University of Groningen Endovascular approaches to complex aortic aneurysms de Niet, Arne DOI: 10.33612/diss.111895510 IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document version below. Document Version Publisher's PDF, also known as Version of record Publication date: 2020 Link to publication in University of Groningen/UMCG research database Citation for published version (APA): de Niet, A. (2020). Endovascular approaches to complex aortic aneurysms. [Groningen]: University of Groningen. https://doi.org/10.33612/diss.111895510 Copyright Other than for strictly personal use, it is not permitted to download or to forward/distribute the text or part of it without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license (like Creative Commons). Take-down policy If you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediately and investigate your claim. Downloaded from the University of Groningen/UMCG research database (Pure): http://www.rug.nl/research/portal. For technical reasons the number of authors shown on this cover page is limited to 10 maximum. Download date: 14-08-2020

Upload: others

Post on 10-Jul-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

University of Groningen

Endovascular approaches to complex aortic aneurysmsde Niet, Arne

DOI:10.33612/diss.111895510

IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite fromit. Please check the document version below.

Document VersionPublisher's PDF, also known as Version of record

Publication date:2020

Link to publication in University of Groningen/UMCG research database

Citation for published version (APA):de Niet, A. (2020). Endovascular approaches to complex aortic aneurysms. [Groningen]: University ofGroningen. https://doi.org/10.33612/diss.111895510

CopyrightOther than for strictly personal use, it is not permitted to download or to forward/distribute the text or part of it without the consent of theauthor(s) and/or copyright holder(s), unless the work is under an open content license (like Creative Commons).

Take-down policyIf you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediatelyand investigate your claim.

Downloaded from the University of Groningen/UMCG research database (Pure): http://www.rug.nl/research/portal. For technical reasons thenumber of authors shown on this cover page is limited to 10 maximum.

Download date: 14-08-2020

Page 2: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

CHAPTER

Summary and future perspectives

10

Page 3: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

Chap

ter

10

172

Page 4: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

10

Summary and future perspectives 173

The advocated treatment of abdominal aortic aneurysms (AAAs) has shifted from an open

surgical repair towards a total endovascular approach. After introduction of endovascular

aneurysm repair (EVAR) for infrarenal AAAs, endografts for juxtarenal AAAs, suprarenal

AAAs and thoraco-abdominal aortic aneurysms (TAAAs) have been developed to enable a

more cranial deployment of an endograft in a healthier part of the aorta while maintaining

visceral artery vascularization. This thesis focused on the clinical outcome of these devices,

designed for infrarenal and complex aneurysms, and the considerations of these endografts

in relation to patients’ anatomy.

Since their introduction, endografts for EVAR have been further developed and newly

designed endografts have been added meanwhile.1 Despite the fact that long-term follow-up

results of infrarenal endografts were not available yet, alterations were already made to these

endografts throughout the whole time period. In Chapter 2 an overview of currently available

endografts for infrarenal AAA repair and their mid- to long-term follow up results is shown.

Main variations between the endografts were the way of proximal fixation (infrarenal or

suprarenal), the use of either Z- or circular stents and the fabric of the endograft. In comparison

to first generation devices, currently available endografts are more flexible, have improve-

ments for a more controlled deployment, and are delivered in smaller sheaths.2,3 The Zenith

SP Flex® (Cook Medical), Incraft® (Cordis corp), Excluder® (Gore), Endurant™ (Medtronic)

and the Anaconda™ (Terumo aortic) have similar results for peri-operative outcome and early

reintervention rate. Although studies with these newer generation devices show good results,

there is still a 10-20% reintervention rate during a mean of three year follow-up, and as a

consequence close follow-up remains necessary. Important factors influencing outcome seem

to be the anatomical suitability, the adherence to the instructions for use, and the experience

with a certain endograft by the treating clinician. The constant change in designs and lack of

large long-term follow-up studies still demand careful conclusions on long-term outcome of

these endografts for EVAR.

Once the AAA extents towards or beyond the visceral arteries or there is an unfavorable

infrarenal neck, infrarenal sealing of EVAR might not be sufficient. To this context, the use

of fenestrated EVAR (FEVAR) allows a more cranial deployment of the endograft while

maintaining vascularization to the visceral arteries. In Chapter 3 an overview is given of

Page 5: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

Chap

ter

10

174

fenestrated endografts with special emphasis on clinical outcome. Most experience had been

gained with the custom-made Zenith® Fenestrated AAA Endovascular Graft (Cook Medical,

Bloomington, IN, USA). The custom-made Fenestrated Anaconda™ System (Terumo Aortic,

Inchinnan, Scotland, UK) was designed a few years later, and in contrast to the fenestrated

Cook device, the fenestrated Anaconda device has an unsupported proximal part enabling

wider options for placement of fenestrations. The off-the-shelf Endologix Ventana™

fenestrated graft (Endologix, Inc. Irvine, CA, USA) was applicable in cases with a specific

location of the superior mesenteric artery and the renal arteries preventing waiting time for

customization, and therefore could be used in emergency cases also. Early results however

were, unsatisfactory, urging the manufacturer to withdraw the device from the market. Only

recently the custom-made E-xtra Design Engineering endograft (JOTEC GmbH, Hechingen,

Germany) was introduced. Consequently, there was very little literature available on results

with this newly introduced device. The available data varied between the four endografts,

but procedural results including procedure time, used contrast volume and technical success

were very similar. The 30-day post-operative mortality varied between 0 to 9% independent

of the used endograft. Longer term follow-up series with FEVAR are mainly available with

the fenestrated Cook device. Only a few series, containing small sample sizes, with variable

follow-up periods are available with the other three devices making them unfeasible for

statistical comparison.

The life expectancy in the Western world is increasing and the number of cases with an AAA

at older age will most likely increase. Chapter 4 focuses on the clinical outcome of octogenari-

ans treated for complex AAAs with FEVAR and comparing those with non-octogenarians. No

statical difference was seen between non-octogenarians and octogenarians with an estimated

survival of 42% in the octogenarian group at five years. Eyeballing the survival curves, it seems

the octogenarians have a lower survival. Except renal function and age, no difference was seen

for other comorbidities between groups and the difference between groups might simply be

related to shorter survival at older age.4 However, survival is not the only relevant outcome

measure. Endograft specific variables such as reintervention rate, target vessel patency and

number of endoleaks during follow-up might be more suitable, and no differences were seen

between the two groups. Older age is considered more challenging because of more angulation

in aortic anatomy, but the current results and previously published results suggest no influence

Page 6: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

10

Summary and future perspectives 175

of age to endograft related outcome.5,6 Age itself should not be a reason to withhold treatment

with FEVAR, but a few questions still remain. FEVAR is an expensive treatment compared

to open surgical treatment, and a balance in cost-effectiveness, gained life years and gained

quality adjusted life years has yet to be settled.7

The commercially available fenestrated endografts from Cook Medical and Terumo Aortic

are both custom-made. To perfectly fit the endograft to the patient’s anatomy, dedicated

software with 3D-reconstructions of computed tomography angiography (CTA) is used. The

use of a central lumen line in the aorta and target vessels enables accurate determination

of the locations of proximal and distal landing zones and fenestrations. Deployment of the

endograft with stents in the fenestrations into the target vessels might, however, influences

native anatomy and thereby may lead to endograft related complications.8 Specifically in

FEVAR, the placement of a stent in a visceral artery could influence the tortuosity of the artery

and give rise to complications such as stenosis, occlusion, kinking or fracture of the placed

stent. In Chapter 5 the tortuosity index (TI) to measure the tortuosity of visceral arteries was

introduced and successfully tested for reproducibility. As discussed in Chapter 3, there are

some differences in used material and design for the Zenith® Fenestrated and the Fenestrated

Anaconda™ endografts. The Zenith® Fenestrated endograft consists of self-expandable

stainless steel Z-stents, while the Fenestrated Anaconda™ contains circular nitinol stents and

has, in contrast to the Zenith® Fenestrated endograft, an unsupported area for the fenestrati-

ons. Chapter 5 focused on the differences between these two endografts and the influence of

the specific endograft on native anatomy. Both fenestrated endografts clearly influenced the

anatomy of the aorta and the stented visceral arteries, but the Zenith® Fenestrated endograft

seemed to influence native anatomy more than the Fenestrated Anaconda™ endograft. In the

Zenith® Fenestrated endograft the angle of both visceral arteries moved more perpendicular.

It was postulated that the Z-stents were more rigid, therefore the docking of the stents in

the fenestrations affect native anatomy more than the unsupported part in the Fenestrated

Anaconda™. On the other hand, the diameter of the aorta at the level of the SMA in the

Zenith® Fenestrated endograft was reduced to the diameter of the endograft, while no such

decrease in aortic diameter was seen with the Fenestrated Anaconda™ endograft. Possibly the

blood pressure pushed the unsupported part against the aortic wall, while the endograft was

designed to be straight. Although this was not shown in this study, this might influence the

Page 7: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

Chap

ter

10

176

conformability of the endograft. Furthermore, the clock position of both the renal arteries

moved more anteriorly in the Zenith™ Fenestrated endograft. After releasing the Zenith®

Fenestrated endograft from the sheath, diameter reducing ties are still present so the location

of the endograft can be changed slightly. These diameter reducing ties are located on the back

of the endograft, and releasing of these ties was usually done after stenting of the fenestra-

tion. After releasing the diameter reducing ties the clock position of the stented fenestra-

tion changes, consequently pushing clock position slightly anteriorly. No large changes in

tortuosity indices for target vessels were seen, but this variable might be beneficial in the

prediction of target vessel related complications. Additional studies should be done to find the

cut-off value for TI change mandating more aggressive intervention or closer follow-up. In

this chapter, the changes in native anatomy were small and it is still unclear in what extent this

influenced clinical outcome. The measurements in this chapter were done in static images,

but might behave very differently during cardiac cycle. Therefore, it was suggested to study

anatomic changes with dynamic CTAs to find the true anatomic conformability of FEVAR.

The clinical applicability of the Fenestrated Anaconda™ endograft is shown in Chapter 6.

Results with this endograft showed a procedural type Ia endoleak of 8.6%, which mostly

disappeared without reintervention during further follow-up leading to 0.3% type Ia endoleak

at one year. These type Ia endoleaks enable persisting blood pressure into the aneurysmal sac,

and therefore the risk of rupture theoretically remains.9 However, the clinical consequen-

ces remain unknown, but seem limited because of the high patient and reintervention-free

survival, at least in the current study.

Chapter 7 focuses on mid-term results with the Fenestrated Anaconda™ endograft including

cases globally and specifically focused on procedural and post-operative endoleaks. In the

335 cases from 11 centers worldwide the applicability of the Fenestrated Anaconda™ for the

treatment of complex AAA was shown. Technical success, mid-term survival and reinterven-

tion rates were comparable to the results with the Zenith® Fenestrated endograft, as previously

reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%,

but also mostly spontaneously disappeared and no type Ia endoleak related aneurysm rupture

was observed. It was hypothesized that sealing of the proximal endoleak takes some time

because the proximal nitinol rings of the endograft fully expand over time. Probably more

Page 8: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

10

Summary and future perspectives 177

anatomical variables, such as oversizing, proximal landing zone diameter and angle, need to

be taken into account to predict a persisting, and clinical relevant, endoleak. Another issue

noted in this chapter was the decline of renal function over time. Renal function naturally

declines when age increases, but high administration of contrast agents and placement of

stents in target vessels, could increase the risk of renal function deterioration.12,13 This chapter

underlined that clinicians should pay special attention to patients with an already borderline

renal function in FEVAR.

Patients treated for an AAA remain in close follow-up. Among other risks, the risk after open

surgical repair is the development of a proximal para-anastomotic AAA. After EVAR possible

risks are the development of neck dilation, AAA growth, endoleaks, endograft migration,

and device failure. An open surgical repair in failed EVAR has a 30 day mortality risk of

3% and in para-anastomotic aneurysm a 14% 30 day post-operative mortality risk has been

described.14,15 In these cases, but also in cases with a primary AAA or TAAA with a narrow or

otherwise healthy distal aorta and not treated before, a fenestrated aortic cuff can be used.16

In Chapter 8 the global mid-term clinical outcomes of the Fenestrated Anaconda™ aortic

cuff are discussed. In this chapter, 29 cases of the 57 globally treated cases were analyzed.

The preoperative patient characteristics possibly made clinicians decide to withhold open

surgical repair and choose FEVAR. Furthermore, these cases were probably technically more

challenging, reflected by more fenestrations and cranial approaches compared to regular

FEVAR. The patient characteristics and complexity of anatomy probably resulted in rather

high mortality and reintervention rates within the early post-operative period. Once through

this early post-operative period survival up to two years remains relatively stable and no

aneurysm related deaths were seen, but the necessity for reinterventions remained an issue.

A reintervention free survival of 79% at two years was seen after primary FEVAR with the

Fenestrated Anaconda™ endograft, but after redo surgery with these fenestrated aortic cuffs

the reintervention free survival was 67%. Problems like failure of the primary endograft or

an endoleak between the primary endograft and the fenestrated cuff were seen, therefore a

complete relining of the endograft might be a more suitable option. The choice of a fenestrated

aortic cuff was a tailor-made option and possibly only for those cases where no other option is

available. Especially in these cases close follow-up will remain necessary.

Page 9: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

Chap

ter

10

178

The aneurysm of the aorta can extend more cranially, consequently the visceral arteries branch

from the aneurysmal sac. In these suprarenal AAAs or TAAAs a branched, or combined

branched and fenestrated EVAR (B/F-EVAR) was developed and has branches attached to

the endograft, to bridge the distance from the endograft to the origin of the visceral artery.

Chapter 9 focuses on the long-term results of these endografts with special attention to

geometrical changes in the bridging stents. The geometrical changes were more outspoken in

branches than in fenestrations, probably related to the longer distance from endograft to the

origin of target vessels. Furthermore, it was postulated that migrating endografts strain the

stents and will be seen more in branched compared to fenestrated endografts. The aortic wall

is adjacent to the endograft in fenestrations allowing less movement. Branches with a bridging

stent to the visceral artery are more complex constructions and are more vulnerable to an

adverse event. Therefore, overall stent patency was lower in branches than in fenestrations.

Meticulous measurement of endograft migration, the angle and length of bridging stents, and

comparing these with previous measurements can predict the risk of disconnection, stenosis

or occlusion and should be part of the routine during follow-up imaging. In this chapter a

substantial number of deaths were seen during long-term follow-up and treatment of TAAAs

remains challenging. The study included fragile patients, but aneurysm related deaths were

not seen until 4-6 years of follow-up and suggest the benefit of endovascular treatment

of patients with a TAAA. Larger multicenter studies, preferably with a prospectively held

database, should be done to show the true benefit of B/F-EVAR.

In conclusion, over the recent decades great steps have been made towards a total endovascu-

lar abdominal aortic aneurysm repair and even extending into the thoracic aorta. Endovascu-

lar treatment modalities have increased in complexity and many more challenges remain to

perfectly customize the treatment to the patients’ anatomy. Although endovascular treatments

are done on a large scale, endograft design keeps on changing making true long-term outcome

of current available devices still unavailable. Only few designs for fenestrated endovascular

aneurysm repair are available with variable applicability and experience. These endografts

take time to customize to the patients anatomy and in urgent cases there is only limited availa-

bility of an off-the-shelf endograft.17 Nevertheless, a shift can be seen towards endovascular

treatment. With more experience, especially in very complex cases, results will most likely

improve for endovascular repair. On the other hand, open surgical repair will lose field. Con-

sequently less experience with open surgical repair is gained by newly trained surgeons, more

Page 10: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

10

Summary and future perspectives 179

specifically the skills necessary in unsuitable cases or in failing endovascular repair.

Future perspectives

Chapters 5 and 9 show us the influence of endografts to native patient anatomy and the

geometrical changes of the endograft over time. These studies are the first steps in gathering

knowledge of endograft influence on native anatomy. Moreover, to improve an even better

understanding of the anatomic changes after FEVAR, iliac artery tortuosity should be taken

into account in further studies.

Larger studies, including multiple endografts and multiple validated anatomic measure-

ments, will allow better understanding of anatomic changes after (infrarenal, fenestrated and

branched) EVAR.

There was no clear explanation why decline in renal function shown in Chapters 4 and 7

occurred after FEVAR. Pre-operative renal function and contrast use might, besides obvious

reasons like renal stent stenosis or occlusion, elicit decline in renal function. Future studies

including multiple possible parameters affecting renal function must be done to be able to

protect renal function in patients treated for an abdominal aneurysm.

The aorta is subject to respiration and the cardiac cycle, therefore an endovascular device is

repeatedly strained.18 To gain greater insight of these dynamic changes over time, ECG-gated

CTA can be done, allowing 4-dimensional imaging of the endograft and patient anatomy.

Before long-term follow-up results are available, they can be obsolete because constant

changes are made to designs of endografts. These endograft alterations can make inter-

pretation of results with (infrarenal, fenestrated and branched) EVAR challenging. As was

discussed in Chapters 2 and 3, in a few currently available endografts there are long-term

results available. To gain knowledge about the durability of the available endografts long-term

follow-up studies need to be conducted, preferably in prospectively held databases.

Reinterventions, because of endograft related complications, remain a problem in (infrarenal,

fenestrated and branched) EVAR. Multicenter studies including clinical outcome and patient

anatomic characteristics, will allow us to enlarge the study cohorts and to correlate certain

anatomic features to long-term complications.

Ultimately, a prediction model allows clinicians to make a tailor-made treatment to patient’s

anatomy and preferences.

Page 11: University of Groningen Endovascular approaches to complex ... · reported in literature.10,11 In all these cases a procedural type Ia endoleak was seen in 6.9%, but also mostly spontaneously

Chap

ter

10

180

1.Tadros RO, Faries PL, Ellozy SH, Lookstein RA, Vouyouka AG, Schrier R, et al. The impact of stent graft evolution on the results of endovascular abdominal aortic aneurysm repair. J Vasc Surg. 2014;59:1518-27.

2.Maudet A, Daoudal A, Cardon A, Clochard E, Lucas A, Verhoye JP, et al. Endovascular Treatment of Infrarenal Aneurysms: Comparison of the Results of Second- and Third-Generation Stent Grafts. Ann Vasc Surg. 2016;34:95-105.

3.Cayne NS, Adelman MA, Veith FJ. Current status of investigational devices for EVAR: similarities and differences. Semin Vasc Surg. 2009;22:127-31.

4.Roy IN, Gharib M, Zerwes S, Jakob R, Torella F, McWilliams RG, et al. Anatomical applicability of endovascular aneurysm sealing techniques in a consecutive cohort of fenestrated endovascular aneurysm repairs. J Endovasc Ther. 2017;24:773-8.

5.Savji N, Rockman CB, Skolnick AH, Guo Y, Adelman MA, Riles T, et al. Association between advanced age and vascular disease in different arterial territories: a population database of over 3.6 million subjects. J Am Coll Cardiol. 2013;61:1736-43.

6.Silveira D, Pitoulias G, Torsello G, Donas KP. Outcomes of total endovascular treatment of juxtarenal aortic aneurysms in octogenarians. J Vasc Surg. 2016;63:909-14.

7.Michel M, Becquemin JP, Marzelle J, Quelen C, Durand-Zaleski I, WINDOW Trial participants. A study of the cost-effectiveness of fenestrated/branched EVAR compared with open surgery for patients with complex aortic aneurysms at 2 years. Eur J Vasc Endovasc Surg. 2018;56:15-21.

8.Ullery BW, Suh GY, Lee JT, Liu B, Stineman R, Dalman RL, et al. Comparative geometric analysis of renal artery anatomy before and after fenestrated or snorkel/chimney endovascular aneurysm repair. J Vasc Surg. 2016;63:922-9.

9.van Marrewijk C, Buth J, Harris PL, Norgren L, Nevelsteen A, Wyatt MG. Significance of endoleaks after endovascular repair of abdominal aortic aneurysms: The EUROSTAR experience. J Vasc Surg. 2002;35:461-73.

10.Verhoeven EL, Katsargyris A, Oikonomou K, Kouvelos

G, Renner H, Ritter W. Fenestrated endovascular aortic aneurysm repair as a first line treatment option to treat Short necked, juxtarenal, and suprarenal aneurysms. Eur J Vasc Endovasc Surg. 2016;51:775-81.

11.Wang SK, Gutwein AR, Gupta AK, Lemmon GW, Sawchuk AP, Motaganahalli RL, et al. Institutional experience with the Zenith Fenestrated aortic stent graft. J Vasc Surg. 2018;68:331-6.

12.Denic A, Glassock RJ, Rule AD. Structural and functional changes with the aging kidney. Adv Chronic Kidney Dis. 2016;23:19-28.

13.Saratzis A, Nduwayo S, Sarafidis P, Sayers RD, Bown MJ. Renal function is the main predictor of acute kidney injury after endovascular abdominal aortic aneurysm repair. Ann Vasc Surg. 2016;31:52-9.

14.Kouvelos G, Koutsoumpelis A, Lazaris A, Matsagkas M. Late open conversion after endovascular abdominal aortic aneurysm repair. J Vasc Surg. 2015;61:1350-6.

15.Locati P, Socrate AM, Costantini E. Paraanastomotic aneurysms of the abdominal aorta: a 15-year experience review. Cardiovasc Surg. 2000;8:274-9.

16.Katsargyris A, Yazar O, Oikonomou K, Bekkema F, Tielliu I, Verhoeven EL. Fenestrated stent-grafts for salvage of prior endovascular abdominal aortic aneurysm repair. Eur J Vasc Endovasc Surg. 2013;46:49-56.

17.Cowan JA,Jr, Dimick JB, Henke PK, Huber TS, Stanley JC, Upchurch GR,Jr. Surgical treatment of intact thoracoabdominal aortic aneurysms in the United States: hospital and surgeon volume-related outcomes. J Vasc Surg. 2003;37:1169-74.

18.Hansen PA, Richards JM, Tambyraja AL, Khan LR, Chalmers RT. Natural history of thoraco-abdominal aneurysm in high-risk patients. Eur J Vasc Endovasc Surg. 2010;39:266-70.

19.Mendes BC, Rathore A, Ribeiro MS, Oderich GS. Off-the-shelf fenestrated and branched stent graft designs for abdominal aortic aneurysm repair. Semin Vasc Surg. 2016;29:74-83.