spintronic lateral flow biochip platform (extended abstract)

10
Spintronic Lateral Flow Biochip Platform (Extended Abstract) Jos´ e Alexandre de N´ obrega Ch´ ıcharo INESC MN Lisbon / Technical University of Lisbon [email protected] Abstract In recent years, the application of lateral flow immunoassays has seen a huge growth on the diagnostic industry. It is considered to be a highly promising class of point-of-care (POC) diagnostic tests and represents a multibillion-dollar market on its own. This development was motivated by the combination of unique properties of lateral flow, which include its simplicity, relative accuracy, fast results, low price, and user friendliness. This technology has been used not only in the academic context but also to support a wide range of commercial applications including glucose tests, pregnancy tests, monitoring and testing of cardiac panels in emergency conditions, infectious disease screening (including Flu and HIV), cancer marker screening, and drugs abuse testing. The work presented in the thesis aims at exploring a new approach for combining lateral flow test with magnetoresistive sensors, with particular focus on: i) designing more compact test platforms that are easier to use and handle; and ii) to further push the detection limit of existing approaches by lowering the mechanical and thermal drift produced by the movement of the lateral-flow strip. In order to achieve the platform with the properties indicated above, it was necessary to combine several components in a single orchestrated architecture. To pave the way for the development of such a platform this work: i) presents a comprehensive study on spin valve sensors; ii) provides an experimental evaluation of the applicability of two magnetic nanoparticles to achieve magnetic labelling employing a permanent magnet for the magnetization method; iii) presents experimental characterization of several nitrocellulose membranes usually employed in lateral flow tests; and finally, iv) proposes an initial design of a physical platform that allows one to demonstrate a way to combine all required elements of the envisioned platform. Experimental measurement using a locking amplifier in combination with a prototype platform that was designed to serve as proof-of-concept for the applicability of the envisioned platform have shown that this approach is very promising, and that the developed platform was able to effectively detect magnetic nanoparticles that were inside a nitrocellulose membrane. 1 Introduction The lateral flow technique allows the extraction of clinical results from the manipulation of small sample volumes (e.g., blood, urine). This enables the design of very small and portable testing units, enabling the miniaturization of complex laboratorial procedures [1, 2]. In a nutshell, lateral flow techniques consists on using nitrocellulose strips to serve as a conductor for the capillary flow of the sample. The antigen (protein, virus, bacteria...) of interest present in the sample can be labeled with a fluorescent label, color dye or magnetic particle in the conjugate pad (Figure 1). The sample flows in one direction in the nitrocellulose (NC) membrane through capillarity towards the absorbent pad. The sample will reach a test line and a positive control line (detection zone). The test line consists of previously immobilized antibodies in the NC-membrane specific to the antigens of interest. When passing through this line, the labeled antigens are captured while the remaining molecules and labels continue flowing along the strip. The control line usually captures the conjugated markers, which passed the test line to testify the activity of the detection conjugate, thus validating the assay. Although the existing lateral-flow tests are user-friendly and fast, most commercial applications (one exception is the glucose tests) only allow for a binary measurement (positive/negative result) besides possessing a lack of sensitivity (hardly go below μM range) [3]. Therefore, additional research efforts are required to design quantifiable tests based on lateral-flow techniques. One of the alternatives of colorimetric or fluorescence labeling is the use of magnetic labeling. The main advantage of this approach is that it potentially enables the performance of quantitative tests without the need of bulky detection equipment. In these analytical assays, magnetic susceptometers are typically used to measure the induction variation due to the magnetic 1

Upload: others

Post on 17-Jul-2022

12 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

Spintronic Lateral Flow Biochip Platform (Extended Abstract)

Jose Alexandre de Nobrega ChıcharoINESC MN Lisbon / Technical University of Lisbon

[email protected]

Abstract

In recent years, the application of lateral flow immunoassays has seen a huge growth on the diagnostic industry. Itis considered to be a highly promising class of point-of-care (POC) diagnostic tests and represents a multibillion-dollarmarket on its own. This development was motivated by the combination of unique properties of lateral flow, which includeits simplicity, relative accuracy, fast results, low price, and user friendliness. This technology has been used not only inthe academic context but also to support a wide range of commercial applications including glucose tests, pregnancy tests,monitoring and testing of cardiac panels in emergency conditions, infectious disease screening (including Flu and HIV),cancer marker screening, and drugs abuse testing.

The work presented in the thesis aims at exploring a new approach for combining lateral flow test with magnetoresistivesensors, with particular focus on: i) designing more compact test platforms that are easier to use and handle; and ii)to further push the detection limit of existing approaches by lowering the mechanical and thermal drift produced by themovement of the lateral-flow strip. In order to achieve the platform with the properties indicated above, it was necessary tocombine several components in a single orchestrated architecture. To pave the way for the development of such a platform thiswork: i) presents a comprehensive study on spin valve sensors; ii) provides an experimental evaluation of the applicabilityof two magnetic nanoparticles to achieve magnetic labelling employing a permanent magnet for the magnetization method;iii) presents experimental characterization of several nitrocellulose membranes usually employed in lateral flow tests; andfinally, iv) proposes an initial design of a physical platform that allows one to demonstrate a way to combine all requiredelements of the envisioned platform. Experimental measurement using a locking amplifier in combination with a prototypeplatform that was designed to serve as proof-of-concept for the applicability of the envisioned platform have shown that thisapproach is very promising, and that the developed platform was able to effectively detect magnetic nanoparticles that wereinside a nitrocellulose membrane.

1 Introduction

The lateral flow technique allows the extraction of clinical results from the manipulation of small sample volumes (e.g.,blood, urine). This enables the design of very small and portable testing units, enabling the miniaturization of complexlaboratorial procedures [1, 2]. In a nutshell, lateral flow techniques consists on using nitrocellulose strips to serve as aconductor for the capillary flow of the sample. The antigen (protein, virus, bacteria...) of interest present in the samplecan be labeled with a fluorescent label, color dye or magnetic particle in the conjugate pad (Figure 1). The sample flowsin one direction in the nitrocellulose (NC) membrane through capillarity towards the absorbent pad. The sample will reacha test line and a positive control line (detection zone). The test line consists of previously immobilized antibodies in theNC-membrane specific to the antigens of interest. When passing through this line, the labeled antigens are captured while theremaining molecules and labels continue flowing along the strip. The control line usually captures the conjugated markers,which passed the test line to testify the activity of the detection conjugate, thus validating the assay.

Although the existing lateral-flow tests are user-friendly and fast, most commercial applications (one exception is theglucose tests) only allow for a binary measurement (positive/negative result) besides possessing a lack of sensitivity (hardlygo below µM range) [3]. Therefore, additional research efforts are required to design quantifiable tests based on lateral-flowtechniques.

One of the alternatives of colorimetric or fluorescence labeling is the use of magnetic labeling. The main advantage of thisapproach is that it potentially enables the performance of quantitative tests without the need of bulky detection equipment.In these analytical assays, magnetic susceptometers are typically used to measure the induction variation due to the magnetic

1

Page 2: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

Absorbent)pad)

Conjugate)pad)

)))Control)line)

)))))Nitrocellulose)membrane)

Test)line)Sample)pad)

Sample)flow) )))))Figure)1:)Lateral'flow+test+membrane+composed+with+a+sample+pad,+conjugate+pad,+and+a+

absorbent+pad+[12].+The+sample+is+loaded+into+the+sample+pad+that+promotes+the+even+and+controlled+ distribu@on+ of+ sample+ onto+ the+ conjugate+ pad.+ In+ the+ conjugate+ pad+magne@c+par@cles+ immobilize+ to+ the+ analyte+ of+ interest+ (labeling).+ The+ sample+ then+ flows+ to+nitrocellulose+membrane+and+is+driven+through+capillarity+to+the+absorbent+pad.+The+labeled+analytes+will+bind+in+the+test+line+and+control+line,+represented+in+the+figure.++++

Figure 1. Schematic diagram of the lateral flow technique.

material [4, 5]. Although being able to quantify the amount of magnetic particles (MNPs) immobilized in the test line,magnetic susceptometers still show low sensitivities.

Considering this, the combination of lateral flow and magnetoresistive sensors is a promising approach to further advancethe state of the art in point-of-care tests. Previous research efforts have already tried to pave the way for developing suchsystems.

2 Theory Overview and Framework

The main contributions of the thesis are multi-disciplinary in nature, and due to this an extensive and comprehensivesurvey on relevant theory is presented in the thesis. In particular the reader can find the relevant contents in the thesis maindocument:

Spintronics Based on the GMR effect, Dieny et. al invented in 1991 at IBM a device called Spin Valve (SV) [6]. Typically,a standard spin valve is composed by two ferromagnetic (FM) layers: a free layer and a pinned layer. As mentionedpreviously, these layers are separated by a conductive spacer layer. Finally, SVs also present an antiferromagnetic(AFM) pinning layer. Figure ?? illustrates the typical layer composition of a Spin Valve. The main proprieties of theaforementioned layers are presented below:

Magnetic Particles As mentioned in the previous section, magnetic nanoparticles (MNPs) can be used as a labeling methodfor biomocules and cells. The detection system in this work aims at enabling not only the detection, but also therealtime quantification of MNPs that will serve as label for biomolecules. Superparamagnetic particles only possessa single magnetic domain bellow a critical size, where all magnetic spins in nanoparticles align unidirectionallly andits magnetic coercivity is zero [7]. These particles are non-magnetic when no external magnetic field is applied but dodevelop an average magnetic moment in an external magnetic field. Therefore, superparamagnetic particles (MNPs)were considered in the work of the thesis.

Lateral Flow Immunoassays LFTs rely on clever and sophisticated technology to build devices which are very simple,enabling them to be operated by almost anyone. The fundamental principle underlying lateral flow immunoassays isbased on the procedure of loading and handling a liquid sample along a single axis combined with a specific immuno-logical assay. In more detail, a sample flows through capillary flow in a microporous membrane in one direction (seeFigure 1). Along the course, analyte-specific antibodies, which are identified targets designated to produce a signal, la-bel the analyte which is free in the sample by forming a complex. Then, the sample with the labeled analyte of interestreaches a detection zone where capture antibodies, bound in the membrane, seize the labeled analyte. The function ofthe antibodies is to detect the analyte, and is analogous to an ELISA (Enzyme-Linked Immunoabsorbent Assay), withthe distinction that in LFTs the detection analyte-specific antibody is labeled with an element/particle that producesa signal, that can be chromogenic, fluorescent, or magnetic field signal. LFTs can operate as either competitive orsandwich assay formats. These formats are chosen considering the proprieties of the target analyte [8].

Additionally, we provide here a brief overview of some works found in the literature from both industry and academicresearch groups that are closer in nature to the goals of the work discussed in the thesis. The first work briefly exposedemploys a magnetic susceptometer, and the remaining employ MR sensors, such as SV.

2.1 Magnetic Assay Reader from Magna BioSciences

For the magnetic particles analytical assays, magnetic susceptometers are typically used to measure the induction variationdue to the magnetic material. An example of this application named Magnetic Assay Reader (MAR), is currently commer-

2

Page 3: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

cially available, and is produced by Magna BioSciences. This enables MAR to detect the presence of molecules with aprecision of 10 to 1000 times higher than conventional visual means [9]. According to the MAR s manual book [10], thesequantitative lateral tests have achieved equivalent, or superior, sensitivity when compared with ELISA tests. It is a compact,portable, simple, cost-effective and presents a result in seconds. A study based on this MICT platform is presented in [11],where a sensitivity and specificity are reported to be 93% and 98% respectively, for the detection of Taenia Solium, whichcauses taeniasis and neurocysticercosis in humans.

2.2 Research Group from Innovative Surface Technologies

A research group from Minneapolis, USA, have conducted research with the aim of improving the sensitivity of magneticquantification platforms [12]. They successfully demonstrated a equivalent range and sensitivity comparing their MR detec-tion platform to colorimetric ELISA assays. Considering the results of their research, they suggest the eventual ability ofsuch an assay to be much faster (15min instead of more than 4h), as well as portable, while achieving detection of 12 pg/mlfor protein detection of rabbit IgG and interferon-gama (Il-γ). The magnetization method of MNPs consisted in a Helmoltzcoil in-plane with the SV sensitive direction.

A similar work was also reported [13], where the same magnetization method as employed and the lateral flow test stripwas moved on top of the sensor at constant velocity of 100µm/s. Although quantification limits were not reported, adetection of a few pg/mL could potentially be achieved.

2.3 Research Group of INESC-MN and University of Zaragoza

In parallel, INESC-MN, in collaboration with INA located in University of Zaragoza, has proposed a detection ar-chitecture, based on the movement of the lateral-flow strip over the sensor, which showed sensitivities below the visiblelimit [12, 14]. Furthermore, biological detection and quantification of the pregnancy hormone has also been demonstratedusing this approach [15]. The method is based on a discrete scan of lateral flow strips, of 70µm steps, and a precise controlof the contact force between sensor and LFT membrane. The membrane was placed parallel and above the sensor chip,however, and in contrast with the approach of the Minneapolis setup, the magnetic particles were magnetized out-of-plane.

2.4 Research Group from Chungnam National University

Another research group conducted a study for devising a magnetic detection systems for real-time LFT for Cardiac Tro-ponin I (cTnI). The real time monitoring of the sensor voltage for the concentrations of 0.12, 0.34, and 0.42nM of cTnIallowed to obtain a faster result within 1min of the loading of the sample. This system supports quantitatively monitoring ofaccumulated labels, above the test line, the authors employ a hybrid AMR/PHR ring sensor [16]. The sensor is mechanicallypositioned over the test zone by cutting the plastic cartridge edge at one side.

In the following sections we identify the focus of the thesis in each of the studied components and present a few relevant ex-perimental results. The interested reader is referred to the main thesis document for additional information and experimentalresults.

3 Magnetoresistive Sensors

Sputtering systems are used to deposit thin films. Sputter deposition is a type of physical vapor deposition (PVD), accom-plished by the condensation of a vaporized form of the desired film material onto a substrate, such as a silicon wafer. Severalsputtering equipment (IBD 3000, UHVII, Nordiko 7000) were employed to build the designed SV biochip composed of anarray of 20 SVs (see Figure 2(a) and fig:SVmask). Patterning techniques such as etching (by ion milling beam in IBD3600)and liftoff techniques were performed in the microfabrication of the final biochip, using the shape manipulation of a film ofphotoresist. Figure 2(c) depicts a developed SV sensor. The MR curve of the patterned SV is depicted in Figure ??. For a totalof 65 measured sensors, a good MR ratio was obtained (7.68±0.41 %), low coercive field (Hc = 0.34±0.14Oe) and a goodlinearity in the region centered at the zero applied field. An important propriety for the measurement of MNPs is the fact thatthe linear region is centered at the zero field (Hf = 0.83±0.82Oe). The resistance at zero fieldR0 was of 140.5±4.6 Ω, andthe sensitivity and normalized sensitivity was estimated to be 0.205± 0.066 Ω/Oe and 0.194± 0.018 %/Oe, respectively.

The noise characterization and magnetic detection limit was performed using the methodology used in [17, 18]. For thispurpose, the developed SV submitted to evaluation. Figure 3(a) and 3(b) depict, respectively, measured noise spectral densityand calculated magnetic detection limit of different applied bias currents from 50 Hz to 10 kHz. At a frequency of operationof 10 kHz, and for for bias current of 1mA, the voltage spectral noise density obtained was 3.58nV/

√Hz and the minimum

mangnetic field detectivity obtained was of 1.66× 10−5Oe/√Hz.

3

Page 4: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

(a) (b)

W"

L"

(c) .

Figure 2. a) Spin valve biochip mask composed of an array of 20 SVs in red displayed horizontally in the centre of the chip. Thecontact leads (two per SV) are presented in blue lines. b) Each SV was design in Yoke shape with width (W) of 3µm and length (L)of 10µm. c) Microfabricated Spin valve. .

102 103 104100

101

102

103

Frequency (Hz)

Spec

tral V

olta

ge S

V (nV/

Hz0.

5 )

Ibias= 5.0 mAIbias= 1.4 mAIbias= 1.2 mAIbias= 1.0 mAIbias= 0.8 mAIbias= 0.6 mA

Theoretical Thermal Noise Level

(a)

102 103 10410−5

10−4

Frequency (Hz)

Fiel

d D

etec

tivity

Hm

in (O

e/H

z0.5 )

Ibias= 5.0 mAIbias= 1.4 mAIbias= 1.2 mAIbias= 1.0 mAIbias= 0.8 mAIbias= 0.6 mA

(b)

Figure 3. a) Measured spectral noise density of the SV sensor. b) Field detectivity of the SV sensor.

4 Magnetic Labeling

The SV sensor, described in the previous section, is sensitive to external magnetic fields along one direction (the directionof the width of the sensor), and thus changing the SV resistance. Assuming this fact, we can evaluate the magnetic fieldof a MNP based only in this direction, which is sensed on the SV. Figure 4(b) depicts the magnetic field of a MNP on thisdirection (HX(x, 0, h)) defined with blue arrows in Figure 4(a), at a constant height (h) in z-direction relative to the sensorssensing layer. The HX(x, y, z) of a MNP vertically magnetized is given by Equation 1, where mP is the magnetic momentof the MNP in the z-direction, i.e., vertically magnetized. Eventually the interaction of the magnetic field of the MNP near aSV sensor will incite a change in its resistance proportional to the average magnetic field over the sensing area.

HX(x, y, z) = mPz3xz

(x2 + y2 + z2)5/210−3 [Oe] (1)

To better understand the effect of the detection of the MNP over the SV sensor, lets consider the simplified example of asingle MNP traveling over the center of the sensor (considering the SV length (L)). This scenario is illustrated in Figure 5.Initially (Figure 5(a)), the MNP will be distant from the sensor (represented in green), thus the magnetic field of the MNP(which intensity is denoted by the blue line) applied on the sensor is very small and negligible. Therefore, in this scenariothe sensor is unable to detect the particle. As the MNP moves towards the sensor (Figure 5(b)), the average magnetic field(< HX >) sensed over the sensor decreases. In this scenario, the < HX > sensed by the sensor is given by the integral ofthe magnetic field of the MNP divided by the sensor area (represented by the red area in the figure). Note that, as the MNPcontinues to move over the sensor, the average magnetic field sensed by the sensor changes. A particular case occurs whenthe MNP is over the center of the sensor (Figure 5(c)). In this case the average magnetic field sensed by the sensor is zero, asthe integral of the magnetic field (HX ) is zero, i.e., the sum of the red areas in the figure. Finally, as the MNP moves awayfrom the sensor, the magnetic field applied (HX ) on the sensor will became dominant on a positive intensity (Figure 5(d)).This happens until the distance between the MNP and the sensor becomes large enough such that the magnetic field of theMNP becomes negligible again, similar to Figure 5(a).

Numerical simulations of vertically magnetized MNPs, at an applied field of 1000 Oe, were conducted in order to evaluateand compare the magnetic field fringe of two types of MNPs (10 nm and 130 nm). These MNPs were measured in a VSMequipment and the magnetic moment of a particle was of 1.0184×10−16 emu, for 10 nm particles, and 1.1375×10−13 emu,

4

Page 5: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

mp#

z"

x"

y"

Happlied#

h#

Hx#

r#

Flow"Direc.on"

(a)

−5 −4 −3 −2 −1 0 1 2 3 4 5−0.1

−0.08

−0.06

−0.04

−0.02

0

0.02

0.04

0.06

0.08

0.1

x (µ m)

HX (x

,0,h

) Oe

h = 1 µ mh = 2 µ mh = 4 µ m

(b)

<HX(x(t),h)>

t

a)

b)

c)

d)

(c)

<HX(x(t))>+

ΔR+

+++

++++

(d)

Figure 4. MNP magnetized vertically relative to the sensor surface. a) MNP with magnetic field lines, centered at the sensorsurface at a height (h) with magnetic moment (mP ) due to the applied field (Happlied). b) Numerical simulation of the magneticfield along the x-direction (HX(x, 0, 1µm)) of a 130 nm particle at different heights. c) Schematic representation of the averagemagnetic field sensed by the sensor relative to position of the MNP over time (x(t)).Red dots represent the instants related toFigure 5. ) Transducing the average magnetic field sensed by the SV will result in change in resistance over time.

mp#

h#

z" x"

y"

L#

W#

(a)

mp#

h#

z" x"

y"

(b)

mp#

h#

z" x"

y"

L#

W#

(c)

mp#

h#

z" x"

y"

L#

W#

(d)

Figure 5. Trajectory of a MNP over the sensor (in green) at a height (h) along the x-direction.

for the 130 nm particles.Figure 4(b) depicts the x-direction field (Hx(x, 0, h)), for different heights relative to the sensor sensing layer surface:

h = [ 1, 2, 4] µm for a 130 nm MNP. One can observe that despite the high magnetization of the MNP the magnetic fieldHx(x, 0, h) decreases drastically for increasing height, the same was observed for 10 nm MNP. In fact, Figure 6(b) depictsthe average magnetic field magnitude according to the height of a 10 nm and a 130nm MNP relative to the center of the sensorat a position of the highest signal. Figure 6(a) depicts the magnetic field along the x and y-direction (HX(x, y, 1µm)) of a130 nm particle. And finally, Figure 6(c) and 6(d) depict the simulated change in signal of a SV a 10 nm and a 130 nm MNP.

The interaction of the MNPs with the ring permanent magnet employed in the work of the thesis was also experimentallystudied and exposed.To better understand the interaction of both types of MNPs and the ring magnet, several experimentswere conducted, and showed the movement of MNPs inside a droplet of 6µL of MNPs with a concentration of 1.16 ×108 particles/ µL (130 nm) (see Figure 7(c)) and 2.31× 1011 particles/ µL (10 nm) (see Figure 7(d)). These were spottedat different locations, considering the plane at half the height of the ring magnet as depicted in Figure 7(b). This included adeveloped platform of PMMA which is depicted in Figure 7(a), were a PMMA support platform was developed in the Micro

(a)

0 20 40 60 80 100 120 140 160 180

10−14

10−12

10−10

10−8

10−6

10−4

10−2

Height (h) of the particle relative to the sensor ( µ m)

<HX(1

.5µ

m,0

, h)>

Oe

10 nm particle130 nm particle

h = 1 µ m<HX>=5.241x10−6 Oe

h = 1 µ m<HX>=5.854x10−3 Oe

(b)

−10 −8 −6 −4 −2 −0 2 4 6 8 10−6

−4

−2

0

2

4

6 x 10−9

Position (x,y=0,1µ m) of a 10 nm particle relative to the sensor ( µ m)

Sign

al (V

)

Ibias = 5 mAIbias = 1 mA

(c)

−10 −8 −6 −4 −2 −0 2 4 6 8 10

−6

−4

−2

0

2

4

6

x 10−6

Position (x,y=0,1µ m) of a 130 nm particle relative to the sensor ( µ m)

Sign

al (V

)

Ibias = 5 mAIbias = 1 mA

(d)

Figure 6. a) Magnetic field along the x and y-direction (HX(x, y, 1µm)) of a 130 nm particle. b) Numerical simulation of theaverage magnetic field magnitude according to the height of a 10 nm and a 130nm MNP relative to the center of the sensor at aposition of 1.5µm. And simulated change in signal of a SV, considering a sensitivity of 0.221 Ω/Oe and two bias currents, relativeto the position of a MNP to the center of the SV c) 10 nm and d) 130 nm MNP.

5

Page 6: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

24.6%mm%

35.4%mm%

8.5%mm%

(a) (b) (c) (d)

Figure 7. a) Dimensions of the ring magnet. At the center a magnitude of 1000 Oe was measured. b) Cross-view of the ringmagnet positioning the surface of the chip (green) with the sensors. The MNPs are being magnetized in this plane. c) Spot of three130nm particle droplets in three locations at the center of the magnet, after 1 minute. The MNPs in the middle droplet are attractedtowards the closest edge of the magnet. c) Same experience considering 10 nm MNP, after 35 minutes.

Milling Machine to hold the ring magnet and expose this plane correctly. Although 130 nm MNPs provide a higher magneticfringe field, and thus higher output signal of the SV, the 10 nm MNPs seem to be the most appropriate for this detection setupas they are less attracted to ring magnet.

5 Lateral Flow Membrane Characterization

Several LF Hi-Flow Plus nitrocellulose membranes were purchased from Millipore, comprised of 6 similar membraneswith different pore sizes, which eventually modifies the capillary flow rates and kinetics of the liquid sample in the membrane.These were characterized by SEM imaging and were subjected to flow rate tests, to determine the position that the test andcontrol line should be placed for the designated membrane (3mm× 20mm). The six different membranes were consideredand the capillary flow rate was extrapolated from measurements of the position of the sample front over the time. This setupconsisted of placing the membrane next to a ruler and a stopwatch, and then a video camera recorder filmed as the sample wasloaded at one edge of the membrane until the flow of the sample stopped. The resulting experiments of 3 tests per differentpore sized membrane is depicted in Figure 8.

The MNP sample have a brownish/yellow color and when the sample was loaded, a clear front of the sample flows in onedirection to farthest side of the membrane, which exhibits a homogenous color along the flow (uniform dispersion of MNPs).Approximately 8 to 11 min after loading the sample a clear line of agglomerated MNPs gradually intensifies at the end offlow front. Contrary this fact, the flow of MNPs employing the ring magnet, showed that the 130 nm MNPs remain in thearea where the sample was loaded due to attraction exerted by the magnet. However, it is relevant to note that contrary toresults presented in the previous section, instead of MNPs concentrating on the edge closest to the magnet, most of themconcentrate at the rightmost side of the sample droplet (see Figure 9(a)). This phenomenon was not observed for the 10 nmMNPs (depicted in Figure 9(c)) but is explained by the fact that there is a drag force induced by the capillary flow of thesample in the membrane. A way to circumvent this problem, to ensure the MNPs flow through the membrane would beto increase the drag force exerted by the capillary flow. To maximize that drag force one should avoid the air to leave themembrane from anywhere else other than its rightmost edge. A recent paper reports a similar development of a method tocontrol the flow rate of fluids within paper-based microfluidic analytical devices [19]. To achieve this, a small PMMA piece

17.59& 17.14& 16.81& 16.18&15.49&

11.77&

0&

2&

4&

6&

8&

10&

12&

14&

16&

18&

20&

HF075& HF090& HF120& HF135& HF180& HF240&

Es#m

ated

)loca#o

n)for)the

)detec#o

n)zone

)(mm))

Membrane)

(a)

11.05%12.34%

14.60%

18.71%20.41%

24.64%

0%

5%

10%

15%

20%

25%

30%

HF075% HF090% HF120% HF135% HF180% HF240%Time%requ

ired%to%re

ach%a%flo

w%ra

te%of%1

cm/m

in%%(s)%

Membrane%

(b)

Figure 8. a) Estimation of the position of the detection zone for all membranes. b) Amount of time took for the sample front flowrate reach 1 cm/s.

6

Page 7: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

(a)

1.8"mm"

2.6"mm"

0.19"mm"

(b) (c) (d)

Figure 9. a) Experimental setup with the ring magnet depicting the MNPs trapped in the area where it was loaded. b) PMMApiece for enhancing the drag force of the liquid in the membrane. c) Experimental setup of 10 nm MNPs with the ring magnet usingthe same PMMA piece. c) Example of a NC membrane after the experimental setup of 130 nm MNPs.

was built in the Milling machine such that it covers most of the membrane surface, as well as its lateral edges (see Figure 9(b)and 9(c)).

It was experimental observated that the flux of samples labelled with smaller MNPs is much uniform (monodisperse) thanwhen using larger MNPs and the flow rate of MNP sample by employing a PMMA piece that limits the exit of air to theextremity of the membrane farther from the inlet. A clearer line is depicted at the end of most tests, as shown by Figure9(d),although for 130 nm MNP samples, a higher background occurred. This happens because smaller MNPs travel faster acrossthe membrane and are less attracted by the permanent magnet. This motivates the use of smaller MNPs to achieve morereliable and predictable tests.

6 Integration Platform

As the sensor surface must be at the center of the permanent ring magnet which only has a 2.5 cm internal diameter, wehad to design a custom made PCB composed of three individual components to allow the wirebonding to be made directlybetween the SV chip and the PCB. The thesis also discussed the detailed information of the developed platform whichincluded a PMMA support to hold the magnet was developed in the MicroMilling Machine.

For the signal acquisition of the voltage tension applied at the sensors terminals a lock-in amplifier (DSP 7265, Signal Re-covery, USA) and an AC voltage source is employed. A lock-in amplifier is a specialized AC voltmeter that uses synchronousdemodulation to measure signal strength or phase, even under severe noise conditions. The position of the magnet was fixedin all experiments resulting in a baseline voltage of 75.840mVrms and 153.300mVrms for Vpp values of 1Vpp and 2Vpprespectively. For each applied Vpp, the resulting bias currents (Ibias) were, respectively of, 0.806mApp and 1.612mApp.

Reported experiments were conducted using three different samples of: control sample, consisting only of phosphatebuffer solution (PB) without MNPs; a sample consisted of 10 nm MNPs in PB suspension with a concentration of 2.31 ×1011 particles/µL; and a sample consisted of 130 nm MNPs in PB suspension with a concentration of 1.16× 108 particles/µL.A volume of 6µL was employed.

The experimental procedures for were tested as follows:

Permanent&Ring&Magnet&

Spin&Valve&Sensor&

Sample&Droplet&

Magne>c&Nano&Par>cle&in&Suspension&

(a)

Permanent&Ring&Magnet&

Spin&Valve&Sensor&

Sample&Droplet&

Magne>c&Nano&Par>cle&in&Suspension&

(b)

0 20 40 60 80 100−80

−60

−40

−20

0

20

Time (s)

Vsi

gnal

− V

base

line (

µV)

10 nm sample droplet130 nm sample dropletPB sample droplet

6V130nm

6 V10nm!(µV r

ms)!

!

(c)

.4.2"

.84" .83.3"

.5"

.113.3"

.153.3".180"

.160"

.140"

.120"

.100"

.80"

.60"

.40"

.20"

0"

Signal'Cha

nge''

ΔV=

Vbaseline'/Vsignal'(m

V rms)''

"""""""""""""""Without"MNPs""""""""""""""10nm""""""""""""""""""""""130nm"

1Vpp""2Vpp""

3"measurements"

5"measurements"

3"measurements"

(d)

Figure 10. Schematic diagram of the experimental setup for a direct measure of a sample droplet. a) The SV sensor is shiftedtowards the magnet and the sample of MNPs is loaded on top of it. b) The MNPs are attracted to the closest edge of the ring magnet,and the output signal is acquired. c) a) Signal change (V output − Vbaseline) of two sample droplets (10nm MNPs in red, 130nmMNPs sample in blue, PB solution in green). d) Signal change ∆ V (denoted in Figure a)) of the output signals for both sampledroplets.

7

Page 8: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

Direct Measurement of a Sample droplet: The sample, in the form of a droplet, was placed directly on top the SV, thesignal was obtained for about 150 s. The goal of this experiment was to ascertain the capacity of measurement setup tosense MNPs.

Moving Membrane Over the Sensor: Samples of 10 nm and 130 nm MNPs were applied to HF135 membranes and al-lowed to flow and dry outside of the platform. The MNPs lines that formed were gently passed over the SV by directmanipulation of the membrane. The goal of this experiment was to further validate the ability of the SV to detect MNPsin NC membrane, by employing an approach similar to the one employed in previous works such as [12] and [14].

6.1 Direct Measurement of Sample droplet

Figure 10(a) and 10(b) depicts a schematic diagram of this experimental setup. After loading the sample on top of the SV,the MNPs tend to travel to the closest edge of the permanet magnet. This effect is transduce to a signal change of the SV, asthe MNPs travel to one of this side. Figure 10(c) depicts the typical SV response observed in terms of signal output voltageto the loading of a sample droplet over it. Presented results covers samples of PB, 10 nm MNPs, and 130 nm. Note that thevalues in the y-axis present the nominal change in signal, i.e., it disregards the baseline signal, only quantifying the absolutevariation in the signal. The first relevant aspect of these results was that contrary to samples that contained MNPs, the controlsample of PB did not resulted in a significant output variation of the signal. This leads us to believe that indeed the MNP isable to sense the MNPs.

6.2 Moving Membrane Over the Sensor

In these experiments, samples were loaded in HF135 NC membrane, and the sample was allowed to flow and then dryon the membrane, outside the platform. This leads to the emergence of a visual line close to the end of the NC membrane,containing a deposition of a high agglomerate of MNPs. After this, the membrane was cut such that this line was positionedexactly at its end. The membrane was then carefully passed over the sensor with the extremity containing the agglomerate ofMNPs in the direction of the SV.

Figure 11(a) shows the signal output when a membrane loaded with a sample of 10 nm MNPs is moved over the sensor,from the right to the left and then from the left to the right. One could expect that if the sensor was indeed able to sense themagnetic field of the MNPs agglomerated in a line on the membrane, the signal output should present a negative variation(when the MNPs are sensed in right side of the sensor) followed by a positive variation (when the MNPs are sensed on theleft side of the sensor). This effect is expected to be inverse when the membrane travels from the left to the right of the sensor.Indeed, the results show this to be true as the sensed signal first increases and then decreases, when the membrane crossesthe SV from right to right. Finally, the output signal drops and then rises as the membrane travels from the left to the right ofthe SV. This shows that the SV is able to detect the presence of the agglomerated MNPs in a dried membrane, as shown bythe works described in Section 2

Figure 11(b) presents summarized results for the maximum variation observed in the output signal of the SV for differentsamples. Similar to the results discussed in the previous section, we observe that the PB sample (without MNPs) did not resultin a significant variation of signal, validating the results obtained with the remaining samples. Surprisingly, the membranesin which the sample containing the 10 nm MNPs obtained the highest variation in the output signal. This can be explained by

0 1 2 3 4 5 61.528

1.529

1.53

1.531

1.532

1.533

1.534

1.535

1.536x 105

Time (s)

Sign

al o

uput

(µV)

!(µV r

ms)!

!

(a)

4"

130.6"

108.6"

0"

20"

40"

60"

80"

100"

120"

140"

160"

180"

Without"MNPs" 10"nm" 130"nm"

Signal'Cha

nge'|Δ

V|'between'the'maxim

um'

and'minim

um'peaks'of'the

'signal'(m

V rms)''

'

3"measurements"

17"measurements"

7"measurements"

(b)

Figure 11. a) Signal output of 10 nm sample in a HF135 membrane, considering two directions when the membrane is crossedthe sensor (the first curve corresponds from the right to the left and the second corresponds from the left to the right). b) Signalvariation for each peak considering both samples of 10 nm and 130 nm of MNPs.

8

Page 9: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

the fact that, the 130 nm MNPs sample is very diluted. Considering the magnetic moment at 1000 Oe of both single MNPs,the total magnetic moment of 6µL for each sample concentration was respectively, 1.411×10−4emu and 7.917×10−5 emufor 10 nm and 130 nm MNPs samples. Also, as shown in shown that the smaller MNPs can more easily flow in the membranetherefore, the total number of MNPs that are agglomerated at the extremity of the membrane is higher for these MNPs.

7 Conclusions and Future Work

This thesis studies and experimentally evaluates several components that are essential for devising a new approach forcombining lateral flow test with magnetoresistive sensors in such a way that more compact and easier to use and handle testplatforms can be devised while simultaneously pushing the detection limit of existing approaches.

In particular, the thesis focus on the following essential components:

Spin valve sensors Single SV sensors that could potentially fulfil the goals of the thesis were successfully designed, pro-duced, and characterized in the INESC-MN laboratory. The MR of the SV was shown to be high enough to enable theadequate transduction of variable magnetic fields into measurable variation in signal as to potentially enable the quan-tification of MNPs successfully. The developed SV are simple yet reliable devices with fast in the microfabricationprocesses, compared to other MR sensor, and served as a first approach towards the detection schemes.

Magnetic nanoparticles Two different types of MNPs that differed in size and magnetic moment were experimentally char-acterized as candidates to support the labeling of relevant agents in the samples. Experimental results using a perma-nent ring magnet to vertically magnetize the MNPs showed that the smaller particles (10 nm) were less attracted tothe magnet therefore could easily flow across the nitrocellulose membrane despite the attraction of the magnet, beingpotentially better candidates to support the envisioned platform. Both MNPs (10 nm and 130 nm) have been proved tobe detected using the SVs.

Nitrocellulose membranes Six different types of nitrocellulose membranes were experimentally characterized to study theircapillary flow rate. Using these experimental results we were able to determine the adequate positioning for thedetection zone in each membrane. These results are relevant as they can allow to select the most appropriate membranefor a particular lateral flow test device according to the dimensioning of the device.

Integration platform An integration platform, based on a preliminary design, that was discussed in the thesis, was devel-oped and proposed here with the goal of serving as an initial proof-of-concept to the envisioned platform. The designedand built platform has shown a successfully way to integrate all the components required by the envisioned platform,showing that by using a component to limit the areas of the membrane where air could leave the membrane, one couldachieve a more stable and complete flow of MNPs across the membrane. Experimental results with samples of 10 nmand 130 nm MNPs were extracted with this platform and a measurement setup using a lock-in amplifier. This resultshave clearly shown the detection of magnetic field of MNPS inside the NC membranes, while the membranes weremoved over the SV. These results are therefore highly supportive for the possibility of achieving a LFT using SVsensors as the one described in the thesis.

7.1 Future Work

The work that is presented in the thesis is currently being further developed with the goal of submitting a patent. Addi-tionally, in the future we plan to pursue the following research vectors:Integration with Biochip platform The developed integration platform was designed with the goal of integrating it with

an existing portable Biochip platform developed previously in INESC-MN [20]. This will allow to perform additionaltypes of tests in that platform which could be beneficial for the academic community that currently leverages thisplatform in their work, while at the same time simplifying the use of this particular type of test, as it able to detectedsimultaneously 32 sensors.

Study the applicability of different MR sensors Because the sensitivity of a single SV may not be enough to accuratelysense the magnetic field of smaller particles, it would be relevant to test different types of magnetoresistive sensors thatcould offer improved sensitivity. In particular, the use of magnetic tunnel junctions, add flux guides, use series of MRsensors, and a Wheatstone bridge chip design are all viable candidates to achieve this and should therefore, be subjectof study in the future.

Medical application to design a relevant assay As a final proof-of-concept of the applicability of the envisioned platformfor biomedical applications, one should develop a fully functional medical application. In particular, it would beinteresting to develop a lateral flow immunoassay for detecting cancer biomarkers present in human whole blood.There is an increasing research effort on the identification and validation of biomarkers relevant to specific cancersaiming for early detection of cancer (prognosis). One example, is the biomarker for prostate cancer (PSA).

9

Page 10: Spintronic Lateral Flow Biochip Platform (Extended Abstract)

Develop a functional mechanical and electrical prototype Such prototype could include means to bias the sensors, am-plify the signal and digitally acquire and process the signals. It could also include a display-screen where the results ofeach assay can be displayed. All the electronics should be battery powered to allow the prototype to be an independenthandheld (portable) reader point-of-care device.

References

[1] E. Fu, P. Kauffman, B. Lutz, and P. Yager, “Chemical signal amplification in two-dimensional paper networks,” Sensors and ActuatorsB: Chemical, vol. 149, no. 1, pp. 325 – 328, 2010.

[2] A. W. Martinez, S. T. Phillips, M. J. Butte, and G. M. Whitesides, “Patterned paper as a platform for inexpensive, low-volume,portable bioassays,” Angewandte Chemie International Edition, vol. 46, no. 8, pp. 1318–1320, 2007.

[3] C. D. Chin, V. Linder, and S. K. Sia, “Commercialization of microfluidic point-of-care diagnostic devices,” Lab on a Chip, 2012.

[4] P. I. Nikitin, P. M. Vetoshko, and T. I. Ksenevich, “Proceedings of the Sixth International Conference on the Scientific and ClinicalApplications of Magnetic Carriers, S,” Journal of Magn. Magn. Mater., no. 311, p. 445449, 2007.

[5] MagnaBioSciences, “Patents 6-046-585(4/4/00) and 6-275- 031(8/14/01) and 6-437-563(8/20/02) and 6-483-303(12/19/02),” 2004.

[6] B. Dieny, V. S. Speriosu, S. S. P. Parkin, B. A. Gurney, D. R. Wilhoit, and D. Mauri, “Giant magnetoresistive in soft ferromagneticmultilayers,” Phys. Rev., vol. B, no. 43, p. 1297, 1991.

[7] Y.-W. Jun, J.-W. Seo, and J. Cheon, “Nanoscaling laws of magnetic nanoparticles and their applicabilities in biomedical sciences,”Acc. Chem. Res., vol. 41, p. 179, Feb. 2008.

[8] G. A. Posthuma-Trumpie, J. Korf, and A. van Amerongen, “Lateral flow (immuno)assay: its strengths, weaknesses, opportunities andthreats. a literature survey,” Anal. Bioanal. Chem., 2009.

[9] MagnaBioSciences, “MICT Troponin I (Ref: 509B030).” http://www.magnabiosciences.com/pdf/Troponin_I_IFU_2010_07_GCMS.pdf, 2010.

[10] MagnaBioSciences, “MAR Assay Development System Operator’s Manual.” http://www.thm.de/ibva/images/stories/doku/MBS-MAR-ADS/magnabiosciences\%20mar\%20ads\%20user\%20maual\%201094-1001.pdf.

[11] S. Handali, M. Klarman, A. N. Gaspard, X. F. Dong, R. LaBorde, J. Noh, Y.-M. Lee, S. Rodriguez, A. E. Gonzalez, H. H. Garciaand,R. H. Gilman, V. C. W. Tsang, and P. P. Wilkins, “Development and evaluation of a magnetic immunochromatographic test to detecttaenia solium, which causes taeniasis and neurocysticercosis in humans,” Clinical and Vaccine Immunology, vol. 17, pp. 631–637,Apr. 2010.

[12] K. Taton, D. Johnson, P. Guire, E. Lange, and M. Tondra, “Lateral flow immunoassay using magnetoresistive sensors,” Journal ofMagn. Magn. Mater., pp. 1679 – 1682, Feb. 2009.

[13] Y. Ryu, Z. Jin, M. S. Kang, and H.-S. Kim, “Increase in the detection sensitivity of a lateral flow assay for a cardiac marker byoriented immobilization of antibody,” Biochip Journal, Sept. 2011.

[14] D. Serrate, J. M. D. Teresa, C. Marquina, J. Marzo, D. Saurel, F. A. Cardoso, S. Cardoso, P. P. Freitas, and M. R. Ibarra, “Quantitativebiomolecular sensing station based on magnetoresistive pattern arrays,” Biosens. and Bioelect., pp. 206–212, May 2012.

[15] C. Marquina, J. M. de Teresa, J. Marzo, F. Cardoso, D. Serrate, D. Saurel, S. Cardoso, P. P. Freitas, and M. R. Ibarra, “GMR sensorsand magnetic nanoparticles for immuno-chromatographicassays,” Journal of Magn. Magn. Mater., 2012.

[16] S. Oha, S. Anandakumara, C. Leeb, K. W. Kima, B. Lima, and C. Kima, “Analytes kinetics in lateral flow membrane analyzed bycTnI monitoring using magnetic method,” Sensors and Actuators: B Chemical, vol. 160, pp. 747 – 752, Dec. 2011.

[17] P. P. Freitas, R. Ferreira, S. Cardoso, and F. Cardoso, “Magnetoresistive sensors,” J. Phys.: Condens. Matter., pp. 1 – 21, Apr. 2007.

[18] A. Guedes, Hybrid Sensors for Ultra Low Magnetic Field Detection. PhD thesis, Feb. 2010.

[19] S. Jahanshahi-Anbuhi, P. Chavan, C. Sicard, V. Leung, S. M. Z. Hossain, R. Pelton, J. D. Brennan, and C. D. Filipe, “Creating fastflow channels in paper fluidic devices to control timing of sequential reactions,” Lab on a Chip, Oct. 2012.

[20] F. Cardoso, Design, Optimization, and Integration of Magnetoresistive Biochips. PhD thesis, 2011.

10