thermal methods of analysis - personal homepage...
TRANSCRIPT
Thermal Analysis
• Thermal analysis is a branch of materials
science where the properties of materials
are studied as they change with
temperature.
Thermoanalytical Methods
• Thermogravimetry (TGA): recording of sample weight changes during controlled temperature programs (dynamical or isothermal)
• Differential thermoanalysis (DTA): recording of temperature difference between sample and reference crucible during controlled temperature programs. After calibration the heat flux into/out of the sample (reaction or phase transition enthalpy) can be calculated.
• Differential scanning calorimetry (DSC):
– heat flux DSC: measurement of temperature difference between sample and reference similar to DTA
– power compensated DSC: sample and reference are kept at the same temperature, the difference in the necessary heating power is recorded.
Thermogravimetry Analysis (TGA)
The mass of a sample is measured
as a function of the time and temperature.
A highly sensitive balance monitors the weight
loss of a sample (verses time and temperature).
Thermal Analysis
• TGA can provide information about
physical phenomena like vaporization,
sublimation, absorption, and desorption.
• Also chemical phenomena including
dehydration, decomposition and oxidation
and reduction reactions.
Material Characteristics
• From decomposition and degradation
patterns.
• Organic content and inorganic content
(ash).
• Especially useful for the study of polymers,
thermosets, plastics, coatings, paints
Material Characteristics
• TGA requires a high degree of precision in
measuring mass change (accurate
balance), programmable temperature and
temperature change over time.
• Therefore need precision balance,
programmable furnace with constant or
programmable heating rate.
Mass Change (TGA)
Thermogram
Polytetrafluoroethylene
(teflon)
Calcium oxalate monohydrate (12.51 mg)
CO2 (39.75%,3.85 mg)
H2O, (11.68%, 1.46 mg)
CO (18.32%,2.92 mg)
Weight decreases as lose H2O, CO and CO2
TGA
Instrumentation
Main components:
•Sensitive analytical balance
•Furnace
•Temperature programming unit
•Recorder
•Sample holder (thermally isolated)
Thermogravimetry (TGA)
Evolved Gas Analysis (EGA)
TGA-FT-IR• A Thermogravimetric Analyzer
(TGA) combined with an Infrared Spectrometer (TG-IR).
• Heating a sample on the TGA, will release volatile materials or generate combustion components as it burns.
• The components can be identified in the IR cell.
• This technique is most useful when the evolved gases are known small compounds such as water, carbon dioxide or common solvents which have characteristic IR spectra.
TGAIR
Evolved Gas Analysis (EGA)
TGA-MS• The combination of a TGA with
a MS allows you to detect very low levels of impurities in real time.
• Heating a sample on the TGA, the sample will release volatile materials or generate combustion components as it burns.
• These gases are transferred to the MS. This technique is most useful when the evolved gases or breakdown products are known in advance but are few in number.Mass-spec TGA
Calcium oxalate monohydrate (12.51 mg)
CO2 (39.75%,3.85 mg)
Mass = 44
H2O, (11.68%, 1.46 mg) mass = 18
CO (18.32%,2.92 mg), mass = 28
Weight decreases as lose H2O, CO and CO2
TGA-MS?
Evolved Gas Analysis (EGA)
TGA-GC-MS• Heating a sample on the TGA
causes gases to be released.
• These gases are then transferred to the GC where the components can be separated and the peaks identified by the MS.
• Because of its ability to detect very low levels of material in complex mixtures, the TG-GC/MS is a powerful tool for quality control, safety, and product development.GC-MS
TGA
Differential Thermal analysis (DTA)
1. DTA involves heating or cooling a test sample and an inert reference under identical conditions, while recording any temperature difference between the sample and reference.
2. This differential temperature is then plotted against time, or against temperature. Changes in the sample which lead to the absorption or evolution of heat can be detected relative to the inert reference.
• 3. DTA can be used to study thermal properties and phase changes (fusion, vaporization, sublimation, desorption, some chemical reactions) which do not lead to a change in enthalpy.
Differential thermal analysis (DTA),
In analytical chemistry, a technique for identifying and quantitatively analyzing
the chemical composition of substances by observing the thermal behaviour of
a sample as it is heated. The technique is based on the fact that as a
substance is heated, it undergoes reactions and phase changes that involve
absorption or emission of heat. In DTA the temperature of the test material is
measured relative to that of an adjacent inert material. A thermocouple
imbedded in the test piece and another in the inert material are connected so
that any differential temperatures generated during the heating cycle are
graphically recorded as a series of peaks on a moving chart. The amount of
heat involved and temperature at which these changes take place are
characteristic of individual elements or compounds; identification of a
substance, therefore, is accomplished by comparing DTA curves obtained
from the unknown with those of known elements or compounds. Moreover, the
amount of a substance present in a composite sample will be related to the
area under the peaks in the graph, and this amount can be determined by
comparing the area of a characteristic peak with areas from a series of
standard samples analyzed under identical conditions. The DTA technique is
widely used for identifying minerals and mineral mixtures.
Differential Thermal Analysis (DTA)
• A DTA consists of a sample holder comprising thermocouples, sample containers and a ceramic or metallic block; a furnace; a temperature programmer; and a recording system.
• The key feature is the existence of two thermocouples connected to a voltmeter. One thermocouple is placed in an inert material such as Al2O3, while the other is placed in a sample of the material under study.
Differential Scanning Calorimetry (DSC)
• DSC analysis operates by determining how much energy is required to heat a pan containing a sample compared to a reference pan.
• If the sample undergoes an endothermic (absorbs heat) or exothermic (gives off heat) reaction, the sample pan will require more or less energy to increase the temperature at the same rate as the reference pan.
• By measuring the difference between the energy applied to each heater as the temperature is increased, the energy consumed or released by the sample can be determined.
Differential Scanning
Calorimeter
Differential Scanning Calorimetry is the most widely used
thermal analysis technique in the world. DSC measures the heat
flow in materials and provides information about phase changes,
such as amorphous and crystalline transitions (glass transition
temperature, melting point, crystallization point, crystallinity) as
well as chemical changes (aging, degradation, chemical
reactions, and thermal history). The data can be used to identify
materials, determine specific heat capacity of materials, degree
of cure, and to characterize the materials for their thermal
performance
Differential Scanning Calorimetry (DSC)
• DSC can determine phase transitions of
materials. These phase transitions include:
• Melting Point (Tm)
• Glass Transition Temperature (Tg)
• Energy Absorbed (Hm) while melting
• Crystallization Point (Tc)
• Energy Released (Hc) during crystallization
∆H(enthalpy) = K (constant) x A(area)
Crystalization
Temperature
Heat
Flow
(or
Flux)
Melting
Glass transition
(Softening)
DSC• The result of a DSC experiment is a curve of heat flux versus
temperature or versus time.
• There are two different conventions: exothermic reactions in the sample shown with a positive or negative peak, depending on the kind of technology used in the experiment.
• This curve can be used to calculate enthalpies of transitions. This is done by integrating the peak corresponding to a given transition.
• ∆H is the enthalpy of transition, K is the calorimetric constant, and A is the area under the curve. ∆H = K*A
• The calorimetric constant will vary from instrument to instrument, and can be determined by analyzing a well-characterized sample with known enthalpies of transition
Measures heat directly
Enthalpy
• Enthalpy is a measure of the total energy of a thermodynamic system.
• The unit of measurement for enthalpy in the International System of Units(SI) is the joule, but other historical, conventional units are still in use, such as the British thermal unit and the calorie.
• The total enthalpy, H, of a system cannot be measured directly. Enthalpy itself is a thermodynamic potential, so in order to measure the enthalpy of a system, we measure is the change in enthalpy, ΔH.
• The change ΔH is positive in endothermic reactions, and negative in heat-releasing exothermic processes.
• For processes under constant pressure, ΔH is equal to the change in the internal energy of the system, plus the work that the system has done on its surroundings.
Two types of methods used in DSC:
i) Power-compensated DSC
Sample and reference temperatures are kept equal
by heating with separate heaters while external
heat is increased or decreased linearly.
ii) Heat flux DSC
Difference in heat flux into the reference and sample is
measured as sample temperature is increased or
decreased linearly
Heat flux DSC
- One furnace system
- Sample and reference cups (Al) are placed on elevated
platforms on thermoelectric disc.
- Heat flows to the sample and reference
- Differential heat flow is monitored by chromel
thermocouples
- Differential heat flow directly proportional to output
DSC analysis to determine the spreading
qualities and wear qualities of lipsticks
Temperature
0-50 +50
Heat flux
DSC
Heat flow
Power-compensated DSC
- Two independent furnaces
- Furnaces are small, allow about 1 g of sample, for
rapid heating, cooling and equilibration.
- Furnaces are embedded in a large heat sink
- Sample and reference holders are equipped with Pt
resistance thermometers to continuously monitor
temperature (T) values.
Two control circuits, one for average T control, the other
for differential T control.
- Acetaminophen exists in three polymorphic forms, but only two of which
can be readily isolated. Form I (monoclinic) is the commercially marketed
version.
- Form II(orthorhombic), however, has distinct advantages in its tableting
properties due to its plasticity, which enables direct compression without
binders .
- The crystal structures of both forms are known. Form II has proven difficult
to make on a large scale,
- Form II can be made on a small scale, either by slow cooling of the Form I
melt or by recrystallization.
- Characterization of the two forms has revealed that Form I is more stable
than Form II under ambient temperature conditions.
Polymorphic Forms
Simultaneous TGA/DTA
• Simultaneous TGA/DTA measures both heat flow and weight changes (TGA) in a material as a function of temperature or time in a controlled atmosphere.
• Simultaneous measurement of these two material properties can help interpretation of the results.
• The complementary information obtained allows differentiation between endothermic and exothermic events with no associated weight loss (e.g., melting and crystallization) and those that involve a weight loss (e.g., degradation).
Some important application areas:
-Compositional analysis (quality control)
-Thermal and oxidative stability (food expirary dating)
-Product life time
-Filler content of materials (quality control)
-Moisture and volatile content
-Effects of reactive atmospheres on materials
-Decomposition kinetics (protein stability)
Mettler TGA/DSC
• The TGA/DSC 3+ uses a TGA balance from the
worldwide leader in weighing technology with position-
independent weighing, automatic internal calibration
weights, a wide measurement range, the best minimum
weight performance and the highest weighing accuracy
and precision.
• It allows users to analyze a wide variety of sample types
up to 1600 °C.
• A complementary DSC heat flow sensor simultaneously
detects thermal events such as melting and
crystallization in addition to providing accurate and
precise transition temperatures.