Step-by-step procedure for use of FTD

diagnostic assays on the ELITe InGenius™ platform (ELITEch Group)

Materials preparation

1. Mastermix:- FTD Enzyme, Buffer and Primer/probe mix are combined in a 2 ml Sarstedt tube

(ref. 72.694.006) according to FTD usual proportions. However, set a final volumefor 2 additional reactions to take into account the dead volume during pipetting bythe instrument (Table 1)

- The Mastermix can be used for up to 3 consecutive runs on the same day ofreconstitution but it needs to be homogenized by short vortexing between eachrun. Mastermix is stored in the cooling block of the instrument at 8°C during all theduration of the experiment following standard recommendations.

- Do not use the Mastermix for more than 1 day and do not store itat -20°C

Table 11 12

PPmix 4.5 µl 21 µl

Buffer 37.5 µl 175 µl

Enzyme 3 µl 14 µl

Total 45 µl 210 µl

FTD-X

Number of reactions

EliteInGenius – Setup manual – v1 – 2016_11 -1-

2. Internal control:- Because 10 µl is the minimal volume that can be pipetted by the ELITe InGenius™,

FTD Internal control pool (IC) is diluted 5 times in TE buffer before the run andaliquoted in the provided 0.5 ml Sartstedt tube (ref. 72.730.005). Set a finalvolume for two additional reactions than requested (Table 2).

- The diluted and aliquoted IC can be stored for several weeks at -20°C. The IC tubeneeds to be changed by a fresh one for every new run to guarantee the stability ofthe IC detection over the different runs.

Table 2

Materials preparation

Number of

reactionsFTD IC pool (µl) TE buffer (µl)

1 6 24

2 8 32

3 10 40

4 12 48

5 14 56

6 16 64

7 18 72

8 20 80

9 22 88

10 24 96

11 26 104

12 28 112

EliteInGenius – Setup manual – v1 – 2016_11 -2-

3. Negative & Positive controls:- Negative control (NC) with the ELITe InGenius™ is a « PCR only » control (water

or TE buffer). Pipet 15 µl of this NC in an elution tube of the ELITe InGenius™. 10µl will be used for the real-time PCR. In case you want to run a NC following boththe extraction and real-time PCR processes, pipet the NC provided in FTD kits in asonicator tube and consider this track as a generic sample. The volume of NCshould be the same as your sample(s).

- 15 µl of the FTD Positive control (PC) is also pipetted in an elution tube of theELITe InGenius™.

Materials preparation

4. Samples:- Handle your samples as for an usual ELITe InGenius™ run.

EliteInGenius – Setup manual – v1 – 2016_11 -3-

1. Upload FTD assay parameters on the ELITe InGenius™- Flash the assay and the reagent list for run of a negative control, a positive

control, or a generic sample (a). The digital file for each FTD assay is provided bythe ELITech support group ([email protected]).

- These assays include (b) general settings about pipetting volumes (Mastermix,Eluates, IC, etc.) and the targeted pathogens; (c) PCR control (RT step, cycles,temperature, timing, etc.); and (d) an interpretation model that will help for theanalysis of amplification data and determine if a detection is positive or negative.

ELITe InGenius™ run

a b

c d

EliteInGenius – Setup manual – v1 – 2016_11 -4-

2. Run set-up- Choose the appropriate name of the FTD assay for each track of your ELITe

InGenius™ run (negative control, positive control or generic) (a)

- Enter lot numbers of the negative control (water or TE buffer) and positivecontrols used (b)

- Enter sample ID (c)

- Place the Mastermix and IC tubes at the appropriate positions in the cooling blockof the ELITe InGenius™ (d)

ELITe InGenius™ run

a b

c d

EliteInGenius – Setup manual – v1 – 2016_11 -5-

- Follow the usual steps for set-up of an experiment on the ELITe InGenius™

platform

- Start the run

ELITe InGenius™ run

a b c

d e f

EliteInGenius – Setup manual – v1 – 2016_11 -6-

3. Data export and analysis- Approve the results for the tracks with NC and PC (a)

- Print pdf report (b)

- Export raw data

- Analyze the data with your usual criteria