Page 1: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

User’sManual Model FC500G

Free Available ChlorineAnalyzer(Non-Reagent Type)[Style: S2]

IM 12F5A2-01E

IM 12F5A2-01E7th Edition

Page 2: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

i

IM 12F5A2-01E

IntroductionThis user’s manual describes the specifications, installation, operation, maintenance, and troubleshooting for Model FC500G Free Available Chlorine Analyzer. Read this manual thoroughly to understand the contents before operating the equipment.

n Notes on Hardware

l Appearance and AccessoriesCheck the following when you receive the product:

• Appearance• Standard accessories

Contact our sales representative or your local distributor if the product’s coating has come off, it has been damaged, or there is shortage of required accessories.

l Model and Suffix CodesThe name plate on the product contains the model and suffix codes. Compare them with those in the general specification to make sure the product is the correct one. If you have any questions, contact our sales representative or your local distributor.

Media No. IM 12F5A2-01E 7th Edition : Feb. 2020 (YK)All Rights Reserved Copyright © 2007, Yokogawa Electric Corporation

Page 3: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

ii

IM 12F5A2-01E

Safety Precautionsn Safety, Protection, and Modification of the Product

• In order to protect the system controlled by the product and the product itself and ensure safe operation, observe the safety precautions described in this user’s manual. We assume no liability for safety if users fail to observe these instructions when operating the product.

• If this instrument is used in a manner not specified in this user’s manual, the protection provided by this instrument may be impaired.

• Be sure to use the spare parts approved by Yokogawa Electric Corporation (hereafter simply referred to as YOKOGAWA) when replacing parts or consumables.

• Modification of the product is strictly prohibited.

n Signal WordsThe following words are used in this manual.

CAUTION This symbol gives information essential for understanding the operations and functions.NOTE This symbol indicates information that complements the present topic.

n Notes on Handling User’s Manuals• Please hand over the user’s manuals to your end users so that they can keep the user’s

manuals on hand for convenient reference.• Please read the information thoroughly before using the product.• The purpose of these user’s manuals is not to warrant that the product is well suited to any

particular purpose but rather to describe the functional details of the product.• No part of the user’s manuals may be transferred or reproduced without prior written

consent from YOKOGAWA.• YOKOGAWA reserves the right to make improvements in the user’s manuals and product at

any time, without notice or obligation.• If you have any questions, or you find mistakes or omissions in the user’s manuals, please

contact our sales representative or your local distributor.

n Warning and DisclaimerThe product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor responsibility to any person or entity with respect to any direct or indirect loss or damage arising from using the product or any defect of the product that YOKOGAWA can not predict in advance.

n Trademark Acknowledgments• EXA FC is a trademark of Yokogawa Electric Corporation.• All other company and product names mentioned in this user’s manual are trademarks or

registered trademarks of their respective companies.• We do not use TM or ® mark to indicate those trademarks or registered trademarks in this

user’s manual.

Page 4: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

iii

IM 12F5A2-01E

After-sales Warrantyn Do not modify the product.

n During the warranty period, for repair under warranty consult the local sales representative or service office. Yokogawa will replace or repair any damaged parts. Before consulting for repair under warranty, provide us with the model name and serial number and a description of the problem. Any diagrams or data explaining the problem would also be appreciated.l If we replace the product with a new one, we won’t provide you with a repair report.l Yokogawa warrants the product for the period stated in the pre-purchase quotation

Yokogawa shall conduct defined warranty service based on its standard. When the customer site is located outside of the service area, a fee for dispatching the maintenance engineer will be charged to the customer.

n In the following cases, customer will be charged repair fee regardless of warranty period.

• Failure of components which are out of scope of warranty stated in instruction manual.• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa

Electric did not supply.• Failure due to improper or insufficient maintenance by user.• Failure due to modification, misuse or outside-of-specifications operation which Yokogawa

does not authorize.• Failure due to power supply (voltage, frequency) being outside specifications or abnormal.• Failure caused by any usage out of scope of recommended usage.• Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots, warfare,

radiation and other natural changes.

n Yokogawa does not warrant conformance with the specific application at the user site. Yokogawa will not bear direct/indirect responsibility for damage due to a specific application.

n Yokogawa Electric will not bear responsibility when the user configures the product into systems or resells the product.

n Maintenance service and supplying repair parts will be covered for five years after the production ends. For repair for this product, please contact the nearest sales office described in this instruction manual.

Page 5: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

Blank Page

Page 6: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

Toc-1

IM 12F5A2-01E

Model FC500GFree Available Chlorine Analyzer(Non-Reagent Type) [Style: S2]

CONTENTS

IM 12F5A2-01E 7th Edition

Introduction ..........................................................................................................i Safety Precautions .......................................................................................ii1. OVERVIEW ................................................................................................ 1-1

1.1 Standard Specifications ...................................................................................1-21.2 Model and Suffix Codes ...................................................................................1-41.3 External Dimensions ........................................................................................1-51.4 Measuring Principle ..........................................................................................1-6

2. COMPONENTS AND FUNCTIONS .......................................................... 2-12.1 Detector ..............................................................................................................2-12.2 Converter ...........................................................................................................2-2

3. INSTALLATION, PIPING, AND WIRING ....................................................3-13.1 Installation .........................................................................................................3-1

3.1.1 Location ..............................................................................................3-13.1.2 Mounting ............................................................................................3-2

3.2 Piping .................................................................................................................3-23.2.1 Sample Water Piping .........................................................................3-23.2.2 Air Purge Piping .................................................................................3-2

3.3 Wiring .................................................................................................................3-33.3.1 Wiring for Power Supply and Grounding ........................................... 3-33.3.2 Analog Output Wiring .........................................................................3-3

4. OPERATION .............................................................................................. 4-14.1 Preparation for Operation ................................................................................4-1

4.1.1 Inspection of Piping and Wiring Status ..............................................4-14.1.2 Fill with Glass Beads for Electrode Polishing .................................... 4-14.1.3 Measuring Range Setting ..................................................................4-24.1.4 Setting of “Applied Voltage Compensation/Applied Voltage Fixation’’ Se-

lection Jumper ....................................................................................4-34.2 Start-Up ..............................................................................................................4-4

4.2.1 Supplying Sample Water ...................................................................4-44.2.2 Confirmation of “Power Supply Voltage” and “Measuring Range” Selec-

tion Jumpers Setting Status ...............................................................4-44.2.3 Supplying Power ................................................................................4-54.2.4 Operation Check ................................................................................4-54.2.5 Calibration ..........................................................................................4-5

4.3 Steady State Operation ....................................................................................4-5

Page 7: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

Toc-2

IM 12F5A2-01E

5. CALIBRATION ........................................................................................... 5-15.1 Zero Calibration .................................................................................................5-1

5.1.1 Zero Calibration Procedure Using the Open Input Circuit Method .... 5-15.1.2 Zero Calibration Procedure Using Chlorine-Free Water Method ...... 5-1

5.2 Span Calibration ...............................................................................................5-25.2.1 Span Calibration Procedure Using the Sampling Method ................. 5-25.2.2 Span Calibration Procedure Using Standard Solution Method ......... 5-2

6. INSPECTION AND MAINTENANCE ........................................................ 6-16.1 Mechanical System ...........................................................................................6-1

6.1.1 Indicator Electrode Polishing .............................................................6-16.1.2 Glass Beads and Measuring Tank Cleaning ..................................... 6-26.1.3 Brush and Slip Ring Checking ...........................................................6-3

6.2 Electrical System ..............................................................................................6-56.2.1 Span Calibration ................................................................................6-56.2.2 Converter Checking ...........................................................................6-56.2.3 Fuse Replacement .............................................................................6-5

7. TROUBLESHOOTING .............................................................................. 7-18. AUXILIARIES AND CONSUMABLES ...................................................... 8-1

8.1 Auxiliaries and Consumables List ..................................................................8-18.2 Auxiliaries and Consumables .........................................................................8-1

8.2.1 Polisher (Part number: K9088PE) ..................................................... 8-18.2.2 Glass Beads (Part number: K9332ZJ)...............................................8-18.2.3 Indicator Electrode (Part number: K9332MB) ................................... 8-18.2.4 Grease (Part number: K9044FX) ....................................................... 8-18.2.5 Counter Electrode (Part number: K9332MK) .................................... 8-18.2.6 Rotating contact (Part number: K9332SR) ........................................ 8-2

Customer Maintenance Parts List ................................... CMPL 12F05A02-02ENRevision Information ...............................................................................................i

Page 8: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<1. OVERVIEW> 1-1

IM 12F5A2-01E

1. OVERVIEWModel FC500G Free Available Chlorine Analyzer (Non-Reagent Type) is used for continuous measurement of free available chlorine in tap water distribution. This analyzer consists of a detector and a converter as shown in Figure 1.1.This chapter describes the FC500G free available chlorine analyzer specifications and its measuring principle.

Detector

Converter

Figure 1.1 Appearance of Model FC500G Free Available Chlorine Analyzer

Page 9: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<1. OVERVIEW>1-2

IM 12F5A2-01E

1.1 Standard SpecificationsMeasured Object: Free available chlorine contained in tap waterMeasuring System: Polarographic method using rotating electrodeMeasuring Ranges: 0 to 1 / 0 to 2 / 0 to 3 mg/L (Model FC500G-1) 0 to 5 / 0 to 10 mg/L (Model FC500G-5) (Air purge is necessary.)Indication: Digital (3 1/2 LCD)Output Signal: 4 to 20 mA DC (Load resistance: Maximum 500 Ω)Operating pH Range of Sample Solution: pH 6.5 to 7.5: For measured liquids whose pH exceeds the above limits,

error increases (refer to the following figure).Sample Solution: Temperature: 0 to 40°C (temperature compensation range 5 to 35°C). Pressure: 1 to 30 kPa (for without VALVE) 1 to 150 kPa (with VALVE) Flow Rate: 0.1 to 2.5 L/min (at tank inlet) Conductivity: 100 to 300 μS/cm SS (Suspended Solid): 10 mg/L or lessAir Purging (for only 0 to 5 / 0 to 10 mg/L Range) Supply Air Pressure: About 50 to 140 kPa Air Consumption: About 5 L/minElectrode: Indicator Electrode: Rotating gold electrode Counter Electrode: Silver electrode (with Pt 1000 installed)Electrode Cleaning: Glass beads used for cleaningWetted Part Materials: Measuring Tank: Acrylic resin (molding) Piping: Flexible PVC and PEColor: Muncell 5Y7/1Finish: Baked polyurethane resin coatingMounting: On the wall or 2 inch pipe (OD 60.5 mm)Ambient Temperature: -5 to 50°CPower Supply: 100 or 110 V AC; or 200 or 220 V AC as per order ±10%, 50 or 60 HzPower Consumption: Approximately 15 VA (100 or 110 V), 20 VA (200 or 220 V)Weight: Approximately 6.5 kgEMC Regulatory Arrangement in Australia and New Zealand (RCM) EN 55011 Class A, Group 1

Characteristics: Reproducibility: 2% of full scale Linearity: ±5% of full scale Stability: Zero Drift: Within ±1% of full scale/month Span Drift: Maximum -10% of full scale/month

Page 10: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<1. OVERVIEW> 1-3

IM 12F5A2-01E

Response Time: Approximately 2 minutes (time required to obtain a value within 90% of final value)

Ambient Temperature Influence: ±0.5% of full scale/10°C Power Voltage Fluctuation Influence: ±0.5% of full scale/10% of rated voltage Sample Solution Temperature Compensation Error: ±3% of full scale/5 to 35°C

3 4 5 6 7 8 9 10pH

pH6.5 pH7.5

Operating range0.5

0

1.0

Indi

cate

d va

lue

(mg/

l)

Figure 1.2 Sensitivity characteristics of diffusion current by pH value

Figure 1.3 Sensitivity characteristics of diffusion current by conductivity

Note: If conductivity exceeds 300 μS/cm, default value of applied voltage should be changed.

n AccessoriesFuse (1 A) .........................4 pcs.Glass Beads .....................1 (2 bags)Allen Wrench ....................2 pcs.Screwdriver (flat blade) ....1 pc.Polishing Powder ..............1 bottleBelt tension adjuster plate...1 pc.

Page 11: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<1. OVERVIEW>1-4

IM 12F5A2-01E

1.2 Model and Suffix Codes [Style: S2]

Model Suffix Code Option Code DescriptionFC500G Free Available Chlorine AnalyzerRange -1

-5

0 to 1 / 0 to 2 / 0 to 3 mg/L Switchable0 to 5 / 0 to 10 mg/L Switchable

Output signal 1 4 to 20 mA DCPower 3

456

100 V ±10%, 50/60 Hz110 V ±10%, 50/60 Hz200 V ±10%, 50/60 Hz220 V ±10%, 50/60 Hz

Inlet valve 01

Without ValveWith Valve

Language -E English

n Spare PartsName Parts No. Note

Indicator Electrode K9332MB Rotating Electrore

Counter Electrode K9332MK Reference Electrode

Glass Beads K9332ZJ Washing Indicator Electrode (2 bags)

Polishing Powder K9088PE Polishing Indicator Electrode

Rotating contact K9332SR Part for electrode mechanism

Grease K9044FX

Page 12: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<1. OVERVIEW> 1-5

IM 12F5A2-01E

1.3 External Dimensions (1) With Inlet Valve (FC500G-¨¨¨1-E)

0.50mg/l

Unit: mm

(48)(154)

Mounting bracket

2-inch pipe (Ø60.5)

Approx.150 265(209)56

Approx.160Maintenance spaceMaintenance space

2.8

49.2

Appr

ox.

5015

4

Drain cockInlet13.7

5.5 76

CapOutlet

22 22(22)22

108

Cable gland30

2226

100

7021

17.5

7052

.512

8.5

48

52.5

105

131

Wiring inlet3-Ø23 holes

App

rox.

250

Mai

nten

ance

spac

e

295

310

App

rox.

150

Mai

nten

ance

spac

e

Mai

nten

ance

spac

e

240

Converter

Cap

InletØ6/Ø4 Union

Purge air inlet1/4-18 NPT(for only FC500G-5-E)

OutletØ12 Nipple

Drain cock

InsertTub

4-Ø8.5 holesfor mounting(U-bolt on bracket)

DrivingBox

3533

.4

17.3

(2) Without Inlet Valve (FC500G-¨¨¨0-E)

0.50mg/l

Unit: mmApprox.150 260

146.5 (5.5)108Approx.160

Maintenance spaceMaintenance space

2.8

49.2

Appr

ox.

5015

4

Drain cockInlet13.7

76

CapOutlet

22 22(22)22

Cable gland

3022

100

128.

5

Wiring inlet3-Ø23 holes

App

rox.

250

Mai

nten

ance

spac

e

295

178

30A

ppro

x.15

0M

aint

enan

cesp

ace

Mai

nten

ance

spac

e

240

Converter

Cap

OutletØ12 Nipple

Drain cock

InsertTub

2-Ø8.5 holesfor mounting

DrivingBox

3533

.4

17.3

1092

75

4

InletØ6/Ø4 Union

Purge air inlet1/4-18 NPT(for only FC500G-5-E)

Figure 1.4 Model FC500G Free Available Chlorine Analyzer External Dimensions

Page 13: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<1. OVERVIEW>1-6

IM 12F5A2-01E

1.4 Measuring PrincipleThe FC500G free available chlorine analyzer (non-reagent type) is used for measuring concentration of free available chlorine present in sample water, such as chlorine (Cl2) hypochlorous acid (HClO), and hypochlorite ion (ClO-).The measuring principle of this analyzer is analysis determination using the rotating electrode polarographic method. Free available chlorine concentration is determined by measuring the current (diffusion current) which flows when free available chlorine is subjected to electrolytic reduction by applying a voltage (applied voltage) between the indicator electrode (rotating electrode) and the counter electrode. The applied voltage is set in a region where concentration polarization occurs (plateau region) in polarography. In the plateau region, the diffusion current does not change even if the applied voltage changes.In measurement practice, since the plateau region changes with free available chlorine concentration, the applied voltage is compensated so that measurement can always be performed in the plateau region. Also, since the diffusion current is affected by the sample water temperature, temperature is compensated using a temperature sensor (Pt1000Ω) incorporated in the counter electrode.Figure 1.5 shows the block diagram for the FC500G free available chlorine analyzer (non-reagent type) operation. The analyzer determines the free available chlorine concentration by applying an applied voltage corresponding to the diffusion current between the indicator and counter electrodes and by applying temperature compensation to the diffusion current.

Applied voltagecompensatingpulse circuiut

Free availablechlorine

detecting circuit

First stageamplifier

Temperaturecompensation

circuit

PWM circuitisolator

Free availablechlorine/

conductivitydisplay

Voltage-currentconversion

circuit

4 to 20 mA DCoutput

Diffusion current

Temperaturesignal

Applied voltage

Indicatorelectrode

Counterelectrode(Pt 1000Ω)

Samplewater

Measuringtank

Drain

F1-5.ai

Figure 1.5 Block Diagram for FC500G Free Available Chlorine Analyzer Operation

Page 14: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<2. COMPONENT NAMES AND FUNCTIONS> 2-1

IM 12F5A2-01E

2. COMPONENTS AND FUNCTIONS

2.1 Detector

Cell motorRotates the indicator electrodeat a constant speed.

Sample inlet

Drain cock When loosened, samplein the measuring tankdrains.

Measuring tank

Counter electrode(Pt 1000 Ω RTD-built in silver chloride electrode)

Indicator electrode (rotating electrode)

Clamp Holds the counter electrode

Gear head

Terminal block for Rotating contact

Drive shaft Driven shaftThe driven shaft is connected tothe indicator electrode by a belt, anddrives it.The signal from this indicator electrodepasses through commutator and Rotating contact.

Rotating contact

Ceramic beads Polish the indicator electrodeto maintain the clean surfaceof the indicator electrode

Measuring tank

Fixing screws

Base

Electrode mechanism block cover prevents the corrosive gas from intruding into analyzer.

Electrode mechanism block

Figure 2.1 Component Names and Functions of the Detector

Page 15: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<2. COMPONENT NAMES AND FUNCTIONS>2-2

IM 12F5A2-01E

2.2 ConverterFigure 2.2 shows the converter component names and their functions.Figure 2.2 is a figure with the cover removed. Measuring range selection should be performed in this state.

1AFUSE

POWER

CHECK

MEAS. ON

SPAN

OFF

SGME RE T1 T2 G C1 C2+ - G G L1 L2

Zero calibratorMEAS / CHECK selection switch

Display (liquid crystal)

Measuring rangesetting switch

Converter cover

Fastener

Span calibrator

Terminal cover

Measuring rangeselecting jumper

Applied voltagecompemsation/fixationselection jumper Power switch

Fuse

External wiringconnection terminals

Power supplyvoltage selectionjumper

Operated whenzero is to becalibrated.(To be operated with the terminal cover removed.)

For converter checking,select CHECK position.

Sets a measuringrange.

The terminal covershould be attachedor detached withthis fastener grippulled out.

Operated when spanis to be calibrated.

mg/l

Figure 2.2 Converter Component Names and Their Functions

NOTEThe measuring range selection jumper and the power supply voltage selection jumper are shipped with these set based on the desired specifications. The specifications cannot be changed by jumper selection.

Page 16: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<3. INSTALLATION, PIPING, AND WIRING> 3-1

IM 12F5A2-01E

3. INSTALLATION, PIPING, AND WIRINGThis chapter describes procedures for the FC500G free available chlorine analyzer installation, piping, and wiring.

3.1 Installation3.1.1 Location

Install the FC500G free available chlorine analyzer (non-reagent type) in a location which meets the following conditions:

Free from rain water such as an indoor location or in a cabinet Low vibration No exposure to direct sunshine Little dust Absence of corrosive gas Easy maintenance

Figure 3.1 shows maintenance spaces.

0.50mg/l

Unit: mmApprox.150 265 Approx.160

Maintenance spaceMaintenance spaceAp

prox

.50

154

App

rox.

250

Mai

nten

ance

spac

e

310

App

rox.

150

Mai

nten

ance

spac

e

Mai

nten

ance

spac

e

240

Figure 3.1 Maintenance Spaces for the FC500G Free Available Chlorine Analyzer

Page 17: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<3. INSTALLATION, PIPING, AND WIRING>3-2

IM 12F5A2-01E

3.1.2 MountingMount the FC500G free available chlorine analyzer (non-reagent type) on a wall surface. For mounting holes, see section 1.3.

3.2 Piping3.2.1 Sample Water Piping

Connect a 6 mm OD X 4 mm ID polyethylene tube to the sample water inlet. Set the sample water pressure at the sample water inlet to 1 to 30 kPa (flow rate: 0.1 to 2.5 l/min).Note: If a sample water inlet valve is used, set the pressure at the valve inlet to 10 to 150 kPa.

Connect a hose of 12 mm lD to the sample outlet. Make piping so that no drain stagnates in the hose.

3.2.2 Air Purge PipingMake the air purge piping only when the FC500G-5 free available chlorine analyzer is used.The air purge piping connecting port (1/8 NPT female) is located at the electrode assembly. Exercise care in piping because the electrode assembly must be removed in implementing inspection or maintenance (see Figure 3.2).

Inspection ormaintenance

Electrodeassembly

Measuringtank

Detector

Pressure regulator

6 mm OD X 4 mm IDpolyyethylene tube

Air supply

Figure 3.2 Air Purge Piping

Page 18: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<3. INSTALLATION, PIPING, AND WIRING> 3-3

IM 12F5A2-01E

3.3 WiringMake wiring for power supply and output signal. Figure 3.3 shows the symbols on the converter terminals and the types of wiring to be connected.

+

TM1

Powersupply

Detector signal(Exclusively used cable: already wired)Ground

to earth4 to 20 mA DCoutput

500Ω or less

Terminal board TM1 (M4 screw)

- G G L1 L2 ME RE T1 T2 SG G C1 C2

F3-3.ai

Note: Make grounding to earth (grounding resistance 100 Ω or less).

Figure 3.3 Symbols on the Converter Terminals and the Types of Wiring to be Connected

3.3.1 Wiring for Power Supply and GroundingUse a three-conductor cable of 9 to 12 mm OD, each conductor having a cross section of 1 mm2 or more.Make grounding to earth (grounding resistance 100 Ω or less).

3.3.2 Analog Output WiringUse a two-conductor shielded cable of 9 to 12 mm OD, each conductor having cross section of 0.5 mm2 or more. The shield should be connected to terminal ‘‘G”.

Page 19: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

Blank Page

Page 20: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<4. OPERATION> 4-1

IM 12F5A2-01E

4. OPERATIONThis chapter chiefly describes the preliminary work to be performed prior to putting into operation.

4.1 Preparation for Operation4.1.1 Inspection of Piping and Wiring Status

Confirm that wiring and piping are made properly.

4.1.2 Fill with Glass Beads for Electrode PolishingPut glass beads (accessories) for cleaning the indicator (rotating) electrode into the beads case in the measuring tank in the following procedure:(1) As shown in Figure 4.1, mount the electrode assembly to the holding stud. The electrode

assembly is fixed to the measuring tank assembly using two screws. In doing this, exercise care so that the indicator (rotating) electrode and the counter

electrode are not struck by the measuring tank. Also, do not touch the surface of the indicator electrode with the fingers.

Fixing screws(2 pcs.)

Electrodeassembly

Measuringtank

Holding stud

F4-1.ai

In maintenance position

Figure 4.1 Removal of Electrode Assembly

(2) Remove the beads case from the measuring tank. The new glass beads are normally filled up to the bottommost notch of the beads case when the indicator electrode is inserted.

Taking care that glass beads do not drop from the beads case. Mount the beads case so that its slits face the counter electrode.

Beads case

Glass beads

Rubber packing

F4-2.ai

Slits

Counter electrode(reference electrode)

Indicator electrode(rotating electrode)

Beads caseGlass beads

The bottommost notch of the beads case

Figure 4.2 Amount of Glass Beads to be Put into the Beads Case

Page 21: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<4. OPERATION>4-2

IM 12F5A2-01E

(3) Return the electrode assembly to the measuring tank and tightly secure it using the fixing screws.

4.1.3 Measuring Range SettingSelect the measuring range using the setting switch. Set it as shown in Table 4.1. Operate the setting switch using a small flat blade screwdriver.

Table 4.1 Switch Setting Corresponding to Measuring Range

Model Measuring range

DIP switch position1 2 3 4

FC500G-10 to 1 mg/l0 to 2 mg/l0 to 3 mg/l

ONOFFOFF

OFFOFFOFF

OFFONOFF

OFFOFFON

FC500G-5 0 to 5 mg/l0 to 10 mg/l

OFFOFF

ONOFF

OFFOFF

OFFON

ONOFF

S31 2 3 4

J5 jumpersee Figure 4.5

Figure 4.4 Measuring Range Setting Switch

Page 22: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<4. OPERATION> 4-3

IM 12F5A2-01E

NOTEFor the FC500G-1, measuring ranges of 0 to 5 and 0 to 10 mg/l cannot be used. For the FC500G-5, measuring ranges of 0 to 1, 0 to 2, and 0 to 3 mg/l cannot be used.

4.1.4 Setting of “Applied Voltage Compensation/Applied Voltage Fixation’’ Selection Jumper

The applied voltage can be fixed to a definite value without applied voltage compensation. When the applied voltage compensation/applied voltage fixation transfer jumper shown in Figure 4.5 is set to the side B (lower), the applied voltage is fixed to the starting voltage value.

Applied voltagecompensation

Side A

J5

Applied voltagefixationSide B

J5

J5

F4-5.ai

Figure 4.5 “Applied Voltage Compensation/Applied Voltage Fixation” Selection Jumper

NOTEIf conductivity exceeds 300 μS/cm, applied voltage should be changed to fixation.

Page 23: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<4. OPERATION>4-4

IM 12F5A2-01E

4.2 Start-Up4.2.1 Supplying Sample Water

<Sample inlet valve is not provided>Confirm that the sample water pressure is within the range of 1 to 30 kPa and supply sample water at the flow rate of 0.1 to 2.5 l/min. After supplying sample water, confirm that the sample water level does not contact the top of the measuring tank.

<Sample inlet valve is provided>First, fully close the sample inlet valve. Then, supply sample water at the pressure of 10 to 150 kPa. After supplying sample water, open the sample inlet valve so that the sample water flow rate is 0.1 to 2.5 l/min.

4.2.2 Confirmation of “Power Supply Voltage” and “Measuring Range” Selection Jumpers Setting Status

If power supply specification for the free available chlorine analyzer (non-reagent type) used is the 100 V system, confirm that the “power supply voltage” selection jumper is set as shown in Figure 4.6 (1).If the power supply specification is the 200 V system, confirm that the jumper is set as shown in Figure 4.6 (2).

100 V System

200 V System

200 V 100 V

200 V 100 V

200 V 100 V

J1

J2

J1

J2

J1

J2

S1

J4

(1)

(2)

Figure 4.6 “Power Supply Voltage” Selection Jumper

Page 24: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<4. OPERATION> 4-5

IM 12F5A2-01E

If the FC500G-1 free available chlorine analyzer is used, confirm that the “measuring range” selection jumper is set as shown in Figure 4.7 (1).For the FC500G-5 free available chlorine analyzer, confirm that the jumper is set as shown in Figure 4.7 (2).

FC500G-1

FC500G-5

Side A

Side B

Measuring range0 to 3 mg/l

Measuring range0 to 10 mg/l

(2)

(1)

Figure 4.7 ‘‘Measuring Range” Selection Jumper

4.2.3 Supplying PowerArrange a power supply whose voltage and frequency confirm to the specifications. Confirm that the fuse holder cap in the converter is tight and then turn ON the power switch.

4.2.4 Operation CheckConfirm that a measured value is displayed in the display.Also, visually confirm that there is no irregularity or significant deflection when the indicator electrode revolves.

4.2.5 CalibrationImplement zero and span calibration referring to chapter 5.

4.3 Steady State OperationThere is no need for special action during analyzer operation except for executing inspection and maintenance described in chapter 6. If a failure occurs, dispose of it referring to chapter 7.

Page 25: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

Blank Page

Page 26: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<5. CALIBRATION> 5-1

IM 12F5A2-01E

5. CALIBRATIONPerform zero and span calibration when the analyzer is to be initially started up and also when re-operated after prolonged shutdown.During operation, perform span calibration periodically (approx. once a month). When span is to be calibrated, first polish the indicator electrode (see subsection 6.1.1) and then perform running-in operation for approx. one hour.Zero calibration is virtually unnecessary during operation if it is performed once at start-up. How-ever, merely confirm periodically that the zero point is correct during operation.

5.1 Zero CalibrationThere are two methods for calibrating zero: the “open input circuit method” using which the calibration is performed by pulling up the rotating and counter electrodes from the sample water and the “chlorine-free water method” using water free from chlorine (e.g., water filtered through activated carbon). Normally calibrate zero using the “open input circuit method”. However, if free available chlorine of low concentration is always to be measured (e.g., approx. 0.1 mg/l), use the “chlorine-free water method” for zero calibration.

5.1.1 Zero Calibration Procedure Using the Open Input Circuit Method

(1) Remove the electrode assembly from the measuring tank with the indicator electrode ro-tated and fix it to the holding studs. (Remove the electrode assembly by loosening the two fixing screws.)

It is not necessary to stop the sample water.(2) After display of the free available chlorine concentration has stabilized, perform zero calibra-

tion. First, remove the terminal cover. Next, slowly turn the zero calibrator (potentiometer) using the attached flat blade screwdriv-

er to adjust display to zero.(3) Mount the terminal cover. Return the electrode assembly to the measuring tank and tightly

fix the assembly using the two fixing screws.

5.1.2 Zero Calibration Procedure Using Chlorine-Free Water Method

(1) Prepare two to three litre of chlorine-free water (prepared by adding 10 ml/l of pH 7 buffer solution and 50 mg/l of NaCl to purified water).

(2) Stop supply of sample water and wait until the sample water level in the measuring tank decreases to below the sample water inlet. Then disengage the sample water inlet piping.

(3) Introduce the chlorine-free water into the measuring tank. Connect a tube to the sample water inlet and pour the chlorine-free water contained in a

suitable vessel into the measuring tank by utilizing a head from the inlet. If a pump is used for supply, use a flow rate of 100 ml/min or more.

(4) After the free available chlorine concentration display has stabilized, perform zero calibra-tion. First, remove the terminal cover.

Next, slowly turn the zero calibrator (potentiometer) using an attached flat blade screwdriver to adjust display to zero.

(5) Mount the terminal cover. Also stop supply of the chlorine-free water. After the solution level in the measuring tank decreases to below the sample water inlet,

remove the tube and restore the sample water piping connection. Supply sample water.

Page 27: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<5. CALIBRATION>5-2

IM 12F5A2-01E

5.2 Span CalibrationThere are two span Calibration methods: “sampling method’’ using which calibration is performed based on the manually analyzed free available chlorine concentration of sample water, and the “standard solution method” which allows calibration using a standard calibration solution having a free available chlorine concentration of approximately 80 % of the measuring range. Calibrate the analyzer span using the “sampling method’’ unless the chlorine concentration of sample water is near zero. (Span calibration can be performed without removing the terminal cover).Note: Before performing span calibration, confirm that the electrode does not remain exposed to

the air as after zero calibration, and the electrode rotates normally in the measuring tank.

5.2.1 Span Calibration Procedure Using the Sampling Method

(1) Confirm that the free available chlorine concentration in the sample water is stable by ob-serving displayed values.

(2) Sample the sample water by loosening the drain cock at the bottom of the measuring tank. When sampling the sample water, exercise care so that the tip of the counter electrode is

not exposed to air above the level of the sample water.(3) Quickly measure the free available chlorine concentration of the sampled sample water, us-

ing a manual analysis such as amperometric titration (AT method).(4) Slowly turn the span calibrator (potentiometer) using the attached flat blade screwdriver to

adjust display to the manually analyzed value.

5.2.2 Span Calibration Procedure Using Standard Solution Method

(1) Prepare diluting solution which is obtained by adding 10 ml/l of pH 7 buffer solution and 50 mg/l of NaCl to purified water (sample water can also be used instead of purified water).

(2) Prepare 2 to 3 liter of the standard solution conditioned to a free available chlorine concen-tration of approximately 80 % of the measuring range. To obtain this solution, dilute commer-cially available sodium hypochlorite (NaClO) solution using the diluting solution prepared in step (1).

(3) Stop supply of sample water and wait until the sample water level in the measuring tank decreases to below the sample water inlet. Then disengage the sample water inlet piping.

(4) Introduce the standard solution into the measuring tank. Connect a tube to the sample water inlet and pour the solution contained in a suitable vessel

into the measuring tank by utilizing a head from the inlet. If a pump is used for supply, use a flow rate of 100 ml/min or more.

(5) After the free available chlorine concentration display has stabilized, loosen the drain cock to sample the standard solution.

In sampling the solution, exercise care so that the tip of the counter electrode is not exposed to air above the level of the solution.

(6) Quickly measure the free available chlorine concentration of the sampled sample water, us-ing a manual analysis such as amperometric titration (AT method).

(7) Slowly turn the span calibrator (potentiometer) using the attached flat blade screwdriver to adjust display to the manually analyzed value.

(8) Stop supply of the standard solution. After the solution level in the measuring tank decreases to below the sample water inlet,

remove the tube and restore the sample water piping connection. Supply sample water.

Page 28: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<6. INSPECTION AND MAINTENANCE> 6-1

IM 12F5A2-01E

6. INSPECTION AND MAINTENANCETo continue good operating status, perform inspections and maintenance as shown in Table 6.1.Determine the implementation cycle in accordance with the individual operation conditions. Table 6.1 is a guide for the implementation cycle.

Table 6.1 Inspection and Maintenance Items and Implementation Cycle

Inspection and maintenance item Implementation cycleIndicator electrode polishing Once/monthGlass beads cleaning Once/three monthsMeasuring tank cleaning Once/three monthsRotating contact Once/yearCalibration Once/monthConverter checking Once/monthFuse replacement Once/year

6.1 Mechanical System6.1.1 Indicator Electrode Polishing

Indicator electrode contamination influences the measured values. Accordingly, the indicator electrode (rotating electrode) is always rotated to polish its surface using glass beads.However, the surface very gradually becomes contaminated with time elapse. Thus, periodically polish the electrode surface using a polisher (alumina).Polish the electrode surface using the following procedure:(1) Remove the converter cover and turn OFF the converter power switch.(2) Remove the electrode assembly from the measuring tank and fix it to the holding stud. It is

not necessary to stop the sample water.(3) Remove the electrode assembly cover.(4) Insert an Allen wrench (accessories) into the hole on the side of the rotating shaft below the

drive belt and manually fix the shaft. Remove the indicator electrode by turning it counter-clockwise. Do not touch the rotating contact or the indicator electrode surface.

(5) Apply polisher (alumina) attached as an accessory to a wet gauze and carefully polish the electrode surface until all contamination is completely removed. After polishing, wash off the polisher using clean water. Once the surface is clean, it will wet uniformly without repelling water.

F6-1.ai

Electrode surface

Figure 6.1 Indicator (Rotating) Electrode

(6) Similar to electrode removal, insert an Allen wrench into the hole and manually fix the rotating shaft so that it does not turn. Fully screw in the indicator electrode (turn clockwise).

Page 29: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<6. INSPECTION AND MAINTENANCE>6-2

IM 12F5A2-01E

CAUTIONDo not touch the rotating contact or the indicator electrode surface.

(7) Mount the electrode assembly cover and restore the electrode assembly in the measuring tank. Tightly secure the assembly using the two fixing screws.

(8) Turn the converter power switch ON and confirm that the indicator electrode revolves normally. Also confirm that the concentration display presents an approximately proper value.

(9) Mount the converter cover. In addition, after the electrode is polished, perform running-in operation for one hour or longer. Then, calibrate span without fail.

Note: Polish the electrode surface (gold electrode) by following the instructions below when it has become corrugated or deformed from the original shape (about 3 mm in diameter).

Polish the electrode surface with sandpaper (about #600) until the surface becomes finally rounded. Then, polish it with sandpaper (about #2000), and lastly with polishing powder (alumina). Make sure the electrode surface is clean and then install the indicator electrode. Reduce the amount of ceramic beads if the electrode has deformed greatly. The ceramic beads should be filled up to 5 mm under the bottommost notch of the beads case when the indicator electrode is inserted.

6.1.2 Glass Beads and Measuring Tank CleaningIf the glass beads are contaminated, cleaning effect deteriorates. Periodically clean the glass beads and the measuring tank.The procedure follows:(1) Turn OFF the converter power switch. Stop sample water supply.(2) Remove the two fixing screws of the electrode assembly and secure the assembly to the

holding studs.(3) Remove the beads case from the measuring tank with the rubber packing attached. In this

case, exercise care so as not to drop glass beads into the measuring tank.(4) Clean the inside of the measuring tank using a brush with detergent on it. Use dilute

hydrochloric acid if manganese or iron adheres. Do not use organic solvent. After cleaning, completely wash out the detergent (or acid) using clean water or sample water.

(5) Move the glass beads to another vessel and clean them using dilute hydrochloric acid. After that, completely wash out the acid using clean water or sample water. For the beads case and rubber packing, remove dirt using detergent or dilute hydrochloric acid. After that, completely wash out the detergent or acid using clean water or sample water.

(6) Set the beads case to the rubber packing and load the cleaned glass beads in the case. If the beads have obviously become smaller in size (much smaller than a diameter of 1.7 to 2.3 mm), replace them with new ones. Glass beads are worn in long-term use, thereby reducing the ability of polishing indicator electrode. For optimum polishing, the new glass beads are normally filled up to the bottommost notch of the beads case when the indicator electrode is inserted (See Figure 4.2). The polishing ability cannot be recovered by replenishing the beads only by volume reduced by wear.

(7) Restore the beads case and rubber packing in the measuring tank. (8) Restore the electrode assembly to the measuring tank and secure it using the two fixing

screws.(9) Supply sample water. Turn ON the converter power switch. Confirm that the indicator

electrode normally revolves and the concentration display presents approx. the proper value. Finally, mount the converter cover. The cleaning effect of the glass beads deteriorates owing to abrasion of the beads themselves over prolonged use. Recommended practice is to replace the beads once or twice a year.

Page 30: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<6. INSPECTION AND MAINTENANCE> 6-3

IM 12F5A2-01E

6.1.3 Checking Rotating contact

CAUTIONNever touch Rotating contact except when replacement is performed.When you hold Rotating contact, pinch it with two fingers and never touch the center contact part. Never drop it or gives a shock.

Contact part

CAUTIONThe storage limit for Rotating contact is 1 year after the purchase, considering degradation of the lubricant used inside. Store it in room temperature and keep it away from direct sunlight.

[Checks and Maintenance]

Rotating contact does not need inspection. Never contact the rotating contact except when the replacement is performed.[Replacing the Rotating contact]

If they are replaced by the customer, follow the steps below.(1) Shut off the power fed to the free available chlorine analyzer to stop the operation.

Gear head

Terminal block for Rotating contact

Motor assembly

Drive shaftDriven shaftDrive belt Pulleys

Clamp

Rotating contact

Plate

Motor adapter plate

Indicator electrode

Counter electrode

Electric holder

Base

Figure 6.2 Electrode Mechanism Block

Page 31: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<6. INSPECTION AND MAINTENANCE>6-4

IM 12F5A2-01E

(2) Remove the two pulleys of the drive shaft (cell motor shaft) and driven shaft (indicator electrode-mounting shaft). The driven shaft pulley can be removed by inserting a 2 mm nominal size Allen wrench (accessory) into the hole in the shaft, which is used as a lock, and unscrewing the end of the driven shaft. The drive shaft pulley can be removed by unscrewing two set screws using a 1.5 mm nominal size Allen wrench (accessory). Remove the pulleys together with the drive belt.

(3) Unscrew the screws fixed on the terminal block of Rotating contact. Unplug the cable terminals of Rotating contact. Remove Rotating contact from Driven shaft. Hold a new rotating contact, with its metal top facing toward you, while the cable being on your right side. Then put the driven shaft throught the center hole.

(4) Install the pulleys and drive belt in place.(Note) Mount drive shaft pulley so that the tip of the shaft can project out 0.3 to 0.7 mm from the end of the pulley.

(5) Install wiring terminals of Rotating contact on the block terminal of Rotating contact. Be sure that the wiring never contacts the drive belt.

(6) Feed power to FC500G. Confirm that the driven shaft/drive shaft/drive belt/rotates smoothly, and that there are no abnormal sounds, like vibration or discontinuous noises.

This completes the replacement procedure.

CAUTIONWhen you shut the cover of electrode mechanism, be careful not to get the cables caught that run inside the analyzer.

Page 32: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<6. INSPECTION AND MAINTENANCE> 6-5

IM 12F5A2-01E

6.2 Electrical System6.2.1 Span Calibration

Perform span calibration at the rate of once every one or two weeks during the initial operation start period and check the sensitivity changes owing to indicator electrode contamination or the like to establish a reference for determining the calibration period. After the calibration period is determined, perform span calibration in an appropriate interval using the above period as a guide. When the indicator electrode is polished, perform span calibration without fail.

6.2.2 Converter CheckingCheck the converter periodically.The converter check function means to check the converter circuit operation.Converter checks can be performed using switch operations. Placing the MEAS./CHECK selector switch in the “CHECK” position transfers the converter input (current input and temperature input) from connection to the detector to connection to the simulated input. When the converter is normal, the free available chlorine concentration display indicates 0.90 to 1.10 mg/l regardless of the measuring range. In this case, the analog output is sent out in the range shown in Table 6.2 corresponding to the measuring range setting.

Table 6.2 Analog Output in Checking

Measuring range Output current Approximate position0 to 1 mg/l 18.4 to 21.6 mA Approx. 100 % point of full scale0 to 2 mg/l 11.2 to 12.8 mA Approx. 50 % point of full scale0 to 3 mg/l 8.8 to 9.9 mA Approx. 33 % point of full scale

6.2.3 Fuse ReplacementFor preventive maintenance, it is recommended to replace the power supply fuse once a year. The fuse rating is 250 V, 1 A.

FUSE

Cap

Power Supply Fuse

Remove the cap byturning it in thedirection of the arrow.

Figure 6.3 Power Supply Fuse

Page 33: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

Blank Page

Page 34: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<7. TROUBLESHOOTING> 7-1

IM 12F5A2-01E

7. TROUBLESHOOTINGIf any operation abnormality appears, take measures using the following procedures.

Abnormalities Cause CountermeasuresNo indication 1. The fuse is blown.

2. Converter connector poor contact1. Replace the fuse.2. Once turn OFF power and re-engage the

relevant connector.Indication shows hunting.

1. Poor contact in the indicator (rotating) electrode mounting position

1. Coat conductive grease (auxiliary).

Abnormal chlorine concentration displayed.

1. Poor contact of the electrode terminals with connecting leadwires

2. Miss-wiring of leadwires to be connected to electrode terminals

3. Poor continuity between the indicator (rotating) electrode and the contactor

4. Deterioration of RTD insulation

5. Sample water flow rate is out of range (0.1 to 2.5 l/min).

6. Calibrated value deviation7. Electrical circuit failure

1. Correctly connect leadwires to the counter electrode terminals (RE, T1, T2) and the indicator electrode terminals (ME).

2. Check the leadwire symbols and correctly connect the leadwires.

3. Clean the rotating contact. Replace the rotating contact. (see subsec-

tion 6.1.3).4. Disconnect leadwires to the counter elec-

trode terminals T1 and T2, and check that the resistance across terminals T1 and T2 is approx. 1 kΩ at room temperature and the resistance across the terminals RE and T1 (or T2) is 1 MΩ or more. If these are not satisfied, replace the counter electrode.

5. Adjust the flow rate. (Adjust so that sample water just overflows the weir in the measuring tank.)

6. Perform zero and span calibration.7. Place the MEAS./CHECK selector switch

in CHECK position and check if the display indicates 0.9 to1.1 mg/l. lf it is not satisfied,contact Yokogawa.

Abnormal analog output

1. The load resistance is larger than 500 Ω.

2. Electrical circuit failure

1. Adjust the load resistance to 500 Ω or less.

2. Contact Yokogawa.The indicator electrode does not revolve or irregularly rotates.

1. The belt is broken.2. The motor fails.3. The rotating shaft bearing has

corroded.

1. Replace the belt.2. Replace the motor.3. Replace the bearing.

Sample water leakage

1. Drain cock is inadvertently not closed.

2. Piping joint loosened.

1. Re-tighten the drain cock.

2. Re-tighten the piping joint fittings.

Page 35: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

Blank Page

Page 36: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<8. AUXILIARIES AND CONSUMABLES> 8-1

IM 12F5A2-01E

8. AUXILIARIES AND CONSUMABLES

8.1 Auxiliaries and Consumables ListTable 8.1 shows auxiliaries and consumables.

Table 8.1 Auxiliaries and Consumables

Item Part number RemarksIndicator electrode K9332MB Rotating gold electrodeCounter electrode K9332MK Reference elctrode

Glass beads K9332ZJ Washing indicator electrode(2 bags)

Polishing powder K9088PE Polishing Indicator electrodealumina, 30 g

Rotating contact K9332SR Part for electrode mechanism block

Grease K9044FX One bottle 3 ml, for continuity stabilization at the base of indicator electrode

Fuse (1 A) A1109EF 250 V, 1 A

8.2 Auxiliaries and Consumables8.2.1 Polisher (Part number: K9088PE)

This is used for polishing the indicator electrode surface.

8.2.2 Glass Beads (Part number: K9332ZJ)The cleaning capability of glass beads in prolonged use deteriorates owing to abrasion. Normally, replace the beads once or twice a year.

8.2.3 Indicator Electrode (Part number: K9332MB)The electrode is gradually abraded by cleaning using glass beads. Generally, the electrode can be used for around three or four years. However, it is recommended that one spare electrode should be stored for unforeseen accident.

8.2.4 Grease (Part number: K9044FX)The grease is applied to stabilize continuity at the mounting base part (screw connection) when the indicator electrode is replaced. Apply the grease to the indicator electrode screw thread referring to the procedure manual attached to the auxiliaries.

8.2.5 Counter Electrode (Part number: K9332MK)Not subject to any abrasion, this electrode can be used for a prolonged period. However, it is recommended to store one electrode as a spare.

Page 37: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

<8. AUXILIARIES AND CONSUMABLES>8-2

IM 12F5A2-01E

8.2.6 Rotating contact (Part number: K9332SR)Normally, one year is considered the replacement interval.

CAUTIONDo not touch the rotating contact when replacement is performed.

The rotating contact is in mechanism consisting of ring and brush together that obtain signals. It becomes worn out over long period of continuous use. Lubricant used for the part also needs to be concerned about its quality degradation after the long period of use. Therefore, standard replacement interval is one year.

CAUTIONThe storage limit for Rotating contact is one year after the purchase, considering degradation of the lubricant used inside. Store it in normal temperature and keep it away from direct sunlight.

Page 38: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

CustomerMaintenanceParts List

All Rights Reserved, Copyright © 2019, Yokogawa Electric Corporation.Subject to change without notice.

Model FC500GFree Available Chlorine Analyzer(Non-Reagent Type)

CMPL 12F05A02-02EN1st Edition : Apr. 2019 (YK)

2nd Edition : Feb. 2020 (YK)

Converter Detector

1

CONVERTER

Item Part No. Qty Description1 A1109EA 1 Fuse (1A)

[Style: S2]

Page 39: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

2

CMPL 12F05A02-02EN 2nd Edition : Feb. 2020 (YK)

8Capacitor

DETECTOR9

10, 11

12

24

23

22

21

2018, 19

13, 1415, 16

Motor Assembly

Motor

Capacitor

1

2

36

7

5

4

17

Item Part No. Qty Description Item Part No. Qty Description1 L9804UK 1 Drive Belt 10 L9831HC 1 Nut2 —— 1 Gear Head 11 L9831HE 1 Sleeve

K9332JP (For 100/110 V AC power supply) 12 L9852CB 1 ValveK9334VA (For 200/220 V AC power supply) 13 K9332KQ 1 Nipple

3 —— 1 Motor Assembly 14 Y9114XB 1 O-Ring

K9334JY (For 100 V AC power supply) 15 K9332KR 1 Cock AssemblyK9334VQ (For 110 V AC power supply) 16 Y9110XB 1 O-RingK9334VR (For 200 V AC power supply) 17 K9332KP 1 InsertK9334VS (For 220 V AC power supply) 18 L9831KL 1 Nut

4 K9334KD 1 Cable 19 L9831KN 1 Sleeve

5 K9332SR 1 Rotating contact 20 K9332KZ 1 Beads Case Cover6 K9332MK 1 Counter Electrode 21 K9332ZJ 1 Beads7 K9332MB 1 Indicator Electrode 22 K9332KX 1 Beads Case8 K9334JV 1 Shaft Assembly 23 K9332KL 1 Measuring Cell9 Y9115XB 1 O-Ring 24 K9332MS 1 Bracket

Page 40: User’s Manual Model FC500G Free Available Chlorine Analyzer · • If this instrument is used in a manner not specified in this user’s manual, the protection provided by this

i

IM 12F5A2-01E

Revision Informationl Title : Model FC500G Free Available Chlorine Analyzer (Non-Reagent Type)l Manual No. : IM 12F5A2-01E

Feb. 2020/7th Edition CMPL 12F05A02-02EN was revised. (2nd edition)

Apr. 2019/6th Edition Style change to S2 (pages 1-4, 2-1, 6-1, 6-2, 6-3, 7-1, 8-2) CMPL 12F05A02-02EN newly released

July 2018/5th Edition Revision according to the suppliment Added EMC(Australia and New Zealand) (p.1-2) Note (p.6-2) Revised description (p.3-3)

Aug. 2012/4th Edition Page layout changed All page layout changed by the standard template; p.1-5, Deletion of conductivity measurement in

subsection 1.4 Measuring Principle; p.2-2, Deletion of some conductivity measurement description in Figure 2.2; p.3-3, Deletion of subsection 3.3.3 Contact Output Wiring (this function doesn’t exist); p.4-2, Deletion of subsection 4.1.3 Setting of Display (Free Available Chlorine or Conductivity) Selection Jumper; p.4-6, Deletion of subsection 4.3.1 “Out-of-Sample Detection” Contact Signal Output; p.7-1, Deletion of “conductivity display shows abnormality”, and “Abnomal out-of sample detection” in chapter 7. TROUBLESHOOTING table; p.8-1, Some revision of subsection 8.1 Auxiliaries and Consumables List (P/N of fuse changed).

Oct. 2009/3rd Edition Editted by InDesign Change the package of glass beads (p.1-3)

Apr. 2008/2nd Edition Everything is reviewed.

Mar. 1993/1st Edition Newly published.

Yokogawa Electric Corporation 2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPAN

http://www.yokogawa.com/