Optimizing Dose in the Interventional Suite:

Do you really need that image quality?

Robert G, Dixon, MDUniversity of North Carolina

Page 2: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Financial Disclosure

Educational Consultant for Bard Access Systems

Page 3: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Personal Disclosure

Page 4: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

What can I teach a room full of physicists?

http://theinvisibleagent.wordpress.com/tag/physics/

Page 5: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Two Perspectives

PhysicistInterventional

Radiologist

Page 6: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Two Perspectives

PhysicistInterventional

Radiologist

Page 7: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Learning Objectives

• Appreciation of the clinical perspective• Need for operator education

• Simple techniques to optimize dose

• Discuss the question: “Do I really need all that image quality?”

Page 8: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Wide Range of Dose

Mettler et al. Radiology 2008

Page 9: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 10: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 11: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

What should we teach them?

1. Basic, simple techniques

2. Target the young physicians

Page 12: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Things you know

• Increase table height

• Add Barriers

• Exit the room for DSA runs

• Limit magnification

Page 13: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 14: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Increase the Table Height

Why?

Page 15: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Source to Skin Distance

• SSD: determined by table heightthe operator’s height

• Skin dose decreases as SSD increases

• Therefore, maximize SSD within reason700 mm (27.6 inches)600 mm (23.6 inches): increase dose 17-29%Simple 10 cm (4 inch) maneuver

Wagner LK, Archer BR Cohen Am.JVIR 2000;11:25-33

Page 16: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 17: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

20% Savings

Page 18: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 19: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Vary the technique

• Operator is in control• Vary:

Frame RatesGantry PositionProtocol UsedCollimation

Page 20: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Optimize

• Prior to the case

• During the case

Page 21: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Prior to the case

• Review old studiesCTMRIUS

• Review prior Interventional Procedures

• Have a clear plan

Page 22: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

During the case

• Ask: do I need great quality to…• Gain access to the celiac artery• Demonstrate IVC Filter Position• Insert a port• Drain an abscess• Perform cerebral angiography• Embolize a bleeding visceral artery

Page 23: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Use it if you need it

Page 24: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Use it if you need it

Page 25: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Consider alternatives if you don’t

Page 26: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Save Fluoro

• LIH

• Fluoro Video Clips

Page 27: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Last Image Hold

Last Image Hold Spot Radiograph

Page 28: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Slow the Frame Rate

• Fluoroscopic Frame Rate

• DSA acquisition Frame Rate

Page 29: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Pulsed Fluoroscopy15 – 30 pps for critical procedures where precision

required

7.5 pps used for many casesUp to 70% dose savings compared with

continuous fluoro *

3 pulses per secondUsed when dose savings are paramount or for

simple procedures

* Miller DL et al. Radiology 2002;225:329-336

Page 30: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

75 yo male ESRD

Left Arm Fistula

Page 31: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Following Angioplasty

Page 32: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

High Dose Procedures

• Embolization

• TIPS

• Renal/Visceral stent placements

• Cardiac RFA

• Coronary angioplasty/stent

• Biliary procedures

Page 33: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

TIPS

Page 34: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Protocols

Page 35: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 36: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 37: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 38: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 39: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 40: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 41: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Eyes: Protect Them!

Page 42: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Cataracts

Page 43: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Ocular Guidelines

• Previously - Deterministic Effect: • 2 – 5 Gy for protracted exposure

• Early Studies• Short follow up• Lacked sufficient sensitivity• Few subjects with dose < few Gy

NCRP 168

Page 44: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Worse Case Scenario

• Source above table

• No Barrier Protection

• Dose to the eye 450 – 900 mSv/yr

Vano Br J Radiol1998Schueler et al. Radiographics 2006

Page 45: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Eye Protection

Typical workloads: dose to eye may exceed the threshold for cataracts after several years of work if radiation protection tools are not used.

Vano et al. Radiology 2008;248:945-953

Page 46: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Newer Data

• Opacification of the lens at lower doses

• Based on • Patients: CT scans, radiation therapy• Atomic-bomb survivors• Residents of contaminated buildings• Chernobyl accident liquidators• Radiologic technologists

NCRP 168

Page 47: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Newer Data

• Chernobyl workers: < 1 Gy

• Rad Techs: risk increases linearly with no apparent threshold

• These challenge prior recommendations

NCRP 168Worgul et al; Radiat Res 2007

Page 48: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

ICRP

• Lens of the eye: threshold for tissue reaction in absorbed dose is now considered to be 0.5 Gy

• Occupational exposure:• Now recommend an equivalent dose

limit of 20 mSv/yr (avg over 5 yrs)• Previously 150 mSv/yr• No single year > 50 mSv

www.ICRP.org – statement on tissue reaction.2011

Page 49: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Lens Exposure Cardiology

• Retrospective cumulative lens dose

• Avg: 22 years working/ 51 years old

• 25 mSv – 1600 mSv

• New Annual dose limit 20 mSv/yr• Exceeded by 60% of cardiologists

Jacob et al. Radiat Prot Dosimetry 2012

Page 50: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Lens Exposure Cardiology

• Several years + no protection may exceed ICRP lifetime dose threshold of 500 mSv

• Risk for developing radiation-induced cataracts

Jacob et al. Radiat Prot Dosimetry 2012

Page 51: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Eye Protection

Page 52: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

21%

Cousins et al. Radiol 1987;22:688-692

Page 53: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Operator Eye Protection

Thornton, Dauer et al.JVIR 2010

Page 54: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Comparing Strategies

Thornton, Dauer et al.JVIR 2010

• Leaded Glasses: reduced by a factor of 5 - 10

• Scatter Shielding Drapes:reduced by a factor of 5 - 25

• Both Together:reduced by a factor of > 25

• Suspended Shielded:Undetectable

Page 55: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

CT Suite

Page 56: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

15 yo with post-op abscess

Page 57: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

CT Abdomen and Pelvis

Page 58: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

CT Suite

Page 59: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 60: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 61: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 62: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 63: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University
Page 64: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Lung Cancer

Page 65: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Lung Cancer

Page 66: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Lung Cancer

Page 67: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Lung Cancer

Page 68: University of North Carolina - AAPM HQamos3.aapm.org/abstracts/pdf/68-19749-234349-90002.pdf · 2012. 7. 28. · Do you really need that image quality? Robert G, Dixon, MD. University

Summary

• Appreciation of the clinical perspective• Need for operator education• Target the young physicians

• Simple techniques to optimize dose

• Discuss the question: “Do I really need all that image quality?”