tg71 and tg 114, monitor units calculations
TRANSCRIPT
Samir Laoui, Ph.D. 2/08/2016
Basic Principles: Equivalent Square
Field o Calculation data is tabulated according to square field size
o Sterling’s Formula
S = 4xAREA/PERIMETER
o The equivalent square is used to look up dosimetric parameters related to the primary collimator settings
L
W
S
S
Basic Principles: Effective Square
Field
o Field defining blocks (or multi-leaf collimators) further modify the field
o The effective square concept allows one to determine a square field size that is “effectively” equal to the blocked field as relates to dosimetry
12% Blocking
Basic Principles: SAD Setup – SSD Setup
oThe PDD is the primary parameter used to
calculate dose for SSD setups oThe TPR is the primary parameter used to
calculate dose for SAD setups
Basic Principles: Sc
o Collimator scatter factor
o Sc: In-air output ratio, The ratio
of the output (i.e., energy fluence)
in air for a given field size to that
for the reference field size
o NOTE: Sc is a function of the
field size defined in the treatment
head, not the final field size that
reaches the patient Scatter off the collimators
Basic Principles: Sp
• Phantom scatter factor
• Quantifies the relationship
between the field size on the
patients surface and the dose
resulting from scatter within the
patient
• NOTE: Sp is a function of the
field size as defined on the
patient, not the field size as
defined in the treatment head
Field size on the patient surface
TG-71: Introduction
Introduction
o The photon beam normalization depth, d0, is
different from reference depth
oProtocol recommends using normalization and reference
depth to be equal to 10 cm
oThis decreases the difference in programmed MUs
when moving patients from one machine to another
(assuming machines’ characteristics are the same?)
o For electron beams, normalization depth is taken to
be the depth of maximum dose along the central
axis
Monitor Unit equations: Photons
oMay be performed using either TPR (isocentric) or
PDD (nonisocentric)
oTPR (isocentric)
oPDD (nonisocentric)
Field size determination
o Equations (1), (2) and (3) are field size dependent
o Field equivalent 4.A/P estimation
o for Sc: 3 methods to determine equivalent square
field size Equivalent square of jaw settings
1. Equivalent square of jaw settings
2. Point’s eye view (PEV) model of collimating jaws
3. PEV model of all collimators
o For Sp, the field size is proportional to that
incident on the patient
Field size determination
oWedge factor
oPhysical Wedges: Investigations have determined that
the WF for rectangular fields is closely approximated by
the WF of the equivalent square for both external and
internal wedges, regardless of orientation
oNonphysical wedges: WF represent the fractional
change in dose per MU at the calculation depth after the
treatment field is completed
Radiological depth determination
o To correct for heterogeneities within the patient, a
correction factor
oMethods to determine CF
1. Method 1: uses water-equivalent or radiological depth,
deff
2. Method 2: The power law TAR
Electrons
o Se: Electron output factor is defined in TG-70
Electron calculation at extended SSDs o The effect of treatment distance not equal to the
standard SSD can be accounted for in two ways, as
described in the AAPM Task Group 70 report
At dmax
Field size determination
o However, the above techniques do not predict accurately the
output factors or percentage depth-dose values for irregularly
shaped electron fields lateral buildup ratios LBR
Arc therapy
Example: Brain Case
78.7 MU
TG 71 Summary
oNormalization depth of 10 cm be chosen for MU
calculations for photons
o For electrons, normalization depth is the depth of
maximum dose along the central axis for the same
field in water phantom
oMU calculations for patients prior to first
treatment, if not, then before third fraction, 10% of
the dose has been delivered, whichever occurs first
o Patient Safety is the ultimate goal for doing MU
verification
o In the modern era, the purpose and methodology
for the MU verification have come into question
oGuidelines are needed to help the physicist set
clinically reasonable action levels for agreement
Scope
o Recommendations on how to perform verification
of MU in modern clinic
oActions levels
oDoes not pertain to IMRT
Objective of MUV
o IAEA: “Results provided by the TPS need to be checked,
and this should include verification by manual calculation
of the treatment time and dose to the selected point. This
verification should be part of the QA programme.”
o ICRP 2009: “A simple secondary MU calculation,
independent from the TPS, has proven for many years to be
an efficient tool for prevention of major errors in dose
delivery.”
Errors o Errors can be classified into random and
systematic errors
o Random: Incorrect energy, wrong dose
o Systematic: defect in part of the calculation
procedure
Limitations of MUV
o The MUV is not a check of the accuracy of the
entire calculated dose distribution, but rather to a
single point
o In addition to the MUV, the physicist’s plan review
should confirm that the dose, beam energy,
fractionation, and dose point location are
consistent with the physician’s prescription
Aspect of a MUV program
o Each center performing after-hours emergency
treatments should have a policy detailing how
planning, including the MUV, is to proceed when a
physicist is not immediately available for plan and
calculation review
Recommended Methods of MUV
o It is recommended that MUV point need not
coincide with the plan normalization point
o Point placement in regions containing a high dose
gradient should be avoided; as well, a calculation
point should not be placed near a field edge, >2 cm
o The factors that are likely to introduce the greatest
uncertainty in the verification calculation are
blocked field scatter, patient contour, and patient
heterogeneity effects
Recommended Methods of MUV
oManual calculational methods using look-up tables
provide the basis for most MUV
oAll radiation therapy clinics should maintain an
accurate set of dosimetry tables
i. Reference dose rate
ii. Output ratios
iii. Attenuation factors
iv. OAR
v. Depth dependence
Action levels and remedial actions
Remediation of MU discrepancies
o Review parameters in both calculations
oAll factors are taken into account
oDosimetric point placement
UCI EXAMPLES
Shoulder Case
The end
“The price of Safety in Radiotherapy is
an Eternal Vigilance”