validating fe predictions of distal radius failure load ... · tested intact wrist joint utilized...
TRANSCRIPT
M Hosseinitabatabaei, N Soltan, M McDonald, C Kawalilak,
G Johnston, S Kontulainen, JD Johnston
Validating FE predictions of distal radius failure load: standard region and 4% region
Fixed Scan Position vs Relative Scan Position
2
Bonaretti et al1 recently recommended imaging at the 4% site of the distal radius instead of the standard fixed scan site
Benefits:
Avoids measurement bias related to bone length variability
Compare to existing reference data (e.g., pQCT data, other HR-pQCT data from 4% site)
1. Bonaretti et al (2017), Osteoporosis Int
Finite Element (FE) Modeling
3
Experimental studies validating finite element (FE) predictions of distal radius bone failure load2 and other strength metrics3 have evaluated the standard site
Unclear if FE predictions acquired from the 4% site accurately predict distal radius failure load
2. Mueller et al (2011), Osteoporosis Int 3. MacNeil & Boyd (2008), Bone
Objective
Primary objective:
Evaluate FE-predictions of distal radius failure load acquired from the standard region and the 4% region in relation to experimental failure load
Secondary objective:
Compare FE-predictions of distal radius failure load acquired from the standard region and the 4% region
4
Objective
Primary objective:
Evaluate FE-predictions of distal radius failure load acquired from the standard region and the 4% region in relation to experimental failure load
Secondary objective:
Compare FE-predictions of distal radius failure load acquired from the standard region and the 4% region
5
Methods – Samples
Acquired 19 female, fresh-frozen cadaveric arms (age: 84 ± 8.3 years)
Arms were intact and extended from finger-tip to mid-humerus
Radius measured from radial head to tip of radial styloid process
Soft tissue was removed from radius and ulna
Forearms were potted (fixated) midshaft in PMMA4
6
4. Edwards & Troy (2012), Medical Engineering & Physics
Methods – Samples
Acquired 19 female, fresh-frozen cadaveric arms (age: 84 ± 8.3 years)
Arms were intact and extended from finger-tip to mid-humerus
Radius measured from radial head to tip of radial styloid process
Soft tissue was removed from radius and ulna
Forearms were potted (fixated) midshaft in PMMA4
7
4. Edwards & Troy (2012), Medical Engineering & Physics
Methods – HR-pQCT Imaging
8
First generation XtremeCT with an isotropic voxel size of 82 µm
First scan: Standard region of interest (ROI) located 9.5mm proximal to the reference line
Second scan: 9 blocks starting from the standard reference line
Employed Bonaretti’s “skier” approach1 to identify the 4% site
Reference line
Standard clinical ROI
9-b
lock
RO
I
1. Bonaretti et al (2017), Osteoporosis Int
Methods – HR-pQCT Imaging
First generation XtremeCT with an isotropic voxel size of 82 µm
First scan: Standard region of interest (ROI) located 9.5mm proximal to the reference line
Second scan: 9 blocks starting from the standard reference line
Employed Bonaretti’s “skier” approach1 to identify the 4% site
1. Bonaretti et al (2017), Osteoporosis Int
Methods – HR-pQCT Imaging
Standard reference line
Proximal tip reference line
4% Region
4%
First generation XtremeCT with an isotropic voxel size of 82 µm
First scan: Standard region of interest (ROI) located 9.5mm proximal to the reference line
Second scan: 9 blocks starting from the standard reference line
Employed Bonaretti’s “skier” approach1 to identify the 4% site
1. Bonaretti et al (2017), Osteoporosis Int
Methods – HR-pQCT Imaging
First generation XtremeCT with an isotropic voxel size of 82 µm
First scan: Standard region of interest (ROI) located 9.5mm proximal to the reference line
Second scan: 9 blocks starting from the standard reference line
Employed Bonaretti’s “skier” approach1 to identify the 4% site
Standard reference line
4% 9.5mm
Standard Region
1. Bonaretti et al (2017), Osteoporosis Int
Methods – FE Analysis x y
z Uz
Employed the single tissue FE model
Young’s Modulus: 6.829 GPa 3
Poisson’s ratio: 0.3
High-friction axial compression
Pistoia’s failure criteria5
Failure load defined as the load leading to 2% of bone tissue exceeding an energy equivalent strain of 7000 µstrain
3. MacNeil & Boyd (2008), Bone 5. Pistoia et al (2002), Bone
Methods – FE Analysis
Employed the single tissue FE model
Young’s Modulus: 6.829 GPa 3
Poisson’s ratio: 0.3
High-friction axial compression
Pistoia’s failure criteria4
Failure load defined as the load leading to 2% of bone tissue exceeding an energy equivalent strain of 7000 µstrain
0 mstrain 7000 mstrain
3. MacNeil & Boyd (2008), Bone 5. Pistoia et al (2002), Bone
Methods - Experimental Testing
14
Forearms mounted vertically (axial compression) with 0° dorsal inclination and 3-6° radial inclination6
Compression testing performed at 3mm/s until failure 4,5
Determined ultimate failure load from experimental data
4. Edwards & Troy (2012), Medical Engineering & Physics 5. Pistoia et al (2002), Bone
6. Lochmuller et al (2008), JBMR
Methods - Experimental Testing
15
Forearms mounted vertically (axial compression) with 0° dorsal inclination and 3-6° radial inclination6
Compression testing performed at 3mm/s until failure 4,5
Determined ultimate failure load from experimental data
4. Edwards & Troy (2012), Medical Engineering & Physics 5. Pistoia et al (2002), Bone
6. Lochmuller et al (2008), JBMR
Methods - Experimental Testing
16
Forearms mounted vertically (axial compression) with 0° dorsal inclination and 3-6° radial inclination
Compression testing performed at 3mm/s until failure
Determined ultimate failure load from experimental data
Ultimate Failure Load
Methods - Statistics
17
Linear regression used to assess relationships between FE-derived and experimentally derived failure loads
Paired t-test used to compare FE-derived failure load of the standard region and the 4% region
Results 14 specimens experienced a Colles-type distal radius fracture
5 specimens were excluded
Wrist dislocation (n=2)
Scaphoid fracture (n=1)
Ulna fracture (n=1)
Hand fracture (n=1)
19
Results
R² = 0.69
0
1
2
3
4
5
0 1 2 3 4 5E
xper
imen
tal
Fai
lure
Load
[kN
]
FE Failure Load [kN]
Bone failure load (kN)
R² = 0.72
0
1
2
3
4
5
0 1 2 3 4 5
Exper
imen
tal
Fai
lure
Load
[kN
]
FE Failure Load [kN]
Bone failure load (kN)
4% VOI Standard VOI
Standard region and 4% region offered similar predictions of distal radius failure load
FE-derived failure load of the standard region and the 4% region did not differ (p = 0.58)
20
Standard region and 4% region explained similar variance in experimentally-derived distal radius failure load
Findings indicate that either region can be used to estimate distal radius failure load using HR-pQCT and FE
Discussion
21
Study Strengths
Representative experimental loading configuration
Tested intact wrist joint
Utilized fresh-frozen specimens
Study Limitations
Small sample size (n=14)
Study only included postmenopausal women
Slow testing speed (3 mm/s)
Measured radius length vs ulna length used by Bonaretti et al1
Discussion
1. Bonaretti et al (2017), Osteoporosis Int
22
Study Strengths
Representative experimental loading configuration
Tested intact wrist joint
Utilized fresh-frozen specimens
Study Limitations
Small sample size (n=14)
Study only included postmenopausal women
Slow testing speed (3 mm/s)
Measured radius length vs ulna length used by Bonaretti et al1
Discussion
1. Bonaretti et al (2017), Osteoporosis Int
23
Future Research
Additional mechanical testing
Investigate different modeling approaches and failure criteria
Repeat study with off-axis experimental data
Discussion
Conclusions
24
Standard region and 4% region explained similar variance in experimentally-derived distal radius failure load
Results indicate that either region can be used to estimate distal radius failure load using HR-pQCT and FE
Acknowledgements
Natural Sciences and Engineering Research Council (NSERC)
Saskatchewan Health Research Foundation (SHRF)
Student assistants from the Musculoskeletal & Orthopaedic Biomechanical Imaging Laboratory (MOBIL), including: Dustin Eichhorn, Nima Ashjaee, Amy Bunyamin, Mehrdad Hosseini
25
Questions?