Name: Farhana Arshad

Roll no: 1215

Page 2: Spectronic 21

“SPECTRONIC 21”INSTRUMENTATION,WORKING & PRINCIPLE

Page 3: Spectronic 21

Instrumentation:

Page 4: Spectronic 21

Instrumentation:

Internal components areLight sourceMonochromatorDispersing devicesFocusing devices detectors

Page 5: Spectronic 21

Light source:The function of the light source is to provide a sufficient light which is suitable for marking a measurement.

Tungsten lamp:The tungsten lamp is commonly used as a source of visible light.Its wavelength range is 330-1000 nm.It has a long life about 1200h.

Page 6: Spectronic 21

Monochromator:Accepts polychromatic input light from a lamp and outputs monochromatic light

All monochromators contain the following component parts;• An entrance slit• A collimating lens• A dispersing device • A focusing lens• An exit slit

Page 7: Spectronic 21

Radiation enters the monochromator through entrance slit. The beam is collimated and then strikes the dispersing element at an angle. The beam is split into its component wavelength by the grating. By moving the dispersing element, radiation of only a particular wavelength

leaves the monochromator through exit slit.

Page 8: Spectronic 21

Dispersion devices:Dispersion devices causes a different wavelength of light to be dispersion at different angles monochromator used for function.Types of dispersion devices:• Prism• Diffraction gratings

Prism:Prism is used to isolate different wavelength.Prism may be made of glass and quartz.Glass prism used in visible range whereas quartz prism used in

ultraviolet spectrum.

Page 9: Spectronic 21

Diffraction Grating:• It is an optical component , which splits light into several beams

travelling in different directions.

Focusing devices:• Combinations of lenses, slits and mirrors.• Relay and focus light through the instrument.

mirrors Type of rays

Mirror material

X-rays ultraviolet aluminum

visible aluminum

Page 10: Spectronic 21

Lenses

Cuvette:Designed to hold sample for spectroscopic experiment.Made of plastic, glass or optical grade quartz. Glass cuvettes are only suitable for the visible region, whereas quartz may be used in both the UV and visible region.

rays material

Ultraviolet Fused silica, sapphire

visible glass

Page 11: Spectronic 21

Detectors:Convert radiant energy into an electrical signal.• The photocell and phototube are the simplest detectors, producing current

proportional to the intensity of light striking them.

Photomultiplier tube detector:It consist of Photo emissive cathode Several dynodesAn anode

Page 12: Spectronic 21

1. A photon of radiation entering the tube strikes the cathode, causing emission of several electrons.

2. These electrons are accelerated towards the first dynodes causing emission of several electrons for each incident electrons.

3. These electrons are then accelerated toward the 2nd dynode, to produce more electrons which are accelerated toward 3rd dynodes and so on.

4. Eventually the electrons are collected at the anode.5. The resulting current is amplified and measured.

Page 13: Spectronic 21
Page 14: Spectronic 21

Principle: Beer lambert law:Discovery:Beer’s law was independently discovered by Pierre Bouguer in 1729, Johann heinrich lambert in 1760 and August Beer in 1852.Background:Many compound absorb light. Beer’s law provide a scientist tool to determine a solution concentration based on the level of light absorbed. The more light a solution absorb ,the higher the solution concentration. For this to work effectively the solution must be clear meaning there can not be any precipitate in the liquid.

Page 15: Spectronic 21

Statement:“The amount of intensity, when passing through any sample decreases exponentially with increase in thickness of the sample and the concentration of the medium. This phenomenon of decrease in intensity with medium and concentration is called Lambert-Beer law”.

Page 16: Spectronic 21

Mathematical form: A c∝ A l∝ A c * l∝ A = ε c l ε = molar absorptivity l = the sample “thickness” c = the molar concentration of the solution

Page 17: Spectronic 21

Transmittance:• The ratio of the intensity of the transmitted light (I) to the intensity of the incident light (Io) is called transmittance (T) .

T = I / IoBecause the intensity of the transmitted light (I) is never greater than the intensity of the incident light (Io), transmittance (T) is always less than 1.

Percent transmittance: Multiplies T by 100 to obtain the percent transmittance (%T), which ranges from 0 to 100%. 

%T = T * 100

Page 18: Spectronic 21

Absorbance: Absorbance is the amount of light absorbed by a sample.  It is calculated from T or %T using the following equations:

A = - log10 T  or   A = log10 (1/T) A = 2 - log10 %TRelationship between transmittance and absorbance: Transmittance and absorbance are inversely related.  That is, the more a particular wavelength of light is absorbed by a substance, the less it is transmitted.  Moreover, the inverse relationship between A and T is not linear, it is logarithmic. 

Page 19: Spectronic 21

• Two Sample Calculations of Absorbance from T and %T• Calculation #1:• if T = I / Io = 0.50• then %T = T * 100 = 50• and A = 2 - log10 %T = 2 - log10 50 = 2 - 1.69897 = 0.301• Calculation #2:• if T = I / Io = 0.999• then %T = T * 100 = 99.9• and A = 2 - log10 %T = 2 - log10 99.9 = 2 - 1.999 = 0.001

Page 20: Spectronic 21

Absorbance spectrum:• An absorbance spectrum is normally measured to determine the optimal

wavelength for measuring the absorbance of a given solution.

Page 21: Spectronic 21

Calibration graph:To determine the concentration of absorbing compound by construction of the calibration graph. Prepare a series of standards of known concentration of the absorbing compound ,measure

absorbance for each of them and plot the absorbance values against their concentration. measure absorbance of the “unknown” sample, its concentration can then be read from

calibration graph.

Page 22: Spectronic 21

WORKING:SET WAVELENGTHTo set the wavelength correctly, view the dial from directly above. Otherwise, reading of the dial may be wrong. Such 'parallax' errors occur if line of sight is NOT PERPENDICULAR to the face of the dial.

FILLING VOLUMEWhen pouring a liquid into the cuvette, the solution must fill the cuvette to a sufficient height so that the internal light beam passes through the solution in the cuvette, and not through air.The Spec 21 cuvettes have a horizontal index mark to show the minimum required filling volume.

Page 23: Spectronic 21

CLEAN WITH KIMWIPESWipe the cuvette first with a Kimwipe and then with a dry tissue.After cleaning the cuvettes, handle cuvette by their tops. Don't touch the lower portion of the glass.

AIR BUBBLESEven after cleaning the cuvette errors may still occur in a reading if air bubbles are present in the solution.Before reading a sample of even a blank, must REMOVING AIR BUBBLESRemoving air bubbles can be done by tapping the bottom of the cuvette to dislodge the bubbles.Remove all air bubbles.

Page 24: Spectronic 21

SAMPLE HOLDEROnce the sample or blank is free from bubbles and in a clean cuvette, it can be inserted into the sample holder.The sample holder is located on the left, top surface of the Spec 21. It is fitted with a cover, which must be closed before taking readings.

INSERTING A CUVETTEWhen inserting a cuvette into the sample chamber, GENTLY push the cuvette into its position. Hard pushing could damage the instrument.

Page 25: Spectronic 21

ADJUSTING THE BLANKThe blank solution is used to calibrate the instrument so that the internal light beam passes through the cuvette to the light-sensing device. This is indicated by a dial reading of 100% Transmittance or zero Absorbance.

REMOVE CUVETTE AFTER READINGAlways remove the cuvette from the sample holder as soon as the necessary adjustment or reading has been completed. Leaving the cuvette in the sample holder for an extended period can damage the light-sensing device.

Page 26: Spectronic 21

MATCHING CUVETTESTo reduce reading errors due to imperfections or scratches on the cuvettes, the sample and blank cuvettes should be matched to one another.This entails finding two cuvettes, which give exactly the same reading when containing the same solution.An alternative to using matched cuvettes is to read all solutions, both blanks and samples, in a single cuvette. This is cumbersome, but it assures accuracy.If the second cuvette reads within 1% of the blank cuvette, then they may be considered as matching.

Page 27: Spectronic 21

LEARNING TO READ THE METERThe meter simultaneously indicates Absorbance (the amount of light absorbed by the sample) on the lower scale and Percent Transmittance (the portion of light passing through the sample) on the top scale.

HOW TO READ THE METERThe top scale (%T), which is divided into increments of constant size, must be read from left to right. This is an easy scale to read.The bottom absorbance scale has increments between tick marks, which vary across the scale, making for some difficult readings. Also, the scale is read from right to left, the opposite of the %T scale.

Page 28: Spectronic 21

PRACTICE TAKING ABSORBANCE READINGS

Select one of the three concentrations of CuSO4 to work with.

The Spec 21 must first be adjusted to 100% T using the blank cuvette. Verify that the Spec 21 is set to 100% T when the blank cuvette is inserted properly into the chamber.

Now prepare the sample cuvette. Pour the water out of the second, matched cuvette, rinse it with the selected blue solution several times, fill it to above the mark, wipe the outside clean, remove any bubbles, gently insert it into the sample chamber, and close the cover.

Page 29: Spectronic 21

CHANGE THE WAVELENGTHChange the wavelength and read the values for Absorbance and %T again for the same sample of copper sulfate. Use the same cuvette to take the reading.However -- and this is important to remember -- every time the wavelength is changed, readjust the Spec 21 using the blank solution.

CHECK ABSORBANCE AT ALL WAVELENGTHSDetermine the absorbance and %T of the sample at all three suggested wavelengths, checking the accuracy of all the readings.