1 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
JEM FPEF OV 21409
Fluid Physics Experiment Facility (FPEF) Overview
Notice : This technical data is furnished on the condition that it will be used by and disclosed to the receiving Cooperating agency and its contractors and sub contractors only for the purposes of fulfilling the cooperating agency’s responsibilities under the Space Station Intergovernmental Agreement (IGA) and Memorandum of Understanding (MOU). It shall not be used for any other purpose, nor disclosed or retransferred to any other entity or government without prior written permission of the Japan Aerospace Exploration Agency (JAXA).
Instructor: Yuki Kawai Takashi Yamasaki
2 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Explain what kind of phenomenon can be observed by FPEF.
Explain the functions of Experiment Cell.
Explain how Gas Ar (GAr) will be supplied to/wasted from Experiment Cell.
List three observation equipments in FPEF.
Explain the function of Lamps.
Explain two functions of FPEF Control Equipment.
Purpose of FPEF
FPEF Interfaces
Functions and components of FPEF
Objectives
After this lesson the trainee will be able to:
3 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Contents:
- Purpose of FPEF
- FPEF Interfaces
- FPEF Major Architecture
- FPEF Major Components
- Interfaces in FPEF
- Acronyms
Appendix:
- FPEF Specifications
- Connectors/QDs Interfaces in FPEF
Page
4
5
6
7
16
18
20
21
Lesson Outline
4 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
The main purpose of FPEF is to observe and measure Marangoni convection in the liquid bridge.
LiquidBridge
Lower
Higher
Heating disk (High Temp.)
Surface Tension
Cooling disk (Low Temp.)
FPEFFPEF
Experiment CellExperiment Cell
Purpose of the FPEF
Marangoni convection is driven by the difference of the liquid surface tension, and surface tension is higher at low temperature part than at high temperature part.
FPEF is used in combination with Experiment Cell, and provides following capabilities.
Form the liquid bridge and generate Marangoni convection. Observe and/or Measure Marangoni convection.
Accomodate Experiment Cell. Observe/Measure Marangoni convection in three ways.
Control and measure temperature of Experiment Cell. Provide resources to Experiment Cell.
Provide telescience operation and sequential control automation.
Experiment CellExperiment Cell
FPEFFPEF
5 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Water
Power
Communication
Avionics Air
Gas Exhaust
Gas Supply
Video
Ryutai rack
PPDB
AAA/SD
MTLModerate Temperature Water Loop
PDH/JCP
JEM System
CGSE
PDU-b2
GAr
PLT
Ground
(Command and Telemetry)
(Command, Telemetry and Parameter file)
Waste Gas Line
FPEF
Note: Interfaces among equipments in Ryutai Rack are not shown in this figure.
PLT is not used in nominal operations.
Avionics Air
Command/Telemetry
Parameter File
IPUVideo
Ryutai rack
FPEF
IPU
PPDB AAA/SD
Main Power
Experiment CellWaste Gas
FPEF Interfaces
Silicone Filter
Silicone Filter
6 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Experiment Cell
3-D Camera
Front Cover
Experiment cover
Control Equipment
Lamps and Fiber Optic Cables
Motor Driver
Observation equipments
Front covers
FPEFFPEF
Cooling Water and Gas Line
IR Imager
2-D Camera
FPEF Major Architecture
Silicone Filter
Video Switcher
Experiment Cell
FPEF
Experiment cover
Ryutai rack
Silicone Filter
7 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Experiment Cell
FPEF Major Components
Experiment Cell contains experiment sample.
Experiment Cell has the heating disk and the cooling disk, forms liquid
bridge, and creates temperature gradient between the disks.
The design of Experiment Cell depends on the experiment purpose.
- has unique observation/measurement system. - has unique disk size. e.g. 30, 50 mm in diameter - has unique sample. e.g. Silicone oil, small particles
Crew member assemble/disassemble Experiment Cell.
Crew exchange Experiment Cell/sample cassette.
Installed to FPEF
Install to Install to FPEFFPEF
Experiment Cell has Ar/waste gas, cooling water,
power, communication and video interfaces with FP
EF.
FPEF
Experiment Cell
Mechanical interfaces
8 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Experiment Cell
Experiment Cell (Cont’d)
Heating disk
Connecters and QDs with FPEF
Liquid bridge
Install to Install to FPEFFPEF
FPEF Major Components (Cont’d)
Cooling disk
Gas QD
Water QD
Cooling diskHeating disk
9 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
3-D Camera
Mirror CCD Camera
2-D Camera
IR Imager
Observation equipments
FPEF Major Components (Cont’d)
CCD Camera
focusdiaphragm
CCD Camera (three)
Aux light
Crew will install IR Imager into FPEF
IR Imager Power Harness
IR Imager Video Harness
(FPEF GM)
10 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
FPEF Major Components (Cont’d)
Observation equipments (Cont’d)
3-D Camera
IR Imager
2-D Camera
Heating disk side (transparent)
Cooling disk side
Liquid bridge
observes small particles in liquid bridge by three CCD cameras. 3-D Camera also can move backward and forward according to depth of field.
observes whole shape of liquid bridge.
measures surface temperature of liquid bridge.
Mirror Case
11 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Lampslight liquid bridge inside of Experiment Cell through two Fiber Optic Cables.
Video Switcherreceives video images from 2-D Camera, 3-D Camera, IR Imager, and Experiment Cell. Some of these images are switched and transferred to IPU.
Fiber Optic Cable (two)
FPEF interfaceExperiment Cell
interface
Motor Driver
drives 3-D Camera and cooling disk in Experiment Cell.
FPEF Major Components (Cont’d)
IPU interfaces (5 channel) (output)Experiment Cell video interfaces(input)
IR Imager video interface(input)
Lamp 1
Lamp 2
Limit Switch
When Experiment Front Cover is attached, Experiment can be conducted.
12 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Control Equipment
Control Equipment provides following functions
- receives power from Ryutai rack PPDB and supplies power to components in FPEF and Experiment Cell.
- Communicates with Payload Data Handling Unit (PDH) and controls components in FPEF and Experiment Cell.
FPEF Major Components (Cont’d)
Control EquipmentExperiment Cell power/ signal interfaces
1553B interface IR Imager power interface Ground use only
Control Equipment
13 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Front Cover
Stowage ConfigurationStowage Configuration
Operations ConfigurationOperations Configuration
IR Imager
IR Imager
Experiment Cell
Experiment cover
FPEF Major Components (Cont’d)
Front covers
Front CoverExperiment Cover Plate
Experiment Cover Body
- is used during launch and storage. - is used during experiment. - has 1553B cable pass through and Silicon Filter interface.
Silicon Filter interface
1553B cable pass through
Velcro
14 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Cooling Water and Gas Line
FPEF Major Components (Cont’d)
- FPEF has cooling water line to cool down the Control Equipment and Experiment Cell.
- FPEF provides Gas Ar supply and waste gas vent functions to replace the air inside of Experiment Cell by Ar gas.
Fluid interfaces
Fluid interfaces
Cooling water interfaces for Experiment Cell
Gas Ar supply interface to Experiment Cell
Waste gas interface from Experiment Cell through Silicone Filter
15 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
FPEF Major Components (Cont’d)
Silicone Filter
- removes silicone oil vapor from waste gas from Experiment Cell.
- is installed in the right side of FPEF.
Silicone Filter FPEF Silicone Filter UnitFPEF Silicone Hose AFPEF Silicone Hose BFPEF Silicone Cover AFPEF Silicone Cover B
Silicone Cover BSilicone Cover A
Ryutai Rack (upper part)
Silicone Hose A
Silicone Filter Unit
Silicone Hose A QD
Rail for Silicone Cover
Silicone Filter Unit
Silicone Hose B
FPEF Silicone Filter
Experiment Cell
GAr IN
GAr OUT
Bolts
Experiment cover
16 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
3-D Camera 3
3-D Camera MotorForword Lim Sw1
Forword Lim Sw2
Backword Lim Sw1
Backword Lim Sw2
3-D Camera 1
Limit Switch
3-D Camera 2
2-D Camera
3-D CameraController
Strobe Controller
Lamp 1
Lamp 2
2-D CameraController
Photochromic CameraController
Sync
Video
Sync
Video
Power/Control
Sync
IR IMAGER VIDEO J 08
IPU INTERFACE
VIDEO SYNC J 09
Control
Power
Power
Control
Video
Power
Control
Video
Power
Control
Video
Video
Video
IR IMAGER POWER J 15
ControlPowerPower Control Power Power/Control/Moniter
VIDEO J 620POWER J 200
Auxlight
Video Switcher
IR ImagerControl Equipment
Moniter
1553B J 301
GROUND USE
ONLY
TRANSFORMER POWER J 10
DIGITAL I/O J 11
TEMP SENSOR
J 12
HEATER POWER J 13
EXP CELL VIDEO J 310
POWER J 01
VIDEO J 03
RS-232C J 02
Motor Driver
Rear Side
Front Side
Interfaces in FPEFPPDB
Experiment Cell ( uses some of these interfaces)
Fiber Optic Cable
Fiber Optic Cable
(to PDH)
17 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
P
Pt
ExperimentCell
Symbol NAME Symbol NAME
Manual Valve QD
Solenoid Valve Orifice
Regulator Pt temp sensor
Relief Valve CP Cold Plate
Filter Accumulator
Pressure Sensor
F
S
S
M
F
Control Equipment
SiliconeFilter Unit
S
M
F
P
Pt
GF2
GPS1
GRV2 GRV1
GPR1GOF1
Low Pressure
GAr Line
Middle Pressure
GAr Line
Cold plate WOF1
WOF2
PT1
Experiment Cell FPEF
SV1
GF1
SV2
WV1WA1
OD1/4”
OD1/4”
OD1/4”
OD1/4”
FPEF Rack
GAr Supply Line
Ar:0.5 ~ 0.79 MPa[73 ~ 114psia]
Waste Gas Line
INCooling Water :0.69 Mpa(100[psia])(MAX)
OUT
Cooling Water
Differential Pressure : 33+0
-1[kPa]
COOLANT WATER SUPPLY
COOLANT WATER RETURN
Interfaces in FPEF (Cont’d)
GAr IN
GAr OUT
18 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Acronyms
AAACCDFPEFGArIPU IRMTLPDHPDUPLTPPDBSD
Avionics Air AssemblyCharge Coupled DeviceFluid Physics Experiment FacilityGas ArgonImage Processing UnitInfraredModerate Temperature LoopPayload Data Handling UnitPower Distribution UnitPayload Laptop TerminalPayload Power Distribution BoxSmoke Detector
19 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Appendix
- FPEF Specifications - Connector Interface Part
20 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Spectral response; 8~ 14μm
Measurement range; 0~ 100℃Time resolution; 30flames/sec
IR Imager (Surface Temperature Measurement Equipment)
CCD Camera (1ch)
Lamp source; 15W×2, 60Hz
2-D Camera (Liquid Bridge Form Observation)
CCD Camera (3ch)
Lamp source; 15W×2, 60Hz
3-D Camera (3-D Flow Rate Measurement)
1 / 4 Payload Drawer: W482×H621×D660mmDimensions
SpecificationsItem
FPEF Specifications
User Interface -Analog/digital interface
-Digital communication interface
-Power interface
-Video output interface
-Gas/ cooling water interface
-Motor drive interface
21 Training Purposes OnlyJEM FPEF OV 21409JMU-064205 (Ver. 1.3)
Front view of FPEF
POWER J01(Exp Cell Motor I/F)
LAMP 1, LAMP 2
VIDEO J620(Exp Cell CCD Camera I/F)
COOLANT WATER SUPPLY
GAr IN
1553B J301
IPU INTERFACE
TRANSFOMER POWER J10,DIGITAL I/O J11,TEMP SENSOR J12,HEATER POWER J13
COOLANT WATER RETURNGAr OUT
IR IMAGER VIDEO J08
IR IMAGER POWER J15GROUND USE ONLY
VIDEO SYNCJ09
EXP CELL VIDEO J310
Connector and QDs Interfaces in FPEF