the technology and applications of protein biochip array ... · biochip array technology ... the...

15
Biochip Array Technology Katie Walker Biochip Array Product Specialist The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing for the simultaneous multi- analyte detection of a wide range of analytes, from a single undivided sample •81mm 2 Solid State Device •Assays contained in discrete reaction sites •Single patient sample per biochip •Multiple assays performed simultaneously

Upload: lamtruc

Post on 17-May-2018

232 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Biochip Array Technology

Katie Walker

Biochip Array Product Specialist

The Technology and Applications of Protein Biochip Array

Technology

Panels of tests, allowing for the

simultaneous multi-analyte detection of

a wide range of analytes, from a single undivided

sample

•81mm2 Solid State Device

•Assays contained in discrete reaction sites

•Single patient sample per biochip

•Multiple assays performed simultaneously

Page 2: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Development of Biochip Array

• 1992 – 2003• Over 185 Scientists involved• $140,000,000.00 Investment• 45 new patents• Three main stages in Biochip development

– Synthetic Chemistry– Nano Dispense– Detection technology

The Biochip• The crux of the technology is the Biochip itself

• Coated in a proprietary silanation process– Activation of the Biochip surface

• Ensures uniformity and reproducibility

– Modification of surface chemistry • Controlled anti-body binding in optimal orientation

Page 3: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Biochip Surface Chemistry Non -activated Surface

Activated Surface

Page 4: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Biochip Manufacturing

• All manufactured by Randox at our HQ in the U.K.

• 20 million chip per year capability• Fully automated clean-room facility

Nano Dispense of Antibody solution

Quality Control• Quality control is at the heart of all that Randox does.• Quality is ensured at every level of product

– Pre process• Quality components • Randox Antibodies

– In process• Manufactured to ISO 13485 • Ensures reproducibility

– Post Process• Internal QC to ensure Functionality, integrity and

stability

Quality Control

Page 5: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Immunoassay Formats

• Tried and trusted EIA• Sandwich immunoassay

– Cytokine and Growth Factor Array• Competitive immunoassay

– Drugs of Abuse Array• Horse Radish Peroxidase Label

Competitive Immunoassay

Y Y

SUBSTRATE

ANTIBODY

ANALYTEE

CONJUGATE

DETECTION

Signal Analysis

• Chemiluminescence• L.O.D. compared

– Fluorescence (ELISA) 2,000,000 zeptomoles– Radio Immunoassay 1,000,000 zeptomoles– Chemiluminescence 25,000 zeptomoles

» Tietz ‘Clinical Chemistry’

• Imaged with super cooled CCD camera• Quantified by custom software

Page 6: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Available Arrays• Drugs of Abuse Array 1• Cytokine and Growth

Factor Array 1• Free Thyroid Array• Total Thyroid Array• Fertility Array• Tumour Monitoring

Array• Tumour PSA Array• Adhesion Molecule

Array

• Cardiac Array• Growth Promoter Array• Anti-Microbial Array 1• Cardiovascular DNA

Array• Colorectal Cancer DNA

Array• Anti-Microbial Array 2*• Anti-Microbial Array 3*• SyntheticSteroidsArray*• Drugs of Abuse Array2*

Drugs of Abuse Array 1

• Amphetamine• Methamphetamine• Barbiturates• Benzodiazepine I• Benzodiazepine II• Cocaine Metabolite

• Methadone• Opiates• Phencyclidine• Cannabinoids• Creatinine (urine

application only)

Drugs of Abuse Array II

• Buprenorphine• Fentanyl• Ketamine• LSD• MDMA• Hydromorphone

• Methaqualone• Oxycodone I• Oxycodone II• Propoxyphene• Creatinine*

Drugs of Abuse Arrays

• Specific matrix optimised kits for– Urine (FDA cleared)– Urine Semi-Quant– Whole Blood Quant– Oral Fluid Quant

Page 7: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Amphetamine Class

50 ng/mL /50 ng/mL

25 ng/mL /50 ng/mL

1000 ng/mL /1000 ng/mL

Cut-off

d-Amphetamine / + Methamphetamine

d-Amphetamine / + Methamphetamine

d-Amphetamine / + Methamphetamine

Target

Oral FluidBloodUrine

36%0.4%MDMA

0.4%544%MDA

100%<0.2%+ Methamphetamine

1.0%100%d-Amphetamine

Methamphetamine assayAmphetamine assayCross reactivity (urine)

Barbiturates

5 ng/mL50 ng/mL200 ng/mLCut-off

PhenobarbitalPhenobarbitalPhenobarbitalTarget

Oral FluidBloodUrine

84%Amobarbital

183%Pentobarbital

172%Butalbital

40%Barbital

512%Secobarbital

100%Phenobarbital

Barbiturates assayCross reactivity (urine)

Benzodiazepines

1 ng/mL /1 ng/mL

50 ng/mL /50 ng/mL

200 ng/mL /200 ng/mL

Cut-off

Oxazepam /Lorazepam

Oxazepam /Lorazepam

Oxazepam /Lorazepam

Target

Oral FluidBloodUrine

Benzodiazepines

<0.1%512%Diazepam

1007%Lorazepam

<0.1%100%Flunitrazepam

1.2%241%Nitrazepam

2%250%Nordiazepam

<0.1%444%Temazepam

23.9%<0.1%Lorazepam glucuronide

68.7%8.4%Clonezapam

<0.1%1818%Alprazolam

2%100%Oxazepam

Lorazepam assayOxazepam assayCross reactivity (urine)

Page 8: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Cannabinoids

4 ng/mL10 ng/mL50 ng/mLCut-off

−9 THC11 nor −9 THC-COOH11 nor −9 THC-COOHTarget

Oral FluidBloodUrine

<5%11 hydroxy −8 THC

3%11 hydroxy −9 THC

49%11 nor −8 THC-COOH

<0.5%Cannabidiol

<0.5%Cannabinol

100%11 nor −9 THC-COOH

Cannabinoids assayCross reactivity (urine)

Methadone

5 ng/mL25 ng/mL300 ng/mLCut-off

MethadoneMethadoneMethadoneTarget

Oral FluidBloodUrine

<1.0%LAAM

<0.1%EMDP

<0.1%EDDP

100%Methadone

Methadone assayCross reactivity (urine)

Opiates

40 ng/mL25 ng/mL300 ng/mLCut-off

MorphineMorphineMorphineTarget

Oral FluidBloodUrine

17%Hydrocodone

27%Hydromorphone

13%Dihydrocodeine

67%Morphine-3-glucuronide

115%Codeine

100%Morphine

Opiates assayCross reactivity (urine)

Phencyclidine

10 ng/mL5 ng/mL25 ng/mLCut-off

PhencyclidinePhencyclidinePhencyclidineTarget

Oral FluidBloodUrine

90%TCP

100%Phencyclidine

Phencyclidine assayCross reactivity (urine)

Page 9: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Cocaine Metabolite

20 ng/mL50 ng/mL300 ng/mLCut-off

BenzoylecgonineBenzoylecgonineBenzoylecgonineTarget

Oral FluidBloodUrine

<0.1%Ecgonine

4.1%Cocaethylene

6.4%Cocaine

100%Benzoylecgonine

<0.1%Norcocaine

<3%Ecgonine methyl ester

Cocaine metabolite assayCross reactivity (urine)

Sensitivities and RangesAssay Type Urine Whole Blood

Sensitivity Assay Range Sensitivity Assay Range

Amphetamine 8.1ng/ml 0-2500ng/ml 0.4 ng/ml 0-125ng/ml

Methamphetamine 45.2ng/ml 0-2000ng/ml 6.25 ng/ml 0-250ng/ml

Cocaine Metabolite 2.3ng/ml 0-800ng/ml 0.01 ng/ml 0-250ng/ml

Methadone 0.6ng/ml 0-800ng/ml 0.13 ng/ml 0-125ng/ml

Cannabinoids 3.6ng/ml 0-150ng/ml 0.28 ng/ml 0-50ng/ml

Barbiturate 9.4ng/ml 0-800ng/ml 0.1 ng/ml 0-250ng/ml

Opiates 0.8ng/ml 0-3500ng/ml 0.12 ng/ml 0-125ng/ml

Benzodiazepine 1 0.6ng/ml 0-500ng/ml 0.04 ng/ml 0-250/ng/ml

Lorazepam 1.1ng/ml 0-500ng/ml 0.1 ng/ml 0-250/ngml

Phencyclidine (PCP) 1.9ng/ml 0-100ng/ml 0.16 ng/ml 0-50ng/ml

Drugs of Abuse Array IIAssay Specificity

Buprenorphine Buprenorphine

Fentanyl Norfentanyl & fentanyl

Hydromorphone Hydromorphone

Ketamine Norketamine & ketamine

LSD Nor-LSD & LSD

MDMA MDMA & MDEA

Methaqualone Free & 2’-hydroxymethaqualone, 3’-hydroxymethaqualone & 4’-hydroxymethaqualone

Oxycodone 1 Noroxycodone & oxycodone & hydrocodone

Oxycodone 2 Oxymorphone & oxycodone

Propoxyphene Norpropoxyphene & propoxyphene

Whole Blood Trials

• MEO USA trial using pre/post-mortem bloods.

• 33 samples x 10 assays = 330 tests completed.

• 1 x Investigator runs.

• Overall completion time = 1 hr 40 mins.• Includes:-

• Sample preparation (dilution ¼)• Calibrator Reconstitution• Controls

Page 10: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

METHAMPHETAMINE AMPHETAMINE BENZODIAZEPINES

+ - + - + -EVIDENCE + 1 0 EVIDENCE + 0 1 EVIDENCE + 7 0

- 0 32 - 0 30 - 0 23

% AGREEMENT 100.0 % AGREEMENT 96.8 % AGREEMENT 100.0

METHADONE OPIATES PHENCYCLIDINE

+ - + - + -EVIDENCE + 4 0 EVIDENCE + 3 0 EVIDENCE + 1 0

- 0 29 - 0 29 - 0 32

% AGREEMENT 100.0 % AGREEMENT 100.0 % AGREEMENT 100.0

BENZOYLECGONINE OVERALL

+ - + -EVIDENCE + 10 0 EVIDENCE + 26 1

- 0 22 - 0 198

% AGREEMENT 100.0 % AGREEMENT 99.6

GC-MS

GC-MS

GC-MS

GC-MS GC-MS

GC-MS GC-MS

GC-MS

Urine Evaluation

• FDA Trials

• To assess the overall performance of drugs of abuse screening on the evidence immunoassay technology

• To compare the evidence performance to an alternate immunoassay system

• To compare the evidence performance to results obtained by GC/MS

Urine Evaluation

• More than 1300 clinical urine samples covering the entire range of possible test results for all ten drugs of abuse

• Comparison carried out by independent clinical toxicology laboratory in Texas

• 10% of samples fell within ±25% of the assay cut-off concentrations

82.1

3.2

96.8 96.7

17.93.3

0

20

40

60

80

100

evidence with CEDIA evidence with GC/MS CEDIA with GC/MS

% AGREEMENT% DISAGREEMENT

% Agreement Total number of samples Discrepant samples

evidence with CEDIA 96.8 1264 40

evidence with GC/MS 96.7 305 10

CEDIA with GC/MS 82.1 234 42

Amphetamines

Page 11: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Methamphetamine

82.2

3.3

96.7 97.0

17.8

3.0

0

20

40

60

80

100

evidence with CEDIA evidence with GC/MS CEDIA with GC/MS

% AGREEMENT% DISAGREEMENT

% Agreement Total number of samples Discrepant samples

evidence with CEDIA 96.7 1249 40

evidence with GC/MS 97.0 305 9

CEDIA with GC/MS 82.2 219 39

Cannabinoids

81.8

2.1

97.993.9

18.2

6.1

0

20

40

60

80

100

evidence with CEDIA evidence with GC/MS CEDIA with GC/MS

% AGREEMENT% DISAGREEMENT

% Agreement Total number of samples Discrepant samples

evidence with CEDIA 97.9 1335 28

evidence with GC/MS 93.9 231 14

CEDIA with GC/MS 81.8 231 42

Opiates

60.0

7.4

92.6 89.1

40.0

10.9

0

20

40

60

80

100

evidence with CEDIA evidence with GC/MS CEDIA with GC/MS

% AGREEMENT% DISAGREEMENT

% Agreement Total number of samples Discrepant samples

evidence with CEDIA 92.6 1334 99

evidence with GC/MS 89.1 320 35

CEDIA with GC/MS 60.0 320 128

Overall Summary

Evidence with CEDIA 96.9

Evidence with GC/MS 90.5

CEDIA with GC/MS 83.4

Total number of tests 11,750

% Agreement

Page 12: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

User Defined Cut-off selection. Analysers and Software

Features

• Semi-automated• Based on Biochip Array Technology

– Multiple test sites per biochip• Fast Results

– 54 biochips ie. 540 Tests in 1 hour• Small sample volumes

– Urine – 3.5μl per test– Whole Blood - 6μl per test– Oral Fluid - 10μl per test

Page 13: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Features

• Quality of results– All Randox products are developed to ISO 13485,

diagnostic standards– Full trace-ability through manufacture– Antibody Specificity – Self sufficient antibody supply– Specific Methamphetamine and Lorazepam

assays

Features

• Cut-off selection– Cut-offs can be selected and reselected without

recalibration– Different cut-offs can be selected for the same

analyte during the same run, ie. Opiates sample A 25ng/mL, sample B 10ng/mL.

• Fully Quantitative Analysis– Identified as positive/negative – Concentration of compound also displayed

Features

• Software– Windows XP® based– Colour coding easy to follow– Multiple search formats– Easy export of results– Fully LIMS connectable– Extensive QC facilities

• Levy Jennings and Multi-point rules.

– Enhance trouble shooting

Page 14: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

Features

• Fully Automated– Walk away system

• High throughput– 130 samples per hour x no.of tests per chip

• Chain of custody Features– Sample access doors lock once sample is loaded

• Password Protected

Features

• Retrospective Testing– Although only invoiced for tests reported, all assays

are carried out simultaneously– If certain assays are not reported they can be recalled

at a later stage.

Page 15: The Technology and Applications of Protein Biochip Array ... · Biochip Array Technology ... The Technology and Applications of Protein Biochip Array Technology Panels of tests, allowing

References

Molloy RM, McConnell RI, Lamont JV, FitzGerald SP. Automation of biochip array technology for quality resultsClin Chem Lab Med; 43 (12):1303-1313

FitzGerald SP, Lamont JV, McConnell RI, Benchikh EO; Development of a high throughput automated analyser using biochip array technology; Clinical Chemistry; 51 (7) 1165-1176

www. .com