynamic pressure.
Table 2.1 Explanation figure 2.2
In the sequel will be referred to the elements in figure 2.2 by their numbers.
8
2.3 Use of the set-up
2.3.1 Start set-up
To be able to start the pump, a small start-up cycle (10-9-11-10) is made in the set-up. At first tank 11
has to be filled with water. When starting the pump, valve 7 is closed and valve 9 is totally opened. The
water sequentially flows from the pump through valve 9, the tank 11 and back to the pump.
2.3.2 Measurements
To let the water flow through the globe valve, after the start-up, valve 7 has to be opened entirely.
Sensors 5, 8, 13, 14 and 15 start to give signals. The signals are send to the computer and can be read
with the program Labview. With Labview the experimental data can be saved.
The flow rate through the globe valve can be increased by, partially, closing valve 9. By closing this
valve, the flow through the start-up loop decreases and thus the flow through the part with the globe
valve increases. The flow through the valve is measured with flow meter 8. The flow is increased from
100 l/min to 250 l/min. Increasing the flow-rate is done in different experiments with different closings
of the globe valve (3). This will be explained extensively in the sequel of the report. The closing of the
valve can be set with the shaft on top (chapter 2.1).
The air supply 12 is controlled by a pump that is connected to the computer. With Labview the air flow
which flows into the set-up can be set. Herewith the homogeneous void fraction is set.
9
3. Experiment strategy
3.1 Operation points
The set-up has different sensors with which process parameters can be measured. Sensors 1 and 2 in
figure 2.1 respectively measure the inlet – and outlet pressure of the valve. The sensor at point 3, point
of local zero velocity, measures the total pressure of the flow. Sensor 4 at last measures the static
pressure. The experiments are done for a range of different flow rates and closings of the Globe Valve.
Each combination of flow rate and valve closing will in the sequel be denoted as ‘operation point’. The
flow rate in the set-up ranges during the experiments from 100 l/min to 250 l/min with steps of 25
l/min. The closing of the globe valve is varied from 0 mm to 25 mm with measurements at 0, 10, 15,
20, 22.5 and 25 mm. For safety reasons the measurements are only carried out if the pressure
(measured by sensor 5 in figure 2.2) in the loop doesn’t exceed 1.3 bars and the water temperature is
less than 30°C. The operation points are the same for both the single- and two-phase flow
measurements. If different it will be denoted.
The different outputs of the sensors which will be presented in the sequel of this report are: the pressure
drop over the globe valve; pitot pressure drop in the globe valve and a single point pressure
measurement (sensor 5 in figure 2.2) in the loop.
3.2 Data processing
At each different operation point, more measurements are done. Every single measurement in the
graphs in the sequel is represented by a red ‘+’. From all the measurements on one operation point,
mean values are calculated, these are represented by a dot. In some graphs, ‘+’ and dots are left out for
a better visibility of the lines. Through the mean values, polynomial functions are fitted with use of a
MATLAB routine. The polynomials are in general 2nd order; if different it will be denoted. The
polynomials are represented in the graphs by the lines. In the legends can be seen which valve closing
is represented by a certain polynomial.
10
4. Single-phase Experiment Results
4.1 Pressure in the set-up
To check the pressure in the loop a single point pressure tap is placed. The pressure tap is placed on the
pressure side of the pump, before the valve. The actual pressure measured with this pressure tap is
represented by Pa. In the graph below, the measurements on Pa are presented.
0 50 100 150 200 2500
0.5
1
1.5
2
2.5
025mm
Qw [l/min]
Pa [
bar]
Pa single phase
open
10mm
15mm
20mm
22.5mm
Figure 4.1 Pa for single-phase flow
In figure 4.1, on the horizontal axis the water flow rate, in l/min, can be seen and on the vertical axis
the pressure Pa in bars. Like stated in chapter two, measurements are only allowed if Pa < 1.3 bars. As
can be seen in figure 4.1, most measurement points satisfy the restriction of Pa < 1.3 bars. The
measurement points that don’t satisfy the restriction exceed 1.3 bars only slightly. This won’t harm the
set-up. The operation points specified in chapter 3.1 can thus safely be applied to the set-up. The only
exception is a valve closing of 25mm. For this closing only measurements are done with Qw = 100
l/min, because for this lowest flow rate the Pa is already much too high. For increasing flow rate Pa
would only increase more.
In figure 4.1 can be seen, as expected, that Pa increases with increasing closing of the valve and
increasing flow rate. Namely, the resistance for the flow increases with increasing fluid velocity and a
smaller orifice through which it flows. Thus when the flow rate is increased and the valve is closed, Pa
also has to increase to maintain the flow rate at the higher value. Pa can be understood as the driving
pressure for the flow through the whole set-up.
Pa is relative to the atmospheric pressure, it can have values smaller than 1 bar. This measurement
shows good accordance with the measurement done on the old set-up (Appendix 12.1.1).
11
4.2. Pressure drop over the valve
The pressure drop over the valve is measured with two different methods. One method measures the
inlet and outlet pressure of the valve and subtracts the outlet from the inlet pressure. The other method
directly measures the difference between the inlet and outlet pressure of the valve. The sensors are
placed at position 1 and 2 in figure 2.1. In the figure of the set-up figure 2.2 the sensors can be seen at
position 13 and 14. In the sequel these different methods will be denoted by ‘DpValve’ and
‘DpASK92’ (the name of this output is equal to the name of the sensor) respectively.
4.2.1 DpValve The measurements on DpValve for a single-phase flow are represented in figure 4.2. On the horizontal
axis the water flow rate, in l/min, can be seen and on the vertical axis the DpValve pressure drop in
mbar.
0 50 100 150 200 2500
200
400
600
800
1000
1200
Qw [l/min]
DpV
alv
e [
mbar]
DpValve single phase
open
10mm
15mm
20mm
22.5mm
25mm
Figure 4.2 DpValve, single-phase flow
DpValve increases with increasing flow rate and increasing closing of the valve. This can be seen
clearly in figure 4.2. An increasing flow rate and increasing closing of the valve result in increasing
resistance to the flow. An increasing resistance results in a higher pressure drop over the valve. The
measurements for a closing of the valve of 25mm are done for smaller flow rates than the other
measurements. For measurements at flow rates higher than 100 l/min (valve closing 25mm), Pa would
immediately become larger than the maximum allowed value of 1.3 bars. The results in Figure 4.2 are
in good accordance to the measurements for DpValve at the old set-up (see Appendix 12.1.2).
4.2.2 DpASK92
The measurements on DpASK92 are shown in figure 4.3. On the horizontal axis the water flow rate, in
l/min, can be seen and on the vertical axis the pressure drop DpASK92 in mbar. The maximum
pressure difference which can be measured by DpASK92, with good accuracy, is about 200 mbar.
Therefore only measurements could be done on valve closings up to 15mm. For higher valve closings,
at the considered flow rate range, the pressure drop over the valve exceeds the scope of DpASK92. The
values found for DpASK92 are in accordance to the values found for DpValve in the scope of
DpASK92. DpValve measured for the new and the old set-up are in good accordance like shown
before. The measurements for DpASK92 thus show accordance to the measurements on the old set-up.
12
0 50 100 150 200 2500
20
40
60
80
100
120
140
160
180
200
Qa [l/min]
DpA
SK
92 [
mbar]
DpASK92 single phase
open
10mm
15mm
Figure 4.3 DpASK92, single-phase flow
4.2 Pitot pressure drop
Pitot pressure drop is measured with use of the body in the valve (see figure 2.1). In order to measure
this Pitot pressure drop, one pressure tap is placed at the head of the body (3, figure 2.1). The other
pressure tap is placed at the wall of the valve (4, figure 2.1). The pitot pressure is thus measured at the
wall. Fluid velocity which will be calculated from the pitot pressure is thus the velocity at the wall. The
static pressure, pstat, on the wall of the valve is subtracted from the total pressure, ptot, which is
measured at the head of the body (where the velocity is locally equal to zero). With this method, the
dynamic pressure, pdyn, can be measured and thus the velocity around the body be determined. In
formulas this can be represented by:
dynstattot ppp += (4.1)
2
2
1vdpppp pitotdynstattot ρ===− (4.2)
It is adopted that the stationary incompressible Bernoulli equation is valid. The density of the fluid is
represented by ρ and the velocity of the fluid by ν. The velocity of the fluid in the valve can now
directly be related to the measured pitot pressure drop (which is equal to the dynamic pressure):
ρpitotdp
v2
= (4.3)
This velocity is thus only valid at the position where the dynamic pressure is measured.
The measurements on DpPitot are represented in figure 4.4. On the horizontal axis the water flow rate,
in l/min, can be seen and on the vertical axis the pressure drop DpPitot in mbar. For valve closings of
22.5mm and 25mm, to less measurements are available to make polynomials. The tendency of the
graphs in figure 4.4 is like expected. Pitot pressure drop increases with increasing flow rate and
increasing closing of the valve. Namely the velocity of the fluid increases with increasing flow rate. In
figure 4.4 also can be seen that, when you leave out the measurements for zero closing, the pitot
pressure drop is almost independent of the valve closing. This is caused by the design of this typical
valve. The closing of the valve only results in a locally, in the holes in the cage, velocity change. At the
location where the pitot pressure drop of figure 4.4 is determined, the flow area is constant and so is the
pitot pressure drop only dependent of the flow rate. The same can later be seen in figure 4.5 for the
fluid velocity. When compared to the old measurements on DpPitot (see Appendix 12.1.3), there is a
large discrepancy. Both measurements show the same tendencies, but the values for DpPitot in the new
13
experiments have significant higher values than the old measured values. Until now this large
discrepancy can’t be declared. Only for a fully opened valve, measurements on DpPitot on the old set-
up are available.
0 50 100 150 200 250
0
5
10
15
20
25
Qw [l/min]
DpP
itot
[mbar]
DpPitot single phase
open
10mm
15mm
20mm
22.5mm
25mm
Figure 4.4 DpPitot, single-phase flow
4.3 Fluid velocity in the globe valve
With use of formula 4.3, the fluid velocity can easily be determined from the measured DpPitot. The
results are shown in figure 4.5. For the same reason as in figure 4.4 no polynomials are made for the
measurements on valve closings of 22.5mm and 25mm. The velocity shows linear dependency on the
flow rate. Since the fluid in these experiments is incompressible, the conservation of mass relation
indeed gives this linear dependency. The flow area at the place where the velocity is measured is
constant; it’s not affected by the closing of the valve. Therefore the lines for the different closings are
almost the same (same as seen in figure 4.4). The kink at a zero flow rate is because of numerical
reasons and of no importance for the discussion here. It’s trivial that the fluid velocity in the new
experiments has just as DpPitot much higher values than in the old experiments.
0 50 100 150 200 2500
0.5
1
1.5
2
2.5
Qw [l/min]
velo
city [
m/s
]
Fluid velocity around body in valve single phase
open
10mm
15mm
20mm
22.5mm
25mm
Figure 4.5 measured fluid velocity around body in valve
14
5. Two-phase flow
5.1 Introduction
The main topic of this study is two-phase flows. The same experiments as with the single-phase flow
are done for two-phase flow. Results of these experiments are presented in chapter 6. Two-phase flows
studied here, consist out of a liquid phase (water) and a gaseous phase (air) flow. The most important
parameter for two-phase flow is the void fraction. It describes the gas to (gas and liquid) ratio. The void
fraction is a locally determined quantity defined as [3]:
A
Ai
ig∑=
,
α (5.1)
When considering for example a tube with a two-phase flow, A is the flow area of the tube and Ag is the
area occupied by the gas in the cross section of that tube. The summation sign is because of the
existence of more bubbles in one cross section. The main difference of two-phase flow to single-phase
flow is its compressibility. The two-phase flow consisting of water and air is compressible; the single-
phase flow consisting solely of water is incompressible. A feature of compressible flows is choking.
This phenomenon, applied to the globe valve, will be discussed in chapter 5.2.2. First the influence of
the void fraction on the pressure drop will be discussed in chapter 5.2.1.
5.2 Theory
5.2.1 Pressure drop
The content of this chapter is derived from v.d Geld [3].
Strongly idealized the Globe Valve can be seen as a tube. This simplification can be used to gain
insight in the relation between the pressure drop over the valve and the void fraction. The pressure drop
gradient in axial direction for a two-phase flow in a tube can be written as:
gravaccfric dz
dp
dz
dp
dz
dp
dz
dp++= (5.2)
The total pressure drop consists of three parts: friction, acceleration and gravitation. In this analysis
gravitation will be left out, because the set-up is horizontal. The pressure drop due to friction can be
written as:
w
fric
t
L
fric
xGc
Ddz
dp
ρ
22 ))1((2 −
Φ=− (5.3)
The symbol x represents the gas mass fraction, defined as:
1
)1(1
−
−+=
a
wxαρ
ρα (5.4)
2
LΦ (increasing with α ) and tD (constant) are positive numbers and fricc is defined as:
15
25.0
)1(−
−=
w
tfric
DxGc
µ (5.5)
In which µw (constant) is also a positive number. The pressure drop due to friction thus increases with
increasing gas mass fraction and thus void fraction.
The acceleration term in the pressure drop is defined as:
−
−+=−
)1(
)1( 222
αραρ waacc
xx
dz
dG
dz
dp (5.6)
Here G, the total mass flux, is a positive number and just like the pressure drop due to friction, the
pressure drop due to acceleration increases with increasing void fraction. From these two defined
pressure drops can be concluded that a two-phase pressure drop increases with the void fraction. The
pressure drop over the globe valve will increase with increasing void fraction.
5.2.2 Choking
Choking is a physical phenomenon that can appear in accelerated compressible fluids. Since the two-
phase flow, considered here, is compressible; choking is a phenomenon that should be taken in to
account. Here not an in depth analysis will be given. Solely the factors that can cause probable choking
will be discussed in a very simplified manner. Evidentially these factors are discussed in view of the
problem here: the valve.
Figure 5.1 contraction-pipe
Consider a gas that is applied to the contraction-pipe in figure 5.1. The mass flow can be increased by
increasing p1. As long as the Mach number at position 0 is smaller than 1, the mass flow will increase
with increasing p1. When M>1, the ratio p1/p0 remains constant with increasing p1. The flow is
choked and the mass flow is blocked. The difference between p0 and p2 (ambient pressure) will be
bridged with a shock-wave. For two-phase flows, another dimensionless number is used to see whether
choking occurs. This number is defined as [2]:
00DM tp α= (5.7)
where suffix 0 refers(in this analysis) to the position 0 in figure 5.1, suffix tp to ‘two-phase’, α is the
void fraction and D represents the momentum flux density, which is defined as [2]:
0
2
000
p
uD mmρ
= (5.8)
here ρ, u and p are respectively the density, velocity and pressure at position 0 in figure 1. The suffix m
denotes that these are averaged values for the whole two-phase flow. The two-phase flow is here
considered as homogenous. The dominating factor that thus can cause a two-phase flow to choke is the
velocity. This is the same as for gasses. If now the valve is taken in consideration (see figure 2.1). The
most critical position where choking can occur; is at the holes in the cage. The flow area in the holes is
the smallest and the velocity the highest. If choking appears, probably the downstream pressure
measurement gets influenced, because of the shock-wave. It is thus important to figure out; whether
there is a chance that choking has appeared in the experiments. Later with use of the numerical analysis
16
a first approximation will be made for the two-phase Mach-number. With this number the possible
appearing of choking can be determined. Note that this is only a first rough approximation. The model
for choking is strongly simplified, the valve has a complicated configuration and the numerical analysis
has only been done for single-phase flow.
5.3 Experiment Strategy
5.3.1 Operation points
Because of limits to the set-up, it is not possible to determine the local void fraction. In these
measurements the homogeneous void fraction is used. This is defined as the fraction of the air flow rate
to the total flow rate:
aw
a
QGVF
+= (5.9)
In the sequel of the report, void fraction refers to GVF and thus isn’t the actual local void fraction as
defined in formula 5.1. If different it will be denoted. The two-phase flow is applied to the set-up by
adding pressurized air in the water flow (12, figure 2.2). Void fraction is kept constant during a single
experiment. But experiments are done with different void fractions. With this method the same
experiments can be done for different void fractions. The only variable is the void fraction and thus the
effect of the void fraction on the measurements can be clearly studied. Since the void fraction and its
effect on the flow are of utmost importance in studying two-phase flows; this experimental strategy is
the most convenient. To maintain these constant void fractions during the experiments, where the water
flow rate is increased, the air flow rate (Qa) also has to be increased. The experiments are done for void
fractions of: 10%, 20% and 30%. The water flow rates used are the same as in the single-phase
experiments: from 100 l/min to 250 l/min with steps of 25 l/min. Only for GVF = 30% measurements
are done for water flow rates of 75 l/min to 175 l/min. It turned out that for air flow rates higher than 75
l/min the flow starts to oscillate and this is harmful for the set-up. Therefore only measurements on
GVF = 30% can be done for water flow rates up to 175 l/min. A maximum water flow rate of 175 l/min
namely requires a maximum air flow rate of 75 l/min. To have sufficient measurement points, here is
chosen to start measuring from a water flow rate of 75 l/min at GVF = 30%. The air flow rates (in
italic) for these different void fractions and water flow rates are presented below in table 5.1. The
values are in l/min.
Table 5.1 GVF, water- and air flow rates
GVF\ wQ 75
l/min
100
l/min
125
l/min
150
l/min
175
l/min
200
l/min
225
l/min
250
l/min
10% 11.11 13.89 16.67 19.44 22.24 25 27.78
20% 25 31.25 37.5 43.75 50 56.25 62.5
30% 32.14 42.86 53.57 64.28 75
These combinations of Qa and Qw are applied to Valve closings of 0mm, 10mm, 15mm, 20mm and
22.5mm. The empty spaces in table 5.1 are operation points which are not used.
Practicalities for two-phase flow experiments are same as for the single-phase experiments.
17
5.3.2 Tylan calibration The air is added to the water flow by an electronically controlled pump: Tylan. A set point for the
Tylan can be entered in Labview. This value has to be corrected, multiplied with a multiplication factor
(correction factor), to get the wished air flow rate into the set-up. In order to determine this correction
factor, the Tylan is calibrated with a rotameter. It has turned out that the correction factor is dependent
on the flow rate. The correction factor as a function of the air flow rate is given in figure 5.2.
10 20 30 40 50 60 70 80
2.5
3
3.5
4
Qa [l/min]
Corr
ection f
acto
r [-
]Tylan Calibration
Figure 5.2 Tylan calibration curve
The rotameter can only measure air flow rates up to a maximum of approximately 28 l/min. The graph
in figure 5.2 for Qa > 28 l/min is an extrapolation of the gradient measured around the value of 28
l/min. After figure 5.2 was determined, it is used in all two-phase flow experiments.
18
6. Two-phase Experiment Results
6.1 Pressure drop over the valve
The pressure drop over the valve is measured with two different methods, DpValve and DpASK92,
identical to the single-phase experiments (see chapter 4.2).
6.2 DpValve
Every single graph in figure 6.1 represents the measurements done for all void fractions and water flow
rates for one specific closing of the valve. Take note that all graphs have a different scale on the
vertical axis. Like with a single-phase flow the pressure drop over the valve increases with increasing
water flow rate and decreasing valve opening. The friction to the flow namely increases for increasing
flow rate and decreasing flow area. As a result the pressure drop over the valve increases. From figure
6.1 can be concluded that the effect of the void fraction on the pressure drop is not consistent. If the
valve is fully opened (0mm), a lower void fraction results in a higher pressure drop. For all other
closings of the valve (10mm to 22.5mm), a higher void fraction results in a higher pressure drop.
0 100 2000
50
100
Qw [l/min]
DpV
alv
e [
mbar]
DpValve 0mm
10%
20%
30%
0 100 2000
100
200
30010mm
Qw [l/min]
DpV
alv
e [
mbar]
0 100 2000
200
400
60015mm
Qw [l/min]
DpV
alv
e [
mbar]
0 100 2000
500
1000
1500
20mm
Qw [l/min]
DpV
alv
e [
mbar]
0 100 2000
1000
2000
3000
22.5mm
Qw [l/min]
DpV
alv
e [
mbar]
Figure 6.1 DpValve for different GVF and Valve Closings, Two-phase flow
The measurements for valve closings from 10mm to 22.5mm are in accordance to the theory of two-
phase flows (chapter 5.2.1). In a two-phase flow, the pressure drop namely increases with increasing
void fraction. Measurements on the fully opened valve are striking with theory.
Measurements for DpValve on the old set-up are only done for a fully opened valve. The results are
presented in Appendix 12.2.1. When the new measurements are compared to the old measurements, it
can be seen that DpValve for both is approximately equal. Though the values are approximately the
same, in the old measurements a higher void fraction results in a higher value for DpValve, like
expected from theory. This is in contrary to the new measurements on the fully opened valve but in
accordance to the new measurements when the valve is partially closed. With the results of the
numerical model (chapter 8) will be paid attention to the case of the fully opened valve. The flow field
can give insight in this remarkable behavior.
19
6.3 DpASK92 The results of the measurements on DpASK92 are presented in figure 6.2. Measurements up to
200mbar are possible with the DpASK92 sensor.
0 100 2000
50
100
Qw [l/min]
DpA
SK
92 [
mbar]
DpASK92 0mm
10%
20%
30%
0 100 2000
50
100
150
10mm
Qw [l/min]
DpA
SK
92 [
mbar]
0 100 2000
50
100
150
20015mm
Qw [l/min]
DpA
SK
92 [
mbar]
0 100 2000
50
100
150
200
20mm
Qw [l/min]
DpA
SK
92 [
mbar]
0 100 2000
50
100
150
200
22.5mm
Qw [l/min]
DpA
SK
92 [
mbar]
Figure 6.2 DpASK92 for different GVF and Valve Closings
The pressure drop over the valve is again plotted as a function of the water flow rate. It shows a similar
behavior as the measurements on DpValve. The pressure drop increases with increasing water flow rate
and increasing closing of the valve. Only for a fully opened valve again the pressure drop clearly
increases with decreasing void fraction. Measurements on DpASK92 for the old set-up are not
available. The measurements on DpASK92 are in principle the same as those for DpValve until
200mbar. The pressure taps used to determine the different pressure drops are the same. When
DpValve and DpASK92 in figures 6.1 and 6.2 are compared for all valve closings, indeed the pressure
drops have approximately the same values.
6.4 Pitot pressure drop
The measurements on the pitot pressure drop for two-phase flow are plotted in figure 6.3. In contrary to
the other pressure drops, DpPitot is plotted as a function of the total flow rate = water flow rate + air
flow rate. This is done because the measurements on the old set-up are also presented like this. A good
comparison is only possible if DpPitot in both cases is plotted as a function of the same variable. The
negative values for the pitot pressure drop for low flow rates are due to numerical reasons.
In figure 6.3 can be seen that a decreasing void fraction results in an increasing pitot pressure drop.
This can be understood from the following analysis. The pitot pressure drop can (simplified) be
expressed as:
2
2
1mixmixpitot vdp ρ= (6.1)
Here the suffix mix denotes that the variables, density and velocity, are average values for the total
mixture of water and air. Though in reality the mixture is compressible, in the analysis the density will
considered to be constant. The density and velocity of the total mixture can be expressed as:
wamix ρααρρ )1( −+= (6.2)
20
A
Q
A
QQv totwa
mix =+
= (6.3)
In which α is the void fraction, A the flow area and Qtot is the flow rate of the total mixture. Relations
6.2 and 6.3 substituted in 6.1 give as a result for the pitot pressure drop:
2
2))1((
2
1totwapitot Q
Adp ρααρ −+= (6.4)
Since water density ρw is about thousand times larger than air density ρa, dppitot will decrease with
increasing void fraction. This is in accordance to the graphs given in figure 6.3.
0 100 200 300
0
10
20
30
40
Qtot [l/min]
DpP
itot
[mbar]
DpPitot 0mm
10%
20%
30%
0 100 200 300
0
10
20
30
4010mm
Qtot [l/min]
DpP
itot
[mbar]
0 100 200 300
0
10
20
30
4015mm
Qtot [l/min]
DpP
itot
[mbar]
0 100 200 300
0
10
20
30
4020mm
Qtot [l/min]
DpP
itot
[mbar]
0 100 200 300
0
10
20
30
4022.5mm
Qtot [l/min]
DpP
itot
[mbar]
Figure 6.3 DpPitot for different GVF and Valve Closings
As stated in chapter 4.2 the pitot pressure drop for single-phase flow is hardly influenced by the closing
of the valve. This is also valid for two-phase flows (figure 6.3).
The big differences between the graphs in figure 6.3 for a valve closing of 22.5mm are caused by a lack
of measurement points. The pressure in the set-up namely became too high, for safety reasons the flow
rate was not increased. Safety restrictions to the pressure in the set-up are stated in chapter 3.1.
The measurements for the pitot pressure drop on the old set-up are presented in appendix 12.2.2.
Measurement results for single and two-phase flow show equal tendencies. Also the effect of the void
fraction is the same. The pitot pressure drop in both figures decreases with increasing void fraction.
The numerical values, in contrary, are totally different. The new measurements have resulted in much
higher values than the old measurements. New measured pitot pressure drops are approximately 5
times higher than the old ones. This difference between new and old measurements has already been
encoutered on the pitot pressure drop measurements for single-phase flow in chapter 4.2.
21
6.5 Fluid velocity in the globe valve An expression for the mean velocity of the fluid mixture can be derived from relation 6.1. It will be
adopted that the mixture density and velocity are homogeneous. The fluid velocity can be written as:
mix
pitot
mix
dpv
ρ
2= (6.5)
The graphs in figure 6.4 are determined with formula 6.5. Here is chosen to plot the velocity as a
function of Qw instead of Qtot. If the velocity is plotted as a function of Qtot, it’s namely not dependent
on the void fraction. Since the effect of void fraction is the main subject in this study; the choice for Qw
is convenient.
0 100 2000
0.5
1
1.5
2
2.5
3
3.5
Qw [l/min]
Velo
city [
m/s
]
GVF = 10%
0mm
10mm
15mm
20mm
22.5mm
0 100 2000
0.5
1
1.5
2
2.5
3
3.5GVF = 20%
Qw [l/min]
Velo
city [
m/s
]
0 100 2000
0.5
1
1.5
2
2.5
3
3.5GVF = 30%
Qw [l/min]
Velo
city [
m/s
]
Figure 6.4 Mean mixture fluid velocity as a function of water flow rate, different GVF and valve closings
In figure 6.4 can be seen that there exists a linear relation between the flow rate Qw and the velocity.
Moreover the gradient of the graph increases with increasing void fraction. This behavior can be
understood from the following simple analysis. The homogeneous void fraction can be written as:
aw
a
QGVF
+= (6.6)
This can be written differently as:
wa QGVF
GVFQ
−=
1 (6.7)
The total flow rate can be denoted, with use of 6.7, in terms of void fraction and water flow rate as:
GVF
QQ w
tot−
=1
(6.8)
22
Relation 6.3 and relation 6.8 together give:
)1( GVFA
Qv w
mix−
= (6.9)
Relation 6.9 shows exactly the behavior of the graphs in figure 5.6. The velocity increases linearly with
the water flow rate and the gradient )1(
1
GVFA − increases linearly with increasing void fraction.
23
7. Two-phase Multiplicator
7.1 Theory
Theoretical relations are available for predicting the pressure drop in single-phase flow through tubes
and appendices. With use of a correlation factor, the single-phase pressure drop can be correlated to the
two-phase pressure drop [3]:
liquidonly
Lodz
dp
dz
dp=Φ2
(7.1)
In which dz
dp is the pressure drop per unit length for a two-phase flow. The pressure gradient
liquidonlydz
dpis the pressure drop per unit length for a flow consisting solely of liquid. This can be
determined with classical relations for single-phase pressure drop. The correlation factor 2
LoΦ can be
expressed in terms of liquid properties, the geometry of the problem and liquid friction factors [3]. The
two-phase pressure drop can thus be determined with only liquid flow characteristics. This factor will
be determined in chapter 7.2 from the experimental results. If this factor is known for the valve, the
pressure drop in two-phase flow can easily be determined with only knowledge of the single phase
flow for different flow conditions.
7.2 Experiments
The correlation factor is determined by dividing the two-phase data for DpValve by the single-phase
data for DpValve for all valve closings:
svalve
tvalve
Lop
p
,
,2
exp,∆
∆=Φ (7.2)
In which ∆pvalve,t and ∆pvalve,s represent the pressure drop over the valve for two-phase and single-phase
flow respectively. Suffix exp indicates experimental. In relation 7.2 DpValve is not per unit length, like
in formula 7.1. This however is not a problem. DpValve is, both for single and two-phase flow,
measured over the same fixed distance. The derivative can thus be cancelled out. The way in which the
correlation factors are determined from all the data, taken from the experiments, is explained in the
sequel.
Measurements are done for different water flow rates, void fractions and valve closings. In order to
determine correlation factors which represent all these data, the correlation factors have to be averaged.
For a briefer notation the two-phase flow pressure drop over the valve will now be represented by ∆pt
and the single-phase flow pressure drop by ∆ps.
∆pt and ∆ps are row vectors. Every element of ∆pt and ∆ps is the pressure drop over the valve for a
specific water flow rate. First the mean is taken of
jis
t
p
p
,
∆
∆ . This mean is calculated for all
combinations of i and j which stand respectively for the different GVF (i = 1,2,3 = 10%, 20%, 30%)
and valve closings (j = 1,2,3,4 = 10mm, 15mm, 20mm, 22.5mm).
Next the root of this mean is taken in order to calculate the correlation factor:
24
5.0
,
,exp,,
∆
∆=Φ
jis
tjiLo
p
p (7.3)
In total there are now thus 3 GVF * 4 closings = 12 mean values for the correlation factor.
The measurements for a fully opened valve are not taken into account in these calculations because
those measurements are striking with theory. For all closings one mean is taken of
5.0
,
∆
∆
jis
t
p
pfor the three different GVF:
5.04
1 ,
1∑
=
=
∆
∆N
j jis
t
p
p
N for i = 1,2,3
These three remaining values represent the correlation factors for the three different GVF. The results
of these calculations are represented in figure 7.1 by the dots. Though there are only three measured
points (the point at GVF = 0 has to be 1 by definition of formula 7.1) a clear linear relation can be seen
between GVF and the correlation factor.
The line, in figure 7.1, is a 1st order polyfit of this data. The relation between the correlation factor and
GVF in the polyfit is:
GVF0.28241)(exp, ⋅+=Φ GVFLo (7.4)
0 0.1 0.2 0.3 0.4 0.5 0.6 0.70
0.5
1
1.5
2
2.5
GVF [-]
ΦLo,e
xp [
-]
ΦLo,exp (GVF) for Mokveld Globe Valve
polyfit of mean data
mean measured data
Figure 7.1 Correction factor as a function of the void fraction.
The correction factors derived from the old data are given in appendix 12.2.3. For GVF up to 0.4 the
graph in figure 7.1 shows an identical behavior to the graph in appendix 12.2.3. For higher GVF
values, the gradient of the graph in appendix 12.2.3 increases. The correction factor for the new data
remains constant, but this is because the 1st order polyfit is only based on the points until GVF = 0.3. In
fact the correction factor for the new data for GVF>0.3 is thus not known and can’t be compared to the
old data. The line (for GVF>0.3) in figure 7.1 is only a first estimate.
25
8. Numerical analysis
The experiments have given insight in the behavior of the valve. The mean velocity in the valve and the
pressure drop are measured for different flow conditions. These data however don’t give information
about the flow field through the valve. Knowledge about the flow field can be helpful in explaining
certain experimental results, also for two-phase flow. To visualize the flow field through the valve, a
numerical model of the valve is made. At first only the incompressible, single-phase, flow will be
modeled. The same data (pressure drop, velocity) as measured during the experiments are calculated
with the model. The numerical model can thus be verified by the experimental results.
A straight tube is placed between the diverging section and the inlet of the valve (2, figure 2.2). The
goal of this tube is to achieve a fully developed flow after the diverging section before entering the
valve. The numerical model can also be used to visualize the flow field inhere. With the model can be
determined whether the tube is functioning like it is supposed to. Moreover the numerical single-phase
results together with the correlation factors, determined in chapter 7.2, enable to approximate the two-
phase pressure drop over the valve.
The finite element computer code that is used to calculate the flow field is XENIOS, written in
FORTRAN. This code, developed at University of Rome “La Sapienza” is particularly used for solving
flow problems. It’s developed for solving 2D/3D and laminar/turbulent flows [6]. The equations that
form the basic of the code are presented in the next chapter.
8.1 XENIOS
The content of this chapter is derived from Andrea Santoriello [5].
The flow problem considered in the numerical analysis is solved with use of a k-ε model in its low
Reynolds extension. Transport of turbulent kinetic energy k is of great importance in turbulent flows.
With a transport equation for k, the turbulent velocity scales can be determined. However the turbulent
length scale l* also has to be determined, locally. The most simple and widespread solution, in order to
determine l*, is to design a transport equation for the viscous dissipation rate ε. This gave rise to the
two equations of the k-ε model. The principle of this model is to take in account the equations for the
turbulent length and velocity scale: l*
and u*
with which the eddy viscosity νt can be calculated. The
expressions are:
ε
2/3* k
l = (8.1)
ku =* (8.2)
εν µµ
2**** ),(
kculcult == (8.3)
cµ is a model constant. The equations that complete the k-ε model are the transport equations for the
turbulent kinetic energy k and the viscous rate of turbulent kinetic energy dissipation, ε. Expression for
transport of turbulent kinetic energy:
jkv
uSku j
k
tjiijtjj ,2 ,,,
++−=
σνεν (8.4)
Where σk in an effective Prandtl number for diffusion, ju a velocity component and Sij the mean strain
rate tensor.
26
Transport equation for the viscous rate of turbulent kinetic energy dissipation:
jv
CuSCu jt
jiijtjj ,1
21
,2,1,
++−= ε
σνε
τν
τε
ε
εε (8.5)
Where Cε1 and Cε2 are constants of which the meaning goes beyond the discussion here, σε is an
effective Prandtl number for dissipation diffusion and τ is the turbulent time-scale. There are different
versions of turbulence models. A model that gives good solutions also near the walls for low, transient
and high Reynolds numbers is the k-ε model in its low Reynolds extension. This model makes use of a
modified dissipation rate ε~ which has homogenous boundary conditions on solid walls. This modified
dissipation rate is defined as:
( ) Dxk i −=∂∂−= ενεε2
/2~ (8.6)
The transport equations for the turbulent kinetic energy and the modified dissipation rate now read:
Djkv
uSku j
k
tjiijtjj −
++−= ,~2 ,,,
σνεν (8.7)
Ejv
fCuSfCu jt
jiijtjj +
++−= ,~1
21~
,22,11, εσ
νετ
ντ
εε
εεεε (8.8)
where fε1 = 1, fε2 and E are defined below:
[ ])Reexp(3.01 2
2 tf −−=ε (8.9)
22 )/(2 kjit xxuE ∂∂∂= νν (8.10)
To solve the complete set of equations, the boundary conditions have to be defined. This will be done
in the sequel of the report. The equations are solved in steady-state so no initial conditions are required.
All equations together with the boundary conditions have to be implemented in the Finite Element
Method which can be solved by XENIOS. For details on the numerical model, the reader is referred to
[1].
8.2 Grid generation
The configuration of the valve and the mesh are made with use of the program ‘Gridgen’. Since the
valve is rotational symmetric, the valve is modeled in only 2 dimensions. The total configuration
contains: a small inlet section, the diverging section, the straight tube for development of the flow, the
valve and an outlet section. In total the mesh consists of +-55.000 nodes. This number has turned out to
result in a sufficient accuracy. The grid is not everywhere equally spaced. Closer to the boundaries, the
grid becomes finer. At the boundaries the highest velocity gradients exist and to be able to calculate
these with a good accuracy, a local fine grid is required.
Circumferentially the cage (in the closing mechanism) has rows of five respectively six holes next to
each other (figure 2.1). Here is chosen to make the cross-section at a row with six holes. The
experiments are done with six different closings of the valve. In reality, with the closings used in the
experiments, the holes sometimes are partially closed. In the model, partially closing of the holes gives
big difficulties for the creation of the mesh. Therefore the different closings are modeled with only
fully closed and fully opened holes.
27
8.3 Boundary conditions
8.3.1 Velocity
Inlet profile
The Inlet velocity profile has to be defined to be able to calculate the flow field in the configuration.
The inlet of the configuration is the left side from the inlet section left from the diverging section (1,
figure 2.2). Fluid velocity in radial direction is in the inlet equal to zero. The velocity profile for the
axial velocity which will be adopted here is :
m
R
rvrv
/1
max 1)(
−= (8.11)
Here r is the radial coordinate, R the inner radius of the tube (R = 25.4mm at the inlet) and vmax the
maximum velocity at r = 0. The parameter m sets the thickness of the boundary layer. The flow is
considered to be turbulent and fully developed at the inlet. The velocity profile that fits best to these
specifications can be approximated with m = 8. The maximum velocity vmax is dependent on the flow
rate. The flow rate can be expressed in terms of velocity by:
θθππ
rdrdR
rvrdrdrvQ
mRR
w ∫ ∫∫ ∫
−==
2
0
/1
0
max
2
0 0
1)( (8.12)
Here θ is the tangential coordinate. The maximum velocity can be left outside the integral because it is
independent of both the radial and tangential coordinate. Relation 8.12 results in a linear relationship
between the maximum velocity and the flow rate:
∫ ∫
−
=π
θ2
0
/1
0
max
1 rdrdR
r
Qv
mR
w (8.13)
The integral in relation 8.13 is numerically approximated with use of a MATLAB routine. The relation
between the flow rate and the maximum velocity that follows from this integration is presented in
figure 8.1.
The velocity profile that has to be entered in XENIOS has to be normalized by the mean velocity. The
mean velocity can easily be determined by:
A
Qv w
mean = (8.14)
In which A is the flow area of the inlet tube. An expression for the velocity profile normalized by this
mean velocity is given in formula 8.13:
m
meanmean R
r
v
v
v
rvrv
/1
max 1)(
)(
−== (8.15)
Substitution of relations 8.13 and 8.14 in relation 8.15 give that the normalized velocity profile is
independent of the flow rate. The normalized velocity is plotted in figure 8.2. In figure 8.2 can be seen
that the boundary layer is very thin, like should be for a developed turbulent flow. The maximum
velocity is indeed at r = 0 and the corresponding value in figure 8.2 is greater than 1, thus vmax > vmean.
28
100 150 200 250
1
1.2
1.4
1.6
1.8
2
2.2
Maximum velocity inlet velocity profile
Qw [l/min]
v ma
x [
m/s
]
-25 -20 -15 -10 -5 0 5 10 15 20 250
0.2
0.4
0.6
0.8
1
r [mm]
v/v
mea
n
Normalized velocity Inlet tube
Figure 8.1 Maximum velocity at the inlet tube as function
of flow rate.
Figure 8.2 Normalized velocity profile at inlet tube as
function of radius.
Walls
At the walls the no-slip condition is applied, both the vertical and horizontal velocities are equal to
zero.
8.3.2 Turbulent kinetic energy, viscous dissipation rate
Inlet profile
Both the kinetic energy and the viscous dissipation rate are normalized in the numerical model. The
kinetic energy inlet profile can be calculated with use of relation 8.2. At the inlet the normalized u* is
equal to the normalized velocity of formula 8.15. The inlet profile for the normalized kinetic energy
now reads:
2
)()(
=
mean
iv
rvrk (8.16)
where the suffix i refers to inlet. For determination of the dissipation rate at the inlet, next to the
turbulent velocity scale, the turbulent length scale is used. Dissipation rate can be calculated with:
*
2/3
l
k=ε (8.17)
The turbulent length scale l* at the inlet, has in the numerical simulation a length of 0.1*Lref. Lref is the
diameter of the inlet tube. The expression for the dissipation rate at the inlet now reads:
ref
mean
iL
v
rv
r1.0
)(
)(
3
=ε (8.18)
29
Walls
For the kinetic energy and viscous dissipation rate at the walls, the no-slip condition is applied. Kinetic
energy and viscous dissipation rate are zero at the walls.
8.4 Pressure and Velocity normalization
The pressure that can be read in the results of the numerical simulations calculated by XENIOS can be
expressed as:
CppXENIOS += (8.19)
where C is a constant value, different for each simulation. The data which can be read from the
numerical analysis is normalized. p is a normalized pressure which is defined as:
2
meanv
pp
ρ= (8.20)
where ρ is the density of water and vmean the actual mean velocity at the inlet of the configuration. The
horizontal bar on top indicates the normalization. Normalized velocity can be expressed as:
meanv
vv = (8.21)
where v is the actual velocity. This normalized velocity is also the velocity which is read in the results
of XENIOS: XENIOSvv = . Since two different flow rates are applied, two different mean velocities
exist. For Qw = 150 l/min, vmean = 1.2335 m/s and for Qw = 250 l/min, vmean = 2.0558 m/s. In the sequel
of the report, the normalized values will be used. It will be indicated if the flow rate is 150 l/min or 250
l/min which implies how the experimental values can be converted to actual values.
8.3 Flow rate and Valve closings
Because of the high number of grid points, the computation time for the numerical model is large.
Therefore with only two water flow rates (Qw = 150 l/min and Qw = 250 l/min) numerical experiments
are done. These flow rates will be applied to meshes which represent two closings of the valve: 0mm
and 20mm.
Next to the mesh, inlet velocity profile and boundary conditions the Reynolds number at the inlet has to
be defined in XENIOS. The Reynolds number here is based on the mean velocity and the diameter of
the inlet tube (D = 2”). The temperature of the water during the experiments was on average 25°C. For
these conditions the Reynolds numbers at the inlet are: Re = 5.9448e3 for Qw = 150 l/min and Re =
7.0214e4 for Qw = 250 l/min.
8.4 Results
The results of the numerical analysis give a view on the flow field in the valve. Special attention will
be paid to two interesting zones: the tube after the diverging section and the section after the cage. The
numerical model calculates: pressure, velocity, kinetic energy and dissipation at every grid point. In
order to determine whether the numerical model has succeeded in achieving a satisfactorily
convergence, some model results are compared to the results of the experiments. The valve pressure
drop and the pitot pressure drop are taken for this comparison.
30
8.4.1 Valve Pressure drop and Pitot pressure drop
Valve pressure drop is presented in table 8.1; numerical results at the left side and experimental results
at the right side. Pitot pressure drop is presented in table 8.2; numerical results at the left side and
experimental results at the right side.
Table 8.1 Valve pressure drop determined by numerical model (left), experiments (right)
Flow rate/Valve closing 0mm 20mm 0mm 20mm
150 l/min 63 mbar 105 mbar 150
l/min
50 mbar 140 mbar
250 l/min 216 mbar 293 mbar 250
l/min
460 mbar 1200 mbar
Table 8.2 Pitot pressure drop determined by numerical model (left), experiments (right)
Flow rate/Valve closing 0mm 20mm 0mm 20mm
150 l/min 24 mbar 14 mbar 150
l/min
5 mbar 16 mbar
250 l/min 80 mbar 40 mbar 250
l/min
10 mbar 22 mbar
From Table 8.1 can be concluded that the model succeeded in achieving a good convergence for Qw =
150 l/min. However for Qw = 250 l/min, the differences between the model and the experiments are too
large. The higher flow rate results in a higher Reynolds number and more iteration steps are needed for
the model to reach convergence. Numerical calculations for the pitot pressure drop give only a good
result for Qw = 150 l/min with a valve closing of 20mm. Reason for this is not encountered in the scope
of this study. Though the numerical model only partially achieved convergence, it will be used for
further analysis. Analysis focuses mainly on flow field, pressure and velocity contours and not on
numerical values.
8.4.2 Flow field
A streamline plot of the inlet section, diverging section and straight tube can be seen in picture 8.4.
Figure 8.3 streamline plot, Qw =150 m/s, fully opened, inlet, diverging section, straight tube
In figure 8.3 can be seen that the flow, for these flow conditions, is perfectly aligned with the
boundary. The diverging section doesn’t cause the boundary layer to get loose from the wall. The
straight tube between the diverging section and the valve is thus not necessary for Qw = 150 l/min. The
flow in this section of the configuration for a partially closed valve shows similar behavior. The flow
field upstream of the cage is not influenced by the closing of the valve. For Qw = 250 l/min the same
steam line profile in this section of the configuration is obtained. From these results can be concluded
that during all experiments the boundary layer doesn’t get loose from the walls of the diverging tube.
The stream line patterns for the different numerical experiments only differ around the cage, the
discussion focuses now on that specific part. Stream line plots for the four different experiments are
shown in figure 8.4. The most interesting feature is the big vortex that appears at the boundary right
from the most right hole of the cage. The vortex appears because the flow has to flow around the sharp
edge. The edge causes the boundary layer to get loose from the wall. Both for Qw = 150 l/min and Qw
31
= 250 l/min partially closing of the valve causes an enlargement of the dimensions of the vortex. The
closing of the valve causes a substantially increase of the fluid velocity at the exit of the holes in the
cage. As a consequence the tangential velocity increases and thus the size of the vortex increases.
When Qw = 150 l/min is compared to Qw = 250 l/min hardly any differences appear in the streamline
patterns. Increasing the flow rate evidently causes also an increase in the fluid velocity at the exit of the
holes. This increase however seems to be too less to cause visible enlargement of the vortex.
Figure 8.4 Stream line pattern, upper row: Qw = 150 l/min, lower row: Qw = 250 l/min, left column: opened valve,
right column: valve closing = 20mm
Pressure measurement
The pressure in the valve is circumferentially measured at four equidistant points in the wall of the
valve. Exit pressure measurement is located at X = 7.5. In the situation that the valve is closed for
20mm, the exit pressure tap measures inside the vortex. This however doesn’t obstruct the pressure
measurement. The pressure is radial approximately constant, see figure 8.4 – 8.6. Though the vortex,
the correct pressure is measured at the exit.
In chapter 6.2 the remarkable behavior of the pressure drop over the valve is encountered for zero
closing. The stream line pattern in figure 8.4 for the zero closing however doesn’t show deviate
behavior. The nature of the valve with zero closing: valve pressure drop increases with decreasing void
fraction can’t be explained with single-phase numerical simulation. Two-phase flow numerical
simulations are necessary for a more extensive study on this topic.
8.4.3 Pressure contour in valve
In this chapter an analysis is given of the pressure contours in the configuration for the different
experiments. Take notice that the values of the pressure are normalized. The read values have to be
multiplied by 2
meanvρ to obtain the actual pressure. The multiplication factor is evidentially different
for the two flow rates: Qw = 150 l/min: 15 mbar and for Qw = 250 l/min: 42 mbar.
In figures 8.5 and 8.6, the pressure contours are presented for both the fully opened and the partially
closed valve at Qw = 150 l/min. Stream upwards of the cage, the pressure is much higher at the partially
closed valve than is the case in the fully opened valve. Moreover the pressure gradient in the flow
stream upwards of the cage is much smaller in case of the partially closed valve. The pressure stream
downwards of the cage is for both open and partially closed approximately the same. These differences
are caused by the difference in pressure drop over the holes in the cage. In case of the partially closed
valve, the fluid is forced to fewer holes than in the case of the fully opened valve. As a consequence the
fluid has to be accelerated to a higher velocity and thus the pressure drop over the holes is much higher.
The higher pressure drop over the holes in the case of the partially closed valve requires a higher
pressure at the inlet of the holes. When all holes are opened the decrease in pressure over the valve is
32
more gradually. As a result a clearly visible pressure gradient in figure 8.5 (open valve) can be seen
and not in figure 8.6 (partially closed valve).
In figure 8.7 the pressure contour in the valve for Qw = 250 l/min is plotted. In absolute sense the
pressure in the valve at Qw = 250 l/min is higher than in the valve at Qw = 150 l/min. To maintain a
higher flow rate, a higher inlet pressure is required. Moreover the same difference between Qw = 150
l/min and Qw = 250 l/min can be seen as is the difference between the fully opened valve and the
partially closed valve. The pressure stream upwards of the cage is relatively higher to the pressure
stream downwards of the cage at Qw = 250 l/min than at Qw = 150 l/min. The velocity through the
holes at Qw = 250 l/min is higher than at Qw = 150 l/min which results in the higher pressure drop over
the holes. The results of the numerical experiment with Qw = 250 l/min and a partially closed valve are
not shown. The differences to the fully opened valve with Qw = 150 l/min are a sum up of the effects
which are explained with figure 8.6 and 8.7 in comparison to figure 8.5.
Figure 8.5 Pressure Contour, Qw = 150 l/min, valve fully opened
Figure 8.6 Pressure Contour, Qw = 150 l/min, valve 20mm closed
Figure 8.7 Pressure Contour, Qw = 250 l/min, valve fully opened
33
8.4.4 Velocity contour in valve
In figures 8.8 and 8.9 the velocity contours in the valve (with a flow rate of Qw = 150 l/min) are shown
for the fully opened valve and the valve with a closing of 20mm. The fluid velocity deceleration (-
12<X<0) of the partially closed valve is much higher than is the case in the opened valve. This is
caused by the higher pressure stream upwards of the cage for the partially closed valve in comparison
to the opened valve (chapter 8.4.3). The velocity in these holes in the case of the partially closed valve
is evidently higher. Stream downwards of the cage, the velocity profile (perpendicular to the flow
direction) in the case of the fully opened valve is more flat than in the partially closed valve. Like
stated in chapter 8.4.2, the closing of the valve causes an increase of the size of the vortex. Which
causes, downstream of the cage, a decrease of the flow area. As a consequence the velocity profile is
sharper when the valve is partially closed. At last the velocity contours of the fully opened valve with
Qw = 150 l/min and Qw = 250 l/min are compared. The velocity contour of Qw = 250 l/min is
presented in figure 8.10. There are hardly differences in the velocity contours. The most important
difference is the higher value for the velocity everywhere in figure 8.10 compared to figure 8.9.
Figure 8.8 Velocity Contour, Qw = 150 l/min, valve fully opened
Figure 8.9 Velocity Contour, Qw = 150 l/min, valve 20mm closed
Figure 8.10 Velocity Contour, Qw = 250 l/min, valve fully opened
Choking
With use of the results of the numerical analysis, probable appearance of choking can be determined.
This will be done with use of the characteristic value Mtp. Like noticed in chapter 5.2.2, this should be
done very carefully. First of all the fluid in the numerical analysis is not compressible. Moreover the
geometry is far more complicated than in figure 5.1, on which the choking criterion is based. The
situation with the largest chance on choking; is the case with high flow rate and small valve opening.
To examine Mtp, experimental data is needed because C (chapter 8.4) has to be determined. C is needed
to determine the normalized velocity and pressure. For Qw = 250 l/min and valve closing 20mm, this
data is not available. Therefore the case with Qw = 150 l/min and a valve closing of 20mm will be
analyzed.
34
The void fraction determines the compressibility of the fluid. The numerical calculations are done for
an incompressible fluid. A low value for α approaches best the incompressible fluid. Here, α = 0.1 is
taken. The constant C is determined with use of the experimental data. Sequentially the velocity and
pressure are determined with C. Normalized density is equal to 1. D is determined at several points
around and in the holes in the cage. In the cage the highest fluid velocities appear; as a consequence
here is the biggest chance on choking. It turned out that the highest value for Mtp (=0.062) appears at
the exit of the most right hole. This value is much smaller than 1 and thus no choking would appear
with these conditions. Since the circumstances are extremely idealized this conclusion may not be
taken. Nevertheless probably choking has the greatest chance to appear at the exit of the most right
hole. If later numerical calculations with compressible fluid are done, extra attention has to be paid to
this particular region.
35
9. Conclusions
Both for the single and the two-phase flows, the new data is compared to the old data. The pressure in
the set-up and the pressure drop over the valve turned out to be almost equal for the new and old data.
Only in two-phase flow for a fully opened valve an error is encountered. The pressure drop over the
valve for the new measurements showed opposite behavior to the old measurements. In the new
measurements the pressure drop decreases with increasing void fraction. This is both striking with the
old measurements and with theory. This striking behavior, for the fully opened valve, is both found for
the measurements DpValve and DpASK92. Both measurement methods use the same pressure taps, but
are differently connected to the computer. The behavior can’t thus be caused by electronics or the
computer. When the valve is, partially, closed (for all applied closings) the measurements are according
to theory. The difference is thus most likely caused by the set-up or valve itself and not by the sensors.
The pitot pressure drop and the there from derived fluid velocity show both in the single as in the two-
phase flow accordance to theory. The old and new measurements also show the same tendencies.
However the numerical values for the pitot pressure drop don’t show accordance. The difference
between the values, gained from the old and the new experiments, increases with increasing flow rate.
This difference can only be caused by the sensors, because the minor changes in the set-up can’t cause
these big differences in the pitot pressure drop. It can be concluded that in general the new
measurements are a reliable base for the development of a model. Beside of some exceptions the new
results showed good accordance to its references (old measurements and theory). The new
experimental results can be used for further calculations and as a reference for the numerical model.
For the considered void fraction range, the liquid only multiplication factor derived from both the old
and new measurements have turned out to be equal. This could also be expected when only the
measurements on the pressure drop over the valve for single and two-phase flow would have been
considered. The coefficient is only dependent on these pressure drops. For these pressure drops, the
same values are found in the old and new measurements and thus the coefficients have to be equal.
With this multiplication factor and single-phase flow data or model a first approximation for two-phase
flow pressure drop can be made.
The first attempt for a model of the valve for two-phase flow is made by means of the numerical model
in XENIOS. The numerical model has succeeded in approximating the pressure drop, in single-phase
flow for low flow rate, over the valve. For predicting the pressure drop, the currently obtained
convergence is sufficient. The numerical calculated pitot pressure drop in contrary differs too much
from the experimental obtained results.
The numerical model does not contain effects of compressibility and the multiplication factor is only
known for a small range of void fractions. The numerical model for single-phase flow however,
together with the multiplication factor can make an approximation of the pressure drop for the two-
phase flow. This is a good starting point for the final development of a model for a two-phase flow
through the valve.
36
10. Recommendations
The difference in measured pitot pressure between new and old data grows with increasing flow rate.
The pitot tube seems to generate an error dependent of the flow rate. This problem has to be
investigated in order to take in account for probable new measurements. If the error is constant (as a
function of the flow rate) this can be subtracted from new measurements. If not, the pitot pressure tube
has to be replaced.
The Tylan Calibration is limited by the limits of the rotameter. The rotameter can only measure until a
maximum flow rate of Qw= 28 l/min. For the measurements, air flow rates up to 75 l/min are needed.
The air flow rates from 28 l/min till 75 l/min are thus estimated. A rotameter with a larger maximum
air flow rate would result in more precise air flow rates in the set-up.
Though the void fraction is a local parameter, the homogenous void fraction is used in the report. It is
advised to use a local void fraction measurement for new experiments. This possibly gives more
precise relations between the void fraction and the different measured pressure drops and velocities.
When the set-up, in two-phase flow, is applied to higher air- and water flow rates the separation tank
starts to flow over. A larger separation tank would prevent the set-up from this problem. Also the water
tanks can better be enlarged. The tanks frequently overflow, because of there small capacity.
It turned out that the numerical model not has reached convergence for all experiments. At first the
number of iteration steps has to be increased for the experiments where convergence has not been
reached.
Only a numerical model for incompressible fluid has been developed. Together with the multiplication
factor (chapter 7) this gave a first attempt for the model of the valve. A numerical model with
compressible fluid however will give more insight in the actual behavior of the two-phase flow in the
valve. The remarkable behavior that the pressure drop over the valve increases with decreasing void
fraction (chapter 6.2) has to be studied more extensively. A numerical model of the valve for
compressible fluid can possibly give more insight in this behavior.
Possible choking can also be studied more carefully with a compressible fluid model. It can be
determined if the pressure measurements are affected by this possible choking. Moreover the vortex in
two-phase flow is an interesting feature to study. The vortex in single-phase flow doesn’t affect the
pressure taps, but for two-phase flow this can be different.
37
11. Bibliography
[1] Borello D, Corsini A and Rispoli F, A finite element overlapping scheme for turbomachinery flows
on parallel platforms, Computers and Fluids, Elsevier, 2003, 32/7, 1017-1047.
[2] Davis M.R, Wang D, Dual pressure drop metering of gas-liquid mixture flows, Int. J. Multiphase
Flow Vol. 20, No. 5, pp. 865-884, 1994
[3] Geld, C.v.d., Meerfasenstromingen met warmte-effecten, 2 November 2000.
[4] MOKVELD www.mokveld.com
[5] Santoriello A. Multiscale Finite Element Methods for turbulence modeling in turbomachinery CFD,
November 2005
[6] XENIOS code hand-book, DMA – University of Rome “La Sapienza”, November 2003
38
12. Appendix
12.1. Single-phase old measurements
12.1.1. Pa
0 50 100 150 200 2500
0.5
1
1.5
2
2.5Pa single phase old data
Pa [bar]
Qw
[l/m
in]
open
10mm
15mm
20mm
12.1.2. DpValve
0
200
400
600
800
1000
1200
0 50 100 150 200 250 300 350
Series1 0 1.5 2
2.5 2.25 2.75
39
12.1.3. DpPitot
0
2
4
6
8
10
12
14
16
18
0,0 50,0 100,0 150,0 200,0 250,0 300,0 350,0
Flow rate (l/min)
Pre
ssu
re d
iffe
ren
ce (
mb
ar)
Model 25.6-1 25.6-3 9.7 11.7 13.7
40
12.2 Two-phase old measurements
12.2.1 DpValve
0
20
40
60
80
100
120
140
160
180
200
0,00 50,00 100,00 150,00 200,00 250,00 300,00 350,00
Flow rate (l/min)
Pre
ssu
re d
iffe
ren
ce
(m
bar)
Liquid only GVF 10% GVF 20% GVF 30% GVF 40%
12.2.2 Pitot Pressure
0
1
2
3
4
5
6
7
8
0,0 50,0 100,0 150,0 200,0 250,0 300,0 350,0
Flow rate (l/min)
Pre
ssu
re d
iffe
ren
ce (
mb
ar)
Liquid only GVF 10% GVF 25% GVF 40% GVF 55%
12.2.3 Two-phase multiplicator
0
0,5
1
1,5
2
2,5
0 0,1 0,2 0,3 0,4 0,5 0,6 0,7
GVF (-)
ΦΦ ΦΦtp
(-)
unfiltered data Chisholm model GVF 0% 100 l/min 167 l/min