td-100 installation manual - ingeniería analítica · qui-1066 td-100 installation manual 1...

40
QUI-1066 TD-100 Installation Manual TD-100 Installation Manual Version 1.8 October 2012

Upload: vohuong

Post on 15-Sep-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual

TD-100

Installation Manual

Version 1.8

October 2012

Page 2: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual

Contents

1111 Installing TDInstalling TDInstalling TDInstalling TD----100100100100................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................ 3333

1.1 Unpacking TD-100............................................................................................................................... 3

1.2 Installing the transfer line................................................................................................................... 3

1.2.1 Installing the Transfer Line onto the GC............................................................................................. 3

1.2.2 Installing the fused silica transfer line insert .................................................................................... 5

1.2.3 Connecting the transfer line to TD-100 ............................................................................................. 5

1.3 Installing the cold trap ........................................................................................................................ 8

1.4 Removing shipping bracket ..............................................................................................................10

1.5 Checking tube lift position................................................................................................................11

1.6 Connections to TD-100 .....................................................................................................................13

1.6.1 TD-100 with Electronic Carrier Control (ECC) ..................................................................................13

1.6.2 TD-100 without Electronic Carrier Control (ECC) / T’d ECC option.................................................14

1.6.3 T’d ECC...............................................................................................................................................14

2222 Installing the softwareInstalling the softwareInstalling the softwareInstalling the software .................................................................................................................................................................................................................................................................................................................................................................................................................................................................... 16161616

2.1 System Configuration........................................................................................................................16

2.2 Running the software........................................................................................................................16

2.2.1 Configuring the software...................................................................................................................16

2.2.2 Detecting the TD-100........................................................................................................................20

2.3 Setting up installed MFC(s) ..............................................................................................................21

2.4 GC interface logic ..............................................................................................................................22

3333 TTTTDDDD----100 Electronic Gas Control 100 Electronic Gas Control 100 Electronic Gas Control 100 Electronic Gas Control –––– instrument specific guides instrument specific guides instrument specific guides instrument specific guides ........................................................................................................................................................................................................................ 24242424

3.1 TD-100 Electronic Gas Control with the Agilent 6890GC ...............................................................24

3.1.1 Procedure ..........................................................................................................................................24

3.1.2 6890 inlet settings............................................................................................................................25

3.2 TD-100 Electronic Gas Control with the Agilent 7890GC ...............................................................26

3.2.1 Procedure ..........................................................................................................................................26

3.2.2 7890 Inlet Setup ...............................................................................................................................27

3.2.3 7890 Inlet Settings ...........................................................................................................................27

3.3 TD-100 Electronic Gas Control with the Shimadzu QP 2010 (Plus) GCMS fitted with AFC ..........28

3.3.1 Gas connections between TD-100 and the Shimadzu AFC ............................................................28

3.3.2 Configuring the Shimadzu GC software...........................................................................................29

3.3.3 Flow stability ......................................................................................................................................30

3.3.4 Ready / Start connections between TD-100 and Shimadzu GC ....................................................31

3.4 TD-100 Electronic Gas Control with the Thermo DPFC...................................................................32

3.4.1 Procedure ..........................................................................................................................................32

Page 3: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual

3.4.2 Configuring the Thermo GC software...............................................................................................32

4444 Installation qualificationInstallation qualificationInstallation qualificationInstallation qualification ........................................................................................................................................................................................................................................................................................................................................................................................................................................................33333333

4.1 Testing the ECC setup .......................................................................................................................33

Appendix A Appendix A Appendix A Appendix A ---- Installing MFC(s) Installing MFC(s) Installing MFC(s) Installing MFC(s) ........................................................................................................................................................................................................................................................................................................................................................................................................................................................................35353535

Page 4: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 1

Preface

This manual provides detailed instructions on the installation of the TD-100. It is suitable for service

engineers with thermal desorption training. Installation should not be attempted by untrained

personnel or the warranty may be invalidated.

I. Regulatory compliance

Our products are thoroughly tested and evaluated to ensure compliance with applicable domestic and

international regulations. This system (hardware and software) is CE compliant and meets

Electromagnetic Compatibility (EMC) Directive and the Low Voltage Directive. The product has been

evaluated to the following safety standards: UL 61010/CSA C22.2 NO 61010-1-04 UPD 1, IEC 61010-1.

II. Warnings

• If the equipment is not used in a way specified by the manufacturer, the protection provided by

the equipment may be reduced. System failures arising from such use may not be covered in

standard warranty and service contract documents.

• Ensure that the plug (electrical isolator) can be easily and quickly accessed during equipment

use.

III. Technical Specifications

Physical Electrical

Height (cm/inches):

Width (cm/inches):

Length (cm/inches):

Mass (kg/lbs):

62/24.40

40/15.7

53/21

39/85.8

Maximum Power (W):

Line voltage:

Frequency (Hz):

Input inrush current (A):

Energy (maximum) (BTU.hour-1):

*automatically selected

650

100-240*

50-60

<40

191

IV. Environment operating conditions

It is advisable to operate the system in a clean laboratory environment, with minimal atmospheric

concentrations of organic vapours. Performance can be affected by sources of heat and cold from

heating, air conditioning systems, or drafts.

Temperature: Recommended operating ambient temperature range is 15 to 30°C.

Humidity: Recommended operating humidity range is 5 to 95% non-condensing.

NOTENOTENOTENOTE: For storage or shipping the allowable temperature range is -40 to 70°C and the allowable

humidity range is 5-95% non-condensing. After instrument exposure to extremes of temperature

or humidity, allow 2 hours for return to the recommended ranges.

Altitude: The recommended operating altitude is up to 8000 ft (2500 m). Higher altitudes pose no

safety risk, but instrument performance may be reduced.

Page 5: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 2

V. Safety alerts

The ‘CAUTION – HOT SURFACE’ symbol indicates a burn hazard. Make sure the instrument is at room

temperature before touching, or you may incur burn injuries.

The ‘BEWARE OF MOVING MACHINERY’ symbol indicates that there are mechanical moving parts which

may cause physical injury.

The ‘LIFTING HAZARD’ symbol indicates that physical injury may occur if the correct lifting procedure for

the instrument is not followed.

Ensure that sufficient resources are available for moving and positioning the unit, and that staff are

competent in manual handling techniques.

VI. Technical support contact details

In the first instance please contact your supplier. If they are unable to resolve your query, please contact

Markes International on the details below.

Address: Gwaun Elai Medi Science Campus, LLANTRISANT, RCT, UK, CF72 8XL

Website: www.markes.com

E-Mail: [email protected]

Telephone: +44 (0) 1443 230935

Fax: +44 (0) 1443 231531

Page 6: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 3

1 Installing TD-100

1.1 Unpacking TD-100

Remove the instrument from its packaging. It is strongly recommended that the instrument packaging

is retained for future use if ever the system is to be shipped using conventional carriers. Shipping the

instrument in non-standard packaging may irreversibly damage the equipment and invalidate the

warranty. It is recommended that 2 people unpack the instrument to minimise the lifting hazard.

Check the packing list included with the instrument to confirm the entire ship kit is present

Before installing the TD-100 determine whether there are any accessories to be installed (such as MFCs

or internal standard accessory). Any accessories should be installed before installing the TD-100 onto

the GC. See appendix A for accessory installation instructions.

1.2 Installing the transfer line

TD-100 is supplied with a universal transfer line (SERUTDSERUTDSERUTDSERUTD----5003500350035003) to convey desorbed analytes from TD-

100 to a gas chromatograph or other analytical system. The sample path utilises a deactivated fused

silica line (0.25 mm I.D. and 0.35 mm O.D.) heated over its entire length by means of a distributed

heater and at the GC end by heat conduction from the GC oven. The line is 1 m long, which is sufficient

to reach most gas chromatographs even when a mass spectrometer is attached.

NoteNoteNoteNote: The parts supplied can be used in other combinations to suit particular instrument

configurations.

1.2.1 Installing the Transfer Line onto the GC

There are two options for transfer line installation, one requires use of a hole in the top, side or back of

the GC oven, while the other option is to gain access via an unused heated injector port.

Option 1 Option 1 Option 1 Option 1 –––– Through Through Through Through Hole in Oven WallHole in Oven WallHole in Oven WallHole in Oven Wall

1. Locate a hole in the inner oven wall with a corresponding hole leading to the outside of the

instrument.

2. Where necessary, displace the oven insulation material such that the transfer line heater can be

fitted directly against the external face of the oven wall.

Photograph

3. Place one of the washers provided in the TD-100 shipping kit on the external face of the oven

wall, over the hole through which the transfer line will be passed.

NOTENOTENOTENOTE: A number of different washers are provided. The washer required will be determined by the

hole in the oven through which the transfer line will pass.

4. Fit the GC end of the TD-100 transfer line trough the hole in the washer and on through the hole

in the GC oven wall.

Page 7: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 4

5. Place a second washer over the TD-100 transfer line so that it can be secured against the inside

wall of the GC oven.

6. Fit the clamp (SERUTDSERUTDSERUTDSERUTD----1125112511251125) onto the transfer line, followed by the spacer (SERUTDSERUTDSERUTDSERUTD----1232123212321232)

secure with the M6 nut supplied with the transfer.

Option 2Option 2Option 2Option 2 –––– Through Heated Injector PortThrough Heated Injector PortThrough Heated Injector PortThrough Heated Injector Port

NOTENOTENOTENOTE: This option is only suitable if the inlet will not be required at any time.

1. Remove components from the injector (e.g. septum, liner etc.) to enable the transfer line to pass

through unobstructed.

2. As the entry hole will generally be larger than the diameter of the metal line sleeve, one or more

of the large washers supplied will be needed.

Oven wall Oven wall

Spacer

Clamp

Washer

Washer

Heated Transfer Line

Nut

Oven wall Oven wall

Page 8: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 5

3. When in operation the injector body should be heated, and should be set to run at a

temperature equal to the TD-100 flow path temperature setting.

1.2.2 Installing the fused silica transfer line insert

Once the heated line has been fitted to the GC, the fused silica plus associated PTFE sleeving (SERUTD-

5093) (see shipping kit) are pushed from the GC end, along the 1/8-inch aluminium tube until they

protrude from the other (TD-100) end of the transfer line.

Trim the PTFE sleeve to ensure enough fused silica is exposed.

Connect the fused silica transfer line to the GC column with a pressfit (C-QSC10) or alternative

connector.

1.2.3 Connecting the transfer line to TD-100

Place TD-100 on the bench on the most convenient side of your gas chromatograph. Ensure that the

transfer line will reach from the top of TD-100 to the selected entry point into the GC oven.

1. Push fused silica

back into sleeve

2. Cut away enough

sleeve so that

~20cm fused silica

Nut

Washer

Injector Body

Washer

Transfer Line Heater

Page 9: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 6

Ensure that TD-100 is switched off and cool.

Remove the front top cover

1) Loosen the thumbscrew

2) Slide the cover off

Remove the top rear cover

1) Loosen the thumbscrew

2) Slide the cover off

Remove the heated valve (HV) cover

1) Loosen the 2 screws

2) Slide the cover off

Place the 1/16 x 0.4 mm (U-FV001 (pk 10)) ferrule

onto the union then thread the 1/16-inch stainless

steel Swagelok type nut (SERZ-0157) on loosely

Page 10: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 7

Pull ~ 20 cm of fused silica from the PTFE line

casing. Cut off the first few mm of fused silica

using an approved capillary cutting tool.

Make a mark 20 mm from the end of the fused

silica using typing white-out fluid or an alternative

marker.

Gently feed the fused silica through the 1/16” nut

and ferrule until the mark is reached.

Tighten the nut to trap the fused silica and then

tighten a further 1/2 turn using one of the 8 mm

wrenches (spanners) provided in the shipping kit.

Do not over tighten or the ferrule will

become distorted.

Carefully bring the clamp plate, PTFE plate and

shield tube down into the position with the shield

tube covering the union nut.

The shield tube should be positioned such that the

1/16th side tubing projects though one of the

cutouts.

Secure the transfer line to TD-100 using four M4

nuts.

20 mm

insertion

depth

Page 11: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 8

Plug the 2 transfer line connectors highlighted into

their respective sockets.

Insert the split filter tube and secure with the red

handle, ensure the heated valve assy. does not

move.

If it does move check the tube alignment.

Replace the HV and rear top covers.

1.3 Installing the cold trap

Ensure the power to TD-100 is

switched OFF.

Remove the front top cover from TD-

100.

Loosen the screw at the bottom of the

trap pneumatics assembly

Disconnect the 2 solenoid

connections.

Gently pull the trap pneumatic

assembly towards the front of the

instrument and lift off.

Split tube

Page 12: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 9

Once clear of the screw, pull the

pneumatics forward and rotate to the

right.

1. Place the trap alignment tool into

the cold trap box.

2. The tool should pass smoothly up

to the o-ring in the heated valve before

further gentle pressure moves the tool

another 1-2 mm.

If the tool does not move

smoothly past the point where it

leaves the cold trap box there

may be a misalignment

problem.

3. Remove the trap alignment tool

NEVER SWITCH TD-100 ON

WITH THE TRAP ALIGNMENT

TOOL INSTALLED

Holding onto the trap as close as

possible to the cold trap box, gently

push the trap into the cold trap box.

DO NOT APPLY EXCESSIVE

FORCE TO THE QUARTZ COLD

TRAP.

If in doubt practice with the trap

alignment tool supplied in the

shipping kit.

NEVER SWITCH TD-100 ON

WITH THE TRAP ALIGNMENT

TOOL INSTALLED.

Push the trap in until it passes the o-

ring.

1-2 mm past

o-ring

Page 13: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 10

Bring the pneumatic assembly back

and relocate the screw in the slot.

Push the assembly gently back in the

horizontal plane guided by the screw,

taking care to align the trap and

stainless steel connector.

MISALIGNMENT CAN CAUSE

THE END OF THE QUARTZ TRAP

TO SNAP

Apply gentle steady pressure to push

the trap into the sealing O-ring located

inside the steel trap connector. If the

pneumatics assembly does slide

easily onto the cold trap remove trap

pneumatics and check seating of the

o-ring.

Keep hold of the pneumatics while re-

tightening the screw firmly.

Refit the front cover

1.4 Removing shipping bracket

TD-100 is shipped with a special

bracket on the tray stack to minimise

risk of damage during transit.

This MUST be removed before

installation or operation.

Open the front door / cover of TD-100

and remove the four screws circled in

red then the transit bracket.

Retain the transit bracket and screws

and re-fit if the system is to be

shipped using conventional carriers.

CautionCautionCautionCaution: No other brackets or clamps

should be removed during installation

Ensure good alignment

here when re-installing

the pneumatics

Page 14: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 11

1.5 Checking tube lift position

Remove the white plastic ‘thumb’ nuts

highlighted then remove the rear

sloping cover.

Remove the rear left top side panel by

pulling the back (narrow) edge of the

panel sideways (away from the

instrument) and then by pushing the

panel backwards and sliding out.

Loosen the white thumbscrew

highlighted then tilt the fan

backwards.

Page 15: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 12

Loosen the two securing screws

holding the oven retaining bracket in

place and slide the bracket

temporarily upwards in the frame to

release the oven.

Rotate the oven assembly around its

hinge point towards the back of the

instrument.

Push the tube lift mechanism fully

down with your finger ensuring that it

is at the bottom of its range of

movement.

Also check the sensor plate is

correctly located and not free to move.

Tilt the oven assembly back again and

secure with the oven retaining

bracket.

Sensor plate

Page 16: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 13

Tilt fan back again and secure with

the white thumbscrew.

Replace covers.

NoteNoteNoteNote: If tube lift is not pushed fully

down then a ‘Clear Tube Lift Fault’ will

be reported when the instrument

initialises.

1.6 Connections to TD-100

1.6.1 TD-100 with Electronic Carrier Control (ECC)

TD-100 ships ready for ECC carrier gas control. A schematic of the ECC connections are shown below.

NoteNoteNoteNote: If ECC is not available please refer to section 1.6.21 to de-configure ECC on TD-100.

NoteNoteNoteNote: As ECC only controls the carrier gas, suitable pneumatic control of the dry gas will still be

required. A U-GAS01 from Markes International includes a carrier gas regulator to step down the carrier

Optional ISDP

Accy.

(see Appendix B)

Autosampler serial

connection to PC

Purge Air In

~ 50 psi

Unused

Carrier Inlet 60

psi max.

Pressure

Outlet (for

ECC control)

Power Cable

TD-100 to GC cable

SERUTE-5142 = AGILENT GC

SERUTE-5143 = THERMO GC

SERUTE-5144 = VARIAN GC

SERUTE-5141 = SHIMADZU /

OTHER GC

Trap module serial

connection to PC

(SERZ-0189)

GCGCGCGC

Septum

Purge

Carrier

TDTDTDTD----100100100100

Pressure

Outlet

Carrier Inlet

Page 17: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 14

pressure and a separate regulator and gauge for control of the dry air or nitrogen, and is therefore

recommended in this case.

1.6.2 TD-100 without Electronic Carrier Control (ECC) / T’d ECC option

If ECC is not available / required then it is possible to reconfigure the TD-100 with the septum purge line

removed.

Remove the rear panel.

Disconnect the 1/16” t-piece

shown.

Replace t-piece with a 1/16”

union (SERZ-0119), supplied

in the shipping kit,

connecting the 2 vertical

lines.

1.6.3 T’d ECC

The two gas lines (carrier and septum purge) from the ECC module on the GC should be connected

together via a 1/8” t-piece as shown below. The third arm of the t-piece should be connected via 1/16”

– 30/1000” i.d. stainless tubing to the ‘Carrier Inlet’ on the back of TD-100.

TDTDTDTD----100100100100

Pressure

Outlet

Carrier Inlet

GCGCGCGC

Carrier

Septum

Purge

Replace t-piece with a

1/16” union connecting

the 2 vertical lines

Page 18: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 15

(NoteNoteNoteNote: if stainless steel tubing is unavailable, green PEEK tubing can be used in its place)

In this alternate configuration there is no need to regulate the carrier gas supply to the EPC module as

close to the TD-100 supply pressure in order to get stable behaviour. This makes it especially suitable

for constant column flow methods where the pressure is ramped during a run to maintain a constant

column flow.

Page 19: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 16

2 Installing the software

Ensure that your PC meets all specifications presented in the site preparation document and is

operating under a 32-bit version of Windows™. It is advised to close down other applications on your PC

while installing the TD-100 software.

Locate the TD-100 software CD supplied in the shipping kit. Insert the CD into the appropriate

compartment of the PC and follow the instructions on the screen. Alternatively, open Windows Explorer

(located under Start > Programs) and access the CD drive. Open the TD-100 folder and scroll down the

content list to Setup.exe. Selecting this option will initiate installation of the software. Proceed as

directed.

Once the TD-100 software has been loaded onto your PC you can access the program from Start >

Programs.

2.1 System Configuration

There are several TD-100 instrument configurations available. The instrument configuration will

determine how the software needs to be configured. Inside the front door of the TD-100 Autosampler is

a label which should be used to determine the system configuration.

TD-100 installed options identification label

Please tick installed options as appropriate

���� Automated re-collection

���� MFC 1 - Split/Re-collection

���� MFC 2 - Trap

Internal Standard addition accessory

Refer to this information when configuring the software.

2.2 Running the software

2.2.1 Configuring the software

With the TD-100 switched off, start the software either by double clicking the TD-100 icon placed on the

desktop or from Start > Programs.

When the TD-100 is not detected choose Select Options?Select Options?Select Options?Select Options?

Page 20: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 17

This will bring up the Configuration page.

For TD-100 there are two available instrument configurations available in the drop down menu; TD-100

No Re-collection and TD-100 Automated Re-collection. TD-100 No Re-collection should be selected if

the Automated re-collection option is NOT ticked on the TD-100 configuration label.

TD-100 installed options identification label

Please tick installed options as appropriate

Automated re-collection

���� MFC 1 - Split/Re-collection

���� MFC 2 - Trap

Internal Standard addition accessory

Page 21: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 18

TD-100 Automated Re-collection should be selected if the Automated re-collection option is ticked on

the TD-100 configuration label.

TD-100 installed options identification label

Please tick installed options as appropriate

���� Automated re-collection

���� MFC 1 - Split/Re-collection

���� MFC 2 - Trap

Internal Standard addition accessory

Page 22: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 19

Ensure that the correct PC communication ports are selected.

NoteNoteNoteNote: the correct port will depend on the PC setup

Change the default Communications ports if necessary by clicking on the down arrow and then clicking

on the correct Com port in the list.

If an Internal Standard accessory is fitted it should also be enabled in the configuration page.

Page 23: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 20

TD-100 installed options identification label

Please tick installed options as appropriate

���� Automated re-collection

���� MFC 1 - Split/Re-collection

���� MFC 2 - Trap

Internal Standard addition accessory

Having setup the instrument configuration click OK, and when prompted by the following message

choose Yes to close the software allowing any changes to take effect.

2.2.2 Detecting the TD-100

Ensure that the gas supplies to the system - especially the dry air or nitrogen purge gas used for TD-100

valve actuation and purging the cold trap box - are on.

Having checked the above, switch TD-100 on using the switch located on the back panel of TD-100 and

Page 24: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 21

re-start the TD-100 software.

If communications between the PC and TD-100 are established successfully the software will open.

If for some reason, communication is not established an ‘Instrument not Detected’ box appears with a

number of options, Cycle Power and retry? Run simulation? Select options? and Exit?.

Choose Select options. This will take you to Configuration tab of Options and change the

communications ports if necessary..

Close down TD-100 software, switch the power to TD-100 off, wait ten seconds and then switch TD-100

back on again.

Re-access the TD-100 program on the PC as described above.

2.3 Setting up installed MFC(s)

For MFC installation instructions see Appendix A.

Upon first start-up of the software the following MFC configuration message will be displayed.

Click ‘Yes’ to go through to the Gas Flow tab to configure the MFCs following the procedure set out

below.

Page 25: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 22

Click ‘Scan for Flow Controllers’.

With 1 MFC setup the designation as

shown.

Note:Note:Note:Note: The full range value does not

need to be entered as this will be

configured automatically.

With 2 MFCs (default for the TD-100

advanced) setup the designation as

shown.

NoteNoteNoteNote: The full range value does not

need to be entered as this will be

configured automatically.

2.4 GC interface logic

The GC interface logic options can be found under the GC InterfaceGC InterfaceGC InterfaceGC Interface tab.

Check the correct operation of the GC ready signal. If the signal appears inverted i.e. software reports

GC not ready when GC is ready, then change the GC Ready (In) logic.

Page 26: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 23

Page 27: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 24

3 TD-100 Electronic Gas Control – instrument specific

guides

3.1 TD-100 Electronic Gas Control with the Agilent 6890GC

NoteNoteNoteNote: o complete this installation the following 6890 parts are recommended. They can be ordered

directly from Agilent.

Part no. Description Quantity

G2131-80500 Replacement 2 port EPC block 1

1390-1022 Screw 1

However, if these parts are not available at installation both the carrier gas and septum purge lines

running between the EPC module and the Split/Splitless injector must be cut and extended using 1/8”

unions and green PEEK tubing in order to facilitate connection to TD-100.

NoteNoteNoteNote: When installing onto an existing 6890GC the firmware on the GC must be A.03.08 or later (for A-

series) or N.04.09 (for N-series). To check the firmware version, use the keyboard on the GC,

press > Options > Diagnostics > Instrument Status and then scroll down to ‘Version’ where you

will see the version of the firmware running on the instrument.

3.1.1 Procedure

1) Remove the top rear cover on the GC to provide access to the EPC module that is going to be used to

regulate the carrier gas supply to TD-100.

Removing the rear panel to access EPC modules

2a) Remove the 3 port EPC block assembly from the split/splitless EPC module.

Fit the replacement EPC block assembly, and route the 1/16” tubing through the rear panel of the GC

and towards the TD-100.

Page 28: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 25

Location of EPC module with the 2 port EPC block fitted

2b) If a replacement EPC block assembly is not available, the septum purge and carrier gas lines must

be cut, extended, using 1/8” brass unions and 1/16” green PEEK tubing, and routed through the rear

panel of the GC towards TD-100. The lengths of PEEK tubing used should be no longer than required to

reach the TD-100.

Carrier and septum purge lines extended with unions and PEEK tubing

The stainless steel tubing for the carrier supply line is 1/16” while the septum purge line is 2 mm and

will require a different ferrule when connected to the 1/8” brass union.

(NoteNoteNoteNote: two 1/8” brass unions, 3 m of 1/16” green PEEK tubing and 1/8” graphitized vespel ferrules,

including one for 2 mm tubing, are all contained within the Agilent GC Installation Kit.)

3) Connect the carrier gas supply line to the ‘Carrier Inlet’ port of TD-100 and connect the septum

purge line to the ‘Pressure Outlet’ port.

3.1.2 6890 inlet settings

Before the column pressure value is set, the MODE of operation of the inlet has to be configured.

For correct control of TD-100 the mode must be set only to SPLITLESS, and not split.

In addition the splitless purge time value must be set to 999.99 minutes DIRECTLY through the

keyboard of the GC (6890 firmware revision A.03.08 or later required). This value cannot be set using

the Chemstation software as the number is rounded to 1000.0 minutes. Setting this value forces the

split/splitless inlet to remain permanently in the splitless mode.

The purge flow value is irrelevant. Once the split/splitless inlet has been configured as described the

column head pressure can be setup to give the desired flow rate/ velocity down the column.

1/8” Unions allowing

extension of gas lines

Replaced 2 port

EPC block

Routed tubing to

TD-100

Standard 3 port

EPC block

Lines cut

here

Page 29: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 26

3.2 TD-100 Electronic Gas Control with the Agilent 7890GC

NoteNoteNoteNote: To complete this installation the following 7890 parts are recommended. They can be ordered

directly from Agilent.

Part no. Description Quantity

G3430-60011 Replacement hoodlum block assemblies 1

However, if these parts are not available at installation both the carrier gas and septum purge lines

running between the EPC module and the Split/Splitless injector must be cut and extended using 1/8”

unions and green PEEK tubing in order to facilitate connection to TD-100.

3.2.1 Procedure

1) Remove the top rear cover on the GC to provide access to the EPC module that is going to be used to

regulate the carrier gas supply to TD-100.

2a) Remove the two hoodlum block assemblies, for the carrier supply and septum purge lines, from the

split/splitless EPC module.

Location of fixing screws for bulkhead connectors

Fit the two replacement hoodlum block assemblies, and route the 1/16” tubing on both of these

assemblies through the rear panel of the GC and towards the TD-100.

Note: the carrier gas supply line has a 1/8” termination at the TD end while the septum purge line

terminates in 1/16” tubing.

2b) If replacement hoodlum block assemblies are not available, the septum purge and carrier gas lines

must be cut, extended, using 1/8” brass unions and 1/16” green PEEK tubing, and routed through the

rear panel of the GC towards TD-100. The lengths of PEEK tubing used should be no longer than

required to reach the TD-100.

Carrier and septum purge lines extended with unions and PEEK tubing

EPC Module viewed

from above

EPC Module

viewed from

rear

Septum Purge

Connector

Lines cut here 1/8” Unions allowing

extension of gas lines

Carrier Supply

Connector

Page 30: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 27

The stainless steel tubing for the carrier supply line is 1/16” while the septum purge line is 2mm and

will require a different ferrule when connected to the 1/8” brass union.

(Note: two 1/8” brass unions, 3 m of 1/16” green PEEK tubing and 1/8” graphitized vespel ferrules,

including one for 2 mm tubing, are all contained within the TD-100 shipping kit.)

3) Connect the carrier gas supply line to the ‘Carrier Inlet’ port of TD-100 and connect the septum

purge line to the ‘Pressure Outlet’ port.

3.2.2 7890 Inlet Setup

There are several settings that must be implemented on the 7890 whenever the split/splitless EPC

module is controlling the carrier gas supply to the TD-100 thermal desorber. In the Inlet settings the

mode must be set to ‘Splitless’, the purge time must be set to exactly ‘999.99’ minutes and Gas saver

should be set to ‘Off’. In the Instrument Configuration menu the Auto prep run should be set to ‘Off’.

Having completed this, the system should now be ready to use.

3.2.3 7890 Inlet Settings

If the pressure is not correctly controlled after inputting the above settings then turn ‘Auto prep run’ to

‘On’.

Page 31: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 28

3.3 TD-100 Electronic Gas Control with the Shimadzu QP 2010

(Plus) GCMS fitted with AFC

NoteNoteNoteNote: To complete this installation the specialised GM - 1/16” fittings must be ordered directly from

Shimadzu prior to the installation. The parts required are as follows:

Part no. Description Quantity

221-35003 Shimadzu fitting GM-1.6L 1

221-35004-91 Shimadzu fitting GM-1.6L 1

228-16000-84 Ferrules 4

228-16002-84 Nuts 2

3.3.1 Gas connections between TD-100 and the Shimadzu AFC

1) The Shimadzu GC is fitted with two molecular sieve filters, one in the carrier gas line and one in the

septum purge line. Remove the filter from the carrier gas line and replace with the Shimadzu fitting GM-

1.6L. The filter on the septum purge line should remain in place to prevent back diffusion of air/water.

Shimadzu fitting on

carrier gas line

Molecular Sieve filter on

septum purge line

Two lengths of PEEK tubing

connecting to TD-100

TDTDTDTD----100100100100

Page 32: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 29

2) The carrier gas out line from the Shimadzu AFC (“CG Out”) is now connected to the “Carrier Inlet”

port on the back of TD-100 using a 1 m length of green PEEK tubing.

3) The “Pressure Outlet” from the back of TD-100 is now connected - with a similar length of green

PEEK tubing - to the molecular sieve filter on the septum purge line (and hence to the (“SP In”)) on the

Shimadzu AFC. It is important that the two lengths of green PEEK tubing are the same length. Carrier

gas is supplied and regulated upstream of the system in the supply line and it’s pressure is monitored

downstream of the system in the septum purge line.

3.3.2 Configuring the Shimadzu GC software

1) The inlet Split Mode on the Shimadzu GC must be set to Direct.

i. To access the Split Mode press the “FLOW” key on the GC keypad.

NoteNoteNoteNote: In the case of GCs with two or more AFC units it may be necessary to press the “FLOW” key

repeatedly until the correct one is displayed.

ii. Move the flashing cursor over the “Split Mode” (type using the arrow keys on the GC keypad).

iii. Using the right arrow key, select “DIRECT” mode as shown, and press the “ENTER” key on the GC

keypad.

Setting Inlet Mode as Direct on Shimadzu GC

2) The AFC module allows read out of the total flow supplied to system - this is ideal for setting split and

trap / desorb flows and for troubleshooting leaks.

i. To view the total flow supplied press the “FLOW” key on the GC keypad as above.

ii. The Total Flow in ml/min is always displayed (above the Split Mode) as shown above.

3) The septum purge flow should be set to a flow rate of around 5 ml/min

i. To set the septum purge flow display the “FLOW” screen as described above.

ii. Press the button to the right of the “PF3” key (below the screen) as shown below. This displays a

second set of commands for the PF keys.

Extending the PF commands

Page 33: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 30

iii. Select the purge command on “PF3” and enter the desired value. A flow rate of 2-5 ml/min is

recommended.

Setting the septum purge flow value

3.3.3 Flow stability

AFC works with both impeded (traps containing porous polymer e.g. Tenax) and un-impeded (traps only

containing carbon sorbents) traps, with no carrier gas instability generally noted (even with 90-100 psi

supplied to the inlet of the AFC system). However should any instability in the carrier gas occur it may be

necessary to dampen the reaction of the AFC controller.

i. Press the “FUNC” key on the GC keypad.

ii. Enter “7”, enter “5”, enter “1201”, enter “1”. This will display the “Flow Adjust” page.

iii. For instruments with two or more AFC units, press the “PF2” button to select the correct flow unit.

iv. Move the cursor to the “I Time” entry in the “Flow” column.

v. Change the value from “1” to “4” then press the “Enter” key.

Damping the AFC reaction

Page 34: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 31

3.3.4 Ready / Start connections between TD-100 and Shimadzu GC

The remote cable connections between TD-100 and the GC are shown below. The TD-100 - GC cable

(SERUTD-5095) has a 25-pin D-type connector at the TD-100 end and free wires at the GC end. Four

wires need to be inserted into the connection block at the rear of the QP2010 as shown below.

Ready / Start connections between TD-100 and QP 2010

When connected to the Shimadzu GCMS system the TD-100 ready (in) logic must be set to Open as

shown below. To set the logic select view>options from the drop down menus to open the options

dialogue box.

Setting the TD-100 GC Ready (in) logic to Open

Page 35: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 32

3.4 TD-100 Electronic Gas Control with the Thermo DPFC

A TD-100 can be integrated into the DPFC system of a Thermo Electron GC. The DPFC module regulates

the carrier gas supply upstream of the TD-100 and its pressure is monitored downstream of the system

in the septum purge line of the DPFC module. The split line of the DPFC module remains unused.

3.4.1 Procedure

1) The carrier gas supply and the septum purge lines that normally run from the DPFC to the

split/splitless injector need to be cut and fitted with 1/8” brass unions (note: the carrier line requires

use of graphitised vespel ferrule with the 2 mm hole – part number Z-0402).

2) Extend these two lines with the length of green PEEK tubing so that they can reach the back panel of

TD-100. The carrier gas line from the DPFC module should be connected to the TD-100 ‘Carrier Inlet’

and the septum purge line should be connected to the ‘Pressure Outlet’.

Schematic of connections made between DPFC module and TD-100

3.4.2 Configuring the Thermo GC software

Having completed the physical connection between the Thermo Focus or Trace GC and TD-100 the next

step is to configure the inlet settings.

1) The GC must be set up for back pressure regulation. To do this the DPFC must be removed and the

switches set in the correct position for purge & trap. Setting up this configuration is described in the kit

adapter for the purge and trap concentrator system installation guide.

2) The inlet that is being used to control the TD-100 carrier gas supply should then be set to operate in

split mode but the split flow should be set to zero (off).

3) The only other important consideration is to ensure that the pressure of the carrier gas supplied to

the DPFC module is only 10-15 psi greater than the pressure that the DPFC module is required to supply

to TD-100.

DPFC DPFC DPFC DPFC

ModuleModuleModuleModule

Septum

Purge

Carrier

TDTDTDTD----100100100100

Pressure

Outlet

Carrier Inlet

Page 36: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 33

4 Installation qualification

4.1 Testing the ECC setup

Once the connections and inlet settings have been correctly set up it is extremely important that the

following test of the ECC carrier gas supply is carried out:

During the desorption of the TD-100 cold trap the total flow and pressure supply of the ECC module

should be monitored. Both pressure and flow readings should be stable.

If there is any variation in the pressure reading of the ECC module, or if the flow fluctuates by more

3mL/min, this indicates that the setup is not optimized. In such circumstances the reproducibility of

analysis is likely to be poor.

ECC instability during trap desorption

Such fluctuations can normally be avoided by regulating the carrier gas supply to the ECC module to 10

psi above the greatest pressure required by the TD-100 in the course of a run, see below. This is almost

always effective in removing any instability when using a constant pressure method, but may not be

effective for constant flow methods.

Correct Setup of ECC Supply

TDTDTDTD----100100100100

GCGCGCGC EPCEPCEPCEPC

Carrier

Septum Purge

Carrier Supply to GC

25 psi

Total Trap Desorption

Standby

Time

30

Inlet

Pressure 15 psi

Total Flow 30 ml/min

Pressure Differential = 25 – 15 = 10psi

Page 37: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 34

In situations where the pressure and/or flow are unstable despite the appropriate regulation of the

carrier gas supply there is an alternative configuration that should be employed. (Refer to section 1.6.21

for further information and installation instructions.)

Page 38: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 35

Appendix A - Installing MFC(s)

Remove front right cover

Remove right hand cover by loosening

the 3 screws highlighted on the front

and 3 (not shown) on the rear then

removing the 2 screws on the side of

the cover.

Remove MFC mounting plate by

removing 2 x screws highlighted then

lifting the plate out.

MFCs should be mounted on the plate

in the positions shown, depending on

the number of MFC present.

1 MFC1 MFC1 MFC1 MFC

Once installed re-attach the MFC

mounting plate.

Split /

Re-coll.

T’d

MFC plate

Page 39: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

QUI-1066 TD-100 Installation Manual 36

2 MFCs2 MFCs2 MFCs2 MFCs

Once installed re-attach the MFC

mounting plate.

The MFCs should be placed upstream

of their respective needle valve (and

t’d if necessary).

For example:

The needle valve(s) should then be

opened so a flow just above the

maximum of the MFC can be

achieved.

The connection(s) to the control PCB

must be made.

The green RJ45 cable should be

routed through and plugged into the

top right of the control PCB as shown.

Page 40: TD-100 Installation Manual - Ingeniería Analítica · QUI-1066 TD-100 Installation Manual 1 Preface This manual provides detailed instructions on the installation of the TD-100

ERROR: stackunderflow

OFFENDING COMMAND: ~

STACK: