Page 1: Steam and Water Analyzer

STEAM & WATER ANALYSERS

Conductivity

pH

Dissolved oxygen

hydrazine

Silica

sodium

April 7, 2

023

1

Page 2: Steam and Water Analyzer

Any variety of analyzer will consist of these sections:

Sample acquisition and conditioning

Predetermined mixing with the reagents

Measurement (via electrochemical or optical method)

Processing of values (electronic unit-alarms, annunciations)

April 7, 2

023

2

BASIC ANALYZER

Page 3: Steam and Water Analyzer

ELECTROCHEMICAL

April 7, 2

023

3

•The mobile ions in the sample tend to move towards the electrodes.

•The ions get reduced at the cathode by chemical reaction and the necessary electrodes are generated at the other end thus leading to flow of electrons (or current) which could then be measured and calibrated in terms of ionic concentration (related by Nernst equation).

the ionization of water changes only with temperature, not with dilute salt concentration.

Page 4: Steam and Water Analyzer

EXPLANATION:nernst equation

E = E0 –(RT/nF) log aiwhere:

E = total potential (in millivolts) between two electrodes

E0 = standard potential of the ion

R = universal gas constant (in Joules/mol-Kelvin)

T = absolute temperature (in Kelvin)

n = charge of the ion

F = Faraday constant (in Coulombs/mol)

ai = activity of the ion

April 7, 2

023

4

Page 5: Steam and Water Analyzer

CONDUCTIVITY

The measure of the ability of a solution to carry an electrical current.

Importance

Conductivity is just a rough estimate to get a quick know-how of the process (irrespective of the alkalinity or acidity), so that any harmful ingress can be detected at the early.

April 7, 2

023

5

Page 6: Steam and Water Analyzer

FACTORS

The accuracy of conductivity measurements can be influenced by the following factors:

Polarisation Contamination Cable resistance Cable capacitance Temperature

April 7, 2

023

6

Page 7: Steam and Water Analyzer

POLARIZATION

Applying an electrical current to electrodes in solution may cause an accumulation of ionic species near the electrode surfaces and chemical reactions at the surfaces.

Thus, a polarisation resistance arises on the electrode surface, which may lead to erroneous results as it is a parasitic component to the solution resistance.

April 7, 2

023

7

Page 8: Steam and Water Analyzer

TEMPERATURE

Ct = C25[1+ a (t-25)]

Where Ct –conductivity at temperature t

a- temp coeff

t- process temp.

April 7, 2

023

8

Page 9: Steam and Water Analyzer

CATION CONDUCTIVITY

Importance

Contamination of feed water or steam is rarely due to a single contaminant. When a condenser tube springs a leak, chloride, sulfate, carbonate, and many other anions contaminate condensate and feed water. The combined effect will significantly raise cation conductivity and provide the desired early warning even when the individual contamination

April 7, 2

023

9

Page 10: Steam and Water Analyzer

EQPT SPECS. ---CONTD. - ANALYSERS

Common specification for analysers:

Field proven monitors/ analyzers based on microprocessors with LCD display and with necessary fault diagnostic features shall be employed.

CONDUCTIVITY ANALYSER:Type For Hotwell, Two removable

Type of Cells.For Others, Continuous Flow Through

Type Accuracy <= +/- 1 %Response Time <= 1 Sec (90 % of Full Scale)Range 0-1,0-10,0-100 micro-Mho/cm(freely

programmable)for Specific Conductivity.0-1 micro-mho/cm log scale for Cation

conductivityTemperature Comp. Automatic

April 7, 2

023

10

Page 11: Steam and Water Analyzer

Corrosion is function of ph and oxygenAlkaline condition increases stability of magnetite in all

volatile treatment

Principle: ph is normally measured with a glass electrode and a reference electrode. Glass electrode reduces H+ ions and thus leads to en emf dependent on electrode used, type of reference electrode used, and temperature related by Nernst equation.(59.16 mv/25 deg.)-nernstian slope.

pH unit measures the degree of acidity or basicity of a solution. pH is the measurement of the hydrogen ion

concentration, [H+].

April 7, 2

023

11

pH Analyser

Page 12: Steam and Water Analyzer

PH SENSOR

pH sensor uses a differential electrode technique which employs two pH glass measuring electrodes. One electrode is used as the active or measuring electrode, and the other is used as part of a reference assembly. The reference assembly consists of a pH glass measuring electrode immersed in a 7 pH buffer solution. which is mechanically isolated from the solution being measured by a double junction "salt bridge"

The reference junction is located at the measuring end of the reference electrode (or reference electrode assembly for a differential pH sensor). The reference junction is also referred to as a "salt bridge," liquid junction, or frit.

Its purpose is to interface physically and electrically with the internal electrolyte and the solution being measured. This reference junction completes the current path from the glass measuring electrode to the reference electrode.

April 7, 2

023

12

Page 13: Steam and Water Analyzer

PH

At 25°C the slope of the pH electrode is 59.16 mV/pH unit.

At 0°C the slope value is approximately 54 mV/pH, and at 100°C the slope value is approximately 74

mV/pH.

The millivolt output of the glass pH electrode will change with temperature in accordance with the Nernst equation. As the temperature increases, so does the millivolt output. Specifically, the slope of the electrode

April 7, 2

023

13

Page 14: Steam and Water Analyzer

EQPT SPECS. ---CONTD. - ANALYSERS

pH ANALYSER:

Type Flow Through Type

Accuracy <= +/- 1 %

Range 0-14 pH freely programmable

Temperature Comp. Automatic

April 7, 2

023

14

Page 15: Steam and Water Analyzer

Based on Clark cell principle.An oxygen permeable membrane, (no interference)Electrode charged from a constant voltage

source( cathode-gold, anode-silver).Electric current due to reduction of oxygen, current

proportional to the concentration.Rates of reaction and diffusion are temperature

dependent, so RTD installed in the cell.Features: range(0-2 ppm); calibration in air;

programmable alarm levels; o/p’s on relays; 4 mA-20 mA or 0 mA-20 mA; RS-485; alarms4; cell current display( micro/ nano amp.); hysteresis settings; timer operation of relays; average options over a user defined number of inputs.

April 7, 2

023

15

Dissolved oxygen

Page 16: Steam and Water Analyzer

DISSOLVED OXYGEN

April 7, 2

023

16

Dissolved oxygen may lead to corrosion on coming in contact with metal surfaces

Page 17: Steam and Water Analyzer

HYDRAZINE ANALYSER Hydrazine dozing is done for the removal of dissolved

oxygen from water. So, to check dosing of hydrazine, hydrazine conc. has to be monitored.

Cell premixed combination of electrolyte and sample is passed

through a sensing chamber having a exposed porous member.

porous member having a platinum wire electrode wound about the outside thereof and having a silver wire inserted into silver chloride salt (gel) on the inside thereof to form a reference half cell. the porous member and electrode upon being placed in the path of the fluid will cause a current to be developed between the electrodes which is proportional to the amount of chemical in said fluid

pH and temperature compensated

April 7, 2

023

17

Page 18: Steam and Water Analyzer

EQPT SPECS. ---CONTD. - ANALYSERSHYDRAZINE ANALYSER:

Type Automatic Continuous Electrochemical Type

Accuracy <= +/- 5 %Response Time <= 3 Min. (90 % of Full Scale)Range 0-50, 0-100 ppb freely

programmable

Temperature Comp. Automatic

Sample shut off solenoid valve to be provided.

April 7, 2

023

18

Page 19: Steam and Water Analyzer

SENSOR

April 7, 2

023

19

Page 20: Steam and Water Analyzer

EQPT SPECS. ---CONTD. - ANALYSERS

PHOSPHATE ANALYSER:

Type ColourimetricAccuracy <= +/- 5 % of reading

Response Time <= 6 min (90 % of Full Scale)

Range 0-10 ppm freely programmable

Temperature Comp. Automatic

April 7, 2

023

20

Page 21: Steam and Water Analyzer

SILICA

Importance

Solubility of silica decreases with decrease in pressure of steam, so even a small quantity of silica can cause appreciable amount of damage to LP turbine blades by deposition. That may lead to damage to seals or surfacing of vibrations in the turbine

Method of analysis: Continuous Colorimetric

April 7, 2

023

21

Page 22: Steam and Water Analyzer

SILICA

Reagents

April 7, 2

023

22

molybdate Citric acid Amino acidmolybdosilicic

and molybdophosph

oric acids

masks any molybdophosphoric

acids &prevents molybdate from producing an

interfering blue colored compound.

Light absorbance is measured to determine a sample blank reference absorbance. Color formed at this point provides a zero reference.The amount of color formed is directly proportional to the silica concentration of the sample. Light absorbance through the sample is measured at 810 nm. This is compared with the sample blank reference absorbance and silica conc. is calculated.

Page 23: Steam and Water Analyzer

SILICAOperation of the HACH analyzer is semi continuous where

discrete portions of sample are captured and analyzed in a timed sequenced.

A constant flow of sample is directed through a two way valve to the sample cell.

sample flow - 2-8 psi Reagents are stored in containers pressurized at 8-16 psi.

Calibration

Standard sol. from the reagent bottle is added to the sample cell in place of normal sample. the standard sol. is analyzed and the result is used to calculate the slope of the calibration curve.

SiO2= slope x log(reference / sample)

April 7, 2

023

23

Page 24: Steam and Water Analyzer

EQPT SPECS. ---CONTD. - ANALYSERS

SILICA ANALYSER:Type Continuous Colorimetric

Type - Multi stream

Accuracy <= +/- 5 % of readingResponse Time <= 15 min( including sample (90 % of Full Scale) switching)

Range 0-50, 0-100 ,0-500 ppb freely programmable

Temperature Comp. Automatic

Sample shut off solenoid valve to be provided.

April 7, 2

023

24

Page 25: Steam and Water Analyzer

SODIUM

Principle:electrolytic-type.

Sodium sensitive glass electrode has been previously conditioned to a pH > 10.0 .sample conditioning is necessary because sodium glass electrodes are not perfectly specific sensors, but are subject to interference by other ions, specifically hydrogen ion .

April 7, 2

023

25

Page 26: Steam and Water Analyzer

EQPT SPECS. ---CONTD. - ANALYSERS

SODIUM ANALYSER:

Type Continuous Flow through Type – Multi stream

Accuracy <= +/- 10 % of readingResponse Time <= 4 Min(90 % of Full Scale)Range 0-1,0-10, 0-200 ppb

(freely programmable)Temperature Comp. Automatic/

April 7, 2

023

26

Page 27: Steam and Water Analyzer

SEQUENCERVia the help of sequencer, a single channel device can

be made to switch between six different samples.

Sequence of samples being analyzed can be fed into the memory via software interface with the user.

Timings can be programmed.

Serial communication with the analyzing medium.

Flow adjustment via valves.

April 7, 2

023

27

Page 28: Steam and Water Analyzer

April 7, 2

023

28

CONDENSATE PUMP DISCHARGE

PARAMETERS TARGETVALUE

---------------------------------------------------------------------------1.SODIUM, PPB < 3

2.CATION CONDUCTIVITY < 0.2 S/CM

3.OXYGEN, PPB < 10

---------------------------------------------------------------------------

Page 29: Steam and Water Analyzer

April 7, 2

023

29

FEED WATER

PARAMETERS SAMPLE AVT OTFREQUENCY

----------------------------------------------------------------------------------------------------------------------

1. CATION CONDUCTIVITY C < 0.15 <0.15 S/CM2.HYDRAZINE, PPB C 10-15 -----

3. pH C 9.0-9.5 7.0-8.5

4. DISSOLVED OXYGEN, PPB C < 5 30-150

5. TOTAL IRON, PPB S < 2 <2

6. SILICA, PPB S <5 <5

7.SODUIM, PPB S < 2 < 2

8.CHLORIDE, PPB S < 2 < 2

Page 30: Steam and Water Analyzer

April 7, 2

023

30

INSTRUMENTATION FOR CORE PARAMETERS

LOCATION PARAMETERS RANGE ------------------------------------------------------1.MAKE UP WATER CONDUCTIVITY 0 - 1 S/CM

2.CEP DISCHARGE SP.CONDUCTIVITY 0 - 10 S/CMCATION CONDUCTIVITY 0 - 1 S/CM

SODIUM 0 -1,0 - 100 ppbD.O. 0 - 100 ppB

-------------------------------------------------------------------------------

Page 31: Steam and Water Analyzer

April 7, 2

023

31

4.FEED WATER pH 6 - 11SILICA 0 -20 -ppbD.O. 0 - 10 PPb

HYDRAZINE 0 -50 -100PppbCAT.CONDUCTIVITY 0 -1 S/CM

SP.CONDUCTIVITY 0 -10 S/CM

5.BOILER DRUMpH 6 - 11 SILICA 0 -500 ppbSP.CONDUCTIVITY 0 - 10 -20 S/CMCAT.CONDUCTIVITY 0 -1 S/CM

6.STEAM SP.CONDUCTIVITY 0 - 10 S/CM CAT.CONDUCTIVITY 0 -1 S/CM SODIUM 0 -10 ppB

----------------------------------------------------------------------------

LOCATION PARAMETERS RANGE--------------------------------------------------------------------

Page 32: Steam and Water Analyzer

MAIN MENUAlarmsAnalog output (I / v)CalibrationDiagnosticsDisplay modeInstallation- program measurement range and no of sample-

underrange (on/off).PrinterSerial interface(RS-232)Sequence- upto a maximum of 15 is possible“Service” optionTime settings RangeRemote alarm board option (to make sampling options LOCAL-

REMOTE).

April 7, 2

023

32

Page 33: Steam and Water Analyzer

FLUE GAS ANALYSER

April 7, 2

023

33

Page 34: Steam and Water Analyzer

OXYGEN ANALYSER

Principle of Measurement

zirconia oxygen analyzers determine oxygen concentration using the conductivity of a zirconia ceramic cell. Zirconia ceramic cells only allow oxygen ions to pass through at high temperatures. With reference gas on one side and sample gas on the other, oxygen ions move from the side with the highest concentration of oxygen to that with the lowest concentration. The movement of ions generates an EMF (Electro Motive Force) which can be measured to determine the oxygen content

April 7, 2

023

34

Page 35: Steam and Water Analyzer

CO ANALYSER

The main components are an infrared source (lamp), a sample chamber or light tube, a wavelength filter, and the infrared detector. The gas is pumped into the sample chamber, and gas concentration is measured electro-optically by its absorption of a specific wavelength in the infrared (IR).

The IR light is directed through the sample chamber towards the detector. In parallel there is an other chamber with an enclosed reference gas, typically nitrogen. The detector has an optical filter in front of it that eliminates all light except the wavelength that the selected gas molecules can absorb. Ideally other gas molecules do not absorb light at this wavelength, and do not affect the amount of light reaching the detector.

April 7, 2

023

Page 36: Steam and Water Analyzer