    D I A M O N D D I A G N O S T I C S

    Service Manual

    Diamond Diagnostics Inc

    Holliston, MA. USAwww.diamonddiagnostics.com

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    3/41

    Service Manual Contents

    Chapter 1........................................................................................................................................................1Product Description ................................................................................................................................1 1.a System Overview .................................................................................................................................11.b Front and Rear Assembly Components ..............................................................................................41.c Electrode Housing Assembly Components .........................................................................................81.d Replacement List...............................................................................................................................10

    Chapter 2......................................................................................................................................................11Troubleshooting......................................................................................................................................11 2.a Troubleshooting Overview.................................................................................................................112.b Troubleshooting Guide ......................................................................................................................122.c DC Power Verification .......................................................................................................................132.d Display Verification............................................................................................................................132.e Probe Motor Verification....................................................................................................................142.f Printer Verification .............................................................................................................................152.g Sample Detector Verification.............................................................................................................162.h Serial Communication Verification.....................................................................................................17

    Chapter 3......................................................................................................................................................18Repair.........................................................................................................................................................18 3.a Repair Overview................................................................................................................................183.b Recommended Tools ........................................................................................................................183.c Rear Housing Removal .....................................................................................................................193.d Rear Housing Installation ..................................................................................................................193.e Power Supply Removal .....................................................................................................................213.f Power Supply Installation ..................................................................................................................213.g Printer Removal.................................................................................................................................223.h Printer Installation..............................................................................................................................223.i CPU Board Removal .........................................................................................................................253.j CPU Board Installation ......................................................................................................................253.k Display Removal................................................................................................................................273.l Display Installation.............................................................................................................................273.m Pump Motor Removal........................................................................................................................303.n Pump Motor Installation.....................................................................................................................303.o Probe Motor Assembly Removal.......................................................................................................323.p Probe Motor Assembly Installation....................................................................................................323.q Guide Rod Removal ..........................................................................................................................353.r Guide Rod Installation .......................................................................................................................353.s Electrode Jack Removal....................................................................................................................373.t Electrode Jack Installation.................................................................................................................37

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    4/41

    Chapter

    1

    P R O L Y T E A N A L Y Z E R

    1

    Product Description1.a System Overview

    TheproLYTE analyzer is designed to make service and repair quick and easy. Modular construction is usedthroughout the design. The basic operating system consists of a motor which drives the peristaltic pump.

    The peristaltic pump draws the desired fluid through the system. A second motor drives the sample probethrough the solutions valve to aspirate the appropriate fluid for pumping: sample; standards; or air. Thesample detector in the fluid path communicates the position of the fluid in the line.

    Please note that mechanical and electrical assemblies can be replaced. Follow the instructions in theTROUBLESHOOTING SECTION to determine if assembly replacement is necessary. Refer to theReplacement List in this section which lists replaceable assemblies, components, and appropriate partnumbers.

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    5/41

    P R O L Y T E A N A L Y Z E R

    2

    YES

    *ANALYZE URINE?*

    *SEE RESULTS?*

    *PACK USAGE?*

    *QUALCONTROL?*

    *OPER FUNCTS?*

    *DIAGNOSTICS?*

    *SECOND MENU?*

    *DATE/RANGES?*

    RUN CALIBRTN

    STANDBY MODE?

    DAILY CLEANER?

    ANALYZE BLOOD?

    YES

    YES

    YES

    YES

    YES

    YES

    YES

    YES

    YES

    NO

    YES

    NO

    NO

    NO

    NO

    NO

    NO

    NO

    NO

    NO

    NO

    Whole Blood, Serum, or Plasma Analysis

    End/Beginning of Day Cleaning

    Fluid Conservation, Maintenance

    Urine Analysis

    Results Displayed, Printed, Deleted

    Check % Pack Used, Install New Pack, Reset % Countdown

    Run Controls, See Last Control, See/Print Statistics,Delete Results

    On Demand Calibration, Wash, Solution Prime, Probe Wiping,I.D.#, Syringe Sample, Diluted Values

    Check: Calibration Values; Fluid Flow; Electrode Values; SampleDetector; Pump Calibration

    Change Date/Time, Cal Frequency, 1stCal, Set Na/K/Cl Ranges

    NOYES

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    6/41

    P R O L Y T E A N A L Y Z E R

    Table of Symbols

    NOTES Helpful information or references toother related sections of the manual.

    KEY INFORMATION Could take theproLYTE analyzer outof operation.

    3Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    7/41

    P R O L Y T E A N A L Y Z E R

    1.b Front and Rear Assembly Components

    This section describes the main assemblies of the proLYTEfront and rear assemblies. Refer to Figures P1,P2, and P3for the location of each component.

    Front Housingfront case of the analyzer

    Power Supply powers the analyzerEntry Moduleconnection for the power cableFuse Bayholds 2 fusesPower Supply Cableconnects the power supply to the CPU assemblyVoltage Selector to select 110V or 220V power source

    CPU Boardcontains the CPU board with digital components and analogsensor inputs:

    Auto-Sampler Portport for Auto-Sampler deviceSerial Portport for RS-232 for serial device

    Display AssemblyshowsproLYTE prompts/results on LCD screenDisplay Screen2 line displayDisplay/Sample Detector Cableconnects the display to the CPU assembly and sample detector tothe display board

    Peristaltic Pump Assemblyperistaltic pump to move fluids through the analyzerPeristalticPump Motorpowers pumpPeristalticPump Head Assemblycontains pump rotor and rollersPeristalticPump Connectionconnects the pump to the CPU assembly

    Probe Motor Assemblymoves the probe armLinear Actuatorpowers lead screwProbe Motor Supportholds probe motor assembly in placeSampleProbe Position Locatorindicates the home position for the sample probe armOptical Switch senses the location of the probe arm using the position locator andconnects to the CPU board

    Probe Armholds and moves the probe to aspirate samples and calibrants

    Printer prints a hard copy of analysis, diagnostics or maintenance results

    Printer Cableconnects the printer to the CPU assembly

    Electrode Jack(s)connects electrodes to the CPU assembly

    Rear Housingrear case

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    8/41

    P R O L Y T E A N A L Y Z E R

    PrinterFront Housing

    Paper Feed Button

    RearHousing

    ProbeArm

    ElectrodeJacks

    SampleProbe

    DisplayScreen

    Yes/NoTouchpad

    SolutionsValve

    PeristalticPump

    Figure P1 Front Assembly

    5Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    9/41

    P R O L Y T E A N A L Y Z E R

    Probe

    PositionLocator

    DisplayCable

    CPUBoard

    Figure P2 Rear Assembly

    LinearActuator

    LinearActuatorConnection

    OpticalSwitchConnection

    PeristalticPumpConnection

    Serial

    Port

    Auto-SamplerPort

    PowerSupplyCable

    VoltageSelector

    PowerSupply

    FuseBay

    EntryModule

    AC/AC Step

    DownTransformer

    PrinterCable

    6Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    10/41

    P R O L Y T E A N A L Y Z E R

    ElectrodeJacks

    OpticalSwitch

    ProbeMotorSupport

    PeristalticPump

    Assembly

    PeristalticPump Motor

    CPUMounts

    Figure P3 Rear Assembly

    7Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    11/41

    P R O L Y T E A N A L Y Z E R

    1.c Electrode Housing Assembly Components

    This section describes the components of theproLYTE electrode assembly.Refer to Figure P4for the location of each component.

    Electrodessodium, potassium, chloride, and reference

    Sample Detectordetects presence of air/liquid, sets sample position inside electrodes

    Internal Filling Solution 2 molar KCL solution, acts as a salt bridge for the referenceelectrode

    Fill Plugprevents the internal filling solution from evaporating

    Membrane Assemblycreates the connection between the reference electrode and the other

    electrodes

    8Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    12/41

    P R O L Y T E A N A L Y Z E R

    Fill Plug

    Fill Solution

    Membrane Assembly

    Sam le Detector

    Potassium Electrode

    Sodium Electrode

    Chloride Electrode

    Electrode Housing

    LocatingPins

    ReferenceElectrode

    Figure P4 Electrode Housing Assembly

    9Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    13/41

    P R O L Y T E A N A L Y Z E R

    1.d Replacement List

    Description Part No.

    Electrode Housing Na/K/Cl ME-0885D

    Power Supply ME-1676DCPU Board ME-1937DDisplay, 2 Line ME-1901DPump ME-1271DProbe Motor Assembly ME-1392DSample Probe Position Sensor ME-1306DStepper Motor Daughter Board ME-1148DPrinter ME-1296DElectrode Jack ME-0173D

    The CPU Board is supplied with the software installed to operate theproLYTE.

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    14/41

    P R O L Y T E A N A L Y Z E R

    Troubleshooting

    2.a Troubleshooting Overview

    This section describes troubleshooting of the mechanical and electrical functions of theproLYTE analyzer.This section is based on the principle of replacing suspected faulty assemblies with new assemblies.Diamond Diagnostics recommends keeping an inventory of functioning test assemblies for

    troubleshooting the proLYTE. Refer to the Troubleshooting Guide in this sectionto determine the cause andcorrective action that must be taken to properly service theproLYTE analyzer.

    Chapter

    2

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    15/41

    P R O L Y T E A N A L Y Z E R

    2.b Troubleshooting Guide

    Refer to the chart below to help determine the problem and corrective action needed.

    ProblemDescription

    ProblemSource

    Corrective Action Steps

    Blank Display Power Malfunction orDefective Display

    AC Power VerificationDC Power VerificationDisplay Verification

    No Responseto YES/NObuttons

    Power Malfunction orDefective DisplayConnection

    Display AssemblyVerificationDC Power Verification

    Blank Display,Beeping

    FaultyEEPROM

    Installation orPower Malfunction

    Inspect EPROMInstallation

    DC Power Verification

    PRB Errors Probe MotorMalfunction

    Probe Motor Verification

    PMP Errors or NoPump Movement

    Pump MotorMalfunction

    Perform PUMP CAL testIn DIAGNOSTICS?Menu Pump MotorVerification

    Printer DoesNot Print

    PrinterMalfunction

    Printer Verification

    Sample

    Detector TestFailure Result255 Or 000

    Sample

    Detector/CircuitMalfunction

    Sample Detector

    Verification

    Bad ElectrodeVoltages

    Electrodes or SignalInput Malfunction

    Check Electrode InstallationAnd PerformanceSignal Input Verification

    Serial TransmissionFailure

    Serial Cable orTransmissionProtocol

    Serial CommunicationVerification

    12 Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    16/41

    P R O L Y T E A N A L Y Z E R

    3

    2.c DC Power Verification

    1. Remove the rear housing. See Rear Housing Removalin the REPAIR SECTIONfordetails.

    2. Connect the power cord first to an outlet, then to the entry module.

    3. Connect the reference side of the voltage meter to TP12 on the CPU board.

    4. Verify the voltage at each test point listed below.

    Test Point Value Tolerance

    TP1 5.00 V +/- 0.15 V

    TP2 36.0 V +/- 2.0 VTP3 5.00 V +/- 0.15 V

    TP10 2.50 V +/- 0.10 V

    TP11 -5.00 V +/- 0.15 V

    5. If any of the voltage values in STEP 4are not within specification, AND the ACpower verification is successful, replace the CPU board.

    2.d Display Verification

    1. Disconnect the display cable from the CPU board. See Display AssemblyRemovalin theREPAIR SECTIONfor details.

    2. Connect a known good display assembly to the CPU board and connect the mainpower cord. Verify that the test display is operative. Based on the test results, take theappropriate corrective action.

    -Display Blank and/or YES/NO buttons DO NOT operate:Remove the test display cable. Perform AC and DC powerverification. If the AC and DC power tests are successful, replace the

    CPU board with a new CPU board, and return to normal operation.See CPU Board Removalin the REPAIR SECTION for details.

    -Display On and YES/NO buttons function:Remove the original display assembly and replace with anew display assembly.See Display Assemblyin the REPAIR SECTION for details.

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    17/41

    P R O L Y T E A N A L Y Z E R

    4

    2.e Probe Motor Verification

    1. Disconnect the power cord, wait ten seconds, then reinstall the power. The probearm should move down and back up. If there are no error messages, resume normaloperation. If a PRB error occurs, proceed to STEP 2.

    2. Remove the sample probe from the probe arm. For details refer to proLYTEOperators Manual. Remove and reinstall the power cord. The probe arm should movedown and back up. If there are no error messages examine and if necessary replacethe sample probe, solutions valve and/or probe wiper. If a PRB error occurs,proceed to STEP 3.

    3. If the probe arm does not move at all perform the following operations.

    A. Visually inspect that the actuator and retaining ring are correctly installed on theprobe motor lead screw. If the installation is incorrect, refer to CPU Board

    Removal and Probe Motor Installationin theREPAIR SECTION

    for details.

    B. Check that the probe motor wires are completely seated inside the 4-pinconnector as follows:

    - Remove the probe motor connector from the bottom left ofthe CPU Board (position P3).

    - Compress the strain-relief cap against the probe motorconnector to firmly seat the wires.

    - Reinstall the probe motor connector into the CPU Board.

    C. Repeat STEP 1. If no errors occur, resume normal operation.

    4. If the probe arm moves and stops anywhere except the top position, or if the probearm vibrates against the top of the proLYTE housing, replace the probe positionsensor. To replace the sensor, first remove the CPU Board and the probe positionlocator. Refer to Probe MotorAssembly Removalin the REPAIR SECTIONfor details.

    5. If the probe does not move at all, check for a short circuit on the CPU Boardbetween TP12 and TP2. If there is a short circuit, replace the Stepper driver daughterboard at positions U35 and U36. See Front and Rear Assembly Componentsin the ProductDescription Section.

    6. If no short circuit is found, replace the probe motor assembly. Refer to the ProbeMotor Assemblyin the REPAIR SECTIONfor details.

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    18/41

    P R O L Y T E A N A L Y Z E R

    5

    2.f Printer Verification

    1. Press the printer paper advance button. If the printer head moves proceed toSTEP 2. If the printer head does not move, proceed to STEP 3.

    2. Verify that the printer is on. Turn the printer on by pressing NO to thePRINTER OFF? question in the USER OPTIONSsection.

    3. If the printer is still not working, remove the power cord from the analyzer andwait for ten seconds, then install the power cord back into the entry module. Theprinter head should move across the paper and the paper should advance forwardprinting proLYTE.

    4. If the printer is still not working, remove the printer cover and check for anypaper obstruction in the printer head area. Remove any obstructions if possible,without damaging the printer assembly or printer window. If an obstruction is

    present and cannot be removed, replace the Printer Assembly. Refer to PrinterRemoval and Installationin the REPAIR SECTIONfor details.

    5. If there are no obstructions, but the printer does not work, perform the followingsteps.

    a. Remove the rear housing as detailed in the REPAIR SECTION.b. Remove the printer cable at the bottom right of the CPU board.c. Compress the strain-relief cap to firmly seat the printer cable.d. Reinstall the printer cable into the CPU Board.

    6. If the printer still does not print, connect a test printer assembly to the CPUBoard. If the test printer works, install a new printer assembly. If the test printerdoes not work, replace the CPU board. Refer to CPU Board Installation in theREPAIR SECTIONfor details.

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    19/41

    P R O L Y T E A N A L Y Z E R

    6

    2.g Sample Detector Verification

    1. Remove the electrode housing from the analyzer. See the proLYTE Operators Manualforfurther details.

    2. Remove the sample detector from the front housing.3. Install the sample detector into the sample detector receptacle. Do not install the electrode housing or pump tubing at this time.

    4. Proceed to the DIAGNOSTICS?menu and perform the SAMPLE DETECTORTEST?.

    -With the sample detector installed, the value on the display screen should be lessthan 200.

    -With the test in progress, remove the sample detector. The display should rapidlychange to 255.

    If the sample detector test meets these specifications, resume normal operation.

    A value of 255 with the sample detector installedindicates a bad connection. Proceedto STEP5 to check the sample detector circuit.

    5. Reinstall the sample detector. Look at the left side of the sample detector. A red lightmust shine from the LED hole. If there is no red light, install a known good sampledetector. If there is still no red light, replace the Display Assembly. Refer to theREPAIR SECTIONfor information.

    6. If the TEST sample detector red light is on, check the four connecting pins on theORIGINAL sample detector. The pins must face straight out forming a rectangle. Ifthe pins are bent against the black plastic body, straighten the pins using a small flathead screwdriver.

    7. Reinstall the original sample detector and repeat STEP 4.-If the red light comes ON: Reassemble the proLYTEand perform a SOLUTIONSPRIME and a SAMPLE DET. TEST. Resume normal operation.

    -If the red light remains OFF: Install a new sample detector.

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    20/41

    P R O L Y T E A N A L Y Z E R

    17

    2.h Serial Communication Verification

    1. Connect the proLYTE analyzer to a device with a known working serial port. Usea known good proLYTE serial cable, part no. ME-1334D. Make cross

    connections as required (e.g. 2 to 3; 4 to 5).

    2. Refer to the following serial interface specifications.

    Serial Interface Specifications

    Serial Cable Settings

    Connector 25-pin, D-Sub Male

    Pin Type RS-232

    Pin AssignmentsPin 1 Chassis, Not Connected

    Pin 2 T x D

    Pin 3 R x D

    Pin 4 RTS

    Pin 5 CTS

    Pin 7 Signal Ground

    RS-232 Protocols

    Transmission Rate 1200 baud

    Bit Configuration 8 data bits, 1 stop

    Bit Parity None

    Control Codes None

    All lines end with Carriagereturn line feed (CRLF)

    All transmission is ASCII string

    3. When correctly linked theproLYTEwill transmit to the external device all of thedata that is printed by theproLYTEprinter, in the same format.

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    21/41

    P R O L Y T E A N A L Y Z E R

    Repair

    3.a Repair Overview

    Always consult theTROUBLESHOOTING SECTIONfirst to determine the problem beforeattempting to repair the analyzer.

    Always remove the solutions pack and the electrode housing from theanalyzer before beginning any repair work. Refer to theproLYTEOperators Manual for detailed instructions.

    Follow the assembly and removal instructions in the order outlined in this section unless otherwisenoted. Always unplug the power cord from the analyzer before beginning any repair work.

    Note that only complete mechanical and electronic assemblies can be replaced. Refer to the ReplacementListin the PRODUCT DESCRIPTION SECTIONwhich lists replaceable assemblies as well asconsumable items. Consult the list to determine the appropriate replacement item and part number.

    3.b Recommended Tools

    The following is a list of the tools required to perform repairs on theproLYTE analyzer.Please refer to this list to determine the appropriate tool and size needed.

    0.050 Hex Driver5/16 Hex Driver3/16 Hex DriverPhillips Screw Driver, One SmallPhillips Screw Driver, One LargeAllen Wrench, 1/16 inchScissorsHex Nut Driver, 11/32 inch, electrode jack installation and removal toolVolt Meter

    Chapter

    3

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    22/41

    P R O L Y T E A N A L Y Z E R

    3.c Rear Housing Removal

    1. Disconnect the power cord from the entry module, see figurebelow.

    2. Detach the four screws from the rear housing using a Phillips head screw driver, asshown below.

    3. Separate the front and back housing. See FigureP5.

    3.d Rear Housing Installation

    1. Align and attach the front and back housing. See figurebelow. Verify that all cableconnectors are attached properly. Refer to figures in the PRODUCT DESCRIPTIONSECTIONfor detail. Verify that all wires are contained within the front housing

    before proceeding.

    2. Perform steps 1 and 2 from Rear Housing Removalin reverse .

    Figure P5 Rear HousingRemoval

    19Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    23/41

    P R O L Y T E A N A L Y Z E R

    2

    Figure P5 Rear Housing Removal (Cont)

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    24/41

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    25/41

    P R O L Y T E A N A L Y Z E R

    3.g Printer Removal

    1. Remove the printer lid from the printer.

    2. Remove the paper roll from the printer bay.

    3. Remove the rear housing.

    4. Detach the printer cable from the retaining clamps located along the right side of theCPU board.

    5. Detach the two screws using a Phillips head screw driver. Remove the printer body, as shown inFigure P7.

    6. Pull the printer cable up from behind the CPU board, while holding the printer body as shown in

    Figure P8.

    3.h Printer Installation

    1. Slide the printer cable behind the CPU board.

    2. Install the printer body into place. See Figure P8.

    3. Attach the two screws using a Phillips head screw driver. See Figure P7.

    4. Connect the printer cable to the CPU board.

    5. Using scissors, cut the beginning of the paper roll into a V shape. Insert the paper into the feederslot inside the printer bay, and press the paper feed button.

    6. Place the printer roll into the printer bay.

    7. Place the printer lid onto the printer.

    8. Attach the printer cable to the retaining clamps located along the right side of the CPU board.

    22Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    26/41

    P R O L Y T E A N A L Y Z E R

    Printer Bay

    23

    Figure P7 Printer Removal/Installation

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    27/41

    P R O L Y T E A N A L Y Z E R

    PrinterConnection

    Figure P8 Printer Removal/Installation

    24 Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    28/41

    P R O L Y T E A N A L Y Z E R

    3.i CPU Board Removal

    1. Remove the display cable from the top of the CPU board shown in Figure P9.

    2. Disconnect the electrode jacks from behind the CPU board. See figure belowfor the location ofthe electrode jacks. To disconnect, gently pull electrode jack straight away from the connector pinof the CPU board.

    The connector pins are extremely delicate. Avoid pin breakage when removing the electrode jacks,gently pull the electrode jacks straight away from the pins. If a pin breaks the entire CPU board willrequire replacement.

    3. Disengage the CPU board from the four plastic mounting posts as shown.

    4. Slide the CPU board off the mounting posts. See Figure P9.

    3.j CPU Board Installation

    1. Align the CPU board with the plastic mounting posts. Snap the right side mounting posts of theCPU board into place.

    2. Connect the electrode jacks to the back of the CPU board. To connect, gently push the electrode

    jack straight into the connector pin. See figure below to determine the proper connector pinposition on the CPU board and the recommended installation order.

    Do not bend the connector pin when installing the electrode jack. If a pin breaks theentire CPU board will require replacement.

    3. Snap the left and center mounting posts of the CPU board into place.

    4. Attach the display cable to the CPU board. Push the display cable tabs to the front to secure cablein place. See Figure P9.

    5. Connect the probe motor, sensor, pump motor, power supply and printer cables to the CPUboard, see Figure P9.

    25Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    29/41

    P R O L Y T E A N A L Y Z E R

    Figure P9 CPU Board Removal/Installation

    26 Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    30/41

    P R O L Y T E A N A L Y Z E R

    3.k Display Removal

    1. Detach the two screws from the sample detector board using a Phillips head screw driver.

    2. Move and hold the display cable to expose the three display assembly screws. Detach the threescrews from the display assembly using a Phillips head screw driver. See Figure P10.

    3. Carefully remove the display cable from the LCD screen, holding both items together. See FigureP11.

    Notice the orientation of the LCD screen which is extremely important when installing the displayassembly. The display has soldered leads located on the bottom of the LCD screen.

    There is an orientation notch on the display adapter. The display mounts on the two posts below

    the window. When installing the display adapter, the orientation notch is located on the top insideedge of the adapter.

    3.l Display Installation

    1. Place the foam pad on the display board between the screen and mounting plate.

    2. Place the LCD screen on top of the foam pad. The orientation of the LCD screen is extremelyimportant, see Figure P11. The display has soldered leads located on the bottom of the LCDscreen.

    3. Position the front housing so that the front of the analyzer is facing down. Align the displayassembly and insert into place, as shown in Figure P11. Use mounting holds as a guide.

    4. Hold the display in place and attach the three screws using a Phillips head Screw driver. See FigureP10.

    5. Align the sample detector board so the sample detector cable is positioned on the right side. Attachthe two screws lightly, using a Phillips head screw driver, see Figure P10.

    6. Install the sample detector in the front of the analyzer to verify the sample detector boardalignment. When properly aligned fully tighten the sample detector screws.

    27 Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    31/41

    P R O L Y T E A N A L Y Z E R

    Figure P10 Display Removal/Installation

    28Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    32/41

    P R O L Y T E A N A L Y Z E R

    Figure P11 Display Removal/Installation

    29 Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    33/41

    P R O L Y T E A N A L Y Z E R

    3.m Pump Motor Removal

    1. Remove the pump tubing from the front of the analyzer.

    2. Detach the four screws from the front of the analyzer, using a Phillips head screw driver or 1.5Hex as applicable. See Figure P12.

    3. Remove the pump motor assembly, as shown in Figure P13.

    3.n Pump Motor Installation

    1. Install the pump motor assembly through the front housing, as shown in Figure P13. The pumpmotor wires are positioned below the pump when properly aligned.

    2. Attach the four screws into the front of the analyzer, using a Phillips head screw driver. See FigureP12.

    3. Using a tie wrap cable, secure the loose wires under the pump motor to the lead screw/probemotor wires and the sensor wires.

    4. Install the pump tubing.

    3 Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    34/41

    P R O L Y T E A N A L Y Z E R

    Figure P12 DisplayRemoval/Installation

    Figure P13 DisplayRemoval/Installation

    31Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    35/41

    P R O L Y T E A N A L Y Z E R

    3.o Probe Motor Assembly Removal

    1. Detach the screw at the top of the probe position locator, using a Phillips head screw driver, as

    shown in figure below. Remove the washer.

    2. Remove the probe position locator.

    3. Remove the retaining ring from the actuator, using the retaining ring pliers. The retaining ring mustbe replaced once removed. DO NOT stretch out the retaining ring with the pliers. Disengage theactuator from the top of the lead screw assembly.

    Save the retaining ring.

    4. Detach the three screws from the probe motor bracket, using a Phillips head screw driver, as

    shown in Figure P14.

    5. Remove the probe motor/lead screw assembly.

    3.p Probe Motor Assembly Installation

    1. Place the probe motor assembly into place. See Figure P14.

    2. Attach the three washers and screws to the probe motor assembly bracket, using a Phillips head

    screw driver. See Figure P14.

    The other two screws are located on the bottom edge of the bracket.

    3. Insert the actuator through the top of the lead screw assembly. Install the retaining ring onto theactuator using the retaining ring pliers. DO NOT stretch out the retaining ring with the pliers.

    4. Place the probe motor cable wires into the retaining clamp. Secure the wires to the sensor cablewires with a tie wrap cable.The optimum location for the tie wrap cable is slightly to the left of the retaining clamp.

    5. Install the solutions valve.

    6. Remove the black thumb grip from the solutions valve. Install the sample probe as normal. Referto Figure P14 for details.

    A. With the probe hole facing to the left, toward the inside body of theproLYTE,align the probe collar with the notch on the top of the probe arm. Whenaligned, snap the probe into the probe arm.

    32 Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    36/41

    P R O L Y T E A N A L Y Z E R

    B. Manually move the probe arm until the sample probe hole is aligned with thetop chamber on the solutions valve.

    When the sample probe is aligned, DO NOT move the probe arm until theprobe motor installation is complete.

    7 Place the locator into position. The locator should be seated inside the probe position sensor andinside the mounting bracket slot.

    8. Attach and slightly tighten the washer and screw at the top of the position locator. Do not fullytighten the screw.

    9. Align the probe position locator slot within the center of the optical switch. Gently move thelocator to the left and visually confirm the alignment. Refer to Figure P2 for correct alignment.

    Confirm that the sample probe hole is aligned with the top chamber on the solutions valvebefore continuing.

    10.When alignment is complete, fully tighten the probe position screw into place using a Phillips headscrew driver.

    33Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    37/41

    P R O L Y T E A N A L Y Z E R

    Figure P14 Probe MotorRemoval/Installation

    34 Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    38/41

    P R O L Y T E A N A L Y Z E R

    3.q Guide Rod Removal

    1. Remove the sample probe from the probe arm and the solutions valve. Remove the solutions

    valve.

    2. Detach the screw from the bottom of the guide rod, using a Phillips head screw driver. Remove thewasher.

    3. Remove the guide rod and probe arm assembly, down and away from the housing.

    3.r Guide Rod Installation

    1. Install the guide rod. To install, first insert the guide rod into the alignment hole located in the top

    wall of the front assembly.

    2. Attach the washer and screw to the bottom of the guide rod, using a Phillips head screw driver.

    35Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    39/41

    P R O L Y T E A N A L Y Z E R

    6

    Figure P15 Guide RodRemoval/Installation

    Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    40/41

    P R O L Y T E A N A L Y Z E R

    3.s Electrode Jack Removal

    1. Insert the electrode jack wires into the supplied 11/32 inch nut driver.

    2. Remove the hex nuts using the nut driver.

    3. Pull the electrode jacks from the front housing of the analyzer, as shown in Figure P16.

    3.t Electrode Jack Installation

    1. Push the electrode jacks through the holes in the front of the analyzer, see Figure P16.

    2. Insert the electrode jack wires into the 11/32 inch nut driver.

    3. Attach the hex nut using the nut driver and tighten.

    37 Back to Table of Contents

  • 8/14/2019 SOP05-5002F Service Manual Prolyte

    41/41

    P R O L Y T E A N A L Y Z E R