APPLICATION NOTE 21735 Keywords Industrial, trace analysis, GC-MS, Trace 1310 GC, ultra-low column bleed, ultra-inert (UI) column technology, silphenylene column chemistry, WCOT, TraceGOLD TG5-SilMS, trimethyl phosphate, 1-propionic acid, 4-picolinic acid, 1,2-pentanediol Author Aaron Lamb, Thermo Fisher Scientific, Runcorn, UK Application benefits • Thermo Scientific TraceGOLD silphenylene (SilMS)-based columns demonstrate 80% lower bleed compared to premium low bleed competitor columns. • Column bleed variation across batches is 80% less than competitor ultra inert low bleed columns. • Superior column inertness was demonstrated for the TraceGOLD TG-5 SilMS column compared to the market-leading UI column with a 50% improvement in inertness measured by the asymmetry value of 1,2-propanediol, a probe to measure silanol activity. Goal To compare the Thermo Scientific TraceGOLD TG-5SilMS column against the market-leading ultra-inert (UI) columns to determine relative inertness and column bleed across multiple batches. Introduction Trace analysis requires GC columns with excellent inertness, low column bleed, and excellent column-to-column reproducibility so that lower limits of detection and quantitation can reliably be achieved. Columns that require lengthy conditioning procedures to achieve stable baselines cause extended Ultra-inert low bleed GC columns with advanced silphenylene polymer technology

Upload: others

Post on 22-Sep-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Ultra-Inert Low Bleed GC Columns with Advanced Silphenylene … · 2017. 11. 23. · H˚CC H ˚ HO H˚C OH H˚CC H˚ ... Data processing The Thermo Scientific ... Batch reproducibility

APPLICATION NOTE 21735

KeywordsIndustrial, trace analysis, GC-MS, Trace 1310 GC, ultra-low column bleed, ultra-inert (UI) column technology, silphenylene column chemistry, WCOT, TraceGOLD TG5-SilMS, trimethyl phosphate, 1-propionic acid, 4-picolinic acid, 1,2-pentanediol

AuthorAaron Lamb, Thermo Fisher Scientific, Runcorn, UK

Application benefits • Thermo Scientific™ TraceGOLD™ silphenylene (SilMS)-based columns demonstrate 80% lower bleed compared to premium low bleed competitor columns.

• Column bleed variation across batches is 80% less than competitor ultra inert low bleed columns.

• Superior column inertness was demonstrated for the TraceGOLD TG-5 SilMS column compared to the market-leading UI column with a 50% improvement in inertness measured by the asymmetry value of 1,2-propanediol, a probe to measure silanol activity.

Goal To compare the Thermo Scientific TraceGOLD TG-5SilMS column against the market-leading ultra-inert (UI) columns to determine relative inertness and column bleed across multiple batches.

Introduction Trace analysis requires GC columns with excellent inertness, low column bleed, and excellent column-to-column reproducibility so that lower limits of detection and quantitation can reliably be achieved. Columns that require lengthy conditioning procedures to achieve stable baselines cause extended

Ultra-inert low bleed GC columns with advanced silphenylene polymer technology

Page 2: Ultra-Inert Low Bleed GC Columns with Advanced Silphenylene … · 2017. 11. 23. · H˚CC H ˚ HO H˚C OH H˚CC H˚ ... Data processing The Thermo Scientific ... Batch reproducibility

2

instrument down time and this reduces productivity. These problems are compounded with high bleed columns, where source contamination is accelerated and results in the need for more frequent preventative MS maintenance due to siloxane build up.

Many GC columns on the market contain original ‘pendent style’ polymer chemistry where functionality (usually mixtures of phenyl and methyl groups) are appended off the silicon as R groups (Figure 1).

This type of column chemistry is particularly prone to high bleed at elevated oven temperatures, which is caused by the breakdown of the column stationary phase. During the breakdown process the formation of characteristic cyclosiloxanes fragments occurs via a chemical reaction known as ‘back-biting’ (Figure 2) and is greatly accelerated by the presence of oxygen in the carrier gas. High column bleed is an issue to chromatographers as it can result in poor method sensitivity, reduced column

Figure 2. Back-biting reaction that occurs during siloxane polymer bleed process.

Figure 1. Original pendent and silphenylene based polymer chemistry.

Si

n

SiO

SiO

OSi

Si

Si

CH³

CH³CH³

CH³CH³H³C

H³C

H³CH³C

H³C

O

Si

O

SiO

Si

O

Si

O

CH³CH³

CH³

CH³

CH³

CH³

CH¹

H¹C

H³C

H³C

H³C

CH³

Si

Si OH +O

CH³

CH³

OSiO

O

Si

OO

Si

Si

HO

Si

O

SiOO

O

Si

CH³

CH³CH³

CH³

CH³

CH³

H³C

H³C

Si OH +O

O

OSi

O

CH³

CH³

Si

CH³

CH³CH³

OSi OH +O

CH³

CH³

Si

CH³

CH³

OSi

CH³

CH³

O

SiO

HO

CH³

Si

CH³

CH³

OSi

CH³

CH³

O Si

HO

CH³

OSi

CH³

CH³

O

SiO

CH³

CH³

Si

CH³

CH³

OOSi

CH³

CH³

O SiO

CH³

CH³

Si

CH³

CH³

OHOSi

CH³

CH³

m/z = 207 m/z = 281 m/z = 355

Polydimethylsiloxane polymer

SiSi

CH³H³C

H³C

CH³ CH³

O

O

O

Si

Si

Si

CH³

CH³

CH³ CH³

CH³

H³C

H³C

SiO

O

OSiSi

Si

O

CH³

CH³

CH³

H³C

H³CH³C

H³C

Figure 3. Advanced silphenylene based polymer chemistry.

O

NH2NH

NHCl

Cl

NH2

NH2

Cl

Cl

N N

CH3

SiO

CH3

O O

CH3

Si

CH3

R2

Si

R2

x y

lifetime, and higher frequency of MS maintenance. TraceGOLD SilMS columns however are based on newer advanced silphenylene polymer chemistry (Figure 3).

The TraceGOLD SilMS column contains modern silphenylene-based chemistry incorporating the phenyl functionality fully integrated into the backbone of the siloxane polymer itself, which provides extra rigidity for enhanced chemical and thermal stability. This results in significantly reduced column bleed and extended useful column lifetime. Thermo Scientific TraceGOLD columns are also preconditioned prior to leaving the factory, which means that out of the box you can expect an exceptionally clean column that requires minimal conditioning prior to sample analysis.

The GC column accounts for greater than 90% of the exposed surface that the sample encounters during analysis; therefore, free silanol sites can cause poor peak shape, compound degradation, and adsorption. Improved inertness in the TraceGOLD SilMS column

Page 3: Ultra-Inert Low Bleed GC Columns with Advanced Silphenylene … · 2017. 11. 23. · H˚CC H ˚ HO H˚C OH H˚CC H˚ ... Data processing The Thermo Scientific ... Batch reproducibility

3

range is achieved by sterically hindering the access of active compounds to the fused silica surface, thereby reducing the likelihood of analyte interaction. The high efficiency and reproducibility of the TraceGOLD manufacturing process provides extensive surface coverage and maximizes the column inertness consistently between columns.

To compare the inertness of TraceGOLD columns, a demanding test mix containing strongly interacting compounds was chosen. The results for the TraceGOLD TG-5 SilMS column and leading competitor UI low bleed columns were then compared directly using the same method. The mix contained probes that measure stationary phase basicity, acidity, and silanol activity. These probes are more indicative of interaction than those used in the more familiar Grob style test mixes as they are more strongly interacting with the stationary phase (Figures 4 and 5). The individual functions of each probe are listed in Table 1.

H3C CH3

HOH3C

OHH3C CH3

NH2

N

CH3

CH3

O

HO

CH3

CH3

P OH3CH3C

OH

OH

Weak basicity probe

2,6-dimethyl phenol

Weak acidity probe

2,6-dimethyl aniline

Weak silanol probe1,2-pentanediol

Figure 4. Structures for Grob style mix probes.

H3C CH3

HOH3C

OHH3C CH3

NH2

N

CH3

CH3

O

HO

CH3

CH3

P OH3CH3C

OH

OH

Strong basicity probe

1,2-propionic acid

Strong acidity probe 4-picolinic

acid

Strong acidity probe

trimethyl phophate

Strong silanol probe 1,2-pentanediol

Figure 5. Structures for Grob mix weak probes and Thermo Scientific UI GC-Mix strong probes.

Probe ng on column

Column functional test

1 1-propionic acid 2.0 Basicity2 1-octene 1.0 Polarity3 n-octane 1.0 Hydrocarbon marker4 4-picoline 2.0 Acidity5 n-nonane 2.0 Hydrocarbon marker6 trimethly phosphate 2.0 Acidity7 1,2-pentanediol 2.0 Silanol8 n-propylbenzene 2.0 Hydrocarbon marker9 1-heptanol 2.0 Silanol10 3-octanone 2.0 Polarity11 n-decane 2.0 Hydrocarbon marker

Table 1. Components and functions for Thermo Scientific UI GC-Mix probes.

In addition to the problems associated with phase activity, the swapping and reconditioning of columns can be a problem for laboratories that wish to maximize their instrument up time and increase lab productivity. In general, this process is lengthy and may introduce contamination to the system that subsequently must be cleaned. The Thermo Scientific Thermo Scientific™ ISQ™, TSQ™, and Orbitrap™ series of GC-MS systems use a vent-free source removal tool that permits source removal during GC column conditioning, reducing the need for preventative source maintenance due to excessive siloxane contamination.

In this application note, the TraceGOLD 5 SilMS column is compared to leading competitor UI low bleed columns in terms of performance. The tests examined were out-of-the-box bleed, absolute column bleed, column-to-column variation in bleed, and inertness using a fit-for-purpose column test mix.

Experimental Consumables Column: TraceGOLD TG-5 SilMS, 30 m × 0.25 mm × 0.25 µm (P/N 26096-1420)

Injection septum: Thermo Scientific™ BTO™, 11 mm (P/N 31303215-BP)

Injection liner: Thermo Scientific™ LinerGOLD™, Straight liner with glass wool (P/N 453A0164-UI)

Page 4: Ultra-Inert Low Bleed GC Columns with Advanced Silphenylene … · 2017. 11. 23. · H˚CC H ˚ HO H˚C OH H˚CC H˚ ... Data processing The Thermo Scientific ... Batch reproducibility

Column ferrules: Thermo Scientific™ 15% Graphite/85% Vespel (P/N 290VA191)

Injection syringe: Thermo Scientific™ 10 µL fixed needle syringe (P/N 365D0291)

Solvent: Fisher Scientific™ HPLC grade dichloromethane (P/N 10784941)

Thermo Scientific™ Virtuoso™ vial, clear 2 mL kit with septa and cap (P/N 60180-VT402)

Thermo Scientific™ Virtuoso™ vial identification system (P/N 60180-VT100)

Sample pretreatment and preparation10 mg/mL primary solution preparation Individual 10.0 mg/mL solutions of probes 1–11 were prepared by weighing 100.0 mg into separate 10.0 mL volumetric flasks and diluting to volume with dichloromethane.

Mixed Thermo Scientific UI GC-Mix probe preparation The test probe was prepared as outlined in Table 2.

Separation conditions Instrumentation • Thermo Scientific™ TRACE™ 1310 gas chromatograph (P/N 14800320)

• Thermo Scientific™ TriPlus RSH™ Autosampler (P/N 1 R77010-0100)

• Thermo Scientific™ Instant Connect Electron Flame Ionization Detector (FID) (P/N 19070001FS)

• Thermo Scientific™ ISQ™ LT Single Quadrupole GC-MS

Probe SolutionUsed

(mg/mL)

Volume pipetted

(mL)

Final volume

(mL)

Final Concentration

(µg/mL)

Solvent

1 1-propionic acid 2.0 2

100

200

Dichloromethane

2 1-octene 1.0 1 1003 n-octane 1.0 1 1004 4-picoline 2.0 2 2005 n-nonane 2.0 2 2006 trimethly phosphate 2.0 2 2007 1,2-pentanediol 2.0 2 2008 n-propylbenzene 2.0 2 2009 1-heptanol 2.0 2 20010 3-octanone 2.0 2 20011 n-decane 2.0 2 200

Table 2. Preparation of the Thermo Scientific UI GC-Mix.

4

Method conditions Instrumentation: TRACE 1310 GC with TriPlus RSH Autosampler Carrier gas: Helium Split flow: 175.0 mL/min Split ratio: 125:1 Column flow: 1.40 mL/min Injector type: Split/Splitless Injection volume: 2 µL Injector mode: Split, constant flow Injector temperature: 220 °C Detector type: Flame ionization detector (FID) Detector temperature: 250 °C Detector air flow: 350 mL/min Detector hydrogen flow: 35 mL/min Detector nitrogen flow: 40 mL/min

Oven temperature profiles: Inertness testing 65 °C (15 min hold) Bleed testing 40 °C to 320 °C (4 °C/min, 30 min hold) 320 °C to 325 °C (4°C/min, 30 min hold) 325°C to 330 °C (4°C/min, 30 min hold)

Data processing The Thermo Scientific™ Chromeleon™ Chromatography Data System (CDS) was used for data acquisition and analysis.

Page 5: Ultra-Inert Low Bleed GC Columns with Advanced Silphenylene … · 2017. 11. 23. · H˚CC H ˚ HO H˚C OH H˚CC H˚ ... Data processing The Thermo Scientific ... Batch reproducibility

Results and discussionColumn conditioning The chromatogram below (Figure 6) shows the differences in the bleed profile for the market-leading low bleed UI column versus the TraceGOLD TG-5 SilMS column. The black trace for the TraceGOLD TG-5 SilMS column shows negligible contamination peaks (compared with the competitor blue trace peaks 1-3), and the column quickly equilibrates to a point of a stable baseline. The TraceGOLD range of columns come preconditioned, meaning that they are quicker to condition, which reduces instrument downtime.

pA

1.7 12.5

Min

25.0 37.5 50.0 62.5 75.0 87.5 100.0 112.5 131.43.28

5.00

6.25

7.50

8.75

10.00

11.25

12.50

13.75

15.00

2

1

3

More contamination peaksfor Competitor X

Lower bleed Quick tostabilize For TG-5 SilMS

Competitor X

TG-5 SilMS

Figure 6. Chromatogram showing column bleed for TraceGOLD TG5-SilMS column and the leading competitor low bleed UI column following solvent injection directly out of the box.

Column bleed The TraceGOLD SilMS columns exhibited exceptionally low bleed when compared to competitor columns following preconditioning as stated by the manufacturer (Figure 7). The bleed was calculated by subtracting the bleed at 40 °C from the bleed at 320 °C and 325 °C.

???

2 10

40°C 320°C

0.2 pA

0.7 pA

1.1 pA

0.2 pA

0.9 pA

1.4 pACompetitor Y

Competitor X

TG-5 SilMS

325°C

Min

20 30 40 50 60 70 80 90 100 100110 1203.00

3.40

3.80

4.20

4.60

5.00

5.40

5.806.00

Figure 7. Overlay chromatogram showing column bleed for TraceGOLD TG5-SilMS column and leading competitor low bleed UI columns.

Batch reproducibility versus the competition The TraceGOLD SilMS columns exhibited excellent batch-to-batch reproducibility compared to leading competitor low bleed columns (Figure 8). Another major advantage of using the TraceGOLD TG-5 SilMS column over some competitors’ columns is that its top isothermal temperature is 330 °C, which means methods can be made faster, increasing lab productivity.

pA

0 10

Min

20 30 40 50 60 70 80 90 100 110 120 130

6.50

3.21

3.50

4.00

4.50

5.00

5.50

6.00

Competitor Y

More bleed variability

Figure 8. Overlaid column bleed for three lots from TraceGOLD TG5-SilMS, competitor X, and competitor Y columns.

5

pA

-1.9 12.5

Min

2.97

3.00

Competitor X

More bleed variability

50.025.0 37.5 62.5 75.0 87.5 100.0 112.5 125.0 137.5 150.0159.4

6.50

6.00

5.50

5.00

4.50

3.50

4.00

pA

-1.9 12.5

Min

3.11

3.50

TraceGOLD TG-5 SiIMS

Much less bleed variability

50.025.0 37.5 62.5 75.0 87.5 100.0 112.5 125.0 137.5 150.0

6.50

6.00

5.50

5.00

4.00

4.50

162.5 176.0

Page 6: Ultra-Inert Low Bleed GC Columns with Advanced Silphenylene … · 2017. 11. 23. · H˚CC H ˚ HO H˚C OH H˚CC H˚ ... Data processing The Thermo Scientific ... Batch reproducibility

The variation in column bleed for three separate lots of the TraceGOLD TG-5 SilMS, competitor X, and competitor Y low bleed columns were examined (Figure 9). The standard deviation for the TraceGOLD TG-5 SilMS columns was 0.02 pA, which demonstrates an adherence to tight controls in manufacture and quality control. Compared to the competitor columns, the variation in column bleed was 80% and 95% lower. The overall bleed was also found to be 80% lower than both competitors at 325 °C, which is the top isothermal temperature for these competitors. These are important factors as column-to-column performance is vitally important for generating robust and rugged analytical data.

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6Low variability between column Lots

∆ pA

320

Temperature °C

325

Competitor X Competitor Y TG-5 SilMS

Figure 9. Variation in column bleed for three separate lots of columns tested identically Inertness compared to the market leader UI column.

Figure 10 shows components 4, 6, and 7 are significantly improved (30, 10, and 50% improved asymmetry values) on the TraceGOLD TG-5 SilMS column versus competitor X. Probes 4 and 6 test the column’s acidity, whereas probe 7 is used to assess the column’s free silanol activity. These probes are an excellent differentiator for assessing column performance due to their high polarity and modes of interaction. The TraceGOLD TG-5 SilMS column shows minimal differences in peak tailing across the remaining components compared to the competitor, indicating that overall the TraceGOLD column has superior inertness.

pA

Min

25.0

2.0

10.0 1

2

10.0

20.0TG-5 SilMS Enhanced inertness

Competitor X

25.0

20.03

54

EP A

s =

3.1

6EP

As

= 2

.32

EP A

s =

6.5

7

EP A

s =

1.9

6

EP A

s =

5.9

5

EP A

s =

1.0

3

8

9 10 11

6

7

1

23

54

8

9 10 11

6

7

Figure 10. Chromatogram showing Thermo Scientific UI GC-Mix for TraceGOLD TG5-SilMS column and market-leading competitor column.

6

Page 7: Ultra-Inert Low Bleed GC Columns with Advanced Silphenylene … · 2017. 11. 23. · H˚CC H ˚ HO H˚C OH H˚CC H˚ ... Data processing The Thermo Scientific ... Batch reproducibility

Find out more at thermofisher.com/chromatography For Research Use Only. Not for use in diagnostic procedures. ©2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN21735-EN 1117S

Conclusions• Thermo Scientific silphenylene (SilMS)-based columns, such as the TraceGOLD TG-5 SilMS column, demonstrate up to 80% lower bleed compared to premium competitor low bleed columns.

• Column bleed variation across batches is 80% less than competitor ultra-inert low bleed columns.

• Superior column inertness was demonstrated for the TraceGOLD TG-5 SilMS column compared to the market-leading UI column with a 50% improvement in the asymmetry value in the case of the silanol probe 1,2-propanediol.