Topic 8. Gamma Camera (II)

• Basic Performance Characteristics• Detector Limitations• Design and Performance Characteristics of

Parallel-Hole Collimators• Performance Characteristics of Converging,

Diverging, and Pinhole Collimators• Measurements of Gamma Camera

Performance.

Page 2: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Intrinsic Spatial Resolution Limits• Intrinsic spatial resolution refers to the limit of spatial

resolution achievable by the detector and the electronics

• Intrinsic spatial resolution is limited primarily by two factors: multiple scattering and statistical fluctuation in the distribution of light photons (the later is the main factor).

• Intrinsic resolution depends on detector crystal thickness and γ ray energy. A thinner crystal is used for Anger camera to achieve better spatial resolution.

• Larger number of PM tubes and improvement of light collection efficiency result in better resolution.

Page 3: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Intrinsic Resolution vs Photon Energy

Page 4: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Intrinsic Resolution vs Crystal Thickness

Page 5: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 6: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Detection Efficiency

• The crystal thickness of NaI(Tl) in Anger camera is smaller (6-12mm) than that of a well counter (2-5cm).

• Anger camera is designed for optimal detection of γ ray energies of 100-200keV.

Page 7: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 8: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Energy Resolution

Page 9: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Problems at High Counting Rates

• Pulse pile-up is the problem at high count rate and it results in count loss and image distortion.

• Count loss depends on the whole energy spectrum but the apparent dead-time depends on the window fraction. Scatter causes narrower of the window fraction therefore longer dead-time.

Page 10: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 11: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Problems at High Counting Rates

• Pulse pile-up causes image distortion. Two scattered event may be added to form a photo-peak which produces a location between the two scattered events.

Page 12: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Problems at High Counting Rates

• The general effect of the pulse pile-up is to cause a loss of image contrast and details

Page 13: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Problems at High Counting Rates

• Pulse pile-up rejection circuit can be used to improve the pulse pile-up but it will reduce the maximum achievable count rate

• Deadtime can be improved by shortening the effective charge integration time for the output signal from the PM tube but it will reduce the amount of light collected by the PM tube, therefore, degrade the intrinsic resolution.

Page 14: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Problems at High Counting Rates

• Count rate performance should be one of the important factors for Anger cameras

Page 15: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Pulse Pile-up Correction

Page 16: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Image Non-Linearity

• Non-linearity refers to that a straight line object appear as curved line image

• Non-linearity occurs when the X and Y positional signals do not change linearly with the displacement distance.

• “Pincushion” distortion is an inward “bowing” image and “Barrel” distortion is an outward “bowing” image

• A PM tube may have high light collection efficiency in the centre which may result in a “pincushion” in the centre and “barrel” between PM tubes. This kind of images could results in a “wavy” line pattern.

Page 17: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 18: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 19: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Nonlinearity Example

Page 20: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Image Non-Uniformity

• Non-Uniformity refers to the intensity variation from a uniform flood source

• The causes could be the non-uniform detection efficiency (small differences in pulse height spectrum) and non-linearity of PM tubes (more server) or instrument malfunctions.

• “Edge packing” refers to the bright ring around the edge of the image. It is caused by the internal light reflection of the detector crystal. “Edge packing” is usually masked and the useful field of view is smaller than the actual detector size

Page 21: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 22: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 23: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 24: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Non-Uniformity Correction Techniques

• Cosmetic approach:– Adjust individual PM tube gains (compensate for the detection

efficiency difference)

– “Correction factors” matrix. Test image is normalised and the correction factor for each matrix element is used to add or subtract from the image.

• Advanced approach:– A set of microprocessors are used to store correction matrices for

regional differences in pulse height spectra and for position distortion.

– Images are corrected on event by event basis (on fly).

Page 25: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Non-uniformity Correction Example

Page 26: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Uniformity Profiles

Page 27: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Gamma Camera Tuning

Page 28: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Design and Performance Characteristics of Parallel Hole

Collimators

• Basic limitations in collimator performance• Septal thickness• Geometry of collimator holes• System resolution

Page 29: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Basic Limitation in Collimator Performance

• Collimator is a limiting factor in camera system performance.

• Collimator resolution refers to the sharpness or details of the γ ray image projected onto the detector --- worse than the intrinsic resolution.

• Collimator efficiency refers to the fraction of γ rays that pass through it -- a few percent or less.

Page 30: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Septal Thickness• Septal thickness is designed to prevent γ

rays from penetrating from one hole to another (allow less than 5% to pass through).

• Septal thickness should be as small as possible in order to gain maximum efficiency

• High atomic number and density material should be used (lead is the choice)

Page 31: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Septal Thickness

• The required thickness t=2dw/(l-w) where (t/w)=(2d+t)/l if θ is small

• If 5% penetration is allowed, t>[6d/μ]/[l-(3/ μ)] because e-μw<0.05 therefore μw>3

Page 32: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

General Comments• The septal thickness depends on the γ ray energies to be

used because μ depends on the γ ray energy.

• Energy ranges in nuclear medicine are often classified as: low energy <150keV; medium energy <400keV and high energy <1Mev.

• Low energy collimators are generally fragile because they are only a few tenth of mm.

• Low energy collimators are used whenever possible to obtain maximum collimator efficiency (“foggy” background may be superimposed on the image for high energy).

Page 33: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Geometry of Collimator Holes

• The performance of a collimator is affected by its shape, length and diameter (round or hexagonal are the best)

• Spatial resolution and detection efficiency are the two important performance parameters

• Collimator spatial resolution is defined as the FWHM of the radiation profile from a point or line source of radiation projected by the collimator onto the detector.

Page 34: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Collimator Resolution

• A parallel holes collimator resolution is given by Rc=[d(le+b)]/le where le

=l-2μ-1 is the effective collimator length (taking into account of the γ ray penetration)

Page 35: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Collimator Resolution (2)

• Spatial resolution of a parallel collimator increases (worse) as the distance between the collimator and the source increased

Page 36: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Collimator Efficiency

• The detection efficiency is given by g=K2(d/le)2[d2/(d+t)2] where K is a constant that depends on the holes shape. The first part is the geometric factor (solid angle subtended by a collimator hole) and the second part is the fraction that is not covered by the septa (the area ratio of the holes and total detector).

Page 37: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Collimator Efficiency

• Collimator efficiency for a source in air is independent of source-to-collimator distance b.

Page 38: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Collimator Efficiency

• Invariance of collimator efficiency with source to collimator distance applies to point sources, line sources and uniform sheet source in air with parallel hole collimators.

Page 39: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Resolution and Efficiency

• The relationship between the resolution and efficiency is gœRc

2

• For a given septal thickness, collimator resolution is improved only at the expense of decreased collimator efficiency and vice versa.

Page 40: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 41: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

System Resolution

• System resolution depends on a number of factors such as, scattering, septal penetration, intrinsic resolution and collimator resolution with the collimator and intrinsic resolutions are the main factors.

• System resolution (consider intrinsic and collimator only) can be expressed as Rs

2=Ri2+Rc

2

• System resolution is determined primarily by collimator resolution.

Page 42: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 43: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Performance Characteristics of Converging, Diverging and Pinhole

Collimators

Page 44: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Performance Characteristics of Converging Collimator

cos

21

'

21

cos

1)(

1

'

2

2

2

22

2

'

'

'

'

l

bff

tdd

l

dK

l

l

llR

e

e

e

e

ec

e

g

f

bd

Page 45: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Performance Characteristics of Diverging Collimator

blf

lf

e

g

f

bd

tdd

l

dK

lllR e

e

ec

2

22

2

'

'

'

'

21

cos

1

Page 46: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Performance Characteristics of Pinhole Collimator

2/tan2

16

1

2

3

cos

dd

g

l

bl

d

bd

dR

e

e

ec

Page 47: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Performance Characteristics of Converging, Diverging and Pinhole

Collimators

Page 48: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Some Comments

• Resolution always is best with the source as close as possible to the collimator

• For point source in air, the efficiency increases as the source to collimator distance increase with converging collimator (maximum at focus point) and decreases for diverging and pinhole collimators. No change for flood source as long as the source cover the entire detector.

• Diverging, converging and pinhole collimators may be useful for the change of field of view but the image distortion caused by the magnification with depth may be a problem.

Page 49: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance
Page 50: Topic 8. Gamma Camera (II) Basic Performance Characteristics Detector Limitations Design and Performance Characteristics of Parallel-Hole Collimators Performance

Measurements of Gamma Camera Performance

• Intrinsic Resolution

• System Resolution

• Spatial Linearity

• Uniformity

• Counting Rate Performance

• Energy Resolution

• System Sensitivity