strategic management of recalled cieds...replace registry prospective, observational study of 713...

69
Extraction for Class II Indications - Strategic Management of Recalled CIEDs - HRS Satellite Symposium 2015 Steven P. Kutalek, MD Director, Cardiac Electrophysiology & Pacing Drexel University College of Medicine

Upload: others

Post on 26-Sep-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Extraction for Class II Indications

-Strategic Management of Recalled CIEDs

-HRS Satellite Symposium

2015

Steven P. Kutalek, MD

Director, Cardiac Electrophysiology & Pacing

Drexel University College of Medicine

Page 2: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Disclosures

Consultant:Spectranetics, St Jude Medical, Boston Scientific,

Medtronic, Sorin

Research:Boston Scientific, Medtronic, Sorin, Biotronik

Page 3: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Do you ever abandon leads?

A. Yes

B. No

C. Sometimes

Page 4: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Background

Of the nearly 500,000 pacemaker & ICD devices implanted

yearly throughout the world, many of these implants involve

the need for new or additional leads, due to………

Patient Issues

– Younger patients receiving devices

– Patients living longer with their existing devices

Lead Issues

– Lead Failure

– Device Upgrade

Page 5: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Lead Abandonment vs. Extraction is a

Risk – Benefit Assessment

Risk Of leaving leads in place

Of Extracting leads

Benefit Of leaving leads in place

Of extracting leads

BackgroundAbandon vs. Extract

Page 6: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Lead extraction risks:

– Rupture of SVC

– Cardiac tamponade

– Failure to extract an infected lead

– Venous thrombosis

– Lead breakage and migration

– Avulsion of veins and myocardial tissue

– Death

Lead Extraction Risks

Page 7: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Extraction Risks Increase

BMI < 25

SVC coil of ICD leadEspecially non-back filled

Increased lead implant durationEspecially ICD leads

Elderly female with multiple leads

Page 8: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 9: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Summary of Significant Indications

HRS 2009

FUNCTIONAL LEAD Due to design or failure, may pose immediate threat I

Leads causing life threatening arrhythmias I

Leads interfering with treatment of a malignancy I

Risk of interference with device operation I or IIb

Due to design or failure poses potential future threat IIb

Functional leads not being used (ICD upgrade) IIb

Need MRI with no other imaging options IIb

NON FUNCTIONAL LEAD Due to design or failure, may pose immediate threat I

Leads causing life threatening arrhythmias I

Leads interfering with treatment of a malignancy I

Risk of interference with device operation I or IIa

If implant would require >4 leads one side or 5 leads via SVC IIa

Need MRI with no other imaging options I or IIa

Non functional lead at device/lead procedure IIb

Page 10: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

What are the Issues ?

Features of Lead Abandonment

Data regarding Lead Abandonment

Risk of Future Lead Extraction

Are there any special patient populations?

Are there any specific lead issues?

Page 11: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 12: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 13: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 14: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 15: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

How Many Leads re Too Many ?

Hayes (2000)3 or more abandoned leads

4 leads total

HRS Guidelines (2009)4 Leads too many in either subclavian

5 Leads too many in the SVC

Page 16: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Risks of Abandoning Leads

Older leads are more difficult to remove

Page 17: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 18: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 19: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 20: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 21: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Risks of Abandoning Leads

Older leads are more difficult to remove

More pocket bulk means a greater likelihood

for erosion

Page 22: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 23: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 24: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Lead Fracture Rates

Increase over time

Are greater in younger patients

Page 25: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 26: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 27: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

http://www.medtronic.com/wcm/groups/mdtcom_sg/@mdt/documents/images/fidelis-6949.jpg

Sprint Fidelis® Model 6949 Lead Performance(Medtronic Report as of December 2014)

Page 28: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Sprint Fidelis® Model 6949 Lead

Age Data(Medtronic Report as of December 2014)

http://www.medtronic.com/wcm/groups/mdtcom_sg/@mdt/documents/images/fidelis-6949age.jpg

Page 29: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Infection Rates relative to Lead

Abandonment

Lead infection likelihood increases with each

generator change or lead addition

Indwelling abandoned leads get older and more fibrosed

This makes them more difficult to remove chronically if they

eventually do become infected, which is more likely to

occur with repeated operations

Page 30: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Replace Registry

Prospective, observational study of 713 patients at 62 centers

Patients undergoing generator change without lead addition or

revision

4.2% major complication rate and a 0.6% mortality rate

Patients undergoing device generator change with planned lead

addition or revision

15.3% major complication rate with a 1.1% mortality rate at 6 months

These results demonstrate an increased complication rate with

device system upgrade with lead abandonment

Poole JE, Gleva MJ, Mela T, Chung M, Uslan D, Borge R, Gottipaty V, Shinn T, Dan D, Holcomb R. Complications associated

with planned pacemaker or defibrillator lead addition at the time of generator change: results from the REPLACE registry.

Abstract. Circulation. 2009;120:1

Page 31: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 32: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 33: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Summary of Significant Indications

HRS 2009

CATEGORY INDICATION CLASS

INFECTION Pocket infection, erosion, skin adherence, or draining fistula I

Occult gram-Positive Bacteremia I

Valvular endocarditis with or without definitive lead involvement

I

Occult gram-negative bacteremia IIa

CHRONIC PAIN Severe Chronic Pain not manageable medically or surgically IIa

1. Wilkoff, B.L., Love, C.J., Byrd, C.L., Bongiorni, M.G., Carrillo, R.G., Crossley, G.H.,et al. (2009). Transvenous Lead Extraction: Heart Rhythm Society Expert Consensus on Facilities, Training, Indications,

and Patient Management. Retrieved May 13, 2009, from www.hrsonline.org/Policy/ClinicalGuidelines/leadextract_cieds.cfm

Page 34: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Risks of Abandoning Leads

Older leads are more difficult to remove

More pocket bulk means a greater likelihood for

erosion

Lead malfunction increases over time

Infection risk increases with each generator

change with retained material

Risk of occlusion and stenosis increasesSubclavian / axillary vein

SVC

TV

Page 35: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Summary of Significant Indications

HRS 2009

CATEGORY INDICATION CLASS

OCCLUSION Bilateral subclavian or SVC occlusion precluding implant I

Planned stent deployment I

SVC stenosis or occlusion with symptoms I

Ipsilateral occlusion with contraindication to opposite side I

Significant thromboembolic events with lead thrombus I

Ipsilateral occlusion w/o contralateral contraindication IIa

Page 36: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 37: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 38: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 39: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 40: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Venous Occlusion in ICD Patients

Venography at the time of ICD generator replacement

9% of patients had complete occlusion of the access

vessel

25% had some degree of stenosis

67 % risk of some obstruction in pts with a pacemaker

before placement of the ICD

17 % risk of obstruction in pts with just single leads

Lickfett L, Bitzen A, Arepally A, Nasir K, Wolpert C, Jeong KM, Krause U, Schimpf R, Lewalter T, Calkins H, Jung W, Luderitz B.

Incidence of venous obstruction following insertion of an implantable cardioverter defibrillator: a study of systematic contrast

venography on patients presenting for their first elective ICD generator replacement. Europace. 2004;6:25–31

Page 41: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 42: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Superior Vena Cava Syndrome Induced by Endocardial Defibrillator and Pacemaker LeadsArash Aryana, MS, MDa,*, Kristi D. Sobota, MDb, Dennis J. Esterbrooks, MDc,

and Andrew I. Gelbman, DO, PhDd

Am J Cardiol 2007;99:1765–1767

Page 43: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Severe tricuspid valve stenosis secondary to pacemaker leads presenting

as ascites and liver dysfunction: a complex problem requiring a

multidisciplinary therapeutic approachAnita Krishnan & Achintya Moulick & Pranava Sinha & Karen Kuehl & Joshua Kanter & Michael Slack & Jonathan Kaltman & Marco Mercader &

Jeffrey P. MoakJ Interv Card Electrophysiol (2009) 24:71–75

Page 44: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Heart 2000;83:351–352

Pacemaker lead related tricuspid stenosis: a report of two casesD J Heaven,M Y Henein, R Sutton

Page 45: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Abandoned Pacing Leads

Of 1,207 leads, 611 became non-functional

531 non-functional leads were abandoned

Reasons for abandonment

capture and/or sensing failure (243)

lead recall (177)

lead fracture (86)

pacing system replacement to the contralateral side (11)

accommodating patient growth (5)

pacemaker function upgrade (5)

replacement with ICD (2)

interference with ICD (1)

unknown (1)

Suga C, Hayes DL, Hyberger LK, Lloyd MA

Is there an adverse outcome from abandoned pacing leads? J Interv Card Electrophysiol. 2000 Oct;4(3):493-9.

Page 46: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Abandoned Pacing Leads

5.5 % total pacemaker related complications

Of 24 complications

8 device infections

16 vascular occlusions complicating device revisions

No interactions between the abandoned lead and the active

leads or other complications

Suga C, Hayes DL, Hyberger LK, Lloyd MA

Is there an adverse outcome from abandoned pacing leads? J Interv Card Electrophysiol. 2000 Oct;4(3):493-9.

Page 47: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Abandoned Pacing Leads

Complication rates were higher in patients with

Three or more Abandoned leads

Four or more Total leads

Three or more Procedures of new lead placement

Younger Age at initial pacemaker implantation

Suga C, Hayes DL, Hyberger LK, Lloyd MA

Is there an adverse outcome from abandoned pacing leads? J Interv Card Electrophysiol. 2000 Oct;4(3):493-9.

Page 48: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

60 patients with abandoned leads (“cut and snap”

technique)

Complications in 12 patients (20%)

5 lead migration (with the cut proximal end of the lead

retracted into the vascular space and ending up

looped in the heart or vasculature)

3 skin erosion

2 subclavian vein thrombosis

2 extracardiac stimulation

Bohm A, Pinter A, Duray G, Lehoczky D, Dudas G, Tomcsanyi I, Preda I. Complications due to abandoned

noninfected pacemaker leads. Pacing Clin Electrophysiol. 2001;24:1721–1724.

Abandoned Pacing Leads

Page 49: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

78 ICD patients with 101 abandoned leads

Mean follow-up of 3.1±2.0 years

No complications that could be attributed to

the abandoned leads

Abandoned ICD Leads

Glikson M, Suleiman M, Luria DM, Martin ML, Hodge DO, Shen WK, Bradley DJ, Munger TM, Rea RF, Hayes

DL, Hammill SC, Friedman PA. Do abandoned leads pose risk to implantable cardioverter-defibrillator

patients?Heart Rhythm. 2009;6:65–68.

Page 50: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

399 pacemakers in pediatric patients

18 patients with abandoned leads

Follow-up mean of 4 years

No reports of venous occlusion or electrical problems

2 patients developed endocarditis (at 5 and 10 years

after abandonment)

No other complications related to the leads

Abandoned Pacing Leads

Silvetti MS, Drago F. Outcome of young patients with abandoned, nonfunctional endocardial leads. Pacing Clin

Electrophysiol. 2008;31:473–479

Page 51: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Risk of Future Lead Extraction

> 3500 leads in 266 centers

2-fold increase in the risk of extraction failure with

every 3 years of implant duration 1

212 consecutive patients

3.5-fold increase in TLE complications per additional

right ventricular lead extracted

50% increase in the need for powered sheath

assistance per year increase in implant duration of the

oldest lead 2

1. Byrd CL, Wilkoff BL, Love CJ, Sellers TD, Turk KT, Reeves R, Young R, Crevey B, Kutalek SP, Freedman R, Friedman R, Trantham J, Watts M,

Schutzman J, Oren J, Wilson J, Gold F, Fearnot NE, Van Zandt HJ. Intravascular extraction of problematic or infected permanent pacemaker

leads: 1994–1996. U.S. Extraction Database, MED Institute. Pacing Clin Electrophysiol. 1999;22:1348–1357

2. Agarwal SK, Kamireddy S, Nemec J, Voigt A, Saba S. Predictors of complications of endovascular chronic lead extractions from pacemakers and

defibrillators: a single-operator experience. J Cardiovasc Electrophysiol. 2009;20:171–175

Page 52: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Lead characteristics (n= 2463)

Lead location

RA

RV

CS

Others

34.8%

53.8%

11.2%

0.2%

Lead duration 61.95± 55.24 months

(Range 0-331)

Lead fixation

Active

Passive

67.9%

32.1%

Abandoned leads 10.2%

Lead polarity

Unipolar

Bipolar

95.3%

4.7%

Page 53: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Abandoned leadsAbandoned leads Non-abandoned

leads

p-value

Implant duration 118.35 ±69.15

months

55.78 ±45.8

months

<0.0001

Unipolar leads 5.9% 4.6% 0.380

Passive leads 50.7% 30.1% <0.0001

RV location 76.2% 51.2% <0.0001

ICD leads 35% 29.2% 0.066

Dual coil among

ICD leads

90.9% 87.1% 0.520

Hahnemann Data 2015

Page 54: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Extraction tools for abandoned leadsAbandoned leads Non-abandoned

leads

p-value

Extraction time 18.8 ±21.2

minutes

10.7 ±7.6 minutes <0.0001

Manual traction only 14% 42.8% <0.0001

Adjunct tools 86% 57.2% <0.0001

Locking stylets 81% 56% <0.0001

Dilator sheaths 48.1% 37.2% 0.001

Excimer laser 38.3% 20.6% <0.0001

Electrosurgical

dissection sheath

5.5% 2.7% 0.023

Femoral work station 4.7% 1.5% 0.002

Evolution sheath 1.3% 0.9% 0.494

Snare 6.8% 2.2% <0.0001

Page 55: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Patient outcomes with abandoned leadsAbandoned leads Non-abandoned

leads

p-value

Number of days inpatient

after procedure

9.3 ±8.6 7.6 ±11.9 0.352

Total extraction time 142.4 ±65.9 110.1 ±62.2 0.242

Incision to 1st lead cut

time

30.8 ±14.2 25.8 ±14.4 0.864

Fluoroscopy time 26.8 ±24

minutes

18.4 ±21.4

minutes

0.047

Total laser time 371.3 ±441.7

seconds

198.9 ±273.4

seconds

<0.0001

Total laser pulses 15117 ±17355 10131 ±12338 0.001

Inpatient mortality 3.6% 5.3% 0.374

Page 56: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Patient outcomes with abandoned leadsAbandoned leads Non-abandoned

leads

p-value

Any procedural

complication

3.6% 2.6% 0.467

Major procedural

complication

3.1% 1.6% 0.240

Minor procedural

complication

0.5% 0.9% 0.571

Retained fragments 7.2% 2.7% 0.002

CCU admission 51.3% 24.9% <0.0001

Need for ionotropes 7.7% 4.2% 0.037

Blood transfusion

during procedure

3.1% 1.3% 0.11

Blood transfusion

during hospital stay

19.1% 12.8% 0.03

Page 57: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Special Patient PopulationsMRI

Need for MRI

– Estimated need for future MRI = 50 % to 85 %

– Abandoned lead(s) preclude MRI scanning

Do we need to replace leads for MRI

compatibility?

Page 58: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Fidelis

– Epstein: High failure rate even with “normal”

appearing lead

Riata

– Carrillo: Thrombus formation on exteriorized

cables

Specific Lead IssuesRecalled Leads

Page 59: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Medtronic FidelisWhen to Consider Removal ?

Functionally normal leads have a finite incidence of “subclinical” fracture

209 leads extracted from 208 pts

Implant duration 38.9 (0.2-67.2) months

Evident Fracture incidence = 36.9 %

99 “Normal Leads” removed

20 had fractures

– 4 had more than one fracture site

– 1 had 3 fracture sites

17 pacing conductor fractures = 10 proximal & 7 distal

6 high voltage fractures = 1 SVC & 5 RV

Maytin M, Epstein L M. Lead electrical parameters may not predict integrity of the Sprint Fidelis ICD lead. Heart Rhythm,

9:9. Pages 1446-1451, September 2012

Page 60: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 61: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

LEAD EXTRACTION - Conductor Exteriorization

Page 62: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition
Page 63: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

So What’s a Person to Do?

There are no controlled clinical trials

comparing the long term effects of lead

abandonment with lead extraction

There are risks to lead abandonment as there

are risks to lead extraction

Page 64: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Guidelines & Considerations (?)

Extract

Young patients should be

offered the option of lead

removal

Nonfunctioning leads should

be removed early in their life

if possible

Abandon

Leads fibrose more heavily in

young patients, making

extraction more risky

There are no controlled data

to suggest capping has a

higher long term risk

Page 65: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

Guidelines & Considerations (?)

Abandon

There is no need to remove

leads when the vasculature is

patent

Alternative access sites exist

when the vasculature is

occluded

Extract

The long term complications

of leaving leads in place do

exist

Accessing an alternate site

may lead to occlusion at that

site also

Page 66: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

The Decision must be

Individualized

Age of the patient

Life expectancy

Age & Functionality of lead

Age & Functionality of other leads

Vascular occlusion

Patient’s wishes & risk tolerance

Page 67: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

RecommendationsFor Extraction

In “young” patients, offer the option of lead

removal rather than adding more leads

To avoid vascular overload & occlusion

To avoid the need to extract older leads later with an

increased risk

To avoid lead-lead interaction, especially with ICD

leads

Page 68: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

RecommendationsFor Extraction

Additional viable indications for extraction

Recalled ICD leads in young patients, with or without

generator replacement

The patient who requires an MRI safe system

The patient who would be excluded from MRI with

abandoned leads

The patient who is deemed to no longer require an

implantable anti-arrhythmia device

Page 69: Strategic Management of Recalled CIEDs...Replace Registry Prospective, observational study of 713 patients at 62 centers Patients undergoing generator change without lead addition

RecommendationsFor Abandonment

Abandon leads

When the risk of extraction exceeds the risk of abandonment

– Elderly diminutive patient with multiple leads

– Reduced life expectancy

– Unacceptable co-morbidities

– Extra-vascular course of leads