van thompson dds petra guess ips e.max cad lithium disilicate study

11
Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate IPS e.max CAD Lithium Disilicate Study Study

Upload: morgan-walsh

Post on 26-Mar-2015

220 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study

Van Thompson DDS

Petra Guess

IPS e.max CAD Lithium Disilicate StudyIPS e.max CAD Lithium Disilicate StudyIPS e.max CAD Lithium Disilicate StudyIPS e.max CAD Lithium Disilicate Study

Page 2: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study

Single crowns made of IPS e.max lithium disilicate can be…

3X3X STRONGER STRONGER

than a Zirconia single unit crowns!

Page 3: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study

Buccal Lingual cusp loaded increasing cyclical load every 10k cycles until failure

Bottom Radial Crack

Use Level Probability Lognormal

Std=0.08; Alpha(0)=10.10; Alpha(1)=0.24

Cycles

Unreliabilit

y

10000 10000001000001.00

5.00

10.00

50.00

99.00Use LevelCB@90% 2-Sided

Emax Ceram HandL Y-TZPCumulative DamageLognormal200 N LoadF=17 | S=1

Data PointsUse Level LineTop CB-I

Page 4: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study

Fatigue Zirconia supported / Press-on

300 N 50 K

200 N 40 K

LoadLoad

Page 5: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study

Fatigue IPS e.max CAD Lithium Disilicate Survivor at 900 N and 180 K cycles (profile I)

Page 6: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study

Fatigue IPS e.max CAD Lithium Disilicate Survivor at 900 N and 180 K cycles (profile I)

Only superficial damage at load surface

Page 7: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study

• No Cracks - only superficial surface damage

• Specimens were tested up to 900 N of force with 180,000 cycles and no cusp shearing !

• No Cracks - only superficial surface damage

• Specimens were tested up to 900 N of force with 180,000 cycles and no cusp shearing !

IPS e.max CAD Lithium Disilicate

Biaxial mouth motion step stress fatigue

IPS e.max CAD Lithium Disilicate

Biaxial mouth motion step stress fatigue

Page 8: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study

Survival of test materials based on Load and Number of Cycles

0 200 400 600 800 1000

Lithium Disilicate

Zirconia pressed

Zirconia veneered

Cro

wn

s

Load (N) and cycles per thousand

Load cycles (x 1000)

Page 9: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study

“This is by far the most robust system we have tested to date”

Van Thompson DDS

IPS e.max CAD Lithium DisilicateIPS e.max CAD Lithium Disilicate

Page 10: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study
Page 11: Van Thompson DDS Petra Guess IPS e.max CAD Lithium Disilicate Study