td-100 installation manual - ingeniería analítica · qui-1066 td-100 installation manual 1...
TRANSCRIPT
QUI-1066 TD-100 Installation Manual
TD-100
Installation Manual
Version 1.8
October 2012
QUI-1066 TD-100 Installation Manual
Contents
1111 Installing TDInstalling TDInstalling TDInstalling TD----100100100100................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................ 3333
1.1 Unpacking TD-100............................................................................................................................... 3
1.2 Installing the transfer line................................................................................................................... 3
1.2.1 Installing the Transfer Line onto the GC............................................................................................. 3
1.2.2 Installing the fused silica transfer line insert .................................................................................... 5
1.2.3 Connecting the transfer line to TD-100 ............................................................................................. 5
1.3 Installing the cold trap ........................................................................................................................ 8
1.4 Removing shipping bracket ..............................................................................................................10
1.5 Checking tube lift position................................................................................................................11
1.6 Connections to TD-100 .....................................................................................................................13
1.6.1 TD-100 with Electronic Carrier Control (ECC) ..................................................................................13
1.6.2 TD-100 without Electronic Carrier Control (ECC) / T’d ECC option.................................................14
1.6.3 T’d ECC...............................................................................................................................................14
2222 Installing the softwareInstalling the softwareInstalling the softwareInstalling the software .................................................................................................................................................................................................................................................................................................................................................................................................................................................................... 16161616
2.1 System Configuration........................................................................................................................16
2.2 Running the software........................................................................................................................16
2.2.1 Configuring the software...................................................................................................................16
2.2.2 Detecting the TD-100........................................................................................................................20
2.3 Setting up installed MFC(s) ..............................................................................................................21
2.4 GC interface logic ..............................................................................................................................22
3333 TTTTDDDD----100 Electronic Gas Control 100 Electronic Gas Control 100 Electronic Gas Control 100 Electronic Gas Control –––– instrument specific guides instrument specific guides instrument specific guides instrument specific guides ........................................................................................................................................................................................................................ 24242424
3.1 TD-100 Electronic Gas Control with the Agilent 6890GC ...............................................................24
3.1.1 Procedure ..........................................................................................................................................24
3.1.2 6890 inlet settings............................................................................................................................25
3.2 TD-100 Electronic Gas Control with the Agilent 7890GC ...............................................................26
3.2.1 Procedure ..........................................................................................................................................26
3.2.2 7890 Inlet Setup ...............................................................................................................................27
3.2.3 7890 Inlet Settings ...........................................................................................................................27
3.3 TD-100 Electronic Gas Control with the Shimadzu QP 2010 (Plus) GCMS fitted with AFC ..........28
3.3.1 Gas connections between TD-100 and the Shimadzu AFC ............................................................28
3.3.2 Configuring the Shimadzu GC software...........................................................................................29
3.3.3 Flow stability ......................................................................................................................................30
3.3.4 Ready / Start connections between TD-100 and Shimadzu GC ....................................................31
3.4 TD-100 Electronic Gas Control with the Thermo DPFC...................................................................32
3.4.1 Procedure ..........................................................................................................................................32
QUI-1066 TD-100 Installation Manual
3.4.2 Configuring the Thermo GC software...............................................................................................32
4444 Installation qualificationInstallation qualificationInstallation qualificationInstallation qualification ........................................................................................................................................................................................................................................................................................................................................................................................................................................................33333333
4.1 Testing the ECC setup .......................................................................................................................33
Appendix A Appendix A Appendix A Appendix A ---- Installing MFC(s) Installing MFC(s) Installing MFC(s) Installing MFC(s) ........................................................................................................................................................................................................................................................................................................................................................................................................................................................................35353535
QUI-1066 TD-100 Installation Manual 1
Preface
This manual provides detailed instructions on the installation of the TD-100. It is suitable for service
engineers with thermal desorption training. Installation should not be attempted by untrained
personnel or the warranty may be invalidated.
I. Regulatory compliance
Our products are thoroughly tested and evaluated to ensure compliance with applicable domestic and
international regulations. This system (hardware and software) is CE compliant and meets
Electromagnetic Compatibility (EMC) Directive and the Low Voltage Directive. The product has been
evaluated to the following safety standards: UL 61010/CSA C22.2 NO 61010-1-04 UPD 1, IEC 61010-1.
II. Warnings
• If the equipment is not used in a way specified by the manufacturer, the protection provided by
the equipment may be reduced. System failures arising from such use may not be covered in
standard warranty and service contract documents.
• Ensure that the plug (electrical isolator) can be easily and quickly accessed during equipment
use.
III. Technical Specifications
Physical Electrical
Height (cm/inches):
Width (cm/inches):
Length (cm/inches):
Mass (kg/lbs):
62/24.40
40/15.7
53/21
39/85.8
Maximum Power (W):
Line voltage:
Frequency (Hz):
Input inrush current (A):
Energy (maximum) (BTU.hour-1):
*automatically selected
650
100-240*
50-60
<40
191
IV. Environment operating conditions
It is advisable to operate the system in a clean laboratory environment, with minimal atmospheric
concentrations of organic vapours. Performance can be affected by sources of heat and cold from
heating, air conditioning systems, or drafts.
Temperature: Recommended operating ambient temperature range is 15 to 30°C.
Humidity: Recommended operating humidity range is 5 to 95% non-condensing.
NOTENOTENOTENOTE: For storage or shipping the allowable temperature range is -40 to 70°C and the allowable
humidity range is 5-95% non-condensing. After instrument exposure to extremes of temperature
or humidity, allow 2 hours for return to the recommended ranges.
Altitude: The recommended operating altitude is up to 8000 ft (2500 m). Higher altitudes pose no
safety risk, but instrument performance may be reduced.
QUI-1066 TD-100 Installation Manual 2
V. Safety alerts
The ‘CAUTION – HOT SURFACE’ symbol indicates a burn hazard. Make sure the instrument is at room
temperature before touching, or you may incur burn injuries.
The ‘BEWARE OF MOVING MACHINERY’ symbol indicates that there are mechanical moving parts which
may cause physical injury.
The ‘LIFTING HAZARD’ symbol indicates that physical injury may occur if the correct lifting procedure for
the instrument is not followed.
Ensure that sufficient resources are available for moving and positioning the unit, and that staff are
competent in manual handling techniques.
VI. Technical support contact details
In the first instance please contact your supplier. If they are unable to resolve your query, please contact
Markes International on the details below.
Address: Gwaun Elai Medi Science Campus, LLANTRISANT, RCT, UK, CF72 8XL
Website: www.markes.com
E-Mail: [email protected]
Telephone: +44 (0) 1443 230935
Fax: +44 (0) 1443 231531
QUI-1066 TD-100 Installation Manual 3
1 Installing TD-100
1.1 Unpacking TD-100
Remove the instrument from its packaging. It is strongly recommended that the instrument packaging
is retained for future use if ever the system is to be shipped using conventional carriers. Shipping the
instrument in non-standard packaging may irreversibly damage the equipment and invalidate the
warranty. It is recommended that 2 people unpack the instrument to minimise the lifting hazard.
Check the packing list included with the instrument to confirm the entire ship kit is present
Before installing the TD-100 determine whether there are any accessories to be installed (such as MFCs
or internal standard accessory). Any accessories should be installed before installing the TD-100 onto
the GC. See appendix A for accessory installation instructions.
1.2 Installing the transfer line
TD-100 is supplied with a universal transfer line (SERUTDSERUTDSERUTDSERUTD----5003500350035003) to convey desorbed analytes from TD-
100 to a gas chromatograph or other analytical system. The sample path utilises a deactivated fused
silica line (0.25 mm I.D. and 0.35 mm O.D.) heated over its entire length by means of a distributed
heater and at the GC end by heat conduction from the GC oven. The line is 1 m long, which is sufficient
to reach most gas chromatographs even when a mass spectrometer is attached.
NoteNoteNoteNote: The parts supplied can be used in other combinations to suit particular instrument
configurations.
1.2.1 Installing the Transfer Line onto the GC
There are two options for transfer line installation, one requires use of a hole in the top, side or back of
the GC oven, while the other option is to gain access via an unused heated injector port.
Option 1 Option 1 Option 1 Option 1 –––– Through Through Through Through Hole in Oven WallHole in Oven WallHole in Oven WallHole in Oven Wall
1. Locate a hole in the inner oven wall with a corresponding hole leading to the outside of the
instrument.
2. Where necessary, displace the oven insulation material such that the transfer line heater can be
fitted directly against the external face of the oven wall.
Photograph
3. Place one of the washers provided in the TD-100 shipping kit on the external face of the oven
wall, over the hole through which the transfer line will be passed.
NOTENOTENOTENOTE: A number of different washers are provided. The washer required will be determined by the
hole in the oven through which the transfer line will pass.
4. Fit the GC end of the TD-100 transfer line trough the hole in the washer and on through the hole
in the GC oven wall.
QUI-1066 TD-100 Installation Manual 4
5. Place a second washer over the TD-100 transfer line so that it can be secured against the inside
wall of the GC oven.
6. Fit the clamp (SERUTDSERUTDSERUTDSERUTD----1125112511251125) onto the transfer line, followed by the spacer (SERUTDSERUTDSERUTDSERUTD----1232123212321232)
secure with the M6 nut supplied with the transfer.
Option 2Option 2Option 2Option 2 –––– Through Heated Injector PortThrough Heated Injector PortThrough Heated Injector PortThrough Heated Injector Port
NOTENOTENOTENOTE: This option is only suitable if the inlet will not be required at any time.
1. Remove components from the injector (e.g. septum, liner etc.) to enable the transfer line to pass
through unobstructed.
2. As the entry hole will generally be larger than the diameter of the metal line sleeve, one or more
of the large washers supplied will be needed.
Oven wall Oven wall
Spacer
Clamp
Washer
Washer
Heated Transfer Line
Nut
Oven wall Oven wall
QUI-1066 TD-100 Installation Manual 5
3. When in operation the injector body should be heated, and should be set to run at a
temperature equal to the TD-100 flow path temperature setting.
1.2.2 Installing the fused silica transfer line insert
Once the heated line has been fitted to the GC, the fused silica plus associated PTFE sleeving (SERUTD-
5093) (see shipping kit) are pushed from the GC end, along the 1/8-inch aluminium tube until they
protrude from the other (TD-100) end of the transfer line.
Trim the PTFE sleeve to ensure enough fused silica is exposed.
Connect the fused silica transfer line to the GC column with a pressfit (C-QSC10) or alternative
connector.
1.2.3 Connecting the transfer line to TD-100
Place TD-100 on the bench on the most convenient side of your gas chromatograph. Ensure that the
transfer line will reach from the top of TD-100 to the selected entry point into the GC oven.
1. Push fused silica
back into sleeve
2. Cut away enough
sleeve so that
~20cm fused silica
Nut
Washer
Injector Body
Washer
Transfer Line Heater
QUI-1066 TD-100 Installation Manual 6
Ensure that TD-100 is switched off and cool.
Remove the front top cover
1) Loosen the thumbscrew
2) Slide the cover off
Remove the top rear cover
1) Loosen the thumbscrew
2) Slide the cover off
Remove the heated valve (HV) cover
1) Loosen the 2 screws
2) Slide the cover off
Place the 1/16 x 0.4 mm (U-FV001 (pk 10)) ferrule
onto the union then thread the 1/16-inch stainless
steel Swagelok type nut (SERZ-0157) on loosely
QUI-1066 TD-100 Installation Manual 7
Pull ~ 20 cm of fused silica from the PTFE line
casing. Cut off the first few mm of fused silica
using an approved capillary cutting tool.
Make a mark 20 mm from the end of the fused
silica using typing white-out fluid or an alternative
marker.
Gently feed the fused silica through the 1/16” nut
and ferrule until the mark is reached.
Tighten the nut to trap the fused silica and then
tighten a further 1/2 turn using one of the 8 mm
wrenches (spanners) provided in the shipping kit.
Do not over tighten or the ferrule will
become distorted.
Carefully bring the clamp plate, PTFE plate and
shield tube down into the position with the shield
tube covering the union nut.
The shield tube should be positioned such that the
1/16th side tubing projects though one of the
cutouts.
Secure the transfer line to TD-100 using four M4
nuts.
20 mm
insertion
depth
QUI-1066 TD-100 Installation Manual 8
Plug the 2 transfer line connectors highlighted into
their respective sockets.
Insert the split filter tube and secure with the red
handle, ensure the heated valve assy. does not
move.
If it does move check the tube alignment.
Replace the HV and rear top covers.
1.3 Installing the cold trap
Ensure the power to TD-100 is
switched OFF.
Remove the front top cover from TD-
100.
Loosen the screw at the bottom of the
trap pneumatics assembly
Disconnect the 2 solenoid
connections.
Gently pull the trap pneumatic
assembly towards the front of the
instrument and lift off.
Split tube
QUI-1066 TD-100 Installation Manual 9
Once clear of the screw, pull the
pneumatics forward and rotate to the
right.
1. Place the trap alignment tool into
the cold trap box.
2. The tool should pass smoothly up
to the o-ring in the heated valve before
further gentle pressure moves the tool
another 1-2 mm.
If the tool does not move
smoothly past the point where it
leaves the cold trap box there
may be a misalignment
problem.
3. Remove the trap alignment tool
NEVER SWITCH TD-100 ON
WITH THE TRAP ALIGNMENT
TOOL INSTALLED
Holding onto the trap as close as
possible to the cold trap box, gently
push the trap into the cold trap box.
DO NOT APPLY EXCESSIVE
FORCE TO THE QUARTZ COLD
TRAP.
If in doubt practice with the trap
alignment tool supplied in the
shipping kit.
NEVER SWITCH TD-100 ON
WITH THE TRAP ALIGNMENT
TOOL INSTALLED.
Push the trap in until it passes the o-
ring.
1-2 mm past
o-ring
QUI-1066 TD-100 Installation Manual 10
Bring the pneumatic assembly back
and relocate the screw in the slot.
Push the assembly gently back in the
horizontal plane guided by the screw,
taking care to align the trap and
stainless steel connector.
MISALIGNMENT CAN CAUSE
THE END OF THE QUARTZ TRAP
TO SNAP
Apply gentle steady pressure to push
the trap into the sealing O-ring located
inside the steel trap connector. If the
pneumatics assembly does slide
easily onto the cold trap remove trap
pneumatics and check seating of the
o-ring.
Keep hold of the pneumatics while re-
tightening the screw firmly.
Refit the front cover
1.4 Removing shipping bracket
TD-100 is shipped with a special
bracket on the tray stack to minimise
risk of damage during transit.
This MUST be removed before
installation or operation.
Open the front door / cover of TD-100
and remove the four screws circled in
red then the transit bracket.
Retain the transit bracket and screws
and re-fit if the system is to be
shipped using conventional carriers.
CautionCautionCautionCaution: No other brackets or clamps
should be removed during installation
Ensure good alignment
here when re-installing
the pneumatics
QUI-1066 TD-100 Installation Manual 11
1.5 Checking tube lift position
Remove the white plastic ‘thumb’ nuts
highlighted then remove the rear
sloping cover.
Remove the rear left top side panel by
pulling the back (narrow) edge of the
panel sideways (away from the
instrument) and then by pushing the
panel backwards and sliding out.
Loosen the white thumbscrew
highlighted then tilt the fan
backwards.
QUI-1066 TD-100 Installation Manual 12
Loosen the two securing screws
holding the oven retaining bracket in
place and slide the bracket
temporarily upwards in the frame to
release the oven.
Rotate the oven assembly around its
hinge point towards the back of the
instrument.
Push the tube lift mechanism fully
down with your finger ensuring that it
is at the bottom of its range of
movement.
Also check the sensor plate is
correctly located and not free to move.
Tilt the oven assembly back again and
secure with the oven retaining
bracket.
Sensor plate
QUI-1066 TD-100 Installation Manual 13
Tilt fan back again and secure with
the white thumbscrew.
Replace covers.
NoteNoteNoteNote: If tube lift is not pushed fully
down then a ‘Clear Tube Lift Fault’ will
be reported when the instrument
initialises.
1.6 Connections to TD-100
1.6.1 TD-100 with Electronic Carrier Control (ECC)
TD-100 ships ready for ECC carrier gas control. A schematic of the ECC connections are shown below.
NoteNoteNoteNote: If ECC is not available please refer to section 1.6.21 to de-configure ECC on TD-100.
NoteNoteNoteNote: As ECC only controls the carrier gas, suitable pneumatic control of the dry gas will still be
required. A U-GAS01 from Markes International includes a carrier gas regulator to step down the carrier
Optional ISDP
Accy.
(see Appendix B)
Autosampler serial
connection to PC
Purge Air In
~ 50 psi
Unused
Carrier Inlet 60
psi max.
Pressure
Outlet (for
ECC control)
Power Cable
TD-100 to GC cable
SERUTE-5142 = AGILENT GC
SERUTE-5143 = THERMO GC
SERUTE-5144 = VARIAN GC
SERUTE-5141 = SHIMADZU /
OTHER GC
Trap module serial
connection to PC
(SERZ-0189)
GCGCGCGC
Septum
Purge
Carrier
TDTDTDTD----100100100100
Pressure
Outlet
Carrier Inlet
QUI-1066 TD-100 Installation Manual 14
pressure and a separate regulator and gauge for control of the dry air or nitrogen, and is therefore
recommended in this case.
1.6.2 TD-100 without Electronic Carrier Control (ECC) / T’d ECC option
If ECC is not available / required then it is possible to reconfigure the TD-100 with the septum purge line
removed.
Remove the rear panel.
Disconnect the 1/16” t-piece
shown.
Replace t-piece with a 1/16”
union (SERZ-0119), supplied
in the shipping kit,
connecting the 2 vertical
lines.
1.6.3 T’d ECC
The two gas lines (carrier and septum purge) from the ECC module on the GC should be connected
together via a 1/8” t-piece as shown below. The third arm of the t-piece should be connected via 1/16”
– 30/1000” i.d. stainless tubing to the ‘Carrier Inlet’ on the back of TD-100.
TDTDTDTD----100100100100
Pressure
Outlet
Carrier Inlet
GCGCGCGC
Carrier
Septum
Purge
Replace t-piece with a
1/16” union connecting
the 2 vertical lines
QUI-1066 TD-100 Installation Manual 15
(NoteNoteNoteNote: if stainless steel tubing is unavailable, green PEEK tubing can be used in its place)
In this alternate configuration there is no need to regulate the carrier gas supply to the EPC module as
close to the TD-100 supply pressure in order to get stable behaviour. This makes it especially suitable
for constant column flow methods where the pressure is ramped during a run to maintain a constant
column flow.
QUI-1066 TD-100 Installation Manual 16
2 Installing the software
Ensure that your PC meets all specifications presented in the site preparation document and is
operating under a 32-bit version of Windows™. It is advised to close down other applications on your PC
while installing the TD-100 software.
Locate the TD-100 software CD supplied in the shipping kit. Insert the CD into the appropriate
compartment of the PC and follow the instructions on the screen. Alternatively, open Windows Explorer
(located under Start > Programs) and access the CD drive. Open the TD-100 folder and scroll down the
content list to Setup.exe. Selecting this option will initiate installation of the software. Proceed as
directed.
Once the TD-100 software has been loaded onto your PC you can access the program from Start >
Programs.
2.1 System Configuration
There are several TD-100 instrument configurations available. The instrument configuration will
determine how the software needs to be configured. Inside the front door of the TD-100 Autosampler is
a label which should be used to determine the system configuration.
TD-100 installed options identification label
Please tick installed options as appropriate
���� Automated re-collection
���� MFC 1 - Split/Re-collection
���� MFC 2 - Trap
Internal Standard addition accessory
Refer to this information when configuring the software.
2.2 Running the software
2.2.1 Configuring the software
With the TD-100 switched off, start the software either by double clicking the TD-100 icon placed on the
desktop or from Start > Programs.
When the TD-100 is not detected choose Select Options?Select Options?Select Options?Select Options?
QUI-1066 TD-100 Installation Manual 17
This will bring up the Configuration page.
For TD-100 there are two available instrument configurations available in the drop down menu; TD-100
No Re-collection and TD-100 Automated Re-collection. TD-100 No Re-collection should be selected if
the Automated re-collection option is NOT ticked on the TD-100 configuration label.
TD-100 installed options identification label
Please tick installed options as appropriate
Automated re-collection
���� MFC 1 - Split/Re-collection
���� MFC 2 - Trap
Internal Standard addition accessory
QUI-1066 TD-100 Installation Manual 18
TD-100 Automated Re-collection should be selected if the Automated re-collection option is ticked on
the TD-100 configuration label.
TD-100 installed options identification label
Please tick installed options as appropriate
���� Automated re-collection
���� MFC 1 - Split/Re-collection
���� MFC 2 - Trap
Internal Standard addition accessory
QUI-1066 TD-100 Installation Manual 19
Ensure that the correct PC communication ports are selected.
NoteNoteNoteNote: the correct port will depend on the PC setup
Change the default Communications ports if necessary by clicking on the down arrow and then clicking
on the correct Com port in the list.
If an Internal Standard accessory is fitted it should also be enabled in the configuration page.
QUI-1066 TD-100 Installation Manual 20
TD-100 installed options identification label
Please tick installed options as appropriate
���� Automated re-collection
���� MFC 1 - Split/Re-collection
���� MFC 2 - Trap
Internal Standard addition accessory
Having setup the instrument configuration click OK, and when prompted by the following message
choose Yes to close the software allowing any changes to take effect.
2.2.2 Detecting the TD-100
Ensure that the gas supplies to the system - especially the dry air or nitrogen purge gas used for TD-100
valve actuation and purging the cold trap box - are on.
Having checked the above, switch TD-100 on using the switch located on the back panel of TD-100 and
QUI-1066 TD-100 Installation Manual 21
re-start the TD-100 software.
If communications between the PC and TD-100 are established successfully the software will open.
If for some reason, communication is not established an ‘Instrument not Detected’ box appears with a
number of options, Cycle Power and retry? Run simulation? Select options? and Exit?.
Choose Select options. This will take you to Configuration tab of Options and change the
communications ports if necessary..
Close down TD-100 software, switch the power to TD-100 off, wait ten seconds and then switch TD-100
back on again.
Re-access the TD-100 program on the PC as described above.
2.3 Setting up installed MFC(s)
For MFC installation instructions see Appendix A.
Upon first start-up of the software the following MFC configuration message will be displayed.
Click ‘Yes’ to go through to the Gas Flow tab to configure the MFCs following the procedure set out
below.
QUI-1066 TD-100 Installation Manual 22
Click ‘Scan for Flow Controllers’.
With 1 MFC setup the designation as
shown.
Note:Note:Note:Note: The full range value does not
need to be entered as this will be
configured automatically.
With 2 MFCs (default for the TD-100
advanced) setup the designation as
shown.
NoteNoteNoteNote: The full range value does not
need to be entered as this will be
configured automatically.
2.4 GC interface logic
The GC interface logic options can be found under the GC InterfaceGC InterfaceGC InterfaceGC Interface tab.
Check the correct operation of the GC ready signal. If the signal appears inverted i.e. software reports
GC not ready when GC is ready, then change the GC Ready (In) logic.
QUI-1066 TD-100 Installation Manual 23
QUI-1066 TD-100 Installation Manual 24
3 TD-100 Electronic Gas Control – instrument specific
guides
3.1 TD-100 Electronic Gas Control with the Agilent 6890GC
NoteNoteNoteNote: o complete this installation the following 6890 parts are recommended. They can be ordered
directly from Agilent.
Part no. Description Quantity
G2131-80500 Replacement 2 port EPC block 1
1390-1022 Screw 1
However, if these parts are not available at installation both the carrier gas and septum purge lines
running between the EPC module and the Split/Splitless injector must be cut and extended using 1/8”
unions and green PEEK tubing in order to facilitate connection to TD-100.
NoteNoteNoteNote: When installing onto an existing 6890GC the firmware on the GC must be A.03.08 or later (for A-
series) or N.04.09 (for N-series). To check the firmware version, use the keyboard on the GC,
press > Options > Diagnostics > Instrument Status and then scroll down to ‘Version’ where you
will see the version of the firmware running on the instrument.
3.1.1 Procedure
1) Remove the top rear cover on the GC to provide access to the EPC module that is going to be used to
regulate the carrier gas supply to TD-100.
Removing the rear panel to access EPC modules
2a) Remove the 3 port EPC block assembly from the split/splitless EPC module.
Fit the replacement EPC block assembly, and route the 1/16” tubing through the rear panel of the GC
and towards the TD-100.
QUI-1066 TD-100 Installation Manual 25
Location of EPC module with the 2 port EPC block fitted
2b) If a replacement EPC block assembly is not available, the septum purge and carrier gas lines must
be cut, extended, using 1/8” brass unions and 1/16” green PEEK tubing, and routed through the rear
panel of the GC towards TD-100. The lengths of PEEK tubing used should be no longer than required to
reach the TD-100.
Carrier and septum purge lines extended with unions and PEEK tubing
The stainless steel tubing for the carrier supply line is 1/16” while the septum purge line is 2 mm and
will require a different ferrule when connected to the 1/8” brass union.
(NoteNoteNoteNote: two 1/8” brass unions, 3 m of 1/16” green PEEK tubing and 1/8” graphitized vespel ferrules,
including one for 2 mm tubing, are all contained within the Agilent GC Installation Kit.)
3) Connect the carrier gas supply line to the ‘Carrier Inlet’ port of TD-100 and connect the septum
purge line to the ‘Pressure Outlet’ port.
3.1.2 6890 inlet settings
Before the column pressure value is set, the MODE of operation of the inlet has to be configured.
For correct control of TD-100 the mode must be set only to SPLITLESS, and not split.
In addition the splitless purge time value must be set to 999.99 minutes DIRECTLY through the
keyboard of the GC (6890 firmware revision A.03.08 or later required). This value cannot be set using
the Chemstation software as the number is rounded to 1000.0 minutes. Setting this value forces the
split/splitless inlet to remain permanently in the splitless mode.
The purge flow value is irrelevant. Once the split/splitless inlet has been configured as described the
column head pressure can be setup to give the desired flow rate/ velocity down the column.
1/8” Unions allowing
extension of gas lines
Replaced 2 port
EPC block
Routed tubing to
TD-100
Standard 3 port
EPC block
Lines cut
here
QUI-1066 TD-100 Installation Manual 26
3.2 TD-100 Electronic Gas Control with the Agilent 7890GC
NoteNoteNoteNote: To complete this installation the following 7890 parts are recommended. They can be ordered
directly from Agilent.
Part no. Description Quantity
G3430-60011 Replacement hoodlum block assemblies 1
However, if these parts are not available at installation both the carrier gas and septum purge lines
running between the EPC module and the Split/Splitless injector must be cut and extended using 1/8”
unions and green PEEK tubing in order to facilitate connection to TD-100.
3.2.1 Procedure
1) Remove the top rear cover on the GC to provide access to the EPC module that is going to be used to
regulate the carrier gas supply to TD-100.
2a) Remove the two hoodlum block assemblies, for the carrier supply and septum purge lines, from the
split/splitless EPC module.
Location of fixing screws for bulkhead connectors
Fit the two replacement hoodlum block assemblies, and route the 1/16” tubing on both of these
assemblies through the rear panel of the GC and towards the TD-100.
Note: the carrier gas supply line has a 1/8” termination at the TD end while the septum purge line
terminates in 1/16” tubing.
2b) If replacement hoodlum block assemblies are not available, the septum purge and carrier gas lines
must be cut, extended, using 1/8” brass unions and 1/16” green PEEK tubing, and routed through the
rear panel of the GC towards TD-100. The lengths of PEEK tubing used should be no longer than
required to reach the TD-100.
Carrier and septum purge lines extended with unions and PEEK tubing
EPC Module viewed
from above
EPC Module
viewed from
rear
Septum Purge
Connector
Lines cut here 1/8” Unions allowing
extension of gas lines
Carrier Supply
Connector
QUI-1066 TD-100 Installation Manual 27
The stainless steel tubing for the carrier supply line is 1/16” while the septum purge line is 2mm and
will require a different ferrule when connected to the 1/8” brass union.
(Note: two 1/8” brass unions, 3 m of 1/16” green PEEK tubing and 1/8” graphitized vespel ferrules,
including one for 2 mm tubing, are all contained within the TD-100 shipping kit.)
3) Connect the carrier gas supply line to the ‘Carrier Inlet’ port of TD-100 and connect the septum
purge line to the ‘Pressure Outlet’ port.
3.2.2 7890 Inlet Setup
There are several settings that must be implemented on the 7890 whenever the split/splitless EPC
module is controlling the carrier gas supply to the TD-100 thermal desorber. In the Inlet settings the
mode must be set to ‘Splitless’, the purge time must be set to exactly ‘999.99’ minutes and Gas saver
should be set to ‘Off’. In the Instrument Configuration menu the Auto prep run should be set to ‘Off’.
Having completed this, the system should now be ready to use.
3.2.3 7890 Inlet Settings
If the pressure is not correctly controlled after inputting the above settings then turn ‘Auto prep run’ to
‘On’.
QUI-1066 TD-100 Installation Manual 28
3.3 TD-100 Electronic Gas Control with the Shimadzu QP 2010
(Plus) GCMS fitted with AFC
NoteNoteNoteNote: To complete this installation the specialised GM - 1/16” fittings must be ordered directly from
Shimadzu prior to the installation. The parts required are as follows:
Part no. Description Quantity
221-35003 Shimadzu fitting GM-1.6L 1
221-35004-91 Shimadzu fitting GM-1.6L 1
228-16000-84 Ferrules 4
228-16002-84 Nuts 2
3.3.1 Gas connections between TD-100 and the Shimadzu AFC
1) The Shimadzu GC is fitted with two molecular sieve filters, one in the carrier gas line and one in the
septum purge line. Remove the filter from the carrier gas line and replace with the Shimadzu fitting GM-
1.6L. The filter on the septum purge line should remain in place to prevent back diffusion of air/water.
Shimadzu fitting on
carrier gas line
Molecular Sieve filter on
septum purge line
Two lengths of PEEK tubing
connecting to TD-100
TDTDTDTD----100100100100
QUI-1066 TD-100 Installation Manual 29
2) The carrier gas out line from the Shimadzu AFC (“CG Out”) is now connected to the “Carrier Inlet”
port on the back of TD-100 using a 1 m length of green PEEK tubing.
3) The “Pressure Outlet” from the back of TD-100 is now connected - with a similar length of green
PEEK tubing - to the molecular sieve filter on the septum purge line (and hence to the (“SP In”)) on the
Shimadzu AFC. It is important that the two lengths of green PEEK tubing are the same length. Carrier
gas is supplied and regulated upstream of the system in the supply line and it’s pressure is monitored
downstream of the system in the septum purge line.
3.3.2 Configuring the Shimadzu GC software
1) The inlet Split Mode on the Shimadzu GC must be set to Direct.
i. To access the Split Mode press the “FLOW” key on the GC keypad.
NoteNoteNoteNote: In the case of GCs with two or more AFC units it may be necessary to press the “FLOW” key
repeatedly until the correct one is displayed.
ii. Move the flashing cursor over the “Split Mode” (type using the arrow keys on the GC keypad).
iii. Using the right arrow key, select “DIRECT” mode as shown, and press the “ENTER” key on the GC
keypad.
Setting Inlet Mode as Direct on Shimadzu GC
2) The AFC module allows read out of the total flow supplied to system - this is ideal for setting split and
trap / desorb flows and for troubleshooting leaks.
i. To view the total flow supplied press the “FLOW” key on the GC keypad as above.
ii. The Total Flow in ml/min is always displayed (above the Split Mode) as shown above.
3) The septum purge flow should be set to a flow rate of around 5 ml/min
i. To set the septum purge flow display the “FLOW” screen as described above.
ii. Press the button to the right of the “PF3” key (below the screen) as shown below. This displays a
second set of commands for the PF keys.
Extending the PF commands
QUI-1066 TD-100 Installation Manual 30
iii. Select the purge command on “PF3” and enter the desired value. A flow rate of 2-5 ml/min is
recommended.
Setting the septum purge flow value
3.3.3 Flow stability
AFC works with both impeded (traps containing porous polymer e.g. Tenax) and un-impeded (traps only
containing carbon sorbents) traps, with no carrier gas instability generally noted (even with 90-100 psi
supplied to the inlet of the AFC system). However should any instability in the carrier gas occur it may be
necessary to dampen the reaction of the AFC controller.
i. Press the “FUNC” key on the GC keypad.
ii. Enter “7”, enter “5”, enter “1201”, enter “1”. This will display the “Flow Adjust” page.
iii. For instruments with two or more AFC units, press the “PF2” button to select the correct flow unit.
iv. Move the cursor to the “I Time” entry in the “Flow” column.
v. Change the value from “1” to “4” then press the “Enter” key.
Damping the AFC reaction
QUI-1066 TD-100 Installation Manual 31
3.3.4 Ready / Start connections between TD-100 and Shimadzu GC
The remote cable connections between TD-100 and the GC are shown below. The TD-100 - GC cable
(SERUTD-5095) has a 25-pin D-type connector at the TD-100 end and free wires at the GC end. Four
wires need to be inserted into the connection block at the rear of the QP2010 as shown below.
Ready / Start connections between TD-100 and QP 2010
When connected to the Shimadzu GCMS system the TD-100 ready (in) logic must be set to Open as
shown below. To set the logic select view>options from the drop down menus to open the options
dialogue box.
Setting the TD-100 GC Ready (in) logic to Open
QUI-1066 TD-100 Installation Manual 32
3.4 TD-100 Electronic Gas Control with the Thermo DPFC
A TD-100 can be integrated into the DPFC system of a Thermo Electron GC. The DPFC module regulates
the carrier gas supply upstream of the TD-100 and its pressure is monitored downstream of the system
in the septum purge line of the DPFC module. The split line of the DPFC module remains unused.
3.4.1 Procedure
1) The carrier gas supply and the septum purge lines that normally run from the DPFC to the
split/splitless injector need to be cut and fitted with 1/8” brass unions (note: the carrier line requires
use of graphitised vespel ferrule with the 2 mm hole – part number Z-0402).
2) Extend these two lines with the length of green PEEK tubing so that they can reach the back panel of
TD-100. The carrier gas line from the DPFC module should be connected to the TD-100 ‘Carrier Inlet’
and the septum purge line should be connected to the ‘Pressure Outlet’.
Schematic of connections made between DPFC module and TD-100
3.4.2 Configuring the Thermo GC software
Having completed the physical connection between the Thermo Focus or Trace GC and TD-100 the next
step is to configure the inlet settings.
1) The GC must be set up for back pressure regulation. To do this the DPFC must be removed and the
switches set in the correct position for purge & trap. Setting up this configuration is described in the kit
adapter for the purge and trap concentrator system installation guide.
2) The inlet that is being used to control the TD-100 carrier gas supply should then be set to operate in
split mode but the split flow should be set to zero (off).
3) The only other important consideration is to ensure that the pressure of the carrier gas supplied to
the DPFC module is only 10-15 psi greater than the pressure that the DPFC module is required to supply
to TD-100.
DPFC DPFC DPFC DPFC
ModuleModuleModuleModule
Septum
Purge
Carrier
TDTDTDTD----100100100100
Pressure
Outlet
Carrier Inlet
QUI-1066 TD-100 Installation Manual 33
4 Installation qualification
4.1 Testing the ECC setup
Once the connections and inlet settings have been correctly set up it is extremely important that the
following test of the ECC carrier gas supply is carried out:
During the desorption of the TD-100 cold trap the total flow and pressure supply of the ECC module
should be monitored. Both pressure and flow readings should be stable.
If there is any variation in the pressure reading of the ECC module, or if the flow fluctuates by more
3mL/min, this indicates that the setup is not optimized. In such circumstances the reproducibility of
analysis is likely to be poor.
ECC instability during trap desorption
Such fluctuations can normally be avoided by regulating the carrier gas supply to the ECC module to 10
psi above the greatest pressure required by the TD-100 in the course of a run, see below. This is almost
always effective in removing any instability when using a constant pressure method, but may not be
effective for constant flow methods.
Correct Setup of ECC Supply
TDTDTDTD----100100100100
GCGCGCGC EPCEPCEPCEPC
Carrier
Septum Purge
Carrier Supply to GC
25 psi
Total Trap Desorption
Standby
Time
30
Inlet
Pressure 15 psi
Total Flow 30 ml/min
Pressure Differential = 25 – 15 = 10psi
QUI-1066 TD-100 Installation Manual 34
In situations where the pressure and/or flow are unstable despite the appropriate regulation of the
carrier gas supply there is an alternative configuration that should be employed. (Refer to section 1.6.21
for further information and installation instructions.)
QUI-1066 TD-100 Installation Manual 35
Appendix A - Installing MFC(s)
Remove front right cover
Remove right hand cover by loosening
the 3 screws highlighted on the front
and 3 (not shown) on the rear then
removing the 2 screws on the side of
the cover.
Remove MFC mounting plate by
removing 2 x screws highlighted then
lifting the plate out.
MFCs should be mounted on the plate
in the positions shown, depending on
the number of MFC present.
1 MFC1 MFC1 MFC1 MFC
Once installed re-attach the MFC
mounting plate.
Split /
Re-coll.
T’d
MFC plate
QUI-1066 TD-100 Installation Manual 36
2 MFCs2 MFCs2 MFCs2 MFCs
Once installed re-attach the MFC
mounting plate.
The MFCs should be placed upstream
of their respective needle valve (and
t’d if necessary).
For example:
The needle valve(s) should then be
opened so a flow just above the
maximum of the MFC can be
achieved.
The connection(s) to the control PCB
must be made.
The green RJ45 cable should be
routed through and plugged into the
top right of the control PCB as shown.
ERROR: stackunderflow
OFFENDING COMMAND: ~
STACK: